Professional Documents
Culture Documents
NP – non-pathologic
P- pathologic Paaño, Stephanie Mae | 3MT-3 1
AUBF Microscopic Examination of Urine Sediment Part 1 LECTURE
SPECIMEN PREPARATION è Gently tap repeatedly the lower part of the tube
1. Fresh and properly preserved specimen with the finger to mix the sediment
- Very hard to identify RBCs and WBCs in the
microscope from old specimen !"#$%& "( $)*+& ,&+-)*($.&/
2. Bring the specimen to room temperature before
Concentration factor =
0&/*%&+- 1"#$%&
centrifugation
- Amorphous materials and their crystals è Concentration factor (CF) using 10 mL urine; 1 mL
usually precipitate in cool temperature sediment = CF is 10
- Room temperature = dissolve è Concentration factor (CF) using 10 mL urine; 0.5
3. Mix specimen well before transferring into a mL sediment = CF is 20
centrifuge tube è 0.02 mL (20uL) – volume of sediment examined
- RBCs ad WBCs settle at the bottom of the
tube = false negative result if not properly *Exam Question: How many mL is centrifuged?
mixed - 10-15 mL; average of 12
- How long? 5 minutes
SPECIMEN VOLUME - Decant (bottom’s up)
è 10-15 mL; usually 12 mL - Flick the Tube
è Less volume in pediatrics patients - 0.5 – 1 ml should remain
- Note the volume on the report form - put one drop on a slide – should be measured
è Or perform correction and should be around 0.02 mL (20uL) =
Example: volume of sediment examined
6mL of urine is centrifuged multiply results by 2
è In actual practice, they only put 2/3 of urine in the
CENTRIFUGATION tube, centrifuge for 5 minutes, decant without
è Centrifugation Time: 5 minutes measuring amount of sediment left, flick the
è Speed of the Centrifuge: tube, and no longer measure amount put in the
- 400 RCF (Relative Centrifugal Force coverslip.
- 1,500-2,500 RPM
- RCF is better because RPM depends on the COMMERCIAL SYSTEM: CEN SLIDE
diameter of centrifuge è Closed system
è Convert RPM to RCF - Does not require manual loading of the
centrifuge specimen in the slide
RCF = (1.118 X 10-5) (radius in cm) (RPM2) è Fast, clean, accurate, microscopic urine analysis
è Specially designed tube
è Braking of the centrifuge: False negative
- If you push the brake, the centrifuge will
suddenly stop (DO NOT DO THIS)
- Doing this could disturb or agitate the
sediments = False Negative
- If you decant, the sediments will be included
è Centrifuge urine in capped tubes
- If no cap, use paraffin seal
- Because it forms biohazardous aerosols
- Plastic tube with cover
SEDIMENT PREPARATION - Put urine up to the mark
è 0.5-1 mL should remain in the tube after - Centrifuge it, in less than 1 minute
decantation è Direct reading of the urine sediment
- “Bottom’s up” – technique used in è No decanting, re-suspending, pipetting,
decantation sediment transfer, microscope slides and
- Book says it is better to pipette off coverslips
o Example: 10mL is measured and then - After centrifugation, the sediment will go to
0.5 sediments remained, therefore, the narrow part of the tube and then put it in
pipette off the remaining 9.5 the slide holder, then examine directly under
o Pipetting have the tendency to disturb the microscope
the sediments (sasama ang - Plastic tube is not that clear (vague)
sediments) compared to the glass slide
o Re-suspend with 0.5-1 mL
- If sediments are too many, it clumps in one è Reported quantitatively – bibilangin talaga;
place when using plastic tube; not distributed report number
equally è In other lab, their manner of reporting follows the
range (from lowest to highest)
EXAMINATION OF THE SEDIMENT Example:
è 10-20 fields low (10x) and high (40x) power 1-6; 2-8 = like they are overlapping in terms of
- one peak in the microscope = one field their values
- 10 fields = change or transfer view 10
times RBCs/WBCs / HPF
- in most laboratories, usually only è Average number per 10 high power fields (hpf)
examines 10 fields
- not use oil immersion None
è low power to detect casts (near the edges of 0-2
the coverslip) 2-5
- usually casts can be seen near edges of 5-10
coverslip so you need to examine the 10-25
sides, especially is positive for protein. 25-50
è high power to identify the type of cast 50-100
- Because if you use low power, you >100 (TNTC)
cannot see if it is RBC cast or WBC cast è TNTC – too numerous to count
- RBC cannot be identified under the low è Quantitative
power è If medications depend on number; adjust dosage
è bright field microscopy: subdued light depending on how many are present
è examine the sediment in the correct plane è Proper mixing is very important especially when
- wrong plane = more artifacts there are outliers
- use an epithelial cell as a point of - If Vortex Mixer is available to mix the
reference (Squamous epithelial cells – sediment, it is better to use it to achieve
can easily be seen in females than in equal distribution of the microscopic
males) elements
- If you already see a squamous epithelial
cell, do not leave that area anymore; do CRYSTALS / HPF
not move the coarse adjustment knob None 0
anymore, rather only move the fine Rare 0-2
adjustment knob. Few 2-5
Moderate 5-20
MANNER OF REPORTING Many >20
REFERENCE INTERVALS
è WBC MICROSCOPY
• Cytoplasm – purple granules
• Nuclei- purple
è Squamous
• Nucleus – orange-purple
• Cytoplasm – light purple or blue
è Bacteria
• (motile) – unstained
• (non-motile) – purple
- Board Exam: Which of the following may be
unstained using Sternheimer-Malbin?
o Bacteria when alive resist stain PARTS OF A MICROSCOPE
Prussian Blue
è Stains structures containing iron yellow-brown
granules of hemosiderin in cells and casts
è Stains hemosiderin – blue
Gram Stain
Bacterial Casts
è Important in differentiating bacterial casts from è It is important to use both eyes because using
granular casts one eye may cause strain.
è Use dried, heat-fixed preparation of the urine
sediment Objectives
è Examination of urine sediment:
- 10× and 40×
è Coarse knob: initial focusing è Light waves pass through the specimen and
è Fine-focusing knob: sharpen the image enter the objective in an inverted cone.
è The higher the numerical aperture:
2 Features of Objective Lenses___________________ • the closer the lens is to the object
1. Parcentered • the better the light-gathering capability of
- Ability to retain the central FOV (when the the lens
user switches from one objective to another, • greater resolving power
it still stays in the center)
2. Parfocal
- Ability of the objective to remain in focus
Illumination system
regardless of the objective used
- Only minor adjustments by fine adjustment Light source
knob - Base of the microscope
Working Distance_____________________________
- Distance between the objective & the coverslip
on the slide
- Decreases as magnification of the objective
increases
- If you are using OIO, don’t use the coarse
adjustment knob because it may possibly break
the slides. Considering that the distance of
working distance is already at its minimum, and
moving the coarse adjustment would only move
it even more near.
Rheostat
- Regulate the intensity of the light
Field Diaphragm
- Controls the diameter of the light beam
Total Magnification___________________________
reaching the slide
è Objective magnification x Ocular Magnification
è Ex: Ocular mag = 10 (pag walang binigay 10 parin Condenser
gagamitin mo) - Below the stage
è LPO = 10x10 = 100 - Focuses the light on the specimen and
è HPO = 40x10 = 400 controls the light for uniform illumination
è OIO = 100x10 = 1000 - Focuses the light towards the specimen
- Needs to be adjusted to achieve optimum
visualization
Aperture diaphragm
- in the condenser controls the amount of light and
the angle of light rays that pass to the specimen
and lens
- Needs to be adjusted to achieve optimum
visualization
Numerical Aperture___________________________
è Has opening where light travels
è Refractive index of the material between the
slide and the outer lens (air or oil) and the angle
of the light passing through it.
è Best percentage is 75%, not too bright, not too Other Ways (Not Recommended)
dim 1. Remove an eyepiece and look down through the
eyepiece tube.
Centering the condenser and Kohler illumination 2. Adjust the aperture diaphragm until
- Provides optimal viewing of the illuminated field approximately 75% of the field is visible
- The condenser and field diaphragms must be 3. Replace the eyepiece
adjusted each time the microscope is used and - In order skip this step, just estimate it to 75%,
each time the objective is changed. because it would still be a hassle to remove it.
- Alignments are adjusted to achieve optimum
visualization Body
è base
è body tube
è nose piece
Mechanical Stage
- Holds the slide on place
Brightfield Microscopy
è Defracted light
- Light that hit the specimen
- Slight bending occurs
è For detection of more translucent/ transparent è Background light
or low- refractile formed elements and living cells - Light that did not hit the specimen
like:
• Hyaline casts è Best contrast
• Mucous threads - When the light that does not pass through
• Mixed cellular casts the specimen is shifted one quarter of a
• Trichomonas wavelength and compared with the phase
• Sperm cells difference of the specimen
è Hardens the outlines of formed elements è Shifted one quater of a wavelength means that
enhances contrast and improves the visibility bumabagal yung light or nagdadarken ng konti
and definition of structures having a refractive
index similar to that of the urine
è Forms a halo of light around the specimen
è Uses condenser ring and phase objective ring
è Hyaline casts (blue picture - under brightfield)
- They are hard to identify
- Center pic: Under brightfield
- Right: Under phase contrast
è The specimens darken even without the stain
because the light weakens.
Condenser Ring
è Light annulus
è Placed on the condenser
è Phase objective ring
- phase plate or dark annulus
è Annulus means area between two concentric
circles
è Placed on the objectives
è These two will be aligned.
Interference Contrast
è Difference in Phase Conrast Miscroscopy is that
in Interference Contrasts, you would distinguish
its thickness, unlike in PCM
è 3D image
è Layer by layer imaging
è Bright against a dark background without halo
Two Types___________________________________
1. Modulation Contrast (Hoffman)
- Uses split aperture, polarizer, filter
2. Differential-interference contrast DIC
Nomarski)
è Uses: Wollaston Prism and Polarizing filter
Analyzer
- Placed between the objectives and the ocular.
Polarized Microscopy
è For lipids (cholesterol) and birefringent crystals
è The lipids/fats are hard to identify when Polarizing filter
unstained, so that’s why stain and Polarized - Placed on the condenser
microscopy are used. - When added with polarizing filter, light will travel
è Dito tinitignan is cholesterol and sa stain is in only one direction from the light source
triglycerides. - Horizontal
Dark-field Microscopy
ISOTROPIC è Light against a black background
è Properties of a material are the same in all è For elements that cannot be easily stained
directions è Spirochetes like Treponema pallidum
è Since they are the same, they are not rotating, so è Condenser of bright-field microscope is replaced
they are not seen under polarized microscope with a dark-field condenser that contains an
opaque disk
ANISOTROPIC è Indirect light is reflected off the object
è When the properties of a material vary with
different orientation
è Could be seen under polarizer
- Positive birefringence
ADDIS COUNT
è Estimate of the formed elements
è 12-hour urine specimen
è You are going to count the elements using
hemocytometer
è First procedure to standardize the quantitation
of formed elements in the urine microscopic
analysis
Normal values
RBCs 0 to 500,000
WBCs and Epithelial 0 to 1,800,000
Cells
Hyaline Casts 0 to 5,000
REFERENCES
NOTES FROM THE DISCUSSION OF:
DEAN MARIA BENILDA DE GUZMAN