This is not a complete Instruction for use. For more detailed instructions, please refer to IFU. β-HCG Plus_Rev01_200706E
Test Components
Matches
Test Procedure
1 Draw 150 μL
(Detector diluent)
2 Add it into
detector tube
*Use within 1 minute
3 Draw 10 μL
(Whole blood/Serum/
Plasma/Control)
4 Add it into
detection buffer
tube
A 5 Shake 20 times
20 times
10 μL
150 μL 150 μL 10 μL
B
Draw 75 μL of the
6 sample mixture
7 Load the sample mixture 8 Wait 12 minutes 9 Insert the test cartridge 10 Press ‘Start’ 11 Read the test result
Result
12 Min
75 μL 75 μL
C 1 Draw 150 μL
(Detector diluent)
2 Add it into detector tube
*Use within 1 minute
3 Draw 10 μL
(Whole blood/Serum/
Plasma/Control)
4 Add it into
dample diluent tube
5 Draw 10 μL 6 Add it into
detection buffer tube
A B
10 μL
150 μL 150 μL 10 μL 10 μL 10 μL Repeat the
test procedure
A and B
For technical assistance; please contact: Boditech Med Inc. ‘ Technical Services ’ This reference guide is for the main steps’ procedure brief reminder only. Please check the full descriptions procedure and whether any changes may have been
Tel: +82-33-243-1400 Fax: +82-33-243-9373 E-mail: sales@boditech.co.kr implemented from the product insert. The company declines responsibility in case the instruction for use provided with the regents are not properly followed.