Document # QAW00130 Revision

Document Title: Replacing column ASSY P/N A


9020207 GE or 9020227 REDLEN instructions
Procedure Page 1 of 17

Replacing column ASSY P/N 9020207 GE or 9020227 REDLEN instructions Procedure

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 2 of 17

Approvals

Prepared By: Checked by: Checked by: Approved By:


Name: Nir Hertzman
Title
Date
Signature

Document Change History

Revision Date Approved by Description of Change


A New document

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 3 of 17

Content

REPLACING COLUMN ASSY P/N 9020207 GE OR 9020227 REDLEN INSTRUCTIONS PROCEDURE 1


APPROVALS .........................................................................................................................................................2
NAME:.....................................................................................................................................................................2
OCUMENT CHANGE HISTORY .......................................................................................................................2
PURPOSE: ..............................................................................................................................................................4
SAFETY:.................................................................................................................................................................4
REQUIRED TOOLS: ............................................................................................................................................4
METHOD: ..............................................................................................................................................................5

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 4 of 17

Purpose:

This Work Instruction describes the column ASSY P/N 9020207 GE or 9020227 REDLEN replacement

Safety:

Turn off the system before you start this procedure.


Turn off the MCB.

Required tools:

2.5 mm Hexagon wrench bar


3 mm Hexagon wrench bar

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 5 of 17

Method:

1. Take off only the Gary cover according to procedure QAW00125 section 1-4

2. You can now see column jig (rack plate column) P/N 9351085

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 6 of 17

3. Release the column jig from the two plastic holders.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 7 of 17

4. Insert the column jig the holders on the head

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 8 of 17

5. Lower the jig until it horizontal to the floor and can be used as a rail for the column.

6. Using a 2.5 mm & 3 mm Hexagon wrench bar release the two screw on the top of the column and the three screws on
The bottom and one screw 3 of the column.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 9 of 17

7. Release the plastic connectors holding the communication (LAN) cables.


It is recommended to release at least three plastic cable holders.

8. Slide the column out on the rail until you reach the slot.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 10 of 17

9. Once the column reaches the slot, turn the column sideway so you can reach the connectors on the electronic board
And release the communication (LAN) cable and both flat cables.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 11 of 17

10. Continue to move the column to the notch at the end of the jig, and tilt the column up, so the column can slide above it.

11. Slide the column above the notch and all the way to the end of the rail and remove the column.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 12 of 17

12. In case you need to replace column parts ASSY referee to the appropriate replacement instruction. There are
instruction for replacement of the following parts:
a. Column flat cable from MCB to encoder replacement procedure QAW00121
b. Column Fan replacement procedure QAW00120
c. Column MCB electronic board replacement procedure QAW00131
d. Column motor replacement procedure QAW00123
e. Column flat cable from MCB to HVCD board replacement procedure QAW00122

13. Take the new column ASSY p/n 9020207 GE or 9020227 REDLEN or the repaired column.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 13 of 17

14. Insert the column ASSY on the column jig. Slide the column above the notch.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 14 of 17

15. Slide the column to the slot, rotate the column and reconnect the communication (LAN) cable and both flat
cables.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 15 of 17

16. Slide the column all the way back into the head.

17. Close the plastic cable holders holding the communication (LAN) cables.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 16 of 17

Using a 2.5 mm & 3mm


Hexagon wrench bar close the two
Screws on the top of the column and
The three screws on the bottom of the column

18. Return the jig and close the holders

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template
QAF 0009
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # QAW00130 Revision
Document Title: Replacing column ASSY P/N A
9020207 GE or 9020227 REDLEN instructions
Procedure Page 17 of 17

19. Calibrate the motor encoder.(according to procedure number QAW00135)


20. Perform D-SPECT system calibration procedure QAW0074 including block alignment.
21. Perform daily QC test.
22. The column replaced procedure is finished.

Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after local date 15 February, 2016 Document Template