Professional Documents
Culture Documents
Instrumentation Spectrophotometry
Chemical analysis of blood samples The quantitative measurement of the reflection or
Differ in their principle and application transmittance properties of a solution as a function of
The most common used in clinical chemistry involves the wavelength
principle of light absorption Allows measurement of light intensity in a much narrow
wavelength range thus it is suitable for both colored and
Colorimetry
colorless solutions
Measured is colored, it absorbs light within the visible Both visible and UV (<400 nm) spectrophotometer
spectrum (400-700 nm Light) depending on the absorbance peak of the analyte
Relies on the human eye for interpretation Two types
Beer’s law states that the concentration of an analyte in a Single Beam (light source, diffraction grating
solution is directly proportional to the amount of light it apparatus, sample detector)
absorbs Double Beam (light source, monochromator,
The more light absorbed the more analyte present standard + sample, sample detector
and standard detector)
Two types Parts of the spectrophotometer
Light source
provide radiant energy
Visible
o Tungsten bulb (400-700 nm)
Beer’s Law (Beer’s-lambert’s Law) The light intensity of the characteristic wavelength produced
by the atom is directly proportional to the number of atoms
States that the concentration of the solution is directly
emitting the energy, which in turn directly proportional to
proportional to the amount of light absorbed and inversely
the concentration of the substance of interest in the sample.
proportional to the light transmitted. Excitation of the analyte
Beer’s Law can be expressed in equation form as shown Visible spectrum
below:
Qualitative (color)
Quantitative (photons of light)
100
𝐴 = 𝑎𝑏𝑐 = log = − log %𝑇 Different electropositive elements when excited by heat
%𝑇 given off by a hot flame emit in the form of a colored light
A=Absorbance
characteristic of that element, when the atoms return to the
a=Is the absorptivity value
ground state
b= Is the light path of the solution in cm
c= Concentration of the substances of interest
Analyte Wavelength (nm) Color
%T= Percent transmittance
Lithium 670 nm Red
Blank Correction Sodium 589 nm Yellow
Magnesium Blue
ensures the accuracy of the results by eliminating interfering Potassium 466 nm Violet
substances Calcium 682 nm Red orange
Rubidium Red
types of blanks Barium 559 nm Lime green
1. Reagent Blank
- Contains only the reagent (no sample)
- Used to correct for the absorbance reading
2. Sample Blank
- Contains only the sample (no reagent)
- Used to correct non-specific absorbance of
other substances in the sample (hemoglobin,
bilirubin, lipids)
Flame Emission Spectroscopy / FES
Quantitative analysis of serum and plasma electrolytes
Main components
1. Burner assemble
2. Optical system
3. Photodetector
Two methods of doing flame spectroscopy
Potential interference
Direct method 1. Chemical interference
- Standards of sodium or potassium is - Fails to dissociate the sample into free atoms
aspirated directly into the flame to provide a e.g. Phosphate interference determination of
series of meter readings against which is as calcium caused by the formation of calcium
unknown solution is compared phosphate.
Indirect or internal - Lanthanum and strontium is added to
- Lithium is used as the internal standard displace the Calcium phosphate or a higher
- It acts as a radiation buffer to prevent mutual temperature may be used instead
co-excitation 2. Ionization interference
- Atoms are in the flame becomes excited
instead dissociated only and the atoms emit
Types of Refractometer
Traditional handheld refractometer
Digital handheld refractometer (LED light)
Inline process refractometer (LED light, monitor
concentration more accurate)
Abbé refractometer (highly precise)
Refractometry
Osmometry
Determination of the concentration of the solution by
measuring its refractive index Use to measure the concentration of solutes particles
(molecular weight) in a solution.
Interferences:
Types
Wavelength of the incident light
Nature of the liquid medium Freezing point osmometer
Concentration Vapor Pressure osmometer
Temperature Membrane osmometer
Colligative properties
Freezing point
- Depressed by 1.86C
Boiling point
- Elevated by .52C
Osmotic pressure
- Increased .3 mm/Hg at 25C
more analytes the lower the freezing point Vapor pressure
- Decreased
Vapor of osmometer
Interferences:
Lipemic samples-freezing point
Volatile solutions- vapor pressure
Sample
the higher the charge, greater is the electrophoretic mobility
lenses on the laminar flow chamber (C), after passing The data gathered can be analyzed statistically by flow
through the camera, the direct light beam is arrested by a cytometry software to report cellular characteristics such as
screen (s), while the scattered light is focused by another lens size, complexity, phenotype, and health.
(L3) on a photodiode (D1), this constitutes the Forward
Scatter detector (FSC).
The side scattered light and the fluorescence are focused by a
lens onto a dichroic mirror (that reflects most of the light of
wavelength equal to that produced by the Source (A), goes
through a filter (Fl) and impinges on a photodiode
detector(D2). This constitutes the Side Scatter detector
(SSC).
Instrument Overview
Flow Cytometers
The fluidic system - which presents samples to the
-are able to analyze several thousand particles every second, in "real interrogation point and takes away the waste.
time, and can actively separate and isolate particles having specified The lasers - which are the light source for scatter and
properties. fluorescence.
A flow cytometer is similar to a microscope, except that, The optics - which gather and direct the light.
instead of producing an image of the cell, flow cytometry The detectors - which receive the light the electronics
offers "high throughput for a large number of cells) The peripheral computer system – which convert the
automated quantification of set parameters. signals from the detectors into digital data and perform the
The flow cytometer performs this analysis through a laser necessary analyses
beam and capturing the light that emerges from each cell as Factors Affecting instrumentation:
it passes through.
Sample pressure
Proper operation of fluidic components
Alignment of the laser beam with the stream
The wavelength of excitation
Sample preparation
Electrical Impedance
Electrical Impedance - is the measure of the opposition that a
circuit presents to the passage of a current when a voltage is
applied.
Principles of Electrical Impedance Internal electrode - an electrode inside the aperture tube.
Histogram – graphical presentation of blood cell
The impedance principle of cell counting is based on distribution; provide information about erythrocyte,
the detection and measurement of changes in electrical leukocyte and platelet frequency and distribution as well as
impedance (resistance) produced by a particle in depict the presence of subpopulations,
conductive fluid as it passes through a small aperture
Counting chambers
Coulter Principle or Impedance method by Wallace
Coultier. Common chambers that uses impedance
RBC/ Platelet chamber
WBC chamber
RBC chamber
Aspirated blood is divided into two separate volumes for
measurements.
One volume is mixed with diluents and delivered to the cell
bath, where erythrocyte and platelet counts are performed.
As cell passes through the aperture, partially occluding it, the
electrical impedance increases, producing a voltage pulse,
the size of which is proportional to the cell size.
The number of pulses is the number of cell counted.
WBC chamber
Number of pulses = blood cell count The other blood volume is mixed with diluent and a
Amplitude (height) of the pulse= volume of the cell cytochemicalytic reagent that lyses only the red blood cells.
A Backwash or sweep flow mechanism has been use A leukocyte count is performed as the remaining cells pass
to prevent recirculation of cell in the aperture through an aperture.
Components:
Factors Affecting Instrumentations
Aperture - the orifice where the particles or blood cells will
pass through Size of platelets
Cells into 2-20 femtoliter range are counted as platelets;
Aperture diluent - a low concentration electrolyte solution
where the blood is diluted. Those counter above 35 femtoliter as counted as red blood
Aperture tube - the glass tube that contains the aperture. cells. This presents a problem when attempting to obtain
External electrode – the electrode immersed in a low platelet counts in individuals that possess small red blood
concentration electrolyte solution outside the aperture tube.
cells, as counting area between platelets and cells may Types of Chromatography
overlap.
INTERFERENCES Paper chromatography
- Based on the nature of solvent, solubility of
Aperture diameter - is crucial, with the red blood cell solute and rate of diffusion.
(RBC)/platelet aperture smaller that the WBC aperture to Gel permeable (size-exclusion) chromatography
increase platelet counting sensitivity - Size or molecular weight of the molecules
Protein buildup - It decreases the diameter of the orifice, Ion-exchange chromatography
slowing the flow of cells and increasing their relative - Net charge of the molecule
electrical resistance. It results to lower cell counts, which
Liquid-liquid chromatography
result to falsely elevated cell volumes,
- Solubility of the solutes (partition-
Count reduction or coincident passage loss - Coincident
passage of more than once cell at a time through the orifice coefficient)
Causes artificially large pulses, resulting in falsely increased Gas-liquid chromatography
cell volumes and falsely decreased cell counts. - Basis of sample volatility
Recirculation of cells back into the sensing zones - It Thin-layer chromatography (TLC)
creates erroneous pulses and falsely elevates cell counts. - Similar to paper chromatography
When the WBCs are present in extremely high numbers, - Uses glass or paper plates with thins silica,
they become a source of interference. polyacrylamide, or starch gel as matrix
High performance liquid chromatography (HPLC)
Chromatography
- Uses pressure (500-5000 lbs psi) for
pumping of aqueous or organic solution
Method of separation based on the specific difference in the
through a column
physico-chemical chemicals e.g. column and the mobile
Gas chromatography
phase e.g. eluent
- Most powerful
Introduced by Russian botanist Mikhail Tswett
- Can separate nanograms or picograms of
Compounds interacting more strongly with the stationary
volatile substances
phase retained longer in the medium than those that favor
the mobile phase.
The smaller the affinity a molecule have for the stationary
phase the shorter the time spent during reaction, and the
longer it travels.