Professional Documents
Culture Documents
NO : NDE/RT/001
NDE PROCEDURE Page : 1 of 15
Revision : 00
SUNGOV
RADIOGRAPHIC EXAMINATION Date : 09.11.11
PROCEDURE FOR
RADIOGRAPHIC EXAMINATION
(Revision based on ASME Section V – 2011a)
1.0 SCOPE:
DOC.NO : NDE/RT/001
NDE PROCEDURE Page : 2 of 15
Revision : 00
SUNGOV
RADIOGRAPHIC EXAMINATION Date : 09.11.11
This procedure covers the radiographic examination of welds, in steel / stainless steel in the
thickness range of 3 to 65 mm with the requirements and acceptance criteria in accordance
with ASME Sections VIII Div. 1, and ASME Sec V .
ASME SEC V
Weld to be radiographed shall not have the weld ripples or weld surface irregularities of
which radiographic images will mask or be confused with the image of any unacceptable
discontinuity. The finished weld surface may have reasonably uniform crowns with reinforcements
not exceeding that specific in reference Code section.
The personnel carrying out radiography shall be NDE Level I or above and the personnel
interpreting the radiograph shall be NDE Level II or Level III qualified in RT ( Radiography
test) as per SEPL written practice SEPL / NDE / WP / 001 Rev 01
The radiation source shall be Ir 192 for Gama ray radiography. X rays shall be used
whenever the thickness is less than 6mm.
5.2 FILM
DOC.NO : NDE/RT/001
NDE PROCEDURE Page : 3 of 15
Revision : 00
SUNGOV
RADIOGRAPHIC EXAMINATION Date : 09.11.11
TABLE – 1
Agfa Kodak
Structurix Industrex
Low Very High Very Low D4 M
5.4 IQI
Wire type IQI as per ASTM E 747 shall be used for ensuring the sensitivity of
Radiography
Lead letter “B” 13mm (½ in.) size and 1.5mm ( 1/16 in.) thick shall always be placed on the
backside of the cassette during each exposure to check for the back scattered radiation.
Placement of “B” letter shall be recorded in RT report.
5.8 Densitometers
Densitometers shall be of digital type in which a transmitted density of 1- 4 can be
measured and able to locate the spot to be inspected.
6.0 CALIBRATION
6.1.1 Manufacturer’s certificate or Data sheet indicating the dimension of source can
be taken as calibration certificate for source.
6.2 IQI
6.2.1 Each set of IQI shall have IQI manufacturer’s calibration certificate.
6.2.2 In the absence of IQI manufacturer’s certificate , the IQI shall be calibrated by a
calibration lab accredited to NABL.
6.2.3 The frequency of calibration shall be prior to first use and whenever any damage is
noticed for the wires.
DOC.NO : NDE/RT/001
NDE PROCEDURE Page : 5 of 15
Revision : 00
SUNGOV
RADIOGRAPHIC EXAMINATION Date : 09.11.11
6.2.4 The condition of IQI for any damages shall be verified once in three months and
recorded in a report form.
6.3 National Standard Step tablet ( Master Strip)
6.3.1 Standard step tablet shall have a valid calibration certificate issued by the film
manufacturer traceable to National standards.
6.3.2 The calibration shall be valid for a period of one year from the first date of
opening the factory packing or the date of film expiry whichever is earlier.
6.4 Densitometer
6.4.1 Film densitometer shall be calibrated once in 90 days or whenever any major
repair is done whichever is earlier.
6.4.2 Calibration of densitometers shall be done with a National standard step tablet
having a valid calibration.
6.4.3 At least 5 steps in the range of 1 to 4.0 shall be selected for calibration
6.4.4 The measured density shall not vary more than ± 0.05
7.0 EXAMINATION AREA
TABLE – 2
In single wall technique, the weld under examination will be exposed to direct
radiation it can be either Panoramic, multiple segment or single segment shot.
The radiation source shall be perpendicular to the area of interest . When multiple
film cassettes are used each film shall overlap at least 50mm.
8.3.1.1 Panoramic Exposure : Panoramic exposure may be used for
exposing complete circumference of a circular weld joint using
multiple films in a single exposure with the radiation at the inside
center.
8.3.1.2 Multiple segment shot : Technique in which multiple segments are
covered in single exposure.
8.3.1.3 Single segment shot : One film segment is radiographed with single
exposure
8.3.2.2.2 Super Imposed shot in which both the welds are superimposed
over each other. Minimum three exposure with either 60 Deg or
120 to each other shall be taken
8.4 LOCATION MARKERS
Location markers shall be placed as detailed below in order to ensure that the required
coverage is obtained
TABLE 3
8.5.2 The selection of IQI shall be based on the weld material thickness as per the
table 4 given below
8.5.3 Weld thickness given in the table shall be based on nominal single-wall thickness
plus the estimated weld reinforcement not to exceed the maximum permitted by
the referencing Code Section
TABLE 4
8.6.1 IQI shall be placed on the source side except double wall single image where
inside surface cannot be accessible.
8.6.2 When IQI is placed in the film side due to inaccessibility then Lead letter “F” shall
be placed as an identification image in the exposed film.
8.6.3 For film side IQI the essential wire for the film side shall be visible on the film
8.6.4 For double wall double imaging only source side IQI shall be used.
8.6.5 IQI shall be placed across the weld.
10.0VIEWING FACILITY
The film shall be viewed in a darkened area with subdued background lighting of an
intensity that will not cause reflections, shadows, or glare on the radiograph that interferes
with the interpretation process. The viewing conditions shall be such that light from the
outer edges or through the low density portions of radiographs does not interfere with
interpretation.
11.0EVALUATION OF FILM
11.1.1 Processing defects : All processed film shall be free from fog, scratches,
watermark, nail mark, kink mark or any mechanical damages and chemical
processing marks. Any process mark at the area of interest shall be re shoot.
11.1.2 Density : The film density through the area of interest of radiographic image shall
be 2.0 minimum. The maximum density shall be 4.0. The density variation
anywhere in the area of interest shall not be more than minus 15% or plus 30%
from the density through the body of the hole IQI.
11.1.3 Sensitivity : The required sensitivity shall be ensured by the minimum size IQI
wire visible on the film. If the required sensitivity is not achieved then the segment
shall be reshoot.
12.1.3 Any group of indications in alignment, the total length of which exceeds “t” in a
twelve times “t” except when the distance between successive indications
exceeds six times the length of the longest imperfection in the group.
t - Thickness of the weld excluding any allowable reinforcement. For a butt
weld joining two members having difference thicknesses at the weld, t is the
thinner of two different thicknesses. If a full penetration weld includes a fillet
weld the thicknesses of throat of fillet shall be included in t.
Where butt-welded joints are spot radiographed the following shall be unacceptable
12.2.2.1 2/3 t.
t - Thickness of the weld excluding any allowable reinforcement. For a
butt weld joining two members having difference thicknesses at the
weld, t is the thinner of two different thicknesses. If a full penetration
DOC.NO : NDE/RT/001
NDE PROCEDURE Page : 13 of 15
Revision : 00
SUNGOV
RADIOGRAPHIC EXAMINATION Date : 09.11.11
12.2.2.3 Any such group indications shorter than 6mm are acceptable for any plate
thickness
13.1 Radiographs shall be retained for a period of at least 5 years in proper storage
condition.
14.0 Documentation
14.12 SFD
14.13 Intensifying Screen Type & thickness
14.14 Film Manufacturer & Designation
14.15 No. of films per cassette
14.16 Technique
14.17 IQI Designation
14.18 Placement of IQI
14.19 Density
14.20 Sensitivity
14.21 Placement of “B” letter
14.22 Placement of “F” letter for film side IQI
14.23 Observation
14.24 Result
14.25 Name & Qualification level of NDE personnel
15.0 REPORT
All the details as required as per 14.0 shall be recorded in the report form
DOC.NO : NDE/RT/001
NDE PROCEDURE Page : 15 of 15
Revision : 00
SUNGOV
RADIOGRAPHIC EXAMINATION Date : 09.11.11
ANNEXURE I
REPORT No
RADIGRAPHY INSPECTION
REPORT DATE
JOB DESCRIPTION
P.O NO Job NO
Vessel No. Joint No
Extent of RT No.of shots
Weld Thickness Diameter
Reference
Procedure Ref. Code
Radiation Source Source Size
IQI Type Designation
Film Manufacturer Film Brand
Intensifying Screen
Material Thickness Front Back
TECHNIQUE
SWSI (Panoramic / Single Segment / Multiple Segment)
DWSI DWDI ( Eliptical / Super Imposed)
Souce Center / Closer than Radius /
SFD Position Beyond Radius
IQI placement Source Side / Film Side "F" Letter
Location Marker Source Side / Film Side "B" Letter
No. of Film /
Cassette
FILM QUALITY
Density At IQI Area Other Area Min Max
Min Wire Wire
Sensitivity Required seen Ug
RESULTS
SL.NO SEGMENT Interpretation Evaluation RESULT
SANGOV REVIEWED BY
NAME
SIGNATURE
AUTHORIZED
NDE LEVEL INSPECTOR