Professional Documents
Culture Documents
JUNO I
image acquisition and processing system with
digital detector for fluoroscopy and radiography
series 02
JUNO I
series 02
CAUTION:
Read all the enclosed documents before using the EM equipment
Part 0 : PREFACE
CONTENTS
The JUNO I (hereafter called “the system”) is a digital image acquisition system in fluoroscopy and radiography mode
that uses a large-size PIXIUM RF4343 solid state detector (43 x 43 cm), specifically designed for use in general
diagnostic X-ray rooms and emergency rooms.
It offers real-time diagnostic X-ray scans of the skeleton, gastrointestinal organs and urogenital apparatus and, with the
DSA option, also the peripheral vascular apparatus.
The images acquired are saved in the system memory (HD), shown on the high-resolution monitor and can also be sent
to a diagnostic workstation (or general archive) via DICOM network.
JUNO I lets you replace image systems using image intensifying (I.I.) tubes and TV cameras, as well as
radiography systems with conventional or phosphorus (CR) film cassette systems.
External interface:
- USB 2.0 for saving images to CD-ROM / DVD,
- Ethernet 10/100/1000 BASE T for connection to networks using the DICOM protocol.
JUNO I also lets you acquire images from the a second Wireless F.P. detector (optional). This lets you acquire single-
shot radiography images either in wireless mode and via a power/transmission cable.
The system is identified by a rating/serial number plate at the back of the cabinet. A copy of this plate is also
found inside the cabinet, on the power supply box, next to the power switch.
JUNO I 4695451300
XX XXX 87 XX
ATS
Via A. Volta,10 - 24060 Torre de' Roveri (BG)-ITALY
For/Per : Villa Sistemi Medicali
Via delle Azalee,3 - 20090 Buccinasco (MI)-ITALY
This product complies with FDA radiation
C US performance standards 21 cfr subchapter
0051 j, in effect at date of manufacture
255691
The system components are identified with their own serial numbers as follows:
The figures below show the position of the component serial number plates.
Fig. 1
PIXIUM RF4343
Detector
Fig. 2
PU RF4343 Detector Processing Unit
Fig. 3
PIDRF43 Image Processor
Fig. 4
Working monitor
Fig. 5
Dose Area Meter
Fig. 6
Ionising chamber
Fig. 7
DRX 1 wireless F.P. detector
Fig. 8
3543EZ wireless F.P. detector
3 LIABILITY STATEMENT
* The manufacturer can only be held liable for the safety of its products if the product maintenance,
repair or modification is carried out by the manufacturer or by qualified and suitably trained
personnel.
* The manufacturer cannot be held liable for any malfunction, damage or danger resulting from
improper use of the system or from non-observance of the maintenance regulations.
* The user of the equipment fitted with the system has sole liability for checking that the system is
used solely by duly trained and qualified personnel.
The equipment conforms to European Directive 93/42/EEC and subsequent amendments, in particular
2007/47/EEC. The norms applicable are:
Technical manual
EN code 87 80 014C
LIST OF CONTENTS
TECHNICAL MANUAL
issued on 08/03/10
revised on 03/12/14
series 02
Each part is preceded by its own list of contents, with an indication of the most recent
revision for each paragraph.
GENERAL INDEX
code 87 80 014 C - Mti 00.doc
TECHNICAL MANUAL
LIST OF CONTENTS
Note: See paragraph 1.1 in Part 6 of this manual for details of the composition of the system when
fitted with 2 detectors (Pixium RF4343 detector plus DRX1 wireless detector).
See paragraph 1.1 in Part 7 of this manual for details of the composition of the system when
fitted with 2 detectors (Pixium RF4343 detector plus PIXIUM 3543EZ wireless detector).
12 12 12
X-ray tube 8
9 17 11
X-ray 7
10
3 13 14
1a
4 12 12 12
15 16
1b
The system lets you acquire large digital images (43 x 43 cm) in fluoroscopy and radiography mode in
all real-time X-ray scans of the skeleton, gastrointestine tract and urogenital system and, with the DSA
option, the peripheral vascular system.
The system has specialised image acquisition, process and display parameters to suit the body part in
question and the size of the patient. This lets you get immediate correct acquisition and display of the
image, thus avoiding the need for later improvements.
There are also post-processing functions available so you can change the look of the image and
handle them as required.
See chapter 2 below for full details of the data and technical characteristics.
1.3 SAFETY
The system must be used solely in accordance with the safety instructions contained in this manual
and must never be used for purposes other than those for which it is intended.
* Since the system is to be used in combination with other equipment, full checks must be
made to assure the patients or attendant staff are not exposed to risks if there are any doubts
as to the compatibility between the system and such equipment. Contact the equipment
manufacturer or an expert.
* This system must never be used if it is faulty or any part of the overall system where the
system is installed malfunctions. Likewise, it must not be used in the event of a faulty alarm
or signal.
* Prior authorisation must be obtained from manufacturer before making any modifications to
this system or its safety devices.
* Do not remove the covers on the equipment unless for the specifically envisaged
maintenance operations described in this manual: removal of these covers or panels may
compromise the electromagnetic compatibility of the system.
* As with any technical apparatus, the system must be used in a proper manner and receive
regular checks and maintenance as specified in the "Maintenance" section of this manual.
* The operator must read all the information in this manual before using the equipment.
System providers must add the following information in their manuals for the complete X-ray system:
- Recommendations for exclusive use by highly skilled and trained staff, having been informed
of the risks involved when using X-rays.
- Warnings concerning the daily X-ray dose checks.
System providers must also check against the risk of liquids penetrating the detector unit and the ion
chamber.
Note: See paragraph 1.2 in Part 6 of this manual for details about the specific safety rules when using
the DRX1 wireless detector.
See paragraph 1.2 in Part 7 of this manual for details about the specific safety rules when using
the PIXIUM 3543EZ wireless detector.
* Important: The X-ray unit hosting the system must have emergency buttons fitted (in
suitable positions in the room) that cut all power to the system.
* All equipment connected to the system must comply with the following standards:
- EN 60 601-1 (3rd ed.)
- EN 60 601-1-2 (2nd ed.)
* The X-ray equipment fitted with the system must not be used in explosive conditions.
* Unplug the equipment before cleaning, disinfecting and sterilising it.
* Cleaning products and disinfectants, including those used for patients, can form explosive
gas mixtures. Therefore, only use products that comply with the relevant safety regulations.
* Take care not to spill conducting liquids on the equipment as these could infiltrate and so
damage the equipment and render it unsafe to use.
Before carrying out any exposure, make sure that all the necessary radiation protections have been
activated.
During X-ray use, the personnel present in the X-ray room must observe the radiation protection
regulations in force. The following rules apply here:
The system has been designed and built in full respect of the safety regulations.
Nevertheless, there are still some risks involved in the use of this system if it is used incorrectly or the
prescribed safety devices are faulty.
Risks owing to the improper use of the equipment: see the instructions and recommendations in the
above paragraphs of this chapter.
Risks owing to defects in the prescribed safety devices: bear in mind that:
- Protection against electric shocks is provided by means of an efficient earth system for all metal
parts covering the equipment.
* The earth circuit (for both internal and external parts) should therefore be checked for
efficiency on a regular basis (see the “routine maintenance” schedule in chapter 1 of part 5
of this manual).
- If the LCD monitor's TV screen is hit hard enough, it could break and scatter liquid crystal, that is
toxic.
Never remove any parts or covers as this may compromise the electromagnetic compatibility
of the system.
Cordless and cell phones may affect the efficiency of the device.
In accordance with standard EN 60601-1-2 (2nd ed.), the system is intended for use in the
electromagnetic environment specified below.
The customer or the user of system should assure that it is used in such an environment.
Tab. 201
Emissions test Compliance Electromagnetic environment - guidance
RF emissions Group 1 The system uses RF energy only for its internal functions.
CISPR 11 Therefore, its RF emission are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions Class A The system is suitable for use in all establishments, other than
CISPR 11 domestic and those directly connected to the public low-
voltage mains supply for residential buildings.
In accordance with standard EN 60601-1-2 (2nd ed.), the system is intended for use in the
electromagnetic environment specified below.
The customer or the user of system should assure that it is used in such an environment.
Tab. 202
Immunity test EN 60601 Compliance level Electromagnetic environment -
test level guidance
Tab. 204
Immunity test EN 60601 Compliance level Electromagnetic environment -
test level guidance
d = 1.2 √P
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
Note 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection by structures, objects and people.
Note 2: Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast can not be predicted in theory or
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters an
electromagnetic site survey should be considered. If the measured field strength in the location in
which the system is used exceeds the applicable RF compliance level above, the system should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the system or the use of a shielded area with greater
RF shielding capacity.
The system is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled.
The customer or the user of the system can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the
system as recommended below, according to the maximum output power of the communications
equipment.
Tab. 206
Rated maximum Separation distance according to frequency of transmitter
output power of m
transmitter
W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = 1.2 √P d = 1.2 √P d = 2.3 √P
0.01 0.12 0.12 0.23
100 12 12 23
Note 1: For transmitters rated at a maximum output power not listed above, the recommended separation
distance in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection by structures, objects and people.
2 TECHNICAL DATA
2.2.1 SYSTEM
Note: See paragraph 1.3 in Part 6 of this manual for details about the technical data of the DRX1
wireless detector (system with two detectors).
See paragraph 1.3 in Part 7 of this manual for details about the technical data of the PIXIUM
3543EZ wireless detector (system with two detectors).
Models Claymount SSMC 613 and SSMC 601 / preamplifier 1001 / 3 measuring areas
Technology Solid state
Dimensions Overall dimensions 472x460x3.3 mm
Useful area 430x430 mm
Weight 1 kg ± 100 g
Power supply Supply rate ± 15V DC (±2V)
Consumption 40 mA
Technical data Sensitivity 0,8 V/μGy
Attenuation coefficient <1.05
(80kV, 25mm Al, FFD 100cm)
Measurement fields Left, Right, Centre
(choice of any combination)
2.2.7 MONITOR
2.2.9 CABLES
Note: - See paragraph 1.4 in Part 6 of this manual for details about the cables used for the wireless
DRX1 detector unit (system with two detectors).
- See paragraph 1.4 in Part 7 of this manual for details about the cables used for the wireless
PIXIUM 3543EZ detector unit (system with two detectors).
2.3 DIMENSIONS
Note: - See paragraph 1.5 in Part 6 of this manual for details about the dimensions of the components
in the Wireless DRX1 detector unit (system with two detectors).
- See paragraph 1.5 in Part 7 of this manual for details about the dimensions of the components
in the Wireless PIXIUM3543EZ detector unit (system with two detectors).
2.3.1 CABINET
Standard cabinet
Weight: 126 kg
locating pin
X-ray
X-ray center
locating pin
Weight: 25 kg
N° 8 fixing holes
(M5, max length: 10 mm)
N° 2 locating pins
(Ø6, length: 6 mm)
thickness: 10 mm
Weight: 1 Kg
P3 P1 P2
P1: Sensitivity for area 1
P2: Sensitivity for area 2
P3: Sensitivity for area 3
thickness: 3.3 mm
weight : 1 Kg
1001 preamplifier
Field 1
Reset Field 2
Field 3
thickness: 3. mm
weight : 1,15 Kg
Field 1
1007 Preamplifier Field 2
Field 3
Reset Field 4
Field 5
N° 4 x M4 L holes
deep 10 mm
For fixing :
use the 4 x M4 holes Standard VESA 100 or Standard VESA 75
deep 8mm
Receiver Transmitter
Thickness: 31 mm Thickness: 22 mm
Weight: 0.08 Kg Weight: 0.12 Kg
Cable
Diameter: 4 mm
Length: 15 m
Weight: 18 Kg Weight: 19 Kg
Monitor trolley for one 18” /19” LCD display Monitor trolley for two 18” /19” LCD displays
Weight: 26 Kg Weight: 27 Kg
Monitor trolley for one 18” /19” LCD display Monitor trolley for two 18” /19” LCD displays
CE 93/42 Certified CE 93/42 Certified
2.4 COMPATIBILITY
The system has been tested using the following generators and collimators:
2.5 FUSES
The system is protected by a series of fuses, as shown in the table and drawings below:
Thermomagnetic SW1
Fig. 1
Cabinet side
F1
F4 F5 F3 F2 F1
Fig. 2
MAIN CONTROLLER
F2 F1 F1 F2
Fig. 3
WARNING PARTS Supply board
ARE LIVE WHEN
SYSTEM IS OFF
F2 F3 F1 F2 F1
RFPS Board
Fig. 4
Detector supply board
F1
Fig. 5
RGBC 01 board
F1
Fig. 6
RDAM board
F1
Fig. 7
RDAMSP board
3.1 PACKAGING
1 Detector
2 Cabinet (Image processor / Main Controller / PU 4343 / Detector feeder / Power unit)
3 - Ion Chamber
- Dose Area Meter (optional)
- CD Re-Writer (optional)
- Remote control IR (optional)
- Cables
- Alphanumeric keyboard and Mouse
- Manuals
4 Monitor
5 Cabinet Top (optional)
box 3 box 5
box 4
box 2
box 1
Nota: - See paragraph 1.7 in Part 6 of this manual for details about packaging of system fitted with
two detectors (with Wireless DRX1 detector).
- See paragraph 1.7 in Part 7 of this manual for details about packaging of system fitted with
two detectors (with Wireless PIXIUM3543EZ detector).
3.2 STORAGE
We recommend the following temperatures and humidity levels for storage of the system, likewise
during its transport:
limits recommended
Part 2: INSTALLATION
LIST OF CONTENTS
1 INTRODUCTION
1.1 Installation procedure 1.1 to 1.2 B 30/08/13
1.2 Unpacking
1 INTRODUCTION
The flow chart below shows the installation procedure (see note at bottom of page):
START
Unpacking
(see para. 1.2 below)
Mechanical installation
(see Chapter 2 below)
Electrical connections
(see Chapter 3 below)
Set-up procedure
(see Chapter 4 below)
Acceptance Test
(see Part 3)
Only qualified personnel with suitable training should be allowed to install and commission the system.
Note 1: It is essential that the mains power rating to be used for the system is checked to guarantee
correct installation and functioning (this must be 230 V ±10%- 50/60 Hz).
*Note 2: - You must also calibrate the Wireless DRX1 detector if the system is fitted with two detectors
(see chapter 3 in Part 6).
- You must also calibrate the Wireless PIXIUM 3543EZ detector if the system is fitted with
two detectors (see chapter 3 in Part 7).
1.2 UNPACKING
● Cabinet: Remove the cabinet from the packing box following the instructions shown in the figure on
the outside of the box:
- Remove the straps securing the box to the wooden pallet below.
- Open the cardboard box and remove the packing blocks.
- Remove the wooden ramp and fix to the base of the pallet.
- Remove the blocks in front of the wheels.
- Gently push the cabinet to slide it off the base of the box.
● Simply open the other packing boxes, without using any tools that could damage the contents.
See “Annex A” at the end of this manual for PIXIUM 4343 detector unpacking and storage.
2 MECHANICAL INSTALLATION
2.1 CABINET
- Mount the wooden base (optional) and fix with the 4 screws provided.
- Position the cabinet and then engage the brake system (two threaded tie-rods, whose base should
touch the ground).
Note: - The threaded tie-rods used for braking can be rotated using a screwdriver in the holes
accessed from the threaded shaft of the tie-rods.
2.2 MONITOR
See paragraph 2.3.4 in Part 1 of this manual for the overall dimensions and for details on how
to fix the monitors using the holes indicated (hanging or on the trolley).
- the ion chamber must be installed in a compartment planned above the detector.
X-Ray tube
Ion chamber
Detector
You also need to take into account the weight, size and envisaged fixing systems when installing the
above components. See the drawings in paragraphs 2.3.2 and 2.3.3 in Part 1 of this manual.
Nota: - See paragraph 2.1 in Part 6 of this manual for details about the Wireless DRX1 detector
installation (for the system fitted with two detectors).
-See paragraph 2.1 in Part 7 of this manual for details about the Wireless PIXIUM 3543EZ
detector installation (for the system fitted with two detectors).
2.3.1 DETECTOR
● Unpacking
Caution: The detector needs a 24-hour period of acclimatisation before you should open the plastic
bag containing it.
See “Annex A” at the end of this manual for PIXIUM 4343 detector unpacking and storage.
● Earth
Caution: Do not use the cable and connector screens as an earth connection.
Use the M4 screw provided on the panel covering the detector to earth the detector (see figure
below).
● Cooling
To ensure proper cooling of the detector, make sure that:
- where possible, air naturally circulates around the detector (min flow: 0.3 m3/min),
- the working room temperature falls within the set limits (see technical data in paragraph 2.2.3 of
Part 1),
- the heat resistance between the room air and the rear surface of the detector is less than
0.6°C/W.
● Fixing
Caution: To avoid any risks during installation of the detector, at least two people must be involved
in fixing the detector to its support.
Install the detector so that the status LEDs are easily visible (see para. 5.5 in Part 5 of this manual).
Caution: The side of the measure chamber that must face the X-ray tube is identified by the words
TUBE SIDE (see figure above for its position). The figure shows the MEDYS chamber,
model MIC-10 COD. 2252611.
3 ELECTRICAL CONNECTIONS
Integration allows for the other devices to be connected to the system to suit the type of exam
programmed by the system.
More specifically:
- Integration with the X-ray generator allows for the transfer of programmed radiological
parameters and reception of exposure values.
- Integration with the X-ray collimator allows for automatic setting of the aperture of the
collimator to suit the type of study.
- Integration with the anti-diffusion grid ensures the correct selection of the grid and focal
distance.
- Integration with the stand means that the stand can be used in set positions to suit the type of
exam foreseen.
The parameters are programmed in the setup folder for each exam (see Chapter 4 below: “Setup
procedure”).
3.2 CONNECTORS
List of connectors:
Note1: The following abbreviations are used here and elsewhere in the manual:
- PIDRF43 abbreviated to: PID
- PIXIUM FE 4343F abbreviated to: PIXIUM
- PU RF4343 abbreviated to: PU
Nota2: - See paragraph 2.2 in Part 6 of this manual for details about electrical connections of the
Wireless DRX1 detector installation (for the system fitted with two detectors).
- See paragraph 2.2 in Part 7 of this manual for details about electrical connections of the
Wireless F.P. detector installation (for the system fitted with two detectors).
Interfaccia USB
(Presente solo per series n° 3xxx)
PU: B1-C1
WIRELESS Network
(optional) B14-C1
Supply
Main controller: CM15
from ASPS00: CP10
Extender for Keyboard + Mouse DICOM Network
Fig. 1
Video Processor PID
Dose Area Meter (optional) PID: B2-C1 RS-232 (for X-Ray generator) Supply: ASPS: CP12
PU: B1-C1 Rx Generator / Accessory Rx Generator / Accessory CAN BUS I.R. receiver (optional)
Fig. 2
Main Controller
CP13: 230VAC 21VAC board CP10: 230VAC PID CM2: RGBC: CM3 CM1: ASPS-CP13
Fig. 3
WARNING PARTS ARE LIVE Main supply board
WHEN SYSTEM IS OFF
RFPS board
Fig. 4
Detector main supply unit
WARNING PARTS ARE LIVE
WHEN SYSTEM IS OFF
Fig. 5
Board RGBC 01
VIDEO IN (VGA) VIDEO IN (VGA)
PID: B4-C1 PID: B5-C5 > DVI/VGA Converter
Working Monitor Reference Monitor
Fig. 6
PIXIUM RF 4343 F3 detector
ASPS: CP9 (230VAC) PID: B11-C2 PID: B14-C1 / Main controller: CM10 PID: B19-C1
Fig. 7
Detector PIXIUM RF4343: PU module
CM1
P3 P1 P2
CM1
TO MAIN CONTROLLER
Fig. 8b
Measure chamber Claymount mod. SSMC 508 / PREAMPLIFICATORE 1007
Note. The figure 8b represents the Claymount mod. SSMC 508 5-area chamber
It is also the same for Claymount models SSMC 613 and SSMC 601 - 3 area chamber
Fig. 9a
19” Tecnint colour monitor
Fig. 9b
19” Invicta monochrome monitor
Fig. 9c
19” Invicta colour monitor
Connect to
Main Controller: CM22
Fig. 10
Dose Area Meter (optional)
Attention:
ONLY connect and/or disconnect the DAM to/from the Main Controller when the machine is
switched off.
Attention:
The DAM is a very FRAGILE device: handle with care.
3.3 CONNECTIONS
See the descriptions below, the figures and block diagram that follow for details of the connections
required during installation.
There is a slot at the bottom of the cabinet for the cables leaving this, as shown in the figure below.
A cabinet can be supplied, on request, with three slots in the side to facilitate the positioning and fixing
of the output cables (see figure below).
Plastic grid “D” near the cables inside the cabinet lets you fix the cables with straps “E” (see figures
below).
C
B
A
The three side slots let you split the output cables into practical groups by function.
We recommend splitting the cables as follows:
- ETHERNET cable,
- X-ray generator/Accessory (Main Controller CM6 and CM8) connection cables,
- RS232/CAN cable for interface with X-ray generator,
- CAN cable(s) for stand and collimator connections.
• Connections between the keyboard and mouse and the rack (see also para. 3.3.3):
- connect the extension cables to the Extender transmitter module
- connect the keyboard to Receiver-KB
- connect the mouse to Receiver-MS.
Note: To ensure that the mouse and keyboard are properly connected, you must check
the colours of the jacks (green for the mouse and purple for the keyboard).
• Monitor connections:
- 230 V AC supply (+ earth) from supply board: TB3.
- VGA cable for working monitor to RGBC: CM2.
• DVD-Recorder connection:
- signal cable to PID: B10-C1.
- 230 VAC power supply from supply board: TB3.
- CAN BUS cable for peripheral devices (collimator, generator, stand) to MainController: CM4
or CM5.
Note: The CAN BUS is connected using a 150 Ohm resistance fitted inside a connector on Main
Controller CM4.
The figure below shows all the mouse and keyboard connections.
Main
Controller 20 m 1,5 m
1,5 m
EXTENDER
RECEIVER
Note: It is crucial that the detector is always left powered up, so that its working temperature is as
constant as possible.
The room power supply system therefore has two different power supplies, with two distinct insertion
systems: one for the acquisition system and another for the detector.
Figure 9 shows the general power supply system including the necessary emergency buttons.
PIXIUM RF4343
PIXIUM RF4343
Fig. 9
The system must be fed with single-phase 230 V~, 50/60 Hz.
Electrical input is 700 VA.
2
Use conductors with a minimum cross-section of 1.5 mm² (max of 4mm ).
- Connect the supply conductors to connector TB1 on the supply board (fig. 10a and 10b).
- Connect the earth conductor to terminal board GP1 on system supply board (using a conductor with
a minimum cross-section of 2.5 mm²).
- If the cables leave the cabinet through the slot at the bottom of this, the power cable must be
positioned as shown in Figure 10c and fixed with a strap to the two cable clips so it cannot come
loose accidentally.
- If the cables leave the cabinet through the side slots (optional), the power cable must be positioned
as shown in Figure 10d and fixed with a strap to the plastic grid so it cannot come loose accidentally.
Note 1: If a supply shielded cable is used, the shield must be earthed, as shown in the figures (10a-
10b).
Note 2: It must be possible for the acquisition system supply to be cut using the room cut-out
command.
Caution: Check that the emergency switches required for the room do actually cut the system power
supply.
TB1
1 1
2 2
3 3
230Vac
Power Supply Unit
Input
GP1
Fig. 10a
Fig. 10b
It is crucial that the detector is always left powered up, so that its working temperature is as constant
as possible.
It must, however, be possible to cut the detector's power supply with a manual switch (see fig. 9
above).
The detector is fed via its supply board with stabiliser inside the system cabinet.
The detector supply board must be fed with single-phase 230 V ~, 50/60 Hz.
Electrical input is max 130 VA.
2
Use conductors with a minimum cross-section of 0.75 mm² (max of 4mm ).
- Connect the supply conductors to connector TB1 on the detector supply board (figures 11a and 11b).
- Connect the earth conductor to the earth terminal on the detector supply board.
- The cable must be laid as shown fig.11c and set at fixed cable to prevent the accidental posting.
Note: If a supply shielded cable is used, the shield must be earthed , as shown in the figures (11a-11b).
TB1
F1
230 Vac
F2
Fig. 11a
Fixed cable
230Vac
Fig. 11b
Fig. 11c
COLOR
WORKING MONITOR
VIDEO SIGNAL
DVD-RECORDER
230Vac
RGBC 01 BOARD 21Vac
CM2 CM1
EXTENDER
RECEIVER
DVI/VGA CONVERTER
KEYBOARD
B12-C1 B11-C1 B10-C1
MOUSE B9-C1
Video Processor
SUPPLY
B9-C2
POWER
B11-C2
DICOM CM5
NETWORK B11-C3
B14-C2 B2-C1
REFERENCE MONITOR
I/O RS-232
INFRARED
REMOTE
RECEIVER (*)
CM22
CM15
CM10 REFERENCE MONITOR
CM6 MAIN CONTROLLER
CM8
ION. CHAMBER
CM24
230Vac
CM4/5
230Vac
DAM (*)
CM11
NET.1
230Vac
B1-C1
+24V OUT
PU4343
PIXIUM FE4343F
DATA LINK +24V IN
B6-C2
RS-232
CP9
CAN BUS
CP10
INJECTOR POWER SUPPLY CP12
TB3 ON/OFF CP13
TB3
TB1
MAIN 230Vac
always present
MAIN 230Vac
Fig. 12
Block diagram
(*)=OPTION
The system mainly uses the following connectors for its peripheral interface:
MainController-CM6 for connection to accessory / X-ray generator,
MainController-CM8 for connection to the X-ray generator.
PREP_RAD 30
PREP_RAD REQ_RAD 12
COM_REQ 31
RAD
ZOOM
ZOOM 1 (30x30) 34
ZOOM 2 (20x20) 16
SEL_FLAT_PANEL 1 17
SEL_FLAT_PANEL 2 13
SEL_FLAT_PANEL_1_CPI 14
SEL_FLAT_PANEL_2_CPI 33
0V1 ACC. 15
0V ACC. 10
+V ACC. 29
+24V
SEL_INJ. 3
SEL_INJ.
SYNC_INJ. 21
SYNC_INJ.
23
CM11
SYNC_INJ_1 1
SYNC_INJ_1
SYNC_INJ_1 2
SYNC_INJ_1
ACCESSORY/X-RAY GENERATOR
INTERFACE
revision G PART 2 page 3. 18
code 87 80 014C - Mti S2 03.doc.doc
TECHNICAL MANUAL
HIRIS RF43
CM8
+24V
+24V GEN. 7-20
PREP_PULSED 14
PREP.HCF
SEL_PULSED_HCF 1
SEL_PULSED_HCF
SEL_PULSED 2
SEL_PULSED
READY_AQ_RAD/HCF 17
READY_AQ_RAD/HCF
READY_ACQ_FL 5
READY_AQ_FL.
ABC_LOW 18
ABS_LOW
ABC_HIGH 6
ABS_HIGH
ABC_ERROR 12
ABS_ERROR
0V 13
+24V
GEN_READY 23
EXON 21
0V GEN. 19
CAN L =2
CAN H =7
0V =3
CM4
124E
CAN L =2
CAN H =7
0V =3
CAN BUS
TERMINATION
ACCESSORY/X-RAY GENERATOR
INTERFACE
4 SETUP PROCEDURE
4.1 INTRODUCTION
The system set-up procedure basically involves the activities listed in the table below in the order in
which they are to be performed:
Note: - You must also calibrate the Wireless DRX1 detector if the system is fitted with two detectors
(see chapter 3 in Part 6).
- You must also calibrate the Wireless PIXIUM 3543EZ detector if the system is fitted with two
detectors (see chapter 3 in Part 7).
The SD4343 menu is found on the desktop (see chapter 5 below), while the other menus used for the
system set-up are all found in the Main Frame, Option / Setup.
The menu can only be opened when all archives are closed: select Option / Setup / General Setup.
Note: - See paragraph 2.3.2 in Part 6 of this manual for details of the General Setup settings if the
system is fitted with two detectors (with Wireless DRX1 detector).
- See paragraph 2.3.2 in Part 7 of this manual for details of the General Setup settings if the
system is fitted with two detectors (with Wireless PIXIUM 3543EZ detector).
The tables below show the options in the General Setup menu:
Min. Time Remaining Threshold for Offset. Set 300 Time in seconds =
in Offset Life Time If the effective time is less than 10 minutes
this parameter, calibration
automatically takes place when
a new exam is opened.
Active Dynamic Enables the logical mode offset If not selected, the
Update Offset refresh function every 5 Recommended system automatically
seconds when study list shown updates the logical
mode offset when
each study is
opened.
Smart PU When enabled, the system Recommended
Connection acquires the offset during fast
rebooting of the application
(within 5 min since last closed).
This function is not performed
in the case of a total system
reboot.
Settings Notes
Default Open Dir Set: D:\Archives
Juke Box Web Page Network connection with the Rebox device (CD/DVD burner).
Settings Notes
Web Page https://192.168.50.108:8080/rebox/view.php?key=hiris Set only if the
Rebox is installed
Specification of the hard disk directory to be used for Acrobat Reader, needed
PDF Documentation Reader
to view online manuals.
Settings Notes
Set:
C:\ProgramFiles\Adobe\Acrobat 7.0\Reader\Acrord32.exe
Commands
Main Controller RAD To open the parameter setup menu for the grids and (see paragraph 4.2.4
Setup define the integrated components (X-ray generator, below).
collimator, etc.)
Factory Settings To set the default values for the fields in the menu Only enable with
shown integrator key inserted.
See 4.2.1.
Language Settings To set the language Enabled after entering
specific integrator
password
See 4.2.2.
Auto Delete Sets the conditions for enabling the automatic study See 4.2.3
deletion function.
Edit Application Name To define the name of the application Parameter set by
system integrator
Customize RX Filter To open the setup menu for the X-ray beam filters (if See paragraph 4.2.6.
fitted) on the X-ray collimator.
Cancel To quit General Setup without saving any changes
OK To quit General Setup saving changes
This function restores the default factory settings for all the General Setup parameters.
Confirm with Yes: the system restores all the default parameters.
This function lets you set the operator interface language. It is enabled after entering an access
password. Password to access this menu should be request to the supplier.
The new language is used the next time you run the application.
This menu lets you set the parameters for the AUTO DELETE function.
If enabled, this function automatically deletes the oldest exams in the current archive.
Options Notes
This function is only enabled with the integrator key. It lets you set a new name for the system (i.e. to
change the default name).
The new name appears on the screen in the operator interface windows.
It is also sent to the following DICOM fields: Device Name and Device Name Serial Number.
Grid Settings This function can only be used if the value of the actual focal
distance is supplied via CAN from the stand/collimator.
Settings Notes
GRID 1: Min focal distance at Set the envisaged value The system allows for integration
DFFmin which exposure is enabled for the radiography with two extractable grids (thus
with grid 1. system exposure without any grid is also
GRID 1: Max focal distance at Set the envisaged value possible).
DFFmax which exposure is enabled for the radiography
with grid 1. system If the actual focal distance falls
GRID 2: Min focal distance at Set the envisaged value outside the accepted range,
DFFmin which exposure is enabled for the radiography image acquisition is inhibited
with grid 2 system (and thus also X-ray emission).
GRID 2: Max focal distance at Set the envisaged value
DFFmax which exposure is enabled for the radiography
with grid 2. system
NO GRID: Min focal distance at Set the envisaged value
DFFmin which exposure is enabled for the radiography
without any grid. system
No GRID: Max focal distance at Set the envisaged value
DFFmax which exposure is enabled for the radiography
without any grid. system
This menu lets you define up to 4 different X-ray beam filters on the collimator (if these are fitted).
The filter to be used can be set in the exam card using the RX filter parameter (see paragraph 4.5.4.2
below).
Details of the filter parameter settings are sent by the equipment via the DICOM DOSE SR service
(see paragraph 4.6.16.3 below).
RX Filter Config
Settings Notes
Filter Use to select the filter Select the relevant
you want to set item in the list.
Thickness Min. Minimum thickness of 0 - 5000 microns These two values will
[micron] the material be the same if the filter
Thickness Max. Maximum thickness of 0 - 5000 microns is the same thickness
[micron] the material throughout.
The system allows for the presence of the integrated peripheral components listed in the table below:
The presence of the integrated components/functions is set using the parameters in the Main
Controller RAD Setup menu - Options:
You can also use the Remote controlled stand option to check the correct position of the grid.
In fact, the system lets you set the type of grid to be used to suit the exam type (GRID 1, GRID 2 or
NO GRID) and then check that this is correctly position before X-ray emission:
• Use the Exam Setup menu to select the type of grid (see paragraph 4.4.4 below)
• Use the Main Controller RAD Setup menu to set the grid acceptance parameters.
For example, if the stand has 2 grids, focussed at 120 and 180 cm, the settings may be:
GRID 1: DFFmin Min focal distance for exposure with grid 1: 110
DFFmax Max focal distance for exposure with grid 1: 150
GRID 2: DFFmin Min focal distance for exposure with grid 2: 140
DFFmax Max focal distance for exposure with grid 2: 200
In which case, if the selected exam foresees the use of GRID1 and the focal distance supplied by the
system (via CAN bus) is not within 110 - 150 cm, image acquisition is not enabled.
.
Remotely controlled stands normally let you position the different system components automatically in
order to get the right layout for the type of exam set in the exam card. You can thus program the type
of projection required in the exam card (see paragraph 4.4 below): the system sends the relevant code
to the stand when you open the exam.
For example, the figures below show the system in the positions for:
- projection 1 (exams with patient lying on bed),
- projection 2 (exams with patient standing, e.g.: chest),
- projection 3 (side projection, with patient lying on bed).
1 2 3
The position is set in the Exam Setup menu using the Accessory Code option in the Miscellaneous
Properties window.
Up to 8 positions are possible. The code for the selected position (1-8) is sent to the stand (via CAN
bus) when you open the exam.
If the Remote controlled stand option is enabled, the system checks that the current position of the
stand corresponds to that programmed for the exam.
You can, however, decide whether any differences in position should inhibit image acquisition (and
thus X-ray emission) or simply generate a warning for the operator.
Check the following option in the Main Controller RAD Setup menu:
● Check Technique
- Checked: when the stand is incorrectly positioned for the type of projection , X-ray emission is
inhibited and the following alarm appears:
SELECT TECHNIQUE #x (if the selected exam type requires projection x),
- Not checked, X-ray emission is enabled, even if the stand position is not correct. The operator is
warned of this fact:
On accessing the Logical Mode Setup menu (after entering the technical password), you can set a
few parameters concerning the detector’s Logical Modes.
Each acquisition mode in the system is identified by a Logical Mode. The following parameters can be
set for each Logical Mode:
Sensitivity
Logical Mode Field View(cm) Logical Mode Id PU Mode
(Isb/nGy)
Continuous Fluoroscopy 43x43 1 3 16.1
Continuous Fluoroscopy Z1 30x30 2 7 10.7
Continuous Fluoroscopy Z2 20x20 3 15 10.7
Continuous Fluoroscopy Z3 15x15 4 27 5.35
Pulsed Fluoroscopy 43x43 5 2 16.1
Pulsed Fluoroscopy Z1 30x30 6 7 10.7
Pulsed Fluoroscopy Z2 20x20 7 22 10.7
Pulsed Fluoroscopy Z3 15x15 8 27 5.35
Rad 3 i/s 43x43 10 17 0.6
Rad 8 i/s 43x43 11 4 1.2
Tomo 4s 43x43 15 17 0.6
Tomo 2s 43x43 16 17 0.6
Border: to define the number of lines or columns that should not be shown on each side of the
detector. Set the detector borders as shown in the table below:
Logical Mode PU Mode Left Border Right Border Top Border Bottom Border
1 3 7 7 2 2
2 7 0 0 2 2
3 15 0 0 2 2
4 27 0 0 0 0
5 2 7 7 2 2
6 7 0 0 2 2
7 22 0 0 2 2
8 27 0 0 0 0
10 17 20 20 3 3
11 4 10 10 2 2
15 17 20 20 3 3
16 17 20 20 3 3
The system offers a series of Reference exams covering a wide range of applications, where the
specific exam parameters have been pre-set to get a correct image.
These parameters are found in the EXAM SETUP card and are split as follows:
• Image processing parameters, under a single heading Post Processing Setup (only accessed by
the system integrator).
• Radiological, image orientation and DICOM parameters, that can be changed to suit the type of
installation and operator requests (see point 4.4.4 below).
Access to the Exam Setup card (and thus the exam) involves two choices: anatomic part and
projection type.
Up to 100 different anatomic parts can normally be defined, with up to 100 different projection types
per zone:
- each anatomic part has an ID (00 - 99),
- and each projection type a second ID (00 - 99).
See annex 7.10 in this Part of the manual for the full list of exams possible with this system.
Example:
Anatomic part Projection type
ID Description ID Description
00 SKULL 00 AP
01 LL
02 …
… …
… … … …
03 ABDOMEN 00 AP
01 Decubitus
02 …
… … … …
During installation, you need to decide which exams are to be added to the “local library” for the
operator.
The following functions let you create this Local exam library, starting from the Reference exams:
o Clone, this generates a new Exam Setup card, identical to the current one, but with a
new ID. You change the name and free parameters in this card (radiological parameters,
image orientation, zoom, etc.).
o Remove, to eliminate the card from the archive of reference exams.
To access the exam card manager, select Option / Setup / RF Exam Setup and enter the technical
password. The RF Exam Setup Manager opens:
Click on Go to card to select the card: the Exam Setup menu opens. This is split into 4 groups of
parameters, as shown in the figure below:
“Miscellaneous”
parameters
Not all the parameters shown in the exam card can be changed by the operator: the ones that can be
changed are described in paragraph 4.4.4 below.
Note: - To select the exams linked to the Wireless DRX1 detector see paragraph 2.3.4 in Part 6 of this
manual (for the system fitted with two detectors).
- To select the exams linked to the Wireless PIXIUM 3543EZ detector see paragraph 2.3.4 in
Part 7 of this manual (for the system fitted with two detectors).
If necessary, you can create a new exam card, starting from an existing one. Of course, the new exam
card needs to be given a different ID. You can then change the name of this new card and the free
parameters (radiological parameters, image orientation, zoom, etc.):
- Click on the Clone command that appears in the window that lets you select the card you want to
clone.
- Confirm with OK: the window for the new card automatically opens, letting you assign the new codes.
Note: You do not usually need to create a new anatomic part, but simply a new exam within a
given anatomic part. We therefore recommend that you keep the same Anatomic Part -
Code/ Description as the cloned exam.
Click on the Organize option in the Exam Setup Manager menu to open the Tech Card Organizer
menu that contains the list of all the reference exams in the system.
• Change name
- Move the cursor to the anatomic part of exam you want
- Use the Change Anatomic Name or Change Card Name commands to open the Change
Name window
- Select the name and change this (NB: no double names are permitted).
- Confirm with OK.
• Define local library: the Tech Card Organizer menu presents the list of all the anatomic parts and
their exams in the system and forms the reference exam archive, available to the installer.
The anatomic parts and the “activated” exams form the Local exam library, available to the installer
(in the working frame).
- To add an anatomic part or exam to the local library, select it and then click on Activate: a
green symbol indicates that it has been activated.
- To remove an anatomic part or exam from the local library, select it and then click on
Deactivate: a red symbol indicates that it has been deactivated.
• Reorganise local library: the Tech Card Organizer menu lists the anatomic parts and their exams
in the local library in the order in which they will be presented to the operator in the working frame.
To change the order:
- Select the anatomic part or exam you want to move.
- Use the Up or Dw commands to move its position within the list.
You need to check (and possibly modify) the free parameters of each selected exam in the card.
Below is a description of each parameter, split into “general parameters” and “specific parameters for
each acquisition mode” (radiography / continuous fluoroscopy / pulsed HCF fluoroscopy).
general parameters
Settings Notes
Size valid To enable the patient Select at least one size If a size is not enabled
types (no check), the relevant
column of parameters
will be disabled.
Miscellaneous
Settings Notes
Accessory Code To set the working 0 - 100 (RF43 det.) Use to set the working
position of the stand 101 - 199 (wireless position of an integrated
det.) stand with programmable
positioning.
See also paragraph 4.3.1.
Shutters X/Y Inversion X / Y inversion of the Use if the current X/Y
electronic shutters on representation of the
the image electronic shutters is
not the actual
situation.
Flip/Rotation Code Digital orientation: set to The value to be set depends
get the right orientation 0-7 on the orientation of the
of the image on the detector with respect to the
monitor. patient.
The table below* explains the
meaning of the code.
Auto Select HCF To select HCF pulsed To select HCF pulsed If not selected, Continuous
fluoroscopy as the fluoroscopy as the Fluoroscopy is the default
default mode when the default mode when the mode when the exam is
exam is opened. exam is opened. opened
Select Fluoro LOCK-IN To select Continuous Choose this function In fluoroscopy with automatic
Fluoroscopy with active in order to activate the kV, the system will adjust the
LOCK-IN acquisition in kV until reaching the correct
Continuous dose value, then it will
Fluoroscopy with maintain the dose value for all
LOCK-IN the acquisition.
Single Exposure With Automatic Check to enable If single exposure is checked,
Procedure modality, if Single Exposure during Automatic Procedure
checked set the exam modality exam with modality, if a single
as a single exposure Automatic Procedure radiography image is
modality exam modality performed, the procedure
otherwise, a multi image step increase automatically.
modality exam is Otherwise, multi image
enabled. exposure can be performed
and the step of the procedure
increase manually by the
operator (see User Manual)
*Image orientation:
Settings Notes
Modality To set the default To set one of the foreseen The 2 parameters
DICOM mode. fields: (radiography /
CR \ RF fluoroscopy) are only
CR \ XA used if the exam
DX \ RF creation mode is not
DX \ XA already defined when
(radiography \ fluoroscopy) the Worklist is received.
Protocol Code identifying the To be agreed with the
procedure network administrator
Description Description of the exam To be agreed with the
network administrator
Body Part Part of the body to which Select one of the set fields:
the exam refers UNDEFINED
SKULL
CSPINE
TSPINE
LSPINE
SSPINE
COCCYX
CHEST
CLAVICLE
BREAST
ABDOMEN
PELVIS
HIP
SHOULDER
ELBOW
KNEE
ANKLE
HAND
FOOT
EXTREMITY
LEG
HEAD
HEART
JAW
ARM
NECK
Laterality Projection side Select one of the set fields:
Undefined
Right
Left
Unpaired
Both L&R
View Envisaged projection Select one of the set fields:
(view) Undefined
AP
PA
LL
RL
RLD
LLD
RLO
LLO
• Radiological parameters
Rad /item Set the parameters for each programmed patient type (Size
valid).
Settings Notes
Max rate To set the max - Max Rate: selection of max rate of acquisition for the
radiography rate possible selected logical mode.
for each size. When the
max rate is selected, the LM10:
max X-ray pulse time is - up to 1i/s with max X-ray pulse lasting 500ms
automatically defined. - up to 3i/s with max X-ray pulse lasting 150ms
LM11:
- up to 1i/s with max X-ray pulse lasting 500ms
- up to 3i/s with max X-ray pulse lasting 250ms
- up to 8i/s with max X-ray pulse lasting 60ms
Rate Mode To set the acquisition - Def Rate: acquisition at a The chosen mode is
mode constant rate throughout X-ray automatically set when
emission you open the exam.
- Sequence: acquisition at
variable rates, as per
programmed sequence.
To select the default Log. Mode 10: Log. Mode 11: Rate set automatically
radiography rate, Single Single when you open the exam.
Default Rate presented to the operator 1/s 1/s You can, however,
when the exam opens. change it manually.
You only can set this
parameter if the Rate 1.5i/s 1.5i/s 2i/s
Mode is set as Def Rate. 3i/s 3i/s 4i/s
The choice of rates 8i/s
depends on the Max rate
parameter (see above).
Phase 1( ..6) : To set the timing and rate The Rate field has the same You can program from 2
Rate / Time of acquisition for each settings as the Default Rate field to 6 phases.
phase (max 6) in a described above. The first phase with the
programmed sequence. The Time field is expressed in time set as 0 will act as
This can only be set if field seconds. the end of the sequence.
Rate Mode is set as Acquisition will continue at
Sequence. the rate set for this last
phase until you release
the X-ray command.
Lockin Enable To enable the Lockin Select to enable the function. Only available with the
function. DSA option.
This calculates the
exposure time from the
start of a radiography run
and keeps it constant until
the X-ray command is
released.
Auto Mask Enable To enable the automatic Select to enable the function. Only available with the
mask pickup function, with DSA option.
the timing programmed in
field Auto Mask Delay.
Auto Mask Delay To set the delay with Set the delay in seconds. Only available with the
which the auto mask is DSA option.
picked up after the start of
X-ray emission.
Injector Enable To enable the start Select if the exam foresees the Only available with the
command for an injector use of the injector. DSA option.
integrated in the X-ray
system (optional).
Injector Delay To set the time (delay or The Injector Delay field is Only available with the
advance) when the expressed in seconds. A DSA option.
injector starts with respect negative time will activate the
to the start of X-ray start injector signal before X-ray
emission. emission starts; a positive value
will delay the activation of the
signal.
Dose uGy To set the X-ray dose in Typical working value:
radiography mode 0.5 - 10 µGy 2.5 µGy.
kV Optimal kV value for 40 - 150 kV This value will be sent to
exposure the X-ray generator when
the exam is opened
mA Optimal mA value for 1 - 1000 mA This value will be sent to
exposure the X-ray generator when
the exam is opened.
Field only active if in AEC
mode.
mAs Optimal mAs value for 1 - 1000 mAs This value will be sent to
exposure the X-ray generator when
the exam is opened.
Field only active if in
manual exposure mode,
i.e. not AEC mode.
AEC Left To enable the automatic Left If you do not select any
AEC Center AEC control by choosing area, the X-ray generator
AEC Right at least one area of the Center performs exposure of a
ion chamber controlled length. (NO
Right AEC)
Focal Spot Focus of the X-ray tube 0 = small focus This value will be sent to
used for exposure the X-ray generator when
1 = large focus the exam is opened.
Focal distance Envisaged focal distance 80 - 250 cm This value will be
for exposure compared with the actual
focal distance in order to
enable exposure
To set the image This factor will be used by
Calibration Factor enlargement factor the system for TRUE SIZE
(depending on the focal 1.0 - 1.5 printing and default
distance and the distance calibration of the
between the object and measurements on the
the detector). image.
Grid Anti-scatter grid to be 0 = no grid This value will be sent to
used 1 = grid 1 the stand when the exam
2 = grid 2 is opened .
the value is then
compared with the actual
grid to enable exposure
Rx Filter To select the X-ray beam 0 = no filter This value will be sent to
filter in the collimator 1 = filter 1 the collimator when the
2 = filter 2 exam is opened
3 = filter 3
Screen selector X Horizontal aperture of the 1 - 43 cm This value will be sent to
X-ray collimator the collimator when the
exam is opened
Screen selector Y Vertical aperture of the X- 1 - 43 cm This value will be sent to
ray collimator the collimator when the
exam is opened
Default Rate Default radiography rate, Log. Mode 10: Log. Mode 11: The rate is set
presented to the operator Single Single automatically when the
when the exam is opened 1/s 1/s exam is opened
1.5/s 2/s
3/s 4/s
8/s
Note: In the case of an integrated X-ray generator, the kV, mA, mAs and Focal Spot parameters are
automatically sent to the X-ray generator. See also paragraph 7.5 below.
The values of the mA and mAs parameters set here must match those envisaged by the X-ray
generator scales.
Even if the generator is not integrated, it is important that these parameters are set correctly, as
they are a guide for the operator (who can view them by pressing a symbol in the working
frame).
• Display parameters
Rad /item This section of the Exam Setup menu groups the
parameters that determine how the radiography image
will be displayed.
Settings Notes
View negative Positive/negative display Select to suit exam Radiography images are
of the image normally negatives.
Automatic enlargement
Fill the Size of the collimated images Select to suit exam
to fit the max size
permitted by the monitor
Default image zoom 1.0 - 2.9 1.0 = full screen (43x43
Geometric mag. factor on monitor cm)
Accessed from the general Fluoro card that opens the specific cards for each acquisition mode:
Continuous Fluoroscopy → 43x43 cm field
Continuous Fluoroscopy Z1 → 30x30 cm field
Continuous Fluoroscopy Z2 → 20x20 cm field
Continuous Fluoroscopy Z3 → 15x15 cm field
Only DSA
Only DSA
Fluoro /item
Settings Notes
Continuous Fluoroscopy /item Set the parameters for each programmed patient type
Continuous Fluoroscopy Z1 /item (Size valid).
Continuous Fluoroscopy Z2 /item
Continuous Fluoroscopy Z3 / item
Settings Notes
Accessed from the general Fluoro card that opens the specific cards for each acquisition mode:
Pulsed Fluoroscopy → 43x43 cm field
Pulsed Fluoroscopy Z1 → 30x30 cm field
Pulsed Fluoroscopy Z2 → 20x20 cm field
Pulsed Fluoroscopy Z3 → 15x15 cm field
Pulsed /item
Settings Notes
View negative Positive/negative display Select to suit exam Fluoroscopy images are
of the image normally positives.
To identify the pulsed Default parameter, to suit
Post Processing fluoro image processing P.P. Name exam type.
parameter for that exam
Pulsed Fluoroscopy /item Set the parameters for each programmed patient type
Pulsed Fluoroscopy Z1 /item (Size valid).
Pulsed Fluoroscopy Z2 /item
Settings Notes
You need to set the post-processing parameters for each exam to suit the acquisition mode used by
the exam.
These parameters are normally set during a sample installation and then “transferred” to subsequent
installations without the need to make any changes.
Below is a process flowchart for each mode showing how the image provided by the detector is shown
on the monitor.
The equipment has a number of menus that let you set the image processing parameters.
These menus for each acquisition mode are accessed by technicians (via Technical Password) from
each exam card.
- Load Profile to recall a set of parameters already saved on the machine. The name of this
parameter set is shown in the Exam Setup menu as “Post Processing”.
- Save Profile to save the parameter set with the same name (i.e. “Post Processing”).
Post-processing of radiography images involves a series of operations, applied in order to the images
acquired by the detector.
- linear post-processing (“1” in the diagram): the image acquired by the detector is shown on
the monitor with the original pixel values. This sequence is used for system quality test
images.
- ATH post-processing (“2” in the diagram): the image acquired by the detector is processed
according to the specific anatomical part.
The result is a good quality image that should require no further processing. This sequence
is used for real images.
RAD Image Presentation Algorithm
ROI &
LINEAR (NO ATH) EDGE LINEAR
IMAGE LINEAR IMAGE FILTER LOOK UP
(NONE)
TABLE
RAW
DETECTOR CORRECTED
IMAGE
PRE- LATITUDE
EQUALIZED REDUCTED EDGE ROI &
IMAGE IMAGE FILTER LOOK UP
(LOG) (ATH) TABLE
PU processes detector image and provide useful raw image to HIRIS RF43
Use the Post Processing Setup command to open the Post Processing Setup menu which
contains the following parameters:
Spatial Filter Use to select the spatial filter NONE: no filter Typical setting: SHARPEN
type
SHARPEN: sharpen filter
(edge enhancement)
Kernel Use to select the filter matrix 3x3 Typical setting: 5x5
size (in pixel) 5x5
7x7
9x9
11x11
Pre-equalization Use to select the type of pre- LINEAR Used for images whose
Type equalization. pixel level you want to
keep unchanged.
Typically used to check
image quality.
LOG8000 Use this when acquiring
real images.
Harmonization Use to select the type of Select to suit the anatomical Harmonization is an
Type harmonization to be part. image processing
associated to each exam algorithm that improves
type. The harmonization algorithms image contrast in the case
are set in the factory and of high “latitude”
cannot be altered.
- if the Harmonization
Filter is enabled, set 30000
(typ.)
Auto Equalization is
- If an image with linear normally always enabled.
post-processing, set 10000
(typ).
Stitching ROI Use to select the type of ROI Select one of the ROI types You can change the
Type used by the equipment to shown to suit the stitching existing ROI types or
perform the Auto type. create new ones, as
Equalization function on the described in paragraph
stitched image. 4.7 below.
Stitching Set Use to select the stitching Select one of the stitching The manufacturer defines
Name algorithm for the stitched algorithms. certain types of stitching
image. algorithms.
Post-processing of continuous fluoroscopy and pulsed HCF images involves a series of operations,
applied in order to the images acquired by the detector.
- linear post-processing (“1” in the diagram): the image acquired by the detector is shown on
the monitor with the original pixel values. This sequence is used for system quality test
images.
- dynamic range + recursive filter post-processing (“2” in the diagram): the image acquired by
the detector is processed by applying the following algorithms:
- FTH post-processing (“3” in the diagram): the image acquired by the detector is processed
by applying the following algorithms:
- PRE EQUALIZATION
- RECURSIVE FILTER
- MOTION DETECTION FILTER
- LATITUDE REDUCTION
- AUTO EQUALIZATION
- W & L SHIFTING
This sequence is normally used to acquire real images. Given the heavy demands
made of the CPU, it is only possible in systems with CPU QUAD CORE.
ROI &
LINEAR NO FILTER EDGE LINEAR
IMAGE LINEAR IMAGE FILTER LOOK UP
(NONE)
TABLE
1
Image Processing for image analisys
Image Processing for real image acquisition without FTH filter (for DUAL CORE CPU Systems)
Image Processing for real image acquisition with FTH filter (for QUAD CORE CPU Systems)
3
PRE- LATITUDE
EQUALIZED REDUCTED EDGE ROI &
IMAGE IMAGE FILTER LOOK UP
(LOG) (FTH) TABLE
PU processes detector image and provide useful raw image to HIRIS RF43
Example of post-processing settings for real continuous fluoroscopy images with dynamic range +
recursive filter algorithms (2):
Example of post-processing settings for real continuous fluoroscopy images with FTH algorithm (3):
Example of post-processing settings for real pulsed HCF images with dynamic range + recursive filter
algorithms (2):
Example of post-processing settings for real pulsed HCF images with FTH algorithm (3):
Spatial Filter Use to select the spatial filter NONE: no filter Typical setting: SHARPEN
type
SHARPEN: sharpen filter
(edge enhancement)
Gamma Type Use to select the gamma Select one of the gamma You can change or create
correction curve. curves shown to suit the a new gamma curve, as
anatomical part. described in paragraph
4.5.7 below.
Typical setting:
LINEAR
You can set an FTH filter for each EXAM CARD and each continuous/pulsed fluoroscopy image
acquisition format. To access the FTH filter setup page, select the filter you want to change (click on
the filter name) and then press the FTH FILTER SETUP button.
- load the settings for another FTH FILTER using the Load Profile button;
- save the FTH FILTER settings using the Save Profile command;
- save the FTH FILTER settings with a different name by entering this in the
Save Profile As field and then clicking on this button (active).
IIR Mode Exec IIR Before Uses the temporal recursive filter Forced to:
before any spatial filters Exec IIR Before
IIR Filter Adaptive Yes / No Uses the adaptive temporal filter
(sensitive to movement) Yes
Recognises movement within the image and enhances the edges of the moving object.
The recommended values shown above help reduce the noise linked to the image resolution.
If you require higher levels of resolution (regardless of an increase in noise), alter the two parameters shown
below (within the accepted ranges):
Upper Break Point 32768 End point in the contrast control Forced to:
curve (lsb) 32768
Lower Break Point 0 Start point in the contrast control Forced to:
curve (lsb). 0
MSE PyrLevels = 9 for formats Number of planes used to break Forced, to suit
672x672 (field down the image acquisition field.
20x20) and
960x960 (field
43x43)
= 10 for formats
1024x1024 (field
30x30 and field
15x15)
Border Size 0..480 Size of border (pixels) that reduces 10 pixels for all
the area used to calculate the acquisition formats
histogram
Left Discard Perc. 0.002 Histogram discard on black side Forced to:
(percentage). 0.002
Right Discard Perc. 0.000 Histogram discard on white side Forced to:
(percentage) 0.000
In an ANGIO exam (DSA), non subtracted images are treated using the radiography post-processing
parameters shown in paragraph 4.4.5.1 above.
The post-processing described below is used for subtracted images (i.e. acquired after mask pick-up):
The Post Processing Setup command opens the Post Processing Setup menu which contains the
following commands and parameters:
Spatial Filter Use to select the spatial filter NONE: no filter Typical setting: SHARPEN
type
SHARPEN: sharpen filter
(edge enhancement)
If the PeakEnable function in the EXAM SETUP frame is enabled in FLUORO mode, fluoroscopy
images can be acquired in three different ways:
- normal fluoroscopy,
- max opacity (first stage of ROAD MAPPING),
- image subtraction (second stage of ROAD MAPPING).
Different image post-processing is envisaged for each mode. See paragraph 4.4.5.2 above for the
post-processing of normal fluoroscopy images.
The Post Processing Setup command opens the Post Processing Setup menu with the following
parameters:
Spatial Filter Use to select the spatial filter NONE: no filter Typical setting: SHARPEN
type
SHARPEN: sharpen filter
(edge enhancement)
Level
Auto Equalization Do not enable
ROI Type N.U.
LUT Window Set: 65536
LUT Level Set: 0
Gamma Type Use to select the gamma Select one of the gamma You can change or create
correction curve. curves shown to suit the a new gamma curve, as
anatomical part. described in paragraph
4.5.7 below.
Typical setting:
LINEAR
The Post Processing Setup command opens the Post Processing Setup menu with the following
parameters:
This setup menu lets you set the parameters for automatic calculation of the W and L levels needed to
get the best presentation of the image. Automatic W&L calculation is only active when the Auto
Equalization field has been enabled in the POST-PROCESSING SETUP menu.
These parameters are set by using a reference image on the monitor, which immediately shows the
effects of each new setting.
You can set the parameters for low exposure “Blackside” regions of the
image and high exposure “Whiteside” regions separately.
The setup lets you define the following for each part:
- calculation area (ROI section),
- algorithm to be used (Algorithm section).
● ROI
The ROI should contain the anatomic part needed to create a meaningful histogram for calculating the
W&L values.
You can repeat this at will; each new rectangle deletes the previous
one.
Use shutters if closed the ROI is the area defined by the X-ray collimators if their aperture is
less than 43x43cm
Intersect shutters alternatively, this defines as the ROI the area where the shutters
intersect with the rectangle drawn using the mouse. If the aperture of
the X-ray collimators falls within the drawn ROI, the area will be
calculated using the X-ray collimators
Detect true shutters this option enables an algorithm that recognises the X-ray collimators
within the image using a margin of pixels with respect to the
theoretical position data received from the collimator itself.
Use always blackside ROI the Blackside ROI setting will always be
used, even for the Whiteside region of
the image.
Below are examples of the effect of the different ROI settings on the calculation of the W&L.
• Example 1.
Here two different ROI areas have been used: one for the dark area of the image (BLACKSIDE ROI –
red frame – small frame) and one for the lighter area (WHITESIDE ROI – pink frame – large frame).
The figure below shows the histogram obtained using these two ROI areas; this will then be used for
the automatic W&L calculation.
• Example 2.
Two different ROI areas have been used again, but this time the “Use shutter if closed” option has
been enabled for the BLACKSIDE ROI. In this case, if the aperture of the X-ray collimators is less than
43x43cm, the BLACKSIDE ROI will correspond to the actual shutter size (yellow rectangle).
• Example 3.
Once again, two different ROI areas have been used, with the “Use shutter if closed” option enabled
for both the BLACKSIDE ROI and the WHITESIDE ROI. In this case, if the aperture of the X-ray
collimators is less than 43x43cm, the actual shutter size will be used for both ROI areas.
In this case, the histograms for both the BLACKSIDE and the WHITESIDE ROI coincide (see
histograms below).
• Example 4.
In this case, two different ROI areas have again been used, with the “Use always Blackside ROI”
option enabled. This means that only one ROI will actually be used.
The type of algorithm applied to the image inside the ROI lets you define the L and W levels for the
best display of the image.
The “L” stands for the lowest pixel value in the histogram. The “W” stands for the difference between
the highest pixel value and the lowest pixel value in the histogram.
● Fixed Window:
This option lets you specify the max W value and its position within the histogram.
If the histogram of the original image has a lower W value, its value remains unchanged.
● Algorithm: this lets you choose the type of algorithm to be used (and
relative parameter settings) in order to define the L and W.
Min/Max:
This algorithm detects the min and max grey values in the image and then assigns them:
- min value = L,
- max value – L = W.
You can correct the detected L and W values using the Parameter 0 field.
Select Blackside (min histogram value shown on the monitor):
- and correct the L value (range: –50% to + 50%) by setting Parameter 0 (–0.5 to +0.5),
Histogram truncation:
Like the previous option, this algorithm detected the min and max grey values in the image and then
assigns them:
- min value = L,
- max value – L =W.
You can correct the detected L and W values using the Parameter 0 fields. You can shrink the part
of the histogram shown by eliminating the top and tail of the histogram:
Select Blackside:
- increase the min value ( L) by setting Parameter 0 (0 to 0.5).
For example, if Parameter 0 = 0.1, the L value is increased until 10% of all the pixels in
the image are eliminated from the “dark” part of the histogram.
Select Whiteside:
- decrease the max value ( L+W) by setting Parameter 0 (0 to 0.5).
For example, if Parameter 0 = 0.1, the L+W value is decreased until 10% of all the
pixels in the image are eliminated from the “light” part of the histogram.
Notes:
- Show Histogram on AutoLut if enabled, this lets you show the ROI histograms graphically,
together with the W and L values obtained by using the
selected algorithms.
Auto LUT this applies the set ROI, L and W parameters to the image.
Track this function is very useful when you want to find the automatic values of Parameter
0 of the algorithms shown correctly on the monitor.
Save to save the setup parameters as a (*.Rendition.prm) file in the relevant directory so
that these can be used when setting up the exam cards.
The grey levels in an image automatically shown on the monitor are mainly defined by the ATH
algorithm used and the Auto LUT function (see paragraph 4.5.6.2).
The “LOOKUP MANAGER” lets you further change the grey levels of an image within the window (W)
using LUT curves (gamma curves) that you can build at will to get the best contrast conditions.
The new curve can then be saved in a file with the *.LutCurve.prm extension in directory
Settings\LutCurves\ . This file can be associated to the exam card during installation.
Use a typical image for the anatomical part of interest to build the gamma curve.
This image should be acquired under normal projection, X-ray dose and patient type conditions.
Linear LUT curve. To create a LUT curve where the To create a LUT curve where the
building points are linked using the building points are linked using the
The linear curve cannot be altered. “Cubic Natural Spline” function. “Cubic Deriv0 Spline” function.
• Build the curve by dragging the building points with the mouse or by moving these points by
selecting a point and then entering its X and Y values and confirming with Set.
selected point
• Once built, the new curve should be saved (with the same name or a new name) as follows:
- right-click on the mouse to open the working menu and then select the Save command,
- the LUT curve file opens: enter the name you want to give to the new curve and then save by
clicking on Save:
o if the curve is saved with the same name, all the exam cards that use this file will be
altered as a result,
o if the curve is saved with a new name, it will appear in the Gamma Curve list and so can
be used when setting up the exam cards (see paragraph below).
Correction curves must be associated to the type of exam for which they have been built.
This is done by using the exam card (see paragraph 4.4 above and paragraph 7.11 below) where the
Gamma type parameter (in the Image Processing Type menu, accessed by authorised installers
only) lets you choose between the saved curve types.
You can change a gamma curve or create a new one by starting from a n existing curve.
To do this:
4.6 DICOM
After making the electrical connections to the DICOM network (see point 3.3.2.1 above), the DICOM
setup involves the following steps:
The menus in the main frame are mainly used for the above operations:
- Option / Setup / Dicom / Setup, for DICOM connection setup, or
- Option / Setup / Dicom / Verify, to check the connection,
DICOM address
(point 4.6.2)
Store
(par 4.6.7)
Dose SR Settings
(point 4.6.16)
- Dicom / Show Spooler, used to manage the transmission of images to the network and
to set up the remote devices.
ATS supplies all its systems with default local network settings. It’s up to the company’s network
manager to alter these to suit the characteristics of the network that will act as host to the system.
Only the Administrator can access this procedure: see Annex 7.3 below.
Note: there is a third wireless connection – called Wireless Network – if the system is fitted with two
detectors
- with wireless DRX1 detector see paragraph 2.3.1 in Part 6 of this manual
- with wireless PIXIUM 3543EZ detector see paragraph 2.3.1 in Part 7 of this manual
After entering and confirming the network settings, you can proceed with the connection test.
- Enter the PING command followed by the IP Address (provided by the network administrator) of
another device in the network at the DOS prompt, e.g.:
C:\Documents and Settings\Admin2> PING 192.168.50.107
- If there’s a TIME OUT error, check both the network cable (see note below) and the Windows
XP network settings (IP address, subnet mask, any active firewalls, etc.).
- We recommend sending other PING commands to other network devices to check that the
entire network is read properly by the system.
Below are details of the typical wiring for cross-connections and otherwise, should you need to check
the network cable connections for some reason:
Name NIC1 NIC2 Name Name Pin Cable Colour Pin Name
TX+ 1 --- 3 RX+ TX+ 1 White/Orange 1 TX+
TX- 2 --- 6 RX- TX- 2 Orange 2 TX-
RX+ 3 --- 1 TX+ RX+ 3 White/Green 3 RX+
RX- 6 --- 2 TX- 4 Green 4
5 White/Blue 5
RX- 6 Blue 6 RX-
7 White/Brown 7
8 Brown 8
In order to identify the system within the DICOM network, you need to address the system using the
first few fields in the Option / Setup / Dicom / Setup menu (NB: the menu only opens once you have
entered the technical password):
• TCP Port Server TCP port code used by the system. Typically: 104.
These two parameters are established in collaboration with the DICOM network administrator.
• General Manufacturer Name: this lets you set the DICOM field for the manufacturer’s name.
The devices are defined using the DicomSpooler application and the Local (local devices) and Rem
(remote device) cards in the Network menu.
- Select Network in the DICOM Spooler menu (you need to enter the Password).
- The DICOM Setup menu opens, letting you define the devices connected to the system in cards:
• Local (DICOM setting for HIRIS RF43)
• Rem 1, 2, … (DICOM setting for any remote devices connected to the HIRIS RF43)
Confirm settings
- The system is normally supplied with the settings for 5 remote devices:
- PACS, Worklist, Printer, Storage Commitment, Query / Retrieve
These provide a setup example. To add (or remove) a device, use the following options:
- Add Remote: This lets you add a “remote” device to the network and create the
corresponding card.
- Remove Remote: This lets you delete a “remote” device from the network by removing its card.
- Verify: This lets you check the connections between the system and the network
station for the “remote” card in question.
Note. The connection check via the Dicom Spooler menu involves using the
specific name in the Local AE Title for quick remote verification (see
below); normal connection checks (see para. 4.6.4 below) are made
using the name specified in the Local AE Title field in the Option /
Setup / Dicom / Setup menu of the . It is therefore crucial that you use
the same Local AE Title for both fields.
- You must define the parameters using the corresponding REM card for each device.
- Each remote device connected to the HIRIS RF43 system is identified and defined by the
parameters in the Rem cards:
- Port: TCP port of the application that will respond on the remote DICOM device.
- Dose SR: DICOM DOSE STRUCTURED REPORT: DICOM service that lets you collect
information on the X-ray dose.
- Print Defaults: This opens the Print Defaults Setup menu that lets you set the default printer
settings:
Image
General Processing
Film
- Set the default devices for each DICOM mode (Storage, Print, Worklist, MPPS, Storage
Commitment) in the Local card.
Options available in the Local card:
- UL Protocol frame: We recommend you do not change the data already entered here
(normally those shown in the figure).
Use Option / Setup / Dicom / Verify in the system main frame to open the AE Title Selection box
with a list of the DICOM devices in point 4.6.3 above:
Select the device you want to check and then click on OK. This way you start testing the device
connections: the test* indicates whether the device is present and active in the DICOM network.
* Note: This check only concerns the connection test. You may therefore find that the check for a
DICOM device (e.g. a printer) is positive, but there are problems when you send an image to
this. In this case, we recommend that you check the Printer Defaults Setup settings (see
paragraph 4.6.3 above).
The DicomSpooler manages the transmission of the images over the DICOM network via a memory
(on HD) where all the images to be sent are queued.
- Select Setup / Options and enter the technical password; the Spooler Settings menu opens:
This menu lets you set the transmission queue management using the parameters in 4 different fields:
Path The path you want for the data transmission queue.
Typically “C:\ProgramFiles\Digitec\Dicom\DicomSpooler\SpoolDir\”
Size The size in MB of the transmission queue (size of the Spooler).
Typically 10000 MB, corresponding to roughly 600 radiography
images in high resolution (2400 images in medium resolution)
After, this specifies the time interval between the last transmission attempt and the next.
Typically 180 seconds.
Note: After 3 failed attempts, the transmission is considered “failed”.
• Failed items, to specify the fate of the exams that for some reason have caused the transmission
to fail:
Keep till manually deleted Keep all the exams in the transmission queue until you delete
these manually.
Keep for XX days Keep all the exams in the transmission queue for the set
number of days and then delete them automatically.
• Warning setup, to select and manage the alarm messages that will be shown in the Dicom
Spooler frame.
Warn if spooldir level greater Warning that the transmission queue filling level has been exceeded.
than xx % Typical value: 80%.
Warn if queue disabled Warning that transmission queue filling has been disabled.
Warn if there is not enough Warning that there is not enough space left in the transmission
spooldir space queue.
Warn if error in item insertion Warning that there are errors when adding of the image to the
transmission queue
Warn if item processing fails Warning that image transmission has failed.
Warn if there are in list items Warning that images have remained in the spooler for more than xx
older than xx days days.
Warn if there are in list more Warning that there are more than xx images in the spooler.
than xx items
Warn if there are in list more Warning that the spooler contains more than xx images whose
than xx failed items transmission has failed.
Check warning status at To enable the alarm status control on starting the system application
startup up: if any alarms exist, the Dicom Spooler frame opens with the
relevant warning(s).
Show spooler on warning To enable the display of the Dicom Spooler frame every time an
activation alarm is enabled (recommended).
Note: The DICOM Spooler menu also lets the operator manage images in the transmission queue.
See point 4.6.11 below.
4.6.6 WORKLIST
In the Setup / Dicom / Setup menu, the Worklist Settings command opens the Dicom Worklist
Setup menu, used to define these options:
- The “Modality Filter” option lets you apply an import filter on the Worklist:
- The Merge modality requests option lets you make a single DICOM di GET LIST query that
incorporates the various different filters selected using the Modality Filter option.
If not selected, you will need to send a query for each of the selected filter options.
- The “On Get List …” option sorts the list of exams received by the date in the Date-Time field
(newest to oldest).
If not selected, the order shown will be that of receipt.
- The “Clear list before “Get List” execution” option lets you delete the list before effecting
the get list command.
If not selected, the new list is added to the existing list.
- The Create studies on “Get List” option lets you create any new studies received from the
WORKLIST automatically. The parameters to differentiate a new study can be set using the panel:
Study Matching conditions: using the panel, set the DICOM parameters to be used to recognise
the “new study”. A study is considered “new” when at least one of these parameters differs from the
studies already in the list.
The PATIENT ID and ACCESSION NUMBER are normally used to do this.
- The Exec “Get List” on open dialog option lets you generate the DICOM “GET LIST” request
automatically:
- if a default WORKLIST SERVER has already been set, the WORKLIST command will be
sent directly to this SERVER,
- if there is no default WORKLIST SERVER, a window appears, letting you:
- The Send Empty AE Title request option lets you send the WORKLIST query with an empty
DICOM tag (0040,0001) “Scheduled Station AE Title” (no set value); otherwise the value will be set
using the LOCAL AE TITLE set in the system.
- The Send empty seq. Req. With 0 length option lets you send an empty WORKLIST query
sequence (length = 0).
If not selected, an empty sequence of indefinite length (0xffffffff) is sent with the delimiter: “end of
sequence delimeter”.
● Study Create – Identifiers to define the configuration of the DICOM MPPS TAGs on the basis of
the DICOM WORKLIST TAGS
- Enable copy from “Study ID” -If the contents of the Scheduled
Procedure Step ID field is empty
during transmission of the data to the
PACS and MPPS server, the contents
of the Study ID field received from the
Worklist will be copied and inserted in
the Performed Procedure Step ID
field.
Performed Procedure Step Description To copy the contents of the
- Enable copy from “Scheduled Procedure Step Description” Scheduled Procedure Step
Description field received from the
Worklist and insert this in the
Performed Procedure Type
Description field during transmission
of the data to the PACS and MPPS
server.
Build description from procedure text This lets you indicate the contents of the DICOM procedure
field in the Description of the exam you want to create.
Build description from step text This lets you indicate the contents of the DICOM step field in
the Description of the exam.
Insert Pr. and St. in description This lets you indicate the two prefixes Pr. and St. (before the
contents of the Procedure field and Step field respectively) in
the Description of the exam.
Lock Patient and Study Data If active, this function makes it impossible to change the data for
the patients/study received from the Worklist. Any created
manually can, of course, be edited.
Create all patient’s studies If the list contains more than 1 exam for the same patient, this
lets you create all the studies with a single command (the
Create exam command will then create a list with all the studies
involving that patient)
Clear item on create study This lets you remove patients from the Worklist as new studies
are created for these.
If this option is not selected, the created study will, in any case,
be checked (see first exam in following figure).
The Study Data menu with the Lock Patient and Study Data option enabled looks like this:
The Study Data menu with the Lock Patient and Study Data option disabled looks like this:
4.6.7 STORE
• Send series/image Select this to transfer the DICOM “series/image number” parameter”.
• Presentation LUT (DX) Only for devices with DICOM code “DX“: select this to get "linear" transfer
of the image, followed by the LUT correction parameters.
• Lut expansion 0-30% To expand the size of the LUT sent to the workstation (0 - 30%) towards
black (B) or white (W): you can thus include in the image external parts
not presently shown on the system monitor.
• If Patient ID is empty transfer options if the “Patient Data” ID field has not been filled in:
Send nothing the exam is sent if the ID field is left empty.
Use patient name the exam will be sent on entering the name of the patient in the ID field.
Use “PATIENT ID” the exam will be sent on entering the word “PATIENT ID” in the ID field.
• If Study ID is empty Transfer options if the “Study Data” ID field has not been filled in:
Send nothing the exam is sent if the ID field is left empty.
Use study UID the exam will be sent on entering the "UID" in the ID field.
Use “STUDY ID” the exam will be sent on entering the word “STUDY ID” in the ID field.
• Auto Store Settings to access the Setup menu for the Auto Store function:
The Auto Store Setup menu lets you define the options and activation modes for the Auto Store
function.
• Automatic mode
Enable just one of the three options in the Image Selection panel. Do not enable the
“Always show Exit without store and Cancel” option.
This way, the images are automatically sent in the mode selected on the Image
Selection panel when you quit the image.
• Manual mode
In this mode, you are asked to select the relevant option before images can be sent.
Transmission only starts after selecting:
- more than one option on the Image Selection panel,
- or the “Always show Exit without store” and “Cancel” option,
The following options are available for these two Auto Store modes:
- Always execute Auto Store on close study: to enable the Auto Store function on closing a
study that contains studies that have not been sent before, even if no new images have been
acquired.
- Send to default store server: the images will be sent directly to the default STORE SERVER
device.
If this option is not selected, the image selection panel appears in the list of STORE devices
when you close the study.
- Max number of frames sent automatically: use this to set the max number of frames that can
be sent without generating an operator warning.
• Graphics on image: Check if you want the DICOM images in the CD archive to contain all the
graphics currently on the image.
• Presentation LUT (DX): only for “DX” mode: select to transfer the "raw" image and, separately,
the correction parameters (LUT). Normally not selected.
• CR/DX Bits/px: To select the number of bits/pixel in the image being transferred to CD
(accepted range: 8 - 16 bits/pixel); typically 16. Valid for DICOM CR/DX
transmission mode.
• XA/RF Bits/px: To select the number of bits/pixel in the image being transferred to CD
(accepted range: 8 - 16 bits/pixel); typically 16. Valid for DICOM XA/RF
transmission mode.
- Single DataSet: Select if you want all the DICOM images to be saved in a single archive.
- Double DataSet: Select if you want the DICOM images to be saved separately, in 2
archives (CR/DX and XA/RF).
4.6.9 PRINT
• Use calibration in true Select this to get a printout of the images in true size mode with corrected
measurements using the Gauge tool in the working frame.
• Start FE in true size Select this to open the Film Editor menu and automatically select the
True size print function.
• Fast Print Settings To open the Fast Print Setup frame which lets you associate the printer
and format for each of the four fast print symbols in the working frame.
It also lets you select the True size option for each printer.
• Single Image Settings To enable the print layout setup frame for the indications to be inserted
on the film.
The menu (see figure below) defines the text, its position and the size of
the text on the film.
Size and position of text
Size and position of fields
General data to be printed Selection and enabling of fields
Hospital/Dept. Text (normally the name of the hospital / ward) to be entered in the
Hospital field.
st
Model Model of the equipment: enter in the 1 line of the Device field.
rd
Manufacturer Name of the manufacturer: enter in the 3 line of the Device field.
st
Serial nu. Serial number of the equipment: enter in the 1 line of the Device field.
- Set the position and size of all the text on the film. The text is split into groups defined using
Select field:
groups
Hospital Hospital name as entered in this menu (if the box has been left empty in
this menu, the name entered as the Hospital Name in the General
Settings menu will be used)
1 line: AreaDose
st
Exposition
nd
2 line: kV / mA / ms (or kV / mAs)
Exposition no dose kV / mA / ms (or kV / mAs)
Note: These two fields are alternatives (you cannot enable both at the same time).
st
Elaboration 1 line: Calibration Factor (K=n.nnn)
nd
2 line: L / W / E / H, i.e.: Level (L=nnnnn)
Window (W=nnnnn)
Edge (E=n-Nxx) *
Harmonization (H=n)
Patient name Name and surname of the patient
Patient data Date of birth / sex
Patient ID Identification code for patient
Physician Name of technician / doctor / physician responsible for exam
Exam date Date and hour of exposure (format: dd-mm-yyyy hh:mm)
st
Device 1 line: Model / Serial n°.
nd
2 line: SW rel.
rd
3 line: Manufacturer.
True size Indication Text/symbol shown in True size Indication
Stitching Indication Text/symbol shown in Stitching Indication
Accession Number Accession Number
Free Text 1 String for free comment
Free Text 2 String for free comment
Free Text 3 String for free comment
Free Text Quick This lets you set a few set strings in the Free Text 1, Free Text 2 and Free
Selection Text 3 fields:
Ser: n/m : n number of the run (series) to which the image belongs, m
total number of runs in the study
Ser: n/m Img: n1/m1 or Frm: n1/m1 : n number of the run (series), m
total number of runs in the study, n1 number of the image (Img:) in
the case of a singleframe run or frame number (Frm:) in the case of a
* Note: The table below shows the meaning of the Edge field indications:
E=n-Nxx
(Edge filter)
n N xx
(kernel) (weight) (filter)
indication meaning indication meaning indication meaning
3 3x3 0 0 SH sharp
5 5x5 1 1 SM smooth
7 7x7 2 2
9 9x9 3 3
11 11x11 4 4
5 5
Dimension V (mm)
External to image
Minimize Height
Minimize width
Offset H (mm)
Interline (mm)
Offset V (mm)
Border (mm)
Enable
Font
Box
HOSPITAL Yes Top Left Yes Yes Yes 200 10 0 0 Left Top Arial 0 0 6
EXPOSITION No
EXPOSITION NO DOSE Yes Bottom Right Yes Yes Yes 60 6 0 6 Right Center Arial 0 0 4
ELABORATION Yes Bottom Right Yes Yes Yes 60 10 0 10 Right Center Arial 0,5 0 2/4
Patient NAME Yes Bottom Center Yes Yes Yes 200 12 0 0 Center Center Arial 0 0 10
Patient DATE Yes Bottom Center Yes Yes Yes 50 6 0 10 Center Center Arial 0 0 5
Patient ID Yes Bottom Center Yes Yes Yes 50 6 0 15 Center Center Arial 0 0 6
EXAM DATE Yes Bottom Right Yes Yes Yes 50 8 0 0 Right Center Arial 0 0 6
DEVICE Yes Bottom Left Yes Yes Yes 70 15 0 0 Left Bottom Arial 0 0 4/4/4
The MPPS Settings command in the Dicom Setup menu opens the MPPS Setup menu used to set
the relevant options:
General Settings:
• Enable MPPS On New Studies: check to enable the MPPS mode whenever a new
study is created.
• Dicom Direct Print is Dicom Transfer: check for the MPPS to treat the Print Dicom
command for an image as a Store Dicom
command.
• On Open Study if MPPS already Exists: check to enable the sending of the MPPS N-
CREATE message when an exam is opened that
has already been associated to an MPPS during a
previous session.
• Before Image Dicom Transfer: check to enable the sending of the MPPS N-
CREATE or N-SET message immediately before a
DICOM transfer to STORE, PRINT or the CD/DVD
burner.
• After Image Dicom Transfer: check to enable the sending of the MPPS N-
CREATE or N-SET message immediately after a
DICOM transfer to STORE. NB: a transfer to
Dicom PRINT or CD/DVD burner is never followed
by MPPS messages to the server.
• On Close Study (N-Create + N_Set): After closing a study, this option sends the MPPS
N-CREATE message (if not already sent). This is a
general option: the details of the MPPS function
upon closing an exam are set out below in the On
Close Study N-SET Notification panel.
• Auto if New Acquired Images: the MPPS N-SET message is only sent
automatically if ne images have been acquired.
• Auto if Dicom Transferred Images: the MPPS N-SET message is only sent
automatically if images have been sent to Dicom
STORE.
• Auto if MPPS already Created on Server: the MPPS N-SET message is only sent
automatically if the MPPS N-CREATE message
has already been sent to the server.
• Ask Operator if No Auto Conditions: asks you for confirmation before sending the N-
SET message to the server.
MPPS Contents:
• Performed Procedure Step ID (If Empty, enable copy from Study ID):
check to use the Study ID field to enter values for
the Performed Procedure Step ID field, if this has
been received empty from the Worklist and no
values have been entered manually.
The Storage Commitment Settings command in the Dicom Setup menu opens the Dicom Storage
Commitment Setup menu used to set the relevant options:
General:
• Enable Storage Commitment on new Studies: check to enable the Storage Commitment
function for each new study. If disabled, the
images for an exam with Storage
Commitment are labelled with the “auto-
committed” marker.
Spooler Notifications:
• Spooler Notifications Expiration Time (1-7 days): max time for a notification (days).
• Spooler Notifications Queue Max Length (10-200): max time for the spooler notification
queue.
• SCT N-EVENT_REPORT Wait T.O. (3-60 sec): timeout for N-EVENT_REPORT response
after sending the N-ACTION message.
• SCT Retry Delay (5-300 sec): pause between one SC attempt and the next
(min time between 2 N-ACTION
transmissions)
Image deleting:
• Non Stored Images: establishes action when you use the delete
command for images not already sent to the
spooler.
• Store Pending / Stg Cmt Pending Images: establishes action when you use the delete
command for images already sent to the
spooler with an SCT (Storage Commitment
Transaction) pending.
The Storage Commitment request can be sent to the same Store periphery or another one.
For this reason, you can associate a dedicated periphery for Storage Commitment to each Store
server during the Spooler remote periphery setup.
If the STORAGE COMMITMENT device is the same that performs the STORE function:
If the STORAGE COMMITMENT device is not the same that performs the STORE function:
In this case, the configured STORAGE COMMITMENT device must be set in the Setup menu for
remote DICOM devices and the St.Cmt flag must be enabled.
The Q/R Settings command in the Dicom Setup menu opens the Query/Retrieve Setup menu used
to set the relevant options:
Calling AE Title: to specify the AE Title used by the system to make the Query / Retrieve (suggested
to use different AE Title than the Local AE Title specified in the DICOM SETUP frame).
Default Called AE Title: to specify the AE Title of the remote device offering the Query / Retrieve
service.
Move Store Server Port: the TCP port where the Query / Retrieve device sends its response.
(different from number 104 and suggested number 107)
Saved File-set Character set: to set the type of font used when saving received files.
Base Retrieve Folder: to specify the directory on the system hard disk containing the exams received
from the remote Query / Retrieve device (typically C:\Retrieve)
Viewer Command Line: to specify the path needed to run the DICOM viewer for exams received from
the Query / Retrieve server.
Typically C:\ProgramFiles\Digitec\DgViewer\Bin\DgViewer.exe
Retrieve Dose SR: if checked, the parameters for the DICOM DOSE STRUCTURED REPORT option
are requested (if the system uses this option).
View Study On Retrieve: if checked, the exam is automatically shown on the monitor after receipt,
using the specified DICOM viewer.
Close Q/R On View Study: if checked, the Query / Retrieve window closes when the image is viewed.
IHE (Integrating the Healthcare Environment) setup lets you integrate the system with various
linked DICOM components.
This function can only be used if the Worklist, MPPS, Store and Storage Commitment DICOM
functions are present.
Various rules set out in official documents (IHE Radiology Technical Framework) and the
manufacturer’s rules (IHE Integration Statement) apply when integrating the system.
SWF:
Depending on these options, certain dialog windows will be different for the operator.
If the SWF profile is enabled, some data cannot be edited, as they are fixed by the IHE standard: e.g.
the Code Sequence access keys only let you view this and not edit it.
The Study Data dialog window with the SWF profile enabled looks like this:
The Study Data dialog window with the SWF profile disabled looks like this:
The DicomSpooler application lets you manage the images remaining in the transmission queue.
Operations Enable/Disable Status
The Status panel to view the status of the images and transmission queue at any time. This
shows:
Use the panel commands to manage the images in the transmission queue:
Enable/Disable to enable/disable the filling of the queue and/or transmission of the images
by using the following options:
Operations
2) Use the Enable/Disable buttons to invert the status and so enable the Operations panel commands:
Status Queue: Disabled
Status Spool: Disabled
3) Operations:
- Delete: to delete the images from the queue,
- Reschedule: to manage a new transmission request (see figure):
If there are problems with DICOM communication, you can call up a LOG file that records all such
events.
This makes troubleshooting easier (especially if also sent to ATS).
Note: The LOG file is always updated until you remove the “Activatelog.txt” file from the above
directory. To avoid creating excessively big LOG files, we recommend disabling the LOG
function after testing.
DICOM Dose SR is a DICOM service that lets you collect X-ray dose data.
This menu lets you set the general parameters for the various items in the radiological system:
- Dose Area Meter,
- Imaging device (Recording Device),
- Irradiation Device.
The menu is opened with the archive closed. Select the Dose SR Settings option in the Dicom Setup
menu:
Dev. Phys. Location Physical position of the Set by the user (e.g. Radiology
device 2)
If the manufacturer of the radiology system has its own registered code, the UID must be set by the
manufacturer itself.
If not, you can use the ATS codes and set the UID as follows:
1.3.76.6.2. a. b. 1. 1. sn
parameter description
Code used for the manufacturer of the Irradiation Device.
a This is set for each manufacturer by ATS.
Available upon request.
Code used for the Irradiation Device.
b This is set for each manufacturer’s product by ATS.
Available upon request..
Serial number of the equipment. This can consist of a different
sn number of alphanumerical characters.
The DICOM Dose SR service collects and transmits the dose parameters for each X-ray emission.
Right-click over the image frame to view a table containing these parameters:
Study data
Report Root Data
Notes
Procedure Reported Pre-set data: Projection X-Ray
Exposure Index Target Set value (target) for the Exposure Index.
Obtained from the dose set for the exam (Dose
µGy x 100/µ Gy)
Value EI value of the exposure
(see paragraph 4.9 below )
Deviation DI value of the exposure
(see paragraph 4.9 below)
Acquired Image SOP Class UID UID for the image automatically assigned by the
device
Acquired Image SOP Instance UID UID for the image automatically assigned by the
device
Irradiation Type - Stationary Acquisition for Radiography
- Fluoroscopy for continuous or pulsed fluoro
Fluoro Type - Continuous for continuous fluoro
- Pulsed for pulsed fluoro
Pulses Number Number of radio or pulsed fluoro exposures saved
to HD for that event
Dose and Source Data DAP (Gy*m²) Exposure data
Dose (RP) (Gy)
KVP (kV)
Exp Time (ms)
Exposure (µAs)
Irradiating Device Irr. Dev. UID Irradiating device parameters
Irr. Dev. Name (see paragraph 4.6.16.1)
Irr. Dev. Manufacturer
Irr. Dev. Model Name
Irr. Dev. Serial Num.
X-ray Filters Type Parameters of the X-ray beam filter (if fitted) on
Material the X-ray collimator and used during acquisition of
Min. Thickness the selected event (see paragraph 4.2.6).
Max. Thickness
The geometry of the monitor needs to be checked on a regular basis, together with its
brightness/contrast control.
There is an image frame for this that can be recalled at any time from the patient list as follows:
Options / Diagnostic Functions / Pattern Test Exam. A test pattern appears as shown below.
B A
Using test pattern, adjust the brightness and contrast to view the entire grey scale (the brightness % is
shown in each square). More specifically:
- you should be able to see the grey square in the white box (positions A),
- you should be able to see the grey square in the black box (positions B).
The entire adjustment procedure is explained in point 4.6.1 below, together with the controls required
should the monitor be replaced.
Control dial
3) Select:
- Analog VGA if the monitor is connected via the
VGA connector, or
- Analog BNC if the monitor is connected via the
BNC connector.
5) Select:
Gamma and set “Dicom”
Once the DICOM curve has been set you can no longer
adjust the brightness and contrast, as these are defined
by the DICOM curve.
Use the 5 buttons shown in the figure here to adjust the monitor settings.
Settings
Monitor status
ON/OFF
• Press the MENU button to open the OSD setup menu (see figure below):
Use the buttons described above to select the INPUT menu and then the VGA1 option.
Use the buttons described above to select the PICTURE menu, highlight the GAMMA field and then
select DICOM9300°K.
You can also use the setup menu in a simpler manner to adjust the monitor BRIGHTNESS and
CONTRAST only.
• Without entering the general setup menu, press one of the two buttons shown below as
required:
BRIGHTNESS
CONTRAST
• then use the “+” and “-“ buttons to increase or decrease the selected value.
The procedure (procedure / step) sent by the RIS via the Worklist function is a code that defines the
individual exams (X-ray projections) needed for the patient study.
The system interprets this code and guides the operator through the study, automatically providing the
set working parameters for the related exams.
The procedure may require 1 or more exams (typically not more than 4-5).
The system searches the following DICOM fields that may contain the code, in order of priority:
RIS systems do not, at present, define the procedures with a harmonized code and so the acquisition
system needs to be adapted to suit the code used during installation, as explained below.
The PROCEDURE SETUP card lets you correlate the procedure codes used by the hospital’s RIS
with the exams (projections) programmed in the machine.
• In order to use the Procedures, you must check the Enable Acquisition Auto Procedure option.
If you do not, the system will ignore the function and so you need to select each exam (projection)
manually, when performing the study, from among those possible with the equipment.
• Click on Add Proc to add a new procedure code, which will then appear in the ProcedureStep
table, called “UNKNOWN”.
• You can also import these codes via a text file including the list of the single Procedure codes
(linked to 2 descriptions separated by using the TAB key, as shown in the example below: the
Procedure Names.txt text file. These codes will then be shown in the main window:
Note: The RIS system administrator is responsible for providing this file.
• Click on Create From File to open the Procedure list File window and then select the path of the
text file containing the list of procedures you want to import:
• After selecting the text file by clicking on Open, the names of the individual procedures are added
to the Procedure Step list and their description appears on the right:
• If you want to delete a code from the list, select it and then click on Remove Proc.
• Click on Insert Proc to add a new code in any position in the list.
• Select either the RF STATION or WD STATION option in the Add Exam Code window to view the
exams foreseen by the relevant equipment: PIXIUM RF4343 detector or WIRELESS detector
respectively (shown in the anatomical technique or single projection boxes). See figure below:
• Once you have made your choice, the exam is shown in the Exam protocol field:
• Repeat, using the Add Exam Code option to add other exams to the procedure.
The exams in a procedure are normally printed on a single film (using the Film Editor composition
tool). You can, therefore, program the composition parameters when constructing the procedure.
• Select Setup to view to the Film Editor PRESET menu and then select the printer, the format, the
orientation and how the film is divided:
Some settings will be shown in the Print Format field of the selected procedure.
Note: In both these cases, acquisitions in fluoroscopy mode (continuous or pulsed) are always
enabled and do not affect the way the procedure progresses.
The two types of projection can be identified by setting the Single Exposure option in the exam card,
as also explained in paragraph 4.4.4 of Part 2.
The DICOM Dose SR service lets you check for any discrepancies between the set dose and the
actual dose of radiography images. It does this by showing (in accordance with standard EN 62494-1):
• The dose set for the exam (Dose Target = x.x µGy).
• The DEVIATION INDEX (DI) indicates how much the detected EXPOSURE INDEX differs from
that set for the exam: DI = 10 x LOG10 (EI / EI target).
The table below gives an example of EI/DI deviation after different exposures (Dose Target set at
5µGy):
The detector conversion factor needs to be set for each type of acquisition in order to calculate the EI.
This is done in the factory. ATS adopts the exposure conditions indicated in standard
EN62494-1.
The resulting conversion factors are set using the logical mode Sensibility parameter:
- RAD 3i/s and
- RAD 8i/s
The Sensibility value for logical mode RAD 3i/s is also set in the two tomography Logical Mode
parameters:
- TOMO 4s
- TOMO 2s.
5.1 INTRODUCTION
The SD4343 software lets you define and calibrate the acquisition modes (logical modes) foreseen by
the system.
This software acts mainly as the user/detector interface (front end + PU processor), by sending the
calibration commands and displaying the corresponding images. The actual calibration is carried out
by the detector itself once it receives these commands.
In order to calibrate the detector, you must first connect the system video processor (PID) and the
detector (PU).
Run the SD4343 software and then:
Use the TAB keys at the bottom to access the following menus from this frame:
- Configuration (see paragraph 5.3 below)
- Logical Mode Configuration (see paragraph 5.3 below)
- Logical Mode Manager (see paragraph 5.4 below)
- Calibration (see paragraph 5.5 below).
1) Select the “Configuration” TAB in the right-hand menu and check that the parameters settings
correspond to those shown below. If not, you will need to reset these manually.
Before calibrating the acquisition modes, we recommend you check the configuration of each logical
mode:
4) The following window appears showing the values of calibration parameters for that Logical Mode:
5) Check that the parameters for each Logical Mode correspond to those in the table below; if they do
not correspond, you must restore the default conditions indicated below.
Logical Mode
Physical Offset
acquisition Field Frequency X-ray windows
ID
mode (mHz) (ms)
Continuous Fluoroscopy
1 3 18000 1
43x43
Continuous Fluoroscopy Z1
2 7 15000 15
30x30
Continuous Fluoroscopy Z2
3 15 30000 1
20x20
Continuous Fluoroscopy Z3
4 27 15000 10
15x15
Pulsed Fluoroscopy
5 2 15000 15
43x43
Pulsed Fluoroscopy Z1
6 7 15000 15
30x30
Pulsed Fluoroscopy Z2
7 22 15000 15
20x20
Pulsed Fluoroscopy Z3
8 27 15000 10
15x15
Offset 0: Offset 0:
Radiography 3i/s 1000 500
10 17
43x43 Offset 1: Offset 1:
3000 150
Offset 0: Offset 0:
Radiography 8i/s 1000 500
11 4
43x43 Offset 1: Offset 1:
8000 60
Tomo 4s
15 17 200 4300
43x43
Tomo 2s
16 17 400 2300
43x43
During installation of the equipment and before calibrating the detector, you need to carry out the
detector working temperature preset procedure.
This procedure involves two settings in the detector temperature control system:
1) Preliminary setting, where the speed at which the fans rotate is constant until the thermal
stability of the detector is reached (i.e. after at least 3.5 hours of use).
Using the SD4343 application, access the Configuration panel and then set the parameters
below:
2) Final setting, where the automatic fan speed control is set to keep the detector temperature
constant at the thermal stability value.
The final setting must be set when the detector has reached its thermal stability (i.e. after at
least 3.5 hours of use).
1) Find the temperature reached by the detector, shown in the Pixium Manager 1:1 page;
2° activate the logical mode you want to calibrate. This is done in two steps: first you need to load
the mode and then activate it.
To do this:
These three stages must be done in order, as each requires that the previous one has been
completed.
We recommend calibrating all the modes at low dose before calibrating them at high dose.
Radiography mode:
• Offset/gain calibration and defective pixel correction - Mode 10 (43 x 43cm)
• Offset/gain calibration and defective pixel correction - Mode 11 (43 x 43cm).
Tomography mode:
• Offset/gain calibration and defective pixel correction - Mode 15 (43 x 43cm)
• Offset/gain calibration and defective pixel correction - Mode 16 (43 x 43cm).
The paragraphs below describe the calibration procedure, making a distinction between the
fluoroscopy modes (continuous/pulsed) and radiography/tomography modes.
Note: See paragraph 5.5 above for details on how to activate the logical mode you want to calibrate.
OFFSET calibration involves the acquisition of 15 images (12 of which must be valid) by the detector
without X-ray emission.
Continuous and pulsed fluoroscopy modes foresee only one type of offset calibration for each mode.
The table below details the modes to be calibrated and the relevant parameters:
Logical Mode
ID Field Offset
Frequency X-ray Windows
(mHz) (ms)
Continuous Fluoroscopy
1 18000 1
43x43
Continuous Fluoroscopy Z1
2 15000 15
30x30
Continuous Fluoroscopy Z2
3 30000 1
20x20
Continuous Floroscopy Z3
4 15000 10
15x15
Pulsed Fluoroscopy
5 15000 15
43x43
Pulsed Fluoroscopy Z1
6 15000 15
30x30
Pulsed Fluoroscopy Z2
7 15000 15
20x20
Pulsed Floroscopy Z3
8 15000 10
15x15
Calibration procedure:
2) Double-click on the logical mode you want to calibrate (this must be the same as the “active” mode;
logical mode 1 in the figure below). Select Offset and then right-click on Offset calibration.
- click on Exit.
5.6.2 FINDING THE EXPOSURE VALUES FOR GAIN AND DEFECTIVE MAP CALIBRATION
GAIN and DEFECTIVE MAP calibration requires that you know the X-ray parameters in order to set
the detector at the dose level required during calibration.
Use a special function in the SD4343 application to find these parameters (X-ray filter and X-ray
generator exposure values):
5) Go to LastFramePixiumManager1.
7) Access Properties.
8) Select High Limit value and set a value twice that of the foreseen acquisition levels. You can
also use the “slider”: select and drag the High Limit value to the value you want.
10) Release the X-ray command and deactivate the image GRAB using the relevant symbol.
11) Access Statistics and move the cursor over the image to check the Mean value of the acquired
image.
12) Adjust the exposure parameters as necessary and repeat the above procedure until you get the
required dose/image levels (see tables in paragraphs 5.6.3, 4, 6 and 7).
Make a note of the X-ray parameter values (X-ray filter and generator exposure).
The next step after OFFSET calibration is GAIN calibration of the current mode.
The dose must fall within the accepted range, otherwise calibration will be aborted, i.e.:
- the set dose must be sufficient to generate the detector image levels within the accepted
range shown in the tables below.
Therefore, you must know the exposure parameters beforehand: check these before starting
calibration by acquiring images with homogeneous filtering and read the resulting grey level (LSB) (as
described in para. 5.6.2 above).
2) Double-click on the logical mode you want to calibrate (this must be the same as the “active”
mode). Select Linear gain and then right-click on Linear gain calibration.
3) Click on Start.
5) To ensure correct GAIN calibration, the aperture of the X-ray collimator must be higher than that
of the input field for the selected logical mode, thus avoiding the risk of the collimator creating
shadow:
6) Press the X-ray emission command and then wait at least 1 second
before pressing OK.
The X-ray command must be held down through-out the procedure
(several seconds).
8) Press OK to stop.
After GAIN calibration, the logical mode is automatically deactivated.
9) Gain calibration is successful unless a message appears, warning you that the X-ray dose was
too high or too low or there was an object in the X-ray beam (collimator shutter or something else.
In which case, calibration is not valid and so you need to reactive the Logical mode by pressing
the “Logical mode manager” tab again and repeat the load/activate/Offset calibration and Gain
calibration operations.
10) The logical mode is now correct (Offset and Gain). We recommend you check the acquired image
to make sure that there are no calibration errors. Re-activate the Logical mode and acquire
several images. Assess these using the SD4343 function (see paragraph 5.6.2 above).
This procedure is similar to gain calibration, but 4 series of acquisitions are required:
- homogeneous X-ray filtering,
- 4 different doses set by the system.
The X-ray data need to be set for each dose in order to get image levels falling within the accepted
range (see tables below).
As with Gain calibration, you must know the exposure parameters beforehand: check these before
starting calibration by acquiring images with homogeneous filtering and read the resulting grey level
(LSB) (see paragraph 5.6.2 above).
Logical mode 1
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Continuous Fluoroscopy 43x43
603-1006
Dose 0 50 56kV, 1.40mA
(typ. 800)
1207-2012
Dose 1 100 61kV, 1.68mA
(typ. 1600)
9660-16100
Dose 2 800 85kV, 3.03mA
(typ. 12800)
14490-24150
Dose 3 1200 92kV, 3.42mA
(typ. 19200)
Logical mode 2
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Continuous Fluoroscopy 30x30
802-1337
Dose 0 100 64kV, 2.04mA eq.*
(typ. 1070)
1605-2675
Dose 1 200 70kV, 2.40mA eq.*
(typ. 2140)
9630-16050
Dose 2 1200 95kV, 3.90mA eq.*
(typ. 12840)
13642-22737
Dose 3 1700 101kV, 4.26mA eq.*
(typ. 18190)
* Note: mA eq. = mean mA value
Logical mode 3
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Continuous Fluoroscopy 20x20
802-1337
Dose 0 100 65kV, 1.90mA
(typ. 1070)
1605-2675
Dose 1 200 73kV, 2.35mA
(typ. 2140)
9630-16050
Dose 2 1200 101kV, 3.93mA
(typ. 12840)
13642-22737
Dose 3 1700 110kV, 4.43mA
(typ. 18190)
Logical mode 4
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Continuous Fluoroscopy 15x15
802÷1337
Dose 0 100 64kV, 2.04mA eq.*
(typ. 1070)
1125÷1995
Dose 1 150 67kV, 2.2mA eq.*
(typ. 1500)
9630÷16050
Dose 2 1200 95kV, 3.90mA eq.*
(typ. 12840)
13642÷22737
Dose 3 1700 101kV, 4.26mA eq.*
(typ. 18190)
Logical mode 5
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Pulsed Fluoroscopy 43x43
603-1006
Dose 0 50 46kV, 60mA
(typ. 800)
1207-2012
Dose 1 100 49kV, 60mA
(typ. 1600)
9660-16100
Dose 2 800 67kV, 60mA
(typ. 12800)
14490-24150
Dose 3 1200 73kV, 60mA
(typ. 19200)
Logical mode 6
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Pulsed Fluoroscopy 30x30
802-1337
Dose 0 100 49kV, 60mA
(typ. 1070)
1605-2675
Dose 1 200 53kV, 60mA
(typ. 2140)
9630-16050
Dose 2 1200 73kV, 60mA
(typ. 12840)
13642-22737
Dose 3 1700 78kV, 60mA
(typ. 18190)
Logical mode 7
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Pulsed Fluoroscopy 20x20
802-1337
Dose 0 100 49kV, 60mA
(typ. 1070)
1605-2675
Dose 1 200 53kV, 60mA
(typ. 2140)
9630-16050
Dose 2 1200 73kV, 60mA
(typ. 12840)
13642-22737
Dose 3 1700 78kV, 60mA
(typ. 18190)
Logical mode 8
Image level range Default X-ray parameters
nGy /frame
(LSB) Filter: 10mm Al - SID 120cm
Pulsed Fluoroscopy 15x15
802÷1337
Dose 0 100 49kV, 60mA
(typ. 1070)
1125÷1995
Dose 1 150 51kV, 60mA
(typ. 1500)
9630÷16050
Dose 2 1200 73kV, 60mA
(typ. 12840)
13642÷22737
Dose 3 1700 78kV, 60mA
(typ. 18190)
To calibrate the DEFECTIVE PIXEL WITH X-RAYS of the active logical mode:
1) Select Calibration.
3) Click on Next.
4) Click on Start.
5) The system warns you that any X-ray
emission is about to be stopped
(confirm with OK) and then tells you
the dose required for calibration
(confirm with OK).
6) To ensure correct Detective Map calibration, the aperture of the X-ray collimator must be higher
than that of the input field for the selected logical mode, thus avoiding the risk of the collimator
creating a shadow:
- max aperture, greater than 43x43cm, for logical modes 1 and 5
- at least 35cm x 35cm for logical modes 2 and 6
- at least 25cm x 25cm for logical modes 3 and 7
- at least 20cm x 20cm for logical modes 4 and 8.
7) Press the X-ray emission command and then wait at least 1 second
before pressing OK: acquisition/emission starts. The X-ray command
must be held down through-out the procedure (several seconds).
11) Defective Map calibration is successful unless a message appears, warning you that the X-ray
dose was too high or too low or there was an object in the X-ray beam (collimator shutter or
something else. In which case, calibration is not valid and so you need to reactive the Logical
mode by pressing the “Logical mode manager” tab again and repeat the load/activate/Offset
calibration and Defective Map calibration operations.
12) The logical mode is now correct (Offset, Gain and Defective Map). We recommend you check the
acquired image to make sure that there are no calibration errors. Re-activate the Logical mode
and acquire several images. Assess these using the SD4343 function (see paragraph 5.6.2
above).
1) To activate Blinking Pixel calibration: click on Next in the Defect map calibration frame: the
Blinking defect map table calibration frame appears.
4) Blinking Pixel calibration is successful unless a message appears, warning you that the X-ray dose
was too high or too low or there was an object in the X-ray beam (collimator shutter or something
else. In which case, calibration is not valid and so you need to reactive the Logical mode by
pressing the “Logical mode manager” tab again and repeat the load/activate/Offset calibration and
Blinking Pixel calibration operations.
This form of calibration involves acquiring a series of images without X-ray emission.
1) Click on Build defect map in the CALIBRATION frame to activate the calibration.
3) Click on Start.
Calibration takes a few seconds.
Calibration of the logical mode is now complete, unless some defects need to be corrected manually.
We recommend you check the acquired image to make sure that there are no calibration errors. Re-
activate the Logical mode and acquire several images. Assess these using the SD4343 function (see
paragraph 5.6.2 above).
OFFSET calibration always requires that the detector acquires 15 images (12 of which must be valid)
without X-ray emission.
• RADIOGRAPHY mode uses two different types of offset correction, which the system
automatically applies to suit the acquisition rate:
o Offset 0 one-shot and 1 i/s
The table below shows the two offset parameters for the two RADIOGRAPHY Logical Modes:
Logical Mode
Offset
ID Field Frequency X-ray Windows
(mHz) (ms)
2880x2880 Offset 10-0 1000 500
10 pixel
43x43cm Offset 10-1 3000 150
The calibration of the two Offset 0 values is sufficient to calibrate the detector.
Logical Mode
Offset
ID Field Frequency X-ray Windows
(mHz) (ms)
Tomo 4 s
15 200 4300
2880x2880 pixel
43x43cm
Tomo 2 s
16 400 2300
2880x2880 pixel
43x43cm
Procedure:
1) Select Calibration.
The next step after OFFSET calibration is GAIN calibration of the current mode, always using Offset 0
for radiography Logical Modes.
The dose must fall within the accepted range, otherwise calibration will be aborted, i.e.:
- the set dose must be sufficient to generate the detector image levels within the accepted
range shown in the tables below.
Therefore, you must know the exposure parameters beforehand: check these before starting
calibration by acquiring images with homogeneous filtering and read the resulting grey level (LSB) (as
described in paragraph 5.6.2 above).
Tomo 4 s 10875÷18125
15 25000 70kV, 160mA, 100ms
43x43 (typ. 14500)
Tomo 2 s 10875÷18125
16 25000 70kV, 160mA, 100ms
43x43 (typ. 14500)
1) Select Calibration.
3) Click on Start.
5) To ensure correct GAIN calibration, the aperture of the X-ray collimator must be higher than that
of the acquisition field (i.e. greater than 43x43cm), thus avoiding the risk of the collimator creating
a shadow.
Set the X-ray generator with the exposure parameters needed for the dose required during
calibration.
Note: - the set radiography time for all Logical Modes must be less than 450 ms,
- A series of 12 exposures now follows: make sure that the energy required for this series of
exposures is compatible with the fitted X-ray tube.
6) Press PREP + RAD, wait at least 1 second and then press OK. The X-
ray commands must be held down throughout the procedure.
7) The X-ray emission sequence automatically stops. The system tells you
so with the warning “Stop XRay Generator”.
9) Gain calibration is successful unless a message appears, warning you that the X-ray dose was
too high or too low or there was an object in the X-ray beam (collimator shutter or something
else). In which case, calibration is not valid and so you need to reactive the Logical mode by
pressing the “Logical mode manager” tab again and repeat the load/activate/Offset calibration
and Gain calibration operations.
10) The logical mode is now correct (Offset and Gain). We recommend you check the acquired image
to make sure that there are no calibration errors. Re-activate the Logical mode and acquire
several images. Assess these using the SD4343 function (see paragraph 5.6.2 above).
Having calibrated the GAIN, you can then calibrate the DEFECTIVE MAP.
This procedure is similar to gain calibration, but 4 series of acquisitions (each with 12 emissions) are
required:
- homogeneous X-ray filtering,
- 4 different doses set by the system,
The X-ray data need to be set for each dose in order to get image levels falling within the accepted
range (see tables below).
As with Gain calibration, you must know the exposure parameters beforehand: check these before
starting calibration by acquiring images with homogeneous filtering and read the resulting grey level
(LSB) (as described in paragraph 5.6.2 above).
Logical mode 11
Image level range Default X-ray parameters
Rad 8i/s nGy /frame
(LSB) Filter: 21mm Al - SID 120cm
43x43
1350-2250
Dose 0 1500 70kV, 25mA, 40ms
(typ. 1800)
2700-4500
Dose 1 3000 70kV, 50mA, 40ms
(typ. 3600)
21600-36000
Dose 2 24000 70kV, 100mA, 160ms
(typ. 28800)
27900-46500
Dose 3 31000 70kV, 100mA, 200ms
(typ. 37200)
TOMOLogical Modes
Logical mode 15
Image level range Default X-ray parameters
Tomo 4 s nGy /frame
(LSB) Filter: 21mm Al - SID 120cm
43x43
1087-1812
Dose 0 2500 70kV, 32mA, 63ms
(typ. 1450)
2175-3625
Dose 1 5000 70kV, 50mA, 80ms
(typ. 2900)
17400-29000
Dose 2 40000 70kV, 160mA, 200ms
(typ. 23200)
21750-36250
Dose 3 50000 70kV, 160mA, 250ms
(typ. 29000)
Logical mode 16
Image level range Default X-ray parameters
Tomo 2 s nGy /frame
(LSB) Filter: 21mm Al - SID 120cm
43x43
1087-1812
Dose 0 2500 70kV, 32mA, 63ms
(typ. 1450)
2175-3625
Dose 1 5000 70kV, 50mA, 80ms
(typ. 2900)
17400-29000
Dose 2 40000 70kV, 160mA, 200ms
(typ. 23200)
21750-36250
Dose 3 50000 70kV, 160mA, 250ms
(typ. 29000)
The procedure for radiography logical mode DEFECTIVE MAP calibration is:
1) Select Calibration.
3) Click on Next.
4) Click on Start.
6) To ensure correct calibration, the aperture of the X-ray collimator must be higher than that of the
acquisition field (i.e. greater than 43x43cm), thus avoiding the risk of the collimator creating a
shadow.
Place a homogeneous filter on the X-ray tube and set the X-ray generator with the exposure
parameters needed for the dose required during calibration.
Note: - the set radiography time must be less than 450 ms for all logical modes,
- a series of 12 exposures now follows: make sure that the energy required for this series of
exposures is compatible with the fitted X-ray tube.
7) Press PREP + RAD, wait at least 1 second and then press OK.
The X-ray commands must be held down throughout the procedure.
8) The X-ray emission sequence automatically stops. The system tells you
so with the warning “Stop XRay Generator”
9) Now repeat the procedure (steps 6, 7 and 8) for the other 3 series of
image acquisitions.
If an alarm appears, e.g. the X-ray dose was too high or too low or there was an
object in the X-ray beam (collimator shutter or something else), then the calibration is not
valid and so you need to reactive the Logical mode by pressing the “Logical mode manager” tab
again and repeat the load/activate/Offset calibration and Defective Map calibration operations.
1) To activate Blinking Pixel calibration: click on Next in the Defect map calibration frame.
4) Blinking Map calibration is successful unless a message appears, warning you that the X-ray
dose was too high or too low or there was an object in the X-ray beam (collimator shutter or
something else). In which case, calibration is not valid and so you need to reactive the Logical
mode by pressing the “Logical mode manager” tab again and repeat the load/activate/Offset
calibration and Blinking Pixel calibration operations.
This form of calibration involves acquiring a series of images without X-ray emission.
1) Click on Build defect map in the CALIBRATION frame to activate the calibration.
3) Click on Start.
Calibration takes a few seconds.
Calibration of the logical mode is now complete, unless some defects need to be corrected manually.
We recommend you check the acquired image to make sure that there are no calibration errors. Re-
activate the Logical mode and acquire several images. Assess these using the SD4343 function (see
paragraph 5.6.2 above).
If the DEFECTIVE MAP calibration procedure has failed to intercept defective pixels, you can identify
these manually, as follows.
6) Go to the defective pixels and zoom into the image using the mouse trackball until you see the
PIXEL grid.
10) Load the Logical Mode again and activate it. Calibrate its offset and then make a new image
acquisition and check that the defective pixels have been corrected.
Note: To compensate for a full row of defective pixels, you need to mark at least 5% of its pixels:
- continuous/pulsed fluoroscopy: mark at least 25 pixels
- radiography: mark at least 75 pixels.
After calibrating the detector, the Defective map image shows the position of the defective pixels on
the surface of the detector.
In order to assess the quality of the detector, the SD4343 application also provides the ECV
parameter (Effective Cluster Value) for the Defective map, where each defective pixel is associated to
a given Defect class, depending on the presence of other defective pixels surrounding it.
Each defect class is given a weight. The sum of the weights of all these classes produces the ECV.
Note: If the ECV exceeds this accepted value, contact the ATS technical service.
The detector calibration is saved directly to the detector (see paragraph 5.9.2).
We recommend making a back-up copy in the system memory so that you can then reload it on the
detector if necessary (see paragraph 2.8 in Part 5).
Use the Whole Configuration option to transfer the entire detector calibration parameters: all the files
will be saved to the path indicated in the Destination Folder field. This is typically:
The detector saves the calibration data as soon as you switch the system off.
We strongly advise that you do this immediately using the Hardware Reset function described below:
1) Service Maintenance Reset and turn off. Then use the Hardware reset function.
2) The SD4343 application shuts down. Then use the Windows Restart function.
3) Once the system has restarted, the software system automatically runs and the new
calibration parameters are automatically loaded.
During calibration of the detector, you may need to save the acquired images in RAW format in order
to export them and so view them on an external PC.
To save an image in RAW format, simply stop any active acquisition and then take the mouse pointer
to the image frame, right-click on the mouse and then select the Export raw image function:
Depending on the acquisition Logical Mode, the formats used to save RAW images are shown in the
table below. These are specific for use of the IMAGE J application:
Acquisition modality Number type Horizontal Vertical Offset from start of file Little Endian White is 0
(LM Mode) size size
43x43 fluoroscopy 16bit Unsigned 960 960 256 Yes No
(LM 1) int
30x30 fluoroscopy 16bit Unsigned 1024 1024 256 Yes No
(LM 2) int
20x20 fluoroscopy 16bit Unsigned 672 672 256 Yes No
(LM 3) int
15x15 fluoroscopy 16bit Unsigned 1024 1024 256 Yes No
(LM 4) int
43x43 pulsed 16bit Unsigned 960 960 256 Yes No
fluoroscopy (LM 5) int
30x30 pulsed 16bit Unsigned 1024 1024 256 Yes No
fluoroscopy (LM 6) int
20x20 pulsed 16bit Unsigned 672 672 256 Yes No
fluoroscopy (LM 7) int
15x15 pulsed 16bit Unsigned 1024 1024 256 Yes No
fluoroscopy (LM 8) int
43x43 HR radiography 16bit Unsigned 2880 2880 256 Yes Yes
(LM 10) int
43x43 MR radiography 16bit Unsigned 1440 1440 256 Yes Yes
(LM 11) int
43x43 tomography
(LM 15,LM 16) 16bit Unsigned 2880 2880 256 Yes Yes
int
The procedures for saving data are explained in the following paragraphs:
- Updating the back-up hard disk (Part 5, para. 5.7)
- Creating a back-up CD/DVD (ghost disk) (Part 5, para. 5.8)
The procedures for restoring data are explained in the following paragraphs:
- Replacing the system hard disk with the back-up hard disk (Part 5, para. 2.1)
- Restoring the video processor system hard disk (Part 5, para. 2.2)
Note: The system comes with a back-up hard disk and a back-up CD/DVD (exact copies of the
system hard disk upon delivery ) and a CD containing the application that lets you make the
back-up CD/DVD.
7 APPENDICES
Should there be a blackout long enough to cause the Personal Computer (PID) to shut down, there's a
tiresome system reboot procedure that's relatively long (about 10 minutes) before the operating
system and the applications are back up. This procedure also ensures complete acquisition of the
detector offset.
Moreover, even though the Windows operating system foresees the possibility of sudden switching-
off, we recommend avoiding this where at all possible for the above reason.
To avoid this problem, the system can be fed via a continuity unit (UPS), which feeds the main
Acquisition System supply line.
The function of the UPS is to cover for any brief cuts in the power supply or at least for the time
needed for the operator to carry out the shutdown procedure required by Windows 2000.
A 1000 VA continuity unit is normally capable of ensuring the power supply for the system for about
15 minutes.
The feasibility of adding a continuity unit (not provided by ATS) should be considered on the basis of
the following:
1) The UPS must be installed outside the acquisition system cabinet and its on/off button must always
be easily accessible.
3) The UPS supply must not come from the main circuit breaker in the room, otherwise whenever the
room power supply is switched off, this will be recognised as a drop in voltage.
The UPS should always be fed via an appropriate backup and circuit-breaking circuit.
4) If the system power supply is cut for a prolonged period (normally the UPS indicates this via an
intermittent buzzer), you should quickly carry out the WINDOWS 2000 "SHUTDOWN" procedure.
After which, you need to switch off the acquisition system cabinet (switch on cabinet) and then
switch off the UPS using its ON/OFF button. The UPS then needs to be switched back on once the
power returns.
5) At the end of each day, the system is normally switched off after the usual SHUTDOWN
procedure, by using the switch on the cabinet.
The UPS stays ON.
ON/OFF
PIXIUM RF4343
PIXIUM RF4343
The user and technical manuals are available on-line in PDF format:
- see the explanation below for details of how to open and use the manuals,
- see paragraph 2.6 (PART 5). for details of how to install the manual upgrades.
- This command opens the Acrobat Reader SW and displays the user manual with its list of contents.
- This command opens the relevant window containing the Technical Manual command.
- Use this command and enter the “technical” password to open the Acrobat Reader SW and display
the manual with its list of contents:
Note: The list of contents has links to each paragraph to speed up the use of the manuals.
You need to access the system as the network administrator in order to set the system up.
To do this, during operation (system already running):
5) Press OK.
Note: - To return to normal operator mode, repeat the above steps, this time entering:
- User name: operator
- Password: operator
01-05 SKULL
06-10 THORAX
11-15 SPINE
16-20 ABDOMEN
21-25 PELVIS
26-30 G.I.
31-35 UPPER EXTREMITY
36-40 LOWER EXTREMITY
41-45 STITCHING
46÷50 URO
51 ANGIO (DSA)
56 TOMO
57-98 TBD
99 TEST
In theory 99 x 99 cards may exist (99 body zones and 99 projections for each body zone).
These cards are available for the installer, who can then enable/disable them (using a password) in
order to create the list of projections available for the operator.
You can also create a new exam by starting from an existing exam and then use the Clone function
(see para. 4.4.2); you can then change the parameters and name in the cloned exam, while keeping
the same image process (“POST PROCESSING SET UP”).
If, on the other hand, you need to create a new exam with a different image processing algorithm, you
will need to construct a new EXAM CARD with a new “POST PROCESSING SET UP”: only ATS and
the authorised system integrator can do this.
The tables below list the sample exams with their main parameters programmed by ATS.
There are special exam cards for tomography exams, clearly indicating the angle, speed and X-ray
emission time. See table below:
• Continuous fluoroscopy,
• Tomography
The paragraphs below describe the characteristics of these X-ray emission modes to ensure correct
integration with the acquisition system.
7.5.1.1.1 Radiography
• The generator has to allow for two X-ray emission modes to suit the request for the selected exam:
AEC mode: where the exposure time is determined by the system via the
READY_ACQ_RAD signal,
NO AEC mode: the generator starts X-ray emission on receipt of the READY_ACQ_RAD
signal, but the actual exposure time is determined by parameters mAs or mA-s
set using the generator console.
• In AEC mode, the generator lets you choose the focus and set/view the kV and mA.
• In NO AEC mode (manual mode), on the other hand, as well as choosing the focus, the generator
also lets you set/view the kV, 3 and 2 point technique, mAs (if you chose the 2 point technique) or
mA – s (if you chose the 3 point technique).
• The X-ray generator back-up time needs to be set, taking into account:
- The duration of a single exposure or a series of 1i/s exposures must not be more than 500
ms
- In high resolution:
o the duration of a series of 1.5 - 3 i/s exposures must not be more than 150 ms.
- In medium resolution:
o the duration of a series of 1.5 - 3 i/s exposures must not be more than 250 ms.
o the duration of a series of 2 - 4 - 8 i/s exposures must not be more than 60 ms.
While the backup time with the Wireless detector may be set to suit the length of exposure, this must
never, in any case, be more than 1 second.
See also point 7.5.2.2 “STAGES IN IMAGE ACQUISITION IN RADIOGRAPHY MODE” for more
details.
The system envisages the same X-ray generator signals for the acquisition request in both continuous
and pulsed fluoroscopy modes. The operator selects which mode to use from the system monitor.
On receiving this signal, the system generates the READY_ACQ_FL output, used by the X-ray
generator to enable X-ray emission.
The choice between manual and automatic dose control is made by the operator directly from the X-
ray generator console.
The operator selects the mode via the monitor by selecting an exam programmed for Max Op / Road
Mapping continuous fluoroscopy.
When the operator selects the high contrast pulsed fluoroscopy (pulsed HCF) acquisition mode via the
system, the X-ray generator is informed of this via the activation of the SEL_HCF output by the system
itself; the X-ray generator then shows the right parameters on the console.
The acquisition request in pulsed HCF fluoroscopy is sent to the system using the same REQ_FL/HCF
signal used for a request in continuous fluoroscopy mode.
As a result, the system sends the READY_ACQ_RAD/HCF output to the X-ray generator, i.e. a series
of pulses (15 ms) to time X-ray emission: the X-ray generator then has to emit pulses of a fixed length
(typically 10 ms) to suit this timing.
As for continuous fluoroscopy, the X-ray generator should allow for manual and automatic X-ray
dose control.
Automatic dose control means that the kV/mA parameters are automatically adjusted to suit the
correction signal provided by the system.
See also point 7.5.2.4 “STAGES IN IMAGE ACQUISITION IN PULSED HCF FLUOROSCOPY MODE”
for more details.
The detector allows for image acquisition in continuous and HCF pulsed fluoroscopy using fields
43x43 cm and 20x20 cm (zoom2); only pulsed fluoroscopy is used for field 30x30 cm (zoom1) and
field 15x15 cm (zoom3).
The PULSED FLUOROSCOPY mode is therefore envisaged in order to simulate (for the operator)
continuous fluoroscopy, but with pulsed X-ray emission, at the max rate for field 30x30 (15 frame/s)
and field 15x15 (15 frame/s).
The system requests this mode from the X-ray generator, via an RS232 serial message or a logic
signal (SEL PULSED) whenever you decide to acquie images in continuous fluoroscopy mode using
fields 30x30 cm and 15x15 cm.
See also point 7.5.2.5 “STAGES IN IMAGE ACQUISITION IN PULSED FLUOROSCOPY MODE” for
more details.
7.5.1.1.6 Tomography
This mode is set when choosing the system exam. The stand and the X-ray generator are
informed of this setting by a series of serial messages (via CAN bus and/or RS232).
• When you choose tomography mode, the X-ray generator also received the NO AEC mode setting
signal.
The generator back-up time must be set to reflect the max acquisition time (4.2s).
See also point 7.5.2.6 “STAGES IN IMAGE ACQUISITION IN TOMOGRAPHY MODE” for fuller details
of how this function works.
This is the min level of integration, basically allowing for synchronisation of the X-ray exposure and
acquisition of the image by the system.
Integration is by means of logic signals.
The setting of the radiological parameters on the generator continues to be manual, as foreseen by
the X-ray generator.
The logic interface signals are found on connectors CM2 and CM1 of the system's main controller.
31 COM_REQ Safety signal for an acquisition request. This signal must be supplied with:
REQ_FL/HCF for continuous or pulsed fluoroscopy, or
PREP_RAD for preparation for radiography acquisition.
34 ZOOM 1 (30X30) Zoom 1 selection (30x30 cm)
Outputs
29 +V ACC +24V (provided by the accessory / generator)
3 SEL_INJ Signal active if the exam is programmed for automatic injector start
PREP_RAD 30
PREP_RAD REQ_RAD 12
COM_REQ 31
RAD
ZOOM
ZOOM 1 (30x30) 34
ZOOM 2 (20x20) 16
SEL_FLAT_PANEL 1 17
SEL_FLAT_PANEL 2 13
SEL_FLAT_PANEL_1_CPI 14
SEL_FLAT_PANEL_2_CPI 33
0V1 ACC. 15
0V ACC. 10
+V ACC. 29
+24V
SEL_INJ. 3
SEL_INJ.
SYNC_INJ. 21
SYNC_INJ.
23
CM11
SYNC_INJ_1 1
SYNC_INJ_1
SYNC_INJ_1 2
SYNC_INJ_1
ACCESSORY/X-RAY GENERATOR
INTERFACE
Outputs
7, 20 +24V GEN. +24 V (supplied by the accessory / X-ray generator)
1 SEL_PULSED_HCF Signal sent by the system to the generator to select the pulsed
HCF fluoroscopy.
14 PREP_PULSED Signal sent by the system to the generator to request preparation
for pulsed HCF fluoroscopy.
2 SEL_ PULSED Signal sent by the system to the generator to select the pulsed
fluoroscopy .
17 READY_ACQ_RAD/HCF Signal for acquisition in pulsed fluoroscopy/ pulsed HCF
fluoroscopy / radiography in progress:
The generator must emit X-rays in these modes while this signal
is active.
5 READY_ACQ_FL Signal for acquisition in continuous fluoroscopy in progress:
The generator must emit X-rays in this mode while this signal is
active.
18 ABC_LOW ABC signal too low in fluoroscopy / pulsed fluoroscopy/ pulsed
HCF:
logic signal to generator to adjust X-ray dose.
6 ABC_HIGH ABC signal too high in fluoroscopy / pulsed fluoroscopy/ pulsed
HCF:
logic signal to generator to adjust X-ray dose.
analogue outputs
12 ABC_ERROR Dose error in fluoroscopy / pulsed fluoroscopy / pulsed HCF:
Analogue output signal to generator to adjust X-ray dose.
13 0V Reference 0 V for analogue output.
HIRIS RF43
CM8
+24V
+24V GEN. 7-20
PREP_PULSED 14
PREP.HCF
SEL_PULSED_HCF 1
SEL_PULSED_HCF
SEL_PULSED 2
SEL_PULSED
READY_AQ_RAD/HCF 17
READY_AQ_RAD/HCF
READY_ACQ_FL 5
READY_AQ_FL.
ABC_LOW 18
ABS_LOW
ABC_HIGH 6
ABS_HIGH
ABC_ERROR 12
ABS_ERROR
0V 13
+24V
GEN_READY 23
EXON 21
0V GEN. 19
CM4
124E
CAN L =2
CAN H =7
0V =3
CAN BUS
TERMINATION
ACCESSORY/X-RAY GENERATOR
INTERFACE
In order to give the image acquisition command, you need to send a preparation signal to the system,
followed by an acquisition request signal.
When acquisition starts, the system activates the READY_ACQ_RAD/HCF output to the X-ray
generator, which should then start emitting X-rays. During X-ray emission, the X-ray generator must
return the active X-ray information via the EXON signal.
Important: The X-ray generator must cut X-ray emission as soon as the preparation/radiography
commands are released or in the event of a generator alarm, regardless of the actual
status of the READY_ACQ_RAD/HCF signal.
The generator back-up time must be set to reflect the max exposure time:
• With PIXIUM RF 4343 detector:
- for single exposures and a series of 1i/s exposures: 500ms
The X-ray generator is informed of the set acquisition rate via a series of serial messages.
NO AEC
t1 = 1 s (max)
t2 = 500 ms max
t3 = 250 / 150 / 60 ms max
AEC
t1 = 1 s (max)
t2 = 500 ms max
t3 = 250 / 150 / 60 ms max
The system tells the X-ray generator that acquisition is in progress via output READY_ACQ_FL.
The X-ray generator only starts X-ray emission in fluoroscopy mode when it receives this signal. It
must stop X-ray emission when:
- there is no READY_ACQ_FL signal,
- as soon as the fluoroscopy footswitch is released or there is an X-ray generator or accessory
alarm, regardless of the READY_ACQ_FL signal.
During acquisition, the X-ray generator receives a series of signals depending on the image levels
acquired by the detector in order to correct the X-ray dose automatically:
- the analogical output ABC_ERROR
- the logic outputs ABC_LOW and ABC_HIGH (when the X-ray dose is correct, both are
deactivated).
The analogical output ABC_ERROR varies from +10 V to -10 V, with X-ray dose corrected at 0V.
During installation you can associate the polarity of the voltage as “high X-ray dose” or “low X-ray
dose t” (see ”General setup” in Part 4 of this manual).
When there is no acquisition, analogical output ABS_ERROR goes to 0 V and the logic outputs for
control are deactivated.
t1 = 500 ms (typ)
t2 = 100 ms (typ)
The X-ray generator is told that this mode has been selected via output SEL_PULSED_HCF.
The pulsed HCF fluoroscopy acquisition request is sent to the system by the REQ_FL/HCF and
COM_REQ signals from the accessory.
The system then sends the X-ray generator the pulsed fluoroscopy preparation command via output
PREP.PULSED.
The X-ray generator must confirm within 3 seconds that preparation is completed by activating input
GEN.READY.
After checking that the GEN.READY signal has been received, the system informs the X-ray
generator that acquisition is in progress using the READY_ACQ_RAD/HCF input (fixed duration of 15
ms for all zoom factors except for zoom factor 15x15, which lasts 10ms, at rates of between 1 and 15
pulses per second, depending on the value you set).
The X-ray generator only starts X-ray emission in fluoroscopy mode when it receives this signal. It
must stop X-ray emission when:
- there is no READY_ACQ_RAD/HCF signal,
- as soon as the fluoroscopy footswitch is released or there is an X-ray generator or accessory
alarm, regardless of the READY_ACQ_RAD/HCF signal.
The X-ray generator must return the EXON signal to the system (X-ray emission).
To get automatic correction of the X-ray dose, the X-ray generator receives the analogical output
ABC_ERROR and the logic outputs ABC_LOW and ABC_HIGH.
As with continuous fluoroscopy, the analogical output varies from +10 V to -10 V, with X-ray dose
corrected at 0 V.
t1 = 3 s (max)
t2 = 150 ms (typ)
t3 = 15 ms for zoom 30x30 (10ms for zoom 15x15)
t4 = 66 + 1000 (depending of selected rate)
t5 = 100 ms (typ)
The X-ray generator is told that this mode has been selected via output SEL_PULSED.
The pulsed HCF fluoroscopy acquisition request is sent to the system by the REQ_FL/HCF and
COM_REQ signals.
The system then sends the X-ray generator the pulsed fluoroscopy preparation command via output
PREP.PULSED
The X-ray generator must confirm within 3 seconds that preparation is completed by activating input
GEN.READY.
After checking that the GEN.READY signal has been received, the system informs the X-ray
generator that acquisition is in progress using the READY_ACQ_RAD/HCF input (fixed duration of 15
ms for zoom factor 30x30 and 10ms for zoom factor 15x15, at a fixed rate of 15 pulses per second).
The X-ray generator only starts X-ray emission in fluoroscopy mode when it receives this signal. It
must stop X-ray emission when:
- there is no READY_ACQ_RAD/HCF signal,
- as soon as the fluoroscopy footswitch is released or there is an X-ray generator or accessory
alarm, regardless of the READY_ACQ_RAD/HCF signal.
To get automatic correction of the X-ray dose, the X-ray generator receives the analogical output
ABC_ERROR and the logic outputs ABC_LOW and ABC_HIGH.
As with continuous fluoroscopy, the analogical output varies from +10 V to -10 V, with X-ray dose
corrected at 0 V.
When there is no acquisition, analogical output ABC_ERROR goes to 0 V and the logic outputs for
control are deactivated.
REQ_FL/HCF
CM6-11
COM_REQ
CM6-31
GEN. READY CM8-23
PREP_PULSED
CM8-14
SEL_PULSED
CM8-2
READY_ACQ_RAD/HCF
CM8-17
ABC_ERROR
CM8-12
ABC_LOW
CM8-18
ABC_HIGH
CM8-6
t1 = 3 s (max)
t2 = 150 ms (typ)
t3 = 15 ms for zoom 30x30
(10ms for zoom 15x15)
t4 = 66 ms
t5 = 100 ms (typ)
revision B PART 2 page 7. 22
code 87 80 014C - Mti S2 07.doc.doc
TECHNICAL MANUAL
Image acquisition in tomography mode is enabled using the same signals as in radiography mode.
The system then enables output READY_ACQ_RAD/HCF, with a max delay of 700 ms after the
acquisition request REQ_RAD.
Signal READY_ACQ_RAD/HCF is used as the command to move the column that will control X-ray
emission directly via the angle signal.
The max acquisition time (READY_ACQ_RAD/HCF active) is 2.2 seconds for high-speed TOMO
exams (using Logical Mode 16) and 4.2 seconds for low-speed TOMO exams (Logical Mode 15) (see
also paragraph 7.4.2 above).
During X-ray emission, the generator must send the system signals containing information on the
active X-rays via signal EXON.
After Tomography angulation, the generator cuts X-ray emission: signal REQ_RAD from the
accessory stops and as a result the system deactivates the output READY_ACQ_RAD/HCF.
The function is deactivated at any time as soon as the PREP_RAD signal stops.
PREP_RAD
REQ_RAD
READY_ACQ_RAD/HCF
700ms
X-Ray
EXON
200ms (max)
This level of integration lets you automatically set the X-ray generator parameters (focus, kV, mA or
mAs) on opening an exam. These parameters are programmed during installation of the exam folders
in the system (see paragraph 4.4 “EXAM SETUP”).
• On selecting an exam, the system sends the parameters saved in the exam card (focus, kV, mA or
mAs) to the X-ray generator. Any changes can be made by the operator using the generator
console.
• After exposure, the system asks the X-ray generator for the parameters used (kV, mAs and time)
which are then sent to the image processor in order to get a print-out.
• The original X-ray generator console lets you check the status and settings of the generator at all
times.
Integration is done in such a way that the safety level of the X-ray generator is unaffected. This means
that the generator always keeps the working parameters under control. The system simply proposes
a setting and displays the actual value received from the generator.
The safety of the data transferred to the generator is guaranteed by both the communication protocol
used and the generator itself, which won't accept any out-of-range parameters.
The system foresees two serial communication channels in addition to the logic signals in paragraph
7.5.2.1. These can be used as an alternative, RS232 or CAN, in order to get this level of integration.
The possible messages are listed below (the system can be adapted to suit the X-ray generator
protocol to code and read these messages).
Contact ATS for details of the communication protocols.
The following messages are sent to the X-ray generator on opening an exam:
- Focus selection (small, large),
- KV setting, for RAD and FLUORO
- Technique setting (AEC, NO AEC),
- mA setting (in AEC mode),
- mAs (in NO AEC mode).
After exposure, the Main controller asks the X-ray generator for details of the exposure parameters:
- kV value request,
- post mAs value request,
- post exposure time request.
- dose patient request.
7.5.3.1.2 Alarms
• Every time the X-ray generator receives a setup message, it responds with the relevant status
message.
The system checks that the response arrives correctly within 1 second, otherwise it sends the
message again. After 3 failed attempts it ends transmission and sends the “SET X-RAY FAULT”
alarm to the video processor.
• The system checks that the post exposure data are correct. If not (reception error...) it sends the
“POST X-RAY DATA FAULT” alarm to the video processor.
• All the other X-ray generator alarms are not managed by the system: they continue to be visible on
the X-ray generator console.
The system can be integrated with a collimator, interface via CAN line:
The collimator is controlled by the remotely controlled stand. This automatically limits the X-ray beam
to suit the selected detector field (zoom): 43 x 43 cm, 30x30 cm, 20x20 cm or 15x15 cm.
The remote controlled stand is responsible for controlling the collimator, which automatically limits the
X-ray beam to suit the selected detector field (zoom): 43 x 43 cm, 30x30 cm or 20x20 cm.
The collimator aperture can be programmed for each system exam card and for each size. This value
is communicated at the start of each exam and is typically the set aperture for exposure in
radiography mode.
At the end of one-shot exposure or a series (both in radiography mode), the system receives the
actual aperture size and uses this for the application of the digital shutters on the image.
Considering the normal way the collimator works with a remote controlled stand:
- during fluoroscopy mode (continuous and pulsed HCF) you can manually adjust the aperture so
that it is less than the max aperture on the selected field (zoom).
- there are two scenarios in radiography mode:
- with the collimator in “automatic” mode, the aperture will be that programmed for the
exam,
- with the collimator in “manual” mode, the aperture will be that set by the operator during
the last fluoroscopy acquisition.
There is also a condition called “Total manual control” of the collimator, set via a CAN message sent
by the system during the detector setup, which makes it possible to open the collimator to its
mechanical limit.
7.6.2 INTEGRATION
In order to enable integration with the collimator, you must first enable its selection in the general
system setup menu (General Setup / Main Controller RAD Setup, see paragraph 4.2.1 above).
When you select an exam, the system sends the collimator the various aperture values and the need
for any X-ray beam hardening filter, as programmed in the exam setup card.
Any corrections in aperture can be made by the operator using the controls on the stand or by the
collimator itself.
After exposure, the actual value of aperture will automatically be used to position the electronic
shutters around the useful part of the image.
The communication protocol also foresees that the collimator supplies info on the focal distance. This
is useful for the system to check that the focal distance is coherent with the programmed exam
settings (otherwise an alarm will be generated and acquisition will be inhibited).
The system can be integrated with various stands (remote controlled, multipurpose, motorised or
manual).
For example, a multipurpose motorised stand lets you program the system so that it tells the stand
the position to be reached for each type of exam selected by the operator (for example CHEST PA).
For example, the figures above show such a system positioned for:
- Fig. 1 - AP/PA projection on a patient lying on the bed,
- Fig. 2 - AP/PA projection on a standing patient,
- Fig. 3 - LL projection on a patient lying on the bed.
7.7.2 INTEGRATION
The reverse image settings can only be sent to the HIRIS RF43 via CAN bus messages. The field
settings are sent to the system via either logical signals or CAN bus messages.
CAN messages allow for details of the general status of the stand (still, moving), the presence of a
grid, stand alarms and warnings, focal distance, stitching parameters, etc.
To enable integration with a motorised multipurpose stand, you need to enable its selection in the
general system setup menu (General Setup / Main Controller Setup, see paragraph 4.2 above).
Foreseen options:
- Integrated Remote controlled stand, to enable transmission of CAN messages to the stand.
- Check Technique. If enabled, the system will inhibit image acquisition should the actual position of
the stand not match the programmed position (if disabled, just a WARNING is given and exposure is
allowed).
The choice of the position of the stand can be programmed in the individual exam setup card (Exam
Setup menu, see paragraphs 4.3 and 4.4 above), letting you set up to 8 positions using the
Accessory Code parameter.
When you open the exam, the system sends the code for the programmed position via the CAN line.
This is used by the stand for positioning.
Once positioned, the system checks that the position reached corresponds with that set for the exam:
a warning may appear if the position is wrong (acquisition is still possible) or an alarm (acquisition
inhibited), depending on the installation setup (Check Technique option in paragraph 7.7.2.1 above).
7.7.2.3 ALARMS
Integration involves a series of alarms that appear on the system monitor, if used (depending on the
type of stand used).
The system allows for the management of 2 anti-scatter grids, focussed at two different focal
distances, via CAN line.
.
Depending on the type of exam, integration allows for:
• Checks that the grid has been inserted correctly.
• Checks that the focal distance has been set correctly (if the X-ray system is capable of sending the
value via CAN protocol).
Please contact ATS for further info concerning the Can communication protocol.
• The exam setup cards let you choose between two grids: Grid 1, Grid 2 and No Grid.
If, when the exam is opened, the grid actually used isn't the one set in the exam setup card, the
system will generate an alarm and prevent exposure.
Possible alarms:
- Remove grid
- Add grid (for systems with only one grid)
- Add grid 1 (for systems with two grids)
- Add grid 2 (for systems with two grids).
• You can set up the system from the General Setup / Main Controller RAD Setup frame by defining
the working range for each grid.
The min and max limits define the range of variation in the focal distance; exposure is inhibited if the
value of the focal distance falls outside these limits.
Possible alarms:
- Increase focal distance
- Decrease focal distance.
The DAM may be connected to either the X-ray generator or the system.
If to the generator, the system asks the X-ray generator for details of the accumulated dose via serial
messages (RS232 or CAN).
In either case, the system receives the dose value at the end of each exposure (one-shot or series
radiography; continuous or pulsed fluoroscopy).
- One-shot radiography:
- the system calculates the overall value of the “Total Exam Dose”,
- this is then associated to the image in real time,
- Run radiography:
- the system calculates the overall value of the “Total Exam Dose”,
- this is then divided by the number of exposures in the run and the resulting mean figure is
then associated to each image in real time,
The “Total Exam Dose” is saved by the system as part of the exam parameters and sent via DICOM
using the relevant functions (MPPS).
7.10.1 INTRODUCTION
The stitching function lets you reconstruct an X-ray image starting from a set of images acquired in
Radiography Mode (min 2 - max 4) by moving the detector with regular steps.
For example, you can obtain the following image sizes with detector steps of 30cm:
43 x 60 cm (2 acquired images)
43 x 90 cm (3 acquired images)
43 x 120 cm (4 acquired images).
The “stitched” image can be shown on the monitor, processed, printed or sent via network using the
normal system functions.
Stitching takes about 15-20 seconds and depends on the number of images involved.
This function is only possible with a stand capable of scanning the patient as required (i.e. moveable).
The approach described here involves moving the detector along the patient, while the X-ray tube is
turned keeping the focus in the same position, as shown in the figures below.
2 image
stitching
3 image
stitching
The angle of rotation and the shutter aperture must be set to ensure consecutive images overlap. The
overlap must be large enough for the stitching algorithm to reconstruct them correctly (normally 3 - 5
cm).
Note: You can reverse the patient scanning direction (bottom up) by selecting the VFlip parameter in
the Option->Setup->Stitching Setup menu.
2 image
stitching
3 image
stitching
● Stand
The stand receives the exam stitching setting from the system and prepares for the procedure. The
information is supplied by a CAN BUS message.
Radiography images are obtained by using the normal PREP and RAD commands.
After the first exposure, the stand starts moving the detector and rotating the X-ray tube: once in the
next position, all movement stops and X-ray exposure is enabled.
This sequence can be repeated for up to 4 acquisitions.
The stand or the system can enable the next acquisitions once the system has checked the movement
via CAN BUS messages.
● X-ray generator
The procedure involves the same interface signals, PREP_RAD and REQ_RAD, as normal image
acquisition in radiography mode.
• Operator interface
Single images are acquired as a single run and the stitched image is placed at the start of the run.
The full image indicates the X-ray dose as the sum of the individual images.
Note: The full image may, in theory, be formed of all the pixels in the various single images. If 3
images are stitched together, the result would be a single stitched image of roughly
2880x2880x3 = 24,883,200 pixels. This would mean images weighing 50MB or so are stored on
the HD and sent via DICOM. To avoid such heavy files, the image pixels are automatically
reduced using a factor (min 1.5 - max 3 ).
This parameter is set in menu Option->Setup->Stitching Setup.
● Exam setup
The exam cards that can be set as Stitching cards are those with ID = 41-45, i.e. cards 41-00 to 45-99.
To select Stitching, use the Post Processing Setup option for the card.
The stitching function and its parameters are set in the Post Processing Setup card
Once enabled, the function is valid for all sizes.
● Set:
- Stitching Component Images [2-4]: Use to define the number of images to be acquired (min 2 –
max 4).
- Stitching Custom Code: This code is sent to the stand on opening the exam (not used directly by
the system). Can be used to preset the stand:
example: 1 Stitching for rachis
2 Stitching for lower limbs
N.B.: Code “0” disables the sending of the STITCHING mode signal to
the accessory.
- Stitching ROI Type: This defines the post processing calculation (ROI) for the stitched image.
Note: the post processing set in the exam card applies to both the final
stitched image and the individual ones; ROIs are differentiated with one for
the stitched image and another for individual images.
- Stitching. Set Name: the stitching setup profile for the exam.
(see paragraph 7.10.5).
The “setup profile” is the combination of settings needed for correct functioning of the stitching
algorithm. These settings mainly depend on the scanning movement (stepping, overlapped area,
precision of positioning …).
- a menu appears,
- select REBUILD STITCHED IMAGE
- another menu appears: Exam Type Selection.
This image shows the first two images in the stitching sequence and the overlap used by the
STITCHING function.
• Using the Low and High parameters you can adjust the brightness of these images. To speed things
up, we recommend clicking on the blue slider bar and then dragging this right or left.
• Use Previous and Next to select other pairs (if any) of single images. For example, in a run of 3
exposures, you can check the stitching results by using the 1st and 2nd images or the 2nd and 3rd
images.
- stitching parameters:
- chessboard:
- images: shows the position and size of the Valid region and Overlap region
Image valid region Width 2700 Set a value just below the basic detector matrix
(2880 x 2880 pixels)
(pixel) (*) Height 2700
Used overlap region Width 2600 Set a value just below: Image valid region -
Width
Height 270 Set a value just below the actual overlap area
(pixel) between two images: 270 ≅ 4 cm
Y Offset 0÷200 This lets you shift the overlap region down to
move it away from the edge of the shutter (if
present).
Travel unit Motion Nominal step Step/0,148 This sets the scanning step (pixels). e.g. 30 cm
= 2027 pixels
(pixel) Max step error 50 Set a value to suit the expected stepping error
by the stand (**).
50 pixels ≅ 7.4 mm
Max orthogonal error 50 Set a value to suit the expected transverse
positioning error by the stand (**). 50 pixels ≅
7.4 mm
Projection center X Offset 0 To define the offset the centre of the acquired
(pixel) Y Offset 0 image and the centre of the detector
Reliability (0..1) 0 = 0% 0.5 To define the minimum valid stitching value. If
1 = 100% the stitched image exceeds this threshold, the
lateral markers for the stitching areas are
shown as triangles (reliable stitching); otherwise
they are rectangles (non reliable stitching)
Note:
(*) The detector size in pixels is 0.148mm.
(**) The error parameters are sent to the stitching algorithm to define the area in which to find the
stitching line. For instance, if Max step error =50, the algorithm looks for the stitching line in the
area defined by the nominal step ±50 pixel.
We recommend setting the smallest possible error values to reduce the search area (and thus the
time needed for stitching) and to avoid the risk of false reconstructions.
Profile test:
Press Test to check the stitching function: the image below indicates how the results are shown:
Validation of results:
Click on View result to open the Stitching result page with various diagrams that let you assess the
stitching results.
• Orthogonal error
Check the Max orthogonal error by making sure
that all the errors shown in the graph are within
the set range. If not, increase the Max
orthogonal error.
You need to check the profile using a few dedicated functions and a stitched image.
Press More >> in the Exam type page:
• Images showing 2 consecutive images (1st and 2nd in the run by default) and highlighting
both the Image valid region and the Used overlap region.
You can increase/decrease the size of these (Zoom factor 11:1 - 1:11).
• Chessboard showing the part of the 2 images used to create the Used overlap region, plus
chessboard option.
Show chessboard click to get a chessboard view: the image consists of small squares
containing elements taken from the 2 images shown in the Images
window (alternating order).
The size of the squares is set using Square size (accepted range: 5 -
100 pixels).
Zoom to fit check to view the Used overlap region image at the max zoom factor
that fits the frame; if not checked, you can use the Zoom factor at will
(accepted range: 11:1 - 1:11).
Ortho move to move the image at right-angles to the scanning movement, i.e. RH/LH
on the screen (accepted range: -100 - 100 pixels).
Change the values to find the best profile parameter settings. Bear in mind that:
- If the area to be stitched is full of detail, we recommend using the most appropriate Zoom factor.
The system supplies a signal for synchronising the start of the injector with the X-ray emission
sequence. The injector can be started up to 30 seconds before (advance) or after (delay) the start of
image acquisition.
The injector start must, in any case, be activated via a device outside the system, using the
commands and safety devices on the injector itself: the injector must stop regardless of the signal
provided by the system as soon as you release the external manual command or the safety devices
trip (or fail).
This function is only enabled in exam cards with the ANGIO function enabled. You must have the
injector key in order to enable an exam card for the ANGIO function.
The SYNC_INJ outputs continue to be active until you release the image acquisition command.
2 SINC_INJ_1 (contact NOb) Timing signal for injector start (normally open relay contact)
PREP_RAD 30
PREP_RAD REQ_RAD 12
COM_REQ 31
RAD
ZOOM
ZOOM 1 (30x30) 34
ZOOM 2 (20x20) 16
SEL_FLAT_PANEL 1 17
SEL_FLAT_PANEL 2 13
SEL_FLAT_PANEL_1_CPI 14
SEL_FLAT_PANEL_2_CPI 33
0V1 ACC. 15
0V ACC. 10
+V ACC. 29
+24V
SEL_INJ. 3
SEL_INJ.
SYNC_INJ. 21
SYNC_INJ.
23
CM11
SYNC_INJ_1 1
SYNC_INJ_1
SYNC_INJ_1 2
SYNC_INJ_1
ACCESSORY/X-RAY GENERATOR
INTERFACE
LIST OF CONTENTS
1 ACCEPTANCE TEST
1.1 INTRODUCTION
The acceptance test is basically a series of controls/checks listed in the TEST SHEET that let you
assess the status of the system.
The test must be carried out during acceptance of the system, after its installation. It's also used as a
checklist during routine maintenance and after extraordinary maintenance.
The system integrator is responsible for filling in the acceptance test sheet after installation.
Mechanical checks:
- Fixing of the main components: look for loose screws (cabinet, detector unit).
- Safety of the protection devices for the main components: look for signs of damage or dents that
may affect the efficiency and safety of the equipment.
Switch the equipment off using the 2 on/off switches on the panel in the control room: acquisition
system power ON/OFF and detector power ON/OFF (see para. 3.3.4 in Part 2 above).
TS31 Check the general state of the boards and cabinet components (look for dirt and corrosion).
TS32 Check the impedence value between the earth terminal protecting power unit GP1 (inside
the cabinet) and the earth screw on the detector (we recommend checking this by first
detaching detector power supply connector).
- Impedence must be less than 0.1 ohm -
Close the cabinet.
TS34 Check that the video processor fan is working properly and that the ventilation holes are not
blocked.
TS35 Check that the PU module fan is working properly and that the ventilation holes are not
blocked.
TS36 Check that the monitors are working properly.
Wait for the HIRIS RF43 application to open and then open a new exam.
Select an exam card, insert a suitable phantom and give an exposure command.
TS42 X-ray dose in the air intensity (automatic exposure meter control)
Procedure: Carry out this check at both 0.5 and 5 µGy using the ion chamber
calibration procedure explained in para. 5.2 of Part 5 of this
manual.
The measurement is taken at the leading edge of the detector.
Reference values: ATS has adopted CEI EN 60601-2-7, para. 50.102.1-b, using the
video levels of the image (with phantoms of different thicknesses)
to measure the results in 3 different situations:
1. changes in the thickness of the phantoms
2. changes in the Kv.
3. repeating exposure at the same conditions of exposure.
Reference values: ATS adopts the "FUNK 18" resolution test phantom: make sure
that the test results are close to those in the Test Report attached
to the system.
Reference values: ATS adopts the "FUNK 18" resolution test phantom: make sure
that the test results are close to those in the Test Report attached
to the system.
Procedure: - Phantom at 45° to the centre of the detector field, field 43x43
cm
- Give the continuous fluoroscopy command and set the kV to get
the best image.
- Find the max image resolution on the monitor and then compare
this value with that shown on the system Test Report.
- Repeat for fields 30x30 cm, 20x20 cm and 15x15 cm.
Checks for pulsed HCF fluoroscopy mode (only for PIXIUM RF43 detector):
Reference values: ATS adopts the "FUNK 18" resolution test) phantom: make sure
that the test results are close to those in the Test Report attached
to the system.
Procedure: - Phantom at 45° on the centre of the detector field, field 43x43
cm
- Give the pulsed fluoroscopy command and set the kV to get the
best image.
- Find the max image resolution on the monitor and then compare
this value with that shown on the system Test Report.
- Repeat for fields 30x30 cm, 20x20 cm and 15x15 cm
Make a few blank copies of the test sheet attached to this manual (“loose pages”).
Part 4 : ADJUSTMENTS
LIST OF CONTENTS
1 ADJUSTMENTS
1.1 WARNINGS
All the envisaged adjustments have been described in Part 2 of this manual, as these are absolutely
necessary during installation.
See Part 2, “INSTALLATION” for details.
Part 5 : MAINTENANCE
LIST OF CONTENTS
5.1.1 Profiles
5.1.2 Histograms
5.1.3 Statistics
5.1.4 Save to file
5.2 Calibrating the AEC measuring chamber
5.2.1 AEC chamber calibration with RF detector
5.2.1.1 KV compensation
5.2.2 AEC chamber calibration with WIRELESS detector
5.2.2.1 KV compensation
5.2.3 Preliminary Claymount AEC settings
5.2.3.1 Claymount 5 area AEC
5.2.3.2 Claymount 3 area AEC
5.3 Activating DAM and Dose Area Meter efficiency test
5.3.1 Activating DAM reading from Main Controller
5.3.2 DAM indication test from Main Controller
5.3.3 Dose Area Meter calibration
5.4 Adjusting the working dose of the system in
fluoroscopy and pulsed fluoroscopy
5.4.1 Adjusting the working dose in continuous
fluoroscopy mode
5.4.1.1 Setting the dose clipping factor
parameter
5.4.2 Adjusting the working dose in pulsed hcf
mode
5.4.2.1 Setting the dose clipping factor
parameter
5.5 Leds on the detector
5.5.1 PIXIUM RF4343 detector unit
5.5.2 PU RF4343 image correction module
5.5.2.1 DYNAMIX board leds
5.5.2.2 ADD ON board leds
5.6 Saving images in raw format to cd
5.7 updating the back-up hard disk
5.7.1 Updating with Acronis True Image
5.8 Image archiving to CD
5.8.1 Creating a ghost disk using acronis true
image
5.9 Video processor dump information
5.10 Study recovery function
1 ROUTINE MAINTENANCE
The system must be serviced at least once a year and, in any case, in accordance with local laws and
regulations: this mainly involves carrying out the Acceptance Test described in Part 3 of this manual.
This procedure includes the performance constancy check, via the checks on the precision and
constancy of the automatic exposure control.
The results must be noted in the TEST SHEET (see paragraph 1.2 of Part 3 of this manual) together
with any action taken.
In addition, the detector must also be re-calibrated on a regular basis (Gain and Detective
Map).
The detector manufacturer recommends that Gain and Defective Map calibration should be
performed every 3 – 6 months.
WARNING
BEFORE OPENING CABINET COVER, POWER OFF THE SYSTEM USING BOTH THE ROOM
SWITCH (SYSTEM POWER LINE AND DETECTOR POWER LINE) TO AVOID ELECTRIC SHOCKS.
Do not use cleaning products that have a high alcohol content. Do not use corrosive detergents,
abrasive products or solvents.
Use only disinfectants and sterilising products, if required, that fully comply with existing disinfection
and explosion prevention procedures.
The equipment must not be used in presence of anaesthetics and/or inflammable disinfectants and
cleaning products .
If the use of the system needs the sterilisation of the device, this must be done by the user.
Note: - See paragraph 4.1 in Part 6 of this manual for details on how to clean the Wireless DRX1
detector (for the system fitted with two detectors)
- See paragraph 4.1 in Part 7 of this manual for details on how to clean the Wireless PIXIUM
3543EZ detector (for the system fitted with two detectors)
2 EXTRAORDINARY MAINTENANCE
Note: - See paragraph 4.2 of Part 6 of this manual for details on how to replace the components in the
Wireless DRX1 detector unit (for the system fitted with two detectors).
- See paragraph 4.2 of Part 7 of this manual for details on how to replace the components in the
Wireless PIXIUM3543EZ detector unit (for the system fitted with two detectors).
2.1 REPLACING THE SYSTEM HARD DISK WITH THE BACK-UP HARD DISK
The system is supplied with a back-up hard disk. This is an exact copy of the system hard disk.
You can thus replace the original hard disk with the back-up HD if necessary.
Note: To ensure that the back-up HD is always ready for use in an emergency, we recommend that it
is always updated to guarantee that it contains an exact copy of the system HD. Therefore, you
must copy to the back-up HD any changes to an application or the system setup after installing
the processor: see the procedure explained in paragraph 5.7.
A A A A B
If you need to restore the video processor hard disk, use the backup disk (i.e. the “ghost disk”)
provided with the system.
This disk is created:
- by ATS after factory testing,
- by the installer after system setup (see paragraph 5.8 in Part 5).
The procedure for restoring the video processor system hard disk is:
1) Insert the “ghost disk” in the DVD driver on the processor (at the front of the processor rack).
8) The video processor reboots with the new setup and automatically loads the software needed to
restore the system hard disk (Norton Ghost or Acronis True Image)
The video processor automatically reboots with the restored system hard disk.
The video processor’s CPU board BIOS can be lost if its buffer battery fails or if there are problems
with the mains supply or the system hardware.
The PID and PU processors contain fans that are maintenance-free, but which should be replaced
every 3-4 years as a precaution.
The detector processor module (PU RF4343) also contains a maintenance-free fan. To ensure against
the risk of a malfunction (less than 1% over 10 years), we recommend changing this every 2-5 years.
To replace this:
- Switch the equipment off and then open the cabinet:
o undo the 4 screws A using an Allen wrench (see figure)
o remove the side panel
- The PID can now be accessed directly, while to access the PU you also need to:
o extract the PU module rack
o remove the detector power unit
- Open the doors fixing the fans (each has 2 screws)
- Replace the fans
- Remount all the above components in the reverse order
A A A A
2) Access the PID as Administrator: username Admin2, password ats_PED.24 (see paragraph 7.3
in Part 2).
5) Copy the PU SW upgrade directory (“THALES TEST TOOL VERSIONE X.X.X.X”) on the CD on
the PID HD “C:\Install”.
7) Enter the PU IP address: “192.168.0.2” in the box shown in the figure below and click on Open.
Note: If the system is fitted with two detectors and THALES SW with release less than 2.3.4.0 , you
will need to set the PU host name: 10.0.1.2 (see paragraph 2.3.1 in Part 6).
9) Unload any active Logical Mode: to do this, access the Logical Mode list in the “Logical mode in
the Dynamix board” window and select the active mode (with (A) next to it) and then click on
Unset Active.
13) The PU is now automatically upgraded (this may last 10-15 minutes).
The system will not respond during this process.
Note: Do not switch off or reset the PID or the PU until the upgrade has been completed.
14) If the upgrade has been completed successfully, the File Transfer window closes.
Now click on Close to close the application.
To upgrade the SW, copy the upgrade SW from the CD to the installation directory:
- CD_HirisRF43_X.X.X
- SD4343 ver X.X.X.X
- DGVIEWER
in directory “C:\ Install \UPGRADE PID“. Then follow the steps below:
1) Access the PID directory C:\ Install\ THALES TEST TOOL VERSIONE X.X.X.X\ software”
and run RF4343 Customer Software Setup.
2) Click on Next (2 times) and make sure that the installation path is:
“C:\ Program Files\ THALES ELECTRON DEVICES\ RF4343”,
Confirm by clicking on Next, then click on Install and finally on Finish.
To upgrade the software you first need to remove the previous version on the PID. To do this:
1) Access the WINDOWS Control Panel, select Add/Remove Programs and remove the HIRIS
Rf43 and DICOM Kit software.
4) Launch the application and then open the Option General Setup menu (only when the archive
is closed) and set the PU Service Data parameters as shown in the figure below:
See paragraph 2.4.2. above for details on how to set the general setup parameters correctly
Note: - if the system has two detectors (with a DRX1 wireless detector), see also paragraph 2.3.2 in
Part 6 of this manual for details of the correct setup for the General Setup parameters.
- if the system has two detectors (with a PIXIUM 3543EZ wireless detector), see also
paragraph 2.3.2 in Part 7 of this manual for details of the correct setup for the General Setup
parameters.
You need to follow the procedure below to install a USB hardware key on a PIDRF43 with a parallel
hardware key:
● Switch the system on and log in as the system Administrator.
● Copy the \Smartkey2003 folder on the CDROM or USB provided or downloaded via FTP into
this folder:
C:\INSTALL
● Access the folder C:\Install\SmartKey2003\Install\ and run the SDI.exe application.
● Click on the “Install” button in the USB panel.
You are now asked to insert the USB hardware key. Insert and then select “OK".
• SD4343:
To upgrade the SD4343 software, you first need to remove the previous version on the PID. To do
this:
1) Access the WINDOWS Control Panel, select Add/Remove Programs and remove the SD4343
software.
5) Click on FILE NEW and select Online using an existing file as template.
The SD4343 software is now fully upgraded and the PU setup updated.
• DGVIEWER:
To upgrade the DGViewer SW (review software attached to every CD/DVD you burn):
To upgrade the Main Controller SW simply replace the EPROM on the Main Controller CPU with the
new one:
CPU board
EPROM
To upgrade the HIC-04 board firmware, simply replace integrated chip IC24 (CHANGES EPC1441):
Board HIC-04
IC24
If the manuals are updated after installation of the system, any obsolete parts need to be replaced.
The manuals to be installed are supplied on CD.
To install them:
- Rename the obsolete manuals (e.g. by adding “old” to their name) if you wish to keep them.
Copy the new manuals to the same directory and save them with the same names as before:
This chapter explains how to replace the PU RF4343 processor and provides instructions for its setup.
- Remove the 2 side panels on the electrical cabinet after undoing the screws.
- Undo the 4 screws fixing the plate supporting the PU rack and the detector power unit to the
shelf.
- Disconnect all the cables on the PU rack and detector power unit, slide the plate supporting these
out and extract from the cabinet.
- Remove the PU rack from the plate and replace with the new one.
- Return the plate to the cabinet and secure with the 4 screws removed previously.
- Reconnect the cables on the PU rack and detector (see paragraph 3.2 in Part 2 of this manual for
details).
- Connect a monitor with VGA input, a PS2 mouse and a PS2 keyboard to the PU in order to carry
out the setup (if not available, you can “borrow” those connected to the PID video processor for
this purpose).
- Switch the cabinet on and make sure that the PU is on (if it does not start automatically, press
the Power ON button above the rack fan).
- Keep the Del key on the keyboard pressed during the PU startup to enter the Bios Setup
(if asked, enter: KAvftwba).
- Select the Advanced Chipset Control option in the Advanced tab and set Legacy USB
support: Enabled.
- Select the Exit saving changes in the Exit tab and confirm with Yes.
- Wait for the processor to restart automatically and then log in (a Login window appears on the
monitor) as follows:
user: Admin password: P77P8CqAam
- Select Start / Settings / Control Panel / User Accounts in the Windows XP task-bar.
- Enter the account name (Admin2) and type (Computer Administrator) and then press
Create Account.
- Enter the password twice (ats_PED.24) and then press Create Password:
2) You can now log in using this new user and perform the following operations:
- Select Start / Shut Down in the Windows task-bar and then Log Off Admin.
- Enter User name: Admin2 and Password: ats_PED.24. Then press OK.
- Open Windows Explorer, take the mouse pointer to disk D and then right-click to select the
Sharing and Security option.
[ModeACPP17]
Windowing = No
DRZ = 0
BinVert = 1
BinHor = 1
Binning2x3to3x3 = No
BackLight = No
ResetLight = Yes
ResetLightDiv = 9
Mode = 1
Gain = 7
PolyGain = No
BigMode = Yes
Refresh = 0.125
- if not, edit the parameters as shown above and then save the file,
- close the file.
4) Select Start / Settings / Network Connections in the Windows task-bar and check that the set
IP ADDRESS for the network connection is 192.168.0.2 .
Note: If the system is fitted with two detectors and THALES SW with release less than 2.3.4.0 ,
you will need to set the PU host name: 10.0.1.2 parameter (see paragraph 2.3.1 in Part 6).
5) Select Start / Shut Down in the Windows task-bar and then select Restart to reboot the PU
processor.
6) Once the PU processor is running, disconnect the mouse, keyboard and VGA monitor cable (and
return to the PID if “borrowed” from this).
1) Close the application and reboot the system as Administrator (see paragraph 7.3 in Part 2 of this
manual) by entering: User name: Admin2 Password: ats_PED.24
- Enter the PU’s IP address: “192.168.0.2” in the box shown in the figure below and then click
on Open.
Note: If the system is fitted with two detectors and THALES SW with release less than 2.3.4.0 ,
you will need to set the PU host name: 10.0.1.2 parameter (see paragraph 2.3.1 in Part 6).
- Return to the main Setup TAB and click on Apply and then Close to break the PID/PU
connection.
- Connection is now made between the PU and the PID video processor. A new page appears:
select the Configuration tab in the right-hand list and make sure that the settings match those
shown below:
To restore the correct detector setup and the parameters of all calibrations previously saved in the
system memory, use the following function in the SD4343 software:
2) Establish connection with the PU and deactivate any active logical mode (see paragraphs 5.2 and
5.5 in Part 2)
4) Find the directory of the files you want to transfer in the Source Folder; the directory was defined
during the previous Upload, normally:
6) Transfer will take about 30s; a message appears in the Status area once transfer has been
completed successfully.
7) Carry out the PU Hardware Reset procedure (see paragraph 5.9.2 in Part 2) and then close the
application.
The procedure described here should be followed when installing an optional Reference Monitor or if
the system setup data is lost in the case of multiple display (working monitor and reference monitor).
If a reference monitor is used, the video processor must have the following monitor setup:
- Close the image acquisition application and switch off the device.
Important: both monitors must be connected to the video processor when you reboot
the system.
User: Admin2
Password: ats_PED.24
- Check that the 2 monitors are correctly set up using the Multiple Displays menu, as shown in
the figure below:
Note: “Primary Display” identifies the WORKING MONITOR connected to the VGA port, while
“Second Display” identifies the REFERENCE MONITOR connected to the DVI port via a
DVI/VGA video adapter.
- Press Apply.
- Select the General Settings menu and check that both the monitors (Monitor and Digital
Display) are correctly set up, as shown in the figure below:
2) A fault in the Main Controller board, but its CPU is working correctly and you can still access
the Main Controller setup data via the HIRIS RF43 application.
3) A general fault with the Main Controller making it impossible to access the setup data.
FAULT N°1
The Main Controller CPU needs to be replaced. No setup operations are required.
FAULT N°2
Before replacing the Main Controller, you need to find and make a note of its setup data; these must
then be re-entered for the new Main Controller.
Follow the steps below to find the setup data (with reference to paragraphs 5.2, 5.3 and 5.4 in the
Annexes to this Part of the manual):
• Create a new study, select exam “99 TEST- 11 Dose Adj. Exam” and then select the M.C. RAD
SETUP menu and load the Logical Mode using the Start Calibration key. Make a note of the
parameters and selected options:
• Select the M.C. FLUORO SETUP menu and make a note of the parameters and selected options
for all the Logical Modes in fluoroscopy mode: 43x43, 30x30, 20x20 and 15x15.
• Select the M.C. PULSE SETUP menu and make a note of the parameters and selected options for
all the Logical Modes in pulsed fluoroscopy mode: 43x43, 30x30, 20x20 and 15x15.
You can now replace the Main Controller board and re-set the parameters and options in the same
menus above.
FAULT N°3
Follow the steps below if there is a general fault in the Main Controller making it impossible to recover
the setup data via the HIRIS RF43 application:
• Remove component IC1 containing the setup parameters from the faulty Main Controller board
(EEPROM 93cX6 positioned under board CPU, as shown in the figure below):
• Fit component IC1 on the new Main Controller board (replacing the existing one). It is important
that the HIRIS RF43 MC software for the new Main Controller has the same SW release as the
previous one.
3 SPARE-PARTS
Note 1: When ordering spare-parts, always quote the system serial number as well as the item code.
Note 2: Contact ATS for details of those parts not included in the list.
Note3: See paragraph 4.3 in Part 6 of this manual for details about spare part list of the Wireless
DRX1 detector (for the system fitted with two detectors).
Note4: See paragraph 4.3 in Part 7 of this manual for details about spare part list of the Wireless
PIXIUM 3543EZ detector (for the system fitted with two detectors).
9
X-ray tube
View from B 8
10
13
X-ray
7 3
Scheda RGBC 01 1a
5
5
6
SD 87 002
1b
View from A
FUS213/630mAT
2
RDAM board
4
6
11
A
87 01 400
ICI board 14
Transf. 7001050B
RS 232 board
8701110 transformer
8701250 stabilizer
RFPS board
ASPS board
trasf.17 10 120A
2 4
87 01 200 87 01 300
revision F PART 5 page 3. 3
code 87 80 014C - Mti S5 03.doc
TECHNICAL MANUAL
4 SCRAPPING
When scrapping the device, special care must be taken when handling the following components as
these can be hazardous:
- The LCD monitor contains a vacuum input screen which, should it break, can send shards
of glass flying and disperse powder and gas that may be toxic.
- The equipment contains printed circuit boards which contain a certain amount of lead: this
is a toxic material and must be treated as such.
- The batteries used must be disposed of in accordance with European Directive 2002/96/EC
(disposal of waste electrical and electronic equipment - WEEE).
All parts should be disposed of in accordance with the prevailing regulations in each country
at the moment of scrapping.
This symbol is placed on the equipment to remind you that the equipment meets
the European Environmental Directives (2002/95/EC, 2002/96/EC, 2003/108/EC)
and so must be scrapped in accordance with the relevant laws for separated waste
disposal.
This equipment must not be disposed of as normal urban solid waste: it must
be taken to an expert waste disposal centre or returned to the dealer, should
you wish to replace it with a new model.
5 ANNEXES
• Selection:
to choose the field in which you want to read the values.
You can read the values in a:
- 9x9 pixel square
- horizontal segment
- vertical segment
rectangle.
• Statistics:
- to measure the raw image (raw = detector acquired
levels) or the processed monitor image (equalized):
- value in centre (square 9x9 only)
- min value
- max value
- mean value
- standard deviation,
- to select the images to be shown (segments only):
- profile
- moving mean
- moving standard deviation,
- to select the reading window for the moving mean and
moving standard deviation (segments only).
5.1.1 PROFILES
The figure here shows an example of the profiles for a horizontal segment.
5.1.2 HISTOGRAMS
The figure here shows an example of the histograms for a horizontal segment.
5.1.3 STATISTICS
Click on Statistics and the graphs selected in field Statistics are displayed.
Press Save to file to open the Save pixel values window and so save the selected statistics.
The system can manage up to 2 ion chambers: one for the dynamic RF43 detector and the other for a
second WIRELESS detector (optional). Below are the ion chamber calibration procedures for these
two different detectors.
The AEC measuring chamber is calibrated by adjusting the two dose values at the leading edge of the
detector:
Fig. 1 Fig. 2
* Note: If the dosimeter cannot be placed on the ion chamber:
- place the dosimeter on the patient bed or detector cover (see figure 2),
- remove the grid,
- calculate the dose value on the leading edge of the detector bearing in mind the distance
between the surface of the detector and the dosimeter and the absorption factors of the
components in-between these.
Before calibration, you must check which type of measuring chamber is installed: MEDYS or
CLAYMOUNT. If a CLAYMOUNT, you need to set a few preliminary settings before calibrating this, as
described in paragraph 5.2.2. below.
Calibration procedure:
• If not already entered, enter the Technical Password: “TechService” (or “985312”)
and wait for the system to complete the Offset calibration procedure for each Logical Mode.
• Click on key on the toolbar and then option M.C. RAD. SETUP to open the calibration page
• Select Id Log Mode 10 (RAD 3i/s) in the “Available Logical Modes” table and load by clicking on
Start Calibration. The “Dose Settings” calibration table is now enabled.
5.2.1.1 KV COMPENSATION
Note: This function can only be used with an integrated X-ray generator (also via serial RS232 or CAN
protocol), i.e. capable of providing the System with details of the kV values set on the console.
Calibration procedure:
4) Give the single radiography exposure command and then use the Image Statistic menu (see
paragraph 5.1 above) to check the mean level (LSB) in a central part of the image measuring
200x200 pixels (this will then be the reference value for future calibration).
5) Set the generator at 60kV-125mA and command another exposure and then check the mean level
(LSB), again in a central part of the image measuring 200x200 pixels.
6) If the measured value differs by more than ±10% from the reference value (at 70 kV):
- If not already entered, enter the Technical Password: “TechService” (or “ 985312”) and wait for
the system to complete the Offset calibration procedure for each Logical Mode.
- Click on key on the toolbar and then option M.C. RAD. SETUP to open the calibration page
for the Main Controller RAD Setup:
- Select Id Log Mode 10 (RAD 3i/s) in the “Available Logical Modes” table and load by clicking on
Start Calibration. The “Ion Chamber KV Compensation” calibration table is now enabled.
- Set the generator at 60kV-125mA and command another exposure and then check the mean
level (LSB), again in the centre of the image.
7) If necessary, repeat the above procedure until you get the measured mean level at 70 kV (±10%).
9) Set the generator at 120kV-10mA and command another exposure and then check the mean level
(LSB), again in the centre of the image.
10) If the measured value differs by more than ±10% from the reference value (at 70 kV), adjust
parameter kV Compensation at 120 kV following the procedure explained above.
11) Repeat the above procedure until you get the measured mean level at 70 kV (±10%).
The AEC measuring chamber is calibrated by adjusting the two dose values at the leading edge of the
detector:
- min dose 0.5 µGy
- max dose 5 µGy.
Fig. 1 Fig. 2
Before calibration, you must check which type of measuring chamber is installed: MEDYS or
CLAYMOUNT. If a CLAYMOUNT, you need to set a few preliminary settings before calibrating this, as
described in paragraph 5.2.3. below.
Calibration procedure:
• If not already entered, enter the Technical Password: “TechService” (or “985312”).
• Click on key on the toolbar to open the calibration window for Main Controller RAD Setup:
5.2.2.1 KV COMPENSATION
Note: This function can only be used with an integrated X-ray generator (also via serial RS232 or CAN
protocol), i.e. capable of providing the System with details of the kV values set on the console.
Calibration procedure:
4) Give the single radiography exposure command and then use the Image Statistic menu (see
paragraph 5.1 above) to check the mean level (LSB) in a central part of the image measuring
200x200 pixels (this will then be the reference value for future calibration).
5) Set the generator at 60kV-125mA and command another exposure and then check the mean level
(LSB), again in a central part of the image measuring 200x200 pixels.
6) If the measured value differs by more than ±10% from the reference value (at 70 kV):
- If not already entered, enter the Technical Password: “TechService” (or “ 985312”).
- Click on key on the toolbar to open the Main Controller RAD Setup calibration window:
- Set the generator at 60kV-125mA and command another exposure and then check the mean
level (LSB), again in the centre of the image.
7) If necessary, repeat the above procedure until you get the measured mean level at 70 kV (±10%).
9) Set the generator at 120kV-10mA and command another exposure and then check the mean level
(LSB), again in the centre of the image.
10) If the measured value differs by more than ±10% from the reference value (at 70 kV), adjust
parameter kV Compensation at 120 kV following the procedure explained above.
11) Repeat the above procedure until you get the measured mean level at 70 kV (±10%).
The equipment can be supplied with two types of Claymount AEC cameras:
- Camera with 5 areas,
- Camera with 3 areas.
See paragraph 2.3.3.2 in Part 1 of this manual for details of the size of the components.
Access the preamplifier board by undoing the 4 screws on the cover and then set the dip-switches as
shown in the table below:
Preamp. 1007
DIP SWITCH “S1”
1 ON
2 OFF
3 OFF
4 OFF
DIP SWITCH “S2”
1 ON
2 OFF
3 OFF
4 OFF
DIP SWITCH “S3”
1 OFF
2 OFF
3 OFF
4 OFF
Access the preamplifier board by undoing the 4 screws on the cover and then set the dip-switches as
shown in the table below:
Preamp. 1001
DIP SWITCH “S1”
1 OFF
2 OFF
3 OFF
4 OFF
DIP SWITCH “S2”
1 ON
2 OFF
3 OFF
4 OFF
DIP SWITCH “S3”
1 ON
2 OFF
3 OFF
4 ON
5 ON
6 ON
7 OFF
8 OFF
5.3 ACTIVATING THE DAM AND DOSE AREA METER EFFICIENCY TEST
The activation procedure described here only applies to a Dose Area Meter connected directly to the
system.
• The system asks you to enter the technical password, if you have not already entered this:
“TechService” (or “ 985312”).
• Wait while the system completes the Offset calibration procedure for the various Logical Modes.
• Click on the button in the toolbar and then select the M.C. RAD. SETUP option to open the
calibration window for the Main Controller RAD Setup:
• Select Id Log Mode 10 (RAD 3i/s) in the “Available Logical Modes” table and load this by clicking
on Start Calibration: the “D.A.M. Settings” calibration table is enabled.
• Now click on the “Dose Area Meter Present” check box and check that the correction
percentage is 100%.
You therefore need to check the connected DAM type selected, using dip switch on the DAM board
inside the Main Controller. There are two kinds of DAM card reading.
LS HS
Position of switch SWx on board RDAM Switch set for high sensitivity DAM (HS)
Use a reference dosimeter to check the DAM efficiency and indication (mGy*cm2) on the monitor.
Measuring conditions:
- Reference dosimeter placed on the patient bed (or surface of the detector cover) and set to read
µGy values,
- Collimators open to get a radiated area of approximately 1000cm2 (33cm x33 cm) on the rest
surface of the dosimeter,
- The dosimeter must be completely within the radiated area.
• The difference between these 2 values should not be more than 20%.
If this is not the case, you need to calibrate the DAM as explained below.
The calibration procedure described here is only valid for a Dose Area Meter connected directly to the
system.
Calibration procedure:
• If not already entered, enter the Technical Password : “TechService” (or “ 985312”) and wait for
the system to complete the Offset calibration procedure for each Logical Mode.
• Click on key on the toolbar and then option M.C. RAD. SETUP to open the calibration page
for the Main Controller RAD Setup:
• Select Id Log Mode 10 (RAD 3i/s) in the “Available Logical Modes” table and load by clicking on
Start Calibration. The “D.A.M. Settings” calibration table is now enabled.
• Depending on the results of the efficiency test (see previous page), adjust the Dose Area Meter
(50-150%) value to suit the deviation that needs to be corrected.
5.4 ADJUSTING THE SYSTEM WORKING DOSE IN FLUOROSCOPY AND PULSED HCF MODES
Fig. 1 Fig. 2
- DC voltmeter connected to board M.CON between Tp7 and Tp6 (0V) (see figure 3) to measure the
ABC_ERROR signal sent to the X-ray generator, reflecting the levels acquired by the detector (this
should be = 0 if the X-ray dose is correct).
Tp7 Tp6
Calibration procedure:
• If not already entered, enter the Technical Password : “TechService” (or 985312) and wait for the
system to complete the Offset calibration procedure for each Logical Mode.
• Click on key on the toolbar and then option M.C. FLUORO. SETUP to open the calibration
page for the Main Controller Fluoroscopy Setup:
• Select Id Log Mode 1 (continuous fluoroscopy on field 43x43) in the “Available Logical Modes”
table and load by clicking on Start Calibration.
• Without X-ray exposure, set the Offset parameter (in the “Logical Modes General Settings” table),
so that the ABC_ERROR signal (between Tp7 and Tp6) is = 0 V (the typical Offset value is about
315).
• Set the Polarity parameter (polarity of the kV correction signal) to suit the type of input foreseen by
the X-ray generator:
Dose OK
Detector levels
Detector levels
Dose OK
• If not already entered, enter the Technical Password: “TechService” (or 985312)
and wait for the system to complete the Offset calibration procedure for each Logical Mode.
• Click on key on the toolbar and then option M.C. FLUORO. SETUP to open the calibration
page for the Main Controller Fluoroscopy Setup:
• Select Id Log Mode 1 (continuous fluoroscopy on field 43x43) in the “Available Logical Modes”
table and load by clicking on Start Calibration.
• Set manual continuous fluoroscopy on the X-ray generator, give the X-ray emission command and
then adjust the kV to get a dose of 35 µR/s on the dosimeter.
• Acquire an image and measure the mean level (LSB): move your pointer to the centre of the image
and read the value in the Mean field in the Pixel Statistics window:
This setting established the working dose (Medium Dose) for the selected acquisition mode.
The mean level supplied by the detector depends on the sensitivity of the detector.
Typical value: roughly 270 LSB.
• The Low Dose working value is normally defined by a value half that of the Medium Dose.
To set this value, simply enter a value in field Low Dose LSB:
Low Dose LSB = ½ Medium Dose LSB.
• The High Dose working value is normally defined by a value twice that of the Medium Dose.
To set this value, simply enter a value in field High Dose LSB:
High Dose LSB = 2 x Medium Dose LSB.
• Set automatic continuous fluoroscopy on the generator, give the X-ray command and check that
the X-ray generator quickly reaches the dose level (kV /ma) you have set manually and stabilises
at this level.
If not, change the two Gain and Hysteresis parameters, bearing in mind that:
- Signal ABC_ERROR = 0 V as long as the actual dose stays within the values: set value ±
Hysteresis.
• If correctly adjusted, close the calibration page for this acquisition mode by clicking on End
Calibration.
• Repeat the same procedure for the other two Logical Modes in continuous fluoroscopy mode:
The table below shows the default dose settings and the typical levels to be measured:
CONTINUOUS FLUOROSCOPY
• Once you have adjusted the working dose, set the Dose Clipping Factor parameter, as explained
in point 5.4.1.1 below.
• Quit the menu (Exit) to save the changed parameters.
• Close the study.
The Dose Clipping Factor parameter is used to set the max image
level needed for calculating the Vabc voltage supplied by the detector
for the automatic kV adjustment function.
If, for example, you set Dose Clipping Factor = 8, when calculating voltage Vabc, the level value will
be limited to a maximum of: dose level x 8.
e.g. If the Medium dose = 250 LSB, any levels higher than 250x8= 2000 LSB will be
considered as equal to 2000 LSB for the purposes of calculating voltage Vabc.
This setting is needed to limit the influence of over-exposed areas within an image when using the
automatic X-ray dose adjustment system.
• If not already entered, enter the Technical Password: “TechService” (or 985312)
and wait for the system to complete the Offset calibration procedure for each Logical Mode.
• Click on key on the toolbar and then option M.C. PULSE SETUP to open the calibration page
for the Main Controller Fluoroscopy Setup:
• Select Id Log Mode 5 (pulsed fluoroscopy on field 43x43) in the “Available Logical Modes” table
and load by clicking on Start Calibration.
• Set manual pulsed fluoroscopy on the X-ray generator, give the X-ray emission command and then
adjust the kV to get a dose of 4 µR/frame on the dosimeter.
• Acquire an image and measure the mean level (LSB): move your pointer to the centre of the image
and read the value in the Mean field in the Pixel Statistics window:
This setting established the working dose (Medium Dose) for the selected acquisition mode.
The mean level supplied by the detector depends on the sensitivity of the detector.
Typical value: roughly 560 LSB.
• The Low Dose working value is normally defined by a value half that of the Medium Dose.
To set this value, simply enter a value in field Low Dose LSB:
Low Dose LSB = ½ Medium Dose LSB.
• The High Dose working value is normally defined by a value twice that of the Medium Dose.
To set this value, simply enter a value in field High Dose LSB:
High Dose LSB = 2 x Medium Dose LSB.
• Set automatic pulsed fluoroscopy on the generator, give the X-ray command and check that the X-
ray generator quickly reaches the dose level (kV /ma) you have set manually and stabilises at this
level.
If not, change the two Gain and Hysteresis parameters, bearing in mind that:
- Signal ABC_ERROR = 0 V as long as the actual dose stays within the values: set value ±
Hysteresis.
• If correctly adjusted, close the calibration page for this acquisition mode by clicking on End
Calibration.
• Repeat the same procedure for the other two Logical Modes in pulsed fluoroscopy mode:
The table below shows the default dose settings and the typical levels to be measured:
PULSED FLUOROSCOPY
• Once you have adjusted the working dose, set the Dose Clipping Factor parameter, as explained
in point 5.4.2.1 below.
• Quit the menu (Exit) to save the changed parameters.
• Close the study.
If, for example, you set Dose Clipping Factor = 8, when calculating voltage Vabc, the level value will
be limited to a maximum of: dose level x 8.
e.g. If the Medium dose = 250 LSB, any levels higher than 250x8= 2000 LSB will be
considered as equal to 2000 LSB for the purposes of calculating voltage Vabc.
This setting is needed to limit the influence of over-exposed areas within an image when using the
automatic X-ray dose adjustment system.
Note: - see paragraphs 4.4 and 4.5 in Part 6 of this manual for a description of the LEDs on the
Wireless DRX1 detector and on the WI FI Access Point (for the system fitted with two detectors)
- see paragraphs 4.4 and 4.5 in Part 7 of this manual for a description of the LEDs on the
Wireless DRX1 detector and on the WI FI Access Point (for the system fitted with two detectors)
The right-hand side of the detector has some 5 LEDs that indicate the power-up and efficiency of the
detector:
LED meaning
POWER This LED is connected to the power unit inside the detector and
supplies information on the detector power supply.
COM This LED comes on when the optical fibre is connected and
communication with the control rack is detected.
RDY This LED comes on when initialisation after switching-on has
been completed successfully.
FREQ This LED comes on to indicate the activation status of the
received “Acquisition request”.
XRAY This LED is switched off during activation of the “X-ray window”
that lets you enable exposure.
The LED is switched on between each “X-ray window”.
FREQ
COM POWER
XRAY
RDY
There are two 7-segment displays on the PU module, used to provide status and acquisition
information.
• When the PU is switched, a series of bips indicated that connection has (or has not) been made with
the detector:
- if connection has failed, you hear 20 bips and then a high bip followed by a low bip,
- if connection is successful, you hear less than 10 bips and then a low bip followed by a
high bip,
- an error message appears on the 7-segment displays in any case if there is a
malfunction.
Table of alarms:
Η7
Detector not available F3
ΧΦ(turned at
Optical fiber Fault F2 and F3
180°)
Ε7(turned at 180°)
Optical fiber Fault F2 and F3
If the system is fitted with the CD Writer (optional), this lets you save the images in a study in “raw”
format to CD; these images can then be used for various purposes, such as display on another HIRIS
RF43 system.
3) The CD-Writer menu now opens with the new directory containing the selected study(ies) ready for
transfer to CD:
4) If necessary, click on Options to open the CD Writer options menu (see note 1) and set the options
as required.
Note 1: The CD-Writer menu lets you set a few options (using the Options menu):
We strongly advise that you update the spare hard disk after every change to the application SW or
the system setup data. This means making a complete backup of the system hard disk on the spare
HD.
To do this:
Close the application and reboot the system as Administrator (see para. 7.3 in Part 2). Enter:
3) Click on Backup.
4) Click on Next.
The archive hard disk now contains a back-up copy of the system hard disk. You now need to
move this to the back-up hard disk.
To do this:
1) Switch the system off and replace the system hard disk (see procedure explained in paragraph 2.1
of Part 5).
2) Reboot the system with the back-up disk inserted, log in as administrator and then launch the
Acronis True Image application.
4) Click on Restore.
5) Click on Next.
Click on Reboot.
At the end of this procedure, the back-up hard disk will be updated with the same settings as the
system and so can be used when necessary.
Switch the video processor off, extract the back-up HD and return the system HD. Log in as User.
You need to use the DVD burner in the video processor to create the ghost disk for the video
processor system (back-up CD/DVD).
You therefore need an empty DVD (or a few CDs) on which to save the ghost image of the system
hard disk.
Note: The ghost disk must be created after installing the equipment (system fully configured) and
must be constantly updated (after any changes to the SW or system setup). Keep this ghost
disk in a safe place, ensuring it is always available to the technical service.
Only the “Administrator” can carry out the following operations. Close the application and reboot the
system as Administrator (see Part 2, paragraph 7.3 ). Enter:
- User name: Admin2
- Password: ats_PED.24
Note: There are two alternative applications for the back-up CD/DVD procedure (Norton Ghost or
Acronis True Image). Check which of these applications is loaded in the PID and then follow the
relevant instructions in the following paragraphs:
- 5.8.1 Creating a ghost disk using Norton Ghost
- 5.8.2 Creating a ghost disk using Acronis True Image
3) Click on Backup.
4) Click on Next.
Click on Retry.
20) Extract the DVD/CD, close the application and reboot the system as User.
The Dump function lets you analysis problems that have caused the video processor application to
crash.
After a crash, this function automatically generates a system status report, concentrating especially
on the stored data.
The following page appears on the monitor after a crash, showing the stages in data collection.
This operation lasts about 30 seconds on average (though it may take up to a minute in some cases).
The application then closes.
If you need to restart the system immediately, use the “Cancel” button to abort the operation; the data
will not be collected in this case.
Each Dump operation produces a directory and a .md5 file, stored within a single compressed .zip file.
The directory and the two files have the same date/time as the corresponding crash.
The video processor LOG files are automatically added to the Dump directories (apart from the
DICOM function LOG file). This makes it easier for the Technical Service to understand what
happened to cause a crash.
Note: Please provide the following data if you require analysis of a DUMP file:
- the .zip file corresponding to the date and time of the CRASH,
- the Settings folder,
- if necessary, the DICOM LOG folder (see specific paragraph).
The STUDY RECOVERY function lets you reconstruct a study that has become corrupted for some
reason. This function is useful in the following situations:
- Corruption or loss of an acquired image: typically, when the “Error loading exam local file”
message appears in the study frame.
- Corruption of the study data. The system indicates this defect by identifying the study as
“UNNAMED”.
This function can also be used to construct an archive, starting from just one folder for a single study.
• Right-click on the corrupted study in the Patient List and then select the Recover Study option:
Work folder: path containing the folder when the recovered study is to be saved. The default path
is “D:\Recovery”, though you can change this using the button. The work folder is always
created automatically, even if not specified.
o Keep corrupted files: if selected, this option lets you use partial decoding results for files with
the “.dcm” extension. This allows some of the parameters to be reused.
• Analyze: this command lets you analyse the files in the corrupted study.
Press this command to perform the following operations:
- creation of the recovery folders;
- copy and analysis of each file in the study.
More specifically, each time you press the Analyze command, a folder called
“yyyymmddhhmmss_Recovery” (where the suffix contains the date and time of creation) is
created in the Work folder together with a series of subfolders, like those of an archive.
• Recover: use this command for the actual recovery procedure. This is only enabled once the
files have been analysed.
The recovery procedura uses the information in the study files.
The Recovery box contains a series of options that let you set certain data that cannot be found
directly from the corrupted study:
• Single fr. modality (CR/DX): to select the DICOM mode to be assigned to a series of single
frames.
• Single fr. equalizer: to select the equalization type to be associated to a series of single
frames. (*)
• Single fr. harm: to select the harmonization type to be associated to a series of single frames.
(*)
• Multi fr. modality (RF/XA): to select the DICOM mode to be assigned to a series of multiple
frames.
• Multi fr. equalizer: to select the equalization type to be associated to a series of multiple
frames. (*)
• Multi fr. harm: to select the harmonization type to be associated to a series of multiple
frames. (*)
• Multi fr. sub equalizer: to select the pre-equalization type to be associated to a series of
multiple frames acquired in SUB mode. (*)
(*) Note: You can select one of the set options in the system; the first available option in
alphabetical order is automatically shown.
• The Show report command lets you view the procedure reports, indicating both the analysis
results and the actual recovery results.
After using the Recovery procedure, open the recovered study as follows:
• Open the recovered archive by using the Browse command in the “Archive List Manager” panel;
Find the recovered study within the “Work folder” (e.g.: “D:\Recovery”) and the subfolder, whose
name indicated the date and time it was created using the recovery procedure.
CONTENTS
Optional devices:
X-ray tube 8 13 13 13
3b
10 11
7 12
X-ray
2 18
9 14 15
1a
5
19
3a 13 13 13
3d
3c
4
16 17
3e
1b
1c
When using the Wireless DRX1 detector, you must bear in mind that:
The detector battery charger is not a medical device and must not be placed in the
vicinity of the patient.
The Wireless Access Point is not a medical device and must not be placed in the vicinity
of the patient.
Keep all electronic devices (wireless or connected with cables) at least 1 metre from the
detector when in use.
The detector must not be used to support the patient: it must be placed on a suitable
surface, such as a table or the floor, before resting the patient on it. The label below
indicates the max limits to avoid damaging the detector:
Observe the following weight limits to prolong the life of the detector and reduce the risk of internal
damage:
- The max weight concentrated on a small area on the surface of the detector (50 mm in
diameter) must not exceed 23 kg.
- The max weight distributed evenly across the surface of the detector must not exceed 125 kg.
See the instructions and warnings in paragraph 2.4 of this part of the manual for details on
how to clean the detector.
Place the detector inside a protective plastic bag if there is any risk of liquids coming into
contact with this. In this case, remove the protection immediately after use to avoid the risk
of the detector overheating.
When using the detector “off-grid” (Potter-Bucky grid), adjust the X-ray collimator to limit
the X-ray beam so that it remains within the sensitive surface of the detector.
Never allow the detector to come into direct contact with the patient when connected via
cable (i.e. via the tether box).
BATTERY
Thickness : 6.7 mm
Weight: 0.3 kg
DETECTOR
Thickness : 15.5 mm
Weight: 3.5 kg
Weight: 0.386 Kg
Weight: 1.04 Kg
Weight: 2 kg
Weight: 2.5 kg
Box Contents
1 PIXIUM RF4343 detector
2 Cabinet (Image processor / Main Controller module / PU 4343 module / Detector power
unit / Power unit)
3 - Ionising chamber
- WI FI Access Point
- Cables
- Alphanumerical keyboard and Mouse
- Manuals
- Dose Area Meter (optional)
- CD/DVD recorder (optional)
- IR remote control (optional)
- Tether Box (optional)
4 Monitor
5 Cabinet top
6 - WIRELESS DRX1 detector
- Battery charger
- Battery (n.2)
box 6
box 2
box 1
System with
1 x PIXIUM RF4343 detector and
1 x Wireless DRX1 detector
The mobile Wireless DRX1 detector unit does not require installation.
- it can be fed via a power cable, i.e. via the tether box (optional).
Given the max length of the cable between the detector and the tether box (max 7.6m), the box
must be placed outside the cabinet.
X-Ray tube
Detector
The Access Point used for the detector’s wireless connection must be placed inside the area where
the detector is used. The distance between the Access Point and the detector is not critical, provided
there are no obstacles between them.
Weight: 0.386 kg
Weight: 1.04 kg
Ethernet port
Reset button
Power socket
LED
LEDs The LEDs indicate the status of the power supply and the activity of the Access Point.
(see paragraph 4.5 for details of these LEDs).
Ethernet port Port used for connection to the connector belonging to the Ethernet POE Switch
(Power Over Ethernet) providing power.
Reset button Use to reset the default settings in the event of a malfunction. Press for at least 5
seconds.
Power socket Power supply socket (NOT USED)
Console port
Ethernet port
Power socket
LED The LEDs indicate the status of the power supply and the activity of the Access Point.
(see paragraph 4.5 for details of these LEDs).
Ethernet port Port used for connection to the connector belonging to the Ethernet POE Switch
(Power Over Ethernet) providing power.
Power socket NOT USED
Console port NOT USED
We recommend connecting the Ethernet cable and then checking the efficiency of the Access Point
status LEDs before fixing the Access Point in a position that is not easily accessible.
Use the 2 screws and 2 wall plugs provided to fix the A.P.
Hang the A.P. on the screws using the appropriate bracket (see fig.1 in paragraph 2.1.1.1).
Connect the Ethernet cable supplied with the system (code CAVORJ45-03) to the Ethernet port on the
AP.
B P
2 WI FI Access Point LAN Mains cable / Ethernet POE switch connection: LAN1
Fig.1
Wireless DRX1
detector
Fig. 2a Fig. 2b
DLINK WI FI Access Point CISCO WI FI Access Point
Fig. 4
Tether Box
(optional)
COLOR
WORKING MONITOR
VIDEO SIGNAL
DVD-
RECORDER
(*)
230Vac
21Vac
RGBC 01 BOARD CM2 CM1
CM2 CM1
EXTENDER
RECEIVER
DVI/VGA CONVERTER
KEYBOARD
B12-C1 B11-C1 B10-C1
MOUSE B9-C1
DICOM B11-C3
NETWORK CM5
B8-C1
B11-C2 B14-C2 B2-C1
INFRARED
REMOTE I/O RS-232 DAM
RECEIVER (*)
NET.1
230Vac
B1-C1
DATA LINK
PU4343
+24V OUT
B6-C2
+24V IN
RS-232
CP9
CAN BUS
CP10 LAN TETHER
INJECTOR POWER SUPPLY BOX (*)
CP12
TB3 ON/OFF CP13
TB3
TB1 TB3
PS
BATTERY
CHARGER
MAIN 230Vac
always present
MAIN 230Vac
MAIN 230Vac
always present
If the optional “second ionising chamber” is fitted, the following electrical connections are foreseen in
the factory:
From To
1 MCON01 - CM14 ICI 00-CM2 Connection to board ICI 00 for the first ionising
(parallel connection on first chamber chamber (PIXIUM RF43 detector).
both boards ICI00) Board ICI 00-CM2 Connection to board ICI 00 for the second
second chamber ionising chamber (Wireless DRX1 detector).
2 ICI 00 – CM1 first ICI 00-CM1 second Power to board ICI 00 for the second ionising
ionising chamber ionising chamber chamber.
From To
1 ICI 00 –CM3 second Preamplifier for the Connection to the preamplifier for the second
ionising chamber second ionising ionising chamber.
chamber
nd
2 ionising chamber
There are three network connections in a system fitted with two detectors:
- DICOM Network to connect to the local hospital network
- PU Network to connect to the PIXIUM RF4343 detector unit
- WIRELESS Network to connect to the Wireless DRX1 detector unit
PU NETWORK
The following setup procedure is normally only required after upgrading the system (single or
double detector) or after replacing the video processor.
If the SW version for the THALES detector is 2.3.4.0 or higher, the 2 network ports (PU and
WIRELESS) may be configured each with a separate IP ADDRESS:
Otherwise, you must give them the same IP ADDRESS = 10.0.1.251, in order to guarantee that the 2
network connections (PU Network and WIRELESS Network) work correctly.
In the latter case, follow the steps below to configure the network connections:
Before attempting to change the IP address of the video processor above, you must
check that IP Address for the PU RF4343 is IP 10.0.1.2.
(If this is not the case, change it as described in paragraph 2.7 in Part 5 of this manual).
• With the cabinet switched off, disconnect the cables attached to the network ports shown
in the figure below.
• Switch the system on and access the video processor, logging on as Administrator:
username Admin2, password ats_PED.24 (see paragraph 7.3 in Part 2).
The two network connections (PU NETWORK and WIRELESS NETWORK) are now set up
with the same IP ADDRESS = 10.0.1.251.
• Select Start / Shut Down / Shut Down in the Windows task-bar and then switch off the
cabinet.
• Reconnect the network connections you removed at the beginning of the procedure to
their respective ports.
In any case, the components of the DRX1 Wireless detector unit will have the same address
type:
- DRX1 Wireless detector = 10.0.1.150
- WI FI Access Point = 10.0.1.249
The figure and tables below show the different options in the General Setup menu for using the
Wireless F.P. detector:
Hardware Settings
Settings
IP AccpBox To set the IP Address of the PU4343 module.
With THALES SW version prior to 2.3.4.0, set the IP address as
10.0.1.2 (see para..2.3.1 above for further details)
Otherwise, with THALES SW version 2.3.4.0 or higher, set the IP
address as given by the manufacturer: 192.168.0.2
Detectors
Settings
Enable Detector 2 Select to enable the 2nd detector and then select the type:
WIRELESS FP
Enable Detector Warnings Select to enable the detector alarm warnings.
Detectors 2 Params Setup Open this menu to set the working parameters for the Wireless DRX1
detector (see paragraph 2.3.3 below).
This lets you define the parameters used by the image acquisition system with the Wireless DRX1
detector.
These parameters are already set in the factory with the following values.
• Battery Low Threshold Battery charge min threshold. The detector status window in the
working frame reads “Low” when the charge level drops below this
value. Typical value: 30
• Sensibility Detector sensitivity: this value must coincide with that shown on the
relevant document in the detector technical bundle.
Do not change.
• Use Preview To use the Preview Image option during acquisition of an image by the
detector. Typically: enabled
• Enable Notify To enable the display of the Gain Calibration request when this has
Rq. Gain Calib expired and the detector needs to be re-calibrated.
Typically: enabled
• Enable Notify To enable the display of the mechanical shock warnings in the
Drop Det. detector status area. Typically: enabled
• Drop Info To open the Drop Information page with a list of the shocks to the
detector:
The system logs each mechanical shock (e.g. a drop or impact), classing them by the force
(magnitude) of the shock (G).
Each type of shock is saved with a number and the date of the event (year : month : day :: hour :
minutes).
Resettable Shock Magnitude Last logged shock. This is reset with each gain calibration.
Access the management function for exam cards linked to the Wireless DRX1 detector by selecting
Option / Setup / WD Exam Setup and then entering the technical Password. The WD Exam Setup
Manager menu opens:
The Exam Setup cards for exposure using the Wireless DRX1 detector are numbered with:
- the anatomical part, with an ID from 100 to 199,
- the projection type, with an ID from 00 to 99.
The “Go to card” function lets you select the card and open the relevant Exam Setup menu. This has
four groups of parameters, as shown below:
Miscellaneous
parameters
You need to check which parameters are free for each exam and change these as required in the
exam card.
Settings Notes
To enable the patient Select at least one size. When a size is not
Size valid types for the exam. enabled (i.e. not
checked), the entire
column is disabled.
Miscellaneous
Properties
Settings Notes
*Image orientation:
Use to set the default Select one of the given This parameter is only
Modality DICOM mode. fields: used when the mode used
CR to create the exam is not
DX already defined upon
receipt of the Worklist.
Protocol Procedure ID code To be agreed with the
network administrator.
Description Description of the To be agreed with the
exam. network administrator.
Body part to which the Select one of the given
Body Part exam refers. fields:
UNDEFINED
SKULL
CSPINE
TSPINE
LSPINE
SSPINE
COCCYX
CHEST
CLAVICLE
BREAST
ABDOMEN
PELVIS
HIP
SHOULDER
ELBOW
KNEE
ANKLE
HAND
FOOT
EXTREMITY
LEG
HEAD
HEART
JAW
ARM
NECK
Projection side Select one of the given
Laterality fields:
Undefined
Right
Left
Unpaired
Both L&R
Projection type Select one of the given
View fields:
Undefined
AP
PA
LL
RL
RLD
LLD
RLO
LLO
Note: In the case of an integrated X-ray generator, the kV, mAs and Focal Spot parameters are
automatically sent to the X-ray generator.
The values of the mAs parameter set here must match that envisaged by the X-ray generator
scales.
Even if the generator is not integrated, it is important that these parameters are set
correctly, as they are a guide for the operator, who can view them by pressing a symbol
in the working frame:
Automatic enlargement
of the collimated images
Fill Size to fit the max size Select to suit exam
permitted by the
monitor.
1.0 = full screen (entire
detector area).
3.1 INTRODUCTION
The Wireless DRX1 detector must be calibrated during installation and at regular intervals (at least
once every 6 months).
This basically consists in updating the gain parameters associated to each pixel of the detector and in
identifying any defective pixels.
Calibration involves acquiring 16 images at the default dose value for the detector (26 μGy),
following a guided procedure.
This calibration is done using 4 X-ray exposures at different default doses set by the manufacturer
of the detector following a guided procedure.
Calibration and the detector checks are done using the CareCal.exe software provided with the
system.
Calibration is done using the CareCal program, which acts as detector/user interface. This involves
sending calibration commands and displaying the resulting images.
The main working frame opens: the centre section shows the images (calibration and acquisition),
while the side and bottom sections contain 5 tabs to view the detector status and setup parameters.
The top section of the main frame also contains a menu bar and toolbar.
In order to calibrate the detector, the video processor and the detector must be connected. To do this:
1) Select Tools / Options in the menu bar to open the window that lets you set the serial connection
parameters with the system.
3) Select Connection / Connect in the menu bar to open the window that lets you set the connection
parameters for the Wireless DRX1 detector.
5) Wait about 15 seconds until the video processor connects to the detector. Connection is successful
when the following data appear in the bottom left-hand bar:
1) Select the “ Detector Status ” tab and check the status of the detector:
2) Then select the “ Power ” tab and check the detector power supply in the 2 cards (Battery and
Voltages):
1) Select the “ Frame ” tab in the Configuration frame to view some basic info on the acquired
images:
2) Then select the “ Misc ” tab in the same window to view the main wireless connection parameters:
3) Then select (always from the same window) the “ General ” tab and check that the settings are the
same as those shown below:
Use the Calibration / Verify function to check for the existence of any calibration files in the folder on
the system hard disk (C:\IMAGERs).
Note: This function only checks for the existence of calibration files; it does not validate their content.
These stages must be done in order, as each stage requires that the previous one has already been
completed.
st
3.5.1 1 STAGE: DAILY DARK CALIBRATION
OFFSET calibration requires the acquisition of an image without any X-ray emission.
- Click on Exit.
3.5.2 FINDING THE EXPOSURE VALUES FOR GAIN AND FLAT FIELD CONFIGURATION
For GAIN and FLAT FIELD calibration you need to know beforehand the useful radiological
parameters for exposing the detector to the required doses during this calibration.
To find these parameters (X-ray filter and X-ray generator exposure values), use the CareCal function
that lets you measure the levels of the acquired image (and thus the exposure dose). To do this:
4) The image is shown a few seconds later in preview format and then in full size format (scale 1:1).
See image below:
5) Move the Zoom slider left to view the full image (scale 1:6).
6) Use the commands in the Display control area to get the best view of the acquired image:
- enter the max value of the window to be displayed in the Max box. For example, a value (LSB)
twice the foreseen acquisition levels.
- the Display enh slider lets you fine adjust the display using your mouse.
7) Move the mouse pointer onto the image and move this within the image to check the pixel levels.
These are shown in the bottom bar, with the corresponding x-y position in the detector matrix:
8) If necessary, you can change the exposure parameters to get the required image levels. These are
listed in the tables below:
2700 - 3300
Exposure 70kV, 25 mAs
(typ. 3000)
2700 - 3300
Exposure 1 70kV, 25 mAs
(typ. 3000)
900 - 1100
Exposure 2 70kV, 8 mAs
(typ. 1000)
145 - 165
Exposure 3 70kV, 1.3 mAs
(typ. 150)
5400 - 6600
Exposure 4 70kV, 50 mAs
(typ. 6000)
9) Make a note of the radiological parameters used (X-ray filter and generator exposure values).
You will be asked to set these during the guided calibration procedure.
nd
3.5.3 2 STAGE: GAIN CALIBRATION
GAIN calibration involves 16 X-ray exposures at a fixed dose, set by the system. Conditions:
- homogeneous X-ray filtering,
- collimator open enough to radiate the entire detector.
These two conditions are indispensable for both correct GAIN calibration and correct detection of
defective pixels.
The X-ray dose must also be that required, i.e. sufficient to generate detector image levels within the
set range. Otherwise calibration may not be effective.
For this reason, you must find the correct dose before starting the calibration procedure by acquiring
several images and then reading their level (LSB), as described in paragraph 6.5.2 above.
- Click on Start.
3) When the system is ready for image acquisition, the Exposure setup window appears showing the
dose level needed for X-ray exposure and the status of the detector.
4) Press the PREP + RAD commands on the X-ray generator to start exposure:
- The Exposure setup page automatically closes upon completion of image acquisition.
- The Detector status: Acquiring image message in the GAIN calibration frame informs you that
the detector is transferring the image to the video processor.
6) Repeat steps 3, 4 and 5 for the other exposures, at the same X-ray dose.
- Click on Exit.
rd
3.5.4 3 STAGE: FLAT FIELD CALIBRATION
FLAT FIELD calibration involves 4 X-ray exposures, each at a different fixed dose set by the system.
Conditions:
- homogeneous X-ray filtering,
- collimator open enough to radiate the entire detector.
As for GAIN calibration, these two conditions are indispensable for correct completion of the FLAT
FIELD calibration procedure.
The X-ray doses must also be those required, i.e. sufficient to generate detector image levels within
the set range.
For this reason, you must find the correct dose before starting the calibration procedure by acquiring
several images and then reading their level (LSB), as described in paragraph 6.5.2 above.
- Click on Start.
3) When the system is ready for image acquisition, the Exposure setup window appears showing the
dose level needed for the first X-ray exposure and the status of the detector.
4) Press the PREP + RAD commands on the X-ray generator to start exposure:
- The Exposure setup page automatically closes upon completion of image acquisition.
- The Detector status: Acquiring image message in the FLAT FIELD calibration frame informs
you that the detector is transferring the image to the video processor.
6) Repeat steps 3, 4 and 5 for the next three exposures each at a different X-ray dose value.
- Click on Exit.
th
3.5.5 4 STAGE: DEFECT MAP ANALYSIS
After completing GAIN and FLAT FIELD calibration, you can check the new defective pixel map:
- Click on Exit.
After detector calibration the defect pixel map image shows the position of the defective pixels on the
surface of the detector.
You can view this map by using the Retrieve combined defect pixel map function in the CareCal
application, as follows:
1) Select Acquisition / Retrieve image / Retrieve combined defect pixel map in the menu bar:
2) The map is sent from the detector and shows the combined defect map in the centre of the
frame:
3) Move the Zoom slider in the Display control section to the left to view the entire image. Set 0 in
the Max field to view all defective pixels on the map.
This function lets you view the status of the detector in terms of any mechanical shocks it has
received.
Whenever the detector receives a mechanical shock (impact/knocks), this is logged with an indication
of the force (G) and the correct date (year : month : day :: hour : minutes).
2) The Detector shock information frame appears showing the number of shocks received (on the
basis of their force) and the date on which these were logged:
4 MAINTENANCE
The Wireless DRX1 detector may be used in direct contact with the patient and so needs to be
cleaned properly.
- Remove dust and dirt from the battery compartment using a soft cloth,
- Use a toothbrush or vacuum cleaner to clean the contacts on the side of the battery compartment
(alternatively, contact the Technical Service).
Each Wireless DRX1 detector is supplied with a DVD (Registration Disk) containing:
- the default calibration files,
- the detector firmware,
- file Registration.xml.
File Registration.xml contains the Serial Number and IP Address of the detector as set in the factory,
which are needed for its connection to the system.
The Wireless DRX1 detector’s IP Address is always 10.0.1.150 (all detectors), while its Serial Number
is unique and specific for each detector.
Directory: <DVD>:\Calibration\000000030054
(where the number 000000030054 is the serial number of the
detector) contains the default calibration files prepared in the
factory and needed to correct the images acquired by the
detector.
1) Do not insert the battery in the new detector until you have completed the configuration procedure
below.
2) Log in as the Administrator on the video processor (see Annex 7.3 in Part 2 of this manual).
4) Copy the new Registration.xml file from the DVD to the same directory: C:\IMAGERs\
thus overwriting the old file.
- select Option / Setup / General Setup and click on Detectors 2 Params Setup
- change the Serial Number and Sensibility parameters (see paragraph 2.3.2 in this part of the
manual) to suit those of the new detector (see its accompanying documents.
6) Close the HIRIS RF43, application and insert the charged battery in the detector. Wait for the
correct sequence of warning LEDs (see paragraph 4.4 below).
7) Open the CARECAL application and effect connection by inserting the new detector Serial
Number. Then perform the complete calibration procedure, as described in chapter 3 above.
Ethernet Switch model 3CNJ2000, on the other hand, requires the following setup operations:
- Temporarily change the IP Address for the Video Processor (PID) used for the
WIRELESS NETWORK, using an address compatible with that of the switch so that the
two devices can communicate.
- Then change the IP Address for the switch.
- End the procedure by restoring the original IP Address for the PID.
To do this:
1) Find the IP Address of the switch indicated in its installation guide (e.g.: 169.254.1.251).
2) With the system switched off, replace the old switch with the new one and then connect this as
explained in paragraph 2.2 in this part of the manual.
4) Switch the system on and log in as Administrator, keeping the Shift key on the alphanumerical
keyboard down when the system boots up (see paragraph 7.3 in Part 2 of this manual).
5) Change the IP Address for the Video Processor used for the WIRELESS NETWORK, by selecting
in the Windows task-bar:
Start / Settings / Network Connections / Wireless Connection
(see procedure in paragraph 2.3.1 in this part of the manual).
9) Select SYSTEM / LAN Settings in the next page to change the Switch’s network settings:
10) Change the parameters in the Change IP Address table as shown below and then click on
Apply.
12) The IP Address of the Ethernet POE Switch is now correctly set: 10.0.1.250
13) Restore the original working IP Address for the Video Processor used for the WIRELESS
NETWORK (changed earlier at point 5 above): 10.0.1.251
Run the HIRIS RF43 application and check that the DRX1 detector connects correctly with the system
(wireless connection).
To set the Wireless network parameters for the Access Point (AP):
- Change the IP Address of the Video Processor (PID) temporarily, using an address that
is compatible with that of the AP so that the two elements can communicate with each
other.
- Using the PID, change the IP Address of the AP.
- Then restore the initial IP Address of the PID.
To do this:
1) Find the IP Address of the AP indicated in its installation guide (e.g.: 192.168.0.50).
2) With the system switched off, replace the old AP with the new one and then connect this as
explained in paragraph 2.2 in this part of the manual.
4) Switch the system on and log in as Administrator, keeping the Shift key on the alphanumerical
keyboard down when the system boots up (see Annex 7.3 in Part 2 of this manual).
5) Change the working IP Address for the Video Processor used for the WIRELESS NETWORK, by
selecting in the Windows task-bar:
Start / Settings / Network Connections / Wireless Connection
(see procedure in paragraph 2.3.1 in this part of the manual).
8) The web browser opens automatically at the AP access page. Enter admin as the User Name,
leave the Password field blank and then click on Login:
10) Select MAINTENANCE / CONFIGURATION FILE to load the wireless network settings via a
configuration file saved on the video processor system disk:
11) You need to specify the path for the AP configuration file in order to load it. Click on Browse:
12) The Choose file window opens so you can find the file:
13) Select the path: My Computer / System (C:) / Program Files / AP Restore DAP2553
and then the file: config.dcf . Then click on Open.
The file path now appears in the main page:
14) After loading the configuration file, click on Upload and confirm with OK when requested:
15) The AP configuration Restore procedure starts. Do not switch the device off and wait for this
operation to finish (about 1 minute).
16) The following page appears once the Restore procedure is completed. You can now close the
configuration window.
17) To see if the network settings have been loaded correctly, access the AP with the new IP
Address. Then:
- Restore the original working IP Address for the Video Processor used for the WIRELESS
NETWORK (changed earlier at point 5 above): 10.0.1.251
- Start Windows Explorer selecting Start / Programs / Accessories / Windows Explorer than
enter the IP Address for the AP (10.0.1.249) in the Address bar again. Press Enter.
18) If the web browser automatically opens at the AP access page, it means that the wireless network
settings have been loaded correctly.
Enter admin as the User Name, leave the Password field blank and click on Login.
19) Select the BASIC SETTINGS / WIRELESS menu in the main page and check that the current
network is CSH_DRX1_FACTORY:
20) Click on Logout to disconnect the AP and then close the setup window.
22) Run the HIRIS RF43 application and check the WI-FI connection of the DRX1 Wireless detector
with the system.
5a 5b
1 4
LED flash
Description Action required
sequence
1 green Detector in Standby mode. No action needed.
Detector can be used.
4 blue Detector upgrading its firmware. Condition not usually visible when using the
detector normally.
Below is a summary of the typical conditions of the LED when the detector is working normally:
1 green Detector connected to the wireless Connection between the detector and the
access point, but not connected to system can only be made in this condition.
3 blue system yet.
If the system is switched off: use the main
switch in the electrical cabinet to switch it on.
STATUS LED
th
5 LED on = Battery fully charged
th
5 LED flashing = 80% charged
th
4 LED flashing = 60% charged
rd
3 ° LED flashing = 40% charged
nd
2 LED flashing = 20% charged
st
1 LED flashing = <20% charged
It takes less than 3 hours to fully charge a completely flat battery if the temperature of the
room where the charger sits is between +10 and +35°C.
Make sure that the battery temperature remains within the accepted range: 0 - 45°C. The
yellow LED will start flashing if the temperature is outside this range when you insert the
battery.
CONTENTS
X-ray tube 13 13 13
8
3b
10 11
7 12
X-ray
2 18
9 14 15
1a
5
3d
3a
13 13 13
3c
4
16 17
3e
1b
1c
When using the PIXIUM 3543EZ wireless detector, you need to bear in mind that:
The battery charger is not a medical device and so must not be placed in the vicinity of
the patient.
The Wireless Access Point is not a medical device and so must not be placed in the
vicinity of the patient.
Keep all electronic devices (wireless or connected via cable) at least 1 metre from the
detector when in use.
The detector must not be used to support the patient: the detector must be placed on a
suitable surface, such as a table, bed or the floor, before applying the weight of the patient.
The label shown here shows the max weight limits to avoid damaging the detector:
100 kg 150 kg
Respect the following weight limits to prolong the life of the detector and reduce any
potential risk of damaging its parts:
- Max weight concentrated in a small area of the surface of the detector (80 mm in
diameter): not more than 100 kg.
- Max weight evenly distrubted across the entire surface of the detector: not more than
150 kg.
See paragraph 4.1 in this part of the manual for details on how to clean the detector.
Place the detector in a protective bag if there is any risk of fluids coming into contact with
the detector. In this case, remember to remove the protective covering immediately after
use to prevent the detector from overheating.
Adjust the X-ray collimator carefully to limit the X-ray beam so that this falls only on the
sensitive surface of the detector when using the detector without a potter-bucky.
diameter length
description code
[mm] [m]
BATTERY
Thickness: 7 mm
Weight: 0.2 kg
DETECTOR
Thickness: 16 mm
Weight: 2.6 kg
Weight: 0.386 kg
Weight: 1.04 Kg
The battery charger is not a medical device and so must not be placed in the vicinity of the patient.
Simply place it on a workbench near a 100-240V AC, 50/60Hz mains socket (max consumption: 1A).
BATTERY CHARGER
Weight: 0.5 Kg
Box Contents
4 3 5
6
2
1
System fitted with
1 PIXIUM RF4343 detector and
1 PIXIUM 3543EZ wireless detector
The mobile PIXIUM 3543EZ wireless detector unit does not require installation.
Its overall dimensions and weight are shown in the figure below:
BATTERY
Thickness : 7 mm
Weight: 0.2 kg
DETECTOR
Thickness : 16 mm
Weight: 2.6 kg
The figure below shows the active area of the PIXIUM 3543EZ wireless detector:
The Access Point used for the detector’s wireless connection must be placed inside the area where
the detector is used. The distance between the Access Point and the detector is not critical, provided
there are no obstacles between them.
Weight: 0.386 kg
Weight: 1.04 Kg
Ethernet port
Reset
Power socket
LEDs The LEDs indicate the status of the power supply and the activity of the Access Point.
(see paragraph 4.5 in this part of the manual for details of these LEDs).
Ethernet port Port used for connection to the connector belonging to the Ethernet POE Switch
(Power Over Ethernet) providing power
Reset Use to reset the default settings in the event of a malfunction. Press for at least 5
seconds.
Power socket Power supply socket (NOT USED)
Console port
Ethernet port
Power socket
LED The LEDs indicate the status of the power supply and the activity of the Access Point.
(see paragraph 4.5 for details of these LEDs).
Ethernet port Port used for connection to the connector belonging to the Ethernet POE Switch
(Power Over Ethernet) providing power.
Power socket NOT USED
Console port NOT USED
We recommend connecting the Ethernet cable and then checking the efficiency of the Access Point
status LEDs before fixing the Access Point in a position that is not easily accessible.
Use the 2 screws and 2 wall plugs provided to fix the A.P.
Hang the A.P. on the screws using the appropriate bracket (see fig.1 in paragraph 2.4.1.1).
Connect the Ethernet cable supplied with the system (code CAVORJ45-03) to the Ethernet port on the
A.P.
B P
2.2 ELECTRICAL CONNECTIONS FOR THE PIXIUM 3543EZ WIRELESS DETECTOR UNIT
WI FI Access Point LAN Mains cable / Ethernet POE switch connection: LAN1
Fig. 1a Fig. 1b
DLINK WI FI Access Point CISCO WI FI Access Point
Fig. 2
Power supply: 230V AC / 48V DC power unit connection
Ethernet POE Switch
If the optional “second ionising chamber” is fitted, the following electrical connections are foreseen in
the factory:
DA: A:
1 MCON01 - CM14 ICI 00-CM2 Connection to board ICI 00 for the first ionising
(parallel connection on first chamber chamber (PIXIUM RF43 detector).
both boards ICI00) Board ICI 00-CM2 Connection to board ICI 00 for the second
second chamber ionising chamber (PIXIUM 3543EZ wireless
detector).
2 ICI 00 – CM1 first ICI 00-CM1 second Power to board ICI 00 for the second ionising
ionising chamber ionising chamber chamber.
DA: A:
1 ICI 00 –CM3 second Preamplifier for the Connection to the preamplifier for the second
ionising chamber second ionising ionising chamber.
chamber
nd
2 ionising chamber
There are three network connections in a system fitted with two detectors:
- DICOM Network to connect to the local hospital network
- PU Network to connect to the PIXIUM RF4343 detector unit
- WIRELESS Network to connect to the PIXIUM 3543EZ wireless detector unit
PU NETWORK
The following setup procedure is normally only required after upgrading the system (single or
double detector) or after replacing the video processor.
If the SW version for the THALES detector is 2.3.4.0 or higher, the 2 network ports (PU and
WIRELESS) may be configured each with a separate IP ADDRESS:
Otherwise, you must give them the same IP ADDRESS = 10.0.1.251, in order to guarantee that the 2
network connections (PU Network and WIRELESS Network) work correctly.
In the latter case, follow the steps below to configure the network connections:
Before attempting to change the IP address of the video processor above, you must
check that IP Address for the PU RF4343 is IP 10.0.1.2.
(If this is not the case, change it as described in paragraph 2.7 in Part 5 of this manual).
• With the cabinet switched off, disconnect the cables attached to the network ports shown in the
figure below.
• Switch the system on and access the video processor, logging on as Administrator: username
Admin2, password ats_PED.24 (see paragraph 7.3 in Part 2).
The two network connections (PU NETWORK and WIRELESS NETWORK) are now set up
with the same IP ADDRESS = 10.0.1.251.
• Select Start / Shut Down / Shut Down in the Windows task-bar and then switch off the cabinet.
• Reconnect the network connections you removed at the beginning of the procedure to their
respective ports.
In any case, the components of the PIXIUM 3543EZ wireless detector unit will have the
same address type:
- PIXIUM 3543EZ wireless detector = 10.0.1.90
- WI FI Access Point = 10.0.1.249
The figure and tables below show the different options in the General Setup menu for using the
PIXIUM 3543EZ wireless detector:
Hardware Settings
Settings
IP AccpBox To set the IP Address of the PU4343 module.
With THALES SW version prior to 2.3.4.0, set the IP address as
10.0.1.2 (see para..2.3.1 above for further details)
Otherwise, with THALES SW version 2.3.4.0 or higher, set the IP
address as given by the manufacturer: 192.168.0.2
Detectors
Settings
Enable Detector 2 Select to enable the 2nd detector and then select the type:
WIFI 3543EZ
Enable Detector Warnings Select to enable the detector alarm warnings
Detectors 2 Params Setup Open this menu to set the working parameters for the PIXIUM 3543EZ
wireless detector (see paragraph 2.3.3 below)
This lets you define the parameters used by the image acquisition system with the PIXIUM 3543EZ
wireless detector.
These parameters are already set in the factory with the following values.
• Sensibility Detector sensitivity: this value must coincide with that shown on
the relevant document in the detector technical bundle.
Do not change.
• Disable All Correction To disable all the corrections applied to the image acquired by the
detector.
Do not select.
• Disable Echo Line Correction To disable the Echo Line correction applied to the image acquired
by the detector.
Do not select.
• Enable Preview Image To enable the Preview Image option during acquisition of an image
by the detector.
Typically: enabled.
Access the management function for exam cards linked to the PIXIUM 3543EZ wireless detector by
selecting Option / Setup / WD Exam Setup nd then entering the technical Password. The WD Exam
Setup Manager menu opens:
The Exam Setup cards for exposure using the PIXIUM 3543EZ wireless detector are numbered with:
- the anatomical part, with an ID from 100 to 199,
- the projection type, with an ID from 00 to 99.
The “Go to card” function lets you select the card and open the relevant Exam Setup menu. This has
four groups of parameters, as shown below:
Miscellaneous
parameters
You need to check which parameters are free for each exam and change these as required in the
exam card.
Settings Notes
To enable the patient Select at least one size. When a size is not
Size valid types for the exam. enabled (i.e. not
checked), the entire
column is disabled.
Miscellaneus
Properties
Settings Notes
* Image orientation:
Use to set the default Select one of the given This parameter is only
Modality DICOM mode. fields: used when the mode used
CR to create the exam is not
DX already defined upon
receipt of the Worklist.
Protocol Procedure ID code To be agreed with the
network administrator.
Description Description of the To be agreed with the
exam. network administrator.
Body part to which the Select one of the given
Body Part exam refers. fields:
UNDEFINED
SKULL
CSPINE
TSPINE
LSPINE
SSPINE
COCCYX
CHEST
CLAVICLE
BREAST
ABDOMEN
PELVIS
HIP
SHOULDER
ELBOW
KNEE
ANKLE
HAND
FOOT
EXTREMITY
LEG
HEAD
HEART
JAW
ARM
NECK
Projection side Select one of the given
Laterality fields:
Undefined
Right
Left
Unpaired
Both L&R
Projection type Select one of the given
View fields:
Undefined
AP
PA
LL
RL
RLD
LLD
RLO
LLO
Note: In the case of an integrated X-ray generator, the kV, mAs and Focal Spot parameters are
automatically sent to the X-ray generator.
The values of the mAs parameter set here must match that envisaged by the X-ray generator
scales.
Even if the generator is not integrated, it is important that these parameters are set
correctly, as they are a guide for the operator, who can view them by pressing a symbol
in the working frame:
Automatic enlargement
of the collimated images
Fill Size to fit the max size Select to suit exam
permitted by the
monitor.
1.0 = full screen (entire
detector area).
3.1 INTRODUCTION
The PIXIUM 3543EZ wireless detector is normally calibrated in the factory to check its efficiency and
functioning.
You do, however, need to update the calibration during installation in order to adapt this to the actual
X-ray beam conditions in the radiological system.
This basically consists in updating the gain parameters associated to each pixel of the detector and in
identifying any defective pixels.
There are 2 stages to the calibration procedure, automatically carried out in this order:
The system-guided calibration procedure therefore involves 10 exposures at the same dose (Gain)
and 3 exposures at differing doses (Defect Map).
The ThalesCal software provided with the system is used to calibrate and check the detector.
The PIXIUM 3543EZ wireless detector is calibrated using the ThalesCal program, which basically acts
as detector/user interface. This involves sending calibration commands and displaying the resulting
images.
In order to calibrate the detector, the video processor must be connected to the detector. After running
the ThalesCal application, proceed as follows:
2) Check that the settings are the same as those shown in the figure below and then click on OK.
5) Wait about 15 seconds for the processor to connect with the detector. Connection is successful
when the application displays the updated detector data in the central part of the screen:
1) Select the “System Status” tab in the left-hand box and check the detector status:
2) Select the “Acquisition Mod ” tab in the same box and check that the settings are the same as
those shown below:
Before actually calibrating the detector, we recommend checking its correct setup:
1) Select the “Logical mode configuration” tab in the right-hand box and check that the current
Logical Mode is only the first in the list (Logical mode 0):
You must first activate the logical mode before you can calibrate this.
To do this:
1) Select the “Logical modes manager” tab in the right-hand box: the only available Logical Mode
appears:
- Select the Logical Mode with your mouse and then right-click to select this command:
Activate logical mode:
- Select the Logical Mode with your mouse and then right-click to select this command:
Deactivate logical mode:
To calibrate the detector, you need to know beforehand the useful radiological parameters for
exposing the detector to the required doses during calibration.
The table below lists the X-ray parameters used in the factory. These may be used for calibration
purposes provided you keep to the same filtering and focal distance (SID) conditions.
To find these parameters (X-ray filter and generator exposure values), you can use a function that lets
you measure the levels of an acquired image (and hence the exposure dose). To do this:
1) Select the “Logical modes manager” tab in the right-hand box and activate the Logical Mode (see
paragrapg 3.5 above):
- image reception
- end of acquisition
5) The image is then shown (enlarged) after a few seconds in the central “LastFrame“ box. Turn the
central scroll wheel on the mouse to view the full image:
6) If you want a clearer view of the image: select the “Properties” tab in the left-hand box and then
use your mouse to move the slider in the High Limit (Contrast window) field:
8) If necessary, you can change the exposure values to get the image levels you require.
9) Make a note of the radiological parameters used (X-ray filter and generator exposure values), as
the system will ask you to set these during the calibration procedure.
These two conditions are indispensable for correct GAIN calibration and the correct detection of any
defective pixels.
The exposure dose must be that specified, i.e. enough to generate image levels within the set range.
Otherwise calibration may not be effective.
You must therefore find the dose level before starting the calibration procedure: acquire some images
and read their level (LSB), as described in paragraph 3.6.1 above.
To calibrate:
1) Select the “Logical modes manager” tab in the right-hand box and activate the Logical Mode
(see paragraoh 3.5 above):
3) Take your mouse pointer to the Logical Mode you want to calibrate and then right-click and select
the “X-Ray calibration” command:
4) Click on Start in the Xray calibration page that now appears to start the actual calibration
procedure:
5) The Exposure setup window appears when the system is ready for acquisition of the 10 GAIN
images, indicating the dose level required for X-ray exposure (25uGy/frame):
6) After setting the exposure parameters (see previous paragraph) on the X-ray generator for the 10
GAIN acquisition, press the PREP + RAD commands (exposure is not activated) to continue.
Press the PREP + RAD commands (within 6 sec) to get X-ray exposure.
Press Yes to repeat the exposure (without cancelling the entire calibration).
10) Once you have acquired the 10 GAIN images, the “Exposure setup” page appears, showing the
parameters previously found for X-ray exposure of the DEFECT MAP 1 (2.5uGy/frame):
11) After setting the correct new parameters on the X-ray generator, press the PREP + RAD
commands (exposure is not activated) to continue.
12) The system now acquires a single image at the set dose.
Press the PREP + RAD commands (within 6 sec) to get X-ray exposure.
14) Repeat steps 10, 11, 12 and 13 above for the other 2 DEFECT MAP exposures, at different dose
values:
15) Once you have acquired the last image, the Calibration terminated successfully message
appears if the calibration has been successful:
16) Nel riquadro di destra Select the “Calibration” tab in the right-hand box and check that the date
and time of the new calibration appear next to the calibrated Logical Mode:
After calibrating the detector, the defect pixel map image shows the position of any bad (defective)
pixels on the surface of the detector.
2) The map for the active Logical Mode is received from the detector within a few seconds and then
displayed in the central DefectMap box:
3) To view all the bad pixels in the map, select the Properties tab in the left-hand box and then:
- set: “False” in the Show low saturation and Show high saturation fields
4 MAINTENANCE
The PIXIUM 3543EZ wireless detector may be used in direct contact with the patient and so needs to
be cleaned properly.
- Disconnect the detector from its power supply by removing its battery.
- Remove dust and dirt from the battery compartment using a soft cloth,
- Use a toothbrush or vacuum cleaner to clean the contacts on the side of the battery compartment
(alternatively, contact the Technical Service).
Each PIXIUM 3543EZ wireless detector is supplied with a conformity assessment document
containing the detector data (needed for its setup):
- its serial number ( e.g.: 1241B0 )
- its sensitivity ( e.g.: 0.58 580 lsb/µGy )
The IP Address for the static PIXIUM 3543EZ wireless detector set in the factory is always
192.168.1.3 (all detectors), which needs to be changed in the case of equipment fitted with two
detectors, using another IP Address compatible with the Wireless Network in the system to ensure
that there is no conflict with the IP Address used for the dynamic PIXIUM RF4343 detector.
1) Switch the equipment on and log in on the video processor as the Administrator (see Annex 7.3 in
Part 2 of this manual).
2) Temporarily disconnect the cables from the DICOM NETWORK, PU NETWORK and WIRELESS
NETWORK sockets and check that the Access Point is off.
3) Connect the “back-up cable” (supplied with the detector) between the detector and the WIRELESS
NETWORK (leave the third circular connector free).
4) Insert a fully charged battery in the detector and wait a minute or so for this to come on properly.
5) Change the IP Address for the Video processor used for the WIRELESS NETWORK by selecting,
in the Windows task-bar:
Start / Settings / Network Connections / Wireless Connection
(see procedure in paragraph 2.3.1 in this part of the manual).
11) Edit the Detector3543EZ.ini file in this folder: “C:\Documents and Settings\All Users\
Application Data\Digitec\Pixrad\Pixarad-3.50.1\config\digitec_3543EZ”,
and set the correct sensitivity parameter (found in the detector documents). Save the file.
12) Open the HIRIS RF43 application and carry out these steps:
- select Option / Setup / General Setup and then the Detectors 2 Params Setup button,
- change the Sensibility (see paragraph 2.3.3 in this part of the manual) using that of the new
detector.
⇒ 1° file: xxx_DefectMapRef_DetMode3_0.fxd
(es: 1241B0_DefectMapRef_DetMode3_0.fxd 1325G4_DefectMapRef_DetMode3_0.fxd)
⇒ 2° file: xxx_ GainLinRef _DetMode3_0.fxd
(es: 1241B0_GainLinRef_DetMode3_0.fxd 1325G4_ GainLinRef_DetMode3_0.fxd)
14) Restore the original working IP Address for the Video Processor used for the WIRELESS
NETWORK (changed at point 5 above): 10.0.1.251
After setting the new IP ADDRESS for the detector, you need to set up the WIRELESS NETWOK
parameters, as follows:
15) Run the “pixrad- viewer.exe” program from the Windows task-bar:
Start / Program Files / ThalesCal:
16) Select the “Open...” command in the System tab in the new Pixrad Demo page:
18) Once the detector is connected (see box below), select the Wifi tab:
19) Enter the wireless network parameters in the Wifi tab: - SSID : TH3543EZ
- Password : pixium3543
20) Once the detector has been updated with the new network settings (see box below), select the
System tab, press the “Close” command and then close the application:
Button
22) Disconnect the “back-up cable” from the detector and the WIRELESS NETWORK socket.
23) Reconnect the cables to the DICOM NETWORK, PU NETWORK and WIRELESS NETWORK
sockets and then switch the Access Point on.
24) Wait about 2 minutes and then switch the detector back on using its ON/OFF button and check
that the warning LEDs are as follows:
- after about 1 minute, the 3 LEDs become steady, with these colours:
- STATUS LED: orange
- BATTERY LED: green
- WIFI LED: green
Note: If the WIFI LED is green, it means connection to the wireless network is successful; if orange,
you need to repeat the entire setup procedure.
25) Perform the entire calibration procedure for the new detector as explained in chapter 3 above.
When setting the Wireless network parameters for the new Ethernet POE Switch, you need to foresee
the use of 2 different models:
- 3CNJ1000
- 3CNJ2000
The Ethernet switch model 3CNJ1000 does not require any setup.
To set up the Ethernet switch model 3CNJ2000 with the wireless network parameters:
- temporarily change the IP Address for the Video Processor used for the WIRELESS NETWORK,
using an address compatible with that for the switch itself, so that the two devices can
communicate with each other.
- Then change the IP Address for the switch.
- End the procedure by restoring the original IP Address for the PID.
To do this:
1) Find the IP Address of the switch indicated in its installation guide (e.g.: 169.254.1.251).
2) With the system switched off, replace the old switch with the new one and then connect this as
explained in paragraph 2.2 in this part of the manual.
3) Temporarily disconnect the cables in the DICOM NETWORK and PU NETWORK sockets.
4) Switch the system on and log in as Administrator, keeping the Shift key on the alphanumerical
keyboard down when the system boots up (see paragraph 7.3 in Part 2 of this manual).
5) Change the IP Address for the Video Processor used for the WIRELESS NETWORK by selecting
in the Windows task-bar:
Start / Settings / Network Connections / Wireless Connection
(see procedure explained in paragraph 2.3.1 in this part of the manual).
9) Select SYSTEM / LAN Settings in the next page to change the Switch’s network settings:
10) Change the parameters in the Change IP Address table as shown below and then click on
Apply.
12) The IP Address of the Ethernet POE Switch is now correctly set: 10.0.1.250
13) Restore the original working IP Address for the Video Processor used for the WIRELESS
NETWORK (changed at point 5 above): 10.0.1.251
14) Reconnect the cables in the DICOM NETWORK and PU NETWORK sockets.
15) Run the HIRIS RF43 application and check that the PIXIUM 3543EZ wireless detector connects
correctly with the system.
To set the Wireless network parameters for the Access Point (A.P.):
- Temporarily change the IP Address for the Video Processor (PID) used for the WIRELESS
NETWORK, using an address that is compatible with that of the A.P. so that the two elements
can communicate with each other.
- Then change the IP Address for the Access Point.
- End the procedure by restoring the original working IP Address for the PID.
To do this:
1) Find the IP Address of the A.P. indicated in its installation guide (e.g.: 192.168.0.50).
2) With the system switched off, replace the old A.P. with the new one and then connect this as
explained in paragraph 2.2 in this part of the manual.
3) Temporarily disconnect the cables from the DICOM NETWORK and PU NETWORK sockets.
4) Switch the system on and log in as Administrator, keeping the Shift key on the alphanumerical
keyboard down when the system boots up (see Annex 7.3 in Part 2 of this manual).
5) Change the working IP Address for the Video Processor used for the WIRELESS NETWORK, by
selecting in the Windows task-bar:
Start / Settings / Network Connections / Wireless Connection
(see procedure explained in paragraph 2.3.1 in this part of the manual).
8) The web browser automatically opens with the A.P. access page. Enter admin as the User Name,
leave the Password field empty and click on Login:
9) The main section of the page shows all the general Access Point information.
10) Select the MAINTENANCE / CONFIGURATION FILE menu to load the wireless network settings
via a setup file saved on the video processor system disk:
11) You need to specify the path for this file in order to load the A.P. settings.
Click on Browse:
13) Select the path: My Computer / System (C:) / AP Restore DAP2553 (detettore 3543EZ)
Then select the file: config.dcf and click on Open.
The file path appears in the main page:
14) After loading the setup file, click on Upload and confirm with OK when asked to confirm this
operation:
15) The A.P. setup “Restore” procedure now starts. Do not switch the equipment off and wait for the
operation to terminate (this takes about one minute).
16) The following page appears once the “Restore” procedure is terminated. You can now close the
setup page:
17) To check whether the network settings have been loaded correctly, access the A.P. with the new
IP address. To do this:
- Restore the original working IP Address for the Video Processor used for the WIRELESS
NETWORK (changed at point 5 above): 10.0.1.251
- Start Windows Explorer by selecting: Start / Programs / Accessories / Windows Explorer,
then enter the new IP Address for the A.P. in the Address bar (10.0.1.249) and press Enter.
18) The Wireless connection settings are correct if the web browser automatically opens with the A.P.
access page.
Enter admin as the User Name, leave the Password field empty and click on Login.
19) Select the BASIC SETTINGS / WIRELESS menu in the main page and then check that the
network called TH3543EZ is listed:
20) Click on Logout to disconnect the A.P. and close the setup window.
21) Reconnect the cables in the DICOM NETWORK and PU NETWORK sockets.
Run the HIRIS RF43 application and check that the PIXIUM 3543EZ wireless detector is
connected to the system.
4.3 SPARE-PARTS LIST FOR THE PIXIUM 3543EZ WIRELESS DETECTOR UNIT
4a 4b
5
2
There are 3 warning LEDs on one side of the detector, together with an ON/OFF button-
Push button
Orange:
- Internal error
slow flashing
Green:
- Operating mode: IN STANDBY
slow flashing
On Connection made
LAN Green
Flashing Connection activity
STATUS LED
green LED
red LED
The life of the battery may be reduced if the battery is removed from its charger before the
battery has been fully charged.
TEST SHEET
Mechanical checks:
- Fixing of the main components: look for loose screws (cabinet, detector unit, etc.).
- Integrity of the protection devices for the main components: look for signs of damage or dents
that may affect the efficiency and safety of the EM equipment.
revision A page 1
code 87 80 014C - Mti test sheet.doc
TECHNICAL MANUAL
Switch the equipment off using the 2 ON/OFF switches on the theatre power unit: acquisition
system and detector power switches (see paragraph 3.3.4 in Part 2)
TS31 Check the general state of the cabinet boards and components (dirt and corrosion)
TS32 Check the impedance between the main earth terminal on power unit GP1 (inside the
cabinet) and the earth screw on the detector (check this by detaching the power connector
on the detector)
- The value should be less than 0.1 ohm -
Close the cabinet
TS34 Make sure that the fan belonging to the video processor is working properly and that its air
vents are not obstructed
TS35 Make sure that the fan belonging to the PU module is working properly and that its air vents
are not obstructed
TS36 Check that the monitors are working properly
Wait for the HIRIS RF43 application to boot and then open a new study
Procedure: - PIXIUM RF4343 detector: see chapter 5 in Part 2 of the Technical Manual
TS42 X-ray dose intensity in the air (automatic exposure meter control)
Procedure: Check at both 0.5 and 5 µGy following the ionising chamber
calibration procedure, explained in paragraph 5.2 of Part 5 in this
manual.
Measure the values at the leading edge (input) of the detector.
revision A page 2
code 87 80 014C - Mti test sheet.doc
TECHNICAL MANUAL
Reference values: ATS has adopted the procedure in the CEI EN 60601-2-7
standard, paragraph 50.102.1-b, by comparing the results of the
measurements of the image video levels obtaining using
phantoms of different thicknesses in 3 cases:
1 – different phantom thicknesses
2 – different kV
3 – repeating exposures under the same exposure conditions.
st
Procedure 1 check: - Set 70 kV and 100 mA.
- Insert the first phantom (ATS uses an Al plate, 10 mm thick) and
give the radiography command and then use the Image Statistic
menu (see Annex 5.1 in Part 5 of this manual) to find the grey
level at the centre of the image.
- Insert a second phantom (ATS uses an Al plate, 21 mm thick)
and again give the radiography command; find the grey level at
the centre of the image.
- Check that the difference between the mean grey level in the
images is not more than 20%.
nd
Procedure 2 check: - Insert a phantom (ATS uses an Al plate, 21 mm thick).
- Set 60 kV, give the radiography command and then use the
Image Statistic to find the grey level at the centre of the image.
- Repeat at 80 kV.
- Repeat at 120 kV.
- Check that the difference between the mean grey level in the
images is not more than 15%.
rd
Procedure 3 check: - Insert a phantom (ATS uses an Al plate, 21 mm thick).
- Set 70 kV and 100 mA, give the radiography command and then
use the Image Statistic to find the grey level at the centre of the
image.
- Repeat 3 more times under the same conditions.
- Check that the difference between the mean grey level in the
images is not more than 10%.
TS44 Spatial resolution
Reference values: ATS has adopted the "FUNK type 18 resolution test" phantom:
check that the test results are close to those indicated in the Test
Report attached to the EM equipment.
Procedure: Measure the working doses as per the procedure in paragraph 5.4 of
Part 5.
revision A page 3
code 87 80 014C - Mti test sheet.doc
TECHNICAL MANUAL
Reference values: ATS has adopted the "FUNK type 18 resolution test" phantom:
check that the test results are close to those indicated in the Test
Report attached to the EM equipment.
Procedure: - Phantom at 45° in the centre of the detector field, 43x43 cm.
- Give the continuous fluoroscopy command, setting the kV to get
the best image.
- Find the best image resolution on the monitor and then compare
the value with that indicated on the system Test Report.
- Repeat for fields 30x30 cm and 20x20 cm.
Procedure: Measure the working doses as per the procedure in paragraph 5.4 of
Part 5.
Reference values: ATS has adopted the "FUNK type 18 resolution test" phantom:
check that the test results are close to those indicated in the Test
Report attached to the EM equipment.
Procedure: - Phantom at 45° in the centre of the detector field, 43x43 cm.
- Give the pulsed fluoroscopy command, setting the kV to get the
best image.
- Find the best image resolution on the monitor and then compare
the value with that indicated on the system Test Report.
- Repeat for fields 30x30 cm and 20x20 cm.
Observations: …………………………………………………………………………………………………………….……
……………………………………………………………………………………………………………..……
……………………………………………………………………………………………………………..……
revision A page 4
code 87 80 014C - Mti test sheet.doc
FID / Document identification CTD / Type VS / Release
1/3
61665525 108 B
PIXIUM FE 4343 [R or F]
Storage, Handling and Packing instructions
This document provides instructions for inspection, storage, unpacking and re-packing
PIXIUM FE 4343 [R or F] detector shipment. Please read this manual carefully before any handling of
the delivered material.
3. Unpacking instructions
The cardboard box and the inner protective bag have to be opened with care in order to be reused.
Unpacking procedure :
a) Open the cardboard box using the two fasteners on the front side of the box.
b) Remove the inner protective bag from the box (See next page for continuation of instructions).
! ! ! CAUTION ! ! !
Do not open the inner protective bag before a sufficient acclimatization time. The acclimatization
time is specified as 7h30 hours (minimum). It is mandatory to remove the detector from the
cardboard box during this time – however do not under any condition open the sealed protective
bag during this time. Refer to the PIXIUM FE 4343 [R or F] specification for storage conditions
(environmental temperature and humidity, etc...) of the detector once the protective bag has been
opened at the end of this acclimatization time.
c) In order to open the inner protective bag, cut off only the sealed part of the bag.
d) Immediately after opening the inner protective bag, check and note the value provided by humidity
indicator placed inside the bag.
CONFIDENTIEL TRIXELL
Copyright © 2007, 2011 TRIXELL. Tous droits réservés. Ce document et l’information qu’il contient sont la propriété exclusive de
TRIXELL. Ils ne peuvent être reproduits ou divulgués sous quelque forme que ce soit sans l’accord préalable écrit de TRIXELL.
TRIXELL PROPRIETARY INFORMATION
Copyright © 2007, 2011 by TRIXELL. All rights reserved. This document and the information contained therein are TRIXELL’s
exclusive property. Any copy and/or disclosure thereof in any form whatsoever are subject to TRIXELL’s prior written consent.
FID / Document identification CTD / Type VS / Release
2/3
61665525 108 B
e) A digital readout thermometer is placed inside the inner cardboard box, to record max and min
temperatures during transport and storage. To read these temperatures, just press MAX/MIN push-
button and note the indications. Refer to the PIXIUM FE 4343 [R or F] specification for specified
temperature values.
f) Keep the following original parts in a secure place in case of detector return :
- cardboard box,
- foams,
- digital thermometer,
as well as a new inner protective bag (with Velcro strip) which contains a small sealed bag with
desiccant material bag and a new humidity indicator. DO NOT OPEN this small sealed bag as it will be
used in the case of any detector return. Take care to not apply undo shocks to the shipping containers
even if empty.
4. Re-packing instructions
All the original parts of the packing have to be used for any shipment of the detector. In case of detector
return, a set of repacking material is delivered within the original detector package. Do not use the old
desiccant material which was found inside the sealed inner protective bag ! Open the small sealed bag
(see above) and place the contents with the detector just before closing the new bag.
Packing procedure (see pictures in appendix) :
a) Introduce the detector inside the new protective bag.
b) Place the new desiccant material bag close to the detector .
c) Open the protective sealed bag of the new humidity indicator, and place it inside the bag.
d) Close the new protective bag (with Velcro strip) as indicated on the return instructions label stuck on
the bag.
e) Reset the digital readout thermometer placed inside the cardboard box by pressing the MEM
pushbutton.
f) Place the protective bag inside the original cardboard box, using the foam provided with original
shipment.
g) Re-close the cardboard box using the two fasteners.
CONFIDENTIEL TRIXELL
Copyright © 2007, 2011 TRIXELL. Tous droits réservés. Ce document et l’information qu’il contient sont la propriété exclusive de
TRIXELL. Ils ne peuvent être reproduits ou divulgués sous quelque forme que ce soit sans l’accord préalable écrit de TRIXELL.
TRIXELL PROPRIETARY INFORMATION
Copyright © 2007, 2011 by TRIXELL. All rights reserved. This document and the information contained therein are TRIXELL’s
exclusive property. Any copy and/or disclosure thereof in any form whatsoever are subject to TRIXELL’s prior written consent.
FID / Document identification CTD / Type VS / Release
3/3
61665525 108 B
Detector in its protective bag (Velcro strip) placed inside the cardoard box :
END OF DOCUMENT
CONFIDENTIEL TRIXELL
Copyright © 2007, 2011 TRIXELL. Tous droits réservés. Ce document et l’information qu’il contient sont la propriété exclusive de
TRIXELL. Ils ne peuvent être reproduits ou divulgués sous quelque forme que ce soit sans l’accord préalable écrit de TRIXELL.
TRIXELL PROPRIETARY INFORMATION
Copyright © 2007, 2011 by TRIXELL. All rights reserved. This document and the information contained therein are TRIXELL’s
exclusive property. Any copy and/or disclosure thereof in any form whatsoever are subject to TRIXELL’s prior written consent.