Aquatron
Automatic Water Stills
A4000 A8000 & A4000D
Assembly and
operating instructions
Contents
Cleaning Instructions 16
Fault Finding 17
List of Spares 18
Limitations of use
This product meets the applicable E.C. harmonised
standards for radio frequency interference and may
be expected not to interfere with, or be affected by
other equipment with similar qualifications. We
cannot be sure that other equipment used in its
vicinity will meet these standards and we cannot
guarantee that interference will not occur in practice.
Where there is a possibility that injury, damage or loss
might occur if equipment malfunctions due to radio
frequency interference, or for general advice before
use, please contact the Technical Service Department
of Barloworld Scientific Ltd.
Location & Services
A4000, A8000, A4000D
1
7
6 10
5 12
4
8 9
11
13
3 14
1 2 15
16
17
18
23
28 27 26 25 24 22 21 19
20
2
Fig 2
A4000 Assembly
Your Aquatron A4000 has been designed
with ease of assembly specifically in mind, Screw thread
Vent pipe connections
but before commencing assembly please
study the illustration, fig. 1 on page 2 to
familiarise yourself with the general lay- Condenser
out. During assembly, follow the sequence distillate outlet tube
of instructions and do not connect the
mains electricity supply until directed.
Rubber sleeve
Item No. Description Catalogue No. Qty. 13 Unscrew the black plastic screwcap,
and rubber ring, form the boiler. Slide
5 Condenser WC48/M2 1 both onto the heating element
27 Heater W48H 1 (W48H, item 27). Insert the heater
25 Boiler Level Control WL48 1 into the boiler ensuring that the end is
4 Funnel WF48 1 supported in the glass nodule of the
boiler.
16 Reservoir Level Control WLS 1
7 Wall Mounting Bracket – 1 Do not over-tighten the screwcap
– Bag of plastic hoses & tie straps – 1 as this may cause the boiler to
break or push the heater in too far
as this may break the nodule.
3 Remove the perspex viewing screen, 8 Take the 1 metre length of 16mm
item 1 fig. 1 by using the finger recess tubing and push through the larger of 14 Pass the electric cable of the heater
to lift the screen out of the lower the three holes, item 20, on the side up into the electrical control
cabinet groove, then remove the of the cabinet. This may be easier if compartment via the available orifice.
internal packaging which has the tubing is lubricated by wetting it Push the spade terminals onto the
protected the glass boiler during with water. Continue until the tube correct pins of the connecting strip
transit. Also sever the temporary tie extends approximately 250mm into (refer to fig. 3).
holding the two boiler retaining the main chamber, roughly level with
springs, item 28 in place. the outlet of the condenser. Blue - neutral
Brown - live
4 Remove the cabinet end plate, item 9 Connect the tubing to the outlet of Earth - yellow/green
17 by unscrewing the six retaining the level control WL48 and secure
screws and ‘shake proof washers’. with a tie-strap. E
N
Similarly, unscrew the six countersunk
Ensure that the Rotaflo PTFE stopcock
screws to remove the cabinet roof.
of the WL48 is closed. PCB
L
5 Position the condenser WC48/M2 by
10 Couple the WL48 to the boiler
mounting it onto the vapour tube of
connection arm with the plastic
the boiler with the distillate outlet
coupling and rubber sleeve item 26. Relay
tube of the condenser facing to the
Ensure there are no stresses from the Heater
front. See fig. 2.
16mm tubing on the connection. lead
3
15 Refit the cabinet end panel with all six The colour code for the mains cable
screws and washers and the cabinet is:–
roof with all six counter sunk screws.
Brown – Live
Replace the perspex viewing screen.
Blue – Neutral
You have now assembled your Yellow/Green – Earth
Aquatron Water Still. Move the
assembled A4000 to its working Mains Cable Replacement
location - using the mounting bracket, If the mains cable requires
item 7 if the unit is to be wall replacement, only specially prepared
mounted. Ensure there is a space of at spare mains lead obtained from
least 50cm to the right of the unit to Barloworld Scientific should be used.
allow access to the heating element.
DO NOT SWITCH ON ELECTRICITY
SUPPLY.
16 Connect the cold water supply to the
mains water inlet, item 18, fig. 1. 18 Connect the reservoir level control
Note the tubing selected should have WLS as detailed on page 14 of the
a safe working pressure of at least manual.
equal to the pressure of the water
supply and should be adequately 19 Refer to page 5 for operating
secured by hose clips. instructions.
DO NOT TURN ON WATER SUPPLY.
17 Electrical Installation
THIS EQUIPMENT MUST BE
EARTHED!
The electrical installation should only
be carried out by a qualified
electrician.
The A4000 is classed as permanently
connected equipment.
The equipment is supplied with 1.5m
of flexible, triple core, circular cable to
the following specification:- 1.5mm2,
to BS 6500 or equivalent and <HAR>
or BASEC approved.
Connection to the mains electrical
supply, should be via a double pole
isolation switch with a continuous
current carrying capacity of 15A at
250V and overcurrent protection
should be provided by a double pole
approved fuse rated at 15A.
These devices should be sited near to
the equipment and clearly marked –
“Disconnect device for Aquatron
Water Still“
4
A4000 Operation
The following instructions apply to the 4 Press the ON/OFF switch to the ON 11 Safety Cut-Outs
A4000 water still where the water feed to position – the switch will illuminate. All Aquatron water stills are protected
the boiler is via the mains supply, or a by the following safety devices.
header tank. 5 Recheck that the CLEAN switch is in
Flow Control Switch (item 15)
the CLEAN position and illuminated.
If the boiler is to be fed with either de- monitors the flow of coolant water
ionised or pre-treated water, please follow into the still and shuts off the heating
6 Turn on the mains water supply and
the instructions on page 15. element, if the flow is insufficient.
using the needle valve item 18 fig. 1
on the side of the cabinet, adjust the Thermostat Cut-Out (item 22)
1 Before switching on either the mains
flow rate to approximately 1 litre/min. In normal operation the water within
electricity or water supply, identify the
the glass will be at 100°C. In single
control switches and indicator lights Check that there are no water leaks
fault conditions, i.e. in the event of
on the front of the cabinet. from the hose connections.
water supply failure, the content will
Green Light increase to 110°C where upon a
7 Observe that the boiler is now filling
When illuminated this shows that resettable thermostat will operate.
with water to cover the heating
there is mains supply voltage to the Once switched, the thermostat has to
element. When the water has reached
unit. Under no circumstances should be manually reset. This is achieved by
the pre-set level, the excess will be
the cabinet end-panel be removed unscrewing the black knob, fig. 1,
discharged to drain. Check that the
when this light is illuminated. item 23, located inside the cabinet
drain water flows freely and does not
and pressing the reset button.
White “On Off” Switch “back-up” into the level control.
This is the primary switch for Reservoir Control (fig. 11)
controlling water and electricity inside 8 Switch the CLEAN button off by is positioned on the collection
the still. The switch illuminates when pressing it for a second time – the reservoir and switches off both the
pressed in the ON position. switch light goes out. electricity and water supply when the
reservoir is full of distilled water. The
Amber Light The heater will start to warm up and
water supply is only shut off around 7
When illuminated, this indicates that eventually run at a red glow. If the
minutes after the reservoir is full to
the distillate collection reservoir is full flow rate is insufficient, then the
allow the heating element to cool.
as detected by the level control. heater will not switch on – this will
require the control needle valve to be It is recommended that the operation
White “Clean” Switch opened further. After about 2 to 3 of the flow switch and reservoir
This shuts off the heating element but mins. of operation distilled water will control be checked on a regular basis.
allows water to flow into the boiler emerge from the outlet pipe, falling
when the still is being commissioned into the collection reservoir. a) Simulation of Mains Water
or cleaned. When pressed in the Supply Failure
CLEAN position the switch illuminates. 9 To avoid excessive wastage of coolant Turn off the mains water supply at
water, make further adjustments to the tap. This should switch off the
2 Carefully check the following:– the needle valve. Slowly reduce the heating element. Turning the tap on
flow-rate until the flow control switch, again will switch the heater on.
– the appropriate electricity, water
and drain services have been item 15, switches off the heaters – b) Simulation of Reservoir Full
provided. If in doubt, consult the then increase the flow until the power Condition
LOCATION & SERVICES section on is restored. Lower the glass pressure bell WPB,
page 1. (fig. 11), into the distillate to a
10 Refit the perspex viewing screen. depth of 150mm. This will switch
– the ON/OFF switch on the control
off the heating element
panel is in the OFF position.
immediately and the cooling water
– the CLEAN switch is pressed in the after 7 minutes. Raising the bell will
CLEAN position. cause both to be re-supplied.
5
7
6 10
5 12
4
8 9 11
13
14
3
1 2 15
16
17
18
24
19
29 28 27 26 25 23 22 20
21
Fig 4: Major component parts A8000
6
Fig 5
A8000 Assembly 4
5
1 From the outer packaging remove all
the components including the metal Condenser
To distillate
cabinet containing the glass boiler. reservoir
distillate outlet tube
Connect No. 2 to lower water inlet. of the cabinet. This may be easier if Do not over-tighten the screwcaps
the tubing is lubricated by wetting it as this may cause the boiler to
Connect No. 4 to upper water outlet. with water. Continue until the tube break or push the heaters in too
extends approximately 250mm into far as this may break the nodule.
7 Position the second condenser the main chamber.
WC48/M2 onto the right hand vapour
tube of the boiler with the distillate
outlet tube facing the front and make
the following tube connections:–
7
17 Pass the electric cables of the heaters Mains Cable Replacement
up into the electrical control If the mains cable requires
compartment via the available orifice. replacement, only specially prepared
Push the spade terminals onto the spare mains lead obtained from
correct pins of the connecting strip Barloworld Scientific should be used.
(refer to fig. 6).
DO NOT SWITCH ON ELECTRICITY
18 Refit the cabinet end panel with all six SUPPLY.
screws and washers and the cabinet
roof with all six countersunk screws. 21 Connect the reservoir level control
Replace the perspex viewing screen. WLS as detailed on page 14 of the
manual.
You have now assembled your
Aquatron Water Still. Move the 22 Refer to page 9 for operating
assembled A8000 to its working instructions.
location – using the mounting
bracket, item 7 if the unit is to be wall
mounted. Ensure there is a space of at Blue - neutral
least 50cm to the right of the unit to Brown - live
allow access to the heating elements. Earth - yellow/green
20 Electrical Installation
Relay
THIS EQUIPMENT MUST BE Heater
EARTHED! lead
(front)
The electrical installation should only Thermostat
be carried out by a qualified
electrician.
Fig 6 Terminations for A8000
The A8000 is classed as permanently
connected equipment.
The equipment is supplied with 1.5m
of flexible, triple core, circular cable to
the following specification:- 4.0mm2,
to BS 6500 or equivalent and <HAR>
or BASEC approved.
Connection to the mains electrical
supply, should be via a double pole
isolation switch with a continuous
current carrying capacity of 30A at
250v and overcurrent protection
should be provided by a double pole
approved fuse rated at 30A.
These devices should be sited near to
the equipment and clearly marked –
“Disconnect device for Aquatron
Water Still“
The colour code for the mains cable
is:–
Brown – Live
Blue – Neutral
Yellow/Green – Earth
8
A8000 Operation
The following instructions apply to the 6 Turn on the mains water supply and Reservoir Control (fig. 11)
A8000 Aquatron water still where the using the needle valve item 18, fig. 4, is positioned on the collection
water feed to the boiler is via the mains on the side of the cabinet, adjust the reservoir and switches off both the
supply, or a header tank. flow rate to approximately 2 electricity and water supply when the
litres/min. reservoir is full of distilled water. The
If the boiler is to be fed with either de- water supply is only shut off around 7
ionised or pre-treated water, please follow Check that there are no water leaks mins. after the reservoir is full to allow
the instructions on page 15. from the hose connections. the heating element to cool.
1 Before switching on either the mains 7 Observe that the boiler is now filling It is recommended that the operation
electricity or water supply, identify the with water to cover the heating of the flow switch and reservoir
control switches and indicator lights elements. When the water has control be checked on a regular basis.
on the front of the cabinet. reached the pre-set level, the excess
will be discharged to drain. Check a) Simulation of Mains Water
Green Light that the drain water flows freely and Supply Failure
When illuminated this shows that does not “back-up” into the level Turn off the mains water supply at
there is mains supply voltage to the control. the tap. This should switch off the
unit. Under no circumstances should heating element. Turning the tap on
the cabinet end-panel be removed 8 Switch the CLEAN button off by again will switch the heater on.
when this light is illuminated. pressing it for a second time – the b) Simulation of Reservoir Full
White “On Off” Switch switch light goes out. Condition
This is the primary switch for The heaters will start to warm up and Lower the glass pressure bell WFB,
controlling water and electricity inside eventually run at a red glow. If the fig. 11, into the distillate to a depth
the still. The switch illuminates when flow rate is insufficient, the heaters of 150mm. This will switch off the
pressed in the ON position. will not switch on – this will require heating element immediately and
the needle valve to be opened further. the cooling water after 7 minutes.
Amber Light Raising the bell will cause both to
When illuminated, this indicates that After about 2 to 3 mins. of operation
distilled water will emerge from the be re-supplied.
the distillate collection reservoir is full
as detected by the level control. outlet pipe, falling into the collection
reservoir. 12 Switching Off
White “Clean” Switch Push the CLEAN switch to the ON
This shuts off the heating element but 9 To avoid excessive wastage of coolant position. Wait until the residual heat
allows water to flow into the boiler water, make further adjustments to contained in the heating elements has
when the still is being commissioned the needle valve. Slowly reduce the dispersed and no boiling is evident.
or cleaned. When pressed in the flowrate until the flow control switch, Turn off water supply. Push ON/OFF
CLEAN position the switch illuminates. item 15, switches off the heaters – switch to OFF position. Isolate from
then increase the flow until the power mains electricity supply.
2 Carefully check the following:– is restored.
– the appropriate electricity, water
and drain services have been 10 Refit the perspex viewing screen.
provided. If in doubt, consult the
LOCATION & SERVICES section on 11 Safety Cut-Outs
page 1. All Aquatron water stills are protected
by the following safety devices.
– the ON/OFF switch on the control
panel is in the OFF position. Flow Control Switch (item 15)
monitors the flow of coolant water
– the CLEAN switch is pressed in the into the still and shuts off the heating
CLEAN position. elements, if the flow is insufficient.
– the Rotaflo stopcock on the boiler Thermostat Cut-Out (item 23)
level control WL48 is closed. In normal operation the water within
the glass will be at 100°C. In single
– the Reservoir pressure switch is fault conditions, i.e. in the event of
fitted to the distillate collection water supply failure, the content will
vessel. increase to 110°C where upon a
– the dummy shorting plug is fitted resettable thermostat will operate.
to socket item 8, fig. 4. Once switched, the thermostat has to
be manually reset. This is achieved by
3 Switch on the electricity supply – the unscrewing the black knob, fig. 4,
green indicator light will illuminate. item 24, located inside the cabinet
and pressing the reset button.
4 Press the ON/OFF switch to the ON
position – the switch will illuminate.
9
9
7
6 8
11
10
12
13
4 14
5
3
1 2 15
16
17
18
30
24
19
29 28 27 26 25 23 20
22 21
10
A4000D Assembly 3
Fig 8a: Rear
To boiler
level
socket
Your Aquatron A4000D has been designed
with ease of assembly specifically in mind,
but before commencing assembly please
study the illustration, fig. 7 on page 10 to
familiarise yourself with the general lay- Vent pipe
out. During assembly, follow the sequence
1
of instructions and do not connect the
mains electricity supply until directed.
11
16 Refit the metal thermostat probe into Blue - neutral 30 Connect the cold water supply to
glass tube of the boiler. Brown - live connection 18 fig. 7.
Earth - yellow/green
Note: The tubing selected should
17 Take the one metre length of 16mm
have a safe working pressure at least
tubing and insert through the hole, N
E
equal to the pressure of the water
item 22 in the cabinet side. The
supply. The tubing should be
tubing should extend approximately PCB
L Heater adequately secured by hose clips.
250mm into the cabinet, roughly level lead
with the outlet of the condenser. (back) DO NOT SWITCH ON.
12
A4000D Operation
The following instructions apply to the – The dummy shorting plug is fitted 10 To avoid excessive wastage of coolant
A4000D double distillation stills where the to socket item 7, fig. 7. water, make further adjustments to
water feed to the first stage boiler is either the needle valve, item 18. Slowly
via the mains supply or a header tank. 3 Switch on the electricity supply. reduce the flowrate until the flow
(Note: the green indicator light will control, item 15, switches off the
If it is desired to feed the first stage boiler
illuminate). heaters - then increase the flow until
with deionised or pre-treated water to
the power is restored.
obviate cleaning, then the instructions on
4 Press the ON/OFF switch to the ON
page 15 should be followed.
position (switch will illuminate). 11 Safety Cut-Outs
All Aquatron water stills are protected
1 Before switching ON either mains
5 Re-check that the Clean switch is in by the following safety devices.
electricity or water supply, identify the
the ‘ON’ position (switch will
control switches and indicator lights Flow Control Switch (item 15)
illuminate).
on the front of the cabinet. monitors the flow of coolant water
into the still and shuts off the heating
Green Light 6 Turn on the mains water supply and
elements, if the flow is insufficient.
When illuminated this shows there is using the needle valve, item 18, fig. 7,
mains supply voltage within the unit. on the side of the cabinet, adjust the Thermostat Cut-Out (item 24)
In no circumstances should the flow rate to approximately 2 In normal operation the water within
cabinet end-panel be removed when litres/minute. the glass will be at 100°C. In single
this light is illuminated. fault conditions, i.e. in the event of
Check there are no water leaks from
water supply failure, the content will
White “On/Off” Switch the hose connectors.
increase to 110°C where upon a
The primary switch for controlling
resettable thermostat will operate.
water and electricity inside the still. 7 Observe that the front boiler is now
Once switched, the thermostat has to
The switch illuminates when pressed filling with water to cover the heating
be manually reset. This is achieved by
in the ON position. elements. When the water has
unscrewing the black knob, fig. 7,
reached the pre-set level excess will be
Amber Light item 26, located inside the cabinet
discharged to drain. Check that the
When illuminated, this indicates that and pressing the reset button.
drain water flows freely and does not
the distillate collection reservoir is full, “back-up” into the level control. Reservoir Control (fig. 11)
as detected by the level control.
is positioned on the collection
White “Clean” Switch 8 Press the Clean button to ‘OFF’ reservoir and switches off both the
This shuts off the heating elements position. (Switch light should go out). electricity and water supply when the
but allows water to flow into the reservoir is full of distilled water. The
The heater on the front boiler should
boiler. Used for filling the boiler on water supply is only shut off around 7
start to warm up and eventually run
commissioning and cleaning purposes. mins. after the reservoir is full to allow
with a red glow. After a short time
The switch illuminates when pressed the heating elements to cool.
distilled water will emerge from the
in the ‘ON’ position. outlet pipe item. This should feed into It is recommended that the operation
the empty rear boiler via the level of the flow switch and reservoir
2 Carefully check the following:– control W4L2A. control be checked on a regular basis.
– That the appropriate electricity,
9 Allow the rear boiler to fill with single a) Simulation of Mains Water
water and drain services have been
distilled water (for approx. 20 to 30 Supply Failure
provided. If in doubt,– consult the
minutes), keeping an eye on the level. Turn off the mains water supply at
“Location and Services” notes on
Note that the heater of the second the tap. This should switch off the
page 1.
stage only comes on when the level of heating element. Turning the tap on
– Water still ON/OFF switch is in the single distilled water in the rear boiler again will switch the heaters on.
OFF position. is at about the equator. It may be b) Simulation of Reservoir Full
i.e. – switch should not be pressed necessary to adjust the level of the Condition
and should not illuminate. float switch of the WL42A to achieve Lower the glass pressure bell WFB,
this condition. Refer to fig. 10. fig. 11, into the distillate to a depth
– Water still CLEAN switch is in the of 150mm. This will switch off the
ON position. Lifting the guide tube increases the
level at which the heating element heating element immediately and
– The Rotaflo stopcock on the front will switch on. the cooling water after 7 minutes.
boiler level control WL48 is closed. Raising the bell will cause both to
Lowering the guide tube decreases be re-supplied.
– The float switch on the second the level at which the element will
stage boiler level control W4L2A is switch on. 12 Switching Off
set at the factory. However, check Push the CLEAN switch to the ON
that this has not been disturbed Double distilled water ultimately flows
position. Wait until the residual heat
during transit. A rough check is that from outlet pipe, item 19. This should
contained in the heating elements has
the distance from the top of the fall away from the still with no kinks,
dispersed and no boiling is evident.
large screw cap of the W4L2A to restrictions or U bends.
Turn off water supply. Push ON/OFF
the top of the glass guide tube switch to OFF position. Isolate from
should be approximately 15-20mm mains electricity supply.
see fig. 10.
13
Distillate Collection 3 Take the bacteriological filter and
connect this to the top of the glass
A4000, A8000, feed-pipe using the vinyl tubing
provided. Fit the feed pipe and filter
A4000D into the reservoir lid via the large
screwcap fitting.
Each Aquatron water still is supplied with a 4 Fit the Reservoir Level Control WLS to
RESERVOIR LEVEL CONTROL, WLS, fig. 11. the lid via the small screwcap fitting.
This automatically switches off the heating Note that the stem of the glass
elements and coolant water supply when pressure bell should extend
the distillate collection reservoir is filled approximately 50mm above the
with distilled water. screwcap (see illustration).
Note that, the control is fitted with a 5 Place the assembled lid into the
‘delay timer’ which switches off the socket of the reservoir.
coolant water supply approximately 7
minutes after the heating elements. This is 6 Take the rubber tubing from the
to allow time for the heaters to cool and reservoir level control WLS and
for boiling to cease. connect to the small plastic nozzle,
The control also automatically switches on item 13, on the side of the water still
the heating elements and coolant water cabinet.
140mm
supply when the level of distillate in the
collection reservoir falls. 7 Take the vinyl plastic tubing from the
distillate outlet of the water still and
It is possible to fit the glass pressure bell of connect to the side-arm of the feed
the control to most types of reservoir and pipe on the reservoir lid. Ensure that
container by means of a metal clip WFC/1 Fig 11 the tubing falls down to the reservoir
which is provided. WLS Reservoir level control without any restrictions, kinks or U
Alternatively, Barloworld Scientific offer a bends.
purpose designed reservoir collection
system WR20 fitted with a special adapter
to hold the bell, see fig. 12.
To pressure switch
Fitting the Reservoir Control WLS to (Item 13 on Aquatron)
Collection Vessels Reservoir System
1 Position the distillate collection WR20 & WS20 50mm
reservoir in a suitable location. This Distillate
from
should be below the water still so that Aquatron
the vinyl rubber tubing transporting Barloworld Scientific offer a purpose
the distilled water from the still can designed reservoir system for the
WLS
fall down to the reservoir without collection of distilled water from your
restrictions, kinks or U bends. Aquatron. This comprises:–
WR20 – 20 litre Pyrex glass reservoir
2 Take the Reservoir Level Control WLS
complete with Rotaflo stop-cock
and connect the free end of the
and special lid.
rubber tubing to the nozzle, item 13,
of the reservoir pressure switch WS20 – Reservoir stand, 43cm high for
located on the side of the cabinet. use with WR20.
The lid of the reservoir has a side arm for
3 Use the metal clip to fit the glass
connection to the distillate feed from the
pressure bell inside the distillate
Aquatron and a vent fitted with a
collection reservoir. Refer to fig. 11.
bacteriological filter to maintain the purity
of the water. A screwthread fitting is also
4 Position the bell so that the open end
provided to support the Reservoir Level
is 140mm below the desired distillate
Control WLS.
level.
Location & Assembly
1 Locate the reservoir in a position
which is below the level of the water
still.
14
Connection of 8 Take the 220mm length of reinforced
plastic tubing from the WCK/N kit.
Operation
Aquatron stills to Connect the free-end to the nozzle of
the solenoid valve, item 1, fig 13. 1 Before switching on either the mains
deionised and Secure with metal hose clip. Fit the electricity or water supply, identify the
other end of the tubing to the side control switches and indicator lights
pre-treated water arm of the Level Control WL48, item on the front of the cabinet.
supplies 3, using the screwthread connector.
Green Light
9 Refit the front switch panel and When illuminated this shows that
secure with screw. there is mains supply voltage to the
Many laboratories prefer to feed their
unit. Under no circumstances should
water stills with deionised or pre-treated
10 Take the plastic screwcap, hose the cabinet end-panel be removed this
water to obviate the need for boiler
connector and rubber ring from the light is illuminated.
cleaning and de-scaling.
WCK/N kit. Fit to the deionised water White ‘ON/Off’ Switch
Your new Aquatron can be easily inlet, item 2, fig. 13, having first This is the primary switch for
converted to deionised water-feed by use removed the red protective cap. controlling water and electricity inside
of a simple accessory – the WATER FEED
the still. The switch illuminates when
CONVERSION KIT, Catalogue No. WCK/N. 11 Unscrew the screwcap, item 4, fig. 13, pressed in the ON position.
This kit permits the still to operate with from the Level Control WL48. Remove
most types of deioniser and also piped the glass funnel WF48 and the rubber Amber Light
supplies of pre-treated water. Please ring which are no longer required. When illuminated this indicates that
contact Barloworld Scientific or your the distillate collection reservoir is full
laboratory trade house for ordering and 12 Take the glass float, guide tube and as detected by the level control.
price information. washer assembly, item 5, and pass the White ‘Clean’ Switch
The sequence of operations to fit the electric lead and plug through the This shuts off the heating element but
WCK/N is detailed below:– hole in the screwcap. Carefully push allows water to flow into the boiler
the washer into the cap. when the still is being commissioned
1 Assemble the Aquatron as described
or cleaned. When pressed in the
under the relevant ASSEMBLY 13 Refit the screwcap and float assembly CLEAN position the switch illuminates.
instructions. to the Level Control WL48 ensuring
that the end of the guide tube locates 2 Carefully check the following:–
A4000 refer to page 3
in the well at the base.
A8000 refer to page 7
– the appropriate electricity, water
A4000D refer to page 11
14 Remove the dummy shorting plug and drain services have been
from the socket item, 6, fig. 13, and provided. If in doubt, consult the
2 Ensure the unit is isolated from the
fit the plug from the Float Switch. LOCATION AND SERVICES section
mains electricity supply.
on page 1.
15 Replace the cabinet end-plate and
3 Remove the perspex viewing screen, – the ON/OFF switch on the control
roof.
the cabinet end-panel and the cabinet panel is in the OFF position.
roof.
16 Connect the hose connector, item 2, – the CLEAN switch is pressed in the
of the deionised water inlet to the CLEAN position.
4 Lift out the front switch panel from
outlet of the deioniser or pre-treated
the cabinet groove by unscrewing the – the Rotaflo Stopcock on the boiler
water supply. Note the tubing selected
counter sunk screw at the top of the level control WL48 is closed.
should have a safe working pressure
panel. This will allow better access to
of at least equal to the pressure of the – the Reservoir pressure switch is
the plumbing system.
water supply and should be fitted to the distillate collection
adequately secured with hose clips. vessel.
5 Identify the length of 9mm vinyl
tubing which connects the side arm of – the dummy shorting plug has been
the Boiler Level Control WL48 and the removed from it’s socket and
outlet of the Water Flow Control replaced with the lead from the
Switch (WS48/1 or WS48/2). Remove WCK/N.
the tubing by unscrewing the plastic
screwthread connector from the 6
WL48 and then severing the tie strap
on the Flow Switch.
15
3 Switch on the electricity supply – the
green indicator light will illuminate.
Cleaning 9 Allow the chemical reaction within the
boiler to continue until all the deposits
position – the switch will illuminate. A4000D Note – if scale has built up above the
operating level, the boiler may be
completely filled by either lifting the
5 Recheck that the CLEAN switch is in When any of the Aquatron stills have been drain pipe or restricting the flow of
the CLEAN position and illuminated. used to produce distilled water directly water to drain. This operation must be
from a mains water supply, there will carefully controlled so that the boiler
6 Turn on the mains water supply and inevitably be a build-up of scale in the does not over-fill and cause water to
using the needle valve on the side of boiler and on the heating element(s). To be forced over the funnel top.
the cabinet, adjust the flow rate to obtain optimum performance from the still
approximately 1 Litre/Min. for A4000 the scale should be removed on a regular 10 Open Rotaflo stopcock and allow
and 2 Litres/Min. for A8000 and basis. The time span between cleaning contents of the boiler to drain
A4000D. Ensure that the mains water depends entirely upon the hardness of the completely.
flows to drain without any restriction. water supply. In very hard water areas it
may be necessary to clean the still once a Note:
7 Turn on the supply of deionised or week, whereas in soft water areas several If the acid added to the boiler has not
pre-treated water. Observe that weeks may elapse before cleaning is been completely neutralised, the
deionised water enters the boiler via necessary. liquid flowing to drain may be
the float system of the WCK/N. strongly acid. Necessary safety
It should not, of course, be necessary to precautions should be observed
8 Allow deionised water to flow into de-scale the second stage boiler on the around the drain position, and any
the boiler until the heating element(s) A4000D or when the stills have been effluent control procedures followed.
is covered by approximately 10mm. At operated from a de-ioniser.
this level the flow of deionised water 11 Close Rotaflo stopcock.
Cleaning the still involves the use of
into the boiler should be automatically
concentrated hydrochloric acid. As a safety
cut off by the WCK/N float system. 12 Push ON/OFF switch to ON position
precaution protective clothing, gloves,
mask and goggles should be worn during and allow boiler to fill with water.
9 Switch off the CLEAN button by Again the boiler may be completely
cleaning.
pressing it for a second time – the filled by following procedure detailed
switch light goes out. Note: in step 8.
The cabinet and perspex screen should be
The heater(s) will start to warm up
cleaned using a dilute detergent solution 13 Push ON/OFF switch to OFF position.
and eventually run at a red glow. (If
only.
heater fails to glow, check that the
14 Drain the boiler by opening the
flow of mains water is adequate). It is possible to descale the Aquatron
Rotaflo stopcock.
Water Still without dismantling the
10 Allow distillation to commence and glassware by following these instructions
15 Repeat steps 11-14 two or three times
observe that the WCK/N float system in conjunction with Control of Substances
until the boiler has been thoroughly
allows fresh deionised water to enter Hazardous to Health regulations (COSHH)
flushed through.
the boiler and maintain a satisfactory 1988.
operating level.
Method 16 The still is re-started by finishing the
If the deionised supply fails then the 1 Push the CLEAN switch to the CLEAN cleaning cycle at step 12, and then
WCK/N will switch off the heater until position – the switch illuminates. pushing the CLEAN switch to the OFF
it is restored. position. (Switch is not illuminated).
2 Observe that the heating element/s
Before collecting the distillate the still
11 Replace the perspex viewing screen. has switched off. Allow boiling to stop
should be allowed to run for about 10
and water to cool.
minutes with the distillate running to
drain, this will ensure that any residual
3 Push ON/OFF switch to OFF position.
acid is removed.
4 Remove perspex screen. Note:
The amount of acid required will
5 Open the Rotaflo stopcock on the depend upon the degree of scaling.
level control WL48, allow the boiler to If after performing the cleaning cycle
empty to 3/4’s of its full capacity, and deposits are left in the boiler, it may
close stopcock. be necessary to repeat the cleaning
cycle using a fresh quantity of acid.
6 Carefully add about 100ml. of
concentrated hydrochloric acid into The benefits gained from regular
the glass funnel of the WL48. cleaning cannot be over emphasised.
16
Fault Finding 3 Heater(s) not on –
Water not flowing
8 Cleaning blocked water filters
Your Aquatron is fitted with a built-in
A4000, A8000, Mains green indicator light on
water filter to prevent the ingress of
A4000D Check that the thermostat has not
operated. Confirm this by pressing the
particulate matter into the plumbing
system.
thermostat reset button located inside
The filter can be examined and
In the event of operating difficulties with the main chamber of the cabinet.
cleaned if necessary by unscrewing
your Aquatron it is suggested that the
If the thermostat has cut-out then the the black plastic screwcap which
following basic checks are made. (Please
cause should be identified at once. secures the needle valve, item 18, to
note that these checks should only be
the right-hand side of the cabinet.
carried out by suitably qualified Check that the boiler contains
personnel). sufficient water. A similar filter is also incorporated on
the same side of the cabinet for use
If these checks fail to identify and remedy Check that the plumbing connections
with pre-treated water supplies.
the problem then you are advised to seek are not leaking.
Access is via the black screwcap which
the help of your supplier or the Technical
Check that the stopcock on the boiler secures the hose nozzle.
Services Department of Barloworld
Scientific Ltd. level control WL48 is closed.
Check the electricity supply to the Check the setting of the boiler level Additional checks
Aquatron and ensure that a fuse has control W4L2A. See page 12 for
details.
for Aquatrons fed
not blown at the plug or supply fuse
board. (This check should be Check thermostat has not operated. with deionised or
conducted by a QUALIFIED
ELECTRICIAN).
See notes under 3. pre-treated water
5 Water level in boiler too high –
2 Heater(s) not on – Choking of condenser
Mains green indicator light on a) Boiler does not fill –
water flowing Check that the flow of water into the
boiler is not excessive. Control flow by Check that supply of pre-treated
a) Check the Aquatron switches are in adjusting needle valve on right-hand water is switched on.
the correct mode, i.e.: side of cabinet. Check input water filter on side of
ON/OFF switch is in the ON Check that the coolant water drain cabinet is not blocked (see note 8).
position. pipe falls away from the Aquatron Check Water Feed Conversion Kit
CLEAN switch is in the OFF position. without restrictions or U bends. WCK/N is correctly fitted and
If level of water in the second stage adjusted. See page 14 for details.
Check that the Reservoir indicator
light is NOT illuminated and boiler of model A4000D is too high, b) Heater(s) not on –
indicating a full collection reservoir. then:–
Check water level in boiler is above
b) Check that the 15 amp fuses, Check setting of the float switch in the heating element.
located inside the electrical boiler level control W4L2A. See page
12 for details. Check Water Feed Conversion Kit
chamber of the cabinet, have not WCK/N is correctly fitted and
blown. (This check should be adjusted. See page 14 for details.
conducted by a QUALIFIED 6 Distillate quality low –
ELECTRICIAN). Check the condition of the boiler and
c) Check that the flow of cooling clean if heavily scaled.
water is sufficient.
7 Distillate collection reservoir
A4000 requires a flowrate of at floods –
least 1 litre/min.
Check the location of the Reservoir
A8000, A4000D requires a flowrate level control WLS inside the reservoir.
of at least 2 litres/min. See page 13 for details.
If the flowrate is NOT sufficient Check that the WLS pipe connections
then:– between the glass bell and Aquatron
are secure and air tight.
Check the mains water supply is
turned on.
Check the needle valve on the
right-hand side of the cabinet is
open.
Check the water inlet filter is not
blocked (see note 8).
Check the flow of water is not
impeded by blockages or
restrictions.
17
List of Spares
Catalogue no.
Components:–
20 litre glass reservoir 1569/22
Bacteriological filter WR20/01
Lid/adapter WR20/02
Feed pipe WR20/03
Stopcock WR20/04
Stopcock retaining clip KC29
18
Circuit diagram for A4000
RESERVOIR
ON/OFF CLEAN FULL
T1 (S) (WHITE) (AMBER)
3,6,9 (WHITE)
IC1
C1 C2 RL4
MR1 1000uF 33uF (E)(S)
25V 16V
2,5,7,8,11 1N4007
D1 12 - WAY
1N4007
1 SOCKET
+12V D3 D2 10
T 125 m AL 1N4007
CLEAN
1N4007 PRESSURE
(S) SWITCH
R1 R2
68K 1K (S)
RL2 RL1
(S) (S) C3
2200uF
TR1 16V
FLOW
(SEE NOTE)* SWITCH
SKT1/2 (S)
4 1 2 3 12 12 - WAY
SOCKET 3
2 4
4 - WAY SOCKET 1 5
(S) A4000/PCB (S)
*NOTE
SKT1 IS USED IN CONJUNCTION WITH THE DE-IONISED FEED CONVERTION KIT, WCK/N.
IF THE STILL IS NOT FITTED WITH WCK/N, PINS 1 & 5 ARE CONNECTED TOGETHER
WITH THE 'DIN' PLUG SUPPLIED WITH THE STILL.
ON/OFF PINS 1 & 5, BOILER WATER LEVEL REED SWITCH.
(S)
2
3
4 PINS 2 & 4 DE-IONISED FEED SOLENOID REED SWITCH.} WHEN USED WITH WCK/N
1 5
SKT1/1 (SEE NOTE)* (E) DENOTES E.M.C. CRITICAL COMPONENT
RL3/1 (E)(S)
THERMOSTAT
E
(S)
SV1 (S) SV2 (S) HEATER
COOLING DE - IONISED MAINS ON
WATER WATER (GREEN)
SOLENOID SOLENOID (S)
PINK
10 (E) DENOTES E.M.C. CRITICAL
11 COMPONENT
(S) DENOTES SAFETY CRITICAL
VIOLET WHITE COMPONENT
12
BLACK 4 5
7 3
2 1
SKT1 (SEE NOTE)*
RED YELLOW
9
1 2
PRESSURE
RED SWITCH
3
CLEAN (S)
RED SWITCH
12 - WAY PLUG 3
4
YELLOW WHITE
WHITE
1
RED
6
RESERVOIR
ON/OFF CLEAN FULL
BLACK
BLACK BLACK
5
1N4007
1 FLOW
BLACK PINK SWITCH
2 2 (S)
RL3 (E)(S) BLUE
3
YELLOW SV2 (S)
3 4
ORANGE BLUE
2 5
4 - WAY PLUG BLUE 6 SV1 (S)
BLUE
4
BROWN BROWN MAINS ON
1 (S) BLUE
ON/OFF
(S) BROWN L
THERMOSTAT (S)
*NOTE BROWN BROWN BLUE N
SKT1 IS USED IN CONJUNCTION WITH THE DE-IONISED FEED CONVERTION KIT,
WCK/N RL3/1 (E)(S) HEATER
IF THE STILL IS NOT FITTED WITH WCK/N, PINS 1 & 5 ARE CONNECTED
TOGETHER WITH THE 'DIN' PLUG SUPPLIED WITH THE STILL.
PINS 1 & 5, BOILER WATER LEVEL REED SWITCH
PINS 2 & 4, DE-IONISED FEED SOLENOID }
WHEN USED WITH WCK/N
19
Circuit diagram for A8000
RESERVOIR
ON/OFF CLEAN FULL
T1 (S) (WHITE) (AMBER)
3,6,9 (WHITE)
IC1
C1 C2 RL3 RL4
MR1 1000uF 33uF (E)(S) (E)(S)
25V 16V
2,5,7,8,11
1N4007 1N4007
D1 12 - WAY
1N4007
1 SOCKET
+12V D3 D2 10
T 125 m AL 1N4007
CLEAN
1N4007 PRESSURE
(S) SWITCH
R1 R2
68K 1K (S)
RL2 RL1
(S) (S) C3
2200uF
TR1 16V
FLOW
(SEE NOTE)* SWITCH
SKT1/2 (S)
12 - WAY
4 1 2 3 12 SOCKET
3
2 4
4 - WAY SOCKET 1 5
(S) A4000/PCB (S)
*NOTE
SKT1 IS USED IN CONJUNCTION WITH THE DE-IONISED FEED CONVERTION KIT, WCK/N.
IF THE STILL IS NOT FITTED WITH WCK/N, PINS 1 & 5 ARE CONNECTED TOGETHER
WITH THE 'DIN' PLUG SUPPLIED WITH THE STILL.
ON/OFF PINS 1 & 5, BOILER WATER LEVEL REED SWITCH.
(S)
2
3
4 PINS 2 & 4 DE-IONISED FEED SOLENOID REED SWITCH. } WHEN USED WITH WCK/N
1 5
SKT1/1 (SEE NOTE)* (E) DENOTES E.M.C. CRITICAL COMPONENT
RL3/1 (E)(S)
E
SV1 (S) SV2 (S) FRONT
COOLING DE - IONISED HEATER
MAINS ON
WATER WATER (GREEN)
SOLENOID SOLENOID RL4/1 (E)(S) (S)
REAR THERMOSTAT
HEATER (S)
PINK
10
(E) DENOTES E.M.C. CRITICAL
11 COMPONENT
(S) DENOTES SAFETY CRITICAL
COMPONENT
VIOLET WHITE
12
BLACK 4 5
7 3
1 SKT1 (SEE NOTE)*
2
RED YELLOW
9 PRESSURE
SWITCH
1 2 (S)
RED
3
CLEAN
RED SWITCH
12 - WAY PLUG 3 4
WHITE
1 YELLOW WHITE
RED
6
ON/OFF RESERVOIR
CLEAN FULL
BLACK
BLACK
5
1N4007
BLACK PINK
2
RL3 (E)(S)
REAR BOILER 1 FLOW
1N4007 FLOAT SWITCH
BLACK SWITCH
8 ORANGE PINK (S)
2
RL4 (E)(S) BLUE
3
YELLOW SV2 (S)
3 4
ORANGE BLUE
2 5
SV1 (S)
4 - WAY PLUG BLUE BLUE 6
4
BROWN BROWN
1 MAINS ON BLUE
(S)
ON/OFF
(S) BROWN L
*NOTE BROWN BROWN BLUE N
SKT1 IS USED IN CONJUNCTION WITH THE DE-IONISED FEED CONVERTION
KIT, WCK/N. RL3/1 (E)(S)
FRONT HEATER
IF THE STILL IS NOT FITTED WITH WCK/N, PINS 1 & 5 ARE CONNECTED
TOGETHER WITH THE 'DIN' PLUG SUPPLIED WITH THE STILL. BROWN
PINS 1 & 5, BOILER WATER LEVEL REED SWITCH WHEN USED
PINS 2 & 4, DE-IONISED FEED SOLENOID }
WITH WCK/N
THERMOSTAT (S)
BROWN BROWN BLUE
20
Circuit diagram for A4000D
RESERVOIR
ON/OFF CLEAN FULL
T1 (S) (WHITE) (AMBER)
3,6,9 (WHITE)
IC1
C1 C2 RL3 RL4
MR1 1000uF 33uF (E)(S) (E)(S)
25V 16V
2,5,7,8,11
1N4007 1N4007
D1 12 - WAY
1N4007
1 SOCKET
+12V D3 D2 10
CLEAN REAR
T 125 m AL 1N4007 1N4007 PRESSURE BOILER
(S) SWITCH FLOAT
R1 R2
68K 1K (S) SWITCH
RL2 RL1
(S) (S) C3
2200uF
TR1 16V
FLOW
(SEE NOTE)* SWITCH
SKT1/2 (S)
12 WAY
4 1 2 3 12 SOCKET 3
2 4
4 - WAY SOCKET 1 5
(S) A4000/PCB (S)
*NOTE
SKT1 IS USED IN CONJUNCTION WITH THE DE-IONISED FEED CONVERTION KIT, WCK/N.
IF THE STILL IS NOT FITTED WITH WCK/N, PINS 1 & 5 ARE CONNECTED TOGETHER
WITH THE 'DIN' PLUG SUPPLIED WITH THE STILL.
ON/OFF PINS 1 & 5, FRONT BOILER WATER LEVEL REED SWITCH.
(S)
2
3
4 PINS 2 & 4 DE-IONISED FEED SOLENOID REED SWITCH. }
WHEN USED WITH WCK/N
1 5
SKT1/1 (SEE NOTE)* (E) DENOTES E.M.C. CRITICAL COMPONENT
RL3/1 (E)(S)
TT1(S)
E
SV1 (S) SV2 (S) FRONT FRONT
COOLING DE-IONISED HEATER THERMOSTAT MAINS ON
WATER WATER (GREEN)
SOLENOID SOLENOID RL4/1 (E)(S) (S)
TT2(S)
REAR REAR
HEATER THERMOSTAT
PINK
10
(E) DENOTES E.M.C. CRITICAL
11 COMPONENT
(S) DENOTES SAFETY CRITICAL
COMPONENT
VIOLET WHITE
12
BLACK 4 5
7 3
1 SKT1 (SEE NOTE)*
2
RED YELLOW
9 PRESSURE
SWITCH
1 2 (S)
RED
3
CLEAN
RED SWITCH
12 - WAY PLUG 3 4
WHITE
1 YELLOW WHITE
RED
6
ON/OFF RESERVOIR
CLEAN FULL
BLACK
BLACK
5
1N4007
BLACK PINK
2
RL3 (E)(S)
REAR BOILER 1 FLOW
1N4007 FLOAT SWITCH
BLACK SWITCH
8 ORANGE PINK (S)
2
RL4 (E)(S) BLUE
3
YELLOW SV2 (S)
3 4
ORANGE BLUE
2 5
SV1 (S)
4 - WAY PLUG BLUE BLUE 6
4
BROWN BROWN
1 MAINS ON BLUE
(S)
ON/OFF
(S) BROWN L
FRONT THERMOSTAT (S)
*NOTE BROWN BROWN BLUE N
SKT1 IS USED IN CONJUNCTION WITH THE DE-IONISED FEED CONVERTION
KIT, WCK/N. RL3/1 (E)(S)
FRONT HEATER
IF THE STILL IS NOT FITTED WITH WCK/N, PINS 1 & 5 ARE CONNECTED
TOGETHER WITH THE 'DIN' PLUG SUPPLIED WITH THE STILL. BROWN
PINS 1 & 5, FRONT BOILER WATER LEVEL REED SWITCH WHEN USED
PINS 2 & 4, DE-IONISED FEED SOLENOID }
WITH WCK/N
REAR THERMOSTAT (S)
BROWN BROWN BLUE