A gas analysis system usually consists of two groups of devices: Contents Page
• The gas analyzer measures the concentrations of the sample
Modular Sample Gas Probe System 2
gas components and converts them into an electrical signal.
FE2 Filter Unit 3
• Supplementary devices sample the sample gas from the
Annular Heating Element 4
process, condition it and feed it in the gas analysis system.
Heating Sleeve 5
In most cases, the sample gas sampled from the process cannot PFE2 Filter Unit 6
be processed by the gas analyzer without further conditioning. Too PFE3 Filter Unit 11
high a dust content, temperature and dew point as well as too Selection Table for Probe Tube Types and Probes 15
high or too low pressure, and not least interference components in Type 40, Type 40 W Probe Tubes 16
the sample gas can impair the operational capability of the gas Type 42 Probe Tube (heated) 17
analyzer and falsify the measuring result. Type 60 Probe Tube (Cement Flue Gas Probe) 18
Probe 2 19
Supplementary devices such as the sampling probe, the sample
Probe 11 20
gas line, the sample gas cooler and pumps and filters ensure that
Ceramic Filter 21
the sample gas inlet conditions of the connected gas analyzers are
Process Pressure Reducers 22
satisfied and faultless measuring results are obtained independ-
Heated Sample Gas Lines 24
ently of the process sequence and the local conditions.
CGWB 13 Heated Sample Gas Line 26
TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines 29
Electronic Temperature Controller and Switching Relay 35
Modular Sample Gas Probe System
Use
The FE2 filter unit is a component of the modular gas sampling
probe system. It is used for filtering gases containing dust.
Filter Characteristics
Connection Ports
Filter Element
Sample Gas / Test Gas
Surface filter
G 1/4 internal thread to DIN ISO 228/1
Filter Fineness
Compressed Air (2 Ports)
0.3 µm
G 1/2 internal thread to DIN ISO 228/1
Retention Rate
Probe Tube
99.98 %
G 3/4 internal thread to DIN ISO 228/1
Seal Integrity
10–4 hPa l/s
Dead Volume Compressed Air Supply for Purge Back
280 ml (See PFE2 Filter Unit, page 6)
Instrument Air
Free of dust, water and oil
Materials of the Gas-carrying Parts
Pressure
Casing and Connections
max. 600 kPa
Special steel Mat. No. 1.4571
Flange
Special steel Mat. No. 1.4541 Dimensional Drawing (Dimensions in mm)
Filter
Aluminum oxide surface filter 14
105
Coating
Thermoset plastic, resistant up to 180 °C
130
160
65
Installation
Flange
DN 65, PN 6, Form B to DIN 2573, flange gasket A;
positioning of holes:
0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°, 360° 280 40...65
Mounting
Flange connect or bolt to the probe tube
Ordering Information
Weight
approx. 8.0 kg Catalog No.
Housing and flange SS, filter fineness 0.3 µm 23042-4-0730709
Housing and flange SS, coated, 0.3 µm 23042-4-0730718
Use
For heating the FE2 filter unit, which is a component of the
modular gas sampling probe system (see page 3).
Heating is necessary if condensation is likely from moist gases
and there is consequently risk of the filter becoming clogged
and corroded.
Data
Power Supply
115 VAC or 230 VAC, 50 to 60 Hz
Heater Output
200 W, 100 W or 50 W – depending on setting (see below)
The heating resistors are factory-set for heat output of 100 W
Temperature of Heater (unregulated)
approx. 120 °C to 180 ° C (measured at ambient temperature
of 20 °C, still air)
Ambient Temperature
0 to 55 °C (still air)
Degree of Protection of Housing
IP50
Materials Dimensional Drawing (dimensions in mm)
Housing: aluminum alloy, heating surface: brass
Weight
approx. 1.7 kg
Electrical Connections
Terminals via M20 cable gland. The electrical supply
conductor must be heat resistant (≥ 200 °C).
Installation
Two half shells are placed round the FE2 filter unit to be heated
and closed with a wing screw.
Option
Regulated Heating with 1 x Pt 100
Ordering Information
Heating Circuit Catalog No.
Ring heater 230 V, 50/60 Hz 23029-5-7865312
Ring heater 115 V, 50/60 Hz 23029-5-8008813
Ring heater with Pt 100, 230 V, 50/60 Hz 23029-5-7865316
Ring heater with Pt 100, 115 V, 50/60 Hz 23029-5-7865317
– Insulating Mat
– Temperature Controller (see page 35)
Use
The PFE2 filter unit consists of the FE2 filter unit with a regulat-
ed heating sleeve, installed in a special-steel protective case. It
is used to filter dust-containing gases and is a component of
the modular gas sampling probe system.
When used for gases with high dust content, the PFE2 can
optionally be fitted with valves for automatic cleaning of the
filter element in the FE2 filter unit.
Protective Case
Dimensions
450 x 450 x 400 mm (H x W x D)
Material
Special steel, Mat. No. 1.4301
Degree of Protection of Housing
IP54
Ambient Temperature Control Valve
–20 to +45 °C
Solenoid Valve
Weight 3/2-way solenoid valve
approx. 18 kg
Pressure Range
600 kPa (6 bar)
Options Nominal Diameter
DN 2.0
– Annular heating element, unregulated (see page 4)
– Heating sleeve, regulated (Pt 100), with insulating mat Connections
(see page 5) G 1/8 internal thread DIN ISO 228/1
Material
Brass
Pneumatic Valve
Mode of Operation
Diaphragm Valve Closed in normal position
2/2-way straight-way valve
Ambient Temperature
Operating Pressure / Control Pressure 0 to +100 °C
600 kPa (6 bar)
Electrical Connections
Nominal Diameter Standard connector
DN 12
Power Supply
Connections 115 VAC or 230 VAC, 50 to 60 Hz
G 1/2 internal thread DIN ISO 228/1
Control air: G 1/8 internal thread DIN ISO 228/1 Power Consumption
8 VA
Material
Brass
Mode of Operation
Closed in the normal position
Ambient Temperature
0 to +100 °C
75 110 75
Sample gas connection
PFE2 Filter Unit
M4 M4 M4 M4
Compressed air
65
Purging
ø6,5
Bulkhead fitting 10/12x1 mm
Test gas connection
210
180
Take holes for cable glands on the left
165
ø11,5 Bulkhead fitting 4/6x1 mm
90
into account
Control air
Terminal box with cable glands Bulkhead fitting 4/6x1 mm
50 80 Use as required
Electrical connections
2x M32
3x M25
2x M20
Probe Protective Case Dimensional Drawing (dimensions in mm)
550
155
155
450
440
17
0
Flange plate with pipe clamp
300
300 450 80
450 400
7
PFE2 Filter Unit
Ordering Information
Catalog No. 23028-0- 0 0 0 0 0 0
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Design
Without protection box 1
with protection box, IP54 (450 x 450 x 400 mm) 2
with protection box, IP54, anti-freeze heating 3
with protection box, IP54, ambient temp. max. 65 °C 4
Heating
Without 1
with heating sleeve and insulation (max. 200 °C) 2
with ring heater, unregulated 3
Control / Monitoring
Without 1
with 1x Pt 100 2
with 2x Pt 100 3
Power supply
230 V, 50...60 Hz 1
115 V, 50...60 Hz 2
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23028-0- 0 0 0 0 0 0
Cleaning Filter Surface and Probe Tube Cleaning Filter G 1/2" -> 12 mm Tube Fitting
Filter Unit
Diaphragm Valve
Puls-Instrumentenluft
-Y1.2
Solenoid Valve Aeration
-Y4
Heated
Check Valve
-Y5 Instrument Air Inlet
9
PFE2 Filter Unit
-X1
J
Power Supply Heating Sleave Filter Unit,
L
1
230 or 115 VAC, 50...60 Hz, 250 W,
N
required fusing 6 A,
2
controller required !
-E1
PE
Sample Line
3
3x1,5 mm²
Heated
Filter Unit
Sleeve
Heating
3x2,5 mm²
-X1
L
Power Supply Sample Line,
4
J
5
-E2
6
J
N
required fusing 16 A,
7
controller required !
PE
21.10.2005
3x1,5 mm²
L
9
-E3.1
10
230 VAC, 50...60 Hz, max. 800 W,
required fusing 6 A,
in case of 115 VAC power supply wire the
J
controller required !
N
11
PE
12
3x1,5 mm²
set @ 30°C
Probe Tube
Heated
3x1,5 mm²
-B01
-X1
Frost Protection
L
Power Supply Frost Protection,
13
230 or 115 VAC, 50...60 Hz, 300 W,
N
required fusing 6 A
14
PE
-E4
15
-X1 Power Supply PTC-Elements and
L
-E5.1
16
Powerpack -G02 / Heated Solenoid Valve
in Sample Outlet,
N
Heated Check
17
230 or 115 VAC, 50...60 Hz, 250 W,
Valve
PE
required fusing 6 A
18
-E5.2
19
20
4x1,5 mm²
-T01
21
115 V~
24 V=
1,3 A
3x1,5 mm²
-X2
required fusing 1 A
PE
...
Puls-Instrument Air,
N
required fusing 1 A
PE
...
Cleaning Filter,
N
required fusing 1 A
PE
9
...
10
Aeration,
N
required fusing 1 A
PE
12
-Y5
3
1
-X2
2
-K01
A2
13
required fusing 1 A
PE
15
x 0,14 mm²
2x0,14 mm²
-X3
3...18
2x0,14 mm²
3-Wire-Connection
3
Sample Line,
5
3-Wire-Connection
6
Probe Tube,
8
3-Wire-Connection
2x0,14 mm²
9
2x0,14 mm²
Wire-Connection
Design
The PFE3 filter unit consists of a modified FE2 filter unit with a
self-regulating heater. It is installed in a stainless steel protec-
tive case.
Filter Characteristics
Filter Element
Surface filter Installation
Filter Fineness Flange
0.3 µm DN 65, PN 6, Form B to DIN 2527;
Seal Integrity positioning of holes: 45°, 135°, 225°, 315°
10–4 hPa l/s Mounting
Dead Volume Flange connect or bolt to the probe tube
280 ml
Dimensions Connection Ports
50 / 20 x 135 mm
Sample Gas
Pipe fitting 6 mm
Heater Test Gas
Type Bulkhead fitting 6 mm (option)
Self-regulating heater (400 W) Purge Gas
Temperature Bulkhead fittings 18 mm (option)
180 °C (ambient temperature 20 °C) Probe Tube
Temperature Alarm G 3/4 internal thread to DIN ISO 228/1
approx. 150 °C
Power Supply Protective Case
115 VAC or 230 VAC, 50 to 60 Hz
Dimensions
Options 270 x 203 x 330 mm (H x W x D)
– Annular heating element (see page 4)
– Heating sleeve (see page 5) Material
Special steel Mat. No. 1.4301
Degree of Protection of Case
Materials of the Gas-carrying Parts IP54
Casing, Flange and Connections Weight
Special steel Mat. No. 1.4571 approx. 17 kg incl. FE2 filter unit
Seals Ambient Temperature
FPM –20 to +45 °C
Tube, SS 316Ti,
6x1 mm
Tube, CU
270
15x1 mm
266
Tube, CU
15x1 mm
Dimensional Drawing, Gas Connections (dimensions in mm)
M20x1,5
Cleaning Filter,
Cleaning
Filter Surface 18 mm
and Probe Tube, Calibration Gas
Junction Box -X1
80
Bulk Head Union Bulk Head Union
18 mm IP66
6 mm
3x 2x 58
M20x1,5 M12x1,5
124
-X1
L
1
Power Supply
J
Heated Aluminium Form Filter Unit,
230 or 115 VAC, 50...60 Hz, 250 W,
required fusing 6 A
N
2
-E1.4
-E1.3
-E1.2
-E1.1
PE
PE
3x1,5 mm²
set @ 140°C
Sample Line
Heated
3x2,5 mm²
-X1
L
230 or 115 VAC, 50...60 Hz,
3
-E2
J
N
4
required fusing 16 A / 25 A,
controller required !
PE
PE
2x0,14 mm²
-X1
3-Wire-Connection
7
Sample Line,
9
3-Wire-Connection
10
Temperature Alarm,
11
13
Ordering Information
Catalog No. 23046-0- 0 0 0 0 0
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Design
with protection box IP54 (330 x 203 x 270 mm) 1
Heating
with self regulated heating and temperature alarm (115 / 230 VAC) 1
with ring heater, unregulated 2
with heated sleeve (max. 190 °C, Pt 100 required) 3
Controlling / Monitoring
without 3
with 1 x Pt 100 (only for heated sleeve) 4
with 2 x Pt 100 (only for heated sleeve) 5
Power supply
230 VAC, 50…60 Hz 1
115 VAC, 50…60 Hz 2
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23046-0- 0 0 0 0 0
Probe system H
silicone carbide
water cooled 3)
water cooled 3)
water cooled 3)
water cooled 3)
partly heated
SS heated
SS 1.4571
SS 1.4571
SS 1.4893
Probe 2
Probe 2
heated
Temperature < 130°C
< 180°C 2)
< 300°C
180–450°C
180–900°C
< 1300°C
< 1400°C
Dust content < 50 mg/m3
< 1000 mg/m3
< 3 g/m3
> 3 g/m3 1) 1) 1) 1) 1)
40–100 g/m3 1) 1) 1) 1) 1) 1)
< 2000 g/m3 1) 1) 1) 1)
Water vapor < 65 °C
dew point > 65 °C
Acidic aerosols (acid dew pt. 130 °C)
Pressure < 2 bar
< 5 bar
Technical data see … p. 16 p. 16 p. 16 p. 16 p. 16 p. 17 p. 18 p. 18 p. 18 p. 18 p. 19 p. 19
1 2 5 3 1 2 5 3 1 4 2 5 3
Flow direction of 1 Lead-through tube Minimum distance xmin of the mounting flange
process gas 2 Mounting flange DN 65, PN 6, from the wall depending on mounting angle α:
Form B to DIN 2573
3 Gasket α 10° 15° 20° 25° 30° 35°
4 Welded-on rectangular block xmin / mm 133 138 143 147 151 153
5 Sampling probe flange
Dimensions in mm
12,6 ø
17,2 ø
G3/4
Pressure pabs = 50 to 500 kPa (0.5 to 5 bar)
Temperature L1
SS Mat. No. 1.4571 max. 450 °C 16
SS Mat. No. 1.4571 (coated) max. 180 °C
SS Mat. No. 1.4893 max. 900 °C
Special Steel Material No. 1.4893 (max. 900 °C)
SS Mat. No. 1.4571 (Type 40 W) max. 300 °C
Silicone carbide (max. length 1.000 mm) max. 1.300 °C
21,4 ø
16,8 ø
Flow Rate max. 300 l/h G3/4
Installation
23 ø
15 ø
34 ø G3/4
G 3/4 external thread DIN ISO 228/1
Mounting Angle 105 L1
any; 8° to 90° downwards from the horizontal is recom- 16
Ordering Information
Catalog No. 23036-0- 0 0 0 0 0 0 0 0 0 0
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Version
Stainless steel (1.4571), up to 500 °C 1
Stainless steel (1.4571), coated, up to 180 °C 2
Silicon carbide up to 1300 °C, max. 1000 mm 4
Stainless steel (1.4893), up to 900°C 5
Stainless steel (1.4571), up to 300 °C, heated (Type 40 W for cement) 6
Nominal length
500 mm 1
1000 mm 2
1500 mm 3
2000 mm 4
2500 mm 5
3500 mm (only for heated version) 6
4000 mm (only for heated version) 7
4500 mm (only for heated version) 8
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23046-0- 0 0 0 0 0 0 0 0 0 0
40
160 17 1000/1500/2000 mm 21
0
13
25 x 1,5
40 x 2
Pt100
Heater
120
123456 789
Bridge:
Line 230 V Terminals 3+4
Line 115 V Terminals 2+3/4+5 Length (mm) 1000 1500 2000
M20 Wattage (W) 400 600 800
Weight (kg) 8 10 12 90
Line Sensor
Temperature Controller
Connection
thickened shank:
800 mm for
L1 ≤ 2000 mm
35
90 1300 mm for
L1 ≥ 2500 mm
L3 L2 3 L3 200 mm with
reinforced
L1 protection tube
Z
Material of the Gas-carrying Parts
Support Tube, Heating Rod, Connecting Pipe
Special steel, Material No. 1.4571
650
Inlet Filter 1
Aluminum oxide
α 1 80
Seals
Ceramic fiber
Z
Flange
Special steel, Material No. 1.4581 1=N
2 = L1 4
260
Mounting
Flange 1 Filter mounting 4 Harting connector
DN 65, PN 6, Form B, to DIN 2573 2 Gas sampling probe tube
position of holes: 45°, 135°, 225°, 315° 3 Sample gas outlet or test gas / purge air inlet
Mounting Angle
any; 10° to 35° downwards from the horizontal recommend- Ordering Information
ed (take direction of flow into account), see dim. drawing
Catalog No.
Sample Gas Outlet
1200 mm, 230 V, 50/60 Hz, 650 W 23022-4-0768301
G 1/4 internal thread DIN ISO 228/1
1200 mm, 230 V, 50/60 Hz, 255 W, part. heat.23022-4-0768305
Test Gas / Purge Air Inlet Prot. box, insulat., IP54, (330x203x270 mm) 23022-4-0730719
G 3/8 internal thread DIN ISO 228/1
Technical Data
Sampling Pressure
Atmospheric pressure
Gas Temperature
–25 to +50 °C, dry
Sample Gas Flow Rate
max. 70 l/h
Dust Content
max. 1 mg/m3
Pressure Drop with New Filter
≤ 10 hPa (mbar) at 70 l/h
Service Life of Filter (estimated)
filter constant (2500)
d (days) =
flow rate (l/h) x dust concentration (mg/m3) Ordering Information
Sample Gas Connection Port
Catalog No.
Bolted connection for flexible pipe with internal diameter
4 mm and external diameter 7 mm Room air probe 11 23021-4-0859585
Materials
Polycarbonate (probe head), aluminum oxide (filter)
Transport and Storage Temperature
–25 to +70 °C
Mounting
On wall or suspended from ceiling
Use
For filtering dust from gases in the gas conditioning systems of
analyzer systems when the gas sampling probe does not have
a filter. The ceramic filter is positioned upstream the other
devices in the sample gas conditioning chain. The ceramic filter
can be heated with the annular heating element if condensation
from the sample gas is likely.
Technical Data
Flow Rate (Air)
max. 300 l/h
Gas Pressure
pabs = 20 kPa to 1.7 MPa (0.2 to 17 bar) at 20 °C filter
temperature, max. 200 kPa (2 bar) at 180 °C
Pressure Drop
0.1 to 0.85 hPa at 30 to 250 l/h (air)
Dynamic Characteristics at
Flow rate (air) 30 l/h 60 l/h 90 l/h
Dead time Tt 3.7 s 2.3 s 1.7 s
90% time T90 11.5 s 5.5 s 3.9 s
Filter Fineness
0.3 µm
Dimensional Drawing (dimensions in mm)
Construction, Mounting
Materials of the Gas-carrying Parts
Housing: steel, St 37 K or special steel, Material No. 1.4571;
Filter: aluminum oxide; Seals: FPM, Cu, Al2O3 fiber
Gas Connection Ports
G 1/4 internal thread DIN ISO 228/1 or
pipe connectors for pipe of internal diameter 4 mm
Ordering Information
Catalog No.
Ceramic filter, steel version 23031-5-8018177
Ceramic filter, stainless steel version 23031-5-8018178
Use
If the pressure of the process gas is too high for the analyzer, a
pressure reducer is required to lower the pressure. The type of
pressure reducer used is determined by the properties of the
process gas and the required sample gas pressure.
Technical Data
Casing: Brass Casing: Brass Casing: Special steel Casing: Special steel
Construction Single-stage with one Single-stage with two Single-stage without Single-stage with two
manometer (outlet) manometers (inlet / outlet) manometer manometers (inlet / outlet)
Inlet Pressure max. 1.5 MPa (15 bar) max. 3.5 MPa (35 bar) max. 3.5 MPa (35 bar) max. 3.5 MPa (35 bar)
Outlet Pressure 0.8 to 10 kPa 10 to 40 kPa 10 to 70 kPa 10 to 70 kPa
(adjustable) (8 to 100 mbar) (0.1 to 0.4 bar) (0.1 to 0.7 bar) (0.1 to 0.7 bar)
Safety Valve Set to 50 kPa 80 kPa 80 kPa
– (0.5 bar) (0.8 bar) (0.8 bar)
Manometer Display 0 to 4 MPa 0 to 4 MPa
Inlet Pressure – (0 to 40 bar) – (0 to 40 bar)
Manometer Display 0 to 250 hPa 0 to 60 kPa 0 to 150 kPa
Outlet Pressure (0 to 250 mbar) (0 to 0.6 bar) – (0 to 1.5 bar)
Flow Rate 25 to 250 l/h 25 to 250 l/h 25 to 250 l/h 25 to 250 l/h
Material Diaphragm Buna rubber Buna rubber Special steel Special steel
Material Gasket Buna rubber Buna rubber EPDM EPDM
Material Housing Brass Special steel; gas- Special steel Special steel
carrying parts: brass
Weight approx. 1.25 kg approx. 2.65 kg approx. 2.35 kg approx. 2.35 kg
Gas Temperature and max. 50 °C max. 50 °C max. 50 °C max. 50 °C
Ambient Temperature
Catalog No. 23035-5-5864958 23035-5-5864951 23422-4-0730699 23422-4-0730700
79
appro
x. 30
Ø4
° (vari
able)
GasTech
105
approx. 94
(depending on handwheel position)
Pressure reducer single-stage with two manometers Pressure reducer single-stage with two manometers
(inlet / outlet) (Catalog No. 23035-5-5864951) (inlet / outlet) (Catalog No. 23422-4-0730700)
118
appro
x. 30
Ø4
°
(varia
148
GasTech
ble)
approx. 94
(depending on handwheel position)
Hard cap
Technical Data
Frost protected line Self-regulating strip heater Regulated parallel strip heater
Outer Covering PVC PVC PVC
Medium-carrying Pipe PTFE PTFE PTFE
Integral Pipe-end Connectors Various options (see ordering Various options (see ordering Various options (see ordering
information, page 28) information, page 28) information, page 28)
Heater self-regulating strip heater self-regulating strip heater CPD heating element, regulated
Power Supply 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz
Heat Output 16 W/m 30 W/m 60 W/m
Current Consumption Making / holding current 0.15 A/m 0.3 A/m
= 65 / 25 mA
Holding Temperature see diagram see diagram 200 °C
Ambient Temperature –30 to +60 °C –30 to +60 °C –30 to +60 °C
Outer Diameter 30 mm 43 mm 40 mm
Flexible Sample Gas Pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm
Maximum Line Length 130 m 90 m 60 m
Line Duct M42 M42 M42
Minimum Bending Radius 300 mm 300 mm 300 mm
Weight approx. 1 kg/m approx. 1 kg/m approx. 1 kg/m
Connect. Lead for Heater (fixed)3 x 2.5 mm2; length 5 m 3 x 2.5 mm2; length 5 m 3 x 2.5 mm2; length 5 m
Connect. Lead for Pt 100 (fixed) – – LIYCI 2 x 0.14 mm2; length 5 m
50
40
30
20
- 30 - 20 - 10 0 + 10 + 20 + 30 + 40
Ambient temperature °C
100
90
80
70
- 30 - 20 - 10 0 + 10 + 20 + 30 + 40
Ambient temperature °C
46
46
42
300 300
Length by customer’s request
max. length depending on MCB
Dimensional Drawings (dimensions in mm)
Power Supply
Cable 3 x 1.5...2.5 mm²
CGWB 13 Heated Sample Gas Line
25
42
49
63
Shrinking hose
POM cap
Pt100
Cable 2 x 0.14 mm²
27
CGWB 13 Heated Sample Gas Line
Ordering Information
Catalog No. 23281-0- 0 0 0 0 0 0 0
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Version
Parallel strip heater 60 W/m, line can be shortened, regulated (Pt 100), 200 °C 2
Self-regulating 16 W/m, 30 °C (antifreeze) 3
Self-regulating 30 W/m, 100 °C 7
Power supply
230 V, 50 to 60 Hz 1
115 V, 50 to 60 Hz 2
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23281-0- 0 0 0 0 0 0 0
Technical Data
TBL 01-S Self-regulating strip heater TBL 01-S Fixed resistor heating conductor
Outer Covering Polyamide corrugated tubing Polyamide corrugated tubing
Medium-carrying Pipe PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm
Integral Pipe End Connectors Probe: metal cap, analyzer: flange or hard caps on both sides (pipe projecting 300 mm) or silicon
caps on both sides (probe: 6/4 stainless steel connections, analyzer: pipe projecting 300 mm)
Power Supply 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz
Heat Output 30 W/m 90 W/m
Holding Temperature see diagram on page 26 200 °C
Ambient Temperature –30 to +60 °C –30 to +60 °C
Outer Diameter 43 mm 43 mm
Maximum Line Length 100 m 40 m
Line Duct M48 M48
Minimum Bending Radius 200 mm 200 mm
Weight approx. 1 kg/m approx. 1 kg/m
Electrical Connection for fixed, 3 x 2.5 mm2, length 5 m fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m
Heater and Pt 100
TBL 01-C Self-regulating strip heater TBL 01-E Fixed resistor heating conductor
Outer Covering Metal corrugated tubing (for ACK) or Polyamide corrugated tubing
PVC outer covering (for ACMP)
Medium-carrying Pipe PTFE pipe 6 x 4 x 1 mm PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm
Integral Pipe End Connectors Probe: metal cap – analyzer: silicon cap Polyamide hard caps
(mandatory for ACK) or POM plastic caps on
both sides (mandatory for ACMP) or probe:
PTFE cap – analyzer: POM plastic cap
(mandatory for ACMP 120°C)
Power Supply 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz
Heat Output 30 W/m or 47 W/m 90 W/m
Holding Temperature 100 °C or 120 °C 200 °C
Ambient Temperature –30 to +60 °C –30 to +60 °C
Outer Diameter 43 mm 43 mm
Maximum Line Length 70 m 40 m
Line Duct M48 M48
Minimum Bending Radius 200 mm 200 mm
Weight approx. 1 kg/m approx. 1 kg/m
Electrical Connection for fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m
Heater and Pt 100
PA corrugated hose
350 Process tube overhang
42
50
Hard cap Hard cap
Power supply
Dimensional Drawings (dimensions in mm)
TBL 01-S, Pipe end connectors with hard caps
60
50
49
63
54
49
TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines
42 24
6 81 81
56
42
25
49
63
54
Power supply
Cable 3 x 1.5...2.5 mm²
Length by customer’s request
max. length depending on MCB or
switching load of the controller respectively
TBL 01-S, Pipe end connectors with metal caps for ACF-NT
Power cupply
2 x M16 Cable 3 x 1.5...2.5 mm²
TBL 01-E
Ø 6 mm
50 42
6
56
Hard cap Hard cap
31
32
Sample take-off side Analyzer side
Metall
corrugated hose
Process tube overhang
63
45
25
54
49
51
81
Silicone cap
Length by customer’s request
max. length depending on MCB or
switching load of the controller respectively 300
Replacable tube nipple,
TBL 01-C, Pipe end connectors for ACK
316ss, 6x1 mm
Dimensional Drawings (dimensions in mm)
Power supply
Cable 3 x 1,5...2,5 mm²
Pt100
Cable 2 x 0,14 mm²
42
TBL 01-S, TBL 01-C, TBL 01-E Heated Sample Gas Lines
Shrinking hose
POM cap
300 300
Length by customer’s request
TBL 01-C, Pipe end connectors for ACMP
Power Supply
Cable 3 x 1.5...2.5 mm²
Pt100
Cable 2 x 0.14 mm²
Material of sheathing
with plastic hose of PA (outside and indoors installation) 1
Connections
Metal caps w. 6/4 SS connectors on both sides (mandatory for ACF-NT) A
Probe: Metal cap / Analyzer: Flange, 6/4 SS connectors on both sides B
Plastic hard caps on both sides (tube 300 mm surplus, self-regulated only) C
Silicone caps both sides / probe 6/4 SS conn. / Analyzer: tube 300 mm surplus D
Control / Monitoring
without Pt 100 (self-regulated version only) 0
with 1 x Pt 100 1
Power supply
230 V, 50 to 60 Hz, for self-regulating strip heater (not for ACF-NT) 1
115 V, 50 to 60 Hz, for self-regulating strip heater (not for ACF-NT) 2
230 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m 3
115 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m 4
230 V, 50 to 60 Hz, for fixed resistor heat. cond. 90 W/m (for ACF-NT, length ≤ 35 m) 5
115 V, 50 to 60 Hz, for fixed resistor heat. cond. 90 W/m (for ACF-NT, length ≤ 15 m) 6
400 V / 230 V, 50 to 60 Hz, for fixed res. heat. cond. 90 W/m (ACF-NT, length 35-60 m) 7
200 V / 115 V, 50 to 60 Hz, for fixed res. heat. cond. 90 W/m (ACF-NT, length 15-40 m) 8
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23283-0- 0 0 0 0 0 0
Material of sheathing
with metal corrugated tubing for ACK (outdoor and indoor installation) 1
with PVC sheathing for ACMP (outdoor and indoor installation) 2
Connections
Probe: Metal cap, Analyzer: Silicone cap (mandatory for ACK) + shortening kit 1
POM plastic caps on both sides (mandatory for ACMP ) + shortening kit 2
Probe: PTFE caps, Analyzer: POM caps (mandatory for ACMP 120°C ) + shorten. kit 3
Control / Monitoring
with 1 x Pt 100 1
Power supply
230 V, 50 to 60 Hz 1
115 V, 50 to 60 Hz 2
↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23285-0- 0 0 0 0 0 0
110...240 VAC
Measurement Input Pt 100
-15/+10%
48...63 Hz
Outputs 2 x relay 3 A, 250 VAC
Output 1
not used
Power Supply 110 to 240 VAC +10 / –15 %, 48 to 63 Hz
Binary
Dimensions 48 x 48 mm, C-rail housing
Weight 140 g
L1 N 342 343 341 142 143
Catalog No. 23345-5-8328907 parameters not set
Alarm 1
23345-5-8328907P parameters set for Temperature
max.
filter unit
230V 0/12V max.
23345-5-8328907L parameters set for
GND
/3A 230V
sample gas line
+
/3A
23345-5-8328907C parameters set for Alarm 2
self-regulating heater Limiter
for ACMP 242 243 81 83 80 112 111 113
Output 2
Outputs
Binary
Logic
Switching Relay
Load Voltage Urms = 24 to 265 V
4+ 3-
Max. Load Current Irms = 30 A (ambient temperature = 25 °C)
Weight 200 g M2A ϑ
Catalog No. 23345-5-8328731
1 2
L Last