marLux® X6 / X8 / H5 / H5 PLUS / H6 / H8
Operating Instructions
Table of Contents
1 Product Liability and Warranty ......................................................5
1.1 General Information .............................................................................5
1.2 Intended Use ......................................................................................5
1.3 Warranty ............................................................................................5
V 2.0 3
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
8 Accessories .............................................................................. 30
10 Troubleshooting ........................................................................ 33
11 Technical Data.......................................................................... 34
11.1 Technical Data for marLux® X8/X6 Models ............................................. 34
11.2 Technical Data for marLux® H8/H6 Models ............................................ 35
11.3 Technical Data for marLux® H5 PLUS/H5 Models .................................... 36
4 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
1.3 Warranty
Gebrüder Martin warrants the products described in this user manual to be free from any
defects in material or workmanship for a period of two years from the date of delivery as
specified on the delivery slip, provided they are used as intended. This warranty shall not cover
consumables such as lamps and sterile handles or potential software updates. In any other
respects, our General Conditions of Sale as updated from time to time shall apply.
Under this warranty, we will remedy free of charge any defects due to faulty workmanship or
the use of faulty materials, either through our Customer Service or directly at the factory.
Important Notices
The product may only be repaired by us or a qualified person or firm expressly authorized by
us to perform such work.
If the repair is carried out by a specially authorized person or firm, the repairer is required to
issue to the unit’s owner/operator a certificate with details about the nature and scope of the
repair work done. This certificate must be dated and signed and include the firm’s details. In
addition, repaired products must be clearly marked with the service technician’s ID label in all
cases where a party other than the product manufacturer performed the work.
V 2.0 5
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
Improper interventions or modifications performed by third parties during the warranty period
shall void the warranty and associated claims. Consequently, any unauthorized action
performed on the operating light shall release Gebrüder Martin from its liability obligations.
User’s Inspection
• Immediately upon receipt, the goods must be checked for completeness and potential
damage in transit.
Hotline
Should you have any technical questions or require spare parts, please contact our hotline.
Phone: +49 7461 706-343
Fax: +49 7461 706-203
msc@klsmartin.com
Our service representatives are available Monday through Friday from 8 a.m. to 5 p.m. (CET).
Should you have any questions concerning maintenance contracts or training courses, please
contact our Technical Service Manager under +49 74 61 706-332, or send an e-mail to:
msc@klsmartin.com
6 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
V 2.0 7
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
Symbol Explanation
CE-mark
8 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
3 Safety Notices
3.1 General Information
KLS Martin operating lights are quality products designed and manufactured in accordance with
the recognized technical rules. The products leave the factory in a perfectly safe working
condition.
To maintain this status, you – as the end user – are required to abstain from or prevent any
action that might have an adverse effect on the lights’ safety characteristics.
Upon completion of any repair or service work, it is mandatory for the service technician to
check all functions of the light, the tracking and spring arms and the power supply units, thus
ensuring their proper working order and operational safety!
It is not permitted to attach to or install on the operating lights any components or systems
other than accessories that have been expressly approved by Gebrüder Martin.
Notice! Due to the eyelid closure reflex, the pilot laser radiation is harmless
for non-narcotized persons.
As long as the impact of the laser radiation is accidental and limited to a maximum period of
0.25 s, Class-2 laser equipment does not pose a danger to the human eye.
V 2.0 9
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
Therefore, Class-2 laser equipment may be used without any special protective measures,
provided that the application requires neither deliberate eye contact for more than 0.25 s nor
repeated eye contact with the direct or reflected laser radiation.
As the eye-protecting lid closure reflex cannot usually be taken for granted, we recommend
closing the eyes deliberately and turning away from the light source immediately if Class-2
laser radiation hits the eye.
These warning signs are attached to the sterilizable handle and next
to the control panel. The laser beam’s exit opening (aperture) is
located inside the central handle.
Note that the pilot laser is automatically activated for 2 seconds
whenever the operating light is repositioned.
It is the user’s own responsibility to observe and implement the above provisions/regulations.
Defective lights can damage the health or even threaten the lives of
users and/or patients!
Therefore, never operate defective lights!
10 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
V 2.0 11
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
3.3 Operation
• Prior to each use, be sure to check the lights for safe functioning and proper operating
condition!
• Be sure to handle the lights in accordance with the instructions provided in this manual
(including startup and withdrawal from service)!
12 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
V 2.0 13
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
• Options:
- pilot laser
- surgiCam camera system (with separate user manual)
- wall operation (wall-mounting control panel)
- backLite
- emergency power (ASPS) changeover module
- LC (Low Ceiling) cardanic suspension for low rooms
- Fully cardanic suspension (H5)
- dimming (H5)
• Available mounting/installation options:
- ceiling-mounting: single lights, light combinations
- wall-mounting (marLux® X6/H6/H5), only with separate power supply on external
mounting plate
- mobile version (stand version) (marLux® X6/H6/H5)
can be mounted either individually (as single lights) or in combinations of two, three or four
lights (optional equipment may differ in each case). The type of combination used has no
influence on light operation.
14 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
Example: X8 centriXbeam®
V 2.0 15
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
During temporary mains power failures, the (optional) emergency power changeover module
automatically switches over to battery operation (additional safety power supply, in short
ASPS, as part of the safety power supply or SPS system). The discharge lamp’s performance
remains totally unaffected by this process.
If the light is switched off and on again (“warm start”), the metal halide discharge lamp will
need approx. 10 minutes before it can be restarted by the ballast once it has cooled down. Just
as in the cold-start scenario, the halogen lamp substitutes for the discharge lamp (indicated by
the green control panel LED – see section 6.3), ensuring 100% illumination.
All marLux® lights can be optionally equipped with an emergency power (ASPS) changeover
module, which can either be integrated into the ceiling tube or installed separately. Where the
light has an external power supply located in a switchbox, the ASPS changeover assembly is
installed on the same mounting plate as the power module.
A dual LED – which can be optionally installed in the canopy or on the mounting plate –
indicates the type of power supply currently used:
16 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
The KLS Martin surgiCamdigital is an imaging system for documentation and/or training
purposes. The system’s high-performance camera can either be integrated into the light head’s
central handle or installed on an arm of its own. The KLS Martin surgiCamdigital comes with a
separate user manual.
For image display, the optional KLS Martin medTFTpro flat-screen monitor system is used.
The pilot laser integrated into the operating light marks exactly the center of the light field.
This makes it very easy to focus the light field on small surgical sites.
The pilot laser is automatically switched on as soon as the light is repositioned.
V 2.0 17
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
5.5.4 backLite
The backLite option comes integrated into the operating light’s tracking arm. It provides the
necessary indirect, diffuse background illumination by way of LEDs (1) whose light is directed
upwards. It is switched on and off automatically in accordance with a preset ambient
illuminance threshold.
18 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
6.2 marLux® H5
The KLS Martin marLux® H5 light can only be operated via the sensoGrip handle (see
section 6.4). The light’s control panel located next to the light arm’s cardan joint features the
ON/OFF switch (1) plus two halogen lamp indicators (2, 3) (for halogen lamp indicator
functions, see section 6.3.)
The various control functions available are marked with symbols. A slight finger touch is
sufficient to operate the pushbutton switches.
V 2.0 19
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
1 ON/OFF switch with standby function: “ON” = green LED lit; “OFF” = LED goes out
2 Halogen lamp (X6/X8)
This LED lights up green as long as the halogen lamp is active. In normal operation, this is the case
during the start-up phase of the metal-halide discharge lamp (cold start: approx. 2 min; warm
start: approx. 10 min). Consequently, this LED goes out as soon as the discharge bulb becomes
active (i.e. takes over).
This LED lights up red in the event of trouble (i.e. halogen lamp failure).
Halogen lamp (H6/H8):
This LED lights up green when the lower halogen lamp is active and the operating light is fully
operational. It lights up red if trouble/malfunction occurs.
3 Metal halide discharge lamp (X6/X8)
This LED lights up green when the discharge lamp is active and the light is fully operational. It
lights up red in the event of trouble (i.e. discharge lamp failure).
Halogen lamp (H5/H6/H8)
This LED lights up green when the upper halogen lamp is active and the light is fully operational. It
lights up red if malfunction occurs.
4 centriXbeam®: deep-cavity illumination ON/OFF (X6CX / H6CX / X8CX / H8CX)
A lit LED indicates that the central depth light has been activated.
Operating the button again switches off the depth light, thus reducing shadow formation.
5 Reduces the light field diameter
6 Reduces illuminance
7 Infrared sensor for optional remote control (currently under development)
8 Increases illuminance
9 Increases the light field diameter
The settings of buttons 4, 5, 6, 8 and 9 are saved when the light is switched off. This last-used
operating status is then automatically activated when turning on the light the next time.
Notice! If the LEDs (2 and/or 3) light up red for more than 10 minutes, both
the discharge lamp and the halogen lamp must be replaced in the X6
and X8 models (for procedure, see section 7.2). As far as the H5,
H6 and H8 models are concerned, only the defective halogen lamp
needs to be replaced (see section 7.3).
20 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
*
Notice! H5 and H5 PLUS lights only offer light field adjustment (by rotating
the sensoGrip handle clockwise or counterclockwise as specified
above).
V 2.0 21
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
7 Replacing Components
Sterilizable handle units for different X8/H8 models: without pilot laser, with pilot laser, and
with camera system integrated into handle (from left to right)
Notice Be sure to observe and follow the cleaning and sterilization instructions given in
section 9.2.
22 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
If the metal-halide discharge lamp has become defective (upper LED (3) lights up red), both
the discharge lamp and the halogen lamp must be replaced.
If the halogen lamp has become defective (lower LED (2) lights up red), only the halogen bulb
needs to be replaced (see section 6.3). Be sure to use only genuine KLS Martin bulbs from the
spare lamp set (89-907-02).
7.2.1 Removing the Lamps (X6/X8 Models)
V 2.0 23
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
The substiLite light source unit consists of a screen unit with six lamellar blades (3) for
scattered light absorption and two lamp supports (2a, 2b) with sockets for the halogen lamp
and the metal-halide discharge lamp, respectively. Each lamp support is connected to the
screen unit with three fastening clips (1).
• Open the three fastening clips (1) on each of the two lamp supports (2a, 2b).
Notice Before removing the lamp supports, take note of the exact mounting positions on
the substiLite light source unit.
• Pull the lamp supports out of the screen unit together with the halogen lamp (5) and the
metal-halide discharge lamp (6), respectively.
• Remove the used halogen lamp (5) and discharge lamp (6) from their sockets and dispose
of them in accordance with local regulations.
• Take the spare lamps from the spare lamp set with the protective covers (1) in place.
• Insert the new halogen lamp (2) and the new metal-halide discharge lamp (5) into the
socket of the appropriate lamp support:
- The support for the halogen lamp (3) is marked by a handle (4).
- The support for the discharge lamp (6) is marked by five bunch plugs (7).
• Remove the protective cover (1) from each of the two lamps.
Notice: When replacing the bulbs, be sure to check the sockets for potential defects. If the
socket is defective as well, the entire substiLite light source unit will need to be
replaced.
24 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
• Insert the lamp support complete with the metal-halide discharge lamp (1) into the screen
unit, making sure that the pin (2) engages with the cut-out provided on the lamp support.
• Using the three fastening clips (3), fix the lamp support in place on the screen unit.
• Install the second lamp support with the halogen lamp (6) in the screen unit, making sure
that the lamellar blade with the arrows (4) engages with the slot provided on the lamp
support (7).
• Using the three fastening clips (5), fix the lamp support in place on the screen unit.
V 2.0 25
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
In the case of marLux® H5/H6/H8 lights, both halogen lamps must always be replaced
together if one of the two control panel LEDs (2 or 3) lights up red (see sections 6.2 and 6.3).
Besides, marLux® H5/H6/H8 lights require the lamp sockets to be replaced along with the
halogen lamps. Be sure to use only genuine KLS Martin bulbs from the spare lamp set (for item
number, refer to section 8).
26 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
The dismounting procedure for the substiLite light source unit is the same as for marLux®
X6/X8 lights (see section 7.2.1).
To remove the halogen lamps, proceed as follows:
• Using the screwdriver from the spare lamp set, loosen the two screws located on the
halogen lamp socket, then remove the lamp complete with socket (see illustration in
section 7.3.2).
The halogen spare bulbs for marLux® H5/H6/H8 lights are delivered in preassembled
condition (complete with socket).
• Take the halogen lamp assembly from the spare lamp set with the protective cover in
place.
V 2.0 27
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
28 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
• Using the handle (4), rotate the substiLite unit so that the arrow-marked lamellar blade (3)
engages with the guide bar (1) located next to the rectangular connector (2).
• Slide the substiLite unit into the filter unit as far as it goes.
Notice! The upper lamp support of the substiLite unit features five bunch
plugs (1) approx. 10 mm long. To ensure that the substiLite unit sits
securely, these plugs must be pressed into the corresponding sockets
of the filter unit. This requires a little force to overcome the
mechanical tension resistance.
• Install the handle unit on the filter unit, making sure that the rectangular connector (3)
engages with the mating connector located on the filter unit.
• Tighten the three screws (2) located on the handle unit (4) by clockwise rotation.
• Check the proper functioning of the light by switching it on. With marLux® X6/X8 lights,
the LED 2 (see section 6.3) must go out after approx. 2 minutes in the case of a cold start.
At the same time, the metal halide discharge lamp will be moved into central position.
V 2.0 29
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
8 Accessories
30 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
The final rinse should be carried out with demineralized water to prevent spotting and the
formation of deposits.
Neutral or alkaline cleaning agents can be used for cleaning. However, cleaners containing
active chlorine should be avoided due to the chemical reaction between chlorine and plastic.
The handle is steam-sterilizable, using a fractionated vacuum process at 134°C (273.2°F) for 5
minutes or 121°C (249.8°F) for 20 minutes (e.g. a sterilizer complying with
EN 285/ANSI/AAMI/ ISO 11134-1993 and ANSI/AAMI ST 46-1993, validated in accordance
with EN 554/ISO 13683 requirements). To prevent permanent (plastic) deformation, the
handle should be kept free from mechanical loads during sterilization.
V 2.0 31
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
9.3 Maintenance
All light functions should be checked for proper working condition and operational safety at
regular intervals.
Every 24 months, a comprehensive safety check (SC) must be carried out by an authorized
person in accordance with the relevant national standard. Please also refer to section 13 for
details.
32 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
10 Troubleshooting
V 2.0 33
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
11 Technical Data
11.1 Technical Data for marLux® X8/X6 Models
marLux® X8 marLux® X6
PSU input voltage: 85–240 VAC, 50–60 Hz
PSU output voltage: (21–28 VDC, adjustable)
Supply voltage at contact block: min. 26 VDC, max. 36 VDC
PSU rated power output 240 VA
Power input (prim.) 128 VA / 218 VA / 330 VA (gas/halogen/mixed operation)
Light sources (lamps) 22.8 V, 150 W IRC / 90 V, 70 W
Max. central illuminance (Ec) 160,000 lx 150,000 lx
Total irradiance at maximum 550 W/m² 515 W/m²
illuminance
Luminous efficacy 291 lm/W 291 lm/W
Light field diameter, d10 20–32 cm 19–30 cm
(small – large)
Light field diameter, d50 10 cm 10 cm
Color temperature 4,300 K 4,300 K
Color rendering index (CRI) 95 95
Shadow dilution**:
- in standard tube with one 57% 37%
mask, relative to Ec
- in standard tube with two
masks, relative to Ec 51% 41%
Illumination depth, L1/L2 450/900 mm 450/650 mm
Starting performance: Cold start: 3.5 min; warm start: 5 min + 3.5 min
Rated life of discharge/halogen approx. 5,000 h / approx. 5,000 h /
lamps 600–1,000 h* 600–1,000 h*
Protection class I (acc. to IEC 60601-1)
Equipment type B (acc. to IEC 60601-1)
Ambient conditions
- Storage temperature -10°C to +50°C (14–122°F) with 20–90% humidity (non-
condensing)
- Operating temperature 5°C to +40°C (41–104°F) with 30–80% humidity (non-
condensing)
- Atmospheric pressure min. 700–1,060 hPa
All values were determined in accordance with IEC 60601-2-41.
* Depending on mechanical shock intensity, supply voltage and number of ON/OFF cycles.
** The shadow dilution values indicated for one mask and two masks, respectively, are the
same in each case when using no tube.
34 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
marLux® H8 marLux® H6
PSU input voltage: 85-240 VAC, 50–60 Hz
PSU output voltage: (21–28 VDC, adjustable)
Supply voltage at contact block: min. 26 VDC, max. 36 VDC
PSU rated power output 240 VA
Power input (prim.) 218 VA
Light sources (lamps) 22.8 V, 150 W IRC 22.8 V, 150 W IRC
Max. central illuminance (Ec) 155,000 lx 145,000 lx
Total irradiance at max. 543 W/m² 508 W/m²
illuminance
Light field diameter, d10
(small – large) 17–30 cm 18–29 cm
Light field diameter, d50 10 cm 10 cm
Luminous efficacy 285 lm/W
Color temperature 4,300 K
Color rendering index (CRI) 93
Shadow dilution**:
- in standard tube with one 57% 32%
mask, relative to Ec
- in standard tube with two
masks, relative to Ec 51% 41%
V 2.0 35
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
36 V 2.0
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
ML X8xxx 00 19 06 C 7777
| | | Software version
| | Hardware version
V 2.0 37
KLS Martin marLux®
X6 / X8 / H5 / H5 PLUS / H6 / H8 Operating Instructions
The performance of the safety checks as well as their results must be documented.
At least every 24 months, the following checks and tests must be performed on the unit by
persons who have the necessary training, knowledge and practical experience to carry out
such safety checks competently and independently:
• Inspect the operating light and its accessories visually for function-impairing mechanical
damage. Check safety-relevant labels and markings for legibility.
• Perform a functional check as specified in the Operating Instructions.
• Perform an electrical check in accordance with the Test Report specifications for Periodic
Safety Checks (see service manual).
• Measure the supply voltage at maximum illuminance.
38 V 2.0
Gebrüder Martin GmbH & Co. KG
Ludwigstaler Straße 132 · D-78532 Tuttlingen
Postfach 60 · D-78501 Tuttlingen/Germany
Tel. +49 7461 706-0 · Fax +49 7461 706-193
info@klsmartin.com · www.klsmartin.com
01.07 · 90-273-52-22 · Printed in Germany · Copyright by Gebrüder Martin GmbH & Co. KG · Alle Rechte vorbehalten · Technische Änderungen vorbehalten
We reserve the right to make alterations · Cambios técnicos reservados · Sous réserve de modifications techniques · Ci riserviamo il diritto di modifiche tecniche