You are on page 1of 14

MAGNETOM

Replacement of Parts
System
Power Distribution

MAGNETOM ESSENZA
2 Document Version / Disclaimer / Copyright

Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.

Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.

Copyright
“© Siemens Healthineers, 2008” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

MAGNETOM M4-060.841.19.04.02 Page 2 of 14 © Siemens Healthineers, 2008


01.12
Table of Contents 3

1 Safety 4

2 FRU Overview 5

3 PDS components 6

3.1 Circuit breakers and contactors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6


3.1.1 Tools and time required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.1.2 Preliminary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Power supply V4FFM 36BB4F 24B5F(Lambda Vega 450) . . . . . . . . . . . . . . . 9
3.2.1 Tools and time required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2.2 Preliminary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2.3 Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2.4 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2.5 Concluding steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4 Changes to previous version 12

5 List of Hazard IDs 13

© Siemens Healthineers, 2008 M4-060.841.19.04.02 Page 3 of 14 MAGNETOM


01.12
4 1 Safety

 WARNING
Voltage continues to be present at the line voltage transformer even after the MR
system has been switched off.
Failure to observe safety precautions may result in death or serious bodily injury.
 Circuit breaker F100 in the mains filter box (ECA, top) has to be switched off.

 WARNING
Voltage continues to be present at the mains filter box even after the MR system
and F100 have been switched off.
Failure to observe safety precautions may result in death or serious bodily injury.
 Always set the on-site circuit breakers to OFF before performing any service
activities on the mains filter box.

Protective earth connection test.


If a protective earth connection has to be disconnected for replacing a component,
measure the protective earth connection resistance <= 100 mOhm with a ground test
 meter (Fluke 1503, or equivalent) between the protective earth connector on the “sub-
component” and the protective earth wire connected to the “main component”.
Document the successful measurement in the service report (measured value and meas‐
ured equipment) as required by IEC 62353.

MAGNETOM M4-060.841.19.04.02 Page 4 of 14 © Siemens Healthineers, 2008


01.12
FRU Overview 2 5

This chapter describes replacement procedures for the FRUs of the Power Distribution.

FRU Time Instructions Final checks Tools


Circuit breaker 20 min. ( Circuit breakers n.a. Normal tool kit
(ECA) and contac‐
tors / Page 6)
Contactor K1, 40 min. ( Circuit breakers n.a. Normal tool kit
K2, K3, K4 and contac‐
tors / Page 6)
Power supply 20 min. ( Power supply n.a. Normal tool kit
V4FFM 36BB4F V4FFM 36BB4F
24B5F 24B5F(Lambda Vega
450) / Page 9)

© Siemens Healthineers, 2008 M4-060.841.19.04.02 Page 5 of 14 MAGNETOM


01.12
6 3 PDS components

3.1 Circuit breakers and contactors

3.1.1 Tools and time required


Time
n Circuit breakers: 20 min.
n Contactors: 40 min.
Tools
Normal tool kit

3.1.2 Preliminary
n Switch off the MR system.
n Switch off F100 (PDS) in the main box.
Fig. 1: Main box on the top of ECA

(1) F100 (PDS)

n Remove the front cover.

MAGNETOM M4-060.841.19.04.02 Page 6 of 14 © Siemens Healthineers, 2008


01.12
PDS components 3 7

Fig. 2: Front cover on the PDS

1
1
2

1 1

1 1

(1) Screws
(2) Front cover

n After the front cover is removed, circuit breakers can be replaced as needed.
Fig. 3: Circuit breakers

n Remove the top cover.

© Siemens Healthineers, 2008 M4-060.841.19.04.02 Page 7 of 14 MAGNETOM


01.12
8 3 PDS components

Fig. 4: Top cover for PDS

1
1
2

1 1

(1) Screws
(2) Top cover

n After the top cover is removed, contactors can be replaced as needed.


Fig. 5: Contactors

1
2

3
4

(1) K2
(2) K1
(3) K3
(4) K4

MAGNETOM M4-060.841.19.04.02 Page 8 of 14 © Siemens Healthineers, 2008


01.12
PDS components 3 9

3.2 Power supply V4FFM 36BB4F 24B5F(Lambda Vega 450)

3.2.1 Tools and time required


Time
20 min.
Tools
Normal tool kit

3.2.2 Preliminary
n Switch off the MR system.
n Switch off F100 (PDS) in the main box.
Fig. 6: Main box on the top of ECA

(1) F100 (PDS)

Fig. 7: Power supply (Lambda Vega 450)

© Siemens Healthineers, 2008 M4-060.841.19.04.02 Page 9 of 14 MAGNETOM


01.12
10 3 PDS components

3.2.3 Disassembly

Fig. 8: Lambda Vega 450 n Remove the cables


n Remove the mounting screws

1
1
(1) Mounting screws

3.2.4 Assembly

Fig. 9: PSU connection n Install the new PSU


n Reconnect the cables.

(1) 100-240V

MAGNETOM M4-060.841.19.04.02 Page 10 of 14 © Siemens Healthineers, 2008


01.12
PDS components 3 11

Fig. 10: PSU connections

3 1

4
2

(1) 36V
(2) GND 36V
(3) 24V
(4) GND 24V

3.2.5 Concluding steps


n Switch on F100 (PDS) in the main box.

© Siemens Healthineers, 2008 M4-060.841.19.04.02 Page 11 of 14 MAGNETOM


01.12
12 4 Changes to previous version

Version 02 VS Version 01
n Change the graphics.( Fig. 6 Page 9)
Version 03 VS Version 02

Chapters Changes
( Safety / Page 4) Add the content about Protective earth connectiont‐
est.

MAGNETOM M4-060.841.19.04.02 Page 12 of 14 © Siemens Healthineers, 2008


01.12
List of Hazard IDs 5 13

There are no Hazard IDs in this document.

© Siemens Healthineers, 2008 M4-060.841.19.04.02 Page 13 of 14 MAGNETOM


01.12
All documents may only be used for
rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec‐
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re‐
served.

siemens-healthineers.com

Siemens Healthineers Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens-healthineers.com

Print No.: M4-060.841.19.04.02 | Replaces: M4-060.841.19.03.02


Doc. Gen. Date: 01.12 | Language: English
Part No.: 10165094
© Siemens Healthineers, 2008

You might also like