You are on page 1of 1

Letter to the Editor

Journal of Indian Orthodontic Society


A Comparative Study of the Inbuilt 55(2) 215, 2021
© The Author(s) 2021
Torque and Slot Size of the MBT Reprints and permissions:
in.sagepub.com/journals-permissions-india
Prescription Bracket of Different DOI: 10.1177/0301574220979766
journals.sagepub.com/home/jio
Manufacturing Companies: A
Stereomicroscopic Study

Mukesh Kumar
MDS in Orthodontics
1048, C-1,Vasant Kunj, New Delhi, Delhi 110070, India
Professor, Department of Orthodontics and Dentofacial Orthopaedics, Teerthanker Mahaveer Dental College, Moradabad,
Uttar Pradesh 244001, India
E-mail: drmukeshortho@yahoo.co.in

Manish Goyal
MDS, PhD, MORTH RCS Ed (UK)
House no 197-198, 2nd floor, Pocket 12, Sector 24, Rohini, New Delhi, Delhi 110085, India
Professor and Head of Department, Department of Orthodontics and Dentofacial Orthopaedics, Teerthanker Mahaveer
Dental College, Moradabad, Uttar Pradesh 244001, India.
E-mail: manishortho@yahoo.com

Ekta Yadav
1265/1B, Friends Colony, Near Alpine Public School, Jhansi, Uttar Pradesh 284002, India.
Post Graduate Student, Department of Orthodontics and Dentofacial Orthopaedics, Teerthanker Mahaveer Dental College,
Moradabad, Uttar Pradesh 244001, India.
E-mail: yadav21ekta@gmail.com

An article was published in October 2020 by Ashish 3. As mentioned in Results (p. 299), a post-hoc power
Mathew et al., with the purpose of comparing the inbuilt analysis was done to determine the built-in torque,
torque and slot size of the MBT prescription bracket of and it was found to be 0.761, but in Table 5 (p. 300),
different manufacturing companies using stereo-microscope.1 the observed power is mentioned as 0.996. Kindly
The article was of great information and appreciated by explain the variation.
most of us, but we have come across certain doubts. Kindly 4. In statistical analysis (p. 298), Kappa statistics is
respond to the same. used for assessing intra-examiner reliability.
1. In Abstract (p. 297), it is mentioned that upper right We feel intraclass correlation coefficient should be used
central incisor brackets of the Mclaughlin Bennet instead of Kappa statistics, as Kappa statistics is used for
and Trevisi (MBT) prescription with a 0.022-inch intra-examiner reliability when data is qualitative and in this
slot were used in this study, while in the methodology study we have quantitative data.
section (p. 298) left upper central incisor brackets are
mentioned. Kindly explain the variation. Reference
2. In Table 3 (p. 300), F-ratio is given as 6.745. In 1. Mathew A, Kiran Kumar HC, Shetty S. A comparative study
statistics, F-ratio is the ratio of the mean square of the in-built torque and slot size of mbt prescription bracket
value of between group to the mean square value of of different manufacturing companies: a stereo-microscopic
within group, but here the mean square value of both study. J Indian Orthod Soc. 2020;54(4):297-303.
the between and within group is given as 0, and so
the F-ratio should also be 0. Kindly clear our doubt.

Creative Commons Non Commercial CC BY-NC: This article is distributed under the terms of the Creative Commons Attribution-
NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-Commercial use, reproduction
and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages
(https://us.sagepub.com/en-us/nam/open-access-at-sage).

You might also like