Stepper Motor
calibration
Procedure
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 2 of 24
Approvals
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 3 of 24
Content
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 4 of 24
Purpose
This Work instruction describes the Stepper Motor Calibration procedure (at the customer's site).
Caution
The motor data is stored in the DB-Hirarchy file which also contains all the patients' information.
At the beginning of the procedure you will prepare a backup DB-Hirarchy file that can be used in case of a problem.
The source DB-Hirarchy should be copied from the Master PC to the Salve PC.
The calibration procedure is done from the Slave PC and then the file is copied back to the Master PC.
Safety
Required tools
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 5 of 24
Procedure
1. Go to My Computer:
2. Go to directory: E:\SRC\D-SPECT\Database:
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 6 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 7 of 24
You have now created a backup copy of the DB_Hirarchy file. This copy will be used in case of a problem!
Important comment: If the system is a D-Spect SOC (Slave on Chip) - please proceed directly to step # 22.
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 8 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 9 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 10 of 24
Notes
• In case you do not see the E-Drive, perform the following steps.
• In case you see the E-Drive, go directly to step 16.
10. From the Master PC press the start and click on Remote Desktop Connection:
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 11 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 12 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 13 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 14 of 24
20. Right click on the empty space in the folder and choose Paste:
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 15 of 24
You have now copied the DB_Hirarchy file from the Master PC to the Slave PC!
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 16 of 24
22. Close the window & on the desktop double click on the Calibration Tools:
Important notes
• If the system is D-Spect SOC- Exit from the DSpect background service:
• On the D-Spect SOC- there’s no Slave. The Calibration Tools to be operated by the following route: E:\SRC\D-
SPECT\LAST_OUTPUT_RELEASE\CALIBRATION TOLLS.EXE.
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 17 of 24
Important Note: While opening the Calibration Tools SW from the Master Explore, Check Local (not Remote),
before pressing initialize!
25. Click Testing & then Stepper Motors Testing & then Motor Encoder Calibration:
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 18 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 19 of 24
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 20 of 24
Notes
• The text changes from Run Encoders Calibration to Abort Encoders Calibration and the calibration
process is running.
• The Calibration process is finished once it reaches 99%:
31. Click on Testing and then on Stepper Motor Testing and then on Motor Write Registers:
32. Verify that Use File for Setting is checked and click on Save Setting:
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 21 of 24
Notes
• The file is being saved in the same name as the old file.
• The old file is being overwritten by the new one.
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 22 of 24
34. Confirm the replacement of the old file with the new one:
36. Uncheck Use File for Setting and click on Save Setting:
Note: You will not receive any confirmation that the file has been saved!
38. From the Slave PC click on Start and then click My Computer:
41. Right click on the empty space in the folder and choose Paste:
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 20 November, 2017 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00135 Revision
Document Title: Stepper Motor calibration C
Procedure Page 24 of 24
You have now copied the DB_Hirarchy file from the Slave PC back to the Master PC!