Document # QAW00135 Revision

Document Title: Stepper Motor calibration C


Procedure Page 1 of 24

Stepper Motor
calibration
Procedure
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 2 of 24

Approvals

Changed By: Checked by: Approved By:


Name: Adi Nahmani Ziv Shefi Igor Naroditsky
Title VP Global Customer Integration Team Engineering Manager
Support Manager
Date 09/01/2017 09/01/2017 09/01/2017
Signature AN ZS IN

Document Change History

Revision Date Approved by Description of Change


A 03/09/2015 Yehuda Otzeri New document

B 09/01/2017 Adi Nahmani Adding additional


step to make sure the
Patients database is “Safe”, P.14
C 09/27/2017 Adding additional step for the D-Spect SOC

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 3 of 24

Content

STEPPER MOTOR CALIBRATION PROCEDURE ................................................................................ 1


APPROVALS ........................................................................................................................................... 2
NAME:...................................................................................................................................................... 2
DOCUMENT CHANGE HISTORY ........................................................................................................... 2
PURPOSE:............................................................................................................................................... 4
SAFETY: .................................................................................................................................................. 4
REQUIRED TOOLS: ................................................................................................................................ 4
PROCEDURE: ......................................................................................................................................... 5

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 4 of 24

Purpose

This Work instruction describes the Stepper Motor Calibration procedure (at the customer's site).

Caution
The motor data is stored in the DB-Hirarchy file which also contains all the patients' information.
At the beginning of the procedure you will prepare a backup DB-Hirarchy file that can be used in case of a problem.
The source DB-Hirarchy should be copied from the Master PC to the Salve PC.
The calibration procedure is done from the Slave PC and then the file is copied back to the Master PC.

Safety

No special safety instructions.

Required tools

No work tools are required.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 5 of 24

Procedure

1. Go to My Computer:

2. Go to directory: E:\SRC\D-SPECT\Database:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 6 of 24

3. Right click on DB_Hirarchy.mdb and choose Copy:

4. Right click on an empty space in the folder and choose Paste:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 7 of 24

You have now created a backup copy of the DB_Hirarchy file. This copy will be used in case of a problem!

Rename: DB_Hirarchy – Copy.mdb, to: DB_Hirarchy – Today’s Date.mdb.

Important comment: If the system is a D-Spect SOC (Slave on Chip) - please proceed directly to step # 22.

5. Close the open window & press the start button,


Choose Remote Desktop Connection:

Connect to the following IP Address 21.0.1.200:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 8 of 24

6. Remote connect to the slave PC:

7. Log in using the following information:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 9 of 24

User name: machine (small letters) ; Password: spectrum (small letters).

8. From the Slave, press start & choose My Computer:

9. Double click E on MASTER-SW-XXXXX:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 10 of 24

Notes
• In case you do not see the E-Drive, perform the following steps.
• In case you see the E-Drive, go directly to step 16.

10. From the Master PC press the start and click on Remote Desktop Connection:

11. Connect to the following IP 21.0.1.200:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 11 of 24

12. Select Options (at the Remote Desktop connection):

13. Click on Local resources & then click More:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 12 of 24

14. Check the Drives box and click OK:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 13 of 24

15. Click Connect:

16. From the Slave PC go to the Master PC: E:\SRC\D-SPECT\Database.


17. Right click on: DB_Hirarchy and choose Copy:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 14 of 24

18. On the Slave Press start & choose My Computer:

19. Go to directory: C:\SRC\D-SPECT\Database:

20. Right click on the empty space in the folder and choose Paste:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 15 of 24

21. Choose Yes to confirm replacement of the DB_Hirarchy file:

You have now copied the DB_Hirarchy file from the Master PC to the Slave PC!
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 16 of 24

22. Close the window & on the desktop double click on the Calibration Tools:

Important notes
• If the system is D-Spect SOC- Exit from the DSpect background service:

• On the D-Spect SOC- there’s no Slave. The Calibration Tools to be operated by the following route: E:\SRC\D-
SPECT\LAST_OUTPUT_RELEASE\CALIBRATION TOLLS.EXE.

23. The next screen appears:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 17 of 24

24. Press Initialize HW.


(The button turns from Gray to Pink and after a while it turns back to Gray):

 
Important Note: While opening the Calibration Tools SW from the Master Explore, Check Local (not Remote),
before pressing initialize!

25. Click Testing & then Stepper Motors Testing & then Motor Encoder Calibration:

26. Click the upper Select csv file:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 18 of 24

Note: This is the Encoder Calibration Motor File.

27. In the OutPut_Release_x.x.x.x, choose StepperCalibrationScanPattern & click Open:

Note: The full path is: C:\SRC\D-SPECT\OutPut_Release_x.x.x.x.

28. Click the lower Select csv file:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 19 of 24

Note: This is the Encoder Calibration Result File.

29. Save the new file as CalibrationresultsFile(x+1):

Note: x is the current version.


Example: if the current version is 3  the current file name to be CalibrationresultsFile3  the new file name will
to be CalibrationresultsFile4.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 20 of 24

30. Click on Run Encoders Calibration:


Notes
• The text changes from Run Encoders Calibration to Abort Encoders Calibration and the calibration
process is running.
• The Calibration process is finished once it reaches 99%:

31. Click on Testing and then on Stepper Motor Testing and then on Motor Write Registers:

32. Verify that Use File for Setting is checked and click on Save Setting:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 21 of 24

 

33. Save the file as StepperMotorSetting:

 

Notes
• The file is being saved in the same name as the old file.
• The old file is being overwritten by the new one.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 22 of 24

34. Confirm the replacement of the old file with the new one:

35. Press OK to confirm:

 

36. Uncheck Use File for Setting and click on Save Setting:

 
Note: You will not receive any confirmation that the file has been saved!

37. Exit from the Calibration tools SW.


Important note
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 23 of 24

If the system is D-Spect SOC (Slave on Chip) - you done!

38. From the Slave PC click on Start and then click My Computer:

39. Go to: C:\SRC\D-SPECT\Database, left click on DB_Hirarchy and choose Copy:

40. From the Slave PC go to the Master PC: E:\SRC\D-SPECT\Database.

41. Right click on the empty space in the folder and choose Paste:

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 24 of 24

42. Choose Yes to confirm replacement of the DB_Hirarchy file:

You have now copied the DB_Hirarchy file from the Slave PC back to the Master PC!

43. The Calibration procedure in now finished.