Service Manual of Video Flexible Laryngoscope

Product model: iS3-F


Release Date: August, 2021 Ver2.0
Copyright © 2019-2020 Insighters Medical. All rights reserved.

Video Flexible Laryngoscope Service Manual iS3-F (V2.0)


Copyright
Statement
SHENZHEN INSIGHTERS MEDICAL TECHNOLOGY CO.,LTD (hereinafter referred to as
Insighters) holds the intellectual property rights in this manual and retains the contents of this
manual as confidential information. This manual is a reference for the operation, maintenance and
cleaning of this product and does not convey any license according to Insighters's patent right or
the rights of others.
This manual contains copyright or patent protection information. Copying, altering or translating this
manual in any way without Insighters's written permission is terminated.
All information contained in this publication is believed to be correct. Insighters shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this material. This publication may refer to information and
protected by copyrights or patents and does not convey any license under the patent rights of
Insighters, nor the rights of others. Insighters does not assume any liability arising out of any
infringements of patents or other rights of third parties.
This manual is based on the maximum configuration, so some contents may not be applicable to
your product. This manual is subject to change without prior notce and legal obligation.

Manufacturer's Responsibility
Insighters will only be responsible for the impact on the safety, reliability and performance of this
product if all the following requirements are met.
 All repairs, installation operations, extensions, changes and modifications of this product are
completed by Insighters authorized personnel.
 The use or application of products or the use of parts or accessories is approved by Insighters.
 The electrical installation in the relevant environment meets the applicable national and local
requirements.
 Use the product according to the instructions.

Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED
OR IMPLIED, INCLUDING WARRANTIES OF MERCHANT ABILITY OR FITNESS FOR ANY
PARTICULAR PURPOSE.
Warranty Period
Party A provides warranty period as described chart below for sold products and Party A offers free
of charge repair service within the warranty period except for the replacement of disposable parts.
The commencement of warranty period is determined by either the installation date on warranty
card or manufacturing date.
Product Name Warranty Period
Insight iS3-F Video Flexible Laryngoscope 1 Year
Insight iS3 Video Laryngoscope 3.5-inch Host 3 Years

Video Flexible Laryngoscope Service Manual iS3-F (V2.0)


Warranty Exemptions
Insighters 's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the improper
use or application of the product or the substitution upon it of parts or accessories not approved by
Insighters or repaired by anyone other than a Insighters authorized representative.

This warranty shall not extend to any instrument which has been subjected to misuse, negligence
or accident; any instrument from which Insighters's original serial number tag or product
identification markings have been altered or removed, or any product of any other manufacturer.

It’ an obligation for Exclusive Distributor to assist Insighters Medical in case of product recall and
after-sales service.

Return Policy
1. If repairs become necessary, for other than damages incurred during initial shipment, the
customer must follow the repair procedure.
2. Warranty repairs will be made without charge. All other repairs are subject to Insighters’
applicable standard repair or exchange charges.
3. If requested, customer will be advised of the estimated cost of the repair work or a product
exchange before it is undertaken.
4. All repairs carry a 90 days warranty.
5. Exchange products carry the applicable Insighters product warranty.
6. If the damaged product is not returned within thirty (30) days of receipt of the replacement
product, customer will be invoiced for the full list price of the replacement.
7. Insighters reserves the right to refuse or return any product sent back without prior
authorization of Insighters Customer Support Department.
Other Terms
1. It is compulsory to provide customer information, troubleshooting description and address to
send back for after-sales repair. The freight cost should be shared by both parties under
warranty period.
2. No matter what causes device damage, please stop usage and contact Party A for repair.
3. Insighters Medical reserves the right of final interpretation.

Contact Information
Manufacturer: Shenzhen Insighters Medical Technology Co., Ltd
Address: The 13th floor of Hengtemei Building, Ganli Road No. 3, 518000 Shenzhen, Guangdong,
PEOPLE’S REPUBLIC OF CHINA
Tel: +86-755-28218609
Fax: +86-755-84509803
E-mail: intl-market@insighters.cn
www.insighters.cn

Video Flexible Laryngoscope Service Manual iS3-F (V2.0)


Contents
1 Safety......................................................................................................................................... 1
1.1 Meaning of Signal Words.............................................. ..........................................................1
1.2 Meaning of Safety Symbols....................................................................................................1
1.3 Safety Precautions...................................................................................................................1
1.3.1 Electrical Safety............................................................................................................ 1
1.3.2 Parts Safety...................................................................................................................3
1.3.3 Biological Risk Considerations................................................................................... 3
1.3.4 Security Symbol............................................................................................................3
2 Specifications............................................................................................................................ 4
3 Checking Procedure.................................................................................................................. 7
3.1 Check Battery................................................................................................................... 7
3.2Check HDMI (Optional)........................................................................................................... 7
3.3 Connect to Host........................................................................................................... 7
3.4 Connect to Blade....................................................................................................................8
3.5 Check Image.................................................................................................................... 8
3.6 Check Record Function.........................................................................................................9
4 Maintenance and Cleaning...................................................................................................10
4.1 Maintenance............................................................................................................................10
4.1.1 Unpacking and Inspection.........................................................................................10
4.1.2 Routine Maintenance.................................................................................................10
4.2 Cleaning for the host.........................................................................................................10
4.2.1 Manual Cleaning Procedure..............................................................................................11
4.3 Disinfection & Sterilization for the Handle................................................................................ 11
4.3.1 Cleaning, Disinfection & Sterilization Methods for the iS3-F Series.............................. 11
4.3.2 Insighters Medical has validated the compatibility of iS3-F series with following AERs......... 12
4.3.3 Sterilization Products for iS3-F Series Handle..............................................................12
4.3.4 Cleaning & Disinfecting procedure the iS3-F Series Handle..................................................13
4.4 Storage and Disposal............................................................................................................ 13
4.4.1 Storage.........................................................................................................................13
4.4.2 Device Disposal.......................................................................................................... 13
7 Troubleshooting...............................................................................................................14

Video Flexible Laryngoscope Service Manual iS3-F (V2.0)


Safety
1.1 Meaning of Signal Words

In this manual, the signal words WARNING, and CAUTION are used regarding safety and other
important instructions. The signal words and their meanings are defined as follows. Please
understand their meanings clearly before reading this manual.

Signal Word Meaning


Indicates a potentially hazardous situation which, if not avoided,
WARNING could result in death, serious injury or other serious adverse reactions
Indicates a potentially hazardous situation which, if not avoided,
CAUTION may result in minor or moderate injury.
Indicates a potentially hazardous situation which, if not avoided,
CAUTION
may result in property damage.

.
1.2 Meaning of Safety Symbols

Symbol Description

Type-BF applied part

"Attention"(Refer to the operation manual.)

1.3 Safety Precautions


Please read and understand all the precautions in this manual before trying to use an laryngoscope.
Regularly review operating procedures and safety precautions.
1.3.1 Electrical safety

WARNING
 Only personnel authorized or trained by the manufacturer can carry out equipment
maintenance. No Team Authorization cannot install or remove equipment.

 The user must receive adequate training and study before use. Official agencies and operators
shall meet the qualification requirements. Otherwise, only medical personnel approved by the
hospital safety administrator or department head can perform the laryngoscope. Only
personnel authorized or trained by the manufacturer can carry out equipment maintenance.
No team authorization cannot install or remove equipment.

 Do not use flammable gasses such as anesthetics, or flammable liquids such as ethanol, near
this product, because there is danger of explosion.

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 1


 Before every usage, ensure the product is operating correctly and has no sign of damage. Do
not use if it appears damaged. Always ensure that alternative airway management methods
and devices are readily and available.

 The insight iS3 video laryngoscope host must be charged with the power adapter supplied by
Insighters, otherwise the device may be damaged.

 The insight iS3 video laryngoscope are delivered non-sterile and require cleaning or
disinfection prior to initial use.

 This product may only be cleaned, disinfected, or sterilized by using the approved methods
provided in this manual. Availability of cleaning, disinfection,and sterilization products varies in
different countries, and Insighters Medical is unable to test products in every market. For more
information, please contact Insighters after-sales team.

 No modification of this equipment is allowed.

 This product and related devices may contain batteries and other environmentally hazardous
materials.

 Electric shock hazard. Do not attempt to open the system components. This may cause
serious injury to the operator or damage to the instrument and will void the warranty.

 Do not use the power adapter in the presence of flammable anesthetics.

 Before the installation and operation, the user should first check the appearance of the
medical device to ensure there are no unintended rough surfaces, sharp edges or protrusions
which may cause HARM.

CAUTION
 Malfunctions due to radio waves:
(1) Use of radio-wave-emitting devices in the proximity of this kind of medical electronic system may
interfere with its operation. Do not bring or use devices which generate radio waves, such as
cellular telephones, transceivers, and radio controlled toys, in the room where the system is
installed.
(2) If a user brings a device which generates radio waves near the system, they must be instructed
to immediately turn OFF the device. This is necessary to ensure the proper operation of the system.

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 2


 Do not allow fluids such as water to contact the system or peripheral devices. Electric shock
may result. The equipment is suitable for use in the environment of professional health
institutions.

 Do not use in home environment or environment directly connected to a low-voltage power


grid that supplies power to residential buildings.

 Video Laryngoscope is a kind of ordinarily sealed device. The user should keep its surface dry
and clean, and prevent any liquid into the internal.

 The user should properly place Video Laryngoscope, preventing it from falling, violent
oscillation, or other damage caused by external mechanical force.

1.3.2 Parts safety


Caution: Only spare parts provided or approved by the manufacturer can be used.Use other spare
parts may cause damage to the equipment and cannot achieve the expected performance
described in this manual.

1.3.3 Biological risk considerations


 Operators should take measures to avoid direct contact with disinfectants or samples of
patients. If there is skin contact, rinse the area with clear water immediately. If the liquid
touches the eyes, rinse with water immediately and seek the help of an ophthalmologist.
 Dispose of disinfectants, detergents and wastes in accordance with local laws or regulations.
Consult your local manufacturer or distributor for more information.

1.3.4 Security symbol

The following table is used to identify important symbols found in labels in laryngoscope.

Symbols Meaning

Date of Manufacture

Serial Number

Serial Number

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 3


2 Specifications
2.1 General
2.1.1 Basics
Product Model iS3-F
Product Name Video Flexible Laryngoscope
Classification Class IIA, Rule 11 (According to MDD 93/42EEC directive)
2.1.2 Safety designations
Safety designations per IEC 60601-1 Standard:
Type of protection against
II (Internally powered equipment)
electric shock
Degree of protection against
Type BF Applied part
electric shock
Degree of EMC Class A
Degree of protection against
Ordinary equipment, not protected
hazards of explosion
Equipment type Hand held
2.2 Display
2.2.1 Display
Host 3.5-inch LCD color touch screen
Resolution of screen 640 x 480
Angle of view 160 degree
Forward 30°/ Backward 90°
Host Tilt & Rotation
Leftward 90º / Rightward 180º
Compatible with insighters video laryngoscope series handle:
Compatibility video laryngoscope iS3-L, video rigid laryngoscope iS3-R,
video flexible laryngoscope iS3-C series & iS3-F series
Micro-USB connection port X 1
Connection port
HDMI connection port X 1
Quick-guide training video.
Manage the photos & videos data.
Software function
Indoor/ outdoor mode of screen brightness adjustment.
Sound volume adjustment.
2.2.2 Buttons
on: short press to power on the host
Power Button (physical button)
Off: press for 2s to power off the host
Short press to take photo. Press 2-3 seconds start to record
Photo/video (touch button)
video, press again to stop recording.
Built-in menu (touch button) Press to enter the built-in menu.Specifications
2.2.3 Power saving features
if no finger is detected within 15 minutes, the laryngoscope automatically shuts down.
2.2.4 Recording
Build-in memory capacity 8 GB (optional, up to 32GB)

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 4


2.3 Handle
2.3.1 Camera
Camera 160,000 pixels camera, apply full digital imaging technology
Field of View ≥90°
Depth of field 3~50mm
Light source ≥ 400 LUX, h=20 mm, 2 LED light source
Deflection angle of insertion
Forward/backward: 150°/150°
tube
Length of insertion tube 600 mm
Aluminium alloy(handle)
Material
Polyurethane (PU)(insertion tube)
Distal end OD(±0.3mm) iS3-F2 2.8 mm
iS3-F3 3.6 mm
2.4 Physical Characteristics
Weight Host: 170g, Handle: 355g
Monitor: 103(Width) x 104(Height) x 20(Depth)
Size
Handle: 840(Length) x 40(Width) x 80(Depth)mm
2.5 Electrical Specifications
2.5.1 Battery
Battery Type Rechargeable build-in lithium battery
Battery capacity 2500 mAH, DC3.7V
Usage time 4 hours
Charging time 4 hours
2.5.2 AC adapter
Input voltage AC 100-240V
Input frequency 50-60Hz
Input current 0.2-0.1A
Out put voltage DC5V
Out put current 1A
fuse T2A, 250V
2.6 Environment and Safety
Temperature
Operation 5℃~40℃
Transportation and
-20℃~60℃
storage
Humidity
Operation ≤80%, non-condensing
Transportation and
≤95%, non-condensing
storage
Atmospheric Pressure
Operation 427.5~806.3mmHg (57.0~107.5kPa)
Transportation and 122.3~806.3mmHg (16.3~107.5kPa)

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 5


storage

2.6 EMC
Guidance and manufacturer’s declaration –electromagnetic emissions
The Video Flexible Laryngoscope is intended for use in the electromagnetic environment
specified below. The customer or the user of the SECP-II should assure that it is used in
such an environment.
Electromagnetic
Emissions test Compliance
environment - guidance
The Video Laryngoscope uses RF energy only
for its internal function. Therefore, its RF
RF emissions
Group 1 emissions are very low and are not likely to
CISPR 11
cause any interference in nearby electronic
equipment.
RF emissions CISPR 11 Class A The Video Laryngoscope is suitable for use in
Harmonic emissions all establishments other than domestic and
Not applicable
IEC 61000-3-2 those directly connected to the public
Voltage fluctuations/ flicker low-voltage power supply network that supplies
Not applicable
emissions IEC 61000-3-3 buildings used for domestic purposes

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 6


3 Checking procedure

3.1 Check Battery


The Video Laryngoscope host includes an internal lithium battery.Insighters Medical recommends
users to charge the battery fully prior to first use.
The system must be operated exclusively on battery power, without a connection to an AC power
source.
Under normal operating conditions, a fully charged battery lasts approximately 240 minutes working
time.

3.1.1 Battery Status Icons


Red: ≤5% battery volume remaining. Battery must be recharged.
Yellow: 5% to 40% battery volume remaining.
Green: ≥40% battery volume remaining.
3.1.2 charge steps
1. Connect the power adapter and host via micro-USB port.
2. Connect the power adapter to proper power supply. Fully charging the battery may
take up to 4 hours.
Note: The system will not operate while charging.

3.2 Check HDMI (Optional)


By using an HDMI cable, users can connect the host to an external monitor that is approved for
medical use for real-time image transmission.
Note: Image quality on the external monitor may vary according to the resolution of the
external monitor.
1. Connect one end of the HDMI cable to the HDMI port on the host.
2. Connect the other end of the cable to the HDMI port on an external monitor approved for medical
use.
Note: The HDMI function support real-time image transmission and video playback.

3.3 Connect to Host


Follow the steps below to install host before use:
1. Hold the iS3-F handle firmly by one hand.
2. Insert the host in the connect port of iS3-F (as shown in Fig 3-3-1).

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 7


Fig 3-3-1

3.4 Perform Functional Check


Before using the device, insighters recommends users to perform following functional check to
ensure that the video laryngoscope is working properly.
1. Turn on the host, ensure the battery volume.
2. Connect the host with handle, check the white LED on distal end of handle whether light up.
3. Check if image from camera are displaying normally and clearly on the screen.
4. Adjust the deflection lever gently, check if the bending angle of distal end working as normal.

Fig 3-4-1
3.5 Check Image
1. Turn on the power. LCD will display start-up image.
2. Connect host with handle, and then the white LED on distal end of handle will light up, host will
display image from camera.

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 8


3.6 Check Record Function
1. When host displays image from camera, short press the Photo/video touch button on screen to
take photo.
2. Press the Photo/video touch button 2-3 seconds to start recording video, there will be prompt
tone if the sound on, and a red dot will flash on the top right corner. Press the button again to finish
recording.
3. Press the built-in menu button to enter built-in menu, press the “photo” or “video” button to
browse and manage the photo/videos taken. (as shown in Fig Fig 3-6-1)

Fig 3-6-1 Built-in Menu

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 9


4 Maintenance/ Cleaning and Disinfection
4.1 Maintenance

4.1.1 Unpacking and inspection


1. Inspect the iS3-F for possible damage during shipment;
2. Check the items according to packing list;
3. Test all the functions working as normal.
If the function have any abnormal, do not operate on the patient, please refer to Troubleshooting or
contact with the Insighters service engineer immediately.

4.1.2 Routine maintenance


1. A thorough examination shall be carried out every 3~6 months or after each maintenance by
qualified personnel, including function and safety test.
2. There shall be no water spilt into the machine. If there is dirt on the surface of the device, it must
be wiped gently with clean cotton dipped in alcohol.
3. If there is quality problem, please contact the company rather than attempt to disassemble the
device.
Battery Maintenance
1. There is a built-in non removable lithium polymer battery in the host. While not using, shutdown
the host to reduce the loss of the battery life.
2. Don’t use the product during charging which may cause unacceptable risk.
3. The users must use the AC adapter configured by Insighters to charge the battery when the
power is low. Avoid using the AC adapter made by other companies.
4. The intend battery lifetime is 5 years and it can be used 4 hours continuously after charge
completely. It means the battery needs changing if the use time shorter than 2 hours after charged.
Please contact the dealer or the company.
5. If Video Flexible Laryngoscope is kept in storage for a long time, the battery should be fully
charged again, and it should be recharged every 3 to 6 months in case of damage to the battery
caused by self-discharge.
6. The battery should be changed only by the manufacture and any action about changement is not
allowed.
7. Don't burn or discard used batteries. Please return the waste battery to the dealer or the
company and the returning is free of charge.Maintenance and Cleaning

4.2 Cleaning for the host


The iS3-F host and handle are considered as non-critical device and normally required a rapid
disinfection.
The equipment should be kept from dust. If the shell or screen needs cleaning, the detergent should
be noncorrosive, such as soap or rinsing etc.
Notes:
1. Do not use strong solvent, such as acetone.
2. Most cleanser need dissolving before use. Keep to the cleanser’s instructions for use.
3. Do not use abrasive materials, such as fine steel wire or silver polishing agent.

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 10


4. Do not let any liquid ingress into the equipment and do not immerge the equipment into
any liquid.
5. Keep the surface of the equipment clean after cleaning.
6.Any port or clearance should be kept at down side to avoid cleaner or water going
inside,results in corrosion or short circuit. Such as USB port, HDMI port, Horn, connector
head, camera head.

Cleaning & Disinfection Methods for the Host


Chemical Concentration Disinfection Level
Enzymatic agent or detergent Clean
Ethyl alcohol 75% Low - Middle
Isopropyl alcohol 70% Low - Middle
Hydrogen peroxide ≤ 3% Low - Middle

Equipment and accessories for cleaning insight iS3-F host


• Soft lint-free cloth
• Detergents (or disinfecting wipes, e.g. 70% Isopropyl alcohol wipes, CaviWipes)
• Clean covered tray/container
• Sterile gauze

4.2.1 Manual cleaning procedure


1. Make sure that the host power is shut off before cleaning for the purpose of safety.
2. Disassemble the blade from the host, dispose the single-use blade* properly.
*If use reusable blade, please refer to User Manual of Reusable Blade to clean, disinfect or sterilize
the reusable blade.
3. Disassemble the host from handle.
4. Use soft lint-free cloth to wipe clean the host and handle thoroughly to remove any visible debris
and soil from all exposed surfaces.
5. Use soft lint-free cloth moistened with detergents or disinfecting wipes to wipe the host and
handle thoroughly. Always follow the detergent manufacturer’s instruction to achieve proper
disinfection.
6. Dry all surfaces with sterile gauze.
7. Store the monitor and handle in aseptic clean pouch for next use.

4.3 Disinfection & Sterilization for the Handle

Following the methods according to Table 10 to reprocessing the video flexible


laryngoscope iS3-F series handle, for detailed process please refer to the Cleaning
and Disinfection Guide of iS3-F series along with the goods.

4.3.1 Cleaning, Disinfection & Sterilization Methods for the iS3-F Series
Table 10
Chemical Concentration Disinfection Level
Enzymatic agent or detergent - Clean

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 11


Isopropyl alcohol 70% Low - Middle
Ethyl alcohol 75% Low - Middle
Hydrogen peroxide ≤7.5% High
Glutaraldehyde ≤3.4% High
Ortho-phthalaldehyde 0.55% High
Peracetic acid 0.2% High/Sterilization
Ethylene Oxide Gas According to the Sterilization
Sterilization manufacturer’s instructions
Low Temperature Plasma According to the Sterilization
Sterilization manufacturer’s instructions

4.3.2 Insighters Medical has validated the compatibility of iS3-F series with following AERs:

Product Disinfectant Cycles Conditions


Medivators Rapicide PA 100 Concentration: 750–950 parts per million
Advantage Plus Exposure: Process the component for 5
minutes at 30°C (86°F) with the following
configuration:
Hookup: 2-8-210HAN or 2-8-245 or 2-8-002
Medivators Disinfectant that is 100 According to the manufacturer’s instructions
DSD/SSD approved from Table 10 Hookup: 2-8-210HAN or 2-8-245 or 2-8-002
series and compatible with the
Medivators system
(CIDEX, CIDEX OPA, or
Rapicide)
STERIS S40 or S20 100 Standard cycle of STERIS System IE
System IE
STERIS Reliance Dry Chemistry 100 Standard cycle of STERIS
Reliance Reliance EPS
EPS

4.3.3 Sterilization Products for iS3-F Series Handle:

Product Level Cycles Conditions


STERIS V-PRO Sterilization 600 Use the Flexible cycle
systems in following STERIS
V-PRO
low-temperature
sterilization system:
V-PRO maX
V-PRO maX 2
V-PRO 60
V-PRO s2

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 12


4.3.4 Cleaning & Disinfecting procedure the iS3-F Series Handle

1.Package the handle in a pouch.


2.Sterilize the handle, follow the manufacturer's instructions for the sterilization system.
3.Store the handle in an environment that is suitable for sterile equipment.
Insighters Medical has validated the products in below table for both chemical compatibility and
biological efficacy when sterilizing the indicated component(s) as instructed in the Conditions
column:

4.4 Storage and disposal

4.4.1 Storage
1. Make sure the host is turned off and the AC adapter is disconnected.
2. Avoid direct sunlight, high temperature, and acid erosion. Store it in a dry, clean place.
3. Avoid storing it in chemicals or at explosible gases.
4. The flexible tube part can’t be twisted when in storage, or it maybe get damaged.

4.4.2 Device disposal


To avoid environmental contamination or infections to people or other equipment, the disposal of
Video Flexible Laryngoscope should be in accordance with relevant provisions of the state on the
purification and disinfection of medical devices.
As for the disposal of the blade, please follow relevant provisions of local hospitals.Troubleshooting

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 13


5 Troubleshooting
Trouble Cause Solution
• Restart the host or disconnect the
handle and then connect it again;
Software bug
• Press RESET with a pin;
• Upgrade software.
The host doesn’t show
Send back to factory to replace
image from camera Camera lens is broken
new lens.
Circuit damage caused by
rotation the shaft of host Send back to factory for repair.
beyond design degree.
The LED light is not connected
LED beads do not Send back to factory for repair.
properly
light up
LED break Send back to factory for repair
Use lint-free cloth to swipe the lens
Dirt cover camera lens with alcohol. (Don 't wet the cloth too
much to avoid water penetration.)
Blur image Camera lens is broken Send back to factory for repair.
The handle is immersed during
disinfection, there's liquid in the Send back to factory for repair.
handle.
The host isn’t turned off during Turn off the host and then recharge
battery charging. battery.
Change to new AC adaptor and
AC adaptor issue
then recharge.
Battery can’t charge
USB port break/ loose/short Send back to factory for repair or
circuit replacement.
Send back to factory for repair or
Main board break/short circuit
replacement.
There’ re blind spots
Send back to factory for repair or
or white and black Bad display quality
replacement.
strips on display
The detach button is Send back to factory for repair or
Bad detach button quality
broken. replacement.
Can’t control rotate
The shaft is loose Send back to factory for repair
position of host
No battery by long time use or
Charge battery
the host can’t not storage
power on Circuit self-protection in Press power button 5 times quickly
sleep-model during power charging to active
Can’t pump up pressure Check the ETO connector/ air pipe/
Air leakage test failure because Tester leaks air airbag leak
Handle leaks air Send back to factory to repair

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 14


The main hose breaks

The bending rubber breaks


The camera assemble out of
shape/ break
The bending part
The wire rope broken one side/
Can’t bending one Send back to factory to repair
both sides
side/ both sides
The position limited screws are
Bending angle not Send back to factory to adjust
loose
enough
The wire rope is loose Send back to factory to adjust
Any appearance
/ Send back to factory to repair
damage or break

Video Flexible Laryngoscope Service Manual iS3-F (V2.0) 15