Professional Documents
Culture Documents
FCO0005 - A - Fans Speed and Dim Instruction Procedure
FCO0005 - A - Fans Speed and Dim Instruction Procedure
Document Title:
Upgrade SoC FW to 0x63 and control fans speed and
led stripe dimming
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
Purpose
This document describes how to upgrade the MSC Firmware Version to 0x63,
setup the fans speed and control LED stripe dimming
General information
Upgrade to FW Ver 0x63 is necessary to run the fans and led control.
One option to verify the firmware version is to open the covers, if the connector (see below
picture) is plugged in to the MSC and the LED is on (on the Gantry cover ) that means that
the firmware is 0x63 and we can skip step 1 ,
If the connector is unplugged and the LED is on it means that the version is older and it is
necessary to upgrade before using the patch
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
Instructions
1. The MSC board operated by FPGA with Firmware and ARM controller running
Linux operating system.
This document describes how to change the MSC FPGA firmware by transferring
new MSC firmware file (*.rbf) the file will be uploaded to the MSC flash memory by
using the Micro SD flash card, this uploaded file will be used to load the MSC board
FPGA.
a. Tools required
• Micro flash SD card 8GB or 16GB FAT32 format.
• Micro SD card to USB adaptor.
• U-boot.scr file
• msc.rbf file
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
See step 5
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
g. Power on the system power from the main switch and wait for the systems fans start
spinning.
NOTE: when systems fans start spinning it indicates the FPGA loaded
h. When firmware upgrade done, take off the micro SD flash card
i. Switch off the D-SPECT power from the main switch
j. Close the system covers
k. Turn on the D-SPECT system main switch
l. Verify the systems fans start spinning
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
2. Download the patch “MSC firmware 0x63” from the service web site
• Click “YES”
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
• Click “NEXT”
• Click “Install”
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
6. Go to folder “E:SRC\D-spect\sw_upgrade\SoC-Slave_patch0001-
LEDS_and_SideFans”
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.
Document # FCO0005 Revision
Unless this printed document bears the controlled copy footnote, the document as a printout is invalid 24 hours after
local date 3 December, 2019
Proprietary and Confidential ~ Property of Spectrum Dynamics Medical, Ltd.