CALIESYS
(Garang TESTING Seton Eee en ene
Batic label iene lelela)
LCE are lee
Halaman (page) 1 deri(out op 2
Nomor(Number):___Losz00s02
Identitas Alat/lastrumont’s (dentiv
Nomor Identifikasi
PTALEPDLITNG ot
‘dontication Nunbor aes
ans, Temperature Gt
vam perature Gun
Merek Pabrik — 7 EE ao o
iRTEK
Manufacturer : a
Tipe Nomor Ser
1
ena 1 60/ e0o0t01547
Rentang Ukur Resotu aaa
Range /Resokiion z
Adenitas Pemilk (Owns deni)
Nena Poesauan PT. PUTRA TUNGGAL MANDIRITAMA
Customer's Name
Alamat oo
nn * Kamplbk Ruko Bebipapan Baru Blok AS2 No. 28 Balkpapan - KALTIMA
‘Tanggal Diterima a
i
Dato of Acceptance ee iH
Tanggat Kaliorasi
28 n'2020
Dato of Cafbration ss
‘Tempat Kalibrasi TESYS INDONESU -
Place of Calibration eo By on
Kondisi Lingkungan suns Ketrbaban|
243 208°C 409335 % RH
Environmental Condition (Temperature) ‘ (Humidity) i )
Ketolusuran ‘Hasil kalbrasi dllaporkan tertelusur ke satuan pengukuran Si melalui (Celbration
Trovoabiy resuts traceabe w the tenatona Syetom of Unts (1) ¥oogh) Lk 280 oN
Diterbitkan Tangget ue a
tasue Date — .
Disetujuioteh
Aoproves By
ree a
Alvan Mansy
Mansemen Tekris
Tochnicallanagement
suns unckuPKALBRAS! coe OF CALERA
KAN ences neopets anarchists atti diene
[eb probed ony rereir andr bl he cret of ate ioe nr po
PT CALTEsys INDONESIAALIESYS
CcaletaTion® FISTING STE Demeester)
etic keli eli)
PCD re adie
Haleman (paps) 2 dari(out of 2
Nomor (Number) :__ 0420 0802
Temporature Gun
HASIL KALIBRASI (Calbration Reports)
Koroksi penunjukan alat (Corrections to instrument's reading)
Penunjukan Alat_ | Penunjukan Standar] Koreksi Ketidakpastian
(instrument's Reading)| (Standard Reading) | (Correction) (Uncertainty)
a co cy £0.
‘50,0 92 “08 35
100.0 987 43 37
1500 148.5 48 38
200,0 1985 “18 40
300.0 239.3 “07 48
400.0 4008 os 85
5000 5026 26 54
Keterangan (Note) :
‘lat ini dikalibrasi menggunskan metode perbandingan langsung terhadap ;
‘Sumber benda hitam standar SMS0Stemf1
berdasarkan prosedur kalibrasi tomperature radiasi : MKSP IRT-O1-L
(This instramentwras calibrated using direct comparison method fo;
Back body source standard SMSOStemtt
‘based on tho tompereture radiation calibration procedure : MKSP IRT-01-L)
2, Emisivitas dari termometer radia! di sat pads 4
(The emissivty ofa radiation thermometer is set at 1)
3. Jatak kalibrasi 140 mm
(The dlstonce of calibration is 140 mm)
4 Nila koveksi bertaku jka nilal omisivitas ala diset pada nilal yang sama dengan emisivit
yang ukur
(The correction value is valid the omissivly value of the instrumentis sett the same
value ofthe abject to ba measured)
Ketidakpastian pengukuran yang dilaporkan adalah ketidakpastian bentangan
ppada tingkat kepercayaan sekitar 95% dengan factoreakupan k=2.
(The report of measurements uncortintyis expanded uncertainty with confidence level
2pproximately 95% with coverage factor k = 2)
benda
Sulu) Tempcratre » Gays} Force » Dimensi/OMmersion » Vokme/ Volume * Tekaran/Prezire + \vaktu& Frekuens/ Time Frecuency
ima Torque = Kelntrhan| letra! » Denstas/Devsiy « Keketatan Hardness + Instumen Antics | Anyi! wstroment