You are on page 1of 4

New technology

Surg Endosc (1999) 13: 528–531 © Springer-Verlag New York Inc. 1999

Trocar and instrument positioning system TISKA


An assist device for endoscopic solo surgery
M. O. Schurr,1 A. Arezzo,1 B. Neisius,2 H. Rininsland,2 H.-U. Hilzinger,3 J. Dorn,3 K. Roth,1 G. F. Buess1
1
Section for Minimally Invasive Surgery, Department of General Surgery, Eberhard-Karls University, Waldhörnlestrasse 22, D-72072
Tübingen, Germany
2
Karlsruhe Research Center, Department of Technical Engineering, Karlsruhe, Germany
3
Karl Storz Endoscopy GmbH & Co., Tuttlingen, Germany

Received: 29 December 1997/Accepted: 24 July 1998

Abstract. The assistance received by the surgeon from sup- are performed with the help of mechanical or robotic de-
port personnel during surgical laparoscopy is extremely im- vices not only in general surgery, but also in neurosurgery
portant. This includes the retracting of instruments and the [1, 3, 4], orthopedics [5, 10], and urology [2].
positioning of the endoscope. However, human assistance is A new instrument positioning system is described, re-
costly and often does not provide satisfaction for the sur- cently developed in cooperation with the Karlsruhe Re-
geon. The aim of the project was to develop a mechanical search Center and Karl Storz Endoscopy, Tuttlingen, Ger-
arm capable of manipulating a laparoscopic instrument un- many. It is provided with a characteristic geometry that
der the control of the operating surgeon. The system design should improve the efficiency of laparoscopic surgical ma-
is based on a particular kinematic construction that main- neuvers by providing assistance to the surgeon.
tains an invariant point of constraint motion coincident with
the trocar puncture site through the abdominal wall. The
guidance system allows transparent and intuitive operation, Materials and methods
and its setup is easy and quick. It may be adapted either as
an instrument retractor or as an optic positioning device. A Technology
new generation of instrument positioning systems, with im-
proved ergonomy, will be a first step toward the diffusion of The aim of the project was to develop a mechanical arm capable of ma-
solo surgery techniques in minimally invasive therapy. We nipulating and fixing a laparoscopic instrument in position under the con-
believe this prototype represents a valid compromise be- trol of the operating surgeon. The geometry of an instrument guidance
system for laparoscopic surgery should respect the principle of the invari-
tween human and robotic control for conventional laparo- ant point of motion [7, 9] where the trocar enters the adominal cavity.
scopic instruments. For the Trocar and Instrument Positioning System Karlsruhe (TISKA)
project, a special kinematic principle was adopted that establishes a remote
Key words: Endoscopic surgery — Retractors — Solo sur- center of instrument motion, ensuring that no lateral force is exerted around
the trocar puncture site. This was achieved by designing a system in which
gery — Surgical robotics the shaft of a laparoscopic instrument can be moved around the point of
trocar insertion as well as rotated and translated in its longitudinal axis. The
principle of maintaining the invariant point of motion through mechanical
constraints has been described by Mueglitz et al. [7]. They described a
robotic motion principle along virtual axes, intersecting at the point of
In recent years minimally invasive surgical techniques have trocar insertion. This is of particular importance for the control scheme of
improved the quality and efficiency of several surgical pro- electrically driven robotic devices [7]. Moreover, it is a helpful kinematic
principle in purely mechanical instrument guidance devices.
cedures. Nevertheless, surgeons sometimes face difficulties
in reaching a stable and optimal position for laparoscopic
instruments and endoscope.
First mechanical retractors and then automated surgical Experimental evaluation
robots were developed to alleviate the problem of instru-
Evaluation of prototypes proceeded in parallel with the technical develop-
ment guidance. At the time of this writing, different tasks ment of the device. We usually assess new technologies under experimen-
tal conditions, using phantom models with integrated animal organs. The
Tübingen Lap-Trainer (Coburger Lehrmittelanstalt, Coburg, Germany)
Correspondence to: G. F. Buess was equipped with porcine liver segments including the gallbladder to
529

Fig. 1. Basic design of the TISKA Endoarm.


Fig. 2. TISKA Endoarm: current prototype.
Fig. 3. The adaptation to the operating table that allows a clear definition of the sterile limit.
530

and the inserted instrument are moved, they are guided pre-
cisely through the invariant point of trocar insertion, with-
out any force being exerted on the abdominal wall. Trans-
lation and rotation around the longitudinal axis of the in-
strument are possible, completing the four degrees of
freedom necessary to guide the instrument in the abdominal
cavity. Furthermore, a spherical wrist joint was included in
the design of the system incorporating the parallelogram
mechanism described earlier. This additional joint is used to
align the position of the device with the anatomy of the
patient.
A prototype of the instrument-positioning system was
created and called the TISKA Endoarm (Fig. 2). The system
is composed of a power supply unit, an operating table
attachment, a positioning device, and a trocar tube adapta-
tion clip. The position of the double parallelogram is locked
and unlocked by electromagnetic brakes controlled by a
footpedal, whereas translation and rotation of the laparo-
scopic instrument are locked by a mechanical clamp.
After the device is attached to the operating table, it can
be covered with a sterile plastic sac. A special technique
allows the precise definition of the sterile area: The carrier
system of the TISKA device is attached to the standard rail
by means of a larger rail screwed under the sterile drape
covering the operating table. This larger rail has rounded
edges to prevent it tearing the drape and to allow easy
translation of the base along the operating table. The whole
arm is covered by a transparent plastic tube fixed to the base
of the carrier system (Fig. 3). Thus a clear limit of the sterile
field is defined. The entire arm is made of stainless steel and
Fig. 4. The invariant point of constraint motion is maintained when the may be gas sterilized if required.
instrument is moved.
Fig. 5. The low space requirements allow the system to be combined with
further positioning systems: the operating field when two TISKA Endoarm Experimental results
systems are used.
The evaluation of the prototype was conducted in phantom
models and confirmed in animal experiments. The overall
simulate laparoscopic cholecystectomy. Conventional laparoscopic operat-
ing room equipment (video unit, HF, etc.) and a regular set of instruments handling of the system was found to be simple and did not
were used. The prototypes were applied to guide assisting instruments for require specific training. The trocar tube can easily be
retracting the gallbladder or to guide the optic during laparoscopic chole- mounted on the instrument guiding device. When the de-
cystectomy. sired position of the instrument is reached, the surgeon may
The experiments were not a standardized series, but tests to clarify the
functionality and handling properties of the new device. Intuitiveness of
lock axes 1 and 2 by simply releasing the pressure on the
handling, practicality of use under operating room conditions, and me- footpedal. This was subjectively judged to be very intuitive
chanical stability of the TISKA system were judged subjectively. The test and easy to perform. It was found to be a particular advan-
results were used to further refine the design of the device until optimal tage of the system that repositioning does not require the
functionality was achieved in the final version. This device was used in involvement of both hands but can be done with only one
three animal experiments (domestic pig) for laparoscopic cholecystectomy
to confirm that the phantom test results could be transferred to surgery. hand.
During phantom and animal experiments, the principle
of the invariant point of motion (Fig. 4) was confirmed, and
Results the device was found to be atraumatic around the trocar
puncture site. Space requirements of the TISKA device at
the operating table are low and allow the combination of
Technological results two systems (Fig. 5) for guiding both the assisting instru-
ment and the laparoscope. For this purpose, both devices
The kinematic design of the system (Fig. 1) comprises a first must be attached to the operating table opposite the surgeon
axis pointing directly through the point of incision. A sec- to avoid conflict.
ond axis intersects the first perpendicularly at the point of
incision. A planar double parallelogram mechanism con-
nects the two axes. This geometry allows movements only Discussion
around an arc whose center corresponds to the point of
incision. Thus, a remote center of motion is established at In recent years, laparoscopic surgery has become increas-
the point where the two axes intersect. When the trocar tube ingly widespread, supported by new developments in in-
531

strumentation. Less attention has been paid to the comfort of tems. The low space requirements of the system allow it to
the first surgeon and his or her assistant, who often are be combined with further positioning arms, even on the
forced into tiring standing positions and monotonous tasks. same side of the patient, which is not always possible with
This can result in an unsatisfactory interaction between the systems moving in a horizontal plane. A combination of two
two, making it difficult to find an optimal working position. TISKA Endoarm systems can be used for both optic and
Moreover, the high costs of the operating theater, even for instrument positioning, thus avoiding the need for an assis-
standard laparoscopic procedures, require the involvement tant surgeon in standard laparoscopic procedures. In phan-
of less experienced fellows, such as residents, which leads tom and animal experiments personally conducted, this
to a further increase in the operating time [11]. Especially in solosurgery solution appeared to be comfortable for the sur-
community hospitals and private institutions, where the role geon. We believe the Endoarm represents a valid compro-
of the surgical assistant is assumed by either assistant phy- mise between human and robotic control [9] of laparoscopic
sicians or trained nurses, the introduction of positioning instruments. Further data concerning the influence of guid-
systems for laparoscopic procedures may alleviate some of ance systems on the course of solo surgery, setup, and in-
the pressure due to limited resources [11]. tervention times will be derived from standardized experi-
Mechanically assisted systems, such as Iron Intern (Au- mental trials currently underway.
tomated Medical Products, New York, NY) and Omnitract
(Minnesota Scientific, St. Paul, MN), incur low costs. On
the other hand, they turn out to be unergonomic as they References
do not offer a true instant release capability to the laparo-
scopic surgeon, and are therefore rarely used [6]. This is 1. Benabid AL, Lavallee S, Hoffmann D, Clinquin P, Demongeot J,
mainly a consequence of their direct derivation from open Danel F (1992) Potential use of robots in endoscopic neurosurgery.
surgery. Pneumatically assisted systems, such as Acta Neurochir Suppl 54: 93–97
2. Davies BL, Hibberd RD, Coptcoat MJ, Wickham JE (1989) A surgeon
Robotrac (Aesculap, Tuttlingen, Germany) or First Assis- robot prostatectomy. A laboratory evaluation. J Med Eng Technol 13:
tant (Leonard Medical, Huntington Valley, PA), also pro- 273–277
vide a significantly increased degree of stability compared 3. Kelly PJ, Kall BA, Goerss S, Earnest F (1986) Computer-assisted
with human assistance, while they do not fatigue or antici- stereotactic laser resection of intra-axial brain neoplasm. J Neurosurg
pate surgical movements during complex operations. Nev- 64: 427–439
ertheless, they require some experience before smooth 4. Koyama H, Uchida T, Funakubo H, Takakura K, Fankhauser H (1990)
Development of a new microsurgical robot for stereotactic neurosur-
working interaction can be achieved [6]. We were able to gery. Stereotact Funct Neurosurg 54/55: 462–467
confirm these impressions from personal experience on 5. Matsen FA, Garbini JL, Sidles JA, Pratt B, Baumgarten D, Kaiura R
phantom models using retractors originally designed for (1993) Robotic assistance in orthopaedic surgery: a proof of principle
open surgery. Therefore we decided to develop a passive using distal femoral arthroplasty. Clin Orthop 296: 178–186
positioning system designed to meet the requirements of 6. Moran ME (1993) Stationary and automated laparoscopically assisted
technologies. J Laparoendosc Surg 3: 221–227
laparoscopy. 7. Mueglitz J, Kunad K, Dautzenberg P, Neisius B, Trapp R (1993).
The geometric design of the new TISKA Endoarm, Kinematic problems of manipulators in minimally invasive surgery.
based on the principle of mechanical maintenance of the Endosc Surg Allied Technol 1: 160–164
invariant point of motion, allows simple, single-handed op- 8. Schurr MO, Breitwieser H, Melzer A, Kunert W, Schmitt M, Voges U,
eration. According to our personal experience, the release of Buess G (1996) Experimental telemanipulation in endoscopic surgery.
Surg Laparosc Endosc 6: 167–175
the brakes through the footpedal does not add significant 9. Schurr MO, Buess G, Rininsland HH, Holler E, Neisius B, Voges U
complexity to system operation. Motion of the abdominal (1996) ARTEMIS-Manipulatorsystem für die endoskopische Chirur-
wall (i.e., as a result of the patient’s breathing), has practi- gie. Endoskopie Heute 9: 245–251
cally no influence on the position of the instrument. The 10. Taylor KS (1993) Robodoc: study tests robot’s use in hip surgery.
position of the trocar sheath is maintained while the instru- Hospitals 67: 46
ment is changed. 11. Traverso LW, Koo KP, Hargrave K, Unger SW, Roush TS, Swanstrom
LL, Woods MS, Donohue JH, Deziel DJ, Simon IB, Froines E, Hunter
Studied mainly as a simple instrument retractor, the J, Soper NJ (1997) Standardizing laparoscopic procedure time and
Endoarm may easily be used as a passive optical positioning determining the effect of patient age/gender or absence of surgical
device, although simpler than advanced robotic guiding sys- residents during operations. Surg Endosc 11: 226–229

You might also like