2C730-532EN*G

INSTALLATION MANUAL
FOR
DIAGNOSTIC ULTRASOUND SYSTEM

MODEL SSA-640A
(2C730-532EN*G)

© TOSHIBA MEDICAL SYSTEMS CORPORATION 2008-2011


ALL RIGHTS RESERVED

YEM55700E-8
No. 2C730-532EN*G

IMPORTANT!

1. No part of this manual may be copied or reprinted, in whole or in part,


without written permission.

2. The contents of this manual are subject to change without prior notice
and without our legal obligation.

YEM55700E-8
No. 2C730-532EN*G

REVISION RECORD

REV. DATE REASON PAGE SER. DOC.


(MM/YY) /AUTHOR CHANGED No. PRODUCT.

INI. 12/'08 Mr. Motizuki ------- TM-WP2

*A 03/'09 Support of V1.1 Mr. Motizuki All pages

*B 04/'09 Support of V1.2 Mr. Motizuki P. 24 to 44

*C 10/’09 Support of V2.0 Mr. Isshiki P.17 to 44

*D 11/’09 Support of V2.1 Mr. Isshiki P.15, 25, 47

*E 01/’10 Support of V2.2 Mr. Isshiki All pages

*F 09/’10 Support of V3.0 Mr. Isshiki P.13,15,16,17,


22, 27,53

*G 03/’11 Support of UEPS-644A Mr. Isshiki P.1, 9

3 YEM55700E-8
No. 2C730-532EN*G

Safety Precautions

1. Meaning of Signal Words

In this installation manual, the signal words DANGER, WARNING, and CAUTION are used
regarding safety and other important instructions. The signal words and their meanings are defined as
follows. Please understand their meanings clearly before reading this manual.

Signal word Meaning

DANGER Indicates an imminently hazardous situation which, if not avoided, will


result in death or serious injury.

WARNING Indicates a potentially hazardous situation which, if not avoided, could


result in death or serious injury.

CAUTION Indicates a potentially hazardous situation which, if not avoided, may


result in minor or moderate injury.

CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in
property damage.

2. Symbols Regarding Safety

Symbol Description
Type-BF applied part
* Type BF when Type-BF applied part is connected.
All ultrasound transducers that can be connected to this system are Type-BF
applied parts.
"Attention" (Refer to the operation manual.)

3. Safety Precautions

To ensure safety, please observe the following precautions when installing this system.

DANGER: Do not use flammable gases such as anesthetics or flammable liquids such
as ethanol near this product, because there is danger of explosion.

4 YEM55700E-8
No. 2C730-532EN*G

WARNING: 1. Connect the power cable and earth terminals correctly as described, to
prevent electric shock.

2. Do not connect this system to outlets with the same circuit breakers and
fuses that control current to devices such as life-support systems. If this
system malfunctions and generates an overcurrent, or when there is an
instantaneous current at power ON, the circuit breakers and fuses of the
building’s supply circuit may be tripped.

3. Do not connect to the system transducers other than those specified by


TOSHIBA, to prevent accidents such as fire.

4. Do not subject the transducers to knocks. Use of defective transducers


may cause an electric shock.

CAUTION: 1. Malfunctions due to radio waves


(1) Use of radio-wave-emitting devices in the proximity of this kind of
medical electronic system may interfere with its operation. Do not bring
or use devices which generate radio waves, such as cellular telephones,
transceivers, and radio controlled toys, in the room where the system is
installed.

(2) If a user brings a device which generates radio waves near the system,
they must be instructed to immediately turn OFF the device. This is
necessary to ensure the proper operation of the system.

2. This system must be placed on a flat surface. If the system is installed on


the pole cart, be sure to securely mount the system to the pole cart and lock
the casters.

3. Before moving the system, be sure to fold down the LCD monitor and
confirm that the LCD monitor is locked. When moving a system that is
installed on the pole cart, hold it by the pole-cart handle. If the system may
fall, possibly resulting in personal injury.

4. To prevent electric shock, do not connect the peripheral units (video printer,
etc.) to an external outlet. Peripheral units should be connected to the
service outlet of the printer support kit. For the connection procedures,
contact your TOSHIBA representative.

5. Do not place any objects on top of the LCD monitor. They may fall, causing
injury.

6. Do not allow fluids such as water to contact the system or peripheral


devices. Electric shock may result.

5 YEM55700E-8
No. 2C730-532EN*G

CAUTION: 1. If the B/W printer is to be used, be sure to connect the printer support kit.
The specified B/W printer must be used.

2. To change the IP address, the network cable must be connected.


If the network cable is not connected, the IP address cannot be changed.

3. When change the setting of image compression.

(a) A image compression ratio can be changed using the DICOM Network Storage setting
(refer to subsection 12.1.4.7), but apply only to the customer who values the size of the
preservation file. The image quality decreases though the size of the preservation file
becomes small if the compressibility is raised.

(b) When change the compressibility setting, have the customer confirm the image
quality. Install the changed setting, after customer’s consent is obtained.

(c) When customer’s consent couldn’t be obtained, return the compressibility to former
setting and install it.

6 YEM55700E-8
No. 2C730-532EN*G

CONTENTS

Page

Safety Precautions --------------------------------------------------------------------------------------------------------------- 4

1. INTRODUCTION ---------------------------------------------------------------------------------------------------------- 9

1.1 Scope ---------------------------------------------------------------------------------------------------------------- 9

1.2 Installation Conditions ------------------------------------------------------------------------------------------- 9

1.3 Operating Requirements ---------------------------------------------------------------------------------------- 9

1.3.1 Power requirements --------------------------------------------------------------------------------- 9

1.3.2 Grounding requirements---------------------------------------------------------------------------10

1.3.3 Ambient conditions----------------------------------------------------------------------------------10

1.3.4 Storage and transport conditions----------------------------------------------------------------10

1.4 Unpacking ---------------------------------------------------------------------------------------------------------10

1.5 Preparation for Installation-------------------------------------------------------------------------------------11

1.6 Symbols ------------------------------------------------------------------------------------------------------------11

2. INSTALLING THE POWER SUPPLY -------------------------------------------------------------------------------12

3. INSTALLING THE B/W PRINTER (OPTION)----------------------------------------------------------------------14

4. CONNECTING THE TRANSDUCER --------------------------------------------------------------------------------15

5. SWITCH AND PERIPHERAL DEVICES SETTINGS ------------------------------------------------------------16

5.1 Setting the User Function Switches (UF) ------------------------------------------------------------------16

5.1.1 Procedures for displaying the Preset window ------------------------------------------------16

5.1.2 Registration procedures ---------------------------------------------------------------------------19

5.2 Setting the System for Use of a B/W Printer --------------------------------------------------------------20

6. INSTALLING THE KEYTOPS PROVIDED-------------------------------------------------------------------------21

7. DATE/TIME SETTING CHANGE -------------------------------------------------------------------------------------22

8. AUTOMATIC MEASUREMENT ITEM START FUNCTION ----------------------------------------------------23

8.1 Selecting the Measurement Items to Be Started Automatically --------------------------------------23

8.1.1 Basic measurement---------------------------------------------------------------------------------23

8.2 Setting Method ---------------------------------------------------------------------------------------------------24

9. INSTALLING THE DICOM KIT ----------------------------------------------------------------------------------------25

10. DATA BACKUP -----------------------------------------------------------------------------------------------------------25

7 YEM55700E-8
No. 2C730-532EN*G

CONTENTS - continued

Page

11. OPERATIONAL CHECKS----------------------------------------------------------------------------------------------26

12. APPENDIX A : INSTALLING THE DICOM KIT--------------------------------------------------------------------27

12.1 Installation Procedures -----------------------------------------------------------------------------------------27

12.1.1 Displaying the maintenance screen ------------------------------------------------------------27

12.1.2 Installing the optional software-------------------------------------------------------------------27

12.1.3 Setting the network environment ----------------------------------------------------------------29

12.1.4 Setting the DICOM information ------------------------------------------------------------------30

12.2 Operational Checks ---------------------------------------------------------------------------------------------50

13. APPENDIX B : UPDATING THE OPTION LICENSE ------------------------------------------------------------51

13.1 Displaying the maintenance screen -------------------------------------------------------------------------51

13.2 Updating the Option License ----------------------------------------------------------------------------------51

13.3 Backup the new option license -------------------------------------------------------------------------------52

14. APPENDIX C : CHARGING THE BATTERY PACK -------------------------------------------------------------53

8 YEM55700E-8
No. 2C730-532EN*G

1. INTRODUCTION

1.1 Scope

This manual describes the following procedures:

(1) Installing the SSA-640A

(2) Connecting the SSA-640A and optional units

1.2 Installation Conditions

Do not install the SSA-640A in a location where it may be exposed to the following adverse
conditions:

(1) Near a source of heat

(2) High ambient humidity

(3) Flammable gases

(4) A strong electric/magnetic field (for example, near a voltage transformer)

(5) A source of high-frequency radio waves

1.3 Operating Requirements

1.3.1 Power requirements

(1) Main unit

• Voltage : 24 VDC
• Capacity : 5.63 A

(2) AC adaptor

• Line voltage : 100 VAC to 120 VAC and 220 VAC to 240 VAC
• Line frequency : 50 Hz/60 Hz
• Line current : 1.8 A to 1.4 A and 0.9 A to 0.8 A

(3) Printer support kit

• Line voltage : 110 VAC to 120 VAC (UEPS-642A)


220 VAC to 240 VAC (UEPS-641A, UEPS-644A)
• Line frequency : 50 Hz/60 Hz
• Line current : 3.95 A to 3.63 A (UEPS-642A)
1.98 A to 1.81 A (UEPS-641A, UEPS-644A)

9 YEM55700E-8
No. 2C730-532EN*G

1.3.2 Grounding requirements

The grounding wire of a three-conductor power cord must be used to ground the SSA-640A.

The ground resistance must be 100 mΩ or less.

• Grounding must be provided in accordance with all applicable legal requirements for
medically used electrical equipment.

1.3.3 Ambient conditions

Ambient temperature : 10 °C to 35 °C

Relative humidity : 35 % to 85 % (no condensation)

Atmospheric pressure : 700 hPa to 1 060 hPa

1.3.4 Storage and transport conditions

Ambient temperature : Main unit -10 °C to 50 °C


Battery pack -10 °C to 50 °C (for up to 30 days)
-10 °C to 40 °C (for up to 90 days)
-10 °C to 30 °C (for up to 1 year)

Relative humidity : 30 % to 85 % (no condensation)

Atmospheric pressure : 700 hPa to 1 060 hPa

1.4 Unpacking

After unpacking the SSA-640A, confirm that the equipment and materials listed below are enclosed
in the packing boxes.

(1) Main unit SSA-640A

(2) Accessories

• Battery pack (standard capacity)


• AC adaptor
• Power cable
• Operation manual (Basic, Application, AOP)
• Installation manual, Installation quality check sheet
• Quick guide
• Microsoft EULA
• User function key top 1(sheet)
• Caution label (for USA)
• Recycle Label (for USA)
• Package list

10 YEM55700E-8
No. 2C730-532EN*G

1.5 Preparation for Installation

(1) Manpower and tools required

Manpower : 1 person

Tools : Keytop exchange tool (KEY EXCHG TOOL, BSM31-7964)


(Other tools are required when the pole-cart is to be used.)

(2) Time required for installation : 1 hour

1.6 Symbols

This system uses the following symbols. For safety symbols, refer to page 4.

Symbol Description
Power switch OFF

Power switch ON

TRANSDUCER A
Transducer connector A

11 YEM55700E-8
No. 2C730-532EN*G

2. INSTALLING THE POWER SUPPLY

(1) Place the battery pack so that it is aligned with the alignment line of the system.

Alignment line

(2) Slide the battery pack in the direction indicated by the arrow and confirm that the battery pack
is locked.

* To remove the battery pack from the system, slide the lock on the left of the system toward
the front of the system.

Lock release

Lock

12 YEM55700E-8
No. 2C730-532EN*G

(3) Connect the AC adaptor to the system. If the printer support kit is not to be used, connect the
supplied power cable to the AC adaptor.

Power cable

AC adaptor

CAUTION: When the caution message [CW020] is displayed during the first installation, see the
Appendix C : CHARGING THE BATTERY PACK.

13 YEM55700E-8
No. 2C730-532EN*G

3. INSTALLING THE B/W PRINTER (OPTION)

CAUTION: If the B/W printer is to be used, be sure to connect the printer support kit.
The specified B/W printer must be used.

(1) Connect the printer signal cable to the USB connector on the rear of the system.

(2) Connect a power supply cable of the printer support kit to the printer.

(3) Connect a power supply cable of the printer support kit to the AC adaptor.

(4) Connect the power cable to the printer support kit.

14 YEM55700E-8
No. 2C730-532EN*G

4. CONNECTING THE TRANSDUCER

Open the transducer connector cover and connect the transducer. Then rotate the handle to the
right to lock the transducer.

CAUTION: To prevent the connection pins being damaged, suit the position of the connector and
do not connect/disconnect the connector at an angle, when connecting or
disconnecting a transducer to/from the system.

15 YEM55700E-8
No. 2C730-532EN*G

5. SWITCH AND PERIPHERAL DEVICES SETTINGS

5.1 Setting the User Function Switches (UF)

Register functions to main panel switches as specified by the user.

5.1.1 Procedures for displaying the Preset window

(1) Turn ON the power switch of the system to start up the system.

(2) Select .

(3) Select .

(4) Click [Maintenance] in the on-screen menu to display the maintenance screen.

NOTE: Information displayed on the screen might be different in each system.

16 YEM55700E-8
No. 2C730-532EN*G

(5) Select [Conf.] in the maintenance screen to display the Conf. window.

(6) Click [UF Key] in the tree to display the password entry window.

17 YEM55700E-8
No. 2C730-532EN*G

(7) Enter the dedicated password and press [OK]. The Preset Key window is displayed.

<Functions of the tabs and switches>

Name of tab or switch Function


Type Item selection field. The options are A, B, C, D, E and F.
Copy Copies the preset key item to another type.
Paste Determines the type for the copy destination. (Displayed after [Copy]
is clicked.)
Copy Cancel Terminates the copying operation. (Displayed after [Copy] is clicked.)
Default Returns the settings to those at the time of shipment from the factory.
Wheel Function Set the function of Wheel key at each Imaging mode.
Save Registers the changes and returns to the maintenance screen.
Quit Returns to the maintenance screen without saving the changes.

18 YEM55700E-8
No. 2C730-532EN*G

5.1.2 Registration procedures

(1) Using the trackball, place the cursor on the word balloon for the desired switch and press
[SET].
The Function List window is displayed.

Display example

(2) After the desired function is selected, click [OK]. The functions can be assigned to the
desired switches by repeating the procedures. The registration of the selected key can be
cancelled by selecting [Delete] and then clicking [OK].

(3) The settings can be saved in the system by clicking [Save] on the Preset Key window.

(4) To restore the default settings, press [Default]. A confirmation dialog with the message
"Are you sure?" is displayed. Press [OK].

19 YEM55700E-8
No. 2C730-532EN*G

5.2 Setting the System for Use of a B/W Printer

Configure the system to enable control of printing to a B/W printer.

(1) Select .

(2) Select .

(3) Click [Maintenance] in the on-screen menu to display the maintenance screen.

(4) Select [Conf.] in the maintenance screen to display the Conf. window.

(5) Select [Special Configuration] in the tree to display the password entry window.

(6) Enter the predetermined password. The Special Configuration window is displayed.

(7) Set the preset conditions as follows.

(a) To connect a B/W printer


Parameter : B/W VCP
Modify Value List : Mitsubishi (for connecting a Mitsubishi printer)
Sony (for connecting a Sony printer)

B/W printer setting window

(8) Select [Save].

(9) A confirmation dialog with the message "Are you sure?" is displayed. Select [OK].

(10) A dialog with the message "Please Power off" is displayed. Push the power switch to reboot
the system.

20 YEM55700E-8
No. 2C730-532EN*G

6. INSTALLING THE KEYTOPS PROVIDED

Some of the user function keys to which functions were assigned in subsection 5.1 are provided with
UF labels. Replace the UF labels as follows.

(1) Using the keytop exchange tool (BSM31-7964), pull each of the relevant keytops in the
direction indicated by the corresponding arrow to remove it.

(2) Remove the tab from the encoder first. Then remove its keytop.

(3) Place the appropriate UF label inside the key-top and remount the keytop on the operating
panel.

21 YEM55700E-8
No. 2C730-532EN*G

7. DATE/TIME SETTING CHANGE

(1) Select .

(2) Select .

(3) Click [Preset] in the on-screen menu to display the preset window.

(4) Click [Time Correction] to display the Date and Time Properties window.

(5) Set the data and time.

(6) Click [OK] to close the Date and Time Properties window.

(7) Click [Exit] to close Preset window.

(8) Check if date and time displayed are correct.

22 YEM55700E-8
No. 2C730-532EN*G

8. AUTOMATIC MEASUREMENT ITEM START FUNCTION

8.1 Selecting the Measurement Items to Be Started Automatically

The measurement item or the measurement menu on the touch panel that is started initially when
the measurement switch on the main panel or the measurement switch /
on the touch panel is pressed can be set for each application preset/each image display mode
individually.

Perform setting as described below according to the user's preferences.

8.1.1 Basic measurement

The measurement item that is started initially when the measurement switch [CALIPER] or
[TRACE] on the main panel is pressed can be selected for each application preset/each image
display mode individually. The default measurement items are shown in the table below.

Image display mode Default measurement function


2D/2DColor Caliper : [Dist] (2D)
Trace : [Area]
M/MColor Caliper : [Dist] (M)
Trace : [Time]
Doppler Caliper : [Vel]
Trace : [Vel Trace]

23 YEM55700E-8
No. 2C730-532EN*G

8.2 Setting Method

(1) Select .

(2) Select .

(3) Click [Preset] in the on-screen menu to display the preset window.

(4) Click [Application Preset] to display the Application Preset window.

(5) Select an Application from the Application Preset list.


The selected Application is highlighted.

(6) Select [Auto Measurement] to display the Auto Measurement window.

(7) Change the settings in the 2D tab, M tab, D tab, and CDI tab as desired.

(a) Select Measure SW and press [SELECT] to open the Measurement Item Select window.

(b) Select Measure Item and press [OK].

(c) Change the other Measure SW settings as desired.

(d) Press [Save] to return to the Exam Preset window.

(8) Change the other Exam settings in the same manner.

(9) Press [Quit] to close the Exam Preset window.

(10) Press [Exit].

24 YEM55700E-8
No. 2C730-532EN*G

9. INSTALLING THE DICOM KIT

If the DICOM kit is to be installed together with the main unit, perform the procedures in
APPENDIX A "INSTALLING THE DICOM KIT".

10. DATA BACKUP

The data stored in the system may be erased due to incorrect operation or an accident. To prevent
data loss, back up the data using the procedure below.

(1) Start the system.

(2) Select .

(3) Select .

(4) Click [Maintenance] in the on-screen menu to display the maintenance screen.

(5) Select the [Utils] tab → [Backup Restore Tools] to display the password entry window.

(6) Enter the predetermined password to display Backup Restore Tools.

(7) Place a radio button for Backup.

(8) Insert the supplied USB flash memory into the port on the left side of the system, and then
specify the storage media (media field) to which preset data is backed up.

(9) Place a check mark for either All Packages in the Packages box.

25 YEM55700E-8
No. 2C730-532EN*G

(10) Select [Go].

CAUTION: Do not press the power button or do not open or close the LCD monitor
during the backup process.

(11) After the Progress status bar reaches 100%, push the [Quit] button to back to the maintenance
screen. Push the [Quit] button again to back to the initial screen.

11. OPERATIONAL CHECKS

Perform the operational checks as described in the installation quality check sheet (2J730-515EN).

When an option has been installed together with the system, also perform operational checks for the
option as described in the installation quality check sheet supplied with the option.

After updating with hardware change, renew an option license. (Refer to appendix B)

26 YEM55700E-8
No. 2C730-532EN*G

12. APPENDIX A : INSTALLING THE DICOM KIT

If the DICOM kit USDI-643A (option) is to be installed in the system, perform the following
procedures.

12.1 Installation Procedures

12.1.1 Displaying the maintenance screen

(1) Select .

(2) Select .

(3) Click [Maintenance] in the on-screen menu to display the maintenance screen.

12.1.2 Installing the optional software

(1) Select [Conf.] in the maintenance screen to display the Conf. window.

(2) Select [Option Management] → [Option Settings] to display the password entry window.

(3) Enter the predetermined password to display the following window.

27 YEM55700E-8
No. 2C730-532EN*G

(4) Select [Add…] in the license setting window. Then, select [Install From Code]. The
following window is displayed.

(5) Enter the license code and license name and click [OK]. The following window is
displayed.

(6) Click [Yes]. The following window is displayed. Click [OK].

28 YEM55700E-8
No. 2C730-532EN*G

12.1.3 Setting the network environment

(1) Select [Conf.] in the maintenance screen to display the Conf. window.

(2) Click [Network settings] in the tree and the following window is displayed.

(3) Set the IP address, subnet mask, default gateway, and, if necessary, the DNS server
address.

Item Description
Obtain an IP Address The IP address can be acquired automatically
Automatically (unselectable).
Use the following IP The specified IP address, subnet mask, and default gateway
address are used.
Obtain DNS server address The DNS server IP address can be acquired automatically
automatically (unselectable).
Use the following DNS The specified preference DNS server and substitute DNS
server address server are used.
[Apply] All the entered information is saved in the system without
closing the window.
[Reset] The entered information is reset.

CAUTION: To change the IP address, the network cable must be connected.


If the network cable is not connected, the IP address cannot be changed.

(4) Press [Apply] to save the entered information.

29 YEM55700E-8
No. 2C730-532EN*G

12.1.4 Setting the DICOM information

12.1.4.1 If the system cannot be connected to the network

(1) Purpose

The DICOM-related settings of the remote nodes can be checked and updated.

To check and update the detailed information items, use the Detailed Settings screen.

(2) Display screen

(a) Select [Conf.] in the maintenance screen to display the Conf. window.

(b) Then, select [DICOM Management] → [Local Node Settings] to display the
password entry window.

(c) Enter the predetermined password to display the following window.

No. Item Description


1 Storage/Query/Retrieve Specify the ultrasound system AE title to be used for the DICOM
SCU Storage, Query, and Retrieve communication sessions.
2 Use the same AE Title for Select this to use the same system AE title for all DICOM services.
all services
3 MWM SCU Specify the ultrasound system AE title to be used for DICOM
MWM communication.
4 MPPS SCU Specify the ultrasound system AE title to be used for DICOM
MPPS communication.
5 Print SCU Specify the ultrasound system AE title to be used for DICOM Print
communication.
6 Verification Specify the ultrasound system AE title to be used for DICOM
Verification communication (Echo).

30 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


7 Storage SCP Specify the ultrasound system AE title to be used for receiving
images and Storage Commitment events in DICOM Storage and
other communication sessions.
8 Port Number Specify the ultrasound system port number to be used for receiving
images and Storage Commitment events in DICOM Storage and
other communication sessions.
Note: Do not change this setting from the factory default (104).
9 [Detailed Settings] Press this switch to display the Detailed Settings screen. (Refer to
subsection 12.1.4.2.)
10 [Apply] Press this switch to save the entered information.
Note: The system must be rebooted to make the changes effective.
11 [Reset] Press this switch to discard all the entered information and load the
information from the system.
Note: Once [Apply] is pressed, the previous information cannot be
restored even by pressing [Reset].

(3) Basic setting procedure

(a) Enter the AE title and select "Use the same AE Title for all services".

(b) Enter the AE title and port number to be used for reception.

(c) Press [Apply] to save the entered information.

31 YEM55700E-8
No. 2C730-532EN*G

12.1.4.2 Setting detailed information related to the local host

(1) Purpose

The detailed DICOM-related settings of the ultrasound system can be checked and
updated.

(2) Display screen

(a) Press [Detailed Settings] in the local node setting window (SCU AE Title window)
described in subsection 12.1.4.1. The following warning dialog is displayed.

(b) Press [OK] to display the following window.

32 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


1 Simultaneous DICOM Specify the maximum number of DICOM associations that can be
Associations performed simultaneously from the ultrasound system.
2 Maximum PDU Size Select the maximum PDU size.
3 Accept Reject Time out The timeout value for Accept/Reject can be set. Enter a value in
the Timeout Value box.
Association Open Time The timeout value for opening an association can be set. Enter a
out value in the Timeout Value box.
Association Close Time The timeout value for closing an association can be set. Enter a
out value in the Timeout Value box.
Accept Message Time The timeout value for accepting a message can be set. Enter a
out value in the Timeout Value box.
Wait for Data Packets The timeout value for data packet reception can be set. Enter a
Time out value in the Timeout Value box.
DNM Response Time out The timeout value for system internal communication can be set.
Enter a value in the Timeout Value box.
HRIS Response Time out The timeout value for system internal communication can be set.
Enter a value in the Timeout Value box.
Accept network connect The timeout value for network connection can be set. Enter a value
Time out in the Timeout Value box.
4 [OK] The entered information is saved in the system and the dialog is
closed.
Note: The settings become effective after the system is rebooted.
5 [Cancel] The dialog is closed without saving the entered information in the
system.
Note: Once [Apply] is pressed, the settings cannot be returned to
the previous settings even by pressing [Reset].
6 [Apply] The entered information is saved in the system.
Note: The settings become effective after the system is rebooted.

(3) Basic setting procedure

(a) Change the settings as required.

(b) Press [Apply] to save the changes.

33 YEM55700E-8
No. 2C730-532EN*G

12.1.4.3 Setting the remote node information

(1) Purpose

The DICOM-related settings of the remote nodes can be checked and updated.

To check and update the detailed information items, use the Detailed Settings screen.

(2) Display screen

(a) Select [Conf.] in the maintenance screen to display the Conf. window.

(b) Select [DICOM Management] → [Remote Node Settings] to display the password
entry window.

(c) Enter the predetermined password to display the following window.

34 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


1 Node List A list of remote nodes used for DICOM Storage communication etc. is
displayed.
Mnemonic Name : The remote node name specified during setting
is displayed.
AE Title : The AE title of the remote node is displayed.
Host Name : The host name of the remote node is displayed.
Port Number : The port number of the remote node is
displayed.
ST : "Y" is displayed when the Storage service is
supported. When it is not supported, "N" is
displayed.
SC : "Y" is displayed when the Storage Commitment
service is supported. When it is not supported,
"N" is displayed.
QU : "Y" is displayed when the Query service is
supported. When it is not supported, "N" is
displayed.
RE : "Y" is displayed when the Retrieve service is
supported. When it is not supported, "N" is
displayed.
MW : "Y" is displayed when the MWM service is
supported. When it is not supported, "N" is
displayed.
MP : "Y" is displayed when the MPPS service is
supported. When it is not supported, "N" is
displayed.
2 [Confirm Settings] Press this switch to display the screen that shows all the DICOM-related
settings for the selected remote node. (Refer to subsection 12.1.4.10.)
3 [Export] Press this switch to display the screen for exporting the DICOM settings
file. The settings file can then be exported.
4 [Import] Press this switch to display the screen for importing the DICOM settings
file. The settings file can then be imported.
5 [Ping] Press this switch to execute the ping command for the selected remote
node and display the result.
The Host is Alive : The selected remote node has been found.
The Host is Not Alive : The selected remote node has not been
found.
6 [DICOM Echo] Press this switch to execute DICOM Echo for the selected remote node
and display the result.
Echo – Succeeded : DICOM Echo has been executed
successfully.
Echo – Failed : No response to DICOM Echo has been
received.
Echo – Error: System : DICOM Echo has failed due to a problem
other than the above.

35 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


7 [Test] Press this switch to open an association in the selected service with the
selected remote node and display the result.
Data is not transmitted during this process.
"Service name" - Succeeded
: The association has been opened successfully.
"Service name" - Failed: Rejected by Remote
: The association has been rejected by the
selected remote node.
"Service name" - Failed: Aborted by Remote
: The association with the selected remote node
has been aborted.
"Service name" - Error: Host Name (Remote)
: The host name is incorrect.
"Service name" - Error: Network
: The association with the selected remote node
is disabled.
"Service name" - Error: System
: The association has failed due to a problem
other than the above.
8 [Add] Press this switch to display the screen for adding a new remote node.
(Refer to subsection 12.1.4.4.)
9 [Edit] Press this switch to display the screen for editing the settings of the
selected remote node. (Refer to subsection 12.1.4.4.)
10 [Remove] Press this switch to delete the information related to the selected remote
node.

(3) Basic setting procedure

(a) Press [Add] to display the Add/Edit DICOM Nodes screen.

36 YEM55700E-8
No. 2C730-532EN*G

12.1.4.4 Adding a new remote node and editing the information for an existing remote node

(1) Purpose

A new remote node can be added or the DICOM settings for an existing remote node
can be edited. When this screen is accessed by pressing [Add] on the remote node list
screen, the fields are blank or filled with the default values. When this screen is
accessed by pressing [Edit], the values currently set for the selected remote node are
displayed.

(2) Display screen

Press [Edit] or [Add] on the remote node list screen shown in subsection 12.1.4.3. The
screen below is displayed.

No. Item Description


1 Mnemonic Name Enter the name of the remote node.
2 AE Title Enter the DICOM AE Title of the remote node (up to 16 characters).
3 Host Name Select the host name of the remote node from the drop-down list. The
options in the list are changed when host names are added or edited
using [Host Add/Del]. (Refer to subsection 12.1.4.6.)
4 [Host Add/Del] Press this switch to display the Host Add/Del screen.
5 Port Number Enter the port number of the remote node.
6 [Ping] Press this switch to execute the ping command for the selected host
and display the result.
The Host is Alive : The selected remote node has been found.
The Host is Not Alive : The selected remote node has not been
found.

37 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


7 [DICOM Echo] Press this switch to execute DICOM Echo for the remote node and
display the result.
Echo – Succeeded : DICOM Echo has been executed
successfully.
Echo – Failed : No response to DICOM Echo has been
received.
Echo – Error: System : DICOM Echo has failed due to a problem
other than the above.
8 [Test] Press this switch to open an association in the selected service with the
selected host and display the result.
Data is not transmitted during this process.
"Service name" - Succeeded
: The association has been opened successfully.
"Service name" - Failed: Rejected by Remote
: The association has been rejected by the
selected remote node.
"Service name" - Failed: Aborted by Remote
: The association with the selected remote node
has been aborted.
"Service name" - Error: Host Name (Remote)
: The host name is incorrect.
"Service name" - Error: Network
: The association with the selected remote node
is disabled.
"Service name" - Error: System
: The association has failed due to a problem
other than the above.
9 [Detailed Info >>>] Press this switch to display the detailed information setting area on the
Add/Edit DICOM Nodes screen. (Refer to subsection 12.1.4.5.)
10 Storage Select this check box to use the DICOM Storage service. When this
box is selected, the [Details] switch is enabled. Press this switch to
display the Storage Details screen. (Refer to subsection 12.1.4.7.)
11 Storage Commitment Select this check box to use the DICOM Storage Commitment service.
When this box is selected, the [Details] switch is enabled. Press this
switch to display the Storage Commitment Details screen. (Refer to
subsection 12.1.4.8.)
12 Query This function cannot be used in this system; therefore, do not place a
check mark in this check box.
13 Retrieve This function cannot be used in this system; therefore, do not place a
check mark in this check box.
14 MWM Select this check box to use the MWM service. When this box is
selected, the [Details] switch is enabled. Press this switch to display the
MWM Details screen. (Refer to subsection 12.1.4.9.)
15 MPPS Select this check box to use the MPPS service. When this box is
selected, the [Details] switch is enabled. Press this switch to display the
MPPS Details screen. (Refer to subsection 12.1.4.9.)
16 [OK] Press this switch to store the entered information and close the screen.
Note: The system must be rebooted to make the changes effective.
17 [Cancel] Press this switch to discard the entered information and close the
screen.
18 [Apply] Press this switch to store the entered information and keep the screen
displayed.

38 YEM55700E-8
No. 2C730-532EN*G

(3) Basic setting procedure

(a) Enter the mnemonic name.

(b) Enter the AE title.

(c) Enter the host name.

(d) Enter the port number.

(e) Select the DICOM services to be used. The screen below is displayed.

The [Details] switches are enabled then push the desired switch.

(f) Press [OK] to store the settings.

39 YEM55700E-8
No. 2C730-532EN*G

12.1.4.5 Editing the information for a new or existing remote node (detailed setting)

(1) Purpose

The detailed DICOM settings (required for DICOM communication) of a remote node
can be edited.

(2) Display screen

(a) Press [Detail Info >>>] in the Add Edit DICOM Nodes window for remote nodes
described in subsection 12.1.4.4. The following warning dialog is displayed.

(b) Press [OK] to display the following detailed settings.

No. Item Description


1 Retry Count Enter the number of retries.
2 Retry Interval Enter the retry interval.

(3) Basic setting procedure

(a) Change the retry settings.

(b) Press [OK] to store the changes.

40 YEM55700E-8
No. 2C730-532EN*G

12.1.4.6 Detailed setting of DICOM host names

(1) Purpose

A host can be added or deleted as a part of the DICOM Node detailed setting that is
required for DICOM communication.

(2) Display screen

Press [Host Add/Del] on the Add/Edit DICOM Nodes screen shown in subsection
12.1.4.4. The screen below is displayed.

No. Item Description


1 Host Name The currently set host names and IP addresses are displayed.
IP Address
2 [Ping] Press this switch to execute the ping command for the selected
host and display the result.
The Host is Alive : The selected remote node has been
found.
The Host is Not Alive: The selected remote node has not been
found.
3 [Add] Press this switch to display the screen for adding a host name.
4 [Edit] Press this switch to display the screen for editing the selected host
name.
5 [Remove] Press this switch to delete the selected host name.
6 [Close] Press this switch to close the screen.

(3) Basic setting procedure


(a) Press [Add] and add a host.

When the right host name is entered, press the Find switch to get the IP Address
automatically. To set the free host name, enter the IP Address manually.

(b) Press [Close] to close the screen.

41 YEM55700E-8
No. 2C730-532EN*G

12.1.4.7 Detailed setting for Storage

(1) Purpose

The detailed DICOM Storage information for a remote node can be edited.

(2) Display screen

(a) Press [Details] in the Add Edit DICOM Nodes window for remote nodes described
in subsection 12.1.4.4. The following warning dialog is displayed.

NOTE: If there are any remote node server settings that have not been saved, the save
confirmation dialog is displayed.
If [Yes] is pressed, the Storage Details window is displayed.

* To return to the Add Edit DICOM Nodes window without saving the remote node
server settings, press [No].

(b) Press [OK] to display the Storage Details window shown below.

42 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


1 Mnemonic Name The logical name of the selected remote node is displayed. This
cannot be changed.
2 Jpeg LossLess Select the compression methods to be supported.
Jpeg Lossy
RLE
3 Max Number of Request Specify the maximum number of connection requests to be sent
simultaneously.
4 Check Duplication Select this check box to skip requests to send the same data to the
same destination at the same time. This prevents duplicate
transmission and transmission errors.
5 Default Send Condition Enter the filtering condition (work status such as archived, printed,
etc.) for the data to be transmitted. Enter an asterisk (*) to transmit
all the data.
6 Monochrome Conversion Select this check box to convert color images to monochrome
images before transmission.
7 Image Data Type The image data types supported by the remote node are displayed
in the list on the left. The list on the right (individual setting list)
shows the settings for each image type. For the image types that
are not shown in the individual setting list, the default settings are
used.
Data : The image type is displayed in this column.
Send : Specify whether an image of the
corresponding type is to be transmitted or not.
Y: Transmitted, N: Not transmitted.
Compression : Specify the default compression type.
Quality Factor : Specify the default compression ratio.

8 [>>> Add >>>] Press this switch to add the image type selected in the list on the
left to the individual setting list.
9 [<<< Del <<<] Press this switch to delete the selected image type from the
individual setting list.
10 [OK] Press this switch to save the entered information.
Note: The system must be rebooted to make the changes effective.
11 [Cancel] Press this switch to discard all the entered information and close
the screen.
12 [Apply] Press this switch to store the entered information and keep the
screen displayed.

CAUTION: When change the setting of image compression.

(a) Apply only to the customer who values the size of the preservation file. The image
quality decreases though the size of the preservation file becomes small if the
compressibility is raised.
(b) When change the compressibility setting, have the customer confirm the image
quality. Install the changed setting, after customer’s consent is obtained.
(c) When customer’s consent couldn’t be obtained, return the compressibility to former
setting and install it.

43 YEM55700E-8
No. 2C730-532EN*G

(3) Basic setting procedure

(a) Change the settings.

(b) Select the image type and add it to the individual setting list.

(c) Change the settings in the individual setting list.

(d) Press [OK] to store the changes.

12.1.4.8 Detailed setting for Storage Commitment

(1) Purpose
The detailed DICOM Storage Commitment information for a remote node can be edited.

(2) Display screen


(a) Press [Details] for Storage Commitment in the Add Edit DICOM Nodes window for
remote nodes in subsection 12.1.4.4. The following warning dialog is displayed.

(b) Press [OK] to display the Storage Commitment Details window shown below.

No. Item Description


1 Mnemonic Name The logical name of the selected remote node is displayed. This
cannot be changed.
2 Commitment Mnemonic Specify whether a Commitment request is sent or not when a
Name Storage request is sent to the node with the name shown above.
Supported Only : A Commitment request is not sent (but
supported).
Current Node : A Commitment request is sent to the same
node to which the Storage request is sent.
"Node name" : A Commitment request is sent to a node other
than the one to which the Storage request is
sent. The drop-down list contains only the
remote nodes that support the Storage

44 YEM55700E-8
No. 2C730-532EN*G

Commitment service.
No. Item Description
3 Timeout Value Specify the timeout value for reception of a Commitment
Completed message after a Commitment request is sent.
4 [OK] Press this switch to save the entered information.
Note: The system must be rebooted to make the changes effective.
5 [Cancel] Press this switch to discard all the entered information and close
the screen.
6 [Apply] Press this switch to store the entered information and keep the
screen displayed.

(3) Basic setting procedure

(a) Change the settings.

(b) Press [OK] to save the changes.

45 YEM55700E-8
No. 2C730-532EN*G

12.1.4.9 Detailed setting for MWM/ MPPS

(1) Purpose

The detailed DICOM MWM/ MPPS information for a remote node can be edited.

(2) Display screen

(a) Press [Details] in the Add Edit DICOM Nodes window for remote nodes described
in subsection 12.1.4.4. The following warning dialog is displayed.

NOTE: If there are any remote node server settings that have not been saved, the save
confirmation dialog is displayed.
If [Yes] is pressed, the MWM( or MPPS) Details window is displayed.

* To return to the Add Edit DICOM Nodes window without saving the remote node
server settings, press [No].

(b) Press [OK] to display the Details window shown below.

MWM MPPS

* Check the “Set as Default” to use it as the default server.

46 YEM55700E-8
No. 2C730-532EN*G

12.1.4.10 Confirming all the settings for a remote node

(1) Purpose

All the DICOM-related settings for a remote node can be confirmed on a single screen.

(2) Display screen

Press [Confirm Settings] on the remote node list screen shown in subsection 12.1.4.3.
The screen below is displayed.

No. Item Description


1 Category Setting category
2 Key Setting item
3 Value Setting value
4 Data Image data type
5 Send Send settings
6 Compression Compression information
7 Quality Factor Quality factor of the compressed data at transmission
8 Retrieve Receive settings
9 [OK] Closing the window

(3) Basic setting procedure

(a) Check the settings and press [OK] to close the screen.

47 YEM55700E-8
No. 2C730-532EN*G

12.1.4.11 Setting the DICOM Print function

CAUTION: When setup the Detailed setting for Storage (refer to 12.1.4.7) and MWM/MPPS (refer
to 12.1.4.9), complete those setting before setup the DICOM Print function.

(1) Purpose

DICOM Printer can be added to the system and images and reports are printed out
through DICOM network.

(2) Display screen

Press button on the OSM to display the Patient Browser screen.

Select [Setting] button on the [Tools] tab and then select [Print] tab. The screen below is
displayed.

48 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


1 Report Printer Set the printer for Report.
2 DICOM/Image Select the DICOM printer or another printer for Images.
Printer
3 Orientation Set the print paper orientation.
Portrait : The data is printed in portrait orientation.
Landscape : The data is printed in landscape orientation.
4 Size Set the print paper size.
A4 : Size A4 is set.
Letter : Letter size is set.
5 Division Set the print layout. A default setting is 1x1.
6 Add DICOM Open the page to set up the DICOM Printer.
Printer
7 [Save] Used to save the changes in the system.
8 [Cancel] Used to close the Settings screen without saving the
changes.

(3) Basic setting procedure

(a) Select [Add DICOM Printer] to display the following screen.

Add new DICOM printer and press [Save] to close the screen.

(b) Setup paper size, paper orientation.

(c) Press [Save] to close the screen.

(d) Register the DICOM Printer to the User Function switch (Refer to subsection 5.1).

(e) Restart the system.

<< DICOM Printer set up page>>

49 YEM55700E-8
No. 2C730-532EN*G

No. Item Description


1 Mnemonic Name Set the printer name to identify it.
2 AE Title Set the DICOM AE title.
3 IP Address Set the IP address of the printer.
4 Border Line Specify whether or not to print a border line.
To print a border line, check the check-box.
5 Color Specify whether or not to print color pages.
Appearance Default is Monochrome.
6 Media Select printing papers. Default is ”Blue Film”.
7 Magnification Select a method of Image interpolation.
REPLICATE
BILINEAR
CUBIC : Default
NONE
8 Film size Set a paper size.
9 Ping Display the result of Ping test.
10 Echo Display the result of Echo test.
11 Max Density Set the maximum density of printing. Default is 200.
12 Min Density Set the minimum density of printing. Default id 20.
13 [All Delete] Delete all printers.
14 [Save] Used to save the changes in the system.
15 [Cancel] Used to close the Settings screen without saving the changes.

12.2 Operational Checks

Perform the operational checks as described in the installation manual and installation quality
check sheet (2J708-841EN).

50 YEM55700E-8
No. 2C730-532EN*G

13. APPENDIX B : UPDATING THE OPTION LICENSE

13.1 Displaying the maintenance screen

(1) Select .

(2) Select .

(3) Click [Maintenance] in the on-screen menu to display the maintenance screen.

13.2 Updating the Option License

(1) Select [Conf.] in the maintenance screen to display the Conf. window.
(2) Select [Option Management] → [Option Settings] and enter the predetermined password
to display the following window.

(3) Select the option license in the option setting window. Then, select [Edit].

51 YEM55700E-8
No. 2C730-532EN*G

(4) The following window is displayed. Then click [Yes].

(5) The following window is displayed. Click [OK].

(6) A new license code is shown.

13.3 Backup the new option license

(1) Select the [Utils] tab → [Backup Restore Tools] to display the password entry window.

(2) Enter the predetermined password to display Backup Restore Tools.

(3) Place a radio button for Backup.

(4) Insert the supplied USB flash memory into the port on the left side of the system, and then
specify the storage media (media field) to which preset data is backed up.

(5) Place a check mark for either All Package in the Package box.

(6) Select [Go].

CAUTION: Do not press the power button or do not open or close the LCD monitor
during the backup process.

(7) After the Progress status bar reaches 100%, push the [Quit] button to back to the
maintenance screen. Push the [Quit] button again to back to the initial screen.

52 YEM55700E-8
No. 2C730-532EN*G

14. APPENDIX C : CHARGING THE BATTERY PACK

Where the battery pack is not in use until the first installation, the caution message may be
displayed when performing installation of the Viamo SSA-640A, this is not failure.

Even the above message is displayed, the Viamo SSA-640A is not failure.

Point the cursor to [OK] in the message box and press [SET] button, and then the Viamo SSA-
640A can be used.

Recharge the battery pack fully, then this message is not displayed.

The battery indicator LED is lighting green and amber alternately, when this message is displayed.

The battery pack should be recharged fully until the LED lights green.

When 90% or more of battery power is remained, recharging the battery pack may fail. So
discharge the battery pack (do battery operation) for more than 5 minutes.

Repeat this procedure three times as the case may be. (Please execute the above steps when not
changing into green within one hour.)

If the LED does not light green, this is possibly resulting in failure, please investigate the cause to
see Service Manual 2.10.8 (Troubleshooting - Battery pack).

NOTE: Causal explanation of a phenomenon (Remained battery power is unknown)

The battery pack has a built-in microcomputer to manage the remained battery
power. For longer storage periods, where the battery pack is recharged fully (100%)
and kept, built-in microcomputer in the battery pack records that remaining charge
in the battery pack is 100%.
However a battery pack has self-discharge characteristics, the remaining charge
should not be 100%; and then built-in microcomputer finds abnormal when
connecting the battery pack to the system next time, causing in the remaining
charge being unknown.
Caution message is displayed on the basis of "the remaining charge which built-in
microcomputer in battery pack recorded", even though the actual remaining charge
might be less. Battery Alarm does not run properly in this case.
To avoid this issue, the Viamo SSA-640A is displayed the caution message and
notices that the battery pack should be recharged fully (100%).
This is designed for the purpose of run down before battery alarm is displayed.

53 YEM55700E-8 E