The information in this article is for “386” MASTER COMPUTER BOARDS only.
After the MASTER COMPUTER BOARD has been removed and reinstalled, Condition Code 4VW
“>SCHEDULER * BOOT * System call failed” will be displayed. To correct the problem, reset the date and the
time:
[1] From the Main Menu, touch:
“OPTIONS”
“SPECIAL SERVICES”
“DATE & TIME”
[2] Follow the prompts on the SCREEN to enter the correct date and time.
[3] Return to the Main Menu.
[4] Touch “SHUTDOWN ANALYZER”.
[5] When the screen displays “This function will require a robot. Do you wish to continue?”, press [Y] and
[Enter].
[6] When the screen displeases analyzer has been shutdown. Reset or power down is required.”, de-energize the
250 Chemistry System.
[7] Wait 30 seconds. Then energize the 250 Chemistry System.
[8] Return the 250 Chemistry System to the normal operating condition.
The Condition Code Message on the 250 System does not include these instructions. The service publication
includes instructions for setting the date and time after the installation of the MASTER COMPUTER BOARD.
Ed Graham,
Technical Assistance Center (TAC)
Rob Peters,
FS, Atlantic Central
Please insert this article with the CIRCUIT BOARDS section of the service manual.