500 ppm 500 503 20 0.60%

500 ppm 500 502 22 0.40%

500 ppm 500 500 25 0.00%

1000 ppm 1000 1005 20 0.50%

1000 ppm 1000 1000 22 0.00%

1000 ppm 1000 995 25 0.50%

5000 ppm 5000 4980 20 0.40%

5000 ppm 5000 4990 22 0.20%

5000 ppm 5000 5003 25 0.06%

The temperature compensation of the HM TDS-3 meter was also verified by


comparing the TDS readings obtained at different temperatures. The results
showed that the TDS readings were consistent within the specified range of 1-
50°C, indicating that the built-in sensor for automatic temperature
compensation is functioning properly.

Conclusion:
Based on the results obtained, the HM TDS-3 meter is accurate and precise
within the specified range of 0-9990 ppm for TDS and 0-80°C for
temperature, with a resolution of 1 ppm for readings between 0-999 ppm and
a resolution of 10 ppm for readings between 1000-9990 ppm. The meter also
features automatic temperature compensation, which was verified to be
functioning properly