You are on page 1of 2

Cold Therapy Protocol*

Treatment Awake/ Inspect Skin


Frequency/Duration
Period Asleep Every:
POLAR CARE CUBE ® TM

Day:
Awake

OPERATING INSTRUCTIONS Through


Asleep
Day:

WARNING Day: Awake


The Polar Care Cube can be cold enough to cause serious injury, including Through
full skin necrosis. Follow these Operating Instructions, and carefully read the
Day: Asleep
Product Insert (see pouch on side of unit) and the Cold Therapy Pad Fitting
Instructions (provided with each Cold Therapy Pad) prior to use.
Day: Awake
Through

Day: Asleep

* To be completed by a licensed Health Care Professional


1 Discuss Treatment with your Licensed Health Care Practitioner
Provide a complete medical history including any reactions to cold. Certain medical conditions make cold-induced injury more likely. Ask your
practitioner about potential adverse reactions and cold-induced injuries.

2 Use Only as Prescribed


Use only according to your practitioner’s instructions regarding the frequency and duration of cold application and length of breaks between uses,
how and when to inspect the skin, and total length of treatment. Do not use this device if you did not receive or do not understand the instructions.
Unless your practitioner provides different instructions, to take a break between uses simply disconnect the power from the unit or remove the pad
from your body for a minimum of 30 minutes. Federal law restricts this device to sale by or on the order of a licensed health care practitioner.

3 Apply Insulation Barrier & Cold Therapy Pad


Do not let any part of the pad touch skin. Always use an insulation barrier (such as Breg Polar Dressing, Webril™, Kerlix™, cast padding or elastic
bandage) between the Cold Therapy Pad and skin. If a sterile dressing has been applied to the treatment site that does not completely cover the
skin under the pad, use an additional insulation barrier. Use only with the Breg Cold Therapy Pads. Other pads may be colder, increasing the risk
of serious cold-induced injury, including full thickness necrosis.

4 Check for Moisture


Check for moisture on the barrier between your skin and the cold pad. Discontinue use if the barrier is moist. Change to a dry skin barrier before
resuming use.

5 Inspect Skin Regularly


Inspect the skin under the Cold Therapy Pad (by lifting the edge) as prescribed, typically every 1 to 2 hours. Ask your practitioner to instruct you
on how to inspect the skin area which is being treated by the device. Do not use the Polar Care unit if dressing, wrapping, bracing, or casting
over the Cold Therapy Pad prevents skin checks. Stop using and contact your practitioner immediately if you experience any adverse reactions,
such as: increased pain, burning, increased swelling, itching, blisters, increased redness, discoloration, welts, other changes in skin appearance, or
any other reaction identified by your practitioner.

Set Up Operation See Product Insert for


FIRST ADD WATER TO electrical precautions.
FILL LINE, THEN ICE
Replace ice when existing
ice is almost melted

Check Skin Frequently

Coupler Connections

Connect

End Session
To stop the pump, disconnect electrical power.
Remove pad.

Over
Page 1 of 2
POLAR CARE CUBE Tips and Trouble Shooting

Usage Tips
1. Use cubed or chunked ice for optimal performance.
OCOL 2. It is recommended to drain the Cold Therapy Pad between uses. To drain the pad, hold the Cold Therapy Pad upright with the
hose pointed toward the ground. Depress the black plunger and allow water to drain out of the pad.
Inspect
Skin 3. You may disconnect the Cold Therapy Pad from the unit without removing the pad from the affected area by depressing the silver
Every: ELECTRICAL
tabs SAFETY
on the hose coupling and gently pulling the hoses apart. The BregTROUBLESHOOTING
Polar Care Cold Therapy Pad GUIDE and unit will seal itself and PROD
Like all electrical products, you must follow
will not leak. Note: Some dripping during release is normal. precautions to Pump not running
Symbo
avoid electrocution, fire, burns, or other injuries. Supervise Water not flowing to the pad
4. DO NOT RUN
children and PUMP
invalidsWITHOUT
near this WATER!
productThe pumpthat
to ensure in this unit is designed to run with water. Running the unit without water will
these Pad is not getting cold Symb
cause permanent
precautions damage to the pump.
are followed. 1. Make sure the transformer is plugged into the wall outlet.
5. Unplug unit beforeconnections
Keep electrical removing lid.dry 2. Confirm that the ice and water are filled to the indicated
Even though the Polar Care unit is designed to hold and pump levels.
water, the electricalGuide
Troubleshooting connections at the power outlet and from 3. Make sure there are no kinks in the pad.
the power cord to the Polar Care unit are not waterproof and 4. Gently pull on the blue tube to make sure the tube/pad
Problem:
must be Pump keptnotdry.running, water not flowing to pad, or pad not cold. junction is straight.
Possible solutions: the transformer or electrical cord with wet
• Do not handle 5. Remove the pad from the treatment
hands.
•• Use larger ice for optimal performance. site and fill it while the pad is lying
ACIER™ • Always keep the unit in a place where the connections will flat; then reapply.
AK® •• Allow not10fallminutes for (e.g.,
into water flow aandtub,pressure to stabilize.
sink, etc.). 6. Confirm that the pad couplings are
E™ •• Ensure
• Do not power use outlet is working and plugs are fully engaged.
outdoors. securely attached to the unit.
•• Ensure unit has both ice and water. Listen for the double clicks
If the electrical connections fall into water, do not touch
•• Ensure
any wet hosepartisofnot
thekinked.
product. Unplug only at dry electrical
(Figure 1) to confirm the pad Physic
couplings are securely attached. (fig. 1)
connections.
•• Disconnect and reconnect the pad and unit. 7. Remove cap, which contains Unit R
•• Release air by depressing the black plus-shaped part inside the unit connector. the foam filter, from the Electr
Avoid
Note: flammables
water and oxidizers
may be released. bottom of the pump (Figure 2)
Do not use in places with flammable vapors or gasses (for Ma
and clean out any debris.
ELECTRICAL example,
• • SAFETY
Place unit flammable anesthetics), high oxygen concentrations,TROUBLESHOOTING
on a table or other raised surface.
Replace cap ensuring the GUIDE PRO Mo
Rat
Like all electrical
•• products,
Decrease you must
tension
or other oxidizers of follow
bandages
(for example, precautions
or strapsoxide).
nitrous to the pad.
around Pump not foamrunning
filter is in place.
Sym
avoid electrocution,
•• Remove
Use fire, burns,
pad andorlay
a grounded other
it flat.injuries.
outlet Allow pad Supervise
to fill; reapply.
Grounded Water not flowing to the pad (fig. 2) Ma
children and invalids near this product to ensure that (Fig. 1) Mo
•• Clean
Connectfilter:this
Disconnect
product topad. unit these
Removegrounded
a properly filter cap fromPad
lid. PullOutlet is not
bottom of lid.getting
Remove foam coldfilter. Rinse Sym
precautions are followed. Rat
filter
outletcaponly.
and filter to remove clogs. Reassemble (Fig. 1). 1. Make sure the transformer is plugged into the wall outlet.
Keep electrical connections dry cord and transformer
Protect the power 2. Confirm that the ice and water are filled to the indicated
Leaks
& Problem:
Even though the Polar Condensation
• Keep theunit
Care cord is
and transformer
designed away and
to hold from pump
heated surfaces. levels.Note: Some condensation on the lines, controller, and pads is
Enviro
Tem
water, the electrical • connections
Possible Set up and useat
solutions: this equipment
the power outletin a low andtraffic location away 3. Make sure
from unavoidable,
there are especially
no kinks in warmer climates.
in the pad.
• • Wrapfrom childrenover
material andpadpets.and hose to minimize air exposure. 1. If using a Breg Polar Care Cold Therapy unit tube/pad
with a detach-
the power cord to Never the Polar Care unit are not waterproof and 4. Gently pull on the blue tube to make sure the Hu
operate this product if it has a damaged cord, able Cold Therapy Pad, and a leak exists (other than conden-
• • Protect the wound site by using a sterile
must be kept dry. transformer, or plug. If the cord, transformer, or plug is dressing with junction is straight.
waterproof barrier.
sation on the lines) disconnect the pad couplings. Make sure
• Do not handle the transformer
damaged, unplugor andelectrical cordCustomer
contact Breg with wetCare, 5. Remove the thesilver
pad from
tabs onthethetreatment
couplings are depressed before
etach- Problem: Unit is leaking Atm
hands. (800) 321-0607. site and fill it while the
reconnecting thepad
padis tolying
the pump hose; then confirm both
•between
Always keepPossible
theStart
unit in a
solutions:place
the product where the connections will flat; then reapply.
sides of the coupling are properly clicked in. Stand
Therapy
nd.
not fall into •water
Use (e.g.,
• Disconnect
only a tub,
the unitsink,
transformeretc.).
connector. Ensurewith
provided metal
the tabs
Breg are pressed
Polar Care down;
6. 2. that
reconnect
Confirm If the coupling
(Fig.
the continues toare
2).pad couplings leak or a leak is detected in the
• Do not use •• Apply
outdoors.
Cold lubricant
Therapy toto
unit, o-rings
meet of connectors.
the requirements of UL 60601-1. securely pad
attacheditself, stop
to using
the it and call Breg Customer Care at
unit.
ain out
•• If leaking
With continues,
dry hands, connector ifthe
a leak is detected
transformer in the
to the pad or unit lid,Listen
electrical for(800)
stop using thethe321-0607.
unit and
double contact
clicks Electr
If the electrical connections
Breg Customer
connection fall into
careon
located water,
at doofnot
thetouch
1-800-321-0607
the top or +1-760-795-5440.
cooler. Next, plug the
unit
any wet part of thetransformer
product. Unplug only at dryelectrical
into an appropriate electrical outlet. The unit will
(Figure 1) to confirm the pad
(Fig. 1)
2)
Phys
(fig.
depress-
connections. automatically start running at this point.
couplingsDISPOSALare securely attached.
pulling Cleaning 7. Remove cap, which
Dispose contains
of product according to local regulations. Unit
ad and AfterUnplug
use, empty to stop thethe
and dry product when
unit with notcloth.
a soft in useWarm water andthe mildfoam
detergent
filter,mayfrombetheused occasionally
ping
Avoid flammables Unplugging
and the
oxidizers Polar Care unit turns it off. Always unplug Elec
Produ
to clean the pump and tubes. bottom of the pump (Figure 2)
immediately after use.vapors
Never leave it plugged WARRANTY M
10601,
Do not use in places with flammable or gasses (for in while and clean
Breg,out Inc. any debris.
warrants that this product is free from defects in 10701M
this unit unattended.
example, flammable anesthetics), high oxygen concentrations, Replace cap
workmanship,ensuring the and fitness for use for 180 days from
materials,
hout
or other oxidizers (forElectromagnetic
example, nitrous Interference
oxide). foam filter
initialispurchase
in place.under normal use for which it was intended 10901R
nplug This Scan
PolarQR code
Care unitformay
Breg’s
cause electromagnetic or other if it has been properly used under direct supervision of a M
Use a grounded interference
outletCube instructional
with othervideo.
electrical devices. To check whether the
Grounded (fig. 2)
licensed health care practitioner. Breg, Inc.’s obligation under
PolartoCare unit is interfering Outlet device, unplug the
with another M
Connect this product a properly grounded this warranty is limited to the replacement or repair of anyBreg, Inc.
PolarBreg.com/PCC
Care unit. If this corrects the problem, move the Polar R
outlet only. defective part or parts of this product. All express or2885 Loker Ave E
implied
Care unit or other device, or use an outlet on a different circuit. warranties, including the warranty of merchantability and92010 USA
Carlsbad, CA
water and
Protect
pump and
the power cord and transformer Leaks fitness for a particular purpose, are limited Toll Free Tel: 1-800-321-0607
to the actual Envi
• Keep the cord and transformer away from heated surfaces. Local: +1-760-795-5440 T
tomer © 2016 Breg, Inc. All rights reserved. AW-1.96015 Rev C 02/16
Note: Some condensation
warranty on
period set forth the lines,
above. Nocontroller, andexpress
other warranty, pads is
Fax: +1-760-795-5295
• Set upPolar
andCare
useisthis equipment
a registered in a and
trademark lowCube
trafficis a location
trademark away
of Breg, Inc. or implied,
unavoidable, is giveninand
especially no affirmation
warmer climates.of or by seller, by
www.Breg.com
from children and pets. words
1. If using or action,
a Breg Polar will
Care constitute a warranty.
Cold Therapy unit with a detach-
Never operate this product if it has a damaged cord,
Page 2 of 2 H
able Cold Therapy Pad, and a leak exists (other than conden-
transformer, or plug. If the cord, transformer, or plug is sation on the lines) disconnect the pad couplings. Make sure