Professional Documents
Culture Documents
Return Policy
the electrical installation of the relevant room complies
Rigid Endoscope
with the applicable national and local requirements, Return Procedure
The device bears CE mark indicating its conformity with the
M 01000A/M 01030A/M 01000 and Product Description
provisions of the Council Directive 93/42/EEC concerning In the event that it becomes necessary to return this device
/M 01030 the device is used in accordance with the instructions
The product is composed of the following parts:
medical devices and fulfills the essential requirements of or part of this device to Mindray, the following procedure
for use.
Rigid Endoscope Annex I of this directive. should be followed:
The device complies with the requirements of EN/IEC Return authorization: Contact the Customer Service
60601-1-2 “Electromagnetic Compatibility – Medical Department and obtain a Customer Service Authorization
Electrical Equipment”. It is important for the hospital or organization that number. This number must appear on the outside of the
th
Operator's Manual employs this device to carry out a reasonable Figure 1Rigid Endoscope
© 2018 Nanjing Mindray Bio-Medical Electronics Co., Ltd. All shipping container. Returned shipments will not be accepted
rights Reserved. service/maintenance plan. Neglect of this may if the number is not clearly visible. Please provide the model
For this Operator’s Manual, the issued date is 2018-05 result in machine breakdown or personal injury. number, serial number, and a brief description of the reason 1. Objective lens 2. Eyepiece
(Version: 2.0). for return. 3. Beam channel interface 4. Richard-wolf beam
Freight policy: The customer is responsible for freight
(used for the ACMI system) channel connection adapter
Intellectual Property Statement This device must be operated by skilled/trained
charges
arges when this device is shipped to Mindray for service
system
(this includes customs charges).
NANJING MINDRAY BIO-MEDICAL ELECTRONICS CO., clinical professionals.
Return address: Please send the part(s) or device to the 5. Storz beam channel 6. Lens
LTD. (hereinafter called Mindray) owns the intellectual
address offered by the Customer Service Department. connection adapter system
property rights to this Mindray device and this manual. This Warranty
manual may refer to information protected by copyrights or
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL Contact Information
patents and does not convey any license under the patent Safety Instructions
OTHER WARRANTIES, EXPRESSED OR IMPLIED,
rights or copyright of Mindray, or of others. Manufacturer: Nanjing Mindray Bio-Medical Safety Markings
INCLUDING WARRANTIES OF MERCHANTABILITY OR
Mindray intends to maintain the contents of this manual as Electronics Co., Ltd. Markings Meaning
FITNESS FOR ANY PARTICULAR PURPOSE.
confidential information. Disclosure of the information in this Address: 666# Middle Zhengfang Road, Jiangning, Indicates the existence of hazards, which, if
manual in any manner whatsoever without the written 211100 Nanjing, Jiangsu, P.R.China not avoided, may cause death or serious
Exemptions
permission of Mindray is strictly forbidden. Tel: +86 25 66082666 injury.
Release, amendment, reproduction, distribution, rental, Mindray's obligation or liability under this warranty does not Indicates the existence of potential
Fax: +86 755 26582680--26666
adaption and translation or any other derivative work of this include any transportation or other charges or liability for hazards, which, if not avoided, may cause
manual in any manner whatsoever without the written direct, indirect or consequential damages or delay resulting personal injury or product/property loss and
permission of Mindray is strictly forbidden. from the improper use or application of the device or the use EC-Representative: Shanghai International Holding damage.
of parts or accessories not approved by Mindray or repairs by Corp. GmbH (Europe) "Caution" indicates an issue that should be
, , are the trademarks, people other than Mindray authorized personnel. Address: Eiffestrasse 80, 20537 Hamburg, noted. Please read the manual carefully
registered or otherwise, owned by Shenzhen Mindray This warranty shall not extend to: Germany before using the system.
Bio-Medical Electronics Co., Ltd. All other trademarks that Malfunction or damage caused by improper use or Tel: 0049-40-2513175
2513175 CF type application part
appear in this manual are used only for informational or man-made failure.
CE logo complies with medical device
Fax: 0049-40-255726
255726
editorial purposes. They are the property of their respective Malfunction or damage caused by unstable or
Directive EEC.
owners. out-of-range power input.
Indicates that the product may be sterilized
Malfunction or damage caused by force majeure
with high pressure and high temperature,
Responsibility on the Manufacturer Party such as fire and earthquake.
including steam sterilization.
Malfunction or damage caused by improper
Contents of this manual are subject to change without prior Manufacturer
operation or repair by unqualified or unauthorized
notice. Date of production
service people.
All information contained in this manual is believed to be
Malfunction of the device or part whose serial
correct. Mindray shall not be liable for errors contained herein Temperature limitations
number is not legible enough.
or for incidental or consequential damages in connection with
Others not caused by device or part itself.
the furnishing, performance, or use of this manual. Storage and transport humidity
IPX7 IPX7 waterproofing rating resonance imaging environment. Before cleaning, disinfecting, or sterilizing the product, test applications. If the medical appliances or accessories being Do not let the far end of endoscope or the beam
WEEE label All devices that connect with this piece of equipment the product using the following steps: used with this product meet CF requirements, then the whole connectors touch a patient's tissues, combustible
The following definition of the WEEE label must be certified with the designated IEC standards Testing of fiber optics endoscopic system can meet the protection requirements for materials, or heat-sensitive materials.
applies to EU member states only: the use (such as IEC 60601-1 Medical Device Standards). CF application. Do not touch the far end of the endoscope or the
1. Point the far end of the endoscope at a bright
of this symbol indicates that this product The operator is responsible for connecting and Medical appliances
pliances or accessories used with this product beam connectors.
light source (not a cold light source) like an
should not be treated as household waste. configuring additional equipment and that the should at least meet the requirements for BF applications. Prevent the surface of the endoscope's far end from
indoor lighting source and move the endoscope,
By ensuring that this product is disposed of system is IEC 60601-1 compliant. For any becoming contaminated, especially the light
questions, please contact the supplier. and then bring the near end up to the eye. Products with this symbol meet the emitting surface.
correctly, you will help prevent bringing
potential negative consequences to the Non-sterile components must be disinfected and 2. The brightness of the fiber optics will change, in requirements for CF applications. The applied part is the metal part of rigid
environment and human health. For more sterilized before use. the even that the fiber has been damaged or endoscope.
detailed information with regard to Improper use or maintenance of the product or use
broken. A single broken fiber will not cause any
returning and recycling this product, please beyond the intended purpose may increase risks to
danger. If 10-20% of the fibers become broken, it
The beam connection port and the endoscopic Routine Maintenance
consult the distributor from whom you patients and result in premature wear and tear of
metal hood on the front of the cold light source will
the product. is suggested you send the endoscope in for
purchased the product. have a surface temperature of over 41 ° c after a
Surgery Cleaning and
After the product has been cleaned, disinfected, and repair. Disassembly Sterilization
long period of use, avoid contact with these parts to pretreatment disinfection
sterilized the product shall be cooled down before Inspection of glass surfaces
Intended Use avoid burns.
use. Do not use the product at high-temperatures. 1. Make a visual inspection of lens surfaces, Using magnetic resonance imaging (NMR),
High energy radiated light may be transmitted from ensuring they are both clean and smooth. Endoscope reprocessing steps are as follows:
endoscopic and accessories can lead to danger and
objective lens of the endoscope, please do not look 2. If damage is found during inspection, the The endoscope should be cleaned, disinfected and sterilized
produce man-made artifacts;
s; please follow the
The product can be used for the following purposes
directly at the light outlet in case of causing eye damage may be remedied by consulting the before and after each use; otherwise it may cause bacterial
manufacturer's instructions and safety precautions.
only.
burn. "Common Faults and Solutions" section infection. The disinfection and sterilization methods should
This product may be used in coordination with a
It is used in the operating room of hospital.
Do not modify this equipment without authorization Inspection of lens surfaces be used only after they are validated by the application unit.
laser and high frequency surgical instruments,
of the manufacturer 1. Make a visual inspection of lens surfaces, pneumatic or electro-hydraulic
hydraulic endoscopic
The product is used for the observation and imaging of Reagents and equipment applied for cleaning, ensuring they are both clean and smooth. Disassembly
accessories in many therapeutic areas. Please
minimally invasive surgery in the abdomen, used in medical disinfection and sterilization procedures must be 2. If bends, depressions, cracks, or other forms of The endoscope should be removed from the camera system
follow the safety instructions of the product and any
checked and accepted, and the intended use must damage are found during inspection, it is before cleaning, disinfection, and sterilization while the beam
institutions accessories used with the product.
be ensured within the period of validity. suggested that you send the endoscope in for connectors should be pulled out of the endoscope. If the
Contraindication: It is prohibited to use this product in If the product is used together with an endoscope
repair. camera system or a cold light source which fails to endoscope is cleaned, disinfected, and sterilized without
circumstances not suitable for endoscopic therapy.
meet the requirements for BF or CF applications, disassembly, it may fail to reach the desired effect.
Protect the product's original packaging in case the Using this product with any sharp edges or broken the possibility of current leakage may increase.
Safety Warning Before use, the operator must check the equipment, Pretreatment of Operation Room
product needs to be returned to the factory for surfaces will put patients at risk. So before each
repair. use, please check to ensure there are no cables and accessories, to ensure they work The time from cleaning and disinfection to operation should
Bending, twisting, pulling or compression may unintended rough surfaces, sharp edges or properly and safely. not exceed 6 hours; otherwise deep stains from operation will
The product must be operated by a trained operator easily damage the product. These types of actions protrusions which may cause harm. Risks associated with the use of a cold light source: increase the difficulty of cleaning or result in ineffective
with the necessary knowledge and experience to do may further damage the optical elements and Cause irreversible damage to a patient's tissues or disinfection or the corrosion of the stainless steel.
so. All operators should read the manual carefully results in surgical failures. Note: Only use the unnecessary coagulation. Immerse endoscope into demineralized or distilled
before using the product. product in designated environments. Cause the combustion or burning-out
burning of surgical water. Wipe the surface soil of device with disposable
Do not use the fiber optics if the glass surface has
The product must be used in the manner shown When disposing of the product and its accessories instruments (such as surgical drapes, plastic soft clothes.
been damaged or if there are stains on it that cannot
within the manual. Any operations of the product do so strictly follow local regulations. materials, etc.). Put the device in the fixed tray to avoid be damaged.
be removed with cleaning. Return damaged
that do not comply with the instructions, warnings If the light source is damaged during use, it may be
products to the manufacturer or authorized repair
and precautions may result in risk and other serious a danger to patients. Therefore, a standby light Cleaning and Disinfection
consequences. Check the product and its
Getting Started center. Information about authorized repair centers
source should be provided or a light source with
may be obtained from the manufacturer. Manual cleaning
accessories before use to ensure they are extra bulbs is used.
1. Cleaning
functioning correctly. Installation Safety precautions:
Disassembly of Component Disassemble adapter and immerse the adapter and
Ensure that all devices operating nearby meet According to guide beam interface forms, select the Avoid using strong light sources for an extended
endoscope into the neutral multienzyme cleaning
related requirements for electromagnetic appropriate adapter, connect the cold light source beam into Remove all adapters. period of time.
agent(3M multienzyme cleaning agent).The
the endoscope-guided beam interface.
recommended reagent concentration is (5-10)mL/1L may worsen stains, such as ethanol and alcohol. regulations. 285 and tested in accordance with DIN EN ISO 17665. image not clean surface as per the
distilled water, immersed for (2-10)min at (25-40) °C. Only cleaning agents and disinfectants permitted for High-temperature disinfection should be performed Steam Sterlizaiton Parameters steps described in
Completely immerse the device into cleaning water use with the product should be used. Follow the with fully demineralized water and the A value must Sterilizaiton Temperature Exposure Pressure "Cleaning and
and clean the inside surface of adapter and connect manufacturer's instructions. Follow the reach above 3,000. process time Disinfection".
o
parts of lens and endoscope rod with a short cleaning manufacturer's instructions for recommended Maintain and inspect the endoscope cleaning and Gravity 121 C 30min 102.9kPa Remove the residues
brush. concentrations, temperature, period of usage. disinfection machine regularly. displacement using the steps
o
Wipe the surface of device with disposable soft clothes, Do not use a metal cleaning brush as it may Put the product in a meshed tray (to ensure all parts Pulsation 132 C 4min 205.8kPa There are deep stains described in
until no visible soil can be observed. damage the product surface and cause corrosion. can be flushed). Vacuum on a glass surface "Cleaning and
2. Intermediate cleaning Strictly follow the processing parameters Sterilization tolerance cycle: 500 times. Disinfection" and
Rinse the device with demineralized or distilled water. recommended by the chemical manufacturers, such check water quality.
Sterilization 2. Gas sterilization
Dry the device external surface with disposable soft as temperature, concentration, exposure time, etc. The lens system is Send back the faulted
Before sterilizing the endoscope, make sure it functions It is recommended to use the following validated ethylene
cloths. Otherwise, the following problems may result: damaged or not endoscope system
normally and goes through the complete cleaning and oxide sterilization parameters:
Wipe the surface of eye lens, objective lens and light The optical performance of materials will decline, sealed properly for repair.
disinfection process, please conduct testing as per the EO concentration Temperature Humidity Dwell time
transmitting lens with sterile swab soaked 70%-alcohol such as tarnishing or fading of titanium or Clean the glass
requirements of "Getting Started" chapter ("Testing" section). 100% 54 oC 50%RH 360min
clockwise direction. aluminum surfaces. For aluminum surfaces, the surface as per the
To ensure the ethylene oxide residues at a level that does no The glass surface is
Rinse the surface of eye lens, objective lens and light chemicals with pH>8 used for processing will steps described in
Endoscopes marked "AUTOCLAVE" may be given harm to the human body, the sterilized endoscope should go not clean
transmitting lens with demineralized or distilled water. cause surface changes. "Cleaning and
priority for using pressure steam sterilization. through a 12-hour resolution at a well- ventilated room with a
Dry the device surface with disposable soft clothes. Material damage, such as corrosion, cracking, Disinfection".
It is not recommended to use several sterilization temperature no less than 40℃ before reuse.
3. Disinfection fracturing, premature aging, or expansion. Remove the residues
methods because it may cause a significant decline Sterilization tolerance cycle: 500 times.
Do not use the preparations that could lead to the using the steps
Only use disinfectants that have been identified to be
of equipment performance. Only use one Note: after long time sterilization, the color of the color ring
synthetic materials splitting, or make the materials There are deep stains described in
effective and compatible with an endoscope. The
recognized method. will fade, which is a normal phenomenon. It will not affect the
brittle. on a glass surface "Cleaning and
recommended sterilization parameters for reference:
The endoscope may be damaged if it is cooled sealing performance and the whole machine function.
After manually cleaning/disinfecting, check whether Disinfection" and
-- 2% glutaraldehyde, disinfection time 20-30 min,
quickly after autoclave sterilization, so it should
there are stains remaining on the visible surfaces. If The image check water quality.
solution temperature 20-25°C;
naturally cool to room temperature after autoclave Maintenance
necessary, repeat the cleaning steps. is too dark Problems with light Use the beams with a
-- 0.55% o-Phthalaldehyde, disinfection time 12 min,
sterilization.
Please use the cleaning, disinfection and The product is maintenance-free and contains no with low beams reliable connection
solution temperature 20-25°C;
The product is very sensitive to impacts at high
sterilization procedure recommended in this components requiring maintenance by the user and brightness. The light beams are
Fully immerse the endoscope in the disinfectant to
temperature, so avoid impacting and vibrating the
manual. not connected with Check the connection
ensure all accessible surfaces are moistened; manufacturer.
product at high temperatures.
Flush the internal cavities and guide slots with a the endoscope of light beams