The brightfield compound microscope uses two lens systems to magnify specimens placed on the stage under illumination. It has four objective lenses with increasing magnifications: scanner at 4x, low power at 10x, high power at 45x, and oil immersion at 100x. Total magnification is calculated by multiplying the eyepiece magnification, usually 10x, by the objective magnification. The oil immersion objective requires using immersion oil of refractive index 1.516 between the objective and specimen slide to improve resolution. Proper use of the microscope involves focusing with the coarse and fine adjustment knobs while changing between objectives.
The brightfield compound microscope uses two lens systems to magnify specimens placed on the stage under illumination. It has four objective lenses with increasing magnifications: scanner at 4x, low power at 10x, high power at 45x, and oil immersion at 100x. Total magnification is calculated by multiplying the eyepiece magnification, usually 10x, by the objective magnification. The oil immersion objective requires using immersion oil of refractive index 1.516 between the objective and specimen slide to improve resolution. Proper use of the microscope involves focusing with the coarse and fine adjustment knobs while changing between objectives.
The brightfield compound microscope uses two lens systems to magnify specimens placed on the stage under illumination. It has four objective lenses with increasing magnifications: scanner at 4x, low power at 10x, high power at 45x, and oil immersion at 100x. Total magnification is calculated by multiplying the eyepiece magnification, usually 10x, by the objective magnification. The oil immersion objective requires using immersion oil of refractive index 1.516 between the objective and specimen slide to improve resolution. Proper use of the microscope involves focusing with the coarse and fine adjustment knobs while changing between objectives.
INTRODUCTION TO BRIGHTFIELD MICROSCOPE OBJECTIVE LENSES
simplest type of compound microscope is the 1. SCANNER: Magnification: 4x
Brightfield compound microscope. 2. LOW POWER OBJECTIVE (LPO): word “compound” from Brightfield Compound -Magnification: 10x Microscope means “specimen positioned properly on -forms general outline or wider portion of the object. the stage of a microscope and illuminated by a light 3. HIGH POWER OBJECTIVE (HPO): -Magnification: 45x source will be magnified by two-lens system.” -longer than the LPO and it forms a bigger image if the word “brightfield” means “magnified objects appear object in focus. as dark objects against a bright background.” A -enlarge specimens are small under LPO. sufficient contrast must exist between the magnified 4. OIL IMMERSION OBJECTIVE (OIO): object and the brightfield background for the objects to -HIGHEST DEGREE OF MAGNIFICATION be visible. -Magnification: 100x THE PATH OF LIGHT -examine stained smear preparations of microorganisms using immersion oil as their medium. OCULAR LENS- image of the specimen is Calculating the Total Magnification magnified - As “eyepiece” TOTAL MAGNIFICATION = Magnifying power of the eyepiece x OBJECTIVE LENS- light rays Magnifying power of the objective used CONDENSER-lenses that direct the light rays Magnifying power of the eyepiece = 10 ILLUMINATOR- light source Magnifying power of the objective used = 100 (OIO) IMPORTANCE OF USING IMMERSION OIL Correct Manipulation of Microscope immersion oil has the same refractive index as the glass slide 1. Place the specimen slide on the stage and secure it with the IO is used, light rays do not refract stage clips. Arrange the portion of the slide to be examined resolving power of the lenses. over the central opening stage. 2. Rotate the low power objective into place under the body When oil is not used with an oil immersion objective tube. You will feel a “click” when it is correctly place. lens, the image becomes fuzzy, with poor resolution. 3. Raise the condenser as high as it will go. HOW TO APPLY? 4. Rotate the coarse adjustment knob clockwise to bring down the LPO close but not touching the slide, until the specimen oil used in oil immersion is “CEDAR WOOD OIL”. is seen through the eyepiece. refractive index is: 1.516. 5. Regulate the intensity of light by opening or closing the iris diaphragm PARTS OF THE MICROSCOPE 6. Sharpen the focus by turning the fine adjustment knob. Structural Components 7. Turn the LPO to HPO without elevating the body tube. Notice that it will be almost in focus because most 1.HEAD/BODY-houses the optical parts microscopes are parfocal. Little adjustments with fine 2.ARM- supports the microscope head adjustment knob will only be needed to clearly view the object in focus. - carry microscope NOTE: The HPO has a smaller aperture than the LPO, hence it will be necessary to open the iris diaphragm further in 3.BASE-supports the microscope and houses the illuminator order to fill the objective aperture with light Optical Components 8. To focus the oil immersion lens, swing the oil immersion objective halfway towards the specimen in focus. Place a 1. EYEPIECE/OCULAR- remagnifies the image formed by the drop of immersion oil on the smear. Swing the oil immersion objective lens; standard magnifying power is 10x. lens in place. Using the coarse adjustment, bring the 2. EYEPIECE TUBE- holds the eyepieces in place above objective down until it touches the oil. Find the object by objective lenses. gently turning the coarse adjustment knob upwards. 3. OBJECTIVE LENSES- primary lenses magnifies specimen. Sharpen the focus with the fine adjustment knob. 4. NOSEPIECE-rotating turret 9. Compare the relative sizes from those seen under the LPO, 5. COARSE ADJUSTMENT KNOB- bigger wheel used to adjust HPO, & OIO. the LPO in focusing; initial focusing. 6. FINE ADJUSTMENT KNOB- smaller wheel for final focusing 10. Clean the stage and the lenses after each use, before of the specimen using HPO and OIO; make the specimen returning the microscope to the stockroom. more vivid. 7. STAGE- slide/specimen is placed for focusing. 8. STAGE CLIPS- hold the slide in place 9. ILUMINATOR- light source 10. CONDENSER- collect and focus the light from illuminator 11. IRIS DIAPHRAGM- controls the amount of light 12. CONDENSER FOCUS KNOB- moves the condenser up or down to control the lighting focus on the specimen.