TFA can be displaced from the ion-pair complex by to determine the actual mechanism. For the scope
exposing the effluent to a higher concentration of of this work, observing elution order change is
a different acid post-column and before the MS sufficient.
detector. One acid that works well for this approach
is acetic acid. Another ion pair displacer is a solu- Figure 4 shows the effect of higher acid concentra-
tion of 75% propionic acid:25% isopropanol. TFA tion on the separation. In almost all cases there
and phosphoric acids yield the best separations as are sharper peaks, the SG peak showing the great-
shown in Figure 3, while commonly used formic est improvement. In Figure 3, the mobile-phase pH
acid and acetic acid concentrations resolve only at was near the pKa values of all compounds except
best five of six components. Furthermore, in the phenol, so the addition of more acid would be
examples using formic acid and acetic acid, 4-HBA expected to lower the solution pH to a point where
and GA switch elution order compared to the TFA the compounds are more completely uncharged,
and phosphoric acid examples. This may be due to causing better peak shape.
changes in pH; however, additional work is needed
10 3 5 3
1 10 1
5 5
0 0
_5
0 1 2 3 4 5 6 7 8 9 min 0 1 2 3 4 5 6 7 8 min
Figure 3. Effect of different acid modifiers on the separation of salicylic acid and its impurities.
3
www.agilent.com/chem
StableBond SB-Aq StableBond SB-Aq
1.0% TFA 0.1% TFA
mAU 2 mAU 2
30 4
3.235
40 4
1. Phenol 1. Phenol
35 25 6
2. 4 HBA 2. 4 HBA
30 3. GA 6 3. GA
20
4. HIPA 4. HIPA
25 5. SG 5. SG
15
0.353
20 6. SA 6. SA
15 5 10
1 5
1 3
10 3 5
5
0
0
0 1 2 3 4 5 6 7 8 9 min 0 1 2 3 4 5 6 7 8 9 min
StableBond columns are ideal for low-pH applica- For More Information
tions. Their unique column chemistry and variety
of bonded phases, such as SB-Aq, provide broad For more information on our products and services,
selectivity, long lifetime, and reproducibility at low visit our Web site at www.agilent.com/chem.
pH.
References Agilent shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this material.
1. M.A. Van Straten, H A. Claessens A. Dams,
Information, descriptions, and specifications in this publication are subject to change
“Analysis of Organic Acids in Aqueous without notice.
Solutions,” Agilent publication 5989-1265EN,
2006. © Agilent Technologies, Inc. 2009