1 Name of the Institution and City New Hope Medical Centre Pvt Ltd,
No 814, Poonamallee High Road,
Kilpauk, Chennai, Tamil Nadu-600010.
2 Type of Equipment C-Arm
3 Model Name Skan C
4 Name of the Manufacturer Skanray Technologies Limited
5 Serial No of the X Ray Equipment MY3422P0170
6 Location of QA test item & QA New Hope Medical Centre Pvt Ltd,
Performed Kilpauk-Chennai.
7 Name(s) of Person(s) testing the N.Sudhakar;
equipment and Name of Advanced QA Services,
Supplier/Service Agency
Chennai.
8 Dates and Duration of the Tests 28/10/2022 - 2 hrs
9 Customer Request Form No: AQAS/CRF-22/0431
10 Report Number: TC945422000000578F
11 Report Type: Original / Correction / Amendment
(Strike out the field not applicable and fill Correction: (original rep No: )Reason for
therelevant information if correction or correction:
amendment) Amendment (Original rep No:
Reason for amendment:
Page 1 of 7
No.70B, Dr. Ambedkar Street, Manali,
ENVIRONMENTAL CONDITION
Radiological Condition Acceptable
Whether two units are in same room Yes No
Is there any radiation interference due to other equipment Yes No
presence
Ambient Temp 280 C
Unit Operational Condition Operational
Not Operational
Central Beam
2
Alignment NA Central Beam Alignment < 1.50 NA
Tolerance :
3
Effective 1. + 0.5 f for f 0.8 mm
focal spot NA 2. + 0.4 f for 0.8 f 1.5 mm NA
measurement 3. + 0.3 f for f 1.5 mm
4 Accuracy of 50 kV 49.33 kV
Operating 70 kV 69.2 kV ± 5 kV PASS
Potential (kV)
90 kV 90.33 kV
5 Accuracy of
Irradiation NA NA % Error < 10 % NA
Time(s)
6 Tolerance :
Total filtration Measurement 1.5 mm Al for kV 70 PASS
at 90 kV 3.2 mm Al 2.0 mm Al for 70 < kV 100
2.5 mm Al for kV 100
Page 2 of 7
No.70B, Dr. Ambedkar Street, Manali,
I hereby undertake that all the information provided above is correct and in accordance with the
detailedQuality Assurance Report enclosed herewith.
Reference Document for test Method - As per AERB prescribed test protocols.
QA test report does not contain any recommendation on the QA test interval except where
there has been agreed with the customer. However, for customer reference generally two year
QA testing interval mentioned in the report unless otherwise requested by the customer as
prescribed by AERB.
Page 3 of 7
No.70B, Dr. Ambedkar Street, Manali,
# Quality Assurance Tests Report shall be signed by Institution’s Representative and duly stamped by the
User’s Institution. The report shall not be reproduced except in full without the written approval of the
agency.
1. CONGRUENCE OF RADIATION & OPTICAL FIELD:
Operating parameters:
-NOT APPLICABLE-
FFD (cm) kV mAs
Page 4 of 7
No.70B, Dr. Ambedkar Street, Manali,
Operating parameters:
-NOT APPLICABLE-
FFD (cm) kV mAs
Observe the images of the two steel balls on the radiograph and evaluate tilt in the central beam.
Observed tilt NA Remark: NA
Tolerance: Central Beam Alignment 1.5 0
Page 5 of 7
No.70B, Dr. Ambedkar Street, Manali,
Operating parameters:
8. OUTPUT CONSISTANCY:
FDD (cm) 100
Diameter of the smallest size hole clearly resolved on the monitor 2.9 mm hole pattern clearly visible
Recommended performance standard 3.0 mm hole pattern must be resolved
Page 6 of 7
No.70B, Dr. Ambedkar Street, Manali,
Distance from Focus Mode Applied kV Applied mAs Exposure rate at table top Remarks
to Table top (cGy/Min)
60 Manual 110 5 3.35 cGy/Min PASS
Tolerance :
1. Exposure Rate without AEC mode ≤ 5 cGy/Min
2. Exposure Rate with AEC mode ≤10 cGy/Min
Minimum focus to table top distance shall be 30 cm for fluoroscopy