UNICORN start

Upgrade Instructions

Step Action
IMPORTANT After the system has been configured, the following message
UNICORN™ start 1.3 is only supported on appears …the application will close now.
Windows® 10.
Upgrading to UNICORN start 1.3 requires a new 7 Click OK and then restart the computer.
license file. Please contact Cytiva service Result:
representatives to get the UNICORN start 1.3 The upgrade to UNICORN start 1.3 is completed.
license.
8 After the computer has restarted, from the desktop, locate and
run the Update Stored Procedure.exe to update the
Archive & Backup stored procedures.
UNICORN start 1.1/1.2 automatic upgrade
Step Action UNICORN start 1.0 manual upgrade
1 In the Administration module of UNICORN start 1.2, click
Step Action
Database Management →Backup Now.
1 Follow steps 1 and 2 described in UNICORN start 1.1/1.2
2 Copy the .bak file and .txt files from C:\Program automatic upgrade.
Files (x86)\GE Healthcare\UNICORN\UNICORN
1.2\Backup to an external folder. 2 Install Windows 10.
Note: 3 Follow steps 3 to 5 in UNICORN start 1.1/1.2 automatic
The UNICORN start 1.3 upgrade creates a backup of the upgrade.
database and restores it after the installation. If necessary, the 4 Locate the backup created in step 1, and copy the .bak file
manually created backup file can be useful.
and associated .txt files to C:\Program Files
3 Start the UNICORN start 1.3 installation: (x86)\Cytiva\UNICORN\UNICORN 1.3\Backup.
• Either download UNICORN start 1.3 from the ESD portal. 5 Go to the Restore tab in Database Management, select the
Log in with your activation ID or username and password. UNICORN start 1.2 backup and click Restore.
• Or insert the UNICORN start 1.3 installation DVD into the 6 Follow the instructions on the screen.
DVD drive. Result:
4 Follow the instructions on the screen. After the database has been restored, the following message
If the message Could not delete services appears, appears …the application will close now.
run the installer again in Repair mode. 7 Click OK and then restart the computer.
Result: Result:
When the installation is completed, the UNICORN start 1.3 The upgrade to UNICORN start 1.3 is completed.
icon appears on the desktop.
8 After the computer has restarted, from the desktop, locate and
5 Double-click the UNICORN start 1.3 icon. run the Update Stored Procedure.exe to update the
Result: Archive & Backup stored procedures.
The following message appears:

You have upgraded from a previous


version of UNICORN start. The system
needs to be configured. This may take
several minutes.
6 Click OK.
Result:

cytiva.com 29688976 AA
cytiva.com/unicornstart
Cytiva and the Drop logo are trademarks of Global Life Sciences IP Holdco LLC or an
affiliate.
ÄKTA and UNICORN are trademarks of Global Life Sciences Solutions USA LLC or an
affiliate doing business as Cytiva.
Windows is a registered trademark of Microsoft Corporation.
All other third-party trademarks are the property of their respective owners.
© 2021 Cytiva
UNICORN © 2021 Cytiva
Any use of UNICORN is subject to Cytiva Standard Software End-User License
Agreement for Life Sciences Software Products. A copy of this Standard Software
End-User License Agreement is available on request.
All goods and services are sold subject to the terms and conditions of sale of the
supplying company operating within the Cytiva business. A copy of those terms and
conditions is available on request. Contact your local Cytiva representative for the
most current information.
For local office contact information, visit cytiva.com/contact
29688976 AA V:1 07/2021

You might also like