Mx TWIN – Inverter FET Check

Rev: April 2002 TRN 29 - Page 1 of 2


Please complete all prerequisite modules. If you cannot comply, see an instructor.

Successful Completion: 1. Ability to check the FETs on the Inverters.

System: Mx Twin

Objectives: Upon successful completion of this module the student will:


1. Perform Inverter FET check.

Lecture/Demo: N/A

Reading Assignment N/A

Lab Assignment: 1. Perform Inverter FET check

Special Tools/Procedures:

——————————————————————————————————— NOTICE ————————————————————————————————


This document is the property of Marconi Medical Systems, and is intended for exclusive use by current Marconi personnel. Copying, disclosure to others, or other use is prohibited without the express written authorization of the Marconi Medical
Systems Law Department. Report violations of this requirement to Marconi Medical Systems., 595 Miner Rd; Highland Heights, OH 44143; Attention: Vice President, Service Division
Mx TWIN – Inverter FET Check
Rev: April 2002 TRN 29 - Page 2 of 2

Procedure:

1. Perform a functional study consisting of a surview and an axial head protocol.

Note:
Inverter FET check should be performed every 6 months!!!!
FET Check
2. There are 40 FETs in each Inverter.
Verify with a multimeter that the voltage between Drain and Source is between 0.3 V – 0.5 V.
(The multimeter should be in diode check position.)

3. If the voltage between Drain & Source exceeds the above voltage, replace the FET or the Inverter.

FET DRIVER Test


4. There is a 8 pin driver near each FET.
Check with an Ohmmeter the resistance between pins 3 – 7 of the driver.
The resistance should exceed 100 ohms.

5. If the resistance is less than the above, replace the Inverters.

SOURCE

DRAIN (Black lead)

GATE

Figure: FET Top View