Warning: To ensure your measuring operates more smoothly and also for the sake
of your personal safety as well as the appliance, please read this manual
carefully before your operation.
2.1 With cases of luxury sectional bar, which is portable, smaller (it is the smallest one in
domestic products of this kind. 355*145*185mm), lighter and easier to take.
2.3 Measurement of various ways: It has ability to measure AC voltage, current, power (active
power, reactive power, apparent power) and electric energy in four-quadrant with
connection modes such as single-phase and four-wire three-phase Y/ three-wire Δ. In
addition, it can measure reactive power and electric energy with ways of true reactive,
cross-phase reactive power, artificial neutral point of reactive.
2.4.1 Fundamental function of measuring for voltage, current, phase, frequency, active
power, reactive power, power factor and error.
Page -1-
SY-3102 Standard Meter Operation Manual
2.4.6 Display of vector diagram. There are a variety of display ways for vector diagram to
meet the different display habits (such as clockwise rotation, counterclockwise
rotation; the choice of the benchmark location between 12:00 o’clock and 3:00
o’clock; reference benchmark can be changeable; phase settings for display ways:
0 - 360°or ±180°).
2.4.13 It can adjust errors of three different kinds of meters (active power, reactive power,
apparent power, voltage, and current).
2.4.14 There are three ports for pulses to output. And its constants can be set
automatically or manually. Types of constants can be set also: active power, reactive
power, apparent power, voltage, and current.
2.4.15 It can check the comprehensive error of device composed of three single-phase
meters.
2.5 Measurement of power for fundamental wave: not only could it measure the full power
(including powers of fundamental wave and orders harmonics), but also the fundamental
power only.
2.6 Adjustment of electric energy error: it can adjust the estimate value of standard deviation
and computation error of electric energy meters as well as calibration device for energy
meter. Moreover, counts of pulse and constants of meters to be checked can be set
automatically so as to work effectively. Especially for the multi-constant standard meter
and the energy meter calibration device, it will be more convenient. By adopting the use
of special compensation techniques, error-beating would not be large because of the
small load (the low frequency of the energy pulses). It can display the current error; the
value of the standard deviation estimates and the average error.
2.7 Harmonic analysis: it can analyze the harmonics of voltage and current under 51 orders.
2.8 Display of data and graphics: it can display the measuring working condition and multiple
measuring data directly. It can also display the waveform of the voltage and current to be
measured, the phase vector graph of the voltage and current and waveform of harmonic
analysis directly.
Page -2-
SY-3102 Standard Meter Operation Manual
2.9 The measuring range can be switched manually or automatically.
2.10 With the function of Pinyin input method: Chinese Character, alphabet and symbol can be
inputted.
2.11 Connection: the input port of the wide-range current is a public port, and there is no need to
change connection mode while the current is changed, so it is very convenient to use.
2.12 Multi-use: it can be used as a standard-meter, also as a calibration device for energy meter.
It can be used alone, with PC by remote communication, also with calibration device.
2.13 Multiple ports for communication: RS485, RS232, USB, and baud rate can be set when
using asynchronous communication.
2.14 With 6.4-inch 640*480-pixel TFT display, it shows clearly and colorfully.
2.15 Operation: human orient design, reasonable combination of the function and simple
operating.
Page -3-
SY-3102 Standard Meter Operation Manual
3.2 Accuracy
3.1.1 Voltage measurement (RMS):
Accuracy: 0.01% (class0.02) / 0.02% (class0.05) of measurement value. Measurement
drift: 50ppm.
Fine resolution: 0.00001V
Measuring range: there are five ranges of 30V, 60V, 120V, 240V and 480V; it can be
changed by automatic or manual.
3.1.4 The measuring accuracy of the appearance power and the appearance electric energy:
30V…576V, 5mA … 120A 0.015% (class0.02) / 0.03% (class0.05)
30V…576V, 1mA … 5mA 0.03% (class0.02) / 0.06% (class0.05)
(Referred to apparent power irrespective of measurement type) Apparent
energy measurement drift:
30V…576V, 5mA … 120A 100ppm (class0.02) / 200ppm (class0.05) 30V…576V,
1mA … 5mA 200ppm (class0.02) / 500ppm (class0.05)
Fine resolution: 0.00001VA/ Vah
Page -4-
SY-3102 Standard Meter Operation Manual
3.5 Others
Dimensions: 355mm (W) × 185mm (D) × 145mm (H)
Weight: 7.5 kilograms.
The arrangement of the buttons in the right side of the front panel:
Numeric buttons: can be used as alphabet keys when in the methods of alphabet and pinyin input;
Decimal key;
Number button ‘0’: under the function interface, pressing the button to enter interface for users’
setting.
Menu button: pressing the button to enter the main menu interface of the apparatus.
Range button: pressing the button to enter the interface for setting of measuring range.
Page -6-
SY-3102 Standard Meter Operation Manual
Hold/Left Backspace button: the display data switch the modes of “refresh” and “hold ”when the
button is pressed repeatedly under the interface for function data
display while under the interface for user parameter settings, it
displays symbols or Chinese characters of the previous screen.
Storage/Right backspace button: under the interface for function data measurement, press the
button to store the data on the display interface. Under the
interface for user’s parameter settings, press the button to
display the previous or next screen of symbols or Chinese
characters.
Button for conventional measuring function;
Button for error measuring function;
Button for harmonic measuring function;
Button for phase measuring function;
The buttons for sub-functions choices are composed of buttons to the right of the screen
and the sub-functions is changing with the primary function, which can be denoted by
symbols or Chinese characters. The current function you choose is displayed on the screen
according to Chinese characters on the screen.
The layout of the rear panel: power socket, switch, and several ports for connection. Refer
figure 5.2. The universal serial communication port complies with
RS-232 standard. As for the definition, you can refer figure 5.2.
Page -7-
SY-3102 Standard Meter Operation Manual
5.2.4 Connect voltage and current wires of calibration device to the appliance according to
phase sequence and the output (Fout1, Fout2, Fout3) for power pulse of the appliance to
the pulse input port Fin of calibration device if adjustment for the appliance is needed,
5.2.5 Connect the ground terminal of the appliance’s case to the PE protective wire of the
user power supply system.
Warning: The range of the access signal must comply with the specifications in
the third chapter of the manual.
Page -8-
SY-3102 Standard Meter Operation Manual
.
Figure 6.1
Note: The appliance must enter the work state before measuring current signal, or
it will lead to a warning of power supply.
Page -9-
SY-3102 Standard Meter Operation Manual
Working condition displayed in the screen: the measurement of the active power
including the full wave and the stability of the active power in the 3P4W mode.
Functions of the six menu buttons abbreviated for menu buttons one to menu button
in the right of the figure are as follow:
, , ,
Under the interface for primary power, press the menu button 2 to choose active power
measurement in the different connection-wire mode.
Press the menu button 3 to measure the full-power including the harmonic or the
fundamental power without the harmonic.
Warning: Before storing the relevant data, you must input the parameter in the
user parameter frame “instrument number “. About the specific operation,
you can refer the directions.
Page -10-
SY-3102 Standard Meter Operation Manual
6.2.2 Reactive power measurement
Under the interface for measuring function, press the second menu button to the right of
the screen to switch the interface for reactive measurement (as shown in figure 6.2.2.1).
You can choose “Qn (true reactive power)” or “Q90 (cross-phase reactive power)” in 4-
wire mode; And you can choose the “natural reactive power”, “Q60 (artificial neutral point
of the reactive power)” or “Q90 (two components Q90 reactive power)”in 3-wire mode.
Interface for other relevant function buttons is same as those for active power.
Interface for other relevant function buttons are same as those for active power.
Page -11-
SY-3102 Standard Meter Operation Manual
6.2.4 Power factor measurement
Under the interface for measuring function, press the button to choosing
“power factor” and switches interface to power factor measurement (shown as figure
6.2.4.1).
The display content of the screen:
L1: R-phase code name (the first component); L2: Y-phase code name (the second
component); L3: Y-phase code name (the third component). Active power, reactive
power, apparent power, active power coefficient CosΦ, reactive power coefficient SinΦ,
total active power coefficient CosΦ, total reactive power coefficient SinΦ, total active
power, total reactive power and total apparent power of each phase.
Under the interface for measuring function, press the button 5 to switches
interface for the demand measurement (shown as figure 6.2.6.1) The demand
measurement can measure the demand of active power,
reactive power and apparent power. The measuring mode is divided into slip way and
interval way. In this interface, press the button 3 to switch. And slip mode is default
mode for the first time to enter the demand measurement state.
6.2.5.1 The slip way to measure the demand
For the first time to enter the demand measurement state, press the button 3 to switch
the slip way. And it will measure the demand by the slip way (shown as figure 6.2.6.1).
Page -12-
SY-3102 Standard Meter Operation Manual
Functions of the menu buttons:
Choosing parameters.
Press button 3 to switch to the , and then you can measure the demand by it
Page -13-
SY-3102 Standard Meter Operation Manual
(shown as figure 6.2.5.2).
In this interface, except the button 4 becomes “Delete” to delete the error characters
inputted, the function of other buttons is the same as that of the slip way.
Operating method:
When the button 2 is at the state “Stop”, the background color of the demand cycle
parameter frame is red. After inputting the demand cycle by pressing the
button, press button 2 to switches the state “Start” and the background color of the
demand cycle parameter frame becomes normal color, then it begins to measure the
demand.
Page -14-
SY-3102 Standard Meter Operation Manual
Warning: The data of the demand interface cannot be stored, because the “storage
button” is invalid.
Page -15-
SY-3102 Standard Meter Operation Manual
Press the button to switch among the s of All and Base. In All
mode, the measuring electric quantity includes harmonic;
while In Base mode, it only contains fundamental wave.
Set parameters.
Page -16-
SY-3102 Standard Meter Operation Manual
Steps of the error checking:
1) Setting the parameters for checking.
2) Pressing button 6 to startchecking.
3) When the error value updates, its background color becomes yellow, then it
restores the background color. Current error “Errp”(Errq, Errs, Erru, or Erri)is
marked as E1, and the previous errors are arrayed in frames of “E1” to
“E10”orderly. The screen can display ten- or five-times continuous error values. And
press button 6 to control the checking starts or stops repeatedly. At the state of
“stop”, the display value of the checking error will not change in order to observe
and record (or press the hold button to freeze and display all the data).
4) During the checking process, press button or to measure
the error in the different connection modes.
5) During the checking process, press button 2 to display different power data: active
power, reactive power (four-wire: natural reactive power, Q90 reactive power;
three-wire: natural reactive power, Q60 reactive power, Q90 reactive power),
apparent power.
In the error checking interface, press button 4 ” to enter the interface for
parameters setting (shown as figure 6.3.1).
6.3.1.1 Functions of the menu-buttons:
The pulse input port (Fin1, Fin2, Fin3) corresponding to channels will be shown
on the screen, for example, “Port3 Error Parameter:” indicates that the relevant
parameters on the interface is the port of Fin1 for which checking error. The
relevant parameters are on the bottom of the screen.
Page -17-
SY-3102 Standard Meter Operation Manual
6.3.1.2.5 Verify time and Pulse: Input one of them is ok, “auto” mode is the default mode
for the first access. Press the number button to input time for checking.
Page -18-
SY-3102 Standard Meter Operation Manual
6.3.1.2.6 CT Ratio: You can input variable rate as the meter to be checked approaches
CT. Default 1.
6.3.1.2.7 Test Function: the following parameters are available:
➢ “Total” (total power of three-phase)
➢ “Phase1”: the power or energy of R-phase.
➢ “Phase2”: the power or energy of Y-phase.
➢ “Phase3”: the power or energy of B-phase.
➢ “3 Ports Add”
Inputting method:
, you can adjust single channel, while selecting for adjusting three
channels. Interfaces are shown as figures respectively; Press button 2 to select channels
number, you can select the channel number according to the number of pulse input.
Enter parameters of the related channel that you’re using in single channel mode, while
you should input parameters of three channels in tri- channels mode. The cursor will
the related parameters button. In the frame of parameters, press number button to
input new value of parameters and press button 4 to cancel the wrong value of
input parameters.
Warning: If the count of adjusting pulses is too few (the interval for error computing
is small than 0.5 sec), the following indication will appear. Several seconds
later, the system will enter interface for parameters input automatically,
just set the value bigger. If time is set too short, that is the meter to be adjust
can’t generate a pulse during the time, the appliance then cannot compute
the error of this meter. You only need set the adjustment time longer under
this circumstance.
Page -19-
SY-3102 Standard Meter Operation Manual
Figure 6.3.1.2
In the interface for the error adjustment, pressing the button to enter the
interface for setting error parameters (shown as figure 6.3.1.2). Press button
to switch “One Port” and press button 2 to select a channel according to number of
pulse input, the type of power can be selected among items of “power, voltage
and current”. The specific type of power to be adjusted is determined by the button
2 in the interface. After inputting the related parameters of a channel, press button
to enter the interface for single channel error adjustment. As shown in figure
6.3.1.
Content shown on the display:
Data of components to be adjusted on the current point:
6.3.2.1 Voltage:
6.3.2.1.1 4-wire: Ua, Ub, Uc (phase to neutral, select active, reactive or apparent
power for pressing button 2); (or Ubc, Uca, Uab, phase to phase)
6.3.2.1.2 3-wire: Uab, ----, Ucb (phase to phase, press button 2 for selecting active,
reactive or apparent power). (or Uc, ---, Ua, phase to neutral, press
button 2 for selecting Q60 reactive power; or Ubc, ---, Uab, line
voltage, press button 2 for selecting Q90 reactive power)
6.3.2.2 Current: Ia, Ib, Ic (Ib don’t display in three-wire, display “---”)
6.3.2.1 Power: active power, reactive power and apparent power.
6.3.2.2 Power factor: display CosΦ in active power mode while SinΦ in reactive power
mode, and show kinds of load: L indicates hunting, C indicates advance.
6.3.2.3 Total power: active power ΣP; reactive power ΣQ; apparent power ΣS.
6.3.2.4 Real time error of meters to be adjusted: Errp for active power and reactive
power for Errq; Errs for apparent power; Erru for voltage and Erri for current.
Page -20-
SY-3102 Standard Meter Operation Manual
6.3.2.5 Unit of error: display “%” or “ppm” according to setting of appliances.
6.3.2.6 Estimated value of standard deviation for meter to be adjusted: S.
6.3.2.7 For computing 10 times (or 5 times) sample value E1 - E10 of S.
6.3.2.8 Average value of 10 times (or 5 times) of error: Eav.
6.3.2.9 Constant of electric meter to be adjusted: C1.
6.3.2.10 Press “hold” button to neglect all data to be shown and press the button once
more to refresh and show all data as adjusting error. Press “store” button to
record the current data of screen.
6.3.3 Adjustment of three channels
Interface for setting is shown as figure 6.3.1.1 when you need to adjust three
Page -21-
SY-3102 Standard Meter Operation Manual
6.4 Measurement of the adjusted meter’s pulse constant
Under the error measurement interface, press button to enter the
measurement interface, as shown in the figure of 6.4.
Press button 1 to select the appropriate connection mode according to
types of pulse constant (active, reactive and apparent power, voltage, current). Press
button 2 to select the type of power (active, reactive and apparent power). Press button
3 to select a channel according to the channel that the adjusted pulse connected. Press
button 4 to select types of constants to be adjusted (constant of active power, constant
of reactive power, constant of apparent power, constant of current and constant of
voltage). And then press button to start measuring. You can enter
the interface of power accumulation. Constants of the adjusted power meter are
measured repeatedly and the measurement time varied according to the pulse
frequency to be measured. Measurement interval is about 1 second when the
frequency is higher than 1HZ, while interval is about period of the pulse when the
frequency is lower than 1HZ.
The adjusted value only can be the reference of the actual value for the sake of error.
You can get the correct pulse value after modifying.
Warning: You should wait for a while when you start to measure before the
constants is displayed.
Page -22-
SY-3102 Standard Meter Operation Manual
control pulse signal. Menu button 6 is in “Stop” state when enter the interface,
and the amount of power accumulation is 0.
Settings for”Pulse” is needed before accumulation. A method can be: press button
to show a settings frame for the amount of control pulse when system enter the
interface, as shown in the figure, and press button 6 to move the cursor into the input
frame for pulse amout. Press button 2 to select channel according to input channel
Press button to return interface of electric power accumulation when you have
set all. Electric power accumulation starts and the system begin to count the control
pulse. System begins to accumulate electric power and to count control pulse. It will
stop when the counts are overflow or button 6 is pressed, and accumulation
stops. Accumulation of electric power repeats if you press button 6 repeatedly. Press
button 1 to change the connection mode (four-wire or three-wire) and press button 2
to select types of electric power. Press button 3 to choose full wave or base wave that
contain in the measuring data.
Warning: data under the interface of electric power cannot be stored , because the
storage button is invalid
Page -23-
SY-3102 Standard Meter Operation Manual
6.6 Measurement of phase (vector)
Press functional button “phase” in front of the panel to enter the interface for
measurement of vector and phase. Press button 1 to select “four-wire” mode and as
is shown in figure 6.6.1.1.
Page -24-
SY-3102 Standard Meter Operation Manual
Data displayed on the screen (4-wire-three-phase):
Vector graphics of voltage and current display on top left of the screen and the left area
Data displayed: values of three-phase voltage and current that you are Measuring
shown
in figure 6.6.3); degree of unbalanced Ness of voltage and (as shown in figure
Page -25-
SY-3102 Standard Meter Operation Manual
Press button 2 to select degree of asymmetry measurement of phase voltage when the
appliance works in four-wire connection mode (as shown in figures 6.6.2, 6.3.3 and 6.3,4
when “active” is selected) or degree of asymmetry measurement of line voltage (as
shown in figures 6.6.5 and 6.6.6 when “Q90 reactive” is selected), and values of voltage
changed correspondingly.
Press button 3 to select a reference for vector (U1 or I1). Press button 4 to measure
different kinds of phase angle: power factor angle ΦUI (as shown in the figure), phase
angle of voltages, phase angle of currents. Phase angle is displayed vary from 0 degree
to 360 degree when button is pressed to select “0 - 360°”; while vary from -180
Page -26-
SY-3102 Standard Meter Operation Manual
Page -27-
SY-3102 Standard Meter Operation Manual
Page -28-
SY-3102 Standard Meter Operation Manual
6.7 Harmonic analysis
Press the button to enter the interface for harmonic analysis, as shown in figure
6.7.
Page -29-
SY-3102 Standard Meter Operation Manual
In the interface, the bottom and top area of graphics display show the current values of
voltage and current (positive peak-to-peak value, negativepeak-to-peak value)
In the interface:
);
);
);
);
);
);
I1: indicating phase current of R-phase;
I2: indicating phase current of Y-phase;
I3: indicating phase current of Y-phase;
Page -30-
SY-3102 Standard Meter Operation Manual
: Press the button once to restart stability measurement, and the sampling
calculation begins from 0.
Data type of stability:
Power: γp1: The active power stability of the R-phase
γq1 : R-phase active power;
γs1 : R-phase apparent power;
γp2 : Active power stability of the Y-phase
γq2 : Y-phase active power;
γs2 : Y-phase apparent power;
γp3 : Active power stability of the Y-phase
γq3 : Y-phase active power;
γs3 : Y-phase apparent power;
Voltage: γu1: The voltage stability of the R-phase
γu2: The voltage stability of the Y-phase
γu3: The voltage stability of the Y-phase
Current: γi1: The current stability of the R-phase
γi2: The current stability of the Y-phase
γi3: The current stability of the Y-phase
Total power: ΣP: Total active power
ΣQ: Total reactive power
ΣS: Total apparent power
Stability of the total power: γp: Stability of the total active power
γq: Stability of the total reactive power
γs: Stability of the total apparent power
Note: stability calibration algorithm is able to be set in system parameters 6.11.2.
Page -32-
SY-3102 Standard Meter Operation Manual
Measuring range switching has two modes: manual and auto. Measuring range of each
phase needn’t be set in “Auto” mode, while settings for measuring range of each phase
are needed in “Man” mode. Measuring range for current needn’t to set in “Auto” mode,
while range can be set respectively in “manual” mode. The appliance can select a range
according to signal amplitude of each phase and make it in a most appropriate range.
Press button on the front panel to enter interface for Measuring range selection,
as shown in figure 6.9. Press button 1 to select settings for measuring range of voltage
and current. Auto mode is default mode. Items with red background in the parameters
frame are the current items that can be selected. You cannot choose items with black
background in the parameters frame in “Auto” mode, while the background will be white
in “Man” mode.
Functions of the menu buttons:
.
Select channel of voltage (U) or current (I).
After setting the range well, press button 5 to return the previous function interface.
Page -33-
SY-3102 Standard Meter Operation Manual
Page -34-
SY-3102 Standard Meter Operation Manual
6.10 Main menu
Press the menu in any interface to enter the menu interface (shown as figure
6.10). In the interface, press number buttons under the pictures to enter the function
interface.
Page -35-
SY-3102 Standard Meter Operation Manual
6.11 Parameters settings for the appliance
Under the main menu interface, press number key to enter interface for settings.
The system enters the interface for output constant settings first (Shown as figure 6.11).
pressing the button 1 to choose “constant”, the system will enter the
interface of output constant settings. As shown in the figure, the output pulse’s relevant
parameter of the three channels can be set respectively.
choosing the pulse port (Fout1, Fout2, Fout3) that needs to set.
Under the interface, settings for relevant items of channel 1(Fout1) at the top of the screen;
channel 2 (Fout2) is in the middle and channel 3 (Fout3) is at the bottom.
Significance of each parameter:
Constant type: corresponding to the constant type of the output channel: active power,
reactive power, apparent power, voltage and current (the constant); and the default is the
“active power”.
Function of the channel: corresponding to the function of constants of output channel.
⚫ “Total”: representing the total power or the total energy (active power, reactive
power and apparent power).
⚫ “Phase1”: Active power, reactive power, apparent power, voltage or current of R-
phase of the appliance.
⚫ “Phase2”: Active power, reactive power, apparent power, voltage or current of Y-
phase of the appliance.
⚫ “Phase3”: Active power, reactive power, apparent power, voltage or current of Y-
phase of the appliance.
The default is “Total”.
Note: When the three channels are set as “Phase1”, “Phase2”, ”Phase3”
respectively, the appliance can be used as three single-phase meters. When
constant type is voltage or current, the function of channel “Total” is invalid
Constant mode: the automatic constant or manual constant. the default is “auto”, the
channel constant input frame becomes black automatically to indicate input is not allowed.
Channel constant: when the constant mode is “manual”, you can input the input frame a
constant you want; if constant you input exceed the range, there will be some cue on the
screen. The ranges of constant are as follow:
Total active (reactive, apparent) constant: 0<constant≤250000.
Page -37-
SY-3102 Standard Meter Operation Manual
Phase (sigle-phase) active (reactive, apparent) constant: 0<constant≤750000.
Voltage (sigle-phase) constant: 0<constan≤7500000.
Current (sigle-phase) constant: 0<constan≤36000000.
For instance: Channel 2 output constant is set as follows:
Constant type: reactive:
Channel function: “Phase1”.
Constant mode: “Man”
Channel constant: “25000”
The meaning is that the pulse constant of pulse output Fout2 is reactive constant of R-phase
of the appliance (single phase). Constant is”25000”.
In auto mode, the appliance works out each output Fout constant according to measuring
range of voltage U and current I automatically (see the table)
Three-phase total active, reactive power, apparent power (energy) constant table:
Page -38-
SY-3102 Standard Meter Operation Manual
Single-phase active, reactive power, apparent power (energy) constant table:
Voltage constant:
Current constant:
Fmax: 72kHz
Page -39-
SY-3102 Standard Meter Operation Manual
5.10.2. Settings system parameters
choosing the pulse port (Fout1, Fout2, Fout3) that needs to set.
Page -40-
SY3102a Standard Meter Operation Manual
⚫ Beeper: turn on or turn off the Beeper. Default is turn on
⚫ Error Time: when you check error and the number of pulses is in auto mode, the time
basis of calculating pulse can choose among”3,5,8,10” second. Default is “3” second.
⚫ S Number: when you check error, it is used to calculate the count of sampling
estimated value. Chosen between 10 and 5 and default is “10.”
⚫ Baud: when you choose serial asynchronous communication, it is used to set
communications baud rate of the appliance. Select among “2400, 4800,
9600,19200,28800,38400, 57600,115200 “. Default is “9600”.
⚫ Refresh time: integration time of the measuring data on the screen and can be set 1
– 99 seconds. Default is “1” second.
⚫ Vector Point: Set a starting position for drawing vector diagram: 12:00 o’clock or 3:00
o’clock
⚫ Vector Display: length mode of vector in vector graphs drawing mode, the vector length
varies according to the signal amplitude in “proportion” mode; the vector length is fixed
in “fixed -length” mode.
⚫ Vector Rotation: vector rotation direction in vector graphs drawing mode and can be
set as “Clockwise” or “Anticlockwise”
⚫ Error unit: the unit of error and can be set as “%” or “ppm” (1ppm=1×106)
⚫ Stability algorithm: The available stability algorithms: DL/T460-2005, JJG597-
2005B1, GB/T11150-2001, JJG597-2005B2, mean square root method A, mean
square root method B.
SAMPLING INTERVAL: the time interval for each sampling data can be set by number
buttons for stability calculating.
SAMPLING FREQUENCY: the sampling of the data frequency (data number) can be set
through number buttons for stability calculating.
Page -41-
SY3102a Standard Meter Operation Manual
Press the menu button to enter the interface for system time settings of the
appliance. As shown in Figure 6.11.3. System time can be set now.
Press button to move the cursor to the input frame for date or time, press button 4
to delete the current date or time. After finishing input, press button 5 to confirm,
press other function button to exit the interface.
Page -42-
HONGKONG SONGYANG INDUSTRIAL LIMITED
HONGKONG OFFICE:
Add: Room 1507, 15/F., Emperor Group Centre, 288 Hennessy Road, Wanchai, Hong
Kong.
Tel: 00852 - 21218691
Fax: 00852 - 21218692
CHINA OFFICE:
Add: 8F, SUNSHINE PLAZA, No.186 HUSHU NAN ROAD, HANGZHOU, CHINA
Tel: 0086 - 571 - 85394178, 85393435, 85391965,
85301329
Fax: 0086 - 571 - 85301324
MANUFACTURE:
HANGZHOU SONGYANG ELECTRONIC TECHNOLOGY CO., LTD.
Add: Building 4, 6 Xiyuan 8th Road, Sandun Xihu Technology and Economic Park, Hangzhou