Test Setup Test Procedure

Test Components In case of Whole Blood 3 Shake the


detection buffer tube 7 Insert the cartridge into
ichroma™ II Reader
1 Turn the power
switch ‘On’ 1 Draw whole blood
10
times

ID Chip
30µL

4
Draw the
detection buffer tube
Detection
Buffer Tube
2 Add it into
detection buffer tube 8 Press ‘Select’

2 Insert ID Chip 30µL


Test Cartridge

75µL

In case of Serum,Plasma,Control

1
Draw
Serum, Plasma, Control
Reader 9 Read the test result

15µL
Just right Too little or Bubble
Result
In the middle or near
the top of the tip

3 Press ‘Multi-test’ 5 Load the


sample mixture

2
Add it into
detection buffer tube

75µL
Important
Allow ‘Detection Buffer Tube’ to attain
room temperature at least for 15µL

Rev.01.190325_AFP
30 minutes before performing the test

Make sure that lot number of the


‘ID Chip’ is exactly same as the lot
number of ‘Test Cartridge’ and
Detection Buffer Tube’
6 Waiting 15 minutes

For technical assistance; please contact: Tel: +82-33-243-1400 This reference guide is only as a quick reminder of main steps of the
Boditech Med Inc.’ Technical Services Fax: +82-33-243-9373 procedure. Please check the instruction for use for full description of
www.boditech.co.kr E-mail: sales@boditech.co.kr procedure and of any changes which may have been implemented
to the procedure. Company declines any responsibility in case of not
fully following the instructions of use provided with the reagents.