1. SYSTEM OVERVIEW
1.1 INTRODUCTION
Allengers 100 is a Diagnostic X-Ray generator, which caters to the general requirements of
Radiography. This generator is also available in other options like Fluoroscopy and Spot
Filming. This equipment can be used with following combinations of stands and tables:
STANDS:
Floor to Ceiling stand (FC Stand)
Floor to Ceiling stand with Fluoroscopy (FCF Stand)
Floor to Ceiling stand with Spot Film Device (FCSFD)
Spring balanced mobile stand (SBM)-Incase of Machines with Radiography only
Counter balanced mobile stand (CBM)- Incase of Machines with Radiography only
TABLES:
Horizon
Multipose
Floatex
Allengers 100 X-Ray generator is ideally suited for Chest/ Extremities/ Skull/ Abdomen and
other routine examinations.
This unit incorporates State - of – the- Art technology and is designed keeping in mind the
diverse requirements of the Radiologists. It is a Full Wave Rectified Unit (Two Pulse) and
consists of Control Panel & Self contained Tube Head Unit with manual collimator.
This machine is Type Approved for Electrical, Mechanical and Radiation Safety Standards
(vide IS: 7620) from the Atomic Energy Regulatory Board (AERB), Govt. of India, thus making
it entirely safe from the Radiation Hazards.
Parameters Specifications
Output (As per IS: 7620) 8 KW
Max. mA output 100 mA
Radiographic KVP 45–100 KVP
(Selectable in Steps by KVP Selector Switch)
Fluoroscopy KVP* 45-100 KVP
(Selectable in steps by KVP Selector Switch)
Radiographic Timer 0.04 to 8 Seconds (in 24 Steps)
Fluoroscopy Timer* Commulative timer of 5 Minutes
Radiographic mA 25 mA, 50 mA & 100 mA
Fluoroscopic mA* Up to 3 mA
Electronic Overload Simultaneous protection from excessive selection of
mA, seconds, KVP and input voltage.
Voltage Compensator Switch Provided for input voltage variations
Tube Head with Collimator Self contained Tube Head consisting of Stationary
Anode X-Ray Tube, Compact Heavy Duty H.V.
Transformer and Silicon Rectifiers all oil immersed
and hermetically sealed. It can be rotated 360 o for use
in operation theatres and in wards for bedside X-
Rays. Manual collimator is provided with Tube Head.
Power Supply 230 Volts AC, 50/60 Hz. 1 Phase. Max. Allowable
Line Regulation ± 10 %
Digital Displays KVP, mA & mAS
Visual Indicators X-Ray ON, Line ON, Overload
13 14 15 16 17
1 12
11
2
3 4 5 6 7 8 9 10
BACKSIDE OF
ON CONTROL
1.3.1 Description of Control Keys
CONTROL
PANEL
1. LINE VOLTAGE COMPENSATOR SWITCH
This is used to adjust the incoming power supply voltage. Using Fine Compensator,
small variations in input voltage to equipment can be adjusted in 12 steps of about 2.5
volts each. The Coarse Compensator switch is also provided at the backside of
control which is having 4 steps of about 25 volts each and is used to adjust large
variations in input voltage.
2. READY SWITCH
When Ready switch is pressed, it boosts X-Ray tube filament voltage from stand by
to operating value determined by the setting of the TECHNIC SELECTOR.
3. X-RAY SWITCH
This switch is used to select Radiographic and Fluoroscopic* KVP Voltage from 45
KVP to 100KVP in steps of 5 KVP each.
6. MACHINE ON SWITCH
This is Push to On type switch used to energize the equipment. The Line Indicator
glows when this switch is pressed.
This is Push to Off type switch used to de-energize the equipment. The LINE
Indicator will turn off when this switch is pressed.
9. LINE INDICATOR
It glows when the selected factors viz. KVP, mA or time exceeds the tube rating
chart.
NOTE: When the overload indicator glows, the X-Ray unit automatically
becomes inoperative.
13. VOLTMETER
This voltmeter indicates input supply to equipment circuits. Red band is provided on
voltmeter at 230 V. Voltage compensator switch is adjusted until pointer on voltmeter
is at this red band.
Displays the mA selected from Technic selector switch during Radiographic Mode.
Displays KVP selected from KVP selector switch during Radio or Fluoro* mode.
17. mA METER
This meter indicates X-Ray tube current (mA) during exposure. The Needle deflects
during exposure & indicates tube current as per setting on technic selector switch.
NOTE: For exposure less than 1.5 sec. duration, pointer may not get enough time to
travel upto the desired position e.g. for an exposure of 100mA, 0.10 sec, mA meter
pointer will not reach upto 100mA mark.
FLOURO mA CONTROL*
This is variable resistor fixed on side of control that allows continuous adjustment of the
fluoroscopic milli ampere up to maximum value of 3mA.
FOOT SWITCH*
This is used to press Ready switch and X-Ray Start switch using retractable type cord.
This is Push to ON type switch. When it is pressed bulb gets ON and remains ON for 1
minute. After 1 minute bulb gets OFF automatically.