Professional Documents
Culture Documents
Compact REFERENCE#MANUAL 20090701
Compact REFERENCE#MANUAL 20090701
Revision Table
Revision Table
1-i
0931033L - July 2009
REFERENCE MANUAL
Revision Table
1-ii
0931033L - July 2009
REFERENCE MANUAL
Revision Table
Additions:
Chapters 1.6, 1.8, 2.7.2 addenda 0931755
Chapter 4.5.2 description of the use of the stylets
Chapter 4.5.3 description of the two types of
knurled nuts fastening the needles
Chapter 1.1 symbols "biohazardous liquid",
"temperature", "precautions", "Caution, hot
surface"
Chapter 1.4 information on the ISI value
Chapter 1.5 On / Off procedure of the analyzer:
wait 30 sec. before turning the analyzer back on
1-iii
0931033L - July 2009
REFERENCE MANUAL
Revision Table
Deletions:
Chapters 2.5.1, 2.7.4 description of the CRT
screen
Chapter 4 use of the metal mandrels
Chapter 12 list of tubes
Modifications:
Chapter 1.1 description of the symbol "warning"
Chapter 2.7.3 minimum diameter for 5 ml tubes:
11 mm (not 11.6 mm)
Chapter 4.1 maintenance schedule for a
laboratory performing up to 200 tests a day and
not more than 200 tests a day
Chapter 4.2.2 description of the preparation of
the decontamination solution
Chapters 7.4.9.2, 7.5.5 precautions concerning
the ISI
Minor modifications of the wording
Modification of some illustrations
1-iv
0931033L - July 2009
REFERENCE MANUAL
Table of contents
Index
1-i
0931033L - July 2009
REFERENCE MANUAL
Table of contents
1-ii
0931033L - July 2009
REFERENCE MANUAL 1
Table of contents
1.1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1-i
0931033L - July 2009
1 REFERENCE MANUAL
Table of contents
1-ii
0931033L - July 2009
REFERENCE MANUAL 1
Symbols and warnings
Off On
Disposables
In the documents
1-1
0931033L - July 2009
1 REFERENCE MANUAL
Symbols and warnings
Initial situation
2
(I) in case of non respect of the Recommendations described in this manual as well as the
use of procedures that are not set forth by Diagnostica Stago,
(II) in case of use of reagents other than those manufactured by Diagnostica Stago even if
the use of said reagents associated with the STA Compact® system is specified in an
adaptation protocol,
(III) in case of use of tubes other than those specifically described in chapter 2.7.3 of this
manual,
(IV) in case of use of washed and/or used cuvettes, the reacting cuvettes being consumables
intended for a single use, as well as cuvettes other than those manufactured by
Diagnostica Stago and commercialized by Diagnostica Stago and/or by its official
distributors,
1-2
0931033L - July 2009
REFERENCE MANUAL 1
Symbols and warnings
(V) in case of non accomplishment of the current and regular maintenance operations,
calibration and support operations set forth in the present manual and permanently
required to ensure the correct functionning and security of the STA Compact® system,
(VI) in case the STA Compact® system has not been decontaminated in accordance with the
decontamination procedures described in this manual.
(II) Make sure that the STA Compact® is decontaminated in conformity with the
decontamination procedures described in this manual,
(III) Observe all standards and precautions that are imposed to laboratories for the
execution of biological assays involving biohazardous products in accordance
with the regulations in effect locally.
1-3
0931033L - July 2009
1 REFERENCE MANUAL
Symbols and warnings
Pre-analytic conditions regarding the samples: in order to maintain the activity of various
coagulation factors, the sample should be taken with care and following the professional
standards in tubes with a specified concentration of citrate.
The quality of the centrifugation and the storage temperature of the sample should also be
carefully insured before the analysis:
- plasma that is hemolysed, partially coagulated (presence of micro-clots), damaged by
temperature changes or with bubbles on its surface may cause inaccurate results.
- plasma that has been frozen may contain interfering precipitates when thawed. These
precipitates should be removed before measurement is performed.
Pre-analytic conditions regarding the products and the reagents: the laboratory must strictly
comply with the instructions given by the manufacturer in the product and reagent inserts. Poor
preparation of the reagent regarding reconstitution volume, stabilization time, stirring as well as
the presence of bubbles or the accidental presence of a magnetic rod may lead to incorrect
results.
The ISI value of the thromboplastin used for the determination of the prothrombin time should
be the one indicated on the assay value insert of the STA product.
Any change in lot, any software update - like any intervention - should give rise to verification of
the ISI value.
Analyzer setup and condition: the laboratory must make sure that the test setting complies with
the instructions given by the reagent manufacturer in the inserts, especially regarding volumes to
be used, incubation times, rinsing solutions, buffers, etc. It is the laboratory's responsibility to
correctly enter the bar codes and to choose the appropriate test setting for a given test.
Moreover, the analyzer brings results from biological materials and although these materials are
implemented and measured by highly sophisticated computer-controlled automated systems in
order to optimize reliability and safety, it is impossible to guarantee an error rate of zero. The
laboratory must ensure that the maintenance is performed on a regular basis and in conformity
with the Recommendations described in this manual.
1-4
0931033L - July 2009
REFERENCE MANUAL 1
Symbols and warnings
All these functions are protected by access codes (see chapter 10.2.2.3). The use of routine functions
has no impact on the results obtained. However, the use of change functions may affect result
accuracy.
Therefore, it is required that each laboratory assesses the consequences of the use of the different
functions prior to the attribution of the access codes.
Make sure that the STA Compact® is positioned in such a way that it can be easily unplugged.
It is absolutely necessary to check that all covers, doors and drawers of the STA Compact® (listed
below) are closed before power is turned on. During operation, it is absolutely necessary to close
them as soon as any handling requiring them to be opened is finished.
- right side door
1-5
0931033L - July 2009
1 REFERENCE MANUAL
Symbols and warnings
The use of this symbol implies that the analyzer must not be disposed of with
household waste; that it must be subject to separate collection and that it has been
put on the market after 13 August 2005.
For more information about the appropriate disposal of the analyzer (waste sorting, collection,
processing of analyzer waste) contact the local authorities, the manufacturer or the distributor who
sold this analyzer.
1-6
0931033L - July 2009
REFERENCE MANUAL 2
Table of contents
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1 Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.2.1 Installation Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.2.2 Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2-i
0931033L - July 2009
2 REFERENCE MANUAL
Table of contents
2-ii
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
2 Introduction
2.1 Definition
The STA Compact® system is a medical device for in vitro diagnosis, composed of a laboratory
analyzer and a software designed to be used with consumables and reagent products.
The STA Compact® system is designed to perform in vitro tests for diagnosis and monitoring of
pathologies linked to hemostasis.
2.2 Installation
Power Supply:
2-1
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
The oscillation amplitude is constant when the environment has a constant viscosity.
The oscillation amplitude decreases when the environment viscosity increases.
Legend:
1 Emitting measurement coil 2 Transmitting measurement coil
3 Motorization coils 4 Ball
5 Cuvette
2-2
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
The intensity of the magnetic field varies depending on the tests to be carried out (PT, APTT...) and
on the expected clot.
The detection system of the oscillation amplitude variations is based on two measurement coils.
The transmitting coil emits an electro-magnetic field. The signal received by the receiver coil depends
on the ball position in the cuvette.
An algorithm uses these magnetic field variations to calculate the oscillation amplitude variations and
to accurately determine the clotting times.
.
2 Clotting
2-3
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
The effect of stray light (Ip) is eliminated by taking two fairly close measurements of the light
transmitted:
I1 = I + Ip (first measurement which includes incident light and stray light),
I2 = Ip (second measurement while blocking the incident light, corresponds to the stray light).
When I2 is substracted from I1, the result is I, which is only the light transmitted from the incident
light. Ip is assumed to remain constant between the two measurements.
Incident light is provided by a tungsten-halogen lamp, and is made monochromatic by passing through
a 405 nm, or 540 nm, interference filter placed in a mobile filter holder.
These items are inside the optical module. A system of optical fibers carries the monochromatic light
from the optical module to the measurement heads. Another set of optical fibers carries the
transmitted light from the measurement head to the photometry measurement board (see figure
hereafter).
2-4
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
1 Measuring head
2 Reference fiber
3 Sensor
6 RS 232
7 PC
From the measurements of optical density that are obtained, the Beer-Lambert law is applied:
A =H l C
2-5
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
All the optical density measurements are done in relation to the reference fiber.
2-6
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
Legend:
1 Transparent panel
2 Right front door (Access to the Syringe and to the Cuvette Roll)
4 Product Drawer
5 Sample Drawer
6 Barcode Reader
8 Keyboard
2-7
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
RISK OF INJURY
Do not touch the moving parts.
2 Barcode Reader
4 Cuvette Roll Drawer (To change the Cuvette Roll, see chapter 3)
8 Product Drawer
9 Sample Drawer
10 Electronic Boards
2-8
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
2 Air Filter
3 Air Vents
2-9
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
Legend:
1 Not used
2 Printer Connector
3 Monitor Connector
4 Keyboard Connector
Switch the devices off (STA Compact®, monitor and printer) when connecting the STA Compact®
to the monitor, the printer or the keyboard.
All the devices connected to the STA Compact®must comply with standards IEC 60950, EN
55022 class B and EN 55024.
2-10
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
Switch off the STA Compact®, the monitor and the printer before replacing the fuses.
Replace the fuses with fuses sharing the same reference.
Legend:
2-11
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
RISK OF INJURY
Do not touch the moving parts.
3 Intermediate Reservoir
4 Vacuum Reservoir
2-12
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
6 Peltier Reservoir
2-13
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
The STA Compact® is an automated laboratory analyzer intended for hemostasis analysis. This
analyzer can carry out clotting assays (clotting time measurements) as well as chromogenic and
immunological assays on plasma samples.
The primary sample tubes and the dilution buffers are loaded in the sample drawer. The position of
each sample tube is automatically detected by the Positive Identification System.
The control plasma vials, the calibration plasma vials as well as the reagent vials are loaded in the
product drawer where the temperature is monitored between 15°C and 19°C by a system based on
Peltier elements. The position of each vial is also automatically detected by the Positive Identification
System.
Sample plasmas, control plasmas as well as calibrator plasmas are pipetted by needle No.1 of the
pipetting head, then they are added in a cuvette at an incubation position.
Reagents to be added before the first incubation are pipetted by needle No.2 of the pipetting head,
then they are distributed in the related cuvette in incubation position.
Reagents to be added after the first incubation (mainly the start reagents) are pipetted by needle
No.3 of the pipetting head. If a preheating to 37°C is necessary, the reagents are moved from needle
No.3 up to heating tube No.3 (this one being just above the tip). Then, with or without preheating,
those reagents are added in the related cuvette in measurement area.
A level detection system on each needle ensures accurate and precise dispensing of fluid volumes.
Carry over is minimized by rinsing the interior as well as the exterior of the needles, each in its own
well.
Test cuvettes are loaded onto the STA Compact® from a roll of 1,000 cuvettes. At the cuvette loading
station, they are placed one-by-one in a shuttle. This shuttle is then moved to the measurement
station by a system based on a pneumatic jack.
At this station, the suction head picks up the cuvette and transfers it to the incubation zone. This
same head then transfers the cuvette from the incubation zone to the measurement zone then from
the measurement zone to the cuvette disposal bag.
The incubation zone and the measurement zone are included in the same block which is monitored
at 37°C r 0.5°C by a system based on Peltier elements.
These operating principles enable the analyzer to process each test independently, and offer a
throughput that is virtually independent of the test time.
2-14
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
RISK OF INJURY
RISK OF BIOLOGICAL CONTAMINATION
Do not touch the drawer while opening or closing.
Observe the usual precautions for biohazardous material while manipulating tubes or cleaning the
sample drawer (for instance: wear disposable gloves).
4 Status LED:
2-15
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
RISK OF INJURY
RISK OF BIOLOGICAL CONTAMINATION
Do not touch the drawer while opening or closing.
Observe the usual precautions for biohazardous material while manipulating tubes or cleaning the
product drawer.
2-16
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
2-17
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
Operating conditions:
• For indoor use
• Impact resistance: according to current standards
• Temperature: + 15°C to + 32°C(+ 59°F to + 89.5°F)
• Humidity: 20 % to 80 %
• Altitude: < 2000 m
Environnement :
• Sound level: 70 dBA (maxi.)
• Heat dissipation when ambiant temperature is 20°C:
- 230V - 100V: 1400 Whr (4778 BTU)
- 115V: 1000 Whr (3413 BTU)
2-18
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
Power supply:
• Voltage and tolerance: 230 V ( 10 %) - 115 V ( 10 %) - 100 V ( 10 %)
• Frequency and tolerance : 50/60 Hz (47 to 65 Hz)
• Rating: 1400 W at 230 V - 100 V
1000 W at 115 V
• Number and type of power sockets: 1 power sockets 16A
• 2 main fuses (refer to paragraph 2.5.4 in this chapter)
• 8 secondary fuses (refer to paragraph 2.5.4 in this chapter)
• Fuses protection: yes
• Inverter needed for micro-cuts: no
• Results of sudden loss of supply: tests in progress lost, possibility of change in hard disk.
Classification :
• Type of protection against electric shock: class I equipment
• Degree of protection against harmful ingress of water: ordinary equipment (enclosed equipment
without protection against ingress of water)
• Pollution level: 2
• Recommended disinfection method: see chapter 4.6
• Degree of safety of application: equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide
2-19
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
All the devices connected to the STA Compact® must comply with standards IEC 60950, EN
55022 class B and EN 55024.
Product management:
• 45 Positions for various vial sizes:
Measurement principle:
• Analytical modes: photometry (colorimetry and immunology) (not available on STA
Compact®CT), clotting
• Clotting: change in medium’s viscosity detected by electromagnetic sensors
2-20
0931033L - July 2009
REFERENCE MANUAL 2
Introduction
Test settings:
• 80 user defined test settings
Pipetting Unit:
• based upon 250 l Hamilton syringe
Cuvettes:
• The cuvette is used for incubation and measurement
• Rolls of 1,000 cuvette with predispensed stainless steel ball
- Maximum usable volume: 400 Pl
- Minimum usable volume: 150 Pl, clotting assays and 250 Pl, photometry assays
- Identification by bar code reader
Patents:
• Allows swinging movement of the ball (STAGO patent)
Patent number :
- FR 8718348
- FR 8809151
- US 4918984
- JP 1939942
- EP 0325874
Test Performance:
• Intra-assay Reproducibility with Diagnostica Stago Products:
Coefficient of Variation
2-21
0931033L - July 2009
2 REFERENCE MANUAL
Introduction
Connexion:
• Serial interface RS232C
• ASTM protocol
LCD screen
Technology LCD
Security IEC60950
Netting <0.30 mm
Horizontal view angle > 120 °
Vertical view angle >90 °
Luminosity > 200 cd/m²
Contrast 100 : 1 to 400 : 1
<40 ms
Response time
Keyboard:
• Cherry keyboard (AZERTY or QWERTY depending on the language)
• Dimensions: 210 x 490 mm(8.3 x 19.3 in.)
Approved: CSA, UL.
Printer:
• List of compatible printers: contact Diagnostica Stago or one of its official distributors
• The printer cable must meet the printer notice requirements
2-22
0931033L - July 2009
REFERENCE MANUAL 3
Table of contents
3-i
0931033L - July 2009
3 REFERENCE MANUAL
Table of contents
3-ii
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
ACTIONS RESULTS
" Switch the ON/OFF switch of The ON/OFF switch lamp on the
STA Compact< to the ON position (I). This STA Compact< lights up (green color) as
switch is located on the left side of the well as the one on the monitor.
STA Compact< (see chapter 2.5.4)
The following messages pop up:
" Switch the monitor on
Please wait during loading databases. It could
take 1 minute ...
" Wait for 25 minutes so that the product STA Compact< is ready for use.
drawer temperature is stabilized
3-1
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
ACTIONS RESULTS
" To proceed, press the key (Continue) The TEST PANEL screen is displayed (see
description chapter 3.2).
The message: “Desorb Solutions (for
cleaning the needles) are missing”
appears.
The date modification can also be made from the SYSTEM STATUS screen (see description
chapter 10.1.2.2).
ACTIONS RESULTS
" To correct the date, press the K key to The following window is displayed:
reach Modify the System Date, then
confirm with the key TYPE NEW DATE
(Format mm * dd * yyyy)
The separator (*) and the date display
format are defined within the Global
Options menu (see chapter 10.2.3.4).
3-2
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
ACTIONS RESULTS
" Type the new date according to the The new date is displayed in the upper right
displayed format and separator, then corner of the screen.
confirm with the key. The GLOBAL VERIFICATION DONE window
In this example: type the month (2
is displayed:
numbers), the displayed separator, the day
(2 numbers), the displayed separator and
the year (4 numbers), then confirm with GLOBAL VERIFICATION DONE
the key Modify the System Date
Modify the System Time
Delete the Working File
Continue
with the cursor on Modify the System
Date
The time modification can also be made from the SYSTEM STATUS screen (see description
chapter 10.1.2.3).
ACTIONS RESULTS
" To correct the time, press the K key to The following window is displayed:
reach Modify the System Time, then
confirm with the key TYPE NEW TIME
(Format hh : mm : ss )
The separator and the time display format
are defined within the Global Options
menu (see chapter 10.1.2.4).
3-3
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
ACTIONS RESULTS
" Type the new time according to the The new time is displayed in the upper right
displayed format and separator, then corner of the screen.
confirm with the key. The GLOBAL VERIFICATION DONE window
In this example: type the hour (2 digits), the
is displayed:
displayed separator, the minutes (2 digits),
the displayed separator and the seconds
GLOBAL VERIFICATION DONE
(2 digits), then confirm with the key
Modify the System Date
Modify the System Time
Delete the Working File
Continue
with the cursor on Modify the System
Time.
Verify that all the “confirm” or “to confirm” files have been transmitted and/or printed before doing
this operation.
There is no selection available here: all the files of the Working File will be deleted if this option is
chosen. To partly delete the Working File, access the TEST PANEL, then see chapter 8.6.
ACTIONS RESULTS
" To erase all the files of the Working File, The following window appears:
press the K key to reach Delete the
Working File, then confirm with the key DELETION OF THE WORKING FILE
Please type Y to confirm
3-4
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
ACTIONS RESULTS
" Press the Y key to confirm deletion of the The GLOBAL VERIFICATION DONE window
Working File is displayed:
3-5
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
Légende :
1 Available keys:
IJ Use these keys to display the following tests (J) or the previous ones (I)
KL Use these keys to display the following files (down: L) or the previous ones (up:K)
To proceed with patient file processing: to add/delete/rerun the tests, etc. (see
description chapter 8.3)
F1 To request opening of the sample drawer so long as the Main Menu is not displayed
(Key not documented on the screen).
F2 To request opening of the product drawer so long as the Main Menu is not displayed
(Key not documented on the screen).
3 Complementary information (Name, First name, ... see description chapter 10.2.3.3) for the file
on which the cursor is positioned.
Grey: File for which a tube has been loaded but the list of tests is unknown (waiting for
downloading from the host computer)
Red: An error has been detected on one or several tests (see description in zone 8 below)
5 Position occupied by each tube within the sample drawer. The P symbol is displayed after the
indication of the tube position if the micro tube option has been chosen.
3-6
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
7 Primary units for reporting results defined in the TEST SETUP screens (see chapter 5.1.1.7)
8 Zone for results display: several codes indicate the test status:
- ”---.-” displayed in red: Test rerun, for example following a technical error (no ball
detected...), less than minimum time (< V min.) or greater
than maximum time (>V max.)
- Result displayed in dark red: An error has been detected for this result
- Result displayed in light red: Result obtained with an out of range quality control or with a
quality control not done
- "Block" displayed in dark Test blocked from the FILE MODIFICATION screen (see
red: description chapter 8.7)
- "> V max." displayed in dark For clotting tests; raw measure above the maximum time
red: defined in the TEST SETUP screens (see description,chapter
5.1.1.6).
For calibration modes graphic 2nd order polynomial or
graphic 3rd order polynomial, raw data greater than raw data
obtained for the highest calibration point.
For colorimetric tests (not available on STA Compact®CT),
raw measure above the calibration maximum O.D.
For immunological tests (not available on STA Compact®CT)
in kinetic 2-points with calibration mode: 2nd order
polynomial or 3rd order polynomial, after redilution with the
parameter “Dilution 1" (if over limit), (see description chapter
5.1.1.6), raw measure obtained once again out of ranges.
This message can and must be interpreted only after the
colorimetric graph and the raw measure of the first point as
defined in the test settings are examined.
3-7
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
- "< V min" displayed in dark For clotting tests, raw measure below the minimum time
red: defined in the TEST SETUP screens (see description chapter
5.1.1.6).
For calibration modes: graphic 2nd order polynomialor
graphic 3rd order polynomial, raw data below the raw data
obtained for the lowest calibration point.
For colorimetric tests (not available on STA Compact®CT),
raw measure below the calibration minimum O.D.
This message can and must be interpreted only after the
colorimetric graph and the raw measure of the first point as
defined in the test settings are examined.
- "Lin" displayed in dark red: Insufficient linearity (see description, chapter 5.1.1.6) for
colorimetric or immunological tests (not available on STA
Compact®CT).
Note: For results displayed in TEST PANEL, the printout limits have not been applied (see description
chapter 5.1.3.3).
9 The “BLOCKED SAMPLE PIPETTING” message is displayed in white on a red background at the
bottom of the screen when the sample pipetting is blocked whether:
- Voluntarily by the operator (Stop Sample Pipetting function, see description chapter 10.1.5.1)
- Or automatically by the STA Compact< after the consistency check; if the required assays cannot
be done.
Note: The bar code tracking option is now active for the cuvette rolls.
To change a roll of cuvettes, the operator should:
" First, access the cuvette roll change window
3-8
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
" Second, identify the new cuvette roll by means of its bar codes
" Third, remove the empty cuvette roll
" Fourth, install the new cuvette roll
The operating mode is detailed in a descriptive note delivered with the analyzer (see Fig.2).
Cuvette roll loading must be followed by cuvette disposal bag replacement (see below chapter 3.4).
The error window is systematically displayed when the theoretical counter for the number of
cuvettes reaches the zero value.
3.3.1.1 Access the cuvette roll change Window from the Error Window
3-9
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
ACTIONS RESULTS
" Press the Space bar The cuvette roll change window appears:
3.3.1.2 Access the cuvette roll change Window from the TEST PANEL
ACTIONS RESULTS
3-10
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
ACTIONS RESULTS
ACTIONS RESULTS
" One after the other, pass the two cuvette When a bar code passes in front of the
roll bar codes in front of the bar code reader, the system beeps.
reader. When a bar code is read, the corresponding
area of the window is filled in.
" In case of failure of the bar code reading, The two bar codes are displayed.
type in the data printed under the bar
codes
3-11
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
ACTIONS RESULTS
" In the field Number of cuvettes in the The following window is displayed:
cuvette roll, indicate the number of
cuvettes remaining in the cuvette roll if it is CUVETTES: CHANGE OF THE CUVETTE ROLL
less than what appears in the window. Please remove the empty roll
" Validate with F10. Space bar : Confirm Esc : Quit
3-12
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
ACTIONS RESULTS
" Turn the reel (arrow direction) until the The cuvettes are in the loading rail.
cuvettes are engaged correctly in the
loading rail
3-13
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
ACTIONS RESULTS
" Push back the cuvette roll drawer If the STA Compact< is in stand-by, the
following window is displayed:
" Close the right side door (step 10, refer to
help diagram No.1) then confirm with the
F10 key CUVETTE ROLL CHANGE
TESTING
= Begin
Space Bar = EMERGENCY STOP
" Press the key to start automatically the The following window is displayed:
loading test
Note: In case of problems, press the CUVETTE ROLL CHANGE
Space Bar to stop the roll motor. TESTING
TEST IN PROGRESS
SPACE BAR = EMERGENCY STOP
Then, the following window is displayed:
After each cuvette roll replacement, the cuvette disposal bag shall be systematically replaced.
3-14
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
Procedure:
Step 1: Remove the cuvette disposal drawer
Step 2: Pull up the folded portion of plastic bag so that the bag can be closed
Step 3: Remove the red band from the plastic bag and close the top of the plastic
bag tightly by tying a knot with the red band
Step 4 : Lift the plastic bag completely off the cuvette disposal drawer and
discard in accordance with local existing regulations.
Do not interfere with the function of the switch designed to detect the presence of the cuvette
disposal drawer.
" Press the F10 key to confirm the replacement of the cuvette disposal bag.
3-15
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
3-16
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
3.5.1.1 Access the WASHING SOLUTION LOADING Window from the Error Window
ACTIONS RESULTS
" Press F1 key to change the Washing The following window is displayed:
Solution.
3-17
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
3.5.1.2 Access the WASHING SOLUTION LOADING Window from the TEST PANEL
ACTIONS RESULTS
" Press the W key (Washing Solution) The following window appears:
" Or press the L key to reach Washing
Solution, then confirm with the key
3-18
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
The empty STA< Cleaner Solution bottle becomes the new used liquid container. The message
“Please, EMPTY then displace this bottle” is displayed.
" Carefully pull out the empty STA< Cleaner Solution bottle (the one on the left)
" As soon as possible, remove the plug from the bottle. To avoid any possible contamination of
tubing, place the end of the tubing in a receptacle
" Empty the STA< Cleaner Solution bottle
3-19
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
" Insert the plug of the right position into the empty bottle
" Then, replace the empty bottle in the right position. The closing of the switch detecting the
presence of the bottle should be heard
Do not interfere with the function of the switch designed to detect the presence of the used liquid
container.
" Press the F10 key to confirm the end of this operation
RISK OF INJURY
Do not swallow the STA< Cleaner Solution (in-vitro usage only).
" Remove the cap of the bottle of the STA< Cleaner Solution
" Insert the system cap (left position) into the STA< Cleaner Solution bottle
" Then replace the bottle in the left position. The switch which detects the presence of the
container will close audibly
" Press the F10 key to confirm the end of this operation
" If using a new bottle: confirm the indicated volume (2500 ml) with the key
Otherwise, enter the remaining volume, then confirm with the key
" Close the right side door of the analyzer
3-20
0931033L - July 2009
REFERENCE MANUAL 3
Preparing the STA Compact<
3.6.1 Reminders
The STA Compact®’s monitor switches off automatically after 20 minutes whether the STA
Compact® analyzer is executing measurements or not. If an error is detected then, the STA
Compact®’s monitor comes back on to display the error message.
When all the tests are completed, the STA Compact® goes to stand-by. All automatic functions stop
except for the temperature regulations.
It is recommended not to switch off the analyzer or to unload the products. The STA Compact® is
then able to automatically manage the stability of the products.
To temporarily stop sample pipetting, the Stop Sample Pipetting function should be used (see
description chapter 10.1.5.1).
ACTIONS RESULTS
3-21
0931033L - July 2009
3 REFERENCE MANUAL
Preparing the STA Compact<
ACTIONS RESULTS
END OF APPLICATION
Saving ...
Please wait
DO NOT SWITCH OFF
END OF APPLICATION
Recording complete
you can switch the system off
" Switch the ON/OFF switch of the The STA Compact<, the monitor and the
STA Compact< to the OFF position (O). printer are switched off.
This switch is located on the left side of the
STA Compact< (see chapter 2.5.4)
" Switch OFF the monitor
" Switch OFF the printer.
3-22
0931033L - July 2009
REFERENCE MANUAL 4
Table of contents
4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
4-i
0931033L - July 2009
4 REFERENCE MANUAL
Table of contents
4-ii
0931033L - July 2009
REFERENCE MANUAL 4
Table of contents
4-iii
0931033L - July 2009
4 REFERENCE MANUAL
Table of contents
4-iv
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
4 Maintenance
Quarterly Replacement of the air filter at the 4.4 Filters air 27420
back of the analyzer (x2)
4-1
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
For example:
- Bleach at 2.6% of active chlorine:
Mix 1 part of 2.6% active chlorine solution with 6 parts of water.
- Bleach at 6.25% of active chlorine:
Mix 1 part of 6.25% active chlorine solution with 16 parts of water.
- Bleach at 9.6% of active chlorine:
Mix 1 part of 9.6% active chlorine solution with 25 parts of water.
or
Mix 1 part of bleach at 9.6% of active chlorine with 3 parts of water to obtain bleach at 2.6% of active
chlorine. Then mix 1 part of bleach at 2.6% of active chlorine with 6 parts of water.
4-2
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
Make sure that the correct assembly direction is observed: the word EXTERIEUR should face
outwards.
4-3
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
Legend:
2 Air filter, in the optical module (not available on STA Compact< CT)
4-4
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
4-5
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
Access the Needle Purge menu: The NEEDLE PURGE screen appears:
" Press the key Needle Purge The following windows appear:
Checks in progress
Please wait...
Positioning in well
Please wait...
Open Drawer
Please wait...
The product drawer is opening.
The pipetting head moves forwards.
4-6
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Close Drawer
Please wait...
The product drawer is closing.
The pipetting head moves above the
washing wells and the 3 needles go down in
their wells.
" Press the keys ALT and PrtSc Printout of the DATES OF RECENT
MAINTENANCE screen.
" Press the K key to Return to the Back to User Maintenance menu.
Previous Menu, then confirm with the
key
4-7
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Run quality controls for the routine tests If the quality control results are consistent
" Compare results with the previous ones with the previous ones, STA Compact® can
be used.
If not, redo the procedure.
In case of problem, call your local authorized
Service Representative.
4-8
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
C C
R = ------------- or N = ---- – 1
N+1 R
where
R = concentration after dilution
C = concentration of the ethanol
N = number of added parts of water
For example:
ACTIONS RESULTS
4-9
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
Legend:
2 Motorization coil
4-10
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
" Clean each measurement channel and each incubation cell with swabs slightly moistened with
ethanol between 20% and 40%
Do not use solvents like Acetone or Trichlorethylene as they may damage the optical fibers.
" Clean the black panel of the measurement plate with a soft cloth or paper tissue slightly
moistened with warm water, then wipe with a dry cloth. Do not use ethanol on the measurement
plate
ACTIONS RESULTS
" Run quality controls for the routine tests If the quality control results are consistent
" Compare results to the previous ones with the previous ones, STA Compact® can
be used.
If not, redo the procedure.
In case of problem, call your local authorized
Service Representative.
4-11
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
Legend:
1 Peltier Reservoir
" If the liquid level is below the mark, switch off the STA Compact®
" Open the Peltier Reservoir
" Fill it with the cooling liquid (Ref.: 38640)*
" Switch on the STA Compact®
RISK OF INJURY
Harmful product that contains methanol !
4-12
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilites
Return to Previous Menu
with the cursor on Rinsing Pump.
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
4-13
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Access the Suction Tip Replacement menu: The following window is displayed:
" Press the L key to reach Suction Tip
Replacement, then confirm with the SUCTION TIP REPLACEMENT
key Initialization in progress
Please wait...
The Suction Tip Replacement menu
appears:
Access the Suction Tip Replacement menu: The following windows are displayed:
" Press the key Suction Tip
Replacement SUCTION TIP REPLACEMENT
Home position in progress...
4-14
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
4-15
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the L key to Suction Tip Test then The following window appears:
confirm with the key
SUCTION TIP REPLACEMENT
Home position in progress
Please wait...
Test of the suction tip on the measurement
plate.
The following window appears:
" Press the Space Bar to proceed The Suction Tip Replacement menu
appears
" Press the K key to Return to Previous The Maintenance menu appears.
Menu, then confirm with the key
" Press the K key to Return to Previous The User Maintenance menu appears.
Menu, then confirm with the key
4-16
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
It is recommended that this replacement be performed monthly, although this frequency can be
adjusted depending on the laboratory workload. In any event, do not exceed the software-
controlled limits, the latter being shown in the SYSTEM STATUS screen, see description in
chapter 10.1.2.1.
If the entire syringe has to be replaced, use glass syringe Cat. No.: 27538 supplied by Diagnostica
Stago only. All other syringes will give the wrong results.
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
4-17
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
Tip or Syringe
Return to Previous Menu
with the cursor on Tip or Syringe.
4-18
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Take a new syringe tip or a new syringe The following windows appear:
according to the element which has to be
changed REPLACEMENT OF TIP OR SYRINGE
" Press the F10 key to proceed Home position in progress...
Please wait...
" Follow the steps shown in help diagram No. 3 (see Figure 3), this help diagram can be found
inside the right front door in a plastic sleeve. For analyzers with serial number above xxx1219, it
can be found inside the right side door.
Procedure:
" Step 1: Open the right front door of the STA Compact®
" Step 2: Lower the syringe piston and loosen the bottom thumb screw
" Step 3: Unscrew the syringe from its support
" Step 4: Remove the syringe from its clamp
" Step 5:
- 1. Pull the piston out of the syringe
- 2. Remove the O-ring
- 3. Remove the Teflon tip with a blunt instrument
" Step 6: Place the new tip on a flat surface as shown in the diagram and press the syringe piston
fully home into the new tip
" Step 7:
- 1. Dip the Teflon tip in distilled water
- 2. Reinstall the piston in the syringe
- 3. Replace the O-ring
4-19
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
" Step 8: Reposition the bottom of the syringe in the clamp while holding it vertically
" Step 9: Screw the syringe on its support by hand, still holding it vertically
" Step 10: Once again position the piston syringe in the upper position
" Step 11: Hand-tighten the thumb screw while supporting the clamp with your free hand
Do not close the right front door since a test is going to be performed on the syringe by the
software.
4-20
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the F10 key to validate this The following window appears:
procedure
Reservoir filling in progress
Please wait
The following screen appears:
" Press the F10 key to proceed The following window appears
" Observe the syringe during the purge and
press the Alt + F10 keys simultaneously in REPLACEMENT OF TIP OR SYRINGE
case of a problem Fluidics circuit purge in process,
please monitor the syringe.
Alt-F10: Emergency Stop
The following windows appear:
Step 12: Close the right front door of the STA Compact®
ACTIONS RESULTS
" Press the Space Bar to proceed In the SYSTEM STATUS screen, the syringe
timer is reset to 100%.
The Syringe Tip/Syringe Replacement
menu appears.
4-21
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the L key to reach Return to The Maintenance menu appears.
Previous Menu then confirm with the
key
" Press the keys Alt and PrtSc Printout of the DATES OF RECENT
MAINTENANCE screen.
" Press the K key to reach Return to The User Maintenance menu appears.
Previous Menu, then confirm with the
key
" Run quality controls for the routine tests If the quality control results are consistent
" Compare results to the previous ones with the previous ones, STA Compact® can
be used.
If not, redo the procedure.
In case of problem, call your local authorized
Service Representative.
4-22
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
Cat. No.: 27420, air filter for rear panel (2 per bag).
Cat. No.: 26538, filter for optical module (see figure hereafter), not available on STA
Compact®CT
OR
Cat. No.: 38125, filter for optical module V3 type (see figure hereafter), not available on STA
Compact®CT
" Switch the STA Compact® off (see Power-off procedure chapter 3.6.2)
" Remove the air filter on the back of the STA Compact® and discard
4-23
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
" Open the right side door of the STA Compact®, remove the optical module air filter (not
available on STA Compact< CT) and discard
Fig. 7 - Air Filter, in the Optical Module (not available on STA Compact< CT)
Make sure that the correct assembly direction is observed: the word EXTERIEUR should face
outwards.
4-24
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
4-25
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
4-26
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Access the Needle Purge menu: The NEEDLE PURGE screen appears:
" Press the key Needle Purge The following windows appear:
Checks in progress
Please wait...
Positioning in well
Please wait...
Open Drawer
Please wait...
The product drawer is opening.
The pipetting head moves forwards.
Close Drawer
Please wait...
The product drawer is closing.
The pipetting head moves above the
washing wells and the 3 needles go down in
their wells.
OR
" Press the F2 key Purge of Needle No.2.
OR
" Press the F3 key Purge of Needle No.3.
4-27
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the keys ALT and PrtSc Printout of the DATES OF RECENT
MAINTENANCE screen.
" Press the K key to Return to the Back to User Maintenance menu.
Previous Menu, then confirm with the
key
" Run quality controls for the routine tests If the quality control results are consistent
" Compare results with the previous ones with the previous ones, STA Compact® can
be used.
If not, redo the procedure.
In case of problem, call your local authorized
Service Representative.
4-28
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
4-29
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
Needle No. 1
Needle No. 2
Needle No. 3
Return to Previous Menu
with the cursor on Needle No. 1.
" Use the L key to select the needle which The following window is displayed:
has to be changed.
" Confirm with the key NEEDLE REPLACEMENT
For this operation, you must have:
- a new needle
- the mapping cuvette
F10: Continue - Esc: Abort
" Take a new needle and the mapping Home positions in progress.
adjustment cuvette
The following window is displayed:
" Press the F10 key
NEEDLE REPLACEMENT
Drawer is going to open
Press any key
to proceed
4-30
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
NEEDLE REPLACEMENT
once the transparent panel is raised,
you can change the needle according
to the procedure described in the
manual. When this operation is
finished, validate by pressing F10.
F10: Validate - Esc: Abort
- Observe the proper precautions (for instance: wear disposable gloves) for handling
biohazardous materials, in accordance with local regulations.
- Observe a procedure in accordance with local regulations to discard the potentially contaminated
needles.
4-31
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
Legend:
" At the bottom of the needle holder assembly, fully unscrew the knurled nut (see Fig.7 below)
" Remove this nut
" Lower the needle-tubing assembly by gently pushing on the top of the tubing
" For Needle No.3: Pull on the needle to disconnect the small tubing from the end of the heating
tube
" Remove the needle-tubing assembly
" Needle with a yellow knurled nut:
- Dispose of the needle-tubing assembly in accordance with the locally applicable regulation for
collection needles.
- Reuse the knurled nut.
4-32
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
Legend:
1 Needle 2 Nut
" If the new needle is provided with no nut: use the yellow nut of the needle that has just been
dismantled
" If the new needle is provided with a metal mandrel:
- remove the mandrel from inside the needle
- discard this mandrel
" Push the tubing into the needle holder from below
" Check that the tubing is not stuck in the needle holder
" Fully retighten the knurled nut by hand while holding the needle holder with the other hand
" Reconnect the tubing to the metal junction tubing at the top of the head (see figure 8 above)
4-33
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
Legend:
3 Needle N°3
" If the new needle is provided with no nut: use the yellow nut of the needle that has just been
dismantled
" If the new needle is provided with a metal mandrel:
- remove the mandrel from inside the needle
- discard this mandrel
" Remove some of the thermal grease in order to clearly see the end of the heating tube
" Make sure that there is no thermal grease on the end of the heating tube
" Attach the tubing of the needle to the end of the heating tube fully home (see figure 10 above)
" Push the needle-tubing assembly fully home into the needle holder
" Fully retighten the knurled nut by hand
4-34
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
" Check the connections of the level detection cable from the needle holder to the electronic
circuit board
" Remove the paper towel and discard it in accordance with the locally applicable regulation
Replacement is completed
ACTIONS RESULTS
" Press the F10 key to validate this Needle replacement is validated.
procedure
The following window is only displayed if the
level detection for the required arm is
managed:
NEEDLE REPLACEMENT
Level detection test.
Please use finger to touch the needle
you have just installed.
You have 19 seconds...
" Take the needle with your fingers (ensure The following screen is displayed whether
the needle is new and not contaminated) the level detection is managed or not:
NEEDLE REPLACEMENT
Please follow these instructions:
There are only 20 seconds to perform this 1° Place the mapping cuvette in
operation. the incubation position No.16.
2° Lower the transparent panel.
F10 to continue
4-35
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press any key to close the product drawer The Drawer is closing
The following window appears:
NEEDLE REPLACEMENT
Please raise transparent panel
to retrieve mapping cuvette.
Then lower transparent panel.
F10 to proceed
" Press the F10 key to proceed The Needle Replacement menu appears
" Press the K key to Return to Previous The Maintenance menu appears.
Menu, then confirm with the key
4-36
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Access the Needle Purge menu: The NEEDLE PURGE screen appears.
" Press the key Needle Purge Home positions.
The needles go to their own washing wells
" Press the keys ALT and PrtSc Printout of the DATES OF RECENT
MAINTENANCE screen.
" Press the K key to Return to the Back to User Maintenance menu.
Previous Menu, then confirm with the
key
4-37
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Run quality controls for the routine tests If the quality control results are consistent
" Compare results to the previous ones with the results run previously, the STA
Compact® is now ready for use.
If not, redo the procedure.
In the event you encounter a problem, call
your local authorized Service
Representative.
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
4-38
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
Access the Suction Tip Replacement menu: The following window is displayed:
" Press the L key to reach Suction Tip
Replacement, then confirm with the SUCTION TIP REPLACEMENT
key Initialization in progress
Please wait ..
The Suction Tip Replacement menu
appears:
Access the Suction Tip Replacement menu: The following windows are displayed:
" Press the key Suction Tip
Replacement SUCTION TIP REPLACEMENT
Home position in progress
4-39
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Do not press the suction tip too hard: the suction tip test may fail.
4-40
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
Legend:
1 Suction tip
ACTIONS RESULTS
" Press the F10 key to validate the The following window appears:
replacement
SUCTION TIP REPLACEMENT
You have replaced the suction tip.
Please lower the transparent panel in
order to run automatic
tests that are necessary to validate
the installation.
F10 to validate the installation
4-41
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the Space Bar to proceed The Suction Tip Replacement menu
appears.
" Press the K key to Return to Previous The Maintenance menu appears.
Menu, then confirm with the key
" Press the keys ALT and PrtSc Printout of the DATES OF RECENT
MAINTENANCE screen.
" Press the K key to Return to Previous The User Maintenance menu appears.
Menu, then confirm with the key
4-42
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
4.5.5 Replacing the Halogen Lamp (not available on STA Compact< CT)
Cat.No.26699
ACTIONS RESULTS
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
4-43
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
Lamp Replacement
Lamp Test
Return to Previous Menu
with the cursor on Lamp Replacement.
" Press the key to Lamp Replacement The following window is displayed:
LAMP REPLACEMENT
For this operation, you must have:
- a new halogen Lamp
- a screwdriver
F10: Continue - Esc: Abort
" Take a Philips screwdriver and a new The following window is displayed:
halogen lamp
" Press the F10 key HALOGEN LAMP REPLACEMENT
You can replace halogen lamp
according to the procedure described in
the manual
Caution!
The lamp may be hot!
F10: Validate - Esc: Abort
4-44
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
RISK OF INJURY
4-45
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
4-46
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
Do not touch the new lamp with your bare fingers. Always use paper tissue or dry cloth when
handling a new lamp.
4-47
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the F10 key to validate the lamp The following window is displayed:
replacement
HALOGEN LAMP REPLACEMENT
You have replaced the lamp.
Please validate in order to
perform a photometric test
F10 to start test
LAMP REPLACEMENT
The procedure was successful
Space Bar to proceed...
The lamp maintenance timer is reset to
100% (see SYSTEM STATUS screen).
If a problem occurs during the procedure,
redo the procedure.
" Press the Space Bar to proceed The Halogen Lamp Replacement menu
appears.
4-48
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the K key to Return to Previous The Maintenance menu appears.
Menu, then confirm with the key
" Press the keys Alt and PrtSc Printout of the DATES OF RECENT
MAINTENANCE screen.
" Press the K key to Return to Previous The User Maintenance menu appears.
Menu, then confirm with the key
4-49
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
RISK OF INJURY
Legend:
1 Main fuses
2 Secondary fuses
The main fuses are in two fuse holders, near the ON/OFF switch on the left hand panel of the STA
Compact® (see figure 12).
" Switch off the STA Compact® (the green indicator switched off)
" Insert a screwdriver into the slot of the fuse holder. Fully unscrew the fuse holder (in direction
shown by the arrow, counterclockwise); the fuse holder will come free
" Remove and replace each fuse with a fuse sharing the same Catalog Number (see above)
" Reinstall each fuse holder by screwing it fully home (reverse arrow direction, clockwise)
" Switch ON the STA Compact® (green indicator switched on)
4-50
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
RISK OF INJURY
The secondary fuses are on the left panel of the STA Compact® (see Fig. 12 -).
" Switch off the STA Compact®
" If necessary, unscrew the protective fuse plate
" Insert a screwdriver into the slot of the fuseholder containing the fuse to be replaced
" Press in fully and rotate in the direction shown by the arrow (counterclockwise): the fuseholder
will come free
" Remove and replace teach fuse with a fuse sharing the same Catalog Number (see above)
" Reinstall the fuseholder by turning in the reverse direction (clockwise)
" If necessary, screw the fuse protective plate back on
" Switch ON the STA Compact®
4-51
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
Cat.No. 27458
4-52
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
Legend:
" Remove the old filter (avoid STA<-Cleaner Solution drop to fall on the electrical parts).
See their location on figure 13.
" Dispose of the filter in accordance with local existing regulations.
" Fit the new filter by positionning it according to figure 13.
" Put the various parts back in place.
" Switch on the STA Compact® and press the Delete key to access the User Maintenance
menu.
" Proceed with the needle purges (see below), those are very important to eliminate the air
bubbles.
4-53
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
Needle Purge
Needle Replacement
Syringe Tip / Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
Checks in progress
Please wait...
Positionning in well
Please wait...
4-54
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the K key to Return to the Back to the User Maintenance menu
previous menu, then confirm with the
key
In order to avoid any risk of contamination and in compliance with international laws and standards
concerning health and security at the working place:
Before any intervention of a field service engineer, perform the decontamination
procedure of the analyzer.
4-55
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
4-56
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the Esc key " The Main Menu appears:
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
ACTIONS RESULTS
" Switch ON the STA Compact® The following window and message appear:
4-57
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the Delete key to access the User The User Maintenance menu appears:
Maintenance menu
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
ACTIONS RESULTS
" Press the Alt + F10 keys simultaneously The User Maintenance menu appears:
Rinsing Pump
Bar Code Test
Printer
Saving
Maintenance
Photometry Graphics
Error History
Host Computer Communications
Data Update
Utilities
Return to Previous Menu
with the cursor on Rinsing Pump.
4-58
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the key Rinsing Pump The following window is displayed:
" Press the L key Verification of the The Rinsing Pump menu appears:
rinsing volume and validate with the
key Rinsing Needle No.1
Rinsing Needle No.2
Rinsing Needle No.3
Return to Previous Menu
with cursor on Rinsing Needle No.1.
ACTIONS RESULTS
" Press the key Rinsing Needle No.1 The following window appears :
" Or press the L key Rinsing Needle No.2
then confirm with the key VOLUME PUMP TEST
INFORMATIONS
" Or press the L key Rinsing Needle No.3
VIAL POSITION
then confirm with the key
Needle used : 1
Vial Position :
Pulse Number :
Pump Volume :
Drawer No.2 will open.
Press a key
to continue
1: Needle Number
4-59
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Place an empty vial in the product drawer The vial position is displayed.
in a position corresponding to its diameter A confirmation message appears at the
bottom of the window:
" Check the vial position then press the F10 A message requesting the closing of the
key to confirm drawer appears at the bottom of the screen:
4-60
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
ACTIONS RESULTS
" Press the K key to reach Return to The User Maintenance menu appears.
Previous Menu, then confirm with the
key
4-61
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the L key to reach Bar Code Test, The BAR CODE READER screen appears (see
then confirm with the key Fig.15 below).
Legend:
4-62
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Take a vial of any STA< product and When the vial is identified:
identify it with the bar code reader
- audible beep;
- display of the product identification
and name;
- display of a cross in the box “STA Codes”.
" Take a label used in the laboratory When the label is detected:
- audible beep;
- display of the identified value.
2BAR CODE READER screen displayed, bar code reader waiting for a test.
ACTIONS RESULTS
" Press the Esc key to return to the User The following window appears:
Maintenance menu
Do you want to return to the
previous menu
Y/N ?
Esc = Cancel
4-63
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
4.7.4 Printer
This program allows to check:
- The printer connection
- The printout of texts or graphics
For the printer maintenance: replacement of the toner cartridge or ribbon, refer to the printer
manual.
ACTIONS RESULTS
" Press the L key to reach Printer, then The Printer menu is displayed:
confirm with the key
Connection Checking
Test of the Characters
Graphics
Return to Previous Menu
with the cursor on Connection Checking.
ACTIONS RESULTS
" Press the key Connection Checking If the printer is correctly connected:
The message "Printer connected" is
displayed.
Otherwise:
An error message is displayed.
4-64
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to reach Test of the The PRINTER CHARACTER SET TEST screen
Characters, then confirm with the key appears (see figure 16 below).
ACTIONS RESULTS
" Press simultaneously the Control + I keys A window is displayed, it indicates that the
to print out this screen printer is connected.
The whole displayed screen prints out.
" Check the printout If problems on the printout occur, the user
should:
• Check printer configuration (switches) by
means of chapter 4.7.4.6.
• Check printer itself (paper, ribbon tape,
cleanliness, etc.).
" Press the Esc key to return to the Printer The Printer menu appears.
menu
4-65
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the L key to reach Graphics, then The GRAPHICS screen appears (see figure 17
confirm with the key below).
ACTIONS RESULTS
" Press simultaneously the Control + I keys A window appears indicating that the printer
to print out this screen is connected.
The whole displayed screen prints out.
" Check the printout If problems on the printout occur, the user
should:
• Check printer configuration (switches) by
means of chapter 4.7.4.6
• Check printer itself (paper, ribbon tape,
cleanliness, etc.).
" Press the Esc key to return to the Printer The Printer menu appears.
menu
4-66
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to Return to the previous The User Maintenance appears.
menu and then, confirm with the key
Emulation mode IBM Proprinter X24E Primordial, any other choice will lead to
unpredictable results in graphic mode.
4-67
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
4.7.5 Saving
This program allows to save or to load the following parameters:
Before saving, format the diskette with the STA Compact® (see below, chapter 4.7.5.5).
To avoid computer viruses, do not use the saving diskette on any other PC system.
When the parameters are loaded, saved data are transfered and data in the base are overwritten.
In order to avoid any risk of loss of data, any risk of wrong results:
Do not load parameters if you do not have the authorized Service Representative
agreement beforehand.
SAVING LOADING
4-68
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to reach Saving, then The SAVE/REREAD THE PARAMETERS screen
confirm with the key appears (see figure 18 below) with the
cursor on Selection of Items.
ACTIONS RESULTS
" Press the key to select the items The cursor goes to the Test
Configurations area.
A help window for selection appears at the
left bottom of the screen.
The following items can be selected:
- Test Configuration
- System Parameters
- Error Messages
4-69
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the F10 key to confirm the selection Selection of Items validated
4.7.5.3 Loading
" Select the items which have to be loaded (see chapter 4.7.5.2).
Loading brings about the deletion of the data of the selected files (Test Configurations, System
parameters and/or Error messages).
In order to avoid any risk of loss of data, any risk of wrong results:
- Do not load parameters if you do not have the authorized Service Representative
agreement beforehand.
- Never load Error messages: this option is exclusively for the authorized Service
Representative agreement. Make sure that the item ERROR MESSAGES has not been selected by
mistake.
ACTIONS RESULTS
" Press the L key to Load then confirm with The following windows will be displayed
the key according to the selected parameters:
4-70
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
SAVING/LOADING PARAMETERS
Please insert load diskette
in the drive.
Return: Valid - Esc: Cancel
" Insert the required diskette into the floppy The following windows appear during the
drive then confirm with the key loading:
Loading
Please wait
Advance %
/__/__/__/__/__/__/__/
Loading OK
Then, the SAVE/REREAD THE PARAMETERS
screen appears at the end of the loading.
4.7.5.4 Saving
Prior: Select the items which have to be saved (see section 4.7.5.2).
ACTIONS RESULTS
" Press the L key to reach Save, then The following window appears:
confirm with the key
SAVING/LOADING PARAMETERS
Do you want to save
selected parameters ?
Return: Valid - Esc: Cancel
4-71
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Insert a 3 1/2", 1.44 Mo diskette into the The following window appears:
floppy drive, then confirm with the key
Diskette Control in Progress
Please wait
If all items have been selected, the following
window is displayed:
SAVING/LOADING PARAMETERS
Valid to start diskette format;
Caution! This operation will
erase all data on the diskette
Return: Valid - Esc: Cancel
" If all items have been selected, press the The following windows appear during
key formatting (if all items have been selected):
Format in progress
Please wait
Advance %
/__/__/__/__/__/__/__/
The following windows appear in any case:
Saving
Please wait
Advance %
/__/__/__/__/__/__/__/
Saving OK
Then, the SAVE/REREAD THE PARAMATERS
screen appears at the end of the Parameters
Saving.
Erases all data (all files) of the diskette. This operation enables to re-use a diskette with full capacity.
4-72
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to reach Format a Diskette, A warning indication appears in the
then confirm with the key displayed window:
Saving/loading parameters
Please insert in the drive
the diskette to format
Caution: this operation
will erase all data
on the diskette !
Return: Valid - Esc: Cancel
" Insert a 3 1/2" into the floppy drive, then The following windows appear while
confirm with the key formatting:
Format in progress
Please wait
Advance %
/__/__/__/__/__/__/__/
Then, the SAVE/REREAD THE PARAMETERS
screen appears at the end of the formatting.
ACTIONS RESULTS
" Press the L key to reach Return to The following window is displayed:
Previous Menu, then confirm with the
key Updating files
Please wait...
The User Maintenance menu appears.
4.7.6 Maintenance
This program helps the operator to replace:
- The Needles
- The SyringeTip and/or the Syringe
- The Suction Tip
- The Halogen Lamp of the optical module (not available on STA Compact< CT)
4-73
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
It also enables to facilitate some maintenance procedures and to have access to a history of the recent
maintenances.
ACTIONS RESULTS
" Press the L key to reach Maintenance, The following window is displayed:
then confirm with the key
Initialization in progress
Please wait...
The maintenance menu appears:
Needle Purge
Needle Replacement
Syringe Tip/Syringe Replacement
Suction Tip Replacement
Halogen Lamp Replacement
Dates of Recent Maintenance
Return to the Previous Menu
with the cursor on Needle Purge.
4-74
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the key Needle Purge The NEEDLE PURGE screen appears.
The following windows appear:
Checks in progress
Please wait...
Positioning in well
Please wait...
4-75
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
ACTIONS RESULTS
4-76
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to reach Suction Tip The following window is displayed:
Replacement, then confirm with the
key Initializing...
The Suction Tip Replacement menu
appears:
ACTIONS RESULTS
" Press the L key to Suction Tip Test then The following window appears:
confirm with the key
Home position in progress
Test of the suction tip on the measurement
plate.
The following window appears:
" Press the Space Bar to proceed The Suction Tip Replacement menu
appears
4-77
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the L key to reach Halogen Lamp The following window is displayed:
Replacement, then confirm with the
key Initializing...
The Halogen Lamp Replacement menu
appears:
Lamp Replacement
Lamp Test
Return to Previous Menu
with the cursor on Lamp Replacement.
Lamp Replacement
See description in chapter Replacing Components, chapter 4.5.6.
Lamp Test
ACTIONS RESULTS
" Press the L key to Lamp Test, then Halogen lamp test in process.
confirm with the key
At the end of this test, the following window
appears:
LAMP REPLACEMENT
The procedure was
successful
Space Bar to proceed...
" Press the Space Bar to proceed The Halogen Lamp Replacement menu
appears.
ACTIONS RESULTS
" Press the L key to reach Return to The Maintenance menu appears.
Previous Menu, then confirm with the
key
4-78
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to reach Dates of Recent The DATES OF RECENT MAINTENANCE
Maintenance then confirm with the key screen appears (see Fig. 20 below).
" Press the keys Alt and Prtsc Printout of the DATES OF RECENT
MAINTENANCE screen.
4-79
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the L key to reach Return to Return to the User Maintenance menu.
Previous Menu, then confirm with the
key
4-80
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to reach Photometry The list of the last 99 files appears. The
Graphics, then confirm with the key following keys are available:
View
K,L Scroll
Esc Quit
Home Beginning of List
End End of List
Previous page
Following page
" Select the required file The required graphics appear (see figure 21
below)
" Confirm with the key
Legend:
1 Available functions:
4-81
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
In OD/Minute kinetic:
Full Calculus: all the points recorded during the test and not deleted are displayed in
yellow.
STA Calculus: only the points used to calculate the regression coefficient are displayed in
yellow.
In kinetic 2 points:
Full Calculus: T0, T1, the point before T0 and the point before T1 are displayed in red.
The other points are displayd in yellow.
STA Calculus: T0, T1, the point before T0, the point before T1, the point after T0, the
point after T1 and the recalculated points are displayed in red. The other points are
displayed in yellow.
F6 Screen print.
I Cursor movement.
J
4-82
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
3 Date and Time of the results saving, name of the file, and channel used for measurement
4 Abbreviation of the test, type of method (OD/minute kinetics or kinetic 2 points), and file
Identification
6 Reject criteria:
Kinetic OD/Minute:
Min OD
Recall of the linearity range of the measurement system.
Max OD
R Regression coefficient for the selected test as defined in the TEST SETUP
screens.
Kinetic 2 points:
Min OD
Recall of the linearity zone of the measurement system.
Max OD
Kinetic OD/Minute:
The equation of the regression curve (OD) and the regression coefficient (R) are calculated with
the yellow points included in the calculation window.
These values are updated at every addition/retrieval of one point in the window calculation or after
using the F3 key.
Kinetic 2 points:
The calculation is updated when the limits of the calculation window are modified.
4-83
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
9 Calculation window
ACTIONS RESULTS
" Press the F3 key All points used by the STA Compact< for the
regression calculation (patient result) appear
in yellow on the screen, points not used are
displayed in red.
The regression curve appears.
ACTIONS RESULTS
4-84
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
Press the F7 key " The list of raw data for the required files
prints out.
" The pair of points [Time (sec); Optical
Density (OD)] displayed in bold are those
used by the STA Compact< for the
regression calculation (patient result).
ACTIONS RESULTS
" Change the final/initial time with the The calculation window and the regression
numerical pad, then confirm with the curve are updated.
key
Note: the initial time cannot be below 2.
" Press the F10 key to validate the
modifications
4-85
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the F10 key The list of the last 99 files appears. The
following keys are available:
View
K,L Scroll
Esc Quit
Home Beginning of List
End End of List
Previous page
Following page
" Select the required file The selected Photometry Graphics appears.
" Confirm with the key
ACTIONS RESULTS
4-86
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the L key to reach Error History The ERROR HISTORY screen appears (see
" Confirm with the key figure 22 below).
ACTIONS RESULTS
4-87
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Enter the selection (2 numbers) then The ERROR HISTORY screen appears.
confirm with the F10 key
" Press the L or K keys to display all the The selected errors can be displayed.
selected error messages
ACTIONS RESULTS
ACTIONS RESULTS
" Press the L key to reach Host Computer The COMMUNICATIONS screen appears.
Communications, then confirm with the
key
" Press the F1 key to run the test The following message appears:
Test
In progress
At the end of this test, the PASSED/FAILED
message appears at the left bottom of the
displayed screen
4-88
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the Esc key to Exit The following window is displayed:
To avoid computer viruses, do not use the diskettes provided by Diagostica Stago on any other
PC System.
4.7.11 Utilities
Patient data of a validated file: transfered even if the corresponding tube is not in the sample drawer.
Patient data of a file that is in error or to be validated: only transfered if the corresponding tube is in
the sample drawer.
The data are saved in a text file (.txt) that can be used with others application softwares (e.g: Excel).
4-89
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
ACTIONS RESULTS
" Press the L key to reach Utilities Menu, The following screen appears: (see Fig.20)
then confirm with the key
1.TDEX
2. Saving Test Configurations disk updating
Your Choice [1,2]
ACTIONS RESULTS
" Enter the name of the files (8 characters Creation of a file "xxxxxxxx.txt" which can
max.) then validate with the key be used in other applications.
" Enter your comments, if necessary. It is
possible to enter 5 lines of comments which
can include up to 75 characters each
" Press the key
4-90
0931033L - July 2009
REFERENCE MANUAL 4
Maintenance
ACTIONS RESULTS
" Press the F10 key to transfer the data The folllowing window is displayed:
Transfer in progress
Transfer of data from the diskette to hard
disk
" Press the ESC key to quit The following window is displayed:
ACTIONS RESULTS
" Press the K key to return to the previous Same results as described in chapter 3.1.
menu, then confirm with the key
4-91
0931033L - July 2009
4 REFERENCE MANUAL
Maintenance
4-92
0931033L - July 2009
REFERENCE MANUAL 5
Table of contents
5 Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
5-i
0931033L - July 2009
5 REFERENCE MANUAL
Table of contents
5-ii
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
5 Test Setup
For each test, there are three screens of test setting parameters:
- Page 1 of 3 for methodology (see chapter 5.1.1)
- Page 2 of 3 for calibration (see chapter 5.1.2)
- Page 3 of 3 for printing, for transmitting results, and for quality control (see chapter 5.1.3)
For each dependent test, there is only one screen of test setting parameters (see chapter 5.1.4).
Legend:
5-1
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
Esc Displays the EXIT OPTIONS window (see description chapter 5.3.2 l)
3 Test Identification
7 Analysis definition
8 Result definition
Note: The definitions of each parameter are described in the following pages.
5-2
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Reminder : 1 / 6
L L
5-3
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
5-4
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
The selection of the incubation time for the sample (see chapter 5.1.1.2) as well as that for each
reagent must be defined in accordance with one of the following methods:
Incubation Methods
5-5
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
5-6
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Clotting
Min. Time Minimum time: inferior time limit for the analysis.
Below this limit, no result is given, “< V min.” is displayed
instead.
Max. Time Maximum time: superior time limit for the analysis.
Above this limit, no result is given, “> V max.” is displayed
instead.
Mean Time Estimated average clotting time of all patients. It is used by the
software to adjust the throughput and to avoid traffic problems in
the measurement zone.
In routine conditions the risk of traffic problems is rather low. On
the other hand, for a given test on large series of plasma with very
long clotting times (anti-vitamin K treatment for example), the
mean time must be increased to allow the STA Compact® to
slow down the throughput.
5-7
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
First Point Time (in seconds) between the dispensing of the start reagent
and the first point of optical reading.
Second Point Time (in seconds) between the dispensing of the start reagent
and the second point of optical reading.
Linearity Min. Minimum value of linearity acceptable for each linear regression.
First Point Time (in seconds) between the dispensing of the start reagent
and the first point of optical reading.
Second Point Time (in seconds) between the dispensing of the start reagent
and the last point of optical reading.
Kinetic 2-point immunology with calibration mode: 2nd or 3rd order polynomial (not
available on STA Compact®CT)
In the special case of an immunological method in kinetic 2-point with calibration mode: 2nd or
3rd order polynomial, redilution level to use for out of range results.
Out of range results are defined as:
- 'O.D. obtained greater than 'O.D. obtained for the highest calibration point
- Or difference between the points of the initial plateau greater than 50 mO.D. (0.05 O.D.)
5-8
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Stirring Type of stirring for the needle No.3 during dispensing of start
reagent (Rd) into the cuvette in measurement position.
The following options are available :
YES: stirring
NO: no stirring
Note: If intermediate reagents are dispensed before first
incubation, then the cuvette content is systematically stirred by a
lateral movement of needle No.2 in the cuvette.
5-9
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
5.1.1.7 Results
5-10
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
5-11
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
The rediluted result is given in the primary unit only (secondary units are not used in that case).
The result is displayed in blue in the test panel.
< "low limit" Low limit of the linearity zone in the primary unit for
reporting results.
> "high Limit" High limit of the linearity zone in the primary unit for
reporting results.
"Dil.1" < For the redilutions, the following options are available:
"Dil.2" > 1/1 1/20 1/2 1/30
1/3 1/40 1/4 1/50
1/5 1/60 1/6 1/80
1/7 1/100 1/8 1/160
1/10 1/200 1/12 1/400 1/15
Example 1:
For a regular methodology with a regular 1/20 dilution and a linearity up to 20g/ml. If the dilution
above 20g/ml is 1/40, the methodology must be set up as follows:
- Dil.1 1/1
- Dil.2 1/40, high limit 20
- save the methodology when quitting.
5-12
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Example 2:
For Fibrinogen, the linearity zone is given from 1.5 g/l to 9 g/l (150 to 900 mg/dl). For a regular 1/20
dilution, the following dilution conditions are defined:
- Dil.1 1/8, low limit 1.5
- Dil.2 1/40, high limit 9.0
- save the methodology when quitting.
It means that:
- For all results lower than 1.5 g/l, the assays will be automatically rerun with a 1/8 dilution
- For all results higher than 9.0 g/l, the assays will be automatically rerun with a 1/40 dilution
In both cases, the results will be given taking into account the dilution change, i.e.:
for a result > 9 g/l, result in primary units x 2.
for a result < 1.5 g/l, result in primary units x 8/20.
5-13
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
5-14
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Legend:
1 Available functions :
Esc Displays the EXIT OPTIONS window (see description chapter 5.3.2)
5-15
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
5-16
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
These parameters are only available within bar coded calibration mode.
View Points Selected points (in primary units for reported results) which the
(8 maximum) user chooses to view on calibration screen (see chapter 7.2.1).
These parameters are only available within graphic calibration modes (linear, 2nd order polynom,
3rd order polynom, or hyperbolic).
Key Only for the Diagnostica Stago products, allows to choose, in a list
corresponding to the order of the parameters printed on the bar
code labels, the test or the reagent concerned.
Dil. Choice of dilutions for each calibrator. There are the following
options:
1/1 1/20 1/2 1/30
1/3 1/40 1/4 1/50
1/5 1/60 1/6 1/80
1/7 1/100 1/8 1/160
1/10 1/200 1/12 1/400 1/15
5-17
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
For a calibration:
It is possible to use different calibrators or a single calibrator. In both cases, the dilution
corresponding to each calibration point must be specified.
CALIBRATOR
DILUTION
ID. Name
CALIBRATOR
DILUTION
ID. Name
5-18
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
These parameters are only available for graphic calibration modes (linear, 2nd order polynom, 3rd
order polynom or hyperbolic) and bar coded calibration mode.
This parameter is only available for graphic calibration modes (linear, 2nd order polynom, 3rd order
polynom, or hyperbolic).
The parameters for “Offset corrector” apply only for chromogenic and immunological tests (only
for linear regression mode). This correction corrects for Y-intercept (the slope is maintained).
5-19
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
Vial. Initial offset corrector plasma volume in each vial (in milliliters).
ml
Min. volume Volume of remaining offset corrector plasma at which the STA
ml Compact® stops running tests.
The following parameters apply to all measurement methods and to all calibration modes.
0 or 2 calibration validation controls can be defined. When 2 controls are used, they enable automatic
validation of the calibration (see chapter 7.3.1). They have to be defined for the bar coded tests and
when an offset corrector plasma is used.
5-20
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Min. volume The volume of remaining control at which the STA Compact®
ml stops measurements.
Legend:
1 Available functions:
Esc Displays the EXIT OPTIONS window (see description chapter 5.3.2)
Unit Besides the primary units, three secondary units may be defined
and used for the printout and/or transmission.
5-21
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
Conver. Factor Conversion coefficient between the primary unit and the
secondary unit concerned, if applicable.
Print For the primary unit or for the secondary units, there are two
options for printing out:
YES: unit used
NO: unit not used
5-22
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Min. Max. Definition of the normal value range as a function of the primary
unit in which the results are expressed.
Min.volume Volume of remaining control at which the STA Compact® will not
ml run another quality control.
5-23
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
Legend:
Esc Displays the EXIT OPTIONS window (see description chapter 5.3.2)
5-24
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Units Reminder of the main and secondary units used for the
dependent test and the main test.
Transmission Test Num- For the use of primary or secondary units for transmission, the
ber transmission test numbers corresponding to the host computer
should be given if the unit is used. If the unit is not used, the rank
must be set to 0 (F1 key).
5-25
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
ACTIONS RESULTS
" Press the key to open the Test Setup The TEST SETUP screen is displayed with the
screen cursor on the first test.
There is a yellow point in front of any
dependent test abbreviations.
If the operator views a test setup and the
following message “Reading in process-
Modification not allowed- Space Bar:
washing window” is displayed at the
bottom of the screen, there are
measurements in progress.
" If the STA Compact® is in stand-by, you can The TEST SETUP screen is displayed with the
create test setups. cursor on the first test.
To modify certain test setup parameters, There is a yellow point in front of any
the following is also necessary: dependent test abbreviations.
- the concerned tests are not present in any
patient file of the working file
- no calibration, no quality control is in
progress for this parameter
- the access code is known
If not, you can only view them.
" Press the key to view or edit a test setup
5-26
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Legend:
Access to the first TEST SETUP screen for the test on which the cursor is positioned
Displace Test
Delete Test
Update
Quit
For the options: Displace Test, Delete Test, Update, please see description in chapter 5.5, 5.6, and 5.7.
3 Complete name corresponding to the cursor position. For each Dependent Test: name of the
main test it is derived from
4 Available test list. There is a yellow point in front of any dependent test abbreviation
5-27
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
2TEST SETUP screen displayed with the cursor on the first test (see chapter 5.2.1).
ACTIONS RESULTS
" Press the Lkey to the required main test Cursor on the required test.
(without yellow point)
" Confirm the choice with the key Page 1 of 3 of the test setup displayed, with
the cursor on the first modifiable setup
parameter.
" Press the PgDn key Page 2 of 3 of the test setup displayed with
the cursor on the first modifiable setup
parameter.
" Press the PgDn key Page 3 of 3 of the test setup displayed, with
the cursor on the first modifiable setup
parameter.
" Press the PgUp key twice Return to page 1 of 3 of the test setup.
" Press the Esc key The EXIT OPTIONS window is displayed:
EXIT OPTIONS
Save before Quitting
Quit without Saving
Do not Quit
To save:
" See the validation procedure in 5.3.2
5-28
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
If the Esc key is pressed by mistake, press it again in order to cancel the instruction to exit from the
setup screens.
5.2.4.1 Access to the DEPENDENT TEST SETUP Screen from the TEST SETUP Screen
2TEST SETUP screen displayed, with the cursor on the first test (see chapter 5.2.1).
ACTIONS RESULTS
" Press the L or the J key to reach the The cursor is on the required test.
required test (with a yellow point)
" Confirm the choice with the key The screen of the dependent test setup is
displayed with the cursor on the required
dependent test of the dependent test list
(see the description of this screen chapter
5.1.4).
Confirm the choice with the key Cursor on the first modifiable parameter of
the dependent test setup.
5.2.4.2 Access to the DEPENDENT TEST SETUP Screen from the Required Main Test Setup
NOTE This access is available only from page 1 of 3 of the main test setup.
5-29
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
ACTIONS RESULTS
" Press the L key to the required dependent Cursor on the first modifiable parameter of
test, then confirm with the key or just the dependent test setup.
confirm by pressing the key
KEYS RESULTS
5-30
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Free entry field Enter the new information on the alphanumeric keyboard.
Confirm with the key.
5-31
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
2One of the setup screens is displayed and all the modifications have been made.
ACTIONS RESULTS
EXIT OPTIONS
Save before Quitting
Quit without Saving
Do not Quit
with the cursor on Save before quitting.
" Press the key to confirm The software checks the consistency of (and
between) the different parameters.
If an inconsistency is revealed [for instance,
total volume (sample + reagents) is higher
than 400 Pl]:
- the error window is displayed giving the
number of the setup page concerned by the
inconsistency and a description
ERROR - Page 1
Sample + Reagents volume
Out of range [150......400] Pl
Esc : Cancel this message
- the software displays automatically the
concerned setup page.
" Read carefully the displayed message, then The cursor is automatically positioned on
press the Esc key the parameter to be modified.
" Proceed with the modification(s) then The software checks the consistency once
press the Esc key once again again.
If no inconsistency is revealed, the following
window is displayed:
Please enter
your access code:
Validate Esc Quit
5-32
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
ACTIONS RESULTS
" Enter your access code, then confirm with The software checks if critical parameters
the key have been modified (for example, incubation
times, reagent volumes).
If critical parameters have been modified:
The following window warns the users of
the consequences of their modifications:
Saving ...
Please wait
- Test setup is saved
- Back to the TEST SETUP screen
" From the TEST SETUP screen, press the A message advising the operator to proceed
Esc key then confirm with the key with the saving of the test configurations is
displayed.
5-33
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
NOTE For the description of the test setup parameters: see chapter 5.1
5-34
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
" When all the parameters are defined, press the F10 key, then press the Esc key and the key
(Save before Quitting), then enter your access code and confirm with the key: return to the
Main Test
" Press the Esc key, then the Q key: return to the TEST SETUP screen
It is not necessary and even not recommended to save the main test since the system will reject this
request if there are still Patient files containing this main test in the Working File.
This procedure is only possible if the STA Compact® is in stand-by (no measurement in process).
2The TEST SETUP screen is displayed, with the cursor on the first test (see chapter 5.2.1).
ACTIONS RESULTS
Displace Test
Delete Test
Update
Quit
with the cursor on Quit.
5-35
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
ACTIONS RESULTS
" Press the L key to the test which has to be The following window appears:
moved then confirm by pressing the key
Move cursor to the new position
Insert
Esc Quit
with the yellow cursor on the test to be
moved.
" Press the Kor the L keys to move the The test order is automatically reorganized.
yellow cursor to the required place, then
confirm with the key
" Same procedure as above to move the The test order is automatically reorganized,
other tests, then exit this procedure by then return to the TEST SETUP screen.
pressing the Esc key
If the deleted test is a main test, then all tests dependent from this test will be also deleted.
2The TEST SETUP screen is displayed, with the cursor on the first test of the list (see chapter 5.2.1).
ACTIONS RESULTS
5-36
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
ACTIONS RESULTS
Displace Test
Delete Test
Update
Quit
with the cursor on Quit
" Press the L key to Delete Test, then The following window appears:
confirm with the key
TEST DELETION
Abbrev. XXX
Name XXX
WARNING:
All dependent tests will be deleted if the
deleted test is a main test
Please type YES to confirm ...
Esc = Quit
with the yellow cursor positioned for the
confirmation.
" Type YES, then confirm with the key The following window appears:
Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The following window appears :
the Key
TEST DELETION
Abbrev. XXX
Name XXX
DELETING...
Please wait ...
When the deletion is done:
Return to the TEST SETUP screen
automatically.
TEST DELETION
Abbrev. XXX
Name XXX
Test used in Working File
CANNOT DELETE
Esc Cancel this message
" Press the Esc key to delete this last The TEST SETUP screen appears.
message
5-37
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
The procedure allows to update or create tests predefined by Diagnostica Stago. This procedure is
only possible if :
- the STA Compact® is in stand-by (no measurement in process);
- no test is present in the Patient Files of the Working File;
- the access code is known.
If tests have to be added, there should be enough empty positions, otherwise the operator will be
asked to delete the number of required tests.
2The TEST SETUP screen is displayed with the cursor on the first test of the test list (see chapter
5.2.1).
ACTIONS RESULTS
Displace Test
Delete Test
Update
Quit
with the cursor on Quit.
5-38
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
ACTIONS RESULTS
" Enter your access code then confirm with The following window appears:
key
Please insert the diskette
TESTS UPDATE
in the floppy drive
Confirm Esc Quit
TEST UPDATING
WORKING FILE NOT EMPTY
Updating not completed
Erase Working File
Abort
with cursor positioned on Abort.
The Working File may be erased by pressing
the L and keys, then by confirming with
YES and
" Insert the diskette in the floppy drive, then The following window appears:
confirm with the key
File reading
Please wait ... ...
Then, the TEST UPDATING screen appears.
The list of tests which are going to be
updated or created is displayed in blue.
The number of tests to be added is also
displayed, as well as the number of positions
currently occupied, the number of available
positions.
5-39
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
ACTIONS RESULTS
Test updating
Please wait ...
On the TEST UPDATING screen, the test
display change from blue to yellow during
the updating.
Then, the window changes into:
Update done
Esc Quit
" Press the Esc key The UPDATE TRACK RECORD screen
appears. The tests previously updated or
created are displayed in blue, those
currently updated or created in yellow.
The following window appears:
OPTIONS
Update
Quit
with the cursor positioned on Quit.
5.8.1 Introduction
The calculated tests are tests for which the result is a calculation expression using one of these
2 formulas:
(R1/R2) x k
or (R1-R2) x k
where:
R1, R2: Results not limited to the printout limits expressed as raw data (sec., 'O.D., O.D./min) for
test [1] and test [2] chosen by the operator.
k: Programmable coefficient.
Thus, one result can be given either as equal to a ratio between two tests or as equal to a difference.
The unit in which the result is given is implicit; i.e.: Ratio for a ratio calculation and raw data for a
difference calculation. The secondary units have not to be defined for calculated tests.
This type of calculation is authorized only when test [1] and test [2] are based on the same type of
measurement method: i.e.: clot-based, chromogenic (not available on STA Compact®CT) or
immunology (not available on STA Compact®CT).
5-40
0931033L - July 2009
REFERENCE MANUAL 5
Test Setup
Calculated tests as well as dependent tests can not be used as test [1] or test [2].
Results from calculated tests are given only when those for test [1] and test [2] have as status “To
Confirm”, “Confirm”, “Dev.> tol.” or “V > V max.” (equivalent to T > T max.) (see description
of these statuses chapters 8.3.2 and 8.3.3).
Moreover, they can not be confirmed manually: their confirmation implicitly follows from those of
test [1] and test [2]. Results from calculated tests have as error code and as alarm code, the most
serious error code and the most serious alarm code among those assigned to test [1] and test [2].
Only alarm code H (result in primary unit limited to printout limit values) escapes this rule because
it directly applies to the values obtained for the calculated results (see below chapter 5.8.2).
Because of their particularity, calculated test can not be rerun, it exists neither calibration nor quality
control for this type of test and those can not have any dependent test.
The procedures to access, to modify and to create these tests are identical to those described for
the other types of test (see chapters 3.2, 5.3, 5.4).
Legend:
1 ESC displays the EXIT OPTIONS window (see description chapter 5.3.2).
5-41
0931033L - July 2009
5 REFERENCE MANUAL
Test Setup
Abbreviation Abbreviations for test [1] and test [2] to be selected in the list of
tests currently available on STA Compact®.
Reminder: Test [1] and test [2] must be of the same type: clot-
based, chromogenica or immunologicala.
Name Names of test [1] and test [2]. They are automatically displayed
after selection of the abbreviations.
a. not available on STA Compact®CT
NOTE The unit is implicit with relation to the type of calculation selected, see chapter 5.8.1.
Trans.: Test Number To be able to transmit the result of the calculated test, the
transmission test number corresponding to the host computer
has to be defined. If the test has not to be transmitted, rank has
to be set to 0 (F1 key).
5-42
0931033L - July 2009
REFERENCE MANUAL 6
Table of contents
6.1 Generalities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
6-i
0931033L - July 2009
6 REFERENCE MANUAL
Table of contents
6-ii
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
6.1 Generalities
The samples and the products (controls, calibrators, diluents, reagents, desorb solutions) can be
loaded in the sample drawer or in the product drawer, see table below:
Controls NO YES
Calibrators NO YES
Pour relecture / For proofreading 11/9/09
Diluents YES NO
Reagents NO YES
Samples YES NO
6-1
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
To comply with package inserts requests related to the intrinsic coagulation pathway
deficient plasmas (VIII, IX, XI, XII) as well as with those related to the reagent STA-
Staclot® Protein S, i.e.: “The factor assays of the patients samples follow the completion
of assay calibration and are performed within the time-frame that is compatible with the
stability of related factor in plasma”, the following procedure must be observed:
- before starting samples loading, block the required test (example: factor VIII), see procedure
chapter 8.7.3;
- run calibration for the required test (example: factor VIII);
- change to STAT (F12 key), any sample tube with the required test (example:
factor VIII) in its working list: whether, while loading (see procedure chapter 6.8.4) or while
processing Patient files (see procedure chapter 8);
as soon as calibration of the required test (example: factor VIII) is validated, unblock the test
(example: factor VIII) (see procedure chapter 8.7.3)
6-2
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
Samples F1
Products F2
Cuvettes
Pour relecture / For proofreading 11/9/09
Washing Solution
with the cursor on Samples.
6-3
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" For the Sample Drawer, press the key 2 possible cases:
" For the Reagent Drawer, press the P key - STA Compact® is pipetting in the selected
drawer or needs the drawer (mainly the
product drawer for the cuvettes still in
incubation); the message “X drawer:
Opening requested” appears at the
bottom of the TEST PANEL screen (X= the
number of the drawer).
6-4
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
RISK OF INJURY
In order to avoid any injury:
While a drawer is open, do not insert the hand or any object inside the STA Compact®.
Reminder If the liquid level detection management has been inhibited for one or several needles, a warning
message will be displayed upon opening of the related drawer (sample drawer for needle No.1,
reagent dawer for the 3 needles).
For the sample drawer, all tubes should be decanted and the operator must insure that
the plasma quantity is sufficient to perform the request assays.
For the product drawer, the theoretical bottle filling volumes as defined while loading shall be
consistent with the actual ones.
6.3
Pour relecture / For proofreading 11/9/09
Legend:
2 Available keys:
6-5
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
3 Identification number for each product. It is displayed in red when the stability is overdue, when
the minimum volume is reached or when a liquid level detection problem is detected.
5 Remaining volume of each product.It is displayed in red when the minimum volume is reached
or when a liquid level detection problem is detected
8 Position of each product in the product drawer. The sign P is displayed at the right side of the
position for products transferred into the microcups.
9 Date and Time of the end of stability for each product. It concerns stability defined within the
Test Setup menu (see chapter 5.1). It is displayed in red when the given time is within the
6-6
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
Legend:
Pour relecture / For proofreading 11/9/09
2 Available keys:
Change profile(s): Creates or modifies Test Profiles (see description in chapter 6.8.3).
Change Prefix (MANUAL Mode) or Change Prefix/First Number (AUTO Mode):
- For Manual Mode, modifies the prefix selected by default (see chapter 6.8.5.2).
- For Auto Mode, modifies the prefix and the first number (see description in chapter 6.8.6.2).
Auto mode: Accesses the AUTO Mode.
Return to Loading: Returns to SAMPLE LOADING with the cursor on zone 11 of the sample
loading screen (see below).
Quit: Closes the sample drawer and displays the TEST STATUS screen.
Modifies several parameters; once the drawer is closed no more modification is possible
6-7
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
4 Selected prefix
8 List of tests requested for each sample. For the diluents, name as defined within the TEST SETUP
screens (see chapter 5.1).
11 Field for entering data concerning the sample which has to be loaded. In the ID field is displayed
the Normal tube or Microtube indication in blue for routine samples and in yellow for STAT
samples.
12 Selected parameters for AUTO Mode: test profile, prefix and first number
6-8
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
In case of lot number change, the calibrations related to these reagents will no longer be valid.
When a stabilization period in STA Compact® analyzer is required for some reagents, it is essential
to block the tests using these reagents before loading them.
See the package inserts of each reagent kit and see blocking procedure chapter 8.7.3.
Pour relecture / For proofreading 11/9/09
2PRODUCT LOADING screen displayed with message “PRODUCT IDENTIFICATION (F8 = Normal
Vol./Microvolume)” at the bottom of the screen and Normal Vol. displayed (see description of
product LOADING SCREEN chapter 6.3).
ACTIONS RESULTS
" If the reagent has been transferred into a The designation of Normal Vol. changes
microcup, press the F8 key into Microvolume (see description of
PRODUCT LOADING screen chapter 6.3).
" Scan the bar code label of the reagent/ Identification (ID), name (Name), volume
desorb solution vial across the bar code (Volume), stability (Stab.), and, if
reader necessary, lot number (Lot) of the reagent/
desorb solution are displayed.
" In case of failure of the Bar Code Reader,
proceed as indicated below (chapter 6.5.2) The cursor is positioned on the Volume
field for correction if any.
The message “INDICATE THE VOLUME
(in ml) - Esc = Correction” is displayed at
the bottom of the screen.
6-9
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" If it is necessary to modify the volume, type The new volume is displayed.
the new value, then confirm with the The cursor is positioned on the Stability
key field.
" Otherwise, confirm directly with the The message “INDICATE THE STABILITY
key and skip to the RESULTS box below
(hours) - Esc = Correction” is displayed
at the bottom of the screen.
" Modify the stability if necessary, then The volume and the stability are displayed.
confirm with the key
The message “LOAD THE LAST IDENTIFIED
" Otherwise, confirm directly with the VIAL - Esc = correction” is displayed at the
key bottom of the screen.
" If the Lot number change is accepted, enter The following window is displayed:
YES then confirm with the key
Please scan the code sheet
in front of the reader
Esc: Quit
6-10
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Scan the bar code sheet in front of the For non-precalibrated reagent, skip to next
reader RESULTS box.
Note: In case of failure of the Bar Code For precalibrated reagents, the two
reader, type the information printed below following windows appear one after another
the bar codes and confirm each line with if necessary:
the key.
BAR CODE READING
End of Data Processing
" Press the Esc key Data for this reagent/desorb solution
transferred to the top part of the screen (see
description on chapter 6.3).
2PRODUCT LOADING screen displayed with message “PRODUCT IDENTIFICATION (F8 = Normal
Vol./Microvolume)” at the bottom of the screen and Normal Vol. displayed (see description of
PRODUCT LOADING screen chapter 6.3).
ACTIONS RESULTS
" If the reagent has been transferred into a The designation of Normal Vol. changes
microcup, press the F8 key into Microvolume (see description of
PRODUCT LOADING screen chapter 6.3).
" Enter the identification of the reagent/ The reagent/desorb solution identification is
desorb solution as defined in the Test displayed.
Setup on page 1 of 3 and confirm with the The cursor is on the Name field.
key
6-11
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Type the name of the reagent/desorb The name of the reagent/desorb solution is
solution as defined in the Test Setup page displayed.
1 of 3, then confirm with the key
The cursor is on the Volume field.
The message “INDICATE THE VOLUME (in
ml) Esc = Correction” is displayed at the
bottom of the screen.
" Type the stability, then confirm with the The stability is displayed.
key The cursor is on the Lot field.
The message" INDICATE THE LOT NUMBER -
Esc = Correction" is displayed at the
" Enter the lot number if necessary and For all reagents which are not from STA<
confirm with the key line of Diagnostica Stago, the following
window is displayed:
REAGENT LOADING
Should the reagent be mixed?
Y = Yes N = No
" For a reagent which needs mixing, type Y The message “LOAD THE LAST IDENTIFIED
VIAL - Esc = Correction” is displayed at
" For a reagent without mixing, type N
the bottom of the screen.
6-12
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" If the lot number change is accepted, type Data for this reagent/desorb solution
YES, then confirm with the key transferred to the top part of the screen (see
description on chapter 6.3).
2PRODUCT LOADING screen displayed with message “PRODUCT IDENTIFICATION (F8 = Normal
Vol./Microvolume)” at the bottom of the screen and Normal Vol. displayed (see description of
PRODUCT LOADING screen chapter 6.3).
ACTIONS RESULTS
" If the control/calibrator has been The designation of Normal Vol. changes
transferred into a microcup, press the F8 into Microvolume (see description of
key PRODUCT LOADING screen chapter 6.3).
" Scan the bar code of the control/calibrator Identification (ID), name (Name), volume
in front of the reader (i.e.: STA< - COAG (Volume), stability (Stab.), and lot number
CONT N) (Lot) of the control/calibrator are displayed.
" In case of failure of the bar code reader, The cursor is positioned on the Volume
proceed as indicated below, chapter 6.6.2) field for correction if any.
The message “INDICATE THE VOLUME (in
ml) - Esc = Correction” is displayed at the
bottom of the screen.
" If the volume has to be modified, enter the The new volume is displayed.
new value, then confirm with the key The cursor is positioned on the Stability
" Otherwise directly confirm with the key field.
and jump to the RESULTS box below
The message “INDICATE THE STABILITY
(hours) - Esc = Correction” is displayed
at the bottom of the screen.
6-13
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Modify if necessary the stability, then The volume and the stability are displayed .
confirm with the key The message “LOAD THE LAST IDENTIFIED
" Otherwise confirm with the key VIAL - ESC= Correction” is displayed at
the bottom of the screen.
" Enter YES then confirm with the key The following window is displayed:
Esc: Quit
6-14
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Scan the bar code sheet in front of the At the end, the following window is
reader displayed:
" In case of failure of the bar code reading,
enter the data printed under the bar codes BAR CODE READING
and confirm with the key at the end of Data Processing End
each line Esc: Cancel this message
" Press the Esc key Data for this control/calibrator transferred
to the top part of the screen (see description
on chapter 6.3).
2PRODUCT LOADING screen displayed with message “PRODUCT IDENTIFICATION (F8 = Normal
Vol./Microvolume)” at the bottom of the screen and Normal Vol. displayed (see description of
PRODUCT LOADING screen chapter 6.3).
ACTIONS RESULTS
" If the control/calibrator has been The designation of Normal Vol. changes
transferred into a microcup, press the F8 into Microvolume (see description of
key PRODUCT LOADING screen chapter 6.3).
" Enter the alphanumeric identification of the The identification of the control/calibrator is
control/calibrator as defined in the TEST displayed.
SETUP screens (including spaces), then
The cursor is positioned on the Name field.
confirm with the key,
The product chosen for this example is the
STA< - COAG CONT N.
6-15
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Enter the product name: STA - COAG The name of the control/calibrator is
CONT N (in this case) as it was defined in the displayed.
TEST SETUP screen (see chapter 5.1), then
The cursor is on the Volume field.
confirm with the key
The message “INDICATE THE VOLUME (in
ml) Esc = Correction” is displayed at the
bottom of the screen.
" Enter the volume, then confirm with the The volume is displayed.
key
The cursor is on the Stab. field.
The message "INDICATE THE STABILITY
(hours) - Esc = Correction" is displayed at
the bottom of the screen.
" Enter the stability, then confirm with the The stability is displayed.
key The cursor is on the Lot field.
The message "INDICATE THE LOT NUMBER -
Esc = Correction" is displayed at the
" Enter the lot number indicated on the vial, The message “LOAD THE LAST IDENTIFIED
then confirm with the key TUBE - Esc = Repeat ident” is displayed at
the bottom of the screen.
6-16
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" If the lot number change is accepted, enter Data for this control/calibrator transferred
YES then confirm with the key to the top part of the screen (see description
on chapter 6.3).
Legend:
1 Identification number for the diluent with indication of the sample type: Normal Tube or
Microtube
4 Diluent volume
5 Diluent stability
2SAMPLE LOADING screen displayed with MANUAL Mode or AUTO Mode (as described in
chapter 6.4).
6-17
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Scan the bar code of the product in front of The STA Compact® beeps and the following
the reader (i.e.: Owren - Koller Buffer) window is displayed:
" In case of failure of the bar code reader,
proceed as indicated below (chapter 6.7.3) Product detected
F1: Diluent
11361
F2: Sample
xxx C2: 6199B
xxx: Recall of the prefix if the latter has been
defined.
" Press the F1 key The diluent name is displayed as well as the
volume and the stability.
The cursor is positioned on the Volume box
(see description above chapter 6.7.1).
" Otherwise directly confirm with the key The cursor is positioned on the Stability
and jump to the RESULTS box below box (area 5, see description above
chapter 6.7.1).
" Modify if necessary the stability, then The message “LOAD THE LAST IDENTIFIED
confirm with the key TUBE - Esc = Repeat ident” is displayed in
" Otherwise confirm with the key yellow at the bottom of the screen.
" Load the diluent vial in the reserved On the sample drawer:
position (No. 8) or in any other position The LED adjacent to the product position
You have to place the vial within 10 seconds, lights up and the STA Compact® beeps =
otherwise the vial will have to be identified position and identification validated.
again.
On the screen:
Automatic display of the position and data
transferred into the zone for samples
already loaded (see description of SAMPLE
LOADING screen chapter 6.4).
2SAMPLE LOADING screen displayed with MANUAL Mode or AUTO Mode (as described in
chapter 6.4).
6-18
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Enter the alphanumeric identification of the The following window is displayed:
product as defined in the TEST SETUP
screens (including spaces), then confirm Product detected
with the key, F1: Diluent
The product chosen for this example is the 11361
Owren - Koller Buffer. F2: Sample
11361
" Press the F1 key to confirm the product The entry zone for diluent is displayed (see
type (F2 allows loading a sample whose description above chapter 6.7.1).
identity will be the same as that of a dilution The cursor is positioned on the complete
buffer) name field (see description above
chapter 6.7.1).
" Enter the product name: Owren - Koller The cursor is positioned on the Volume
(in this case) as it was defined in the TEST field (see description above chapter 6.7.1).
SETUP screen (see chapter 5.1), then
Pour relecture / For proofreading 11/9/09
" Enter the volume, then confirm with The cursor is positioned on the Stability
the key field (see description abov, chapter 6.7.1 of
this manual).
" Enter the stability, then confirm with The message “LOAD THE LAST IDENTIFIED
the key TUBE - Esc = Repeat ident” is displayed in
yellow at the bottom of the screen.
" Load the diluent vial in the reserved On the sample drawer:
position (No. 8) or in any other position The LED adjacent to the position lights up
and the STA Compact® beeps = position
and identification validated.
On the screen:
Automatic display of the position and data
transferred into the zone for samples
already loaded (see description of SAMPLE
LOADING screen chapter 6.4).
6-19
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
6-20
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
TEST Selection
Host Automa-
Individual Profile computer tic Dis-
teleloading play
Manual X X X X
with prefix
Sample Manual X X X X
without prefix
Bar codes X X X X
without prefixes
The AUTO mode is dealt with separately because it incorporates the notions of prefix and test
Pour relecture / For proofreading 11/9/09
selection.
For the special case of the automatic display of the test list, see chapter 6.8.7.
6-21
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
The number of tests per profile is limited to 6. Several profiles can be cumulated to the amount of
12 tests per Patient file (Manual Mode).
2 List of profiles
6-22
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
4 Available functions:
Esc Abandon Returns to the first profile of the list of profiles without saving the
modifications
2SAMPLE LOADING screen displayed with MANUAL Mode or AUTO Mode (as described in chapter
6.4).
ACTIONS RESULTS
" Press the Esc key (Options) For the MANUAL Mode, the following
window is displayed:
Change Profiles
Change Prefix
AUTO Mode
Return to Loading
Quit
For the AUTO Mode, the following window is
displayed:
Change Profile
Change Prefix/First Number
MANUAL Mode
Return to Loading
Quit
In both cases, the cursor is positioned on
Quit.
" Press the C key (Change Profile) The PROFILE MODIFICATION screen appears
" Or move the cursor to Change Profile without the list of available tests (see
with the L key then confirm with the key description above).
" Move the cursor after the last profile using The list of available tests appears with the
the Lkey, thenconfirm with the key cursor on the first test.
6-23
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Select the desired tests by moving the The selected tests appear in the PROFILE
cursor keys L, J ,K, and I and confirm MODIFICATION screen (see above).
the choice of each test with the key
By teleloading is displayed in the PROFILE
" Or press the 0 key (Teleloading) MODIFICATION screen (see above).
" Save the new profile by pressing the The new profile is validated.
F10 key
" Press the Esc key Return to the SAMPLE LOADING screen.
2SAMPLE LOADING screen displayed with MANUAL Mode or AUTO Mode (as described in
chapter 6.4).
" Press the Esc key (Options) The following window is displayed:
Change Profiles
Change Prefix
AUTO Mode
Return to Loading
Quit
For the AUTO Mode, the following window is
displayed:
Change Profile
Change Prefix / First Number
MANUAL Mode
Return to Loading
Quit
In both cases, the cursor is positioned on
Quit.
" Press the C key (Change Profile) The PROFILE MODIFICATION screen appears
" Or move the cursor to Change Profile without the list of available tests (see
with the Lkey then confirm with the key description above).
" Move the cursor to the desired profile using The list of available tests appears with the
the L key, then confirm the choice with the cursor on the first test.
key
6-24
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Select the desired tests by moving the The selected tests appear in the PROFILE
cursor using the cursor keys L, J, K and I MODIFICATION screen (see above).
and confirm the choice of each test with
By teleloading is displayed in the PROFILE
the key
MODIFICATION screen (see above).
" or press the 0 key (Teleloading)
" Save the profile by pressing the F10 key The profile is validated.
" Press the Esc key Return to the SAMPLE LOADING screen.
The STAT samples are identified in yellow. For the STAT sample processing, see description
chapter 6.12.
2LOADING SAMPLE screen displayed in MANUAL Mode or in AUTO Mode (see description
chapter 6.4).
6-25
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" If it is a STAT sample, press the F12 key The designation of Normal tube changes
from blue to yellow (see description of
SAMPLE LOADING screen in chapter 6.4).
" If the sample has been transferred into a The designation of Normal tube changes
microcontainer, press the F8 key into Microtube (see description of SAMPLE
LOADING screen in chapter 6.4).
" If the sample is a STAT sample which has The designation of Microtube is displayed
been transferred into a microcontainer, in yellow (see description of SAMPLE
press both keys F8 and F12. LOADING screen in chapter 6.4).
2SAMPLE LOADING screen displayed with AUTO MODE (as described in chapter 6.4).
ACTIONS RESULTS
Change Profile
Change Prefix / First Number
MANUAL Mode
Return to Loading
Quit
with cursor on Quit.
6-26
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Press the M key (MANUAL MODE). Return to SAMPLE LOADING with the
" Or move the cursor to MANUAL Mode with indication: MANUAL MODE displayed (see
the L key then confirm with the key description of SAMPLE LOADING screen
chapter 6.4).
The message “SAMPLE IDENTIFICATION
(F12 = in priority F8 = Normal / Micro
Tube)” is displayed at the bottom of the
screen.
2SAMPLE LOADING screen with MANUAL Mode is displayed (as described in chapter 6.4) with the
message “SAMPLE IDENTIFICATION (F12 = in priority F8 = Normal/Micro Tube)” displayed
at the bottom of the screen.
Pour relecture / For proofreading 11/9/09
ACTIONS RESULTS
" Press the Esc key (Options) The following window is displayed:
Change Profiles
Change Prefix
AUTO mode
Return to Loading
Quit
with the cursor on Quit.
" Move the cursor with the L key to Change The following window is displayed:
prefix, then confirm with the key
New Prefix
[ - ] Save
Esc Abandon
with the cursor in the entry zone for the
prefix.
" To delete the current prefix, press the Go to the last RESULTS box below.
= key until the prefix is completely erased,
then confirm with the key
" To define a free prefix, type the free Return to SAMPLE LOADING. The chosen
desired prefix then confirm with the key prefix is displayed in the sample
identification area (see description of the
SAMPLE LOADING screen chapter 6.4).
6-27
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
2SAMPLE LOADING screen with MANUAL Mode displayed and the cursor is in the sample entry
zone (as described in chapter 6.4).The sample type has been defined (see chapter 6.8.4.2).
ACTIONS RESULTS
" Scan the sample tube in front of the bar The tube identification appears on the
code reader screen in the ID zone (see description of the
SAMPLE LOADING screen, chapter 6.4 ) and
the message “LOAD THE LAST IDENTIFIED
TUBE - Esc = Repeat ident” appears in
" Place the sample tube in any position of the On the sample drawer:
sample drawer The LED adjacent to the tube position lights
up and the machine beeps = position and
identification validated.
2SAMPLE LOADING screen with MANUAL Mode displayed and the cursor is in the sample entry
zone (as described in chapter 6.4).
The sample type has been defined (see chapter 6.8.4.2).
6-28
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Enter the sample identification on the The message “LOAD THE LAST IDENTIFIED
keyboard and confirm with the key TUBE - Esc = Repeat ident” is displayed in
yellow at the bottom of the screen.
" Place the sample tube in any position of the On the sample drawer:
sample drawer The LED adjacent to the tube position lights
up and the machine beeps = position and
identification validated.
SMITH SMITH 1
100 101
MAT100 MAT101
2SAMPLE LOADING screen with MANUAL Mode displayed and the cursor is in the sample entry
zone (as described in chapter 6.4).The sample type has been defined (see chapter 6.8.4.2).
ACTIONS RESULTS
6-29
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Confirm the generated identification with The message “LOAD THE LAST IDENTIFIED
the key TUBE - Esc = Repeat ident” appears in
yellow at the bottom of the screen.
" Place the sample tube in any position of the On the sample drawer:
sample drawer The LED adjacent to the tube position lights
up and the machine beeps = position and
identification validated.
2Tube in progress , already identified and loaded (see above). SAMPLE LOADING screen with
message “INDICATE TESTS TO PERFORM - F10 = VALIDATE” already displayed and cursor
positioned on the first test of the Selection zone (see description of the SAMPLE LOADING screen
chapter 6.4).
ACTIONS RESULTS
" Select the desired tests by moving the The selected tests appear in the Test
cursor to them using the cursor keys L, K, Selection zone for the sample being loaded
I and J and confirm the choice of each (see description of the SAMPLE LOADING
test with the key screen chapter 6.4).
" Save the test selection with the F10 key Data for this sample are transferred into the
zone for samples already loaded (see
description of SAMPLE LOADING screen
chapter 6.4).
2Tube in progress , already identified and loaded (see above). SAMPLE LOADING screen with
message “INDICATE TESTS TO PERFORM - F10 = VALIDATE” already displayed and cursor
positioned on the first test of the Selection zone (see description of the SAMPLE LOADING screen
chapter 6.4).
6-30
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Press on key (1 to 7) corresponding to The selected test profiles appear in the Test
desired profile Selection zone for the sample being loaded
(see description of SAMPLE LOADING screen
chapter 6.4).
" Validate the sample with the F10 key Data for this sample are transferred into the
Samples already loaded zone (see
description of SAMPLE LOADING screen,
chapter 6.4).
2Tube in progress, already identified and loaded (see above). SAMPLE LOADING screen with
message “INDICATE TESTS TO PERFORM - F10 = VALIDATE” already displayed and cursor
Pour relecture / For proofreading 11/9/09
positioned on the first test of the Selection zone (see description of the SAMPLE LOADING screen
chapter 6.4).
ACTIONS RESULTS
" Press the I key The cursor is positionned on the first profile
of the Recorded profiles area (see
description of the SAMPLE LOADING screen
chapter 6.4).
" Move the cursor to each selected profile The selected test profiles appear in the Test
then save each profile with the key Selection zone for the sample being loaded
(see description of SAMPLE LOADING screen
chapter 6.4).
" Save the sample with the F10 key Data for this sample are transferred into the
Samples already loaded zone (see
description of SAMPLE LOADING screen
chapter 6.4).
2Tube already in progress identified and loaded (see above). SAMPLE LOADING screen with
message “INDICATE TESTS TO PERFORM - F10 = VALIDATE” already displayed and cursor
positioned on the first test of the Selection zone (see description of the SAMPLE LOADING screen
chapter 6.4).
6-31
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Press the key (1 to 7) corresponding to the BY TELELOADING is displayed in the Test
BY TELELOADING profile Selection zone for the sample being loaded
(see description of the SAMPLE LOADING
" Or press the I key, then the L key to
move the cursor to the BY TELELOADING screen chapter 6.4).
profile and confirm with the key
" Confirm the sample with the F10 key If the VERIFY PATIENT DATA option is set to
NO, data for this sample are transferred into
the samples already loaded zone (see
description of SAMPLE LOADING screen
chapter 6.4).
If the VERIFY PATIENT DATA option is set to
YES, the following windows are displayed:
TELELOADING
Please wait
TELELOADED FILE
ID: XXX 1
Patient: XXX XXX
XXX 2 XXX
F10: Save Esc: Cancel
1 Patient File identification
2 Complementary Information (Name, First
name...) (see definitions chapter 10.2.3.3).
" If the data sent back by the Host Computer Data for this sample are transferred into the
are correct, press the F10 key Samples already loaded zone (see
description of SAMPLE LOADING screen
chapter 6.4).
" If the data sent back by the Host Computer The following window is displayed:
are not correct, press the Esc key
SAMPLE LOADING
MAINFRAME
Can you verify if the tube is removed
F10: Remove
The LED near the tube blinks.
6-32
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
2SAMPLE LOADING screen displayed with MANUAL MODE (as described in chapter 6.4).
Pour relecture / For proofreading 11/9/09
ACTIONS RESULTS
Change Profiles
Change Prefix
AUTO Mode
Return to Loading
Quit
with the cursor on Quit.
" Press the A key (AUTO Mode) Return to SAMPLE LOADING screen with
" Or move the cursor to AUTO Mode with Auto Mode displayed (see description of
SAMPLE LOADING screen chapter 6.4 ).
the L key, then confirm with the key
The message: “ = FILE EDITING - (F12 =
in priority - F8 = Normal/Micro Tube)”
displayed at the bottom of the screen.
6.8.6.2 Definitions of the Prefix and of the First Number for the AUTO Mode
2SAMPLE LOADING screen displayed with AUTO Mode (as described in chapter 6.4) and Message “
=FILE EDITING (F12 = in priority F8 = Normal/Micro Tube)” displayed at the bottom of the
screen.
6-33
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Press the Esc key (Options) The following window is displayed:
Change Profile
Change Prefix / First Number
MANUAL Mode
Return to Loading
Quit
with the cursor positionned on Quit.
" Move the cursor to Change Prefix/First The New Parameters window is displayed
Number with the L key, then confirm with with the cursor on the entry zone for the
the key prefix.
" To delete the current prefix, press the Skip to RESULTS box below.
= key until the prefix is completely erased,
then confirm with the key
" Type the first number, then confirm with Return to SAMPLE LOADING.
the F10 key
The parameters defined for the AUTO Mode:
Profile, prefix and first number, are displayed
in the Parameters zone (see description of
the SAMPLE LOADING screen chapter 6.4).
The identification of the first sample is
displayed in the ID zone for the sample
being loaded (see description of the SAMPLE
LOADING screen chapter 6.4).
2SAMPLE LOADING screen displayed, with AUTO MODE and cursor positioned on the entry zone
for the sample (as described in chapter 6.4).
The sample type has been defined (see chapter 6.8.4.2).
The test profile has been defined (see chapter 6.8.3).
The prefix and the first number have been selected (see procedure chapter 6.8.6.2).
ACTIONS RESULTS
" Press the key The message “LOAD THE LAST IDENTIFIED
TUBE” is displayed in yellow at the bottom
" Or scan the sample tube in front of the bar
code reader of the screen.
6-34
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" Place the sample tube in any position of the On the sample drawer:
drawer The LED adjacent to the tube position lights
up and the machine beeps = position and
identification validated.
Modifying the sample type or the list of tests will only be possible after loading of the sample tube
by means of the K key List/Change (see description chapter 6.9.2).
2SAMPLE LOADING screen displayed (MANUAL Mode or AUTO Mode), as described in chapter 6.4.
ACTIONS RESULTS
" Enter the sample identification on the The list of tests is automatically displayed in
keyboard, then confirm with the key the TEST SELECTION zone for the sample
being loaded (see description of the SAMPLE
" Or scan the sample tube in front of the bar
LOADING screen chapter 6.4).
code reader
The message “LOAD THE LAST IDENTIFIED
TUBE - Esc = Repeat ident (ALREADY
EXISTS)” is displayed in yellow at the
bottom of the screen.
6-35
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" Place the sample tube in any position of the On the sample drawer:
sample drawer The LED adjacent to the tube position lights
up and the machine beeps = position and
identification validated.
This procedure is only possible for controls, calibrators, reagents or desorb solutions loaded during
the loading session in progress. The possible modifications concern:
- the volume;
- the stability;
- the indication Normal Vol./Microvolume
2PRODUCT LOADING screen displayed with message “PRODUCT IDENTIFICATION (F8 = Normal
Vol./Microvolume)” at the bottom of the screen and Normal Vol. displayed (see description of
PRODUCT LOADING screen chapter 6.3).
ACTIONS RESULTS
" Press the K key The cursor is positioned on the last line of
the zone of products already loaded (see
description of the PRODUCT LOADING screen
chapter 6.3).
" Move the cursor on the required control/ Cursor positioned on the required control/
calibrator/reagent/desorb solution with the calibrator/reagent/desorb solution.
K key
6-36
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
ACTIONS RESULTS
" If it is necessary to modify the volume, type The new volume is displayed.
the new value, then confirm with the key
The cursor is positioned in the Stability
" Otherwise, confirm directly with the key field.
and jump to the RESULTS box below
The message “INDICATE THE STABILITY
(hours) - Esc = Cancel” is displayed at the
bottom of the screen.
" Modify, if necessary, the stability, then Volume and stability are displayed.
confirm with the key
Pour relecture / For proofreading 11/9/09
This procedure is only possible for samples or diluents already loaded during a loading session.
For samples, the possible modifications are:
- the list of tests for Manual Mode;
- the indication Normal Tube/Microvolume (F8 key);
- the indication Normal/STAT (F12 key).
6-37
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
2SAMPLE LOADING screen displayed with MANUAL Mode or AUTO Mode (as described in
chapter 6.4).
ACTIONS RESULTS
" Press the K key The cursor is positioned on the last line of
the zone for samples already loaded (see
description of SAMPLE LOADING screen
chapter 6.4).
" Move the cursor to the requested sample Cursor positioned on the requested sample
or product or diluent.
" Press the key All data from this sample (diluent) are
displayed in the entry zone (see description of
SAMPLE LOADING screen chapter 6.4).
" If necessary and depending on cases, press The (Normal tube) indication changes to
the F8 key (Microvolume) or reversely.
" Or/and press the F12 key The (Normal tube) or (Microvolume)
indication display changes from blue
(normal sample) to yellow (STAT sample).
" Then, press the key For samples, possibility to modify the list of
tests (see procedures chapters 6.8.5.6,
6.8.5.7 and 6.8.5.8).
For AUTO Mode, it is not possible to change the For diluents, possibility to modify the
list of tests, automatic return to loading. volume and/or the stability (see procedures
chapter 6.7).
" For samples, validate the modifications by Return to loading procedure: cursor
the F10 key positioned on the sample identification entry
zone.
" For diluents, validate the modifications by
the key Message “SAMPLE IDENTIFICATION
(F12 = in priority F8 = Normal/ Micro
Tube) displayed at the bottom of the screen.
6-38
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
This procedure is only possible for samples or products (reagents, controls, calibrators, desorb
solutions or diluents) loaded with positive identification; i.e.: automatic acknowledgement of loading
position within drawer by STA Compact®.
Pour relecture / For proofreading 11/9/09
2SAMPLE LOADING screen with MANUAL Mode or AUTO Mode displayed (as described in
chapter 6.4) or PRODUCT LOADING screen displayed (as described in chapter 6.3).
ACTIONS RESULTS
" Remove the tube (or vial) The LED adjacent to the tube (vial) lights off.
The identification of the tube (vial)
disappears from the list of already loaded
samples (products).
This unloading procedure is only possible for the products or samples of the open drawer.
This procedure is absolutely necessary for products or samples loaded without Positive
identification.
2SAMPLE LOADING screen with MANUAL Mode or AUTO Mode displayed (as described in
6-39
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
chapter 6.4)
or PRODUCT LOADING screen displayed (as described in chapter 6.3).
ACTIONS RESULTS
" Press the K key (List/Change) The cursor is positioned on the last line of
the zone for already performed loadings (see
description of SAMPLE LOADING screen
chapter 6.4 or description of PRODUCT
LOADING screen chapter 6.3).
" Move the cursor to the requested Patient Cursor positioned on the requested patient
file or product with the K key file or product.
" Press the Delete key For the sample drawer, the following
window is displayed:
SAMPLE LOADING
Are you sure you want to remove the
sample (product) from the drawer?
PRODUCT LOADING
Are you sure you want to remove the
product from the drawer?
Y=YES N=NO
" Remove the tube (or vial), then press the If the LED adjacent to the tube (vial) was lit,
Y key it goes off.
The identification of the tube (vial)
disappears from the list of already loaded
samples (products).
For all tubes or vials loaded with this procedure, it is absolutely necessary to observe the
unloading procedure without positive identification (see chapter 6.10.2).
6-40
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
2SAMPLE LOADING screen is displayed (with MANUAL Mode or AUTO Mode). Sample already
identified, tube ready to be loaded. Message “LOAD THE LAST IDENTIFIED TUBE - Esc = Repeat
ident” displayed at the bottom of the screen and cursor positioned in the “Pos.” box (see
description of the SAMPLE LOADING screen chapter 6.4).
ACTIONS RESULTS
" Load the sample tube (or the sample The tube position is not displayed on the
adaptator) in any position of the sample screen.
drawer On the sample drawer, the LED adjacent to
the tube does not light up.
" Enter the tube position on the keyboard by Position (i.e.: G4) displayed on the screen.
typing first the letter indicated at the left
The LED adjacent to the tube position lights
Pour relecture / For proofreading 11/9/09
6-41
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
2PRODUCT LOADING screen is displayed. Reagent, desorb solution, control or calibrator already
identified (Identification, Volume, Stability, Lot number) and vial ready to be loaded. Message
“LOAD THE LAST IDENTIFIED VIAL - Esc = Correction” displayed at the bottom of the screen
and cursor positioned in the “Pos.” Box.
ACTIONS RESULTS
" Place the vial in a position corresponding to The vial position is not displayed on the
its diameter while considering the positions screen.
with or without mixing
On the product drawer, the LED adjacent to
the vial does not light up.
" Enter directly on the keyboard the vial On the screen, display of the position and
position, as indicated on its right side data for this reagent, desorb solution,
Running tests or stopping tests can also be performed from the Status menu (see description
chapter 10.1.5.1).
As soon as the running of tests is confirmed by the operator, the samples are processed as described
below:
- Pipetting of calibrators for all tests
- Pipetting of calibration controls for all tests
- Pipetting of quality controls for all tests
- Pipetting of STAT samples
- Pipetting of samples for all tests in chronological loading order
6-42
0931033L - July 2009
REFERENCE MANUAL 6
Loading Products and Samples
2SAMPLE LOADING screen displayed with MANUAL Mode or AUTO Mode (see description
chapter 6.4); or PRODUCT LOADING screen displayed (see description chapter 6.3).
ACTIONS RESULTS
" Press the Esc key For the sample drawer, one of the following
windows is displayed:
For MANUAL Mode:
Change Profiles
Change Prefix
AUTO Mode
Return to Loading
Quit
Change Profile
Pour relecture / For proofreading 11/9/09
Quit
YES
NO
" Confirm with the key The sample drawer (product drawer) closes
automatically.
When all conditions are met (quality controls
or calibrations OK, no stability is overdue,
and no volume is insufficient) for all reagents
and products, the sample pipetting starts
again and the TEST STATUS screen is
displayed. The Esc key allows to return to
the TEST PANEL.
If one of the conditions previously described
is not met, for one or several tests, the
sample pipetting is blocked and the TEST
STATUS screen is displayed with the
message “BLOCKED SAMPLE PIPETTING”
displayed in red at the bottom of the screen
(see description of TEST STATUS screen
chapter 10.1.3).
6-43
0931033L - July 2009
6 REFERENCE MANUAL
Loading Products and Samples
ACTIONS RESULTS
" If sample pipetting has been blocked, press The following window is displayed:
the Esc key
WARNING
New tests are delayed
Reactivate?
Y=YES N=NO
" To reactivate tests, press the Y key The authorized sample pipetting starts again.
Return to TEST PANEL.
6-44
0931033L - July 2009
REFERENCE MANUAL 7
Table of contents
7 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
7-i
0931033L - July 2009
7 REFERENCE MANUAL
Table of contents
7-ii
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
7 Calibration
ACTIONS RESULTATS
" Move the cursor to the test for which The selected calibration screen is displayed
calibration is required using the cursor (see description below chapter 7.2).
keys L ,K , Jand confirm with the key
7-1
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
Legend:
2 Available keys:
F1 Select Calibrations to Run: requests calibration run for the test on which the cursor
is positioned
F10 Run Selected Calibrations: runs the calibrations selected by the F1 key (see
chapter 7.3.7)
3 List of test abbreviations. A colored square displayed at the left of the abbreviation indicates the
calibration current status (see point 5 below).
White: Validated
Blue: To be validated
Yellow: In progress
Red: Not calibrated
7-2
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
Legend:
7-3
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
OPTIONS
Information
Modify 100% Point - ISI
Print
Run Controls
Quit
Information: displays the second screen (Information screen) of the calibration.
Modify 100% Point - ISI: this option is only possible for clotting tests with % as the primary unit
and if the measurement scale is linear and the concentration scale is inverse.
It moves the calibration curve in a parallel direction to itself in such a way that the calibration passes
through the new point given in seconds and which corresponds to 100%.
It also allows access for ISI modification (International Sensitive Index) if INR (International
Normalized Ratio) has been chosen as the unit.
Print: (only possible if STA Compact< is in stand-by mode). Printout of the two calibration screens:
graphic and information.
Run Controls: modifies the calibration control ranges if those have been defined and runs them
(F10 key).
Quit: returns to CALIBRATION screen (test list).
6 Value in seconds for the point corresponding to 100%. It is displayed only if it has been modified
and if the modification is possible, see above, point 2
8 View points as defined in the TEST SETUP screens (see chapter 5.1.2.2)
7-4
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
NOTE The parameters come from the bar code reading performed if a lot number change has been
detected during reagent loading (see chapter 6.5).
21 Lot number of each reagent used for the calibration. An empty box means that the reagent has
no lot number
7-5
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
Legend:
Run Controls
Print
Quit
Run Controls: modifies the calibration control ranges if those have been defined and runs them
(F10 key).
Print: (only possible if STA Compact< is in stand-by mode). Printout of the calibration screen as
defined above
Quit: returns to CALIBRATION screen (test list)
4 Date and Time of validation for the calibration controls if those are used. Otherwise, the box is
empty.
7-6
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
13 Lot number of each reagent used for the calibration. An empty box means that the reagent has
no lot number
7-7
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
Legend:
4 Date and Time of validation for the calibration controls if those are used. Otherwise, the box is
empty
7-8
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
- Results displayed in red Calibration control results are validated although those were
and preceded by an asterisk: out of range.
14 Lot number of each reagent used for the calibration. An empty box means that the reagent has
no lot number.
15 ISI Value (International Sensitive Index), is only displayed if INR (International Normalized Ratio)
has been chosen as the unit.
7-9
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
7-10
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
Legend:
Options Options
Information Information
Modify 100 %. Point - ISI Modify Offset
Print Print
Calibrate Calibrate
Modify Raw Data Modify Raw Data
Quit Quit
7-11
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
6 Value in seconds for the point corresponding to 100%. It is displayed only if it has been modified
and if the modification is possible, see above, point 2
7 Regression coefficient
9 For clotting tests: clotting time measured in seconds. For assays in photometry (not available on
STA Compact®CT): ' O.D. or O.D./min. An S is displayed at the left of the suppressed values
and an M at the left of the modified values
10 Value of standards after interpolation from the equation found for the calibration curve (in the
primary unit)
13 Offset (not available on STA Compact®CT) : Value in 'O.D. or in O.D./min for the offset
corrector plasma if the latter has been run (see description of the Modifying Offset option
chapter 7.5.2)
7-12
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
Fig. 9 - CALIBRATION INFORMATION Screen for a Photometric Test (not available on STA
Compact®CT) - Graphic Mode
Legend:
16 Not applicable
20 Dilution ratio for each standard, defined in the Test Setup menu (see chapter 5.1.2.3).
7-13
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
29 Lot number of each reagent used for the calibration. An empty box means that the reagent has
no lot number.
30 Identification number, complete name and lot number of the offset corrector plasma.
For clotting tests, this box is always empty.
The calibrations can be run only if the sample pipetting has not been blocked either within the TEST
STATUS screen (see description chapter 10.1.3) or by the Stop Sample Pipetting function of the
Status sub-menu (see description chapter 10.1.5.1).
To block tests from the Global Modification menu does not block the calibrations (see description
chapter 8.7.3).
Mode Calibrate
7-14
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
Use of calibration controls authorizes automatic validation of the calibrations. There must be 2 levels
of controls because a slope cannot be validated as fully reliable with only one level of controls.
NOTE The calibration modes as all the calibration parameters (standard definition, calibration control
definition, scales) are defined within the TEST SETUP screens (see chapter 5.1.2).
Standards are run in single or in duplicate according to the determination mode chosen within the
TEST SETUP screens (see description chapter 5.1.2).
When they are defined, the calibration controls are run in single or in duplicate according to the
determination mode chosen for the sample (see Test Setup menu chapter 5.1.1.7).
As soon as calibration control results are obtained, they are compared to the acceptance range; if
those are outside of the acceptance range, STA Compact< automatically reruns controls according
to principle described below:
Control run in single Outside the defined range Rerun of control in duplicate
If the calibration controls are confirmed as outside the acceptance range, the calibration will have to
be validated manually.
If the operator decides to accept out of range calibration controls, the results of these will be
displayed in red, preceded by an asterisk in the related CALIBRATION screen (see chapter 7.2).
If the same controls are used as calibration controls and as quality controls (same identification
numbers), then the calibration control results will be integrated in the quality control results and the
clock for the quality controls will be reset while running the calibration controls (see chapter 9.4.1).
If running has been requested for one or several calibrations or calibration controls, when existing
the CALIBRATION screen (test list), STA Compact< performs a consistency check and displays the
TEST STATUS screen.
If while running a calibration, some products necessary to perform the calibration are missing
(missing or on-board but with volume or stability not usable), then the calibration will not be run, the
missing products will be displayed in red in the TEST STATUS screen and the sample pipetting will be
blocked (see description chapter 10.1.3).
7-15
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
At any time, the operator can request calibration controls, see below.
ACTIONS RESULTATS
OPTIONS
Information
Modify 100% Point - ISI
Print
Run Controls
Quit
with cursor positioned on Quit.
7-16
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Press the R key (Run Controls) The following window is displayed
" Or move the cursor to Run Controls with
the Kkey then confirm with the key Please enter
your access code:
Validate Esc Quit
" Type your access code then confirm with The second screen (Information screen) is
the key displayed (see description chapter 7.2.1).
The cursor is positioned to modify, if
necessary, the minimum value of the range
of the first calibration control (see description
of this screen chapter 7.2.1).
" If necessary, modify each range value, then Cursor positioned on the following range
confirm with the key value.
" Otherwise, use the key to move from a
range value to the other one
" When all range values are correct, press Back on the selected calibration screen.
the F10 key
The calibration controls are automatically
run by STA Compact< so long as the
consistency check does not lead to a
blocking of the sample pipetting.
" Press the Esc key Back to CALIBRATION screen (test list).
The colored square at the left of the
abbreviation for the selected test is
displayed in yellow (calibration in progress).
If the calibration control results are within
the expected ranges:
- automatic calibration validation;
- the colored square at the left of the
abbreviation for the selected test is
displayed in white (confirmed calibration)
when the CALIBRATION screen (test list) is
visualized.
if the calibration control results are not
within the expected ranges: an error
message is displayed (see chapter 7.4).
7-17
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
NOTE For raw mode, there exists no calibration; for a new lot, the only possible function is to validate the
reagent(s) of the method by running the 2 calibration controls if those have been defined.
If both calibration control results are within the expected ranges, the calibration is automatically
validated, patient assays may start.
If calibration control results are outside the expected ranges, the operator will be informed by an
error message and will be offered several possibilities (see chapter 7.4).
If calibration controls are not used, the calibration status is confirmed by default.
2Screen for a calibration in raw mode displayed (as described in chapter 7.2.2).
ACTIONS RESULTATS
Run Controls
Print
Quit
with cursor positioned on Quit.
" Press the R key (Run Controls) The following window is displayed:
" Or move the cursor to Run Controls with
the L key, then confirm with the key Please enter
your access code:
Validate Esc Quit
" Type your access code then confirm with The cursor is positioned to modify, if
the key necessary, the minimum value of the range
of the first calibration control (see description
of the screen chapter 7.2.2).
" If necessary, modify each range value, then The cursor is positioned on the following
confirm with the key range value.
" Otherwise, use the key to move from a
range value to the other one
" When all range values are correct, press Back on the selected calibration screen.
the F10 key
The calibration controls are automatically
run by STA Compact< so long as the
consistency check does not lead to a
blocking of the sample pipetting.
7-18
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Press the Esc key Return to the CALIBRATION screen (test
list).
The colored square at the left of the
abbreviation of the selected test is displayed
in yellow (calibration in progress).
If the calibration control results are within
the expected ranges:
- automatic calibration validation;
- the colored square at the left of the
abbreviation for the selected test is
displayed in white (confirmed calibration)
when the CALIBRATION screen (test list) is
visualized.
If the calibration control results are not
within the expected ranges: an error
message is displayed (see chapter 7.4).
NOTE For Ratio Mode, calibrations mean entering the reference time, then validating the reagent(s) used
for the method by running the two calibration controls if they have been defined.
If both calibration control results are within the expected ranges, the calibration is automatically
validated, patient assays may start.
If calibration control results are outside the expected ranges, the operator will be informed by an
error message and will be offered several possibilities (see chapter 7.4).
7.3.5.1 Running and Validating Calibrations in Ratio Mode with Calibration Controls
ACTIONS RESULTATS
7-19
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" Press the M key (Modify Reference The following window is displayed:
Time / Range)
" Or move the cursor to Modify Reference Please enter
Time / Range with the L key then confirm your access code:
with the key Validate Esc Quit
" Type your access code, then confirm with The following window is displayed:
the key
CALIBRATION CONTROL
F10 Run
Esc Quit
The cursor is positioned to modify, if
necessary, the reference time (see
description of the screen chapter 7.2.3).
" If necessary, modify the value of the The cursor is positioned to modify, if
reference time, then confirm with the necessary, the minimum value of the range
key of the first calibration control (see description
of the screen chapter 7.2.3).
" Otherwise, confirm with the key
" If necessary, modify each range value, then The cursor is positioned on the following
confirm with the key range value.
" Otherwise, use the key to move from a If INR unit has not been chosen, skip to last
range value to the other one but one ACTIONS box below.
If INR unit has been chosen among the
secondary units, the cursor is positioned to
modify, if necessary, the ISI value (see
description chapter 7.2.3).
" If necessary, modify the ISI value, then ISI value is displayed.
confirm with the key
" When all modifications are performed, Back on the selected calibration screen.
press the F10 key
The calibration controls are automatically
run by the STA Compact< so long as the
consistency check does not lead to a
blocking of the sample pipetting.
7-20
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Press the Esc key Return to CALIBRATION screen (test list).
The colored square at the left of the
abbreviation for the selected test is
displayed in yellow (calibration in progress).
If the calibration control results are within
the expected ranges:
- automatic calibration validation;
- the colored square at the left of the
abbreviation for the selected test is
displayed in white (confirmed calibration)
when the CALIBRATION screen (test list) is
visualized.
If the calibration control results are not
within the expected ranges: an error
message is displayed (see chapter 7.4).
7.3.5.2 Running and Validating Calibrations in Ratio Mode without Calibration Controls
ACTIONS RESULTATS
" Press the M key (Modify Reference The following window is displayed:
Time / Range)
" Or move the cursor to Modify Reference Please enter
Time / Range with the Lkey then confirm your access code:
with the key Validate Esc Quit
" Type your access code, then confirm with The following window is displayed:
the key
REFERENCE TIME MODIFICATION
F10 Save
Esc Quit
The cursor is positioned to modify, if
necessary, the reference time value (see
description of the screen chapter 7.2.3).
7-21
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" If necessary, modify the reference time If INR unit has not been chosen, skip the
value, then confirm with the key ACTIONS cell below and move to the next
" If necessary, modify the ISI value, then ISI value is displayed.
confirm with the key
" When all modifications are performed, Return to CALIBRATION screen (test list).
press the F10 key
The colored square at left of the
abbreviation for the selected test is
displayed in white (confirmed calibration).
NOTE For graphic mode, calibration means running standards, then calibration controls if those have been
defined.
If both calibration control results are within the expected ranges, the calibration is automatically
validated, patient assays may start.
If calibration control results are outside the expected ranges, the operator will be informed by an
error message and will be offered several possibilities (see chapter 7.4).
If calibration controls are not used, the calibration validation will be done manually (see chapter 7.4).
7.3.6.1 Running and Validating Calibrations in Graphic Mode with Calibration Controls
7-22
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Press the Esc key For clotting tests, the following window is
displayed:
OPTIONS
Information
Modify 100% Point - ISI
Print
Calibrate
Modify Raw Data
Quit
For photometric tests (not available on STA
Compact®CT), the following window is
displayed:
OPTIONS
Information
Modify Offset
Print
Calibrate
Modify Raw Data
Quit
In both cases, the cursor is positioned on
Quit.
" Type your access code, then confirm with The second screen (INFORMATION screen)
the key is displayed, (see description chapter 7.2.4).
The cursor is positioned to modify, if
necessary, the concentration for the first
standard (see description of this screen
chapter 7.2.4).
" If necessary, modify the concentration of For all standards having the same
the first standard, then confirm with the identification number, the concentrations
key are automatically calculated and displayed.
" Otherwise, confirm with the key The cursor is positioned on the second
standard.
" If other standard concentrations have to be For all standards having the same
modified, modify them, then confirm with identification number, the concentrations
the key are automatically calculated and displayed.
The cursor is positioned on the next
standard or on the minimum value of the
range of the first calibration control.
7-23
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" Move the cursor to the minimum value of The cursor is positioned on the next range
the range of the first calibration control, value.
modify, if necessary, each range value, then
confirm with the key
" Otherwise, use the key to move from a
range value to the other one
" When all parameters are set, confirm with Back on the selected calibration screen.
the F10 key The standards, then the calibration controls
are automatically run by the STA Compact<
so long as the consistency check does not
lead to a blocking of the sample pipetting.
" Press the Esc key Back to CALIBRATION screen (test list).
The colored square at the left of the
abbreviation for the selected test is
displayed in yellow (calibration in progress).
If the calibration control results are within
the expected ranges:
- automatic calibration validation;
- the colored square at the left of the
abbreviation for the selected test is
displayed in white (confirmed calibration)
when the CALIBRATION screen (test list) is
visualized.
If the calibration control results are not
within the expected ranges: an error
message is displayed (see chapter 7.4).
7.3.6.2 Running and Validating Calibrations in Graphic Mode without Calibration Controls.
7-24
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Press the Esc key For clotting tests, the following window is
displayed:
OPTIONS
Information
Modify 100% Point - ISI
Print
Calibrate
Modify Raw Data
Quit
For photometric tests (not available on STA
Compact®CT), the following window is
displayed:
OPTIONS
Information
Modify Offset
Print
Calibrate
Modify Raw Data
Quit
In both cases, the cursor is positioned on
Quit.
" Move the cursor to Calibrate with the The following window is displayed:
n key, then confirm with the key
Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The second screen (INFORMATION screen)
the key is displayed, (see description chapter 7.2.4).
The cursor is positioned to modify, if
necessary, the concentration for the first
standard (see description of this screen
chapter 7.2.4).
" If necessary, modify the concentration of For all standards having the same
the first standard, then confirm with the identification number, the concentrations
key are automatically calculated and displayed.
" Otherwise, confirm with the key The cursor is positioned on the second
standard.
" If other concentrations have to be modified, For all standards having the same
modify them, then confirm with the key identification number, the concentrations
are automatically calculated and displayed.
7-25
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" When all parameters are set, press the Back on the selected calibration screen.
F10 key
The standards are automatically run by the
STA Compact< so long as the consistency
check does not lead to a blocking of the
sample pipetting.
" Press the Esc key Return to CALIBRATION screen (test list).
The colored square at the left of the
abbreviation for the selected test is
displayed in yellow (calibration in progress).
When the calibration is complete, an
information message is displayed.
This function is reserved to tests predefined by Diagnostica Stago because it implies that no
operator’s intervention is requested to modify standard concentrations and/or calibration control
concentrations. And this is only possible for the bar coded products (standards, controls) for which
the value transfer is automatically performed.
ACTIONS RESULTATS
" For each test whose calibration is The letter S is displayed in grey at the left of
requested, move the cursor to its each test abbreviation for which the
abbreviation then press the F1 key calibration has been requested.
Note: To cancel a running request, press
the F1 key again.
7-26
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" When all required calibrations are marked The following window is displayed:
with the letter S, press the F10 key
Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with On CALIBRATION screen (test list):
the key
- the S letters disappear;
- the colored squares at the left of the
abbreviations for the selected tests are
displayed in yellow (calibration in progress).
The standards and the calibration controls (if
necessary) are automatically run by the
STA Compact< so long as the consistency
check does not lead to a blocking of the
sample pipetting.
" Press the Esc key The TEST STATUS screen is displayed (see
description chapter 10.1.3).
If the calibration control results are within
the expected ranges:
- automatic calibration validation;
- the colored square at the left of the
abbreviation for the selected test is
displayed in white (confirmed calibration)
when the CALIBRATION screen (test list) is
visualized.
If the calibration control results are not
within the expected ranges: an error
message is displayed (see chapter 7.4).
If calibration controls are not used, please
see chapter 7.4.
For calibrations with graphic mode, as long as the reagent lot numbers have not changed, the new
calibration curve is displayed in yellow and the previous one in grey. Raw data, curve equation and
regression coefficient are those of the new calibration curve.
7-27
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
MEASURING MANAGEMENT
Error 05.06.06
CALIBRATION CONTROL
Control out of limit
for the test xxx
Esc=continue
xxx: Abbreviation of the related test.
Absence of control:
MEASURING MANAGEMENT
Error 05.06.10
CALIBRATION
Calibration is ready to be confirmed
for the test xxx
Esc=continue
xxx: Abbreviation of the related test.
7-28
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
7.4.2.1 Description of the CALIBRATION NOT CONFIRMED Window (out of Range Controls)
Legend:
3 Recall of the minimum and maximum ranges for each calibration control
7-29
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
This procedure is only possible if the operator knows the access code to accept out of range
controls (see chapter 1.5).
NOTE If for a given test, the calibration controls are confirmed while out of range, then results of these
controls will be displayed in red with an asterisk in the related CALIBRATION screen (see chapter
7.2).
In the CALIBRATION NOT CONFIRMED window (out of Range Controls), the out of range controls
are displayed in red with an asterik and the ranges of each control are displayed (see description
chapter 7.4.2.1).
2The screen of the selected calibration is displayed with the CALIBRATION NOT CONFIRMED
window (out of Range Controls), cursor is positionned on Postpone Decision (see description
chapter 7.4.2.1).
7-30
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Move the cursor to Confirm with the The following window is displayed:
L key, then confirm with the key
Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The selected CALIBRATION screen is
the key updated for the date, the time, the
calibration control results.
Out of range calibration control results are
displayed in red with an asterisk.
" Press the Esc key then confirm with the Return to CALIBRATION screen (test list).
key
The colored square at the left of the test
abbreviation is displayed in white (confirmed
calibration).
NOTE This function is possible only if calibration controls have been defined.
2The screen of the selected calibration is displayed with the CALIBRATION NOT CONFIRMED
window (out of Range Controls), the cursor is positioned on Postpone Decision (see
description chapter 7.4.2.1).
ACTIONS RESULTATS
" Type your access code, then confirm with The CALIBRATION CONTROL RUNNING
the key window appears.
Controls are rerun by STA Compact< so
long as the consistency check does not lead
to a blocking of the sample pipetting.
7-31
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" Press the Esc key Return to CALIBRATION screen (test list).
The colored square at the left of the test
abbreviation is displayed in yellow
(calibration in progress).
" Press the Esc key The TEST STATUS screen is displayed (see
description chapter 10.1.3).
This procedure is possible only if the calibration controls were not used.
2Screen of the selected calibration displayed with the CALIBRATION NOT CONFIRMED window
(Absence of Controls), the cursor is positioned on Postpone Decision (see description
chapter 7.4.2.2).
ACTIONS RESULTATS
" Move the cursor to Confirm with the The following window is displayed:
L key, then confirm with the key
Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The CALIBRATION NOT CONFIRMED
the key (Absence of Controls) window disappears.
" Press the Esc key then confirm with the Return to CALIBRATION screen (test list).
key
The colored square at the left of the test
abbreviation is displayed in white (confirmed
calibration).
7-32
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
This procedure is possible only if the operator knows the access code to critical functions (see
chapter 1.5).
NOTE This procedure is possible only if the calibration controls were not used.
2The screen of the selected calibration is displayed with the CALIBRATION NOT CONFIRMED
window (Absence of Controls), the cursor is on Postpone Decision (see description
chapter 7.4.2.2).
ACTIONS RESULTATS
" Move the cursor to Modify/Validate with The following window is displayed:
the K key, then confirm with the key
CORRECTION
DEL DELETE
Ins RESTORE
F10 CONFIRM
ESC QUIT
7-33
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" When all modifications are done, press the The following window appears:
F10 key
Please enter
your access code:
Validate Esc Quit
" Type your access code then confirm with The new calibration is recalculated then the
the key following are displayed: the new calibration
curve, the new equation and the new
regression coefficient.
A yellow S is displayed near the deleted
value(s) and a yellow M near the modified
value(s).
Deleted values are still displayed in grey.
" Press the Esc key then the key Return to CALIBRATION screen (test list).
The colored square at the left of the test
abbreviation is displayed in white (confirmed
calibration).
This procedure is possible only if the operator knows the access code to critical functions (see
chapter 1.5).
NOTE This procedure is only possible for calibration curves performed with the same reagent lot numbers
as previously used.
2Screen of the selected calibration displayed with the CALIBRATION NOT CONFIRMED window
(Out of Range Controls or Absence of Controls), the cursor is on Postpone Decision.
7-34
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Move the cursor to Restore Previous The following window is displayed:
Curve with the K key, then confirm with
the key Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The CALIBRATION NOT CONFIRMED
the key window disappears.
The screen of the selected calibration is
displayed again with the previous calibration
and all associated parameters (equation,
regression coefficient, measured values,
interpolated concentrations).
2Screen of the selected calibration displayed with the CALIBRATION NOT CONFIRMED window
(Out of Range Controls or Absence of Controls), the cursor is positioned on Postpone
Decision.
ACTIONS RESULTATS
" Type your access code, then confirm with The CALIBRATION RUNNING window
the key appears.
The standards (and the calibration controls if
those are used) are rerun by STA Compact<
so long as the consistency check does not
lead to a blocking of the sample pipetting.
7-35
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" Press the Esc key Return to CALIBRATION screen (test list).
The colored square at the left of the test
abbreviation is displayed in yellow
(calibration in progress).
" Press the Esc key The TEST STATUS screen is displayed (see
description chapter 10.1.3).
This option is only available if the calibration has been run with new reagent lot numbers.
2Screen of the selected calibration displayed with the CALIBRATION NOT CONFIRMED window
(out of Range Controls or Absence of Controls), the cursor is positioned on Postpone
Decision.
ACTIONS RESULTATS
" Move the cursor to Cancel the The CALIBRATION NOT CONFIRMED
Calibration with the K key, then confirm window disappears.
with the key
The calibration curve disappears.
A red square is displayed before the
abbreviation of the test with calibration
status: not calibrated.
7-36
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
This procedure is only possible if the operator knows the access code allowing critical functions (see
chapter 1.5).
NOTE This option is only possible for clotting tests with % as primary unit and if the measurement scale
is linear and the concentration scale is inverse.
After the 100% point is entered, a parallel curve is displayed that passes through the new 100%
point..
2Screen of the selected calibration displayed with the OPTIONS window (cursor positioned on
Quit).
7-37
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" Move the cursor to Modify 100% Point - The following window is displayed:
ISI with the K key, then confirm with the
key Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The following window is displayed:
the key
MODIFICATION
F10 Save
Esc Quit
The cursor is positioned on the entry field
for:
• The ISI value if INR has been chosen as an
auxiliary unit [see description of the
CALIBRATION screen (bar coded mode)
chapter 7.2.1 and of the CALIBRATION
screen (graphic mode), chapter 7.2.4].
• Modifying the 100 % point [see description
of the CALIBRATION screen (bar coded
mode) chapter 7.2.1 and of the
CALIBRATION screen (graphic mode),
chapter 7.2.4].
In this last case, skip the ACTIONS box below.
" Move the cursor with the L key The cursor is positioned on the entry field
for:
• Modifying the 100 % point [see description
of the CALIBRATION screen (bar coded
mode) chapter 7.2.1 and of the
CALIBRATION screen (graphic mode )
chapter 7.2.4].
" Type in seconds the time corresponding to The screen of the selected calibration is
the 100 % of the lab, then confirm with the displayed with previous calibration drawn in
key stippled yellow and the new one in yellow.
In the curve equation, the slope is kept and
the Y intercept is recalculated.
The interpolated concentrations are
recalculated.
" Save the modificatons with the F10 key If calibration controls are defined, they are
automatically run by STA Compact<so long
as the consistency check does not lead to a
blocking of the sample pipetting.
7-38
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Press the Esc key Return to CALIBRATION screen (test list).
If calibration controls are defined, the
colored square at left of the abbreviation of
the selected test is displayed in yellow
(calibration in progress).
If calibration controls are not defined, the
colored square at left of the abbreviation of
the selected test is displayed in white
(confirmed calibration).
" Press the Esc key If calibration controls are defined, the TEST
STATUS screen is displayed (see description
chapter 10.1.3).
If calibration controls are not defined, return
to TEST PANEL.
NOTE This function is only possible for photometric (chromogenic or immunological) tests with linear
graphic mode and if the offset corrector and the calibration controls have been defined. It moves
the calibration curve in parallel direction to itself in such a way that it passes through the new points
defined by the offset corrector plasma.
2CALIBRATION screen displayed for a photometric test with the OPTIONS window (cursor
positioned on Quit).
7-39
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" Move the cursor to Modify Offset with the The following window is displayed:
L key, then confirm with the key
Please enter your
access code:
Validate Esc Quit
" Type your access code, then confirm with A screen similar to the information screen is
the key displayed (see description chapter 7.2.4) with
information on the offset corrector plasma
instead of information on the standards.
The cursor is positioned to modify, if
necessary, the concentration of the offset
corrrector plasma.
" If necessary, modify the concentration of The cursor is positioned on the minimum
the offset corrector plasma then confirm value of the range of the first calibration
with the key control.
" If necessary, modify each range value, then The cursor is positioned on the following
confirm with the key range value.
" When all parameters are set, press the Back on the selected calibration screen.
F10 key The offset corrector plasma and the
calibration controls are automatically run by
STA Compact< so long as the consistency
check does not lead to a blocking of the
sample pipetting.
" Press the Esc key Return to CALIBRATION screen (test list).
The colored square at the left of the
abbreviation for the selected test is
displayed in yellow (calibration in progress).
" Press the Esc key The TEST STATUS screen is displayed (see
description chapter 10.1.3).
7-40
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
This procedure is possible only if the operator knows the access code to critical functions (see
chapter 1.5).
NOTE This procedure is also possible on calibrations whose status is to be confirmed (see procedure
chapter 7.4.6).
ACTIONS RESULTATS
" Move the cursor to Modify Raw Data The following window is displayed:
with the K key, then confirm with the
key Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The following window is displayed:
the key
CORRECTION
DEL DELETE
Ins RESTORE
F10 CONFIRM
ESC QUIT
7-41
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
ACTIONS RESULTATS
" When all modifications are done, validate The new calibration is recalculated then the
with F10 key following are displayed: the new calibration
curve, the new equation, and the new
regression coefficient.
A yellow S is still displayed near deleted
value(s) and a yellow M appears near
modified value(s)
Deleted value(s) are still displayed in grey.
" Press the Esc key, then the key Return to CALIBRATION screen (test list).
If calibration controls are defined, they are
automatically run by STA Compact< so long
as the consistency check does not lead to a
blocking of the sample pipetting.
" Press the Esc key If calibration controls are defined, the TEST
STATUS screen is displayed (see description
chapter 10.1.3).
If calibration controls are not defined, return
to TEST PANEL.
a. not available on STA Compact®CT
7-42
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
This function is possible only if the STA Compact< is in stand-by mode (no measurements in
progress) and if all printouts of patient files previously generated have been performed.
NOTE Depending on the calibration mode, one or two calibration screens are printed (for details, see
calibration screen descriptions chapter 7.2.1 to 7.2.4).
ACTIONS RESULTATS
" When the printout is complete, press the The Printout window disappears.
Esc key
Back to the selected calibration screen.
7-43
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
The ISI value for the Prothrombin must be the ISI value indicated on the insert of the STA
product.
it is imperative that the user check the ISI value before leaving the menu if there has been
a lot change, a software update, or any major change.
An incorrect ISI value can lead to inaccurate INR (International Normalization Ratio)
results.
This procedure is only available if the operator knows the access code to critical functions, (see
chapter 1.5).
The ISI (International Sensitivity Index) value is modified differently depending on the calibration
mode:
- for ratio modes, it can be accessed through the Modify reference time function (see description
in chapter 7.3.5),
- for graphic and bar coded modes, see procedure below.
2Screen of the selected calibration displayed with the OPTIONS window (cursor on Quit).
ACTIONS RESULTATS
" Move the cursor to Modify100% Point - The following window is displayed:
ISI with the L key, then confirm with the
key Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The following window is displayed:
the key
MODIFICATION
F10 Save
Esc Quit
The cursor is positioned on the entry field
for the ISI value (see description of the
CALIBRATION screen (bar coded mode),
chapter 7.2.1 and of the CALIBRATION
screen (graphic mode), chapter 7.2.4.
7-44
0931033L - July 2009
REFERENCE MANUAL 7
Calibration
ACTIONS RESULTATS
" Type the new ISI value, then confirm with The new ISI value is displayed near ISI
the F10 key Coefficient (see description of the
CALIBRATION screen (bar coded mode),
chapter 7.2.1 and of the CALIBRATION screen
(graphic mode), chapter 7.2.4).
NOTE This function is available only when there are two screens to define a calibration: bar coded mode
or graphic mode (linear, hyperbolic, polynomial 2nd order or polynomial 3rd order).
2Screen of selected calibration displayed with the OPTIONS window, cursor is positioned on Quit.
ACTIONS RESULTATS
" Move the cursor to Information with the The Information screen is displayed (see
L key, then confirm with the key description of this screen in chapters 7.2.1 and
7.2.4).
7-45
0931033L - July 2009
7 REFERENCE MANUAL
Calibration
7-46
0931033L - July 2009
REFERENCE MANUAL 8
Table of contents
8.1 Generalities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
8-i
0931033L - July 2009
8 REFERENCE MANUAL
Table of contents
8-ii
0931033L - July 2009
REFERENCE MANUAL 8
Table of contents
8-iii
0931033L - July 2009
8 REFERENCE MANUAL
Table of contents
8-iv
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
8.1 Generalities
STA Compact® manages one file containing 600 Patient Files: the Working File.
As soon as a Patient file is created, it is stored in the Working File whether the corresponding sample
tube is loaded or not in the sample drawer.
On the Patient files with the corresponding sample tubes loaded in the sample drawer, the following
operations are possible:
- Confirmation of tests or files
- File printout
- File transmission to a host computer
- Deletion of tests within Patient files
- Rerun of tests
- Insertion of tests in files
- Change a file from non-urgent to STAT and vice versa
- Deletion of a file selection
- Addition of complementary information (name, first name, ... ) on the files
All these operations can be performed directly from the TEST PANEL screen by positioning the cursor
on the desired file or from the Files menu, CONFIRMATION/PRINTOUT screen.
Tests may also be blocked, unblocked, deleted or rerun as a global action (see chapter 8.7).
NOTE This menu allows access to the Patient files of the Working file
ACTIONS RESULTS
8-1
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
Confirmation/Printout/Transmission
Delete Patient Files
Customer Printout
with the cursor on Confirmation /
Printout/Transmission.
Legend:
1 Available keys:
Processes the file on which the cursor is positioned (see chapter 8.3)
8-2
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
2 Complementary information (name, first name ... see description chapter 10.2.3.3), concerning
the file on which the cursor is positioned
3 Recall of color code signification as well as the number of files corresponding to each category
for the entire Working File
- Two letters (T and P) confirm if the transmission (T) or the printout (P) have been done. The two
letters may not appear at all (no transmission done nor printout) or simultaneously (file
transmitted and printed out)
- A colored square provides the status of each file:
Blue: Confirmed file
White: Complete file
Red: Complete file with error
Yellow: Incomplete file (including non loaded tubes and loaded tubes not yet run)
ACTIONS RESULTS
" Press the Esc key (Options) The following sub-menu is displayed:
" Press the C key (Change File Selection) The FILE SELECTION window is displayed
(see description chapter 8.2.3.2) with cursor
" Or move the cursor to Change File
positioned in area 1.
Selection with the L key then confirm
with the key
8-3
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
Legend:
3 Available functions:
F1 “Beginning of List” displays the first file of the chronological file listing in the Working File
F2 “End of List” displays the last file of the chronological file listing in the Working File
5 Category of the required files, the key allows display or deletion of the colored square
corresponding to the category of the required file. This functions allows a better filtering.
8-4
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
Type the identification of the first file of the The identification of the first file of the list is
desired selection then confirm with the displayed in area 1 or “Beginning of List” is
key displayed in area 1 and the cursor is
Or press the F1 key to display “Beginning positioned in area 2 (see description of the
of List” (corresponding to the first file (the FILE SELECTION window chapter 8.2.3.2).
oldest) of the Working File)
Or press the F3 key to select the files with
the cursor. In this case, use the moving
cursor keys to move it to the required file,
then confirm with the key
" Type the identification of the last file of the The identification of the last file of the list is
desired selection then confirm with the displayed in area 2 or “End of List” is
key displayed in area 2 and the cursor is
" Or press the F2 key to display “End of positioned in area 4 (see description of the
List” (corresponding to the last file (most FILE SELECTION window chapter 8.2.3.2).
recent file) of the Working File)
" To select the last file with the cursor, use
the moving cursor keys to move it to the
required file, then confirm with the key
" To filter: type the prefix identification, then Filtering prefix identification displayed in the
confirm with the key (If the cursor is not area 4 (if defined) with the cursor
positioned in area 4, move it with the positioned in area 5 (see description of the
L key FILE SELECTION window, chapter 8.2.3.2 ).
" Or type the first number(s) or letter(s) of
the patient files identification, then confirm
with the key
" Or leave the area empty (no prefix
definition) and confirm with the key
8-5
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" To improve the filtering, use the key to The colored square appears near the type of
display a colored square near the type of desired files.
desired files. If the square is already
displayed, move the cursor with the L key
" If a type of file is not desired, use the key The colored square disappears.
to remove the colored square
" When all the parameters of the desired The new selection is displayed on the
selection are set, press the F10 key CONFIRMATION/PRINTOUT screen (see
description chapter 8.2.2).
ACTIONS RESULTS
" Press the Esc key (Options) The following sub-menu is displayed:
" Confirm with the key If no test has been added in patient files, back
on the TEST PANEL screen.
If tests have been added or rerun in patient
files, TEST STATUS screen is displayed (see
chapter 10.1.3).
8-6
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
" Move the cursor to the desired file with the The cursor is positioned on the desired file.
K and J keys
" Confirm with thekey The screen FILE PROCESSING for the
Working File is displayed (see description
below).
8.3.2 Description of the FILE PROCESSING Screen for the Working File
Fig. 3 - FILE PROCESSING screen
(from CONFIRMATION/PRINTOUT screen)
8-7
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
Legend:
2 Recall of the validation limits, for the test on which the cursor is positioned
3 Raw data (time or O.D.1/min. or ' O.D.1) and final results in primary units for each
determination and for the mean, for the test on which the cursor is positioned.
In case of duplicate determination, STA Compact® compares the difference between the two
results by means of the PRECISION parameter defined within the page 1 of 3 of the TEST SETUP
screens (see description chapter 5.1.1.7).
If the calculated difference is greater than the defined one, STA Compact® automatically reruns
a third test, compares the three results and keeps the best result pair.
5 File status:
8-8
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
6 File Identification:
7 Test list of the file. An asterisk appears on the left for tests with alarm(s) (see point 12 hereafter).
9 For each test, mean of the result, in the first secondary unit
10 For each test, mean of the result, in the second secondary unit
8-9
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
For each test, the areas 7, 8, 9, 10, and 11 are displayed with a specific color depending on the
results status:
Blue: Confirmed
White: To be confirmed
Black: Waiting;
Grey: Rerun
Red: Error
12 For the test on which the cursor is positioned, type of alarm code if necessary:
The alarms are classified by priority order (from A to L). The first one appears totally, the other ones
are represented by alphabetical codes and can be viewed by pressing the * key.
13 Adds or selects a note among the standard ones by pressing the F9 key (see procedure in
chapters 8.3.6 to 8.3.8).
Esc List: Returns to the CONFIRMATION/PRINTOUT screen (not available from TEST PANEL)
Esc Quit: Returns to TEST PANEL if no tests have been inserted or rerun; otherwise displays the
TEST STATUS screen if one or several tests have been inserted or rerun (see description of this
screen chapter 10.1.3), not available from CONFIRMATION/PRINTOUT screen
8-10
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
F1 Confirm File - All results with status To confirm are changed to status Confirm; upon
saving (F10 key), if all result status are Confirm, then the file status is changed to Confirm
F2 Accept Test - The test status on which the cursor is positioned is changed to Confirm if its
original status allows it: To confirm or Dev. > tol. or Value > max. (when it is equivalent to
time > T max.)
F3 Rerun Test - The test on which the cursor is positioned will be rerun. Its status is changed to
rerun.
For duplicate determination, the tests are rerun in duplicate. The current results have to be
printed out because they will disappear.
For single determination, the previous value is stored and a new single determination is run. The
parameter PRECISION defined on the first page of the TEST SETUP screens (see description
chapter 5.1.1.7), is used to compare both results. If the difference is greater than the defined
one, the STA Compact® automatically reruns the test a third time, compares the three results
and keeps the best pair of results.
- When returning to the TEST PANEL, the STA Compact® automatically displays the TEST STATUS
screen (see chapter 10.1.3).
- Test rerun is also possible in File Modification screen (see chapter 8.7).
F4 Delete Test - The test on which the cursor is positioned will be deleted. Its status is changed
to erased.
Note:
Test deleting can also be done in File Modification screen (see chapter 8.7).
F5 Insert Test - Accesses a TEST SELECTION window; move the cursor to the required test and
confirm with the key. This is only possible for tests which are not already selected.
When returning to the TEST PANEL, the STA Compact® automatically displays the TEST STATUS
screen (see chapter 10.1.3).
F10 Save - Records all the modifications applied to the displayed patient file
F11 Save + Transmit - Records all the modifications applied to the displayed patient file and
transmits the file (for more details on transmission see chapter 8.4)
KL Moves cursor
8-11
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
NOTE Only the complementary information (demographic parameters) as defined in the Global Options
menu can be added (see chapter 10.2.3.3).
ACTIONS RESULTS
" With the first test in the list highlighted, use The cursor is positioned on the first area
the K key to move the cursor to area 6, (see available for complementary information
description chapter 8.3.2) (see description chapter 8.3.2).
" Type the desired information then confirm The cursor moves to the following area.
with the J key
" When all complementary information is The cursor moves back to the first test of
defined, confirm with the key the test list (see description chapter 8.3.2).
8-12
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
" With the last test in the list highlighted, use " The cursor is positioned in the “Note” area
the L key to move the cursor to the “Note” (see chapter 8.3.2).
area (see chapter 8.3.2 ) " The message “Optional notes - F9: using
standard notes” is displayed at the bottom
of the screen.
" Type the required note with the keyboard " Cursor is positioned back on the last test of
then confirm with the key the list of the tests of the patient file.
2A FILE PROCESSING screen displayed, with the cursor in the “Note” area (see description
chapter 8.3.2) and the message “Optional notes - F9 = using standard notes” is displayed at
the bottom of the screen.
ACTIONS RESULTS
" Move the cursor to the desired note with The list of standard notes disappears and the
the L key then confirm with the key chosen note is displayed in the “Note” zone
(see description chapter 8.3.2).
" Press the key Cursor is positioned back on the last test of
the list of the tests of the patient file.
2A FILE PROCESSING screen displayed, with the cursor in the “Note” field (see description
chapter 8.3.2) and the message “Optional notes - F9 = using standard notes” is displayed at
the bottom of the screen.
ACTIONS RESULTS
" Move the cursor with the L key to a free The cursor is positioned in the desired field.
field to create a standard note or to an
existing note to modify it
8-13
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" Press the F2 key, type the required text, The new (or modified) note is now included
then confirm with the key in the list of standard notes.
" Press the Esc key Back to the FILE PROCESSING screen.
8.4.1 Conventions
All the results are transmitted according to an ASTM protocol. For more information, see document
“User Interface RS 232-C STA Compact< ASTM”, Cat. No. 28062.
All results are transmitted with two codes. The first one defines the result status (see description
chapter 8.3.2, point 11), and the second one the type of alarm if necessary (see description chapter
8.3.2, point 12).
8.4.2 Generalities
According to the conventions defined above, there are three options for transmitting results:
- On-line Transmission: results are sent test by test as soon as they are obtained (see description
below chapter 8.4.3).
- Uploading File by File with the F11 key from different FILE PROCESSING screens, access by the
CONFIRMATION/PRINTOUT screen or by the TEST PANEL screen (see chapter 8.3.3);
8-14
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
System
Tests
Products
Stop Sample Pipetting NO
On-Line Transmission NO
On-Line Printout NO
with the cursor on System.
ACTIONS RESULTS
" Press the Esc key (Options) The following window is displayed:
8-15
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
Legend:
4 Available functions:
F10 Runs the Patient file transmission (or starts it again if it was stopped)
8-16
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
" To include patient files which have been The identification of the patient files being
already transmitted: transmitted are displayed near Processed
press the Y key, then confirm with the File (see the FILE TRANSMISSION window)
F10 key; otherwise, press the F10 key and the number of files remaining to
transmit near Rest (see description above
chapter 8.4.4.2).
The message “TRANSMITTING” then
“TRANSMISSION COMPLETE” are displayed
above Processed File (see description above
chapter 8.4.4.2).
" When the transmission is done: Returns to the list of selected files.
press the Esc key The letter T is displayed in front of the
identification of each file actually
transmitted.
8.5.1 Generalities
There are three options for printing out patient files:
On-Line Printout: the patient files are printed file by file as soon as they are completed (see
description, chapter 8.5.4)
Printout of an individual file by means of the F6 key:
- Directly from the CONFIRMATION/PRINTOUT screen for the file corresponding to the position of
the cursor (see chapter 8.2.2)
- Or from the FILE PROCESSING screens accessible either from the CONFIRMATION/PRINTOUT
screen or from the TEST PANEL screen (see chapter 8.3.3)
Printout of a selection of files (see chapter 8.5.5)
For all these cases, the printout parameters are defined from the PRINTOUT window accessible as
soon as the Print option is selected (see chapters 8.5.2 to 8.5.4).
To check the printer parameters or the printer connection, see chapter 4.
In addition, patient file printout can also be custom-built (see description chapter 8.8.).
8-17
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" Press the Esc key (Options) The following window is displayed:
" Press the P key (Print) The PRINTOUT window is displayed, see
" Or move the cursor to Print by pressing description below.
the L key, then confirm with the key
Legend:
1 Prints only the unprinted files from a selection of Patient files (see procedure in chapter 8.5.5)
5 Printout parameters
Three parameters allow definition of the printout (area 5 of the PRINTOUT window):
8-18
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
- all files: the patient files are printed one after the other in a table, the lab name is printed on top
of each page.
- single file: only one file per printed page, the lab name is printed for each file.
Quality: standard or presentation
- standard: allows a rapid printout, unable to print certain special character forms
- presentation: allows a better printout quality but is slower
Page Length: defined in line quantity without taking into account the real length of paper
To avoid frequent cuts in files, apply the following principle :
- 1 inch = 6 lines
- 11" paper = 66 lines in a real page
- To print 2 patient files per page = set Page Length to 22 lines
- To print 3 patient files per page = set Page Length to 33 lines
If a printed file is divided:
- The message “Continued: next page” is printed on the lower right corner of the first part.
- The complete patient file identification (including the complementary information) is recalled on
top of the second part.
- The message “CONTINUED” is printed right under the patient file identification for the second
part.
NOTE The lab name is defined within the Global Options menu (see chapter 10.2.2.2).
2PRINTOUT window displayed with cursor positioned on Quit (see chapter 8.5.2).
²
ACTIONS RESULTS
" Press the P key (Printer Setup) The following window is displayed
" Or move the cursor to Printer Setup by
pressing the L key twice, then confirm with
the key
" To modify a given parameter: move the For “Mode” parameter, definition changes
cursor on it with the L key then, confirm from all files to single file or inversely.
with the key.
For “Quality” parameter, definition changes
from standard to presentation or
inversely.
For “Page Length” parameter, the cursor is
positioned on the number of lines (see
description of the printout parameters in
chapter 8.5.4).
8-19
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" Type with the keyboard the number of The parameters are displayed.
desired lines then confirm with the key
" When all the printout parameters are set, Back on PRINTOUT window.
press the Esc key
ACTIONS RESULTS
" Press the Esc key (menu) " The Main Menu is displayed:
System
Tests
Products
Stop Sample Pipetting NO
On-Line Transmission NO
On-Line Printout NO
with the cursor on System.
" Move the cursor to On-Line Printout No The option On-Line Printout No
with the L key, then confirm with the changes to On-Line Printout Yes.
key
8-20
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
If the On-Line Printout has been chosen, this option is available only if the STA Compact® is in stand-
by mode (no measurement in progress) and if all the printouts previously scheduled have been done.
If the On-Line Printout has not been chosen, this option is always available.
With this procedure, all patient files of a given selection are printed; it is first necessary to proceed
to file patient selection (see chapter 8.2.3).
2PRINTOUT window displayed (see procedure chapter 8.5.2) with the cursor positioned on Quit.
ACTIONS RESULTS
" Change the printout parameters if A window is displayed in the upper left
necessary (see chapter 8.5.3) corner of the screen. This window indicates
the identification of the file being processed
" Depending on the required patient files:
as well as the number of files remaining to
- All the files not yet printed, print.
press the Lkey Unprinted Files) then
At the same time, the file printout starts.
confirm with the key
- All the selected files:
press the L key to All Files, then confirm
with the key
" To stop the current printout, press the The files already processed are being
Esc key printed.
Then, the printout stops.
" When all files are printed Automatic return to the list of selected files.
The letter P is printed in front of the
identification of each file actually printed
8-21
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
Legend:
1 Name of the lab, as defined within the Global Options menu (see chapter 10.2.2.2)
3 Patient file identification including, if necessary, the complementary information (name, first
name, etc.); on the second line, the standard notes if necessary
8-22
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
4 Complete name of the test; it is preceded by an asterisk (*) if the test is concerned by one or
several alarms
9 Result status:
- Blocked
- Waiting
- In process
- To confirm
- Error (corresponds to the following results:
technical error, Value > Max., Value < Min., Dev. > tol, linearity, plasma quantity not sufficient)
- Confirm, the result is validated
10 Usual values for each test as defined within the Test Setup menu (see chapter 5.1.3.2).
12 Page Number - In “Single File” mode, the page number starts at 1 with each File printout
This option is only accessible if the STA Compact® is in stand-by mode (no measurement in
progress) and if the printouts previously scheduled have been done.
When the Working File is full (more than 600 records), the operator is requested to delete the
oldest file whose sample tube has been removed.
NOTE The option Delete Files from the Working File is also available when switching on the STA
Compact® (see procedure chapter 3.1.4). But in this case, no selection is possible.
8-23
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
Confirmation/Printout/Transmission
Delete Patient Files
Customer Printout
with the cursor positioned on
Confirmation/Printout/Transmission.
" Press the D key (Delete Patient Files) The FILE DELETION window is displayed (see
description below chapter 8.6.2).
" Or move the cursor to Delete Patient
Files with the L key then confirm with the
key
8-24
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
Legend:
3 Prefix identification
5 Available functions:
F10 Displays the list of selected files then deletes the selection
6 List of selectable files (selection procedure in progress) or list of selected files (selection
procedure done)
8.6.3 Procedure of Deletion of the Patient Files from the Working File
ACTIONS RESULTS
" Press the F3 key The message “Please select the First File” is
displayed.
All the available patient files of the Working
File are displayed (see above FILE DELETION
window description chapter 8.6.2).
" Use the moving cursor keys to move it to The message “Please select the Last File” is
the required file, then confirm with the displayed.
key Identification of the first file is recalled.
8-25
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" Use the moving cursor keys to move it to The FILE DELETION window is displayed
the required file, then confirm with the with the cursor positioned near Prefix and
key the identifications of the first selected file
and of the last selected file displayed (see
description of the FILE DELETION window
chapter 8.6.2).
" Type the prefix identification then confirm The prefix identification is displayed and the
with the key cursor is positioned on area 4 (see
" Or type the first number(s) or letter(s) of description of the FILE DELETION window
the required patient file identifications, then chapter 8.6.2).
confirm with the key
" Or leave this area empty (no prefix
identification) and confirm with the key
" If a type of file is not desired, use the L key The colored square disappears.
to move the cursor to it, then confirm with
the key to remove the colored square
" When all the parameters of the desired The following window appears furtively:
selection are set, press the F10 key
Research in progress
Please, wait.
The list of selected files is displayed (see
description of the FILE DELETION window
chapter 8.6.2).
The message “Please, type YES to confirm”
is displayed.
" Type the letters: YES and confirm the file The patient files are deleted from zone 6
deletion with the key (see description of the FILE DELETION window
chapter 8.6.2).
ACTIONS RESULTS
" Press the F4 key All the available patient files of the Working
File are displayed(see above FILE DELETION
window description, chapter 8.6.2).
The message “Please, type YES to confirm”
is displayed.
8-26
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
" Type the letters: YES and confirm the All files of the Working File are deleted from
deletion with the key zone 6 (see description of the FILE DELETION
window, chapter 8.6.2).
8-27
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" Press the / key (Options) of the keypad The following window is displayed:
(right side of the keyboard)
Help
File Modification
Quit
with the cursor positioned on Quit.
" Press the F key (file modification) The FILE MODIFICATION screen is displayed
" Or move the cursor to File Modification (see description below chapter 8.7.2).
with the L key, then confirm with the key
Legend:
8-28
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
1 F1 Change: opens the following window which allows to choose the type of file
modification to apply:
Block
Unblock
Rerun
Delete
Block: Blocks tests in pending status:
Unblock: Unblocks blocked tests;
Rerun: Reruns tests for the selected files;
Delete: Deletes tests for the selected files.
Note: There is only one type of modification possible at one time.
2 Identification of the type of modification chosen; the first letter of the name of the modification
is displayed in yellow
3 F2 Selects the patient files on which the file modification will be applied (see procedure below
chapter 8.7.4). This function is not available when the chosen file modifications are Block or
Unblock.
6 Available functions:
Applies the chosen file modification (or cancels it) for the test corresponding
to the position of the cursor
8 List of the abbreviations of the available tests. The letter B (Blocked) is displayed in grey at the
left of the tests currently blocked. When chosing the application of the file modification, a letter
is displayed in yellow at the left of the selected tests:
B for block
U for unblock
R for rerun
D for delete
Note: The “BLOCKED SAMPLE PIPETTING” message is displayed in white on a red background at the
bottom of the screen when the sample pipetting is blocked whether:
- Voluntarily by the operator (Stop Sample Pipetting function, see description in chapter 10.1.5.1)
- Or automatically by the STA Compact® after the consistency check; if the required assays cannot
be done
8-29
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
The Block or Unblock options concern the tests of any patient files present in the Working File and
the quality control automatic (periodic) running function. It does not concern the calibrations nor
the calibration controls. Consequently, the file selection is not available.
Block is only possible for pending tests and Unblock for blocked tests.
Indeed, the operator can force any quality control running for a blocked test by using the Run
function of the Quality Control menu (see chapter 9.3).
After unblocking tests, while returning to the TEST PANEL, the STA Compact® automatically
displays the TEST STATUS screen (see description chapter 10.1.3).
ACTIONS RESULTS
Block
Unblock
Rerun
Delete
with the cursor positioned on Block
" To block, press the key Block is displayed in the ACTION area (see
" To unblock, press the U key (Unblock) or description of the FILE MODIFICATION screen
move the cursor to Unblock with the chapter 8.7.2).
L key, then confirm with the key Unblock is displayed in the ACTION area
(see description of the FILE MODIFICATION
screen chapter 8.7.2).
The cursor is positioned on the first test (see
description of the FILE MODIFICATION screen
chapter 8.7.2).
" Use the moving cursor keys to move it to A letter is displayed in yellow at the left of
each related test then select each test with the selected test; B for Block or U for
the key Unblock.
8-30
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
" When all parameters are set, press the The file modification is applied.
F10 key For the Block function, the TEST PANEL is
displayed:
The message Block appears in red replacing
requests for the tests concerned by the
Block option.
For the Unblock function, the TEST STATUS
screen is displayed (see description
chapter 10.1.3).
NOTE For duplicate determination, the tests are rerun in duplicate. The current results have to be
printed out because they will disappear.
For single determination, the previous value is stored and a new single determination is run. The
PRECISION parameter defined on the first page of the TEST SETUP screens (see description
chapter 5.1.1.7) is used to compare both results. If the difference is greater than the defined one,
the STA Compact® automatically reruns the test a third time, compares the three results and keeps
the best pair of results.
After chosing to rerun tests, while returning to the TEST PANEL, the STA Compact® automatically
displays the TEST STATUS screen (see description chapter 10.1.3).
ACTIONS RESULTS
Block
Unblock
R erun
Delete
with the cursor positioned on Block.
8-31
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" To rerun tests, press the R key (Rerun) or Rerun is displayed in the ACTION area (see
move the cursor to Rerun with the L key, description of the FILE MODIFICATION screen
then confirm with the key chapter 8.7.2).
" To delete tests, press the D key (Delete) Delete is displayed in the ACTION area (see
or move the cursor to Delete with the L description of the FILE MODIFICATION screen
key, then confirm with the key chapter 8.7.2).
" Use the moving cursor keys to move it to The message “Please select the Last File” is
the required file then confirm with the displayed.
key
The identification of the first selected file is
recalled.
Use the moving cursor keys to move it to The FILE MODIFICATION screen is displayed
the required file then confirm with the with the cursor positioned on the first test of
key the list of available tests and the
identifications of the first selected file and of
the last selected file are displayed (see
description of the FILE MODIFICATION screen
chapter 8.7.2).
" Use the moving cursor keys to move it to A letter is displayed in yellow at the left of
each related test, then confirm with the each selected test: R for Rerun, D for
key Delete.
8.8.1 Generalities
This menu gives the opportunity to print the Patient Files according to each laboratory specific needs.
All the confirmed Patient Files present in the Working File are available for this printout.
NOTE A confirmed or validated Patient File is a Patient File for which all tests are done and confirmed
8-32
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
The location and the type of information related to the Patient Files (Patient ID, test name, results)
are fixed ; only the headings or the additional remarks are defined by each laboratory.
Only one file is printed per page.
This access is only possible if STA Compact® is in stand-by (no measurement in progress).
ACTIONS RESULTS
Confirmation/Printout/Transmission
Delete Patient Files
Customer Printout
with the cursor on Confirmation/
Printout/Transmission.
" Move the cursor to Customer Printout The CUSTOMER PRINTOUT OF VALIDATED
with the K key, then confirm with the FILES screen is displayed (see description
key chapter 8.8.3).
8-33
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
Legend:
1 Available functions:
F6 Prints out the confirmed Patient File on which the cursor is positioned
2 Complementary information (name, first name ... see description chapter 10.2.3.3) for the file
on which the cursor is positioned
3 Selected File List. The letter (P) shows that the file has already been printed by means of the
CUSTOMER PRINTOUT OF VALIDATED FILES screen
4 Warning Message
8-34
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
To determine the length of the Patient File, the Customer Printout menu uses the Page Length
parameter defined within the Files menu (PRINTOUT window, CONFIRMATION/PRINTOUT screen)
on condition that this parameter is at least greater than 54 (see chapter 8.5).
Legend:
1 Data defined by the laboratory (Titles, data, name, notes ...); for these data definitions, see
chapters 8.8.4.3 and 8.8.4.4)
8-35
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
2 Patient File Identification (16 characters max.), Field No.1 as defined within the Global Options
menu, see chapter 10.2.3.3
3 Contents of fields No.2 and No.3 as defined within the Global Options menu, see chapter
10.2.3.3, (16 characters and 12 characters max.).
The field No.2 content is printed in upper case character; the first character of the field No.3
content is printed in upper case character, the remaining in lower case character
4 Contents of field No.4 as defined within the Global Options menu, see chapter 10.2.3.3
(6 characters max.)
5 Content of field No.5 as defined within the Global Options menu, see chapter 10.2.3.3
(4 characters max.)
6 Printout Date; the used format is defined within the Global Options menu, see chapter 10.2.3.4
7 Test complete name as defined within the Test Setup menu (see chapter 5.1.1.1)
The result table can contain an entire file, i.e.: 12 tests
Note: Results for secondary units can be printed only if YES has been selected for the printout
within the TEST SETUP screens, see chapter 5.1.3.1.
Note: The fields 8, 9, 10 and 11 are replaced by the message “Out of Range” if the result
has been confirmed even though its status was “Time > Max Time”, see chapter 8.3.3.
12 Usual values for each test as defined with the TEST SETUP screens, see chapter 5.1.3.1
13 Standard note as defined within the FILE PROCESSING screens, see chapters 8.3.6 to 8.3.8
8-36
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
8-37
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
Legend:
1 Available functions:
LK Moves the cursor to the following line (Lkey) or to the preceding line (K key)
Note: General data names No.1, No.2 and No.5 are mandatory acquisition fields.
8-38
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
Legend:
1 Available functions:
LK Moves the cursor to the following line (Lkey) or to the preceding line (K key)
3 Constant remark:
2 lines of 60 characters printed on one line in condensed format
8-39
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
" Move the cursor to the requested The cursor is positioned on the requested
parameter with the L key parameter.
" Enter the new information on the The new information is displayed.
alphanumeric keyboard, then confirm with
The cursor moves to the next data.
the key
" When all information is entered, press the Customer printout definition parameters
F10 key are saved.
Back to the CUSTOMER PRINTOUT OF
VALIDATED FILES screen.
ACTIONS RESULTS
8-40
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
ACTIONS RESULTS
" Press the C key (Change File Selection) The FILE SELECTION window is displayed
" Or move the cursor to change File (see description, chapter 8.8.5.2).
Selection with the L key, then confirm
with the key
Legend:
3 Available functions:
F1 “Beginning of List” displays the first file of the confirmed files listing in chronological
order in the Working File
F2 “End of List” displays the last file of the confirmed files listing in chronological order in
the Working File
F3 “By cursor” selects the files with the cursor in the available files
8-41
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
2FILE SELECTION window displayed from CUSTOMER PRINTOUT OF VALIDATED FILES screen, (see
description on chapter 8.8.5.1).
ACTIONS RESULTS
" Type the identification of the first file of the The identification of the first file of the list is
desired selection then confirm with the displayed in area 1 or “Start ” is displayed in
key area 1 and the cursor is positioned in area 2,
" Or press the F1 key to display “Start” see description of the FILE SELECTION window
(corresponding to the first confirmed file (the chapter 8.8.5.2 .
oldest) of the Working File)
" Or press the F3 key to select the files with
the cursor. In this case, use the moving
cursor keys to move it to the required file,
then confirm with the key
" Type the identification of the last file of the The identification of the last file of the list is
desired selection then confirm with the displayed in area 2 or “End of File” is
key displayed in area 2 and the cursor is
" Or press the F2 key to display positioned in area 4, see description of the
“End of File”(corresponding to the last FILE SELECTION window chapter 8.8.5.2.
confirmed file (the latest) of the Working File)
" To select the last file with the cursor, use
the moving cursor keys to move it to the
required file, then confirm with the key.
" To filter: type the prefix identification, then Filtering prefix identification displayed in the
confirm with the key area 4 (if defined), see description of the
FILE SELECTION window chapter 8.8.5.2.
" Or type the first number(s) or letter(s) of
the patient files identification, then confirm
with the key
" Or leave the area empty (no prefix
definition) and confirm with the key
" When all the parameters of the desired The new selection is displayed on the
selection are set, press the F10 key CUSTOMER PRINTOUT OF VALIDATED FILES
screen, see description chapter 8.8.3.
8-42
0931033L - July 2009
REFERENCE MANUAL 8
Patient File Management
2CUSTOMER PRINTOUT OF VALIDATED FILES screen displayed, see access chapter 8.8.2.
ACTIONS RESULTS
" Depending on the required Patient files: A window is displayed in the upper left
corner of the screen. This window indicates
- All the files not yet printed,
the identification of the file being processed
move the cursor to Print Unprinted
as well as the number of files remaining to
Files with the Lkey, then confirm with
print.
the key
At the same time, the file printout starts.
- All the selected files:
move the cursor to Print all Files withthe
K key, then confirm with the key
" To stop the current printout, press the The files already processed are being
Esc key printed.
Then, the printout stops.
" When all files are printed Automatic return to the list of selected files.
The letter P is printed in front of the
identification of each file actually printed.
8-43
0931033L - July 2009
8 REFERENCE MANUAL
Patient File Management
ACTIONS RESULTS
Downloading
File: xxxx
ESC = Abort
xxxx = identification of the Patient File being
downloaded.
The Patient File working lists are received.
8-44
0931033L - July 2009
REFERENCE MANUAL 9
Table of contents
9 Quality Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
9.1 Generalities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
9-i
0931033L - July 2009
9 REFERENCE MANUAL
Table of contents
9-ii
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
9 Quality Control
9.1 Generalities
For each test, at least one quality control has to be run daily if at least one patient results is requested
for this test.
Up to three quality controls can be defined for each test within the TEST SETUP screens (see
chapter 5.1.3.4).
Quality controls are loaded in the product drawer and are automatically run by STA Compact® (see
details, chapter 9.4).
In the Quality Control sub-menu, the quality control results are treated on 3 zones:
- CONTROLS FOR TODAY zone
- CUM DAILY zone
- CUM MONTHLY zone
(See description of these zones in chapter 9.3.2 and 9.3.3).
In these zones, up to 2 lot numbers can be stored.
When controls from a new lot are loaded, values of the oldest lot are automatically deleted, the
preceding lot changes into the old lot and the new lot into the current lot.
The treatment of the quality controls is based on the Levey-Jennings statistical model:
i=n
¦ xi
Mean : m = i-------------
=1 - xi = result in primary unit
n
9-1
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
Zone n
CUM DAILY 1 to 31
CUM MONTHLY 1 to 12
ACTIONS RESULTATS
Calibration
Quallity Control
with cursor on Calibration.
" Press the L key to Quality Control, then The QUALITY CONTROL screen is displayed
confirm with the key (see description chapter 9.3.1).
The cursor is positioned on the first test of
the list.
" Press the L or J keys to move the cursor Results of quality control for the first control
to the selected test, then confirm with the displayed in graphic form (see description
key chapter 9.3.2 ).
" Press the F3 key Results of quality control for the next control
displayed in graphic form (see description
Note: If only one quality control level has
chapter 9.3.2).
been defined; with F3, the same level is
displayed. For 2 or 3 levels, with F3, the
different levels are looped.
9-2
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
ACTIONS RESULTATS
" Press the F1 key Results of quality control for the control
selected displayed in tabular form (see
description chapter 9.3.3).
" From the screen of quality control results in Return to QUALITY CONTROL screen (test
graphic form, press the Esc key, then list).
confirm with the key
Legend:
1 Available keys:
9-3
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
3 List of test abbrevations. A colored square displayed at the left of the abbreviation indicates the
quality control current status (see point 5 below)
White: Validated
Blue: To be validated
Yellow: In progress
Red: Not done
Legend:
1 Available keys:
9-4
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
F2/F3 Accesses the other levels; the results are always displayed in graphic form.
Esc Options; this key opens the following window:
Print
Run
Cancel
Change Range
Delete All Results
Return to Result List
When an option is not available, its name is displayed in grey instead of blue
2 Level indication: 1, 2 or 3
The points do not necessarily correspond to consecutive days, they come from several months.
The day corresponding to each mean is displayed at the bottom of the CUM DAILY zone.
Adding a new point is made from the right side of the existing points; shifting those points to the
left when there are more than 31 values. In this case, visualization of the first point is lost.
The means of the daily controls are displayed according to the following color codes:
+++(green): Daily control means obtained during preceding months which are within
the range of acceptable results
+++(red): Daily control means obtained during preceding months which are outside
of the range of acceptable results
9-5
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
xxx (yellow): Daily control means for the current month which are within the range of
acceptable results
xxx (red): Daily control means for the current month which are outside of the range
of acceptable results
When the range of acceptable results is changed, the colors of the various points are reviewed
according to the range of acceptable results (see description chapter 9.4.3).
10 Green vertical bar indicating a lot change (up to 2 lots are shown)
11 Range of acceptable results for the quality control. It is represented by white dotted lines which
are displayed in the 3 zones: CONTROLS FOR TODAY/CUM DAILY/CUM MONTHLY.
12 Mean of range of acceptable results for the quality control. It is represented by a white solid line
which is displayed in the 3 zones: CONTROLS FOR TODAY/CUM DAILY/CUM MONTHLY.
13 Mean ± 2 K (V = standard deviation) for quality control results of the CUM DAILY zone. It is
represented by yellow dotted lines.
14 Mean of quality control results of the CUM DAILY zone, it is represented by a solid yellow line.
15 Glossary explaining the meanings of the color codes displayed on the screen.
9-6
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
Legend:
1 Available keys:
Esc Back to Graph: returns to graphic form for quality control results of the selected level
F4 Suppress current Controls: suppresses one or several results from the CURRENT
CONTROLS zone
F6 Print: prints results as tabular form
F11 Transmit Current Controls: Transmits all today’s quality control results to the Host
computer
2 Level indication: 1, 2 or 3
9-7
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
10 Range of acceptable results for the current lot of the quality control in the primary unit
11 Mean (Mean), Standard Deviation (K) and coefficient of variation (CV) for the results of the
today quality controls
13 Mean (Mean), Standard Deviation (K) and coefficient of variation (CV) for the daily means from
the old lot (displayed in dark blue).
14 Mean (Mean), Standard Deviation (K) and coefficient of variation (CV) for the daily means from
the current lot (displayed in light blue).
9-8
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
For the month corresponding to the lot change, the displayed mean corresponds to that with the
more representative points.
Running controls on request is possible only if the sample pipetting has not been blocked either from
the TEST STATUS screen ( see description chapter 10.1.3) or by the Stop Sample pipetting
function of the Status sub-menu (see description chapter 10.1.5.1).
To block tests by means of the FILE MODIFICATION screen blocks the quality control automatic
(periodic) running function but not a running decided by the operator (see description chapter 8.7.3).
Quality controls are run in single or in duplicate depending on the determination type chosen for the
sample (see description chapter 5.1.1.7 ).
9-9
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
As soon as quality control results are obtained, they are compared to the range of acceptable results,
if they are outside of the range, the STA Compact®automatically reruns the controls according to
the principle below:
Automatic Rerun
First run Result(s)
by STA Compact®
Control run in single Outside of the defined range Control rerun in duplicate
If the quality control is confirmed as outside of the range of acceptable results, all the patient results
for the related test will be given with an alarm code: «Quality Control: out of range or not done»;
they will be displayed in light red on the TEST PANEL screen.
If the operator decides to accept quality control results declared as out of range, all patient results
for the related test will be given with an alarm code: «Quality Control: overridden».
If quality control has been run or rerun, when returning to the TEST PANEL screen, STA Compact®
performs a consistency check and displays the TEST STATUS screen (see description, chapter 10.1.3).
If upon running quality controls (automatically or on request), those are not present in the drawer
(not present or present but with volume or stability not usable) and if the time elapsed since the last
quality control run is less than 24 hours, then STA Compact® does not perform the control. It
proceeds with the assays for the related test and all patient results for the related test are given with
the alarm code: «Quality Control: out of range or not done». In the same case, with an elapsed
time greater than 24 hours, the sample pipetting for the related test is blocked.
If the control lot number has changed or if the quality control results have been reset, the sample
pipetting for the related test will start again only when the products necessary to perform the first
controls will have been loaded. In the same way, the sample pipetting for the related test will be
blocked as long as the range values have not been defined.
Reminder When the same controls are used (same identification Numbers) as calibration controls and as
quality controls, the results of the calibration controls are added into the results for quality controls
and the time at which the calibration controls have been run resets the clock for the quality
controls.
If the On-line Transmission option is enabled (see description chapter 10.1.5.2), then all quality
control results will be sent with their error codes as soon as the measurements are obtained.
For description of the error and alarm codes, see chapter 8.3.2. Each quality control result will be
sent with the run date and time. If the On-Line Transmission is not enabled , all today’s confirmed
quality control results can be sent from the quality control result table, see chapter 9.6.5.
9-10
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
Reminder For Diagnostica Stago controls, the values for the range of acceptable results are properly entered
for the related tests, on condition that for those, the primary unit chosen is identical to that defined
on the package inserts of each control box (except Fibrinogen, for g/l primary unit, values are
properly entered even if on the package insert, they are given in mg/dl).
For controls which are not from Diagnostica Stago, the range of acceptable results has to be
absolutely defined at each lot change so that controls can be run.
For controls from Diagnostica Stago, the range of acceptable results can only be reduced, it cannot
be in any case increased.
2QUALITY CONTROL RESULTS screen (graphic form) displayed for the selected level (see access
chapter 9.2).
9-11
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
ACTIONS RESULTATS
Print
Run
Cancel
Change Range
Delete All Results
Return to Result List
with the cursor positioned on Return to
Return List
" Move the cursor to Change Range with The following window is displayed:
the K key, then confirm with the key.
Range Modification
New values 1 xxx - xxx % 3
(Original Values 2 xxx - xxx)
F10 Save Esc Abort
with cursor positioned on the Entry zone for
new values.
1 Entry zone for the new range values in
primary unit
" Type the new range values, confirm with The Range Modification window
the key then save the new values with disapears.
the F10 key
The following window is displayed:
Please enter
your access code:
Validate Esc: Quit
" Type your access code, then confirm with The screen for quality control results is
the key displayed with the new range of acceptable
results.
In the 2 zones: CONTROLS FOR TODAY and
CUM DAILY, all points outside of the new
range are displayed in red.
9-12
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
Running is impossible:
- if control is already running or if its status is «to be validated»;
- if the calibration for the selected test is not validated;
- if products necessary to perform the control are not present;
- if the values for the range of acceptable results have not been defined (see chapter 9.4.3).
2Screen of quality control results (graphic form) displayed for the selected level (see access
chapter 9.2).
ACTIONS RESULTATS
Print
Run
Cancel
Change Range
Delete All Results
Return to Result List
with the cursor positioned on Return to
Result List
" Move the cursor to Run with the L key, The following window is displayed
then confirm with the key
Please enter
your access code:
Validate Esc: Quit
" Type your access code, then confirm with STA Compact® runs the quality control level
the key as long as the consistency check does not
lead to a blocking of the sample pipetting.
" Press the Esc key then confirm with the Return to QUALITY CONTROL screen (test
key list).
9-13
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
Running is impossible:
- for tests with one control already running or already with status «to be validated»;
- if the calibration for the selected test is not validated;
- if products necessary to run the quality controls are not present, (see chapter 9.4.1).
- for test with undefined range of acceptable results of quality controls (see chapter 9.4.3).
2QUALITY CONTROL screen (test list) displayed (see access chapter 9.2).
ACTIONS RESULTATS
" To select a test for quality controls to be run If running is possible (all the necessary
on, move the cursor on its abbreviation, products are loaded and the control ranges
then press the F1 key have been defined), the letter S is displayed
at the left of each test abbreviation for which
Note: To cancel a selection, press the F1
controls have been selected.
key again.
If running is impossible, STA Compact®
beeps and the letter S is not displayed.
" When all the requested controls are The following window is displayed:
selected, press the F10 key
Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with The letters S at the left of the abbreviations
the key disappear.
Controls are run by STA Compact® as long
as the consistency check does not lead to a
blocking of the sample pipetting.
The colored squares at the left of the
abbreviations for the selected tests are
displayed in yellow.
" Press the Esc key The TEST PANEL screen is displayed (see
description chapter 10.1.3).
9-14
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
9.5.1 Alarm Code and Alarm Message Displayed for out of Range Quality Controls
When a quality control result is outside of the range of acceptable results (out of range), the operator
is informed by the following message:
MEASURING MANAGEMENT
Error 05.06.09
QUALITY CONTROL
Control out of limits
for the test xxx
Esc = continue
xxx = Test Abbreviation
For the related test, all patient results are given with an alarm code: «Quality Control: out of range
or not done» and on the TEST PANEL screen, all results are displayed in light red.
Legend:
1 Test abbreviation
3 Results obtained for the quality control (in the primary unit)
4 Recall of the range of acceptable results for the quality control (in primary unit)
9-15
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
7 Available functions:
Accept: accepts an out of range quality control (see description chapter 9.5.3)
Rerun: reruns the quality control (see description chapter 9.5.4)
Postpone Decision: returns to QUALITY CONTROL screen (test list)
This procedure is only possible if the operator knows the access code which allows to accept out
of range controls (see chapter 1.5).
Reminder If the operator accepts quality control results out of range, all patient results for the related test will
be given with the alarm code: «Quality Control: overridden».
In the CONTROL OUT OF RANGE window, out of range results are displayed in red and the range of
acceptable results is displayed (see description chapter 9.5.2).
2Screen of quality control results displayed with with the CONTROL OUT OF RANGE window,
cursor positioned on Postpone Decision.
ACTIONS RESULTATS
" Press the A key (Accept) The CONTROL OUT OF RANGE window
disappears.
" Or move the cursor to Accept with the
L key then confirm with the key The following window is displayed:
Please enter
your access code:
Validate Esc Quit
" Type your access code, then confirm with Depending on the value of the out of range
the key result, a red o or a red '(out of scale) is
displayed in the CONTROLS FOR TODAY
zone.
" Press the Esc key, then the key Return to QUALITY CONTROL screen (test
list).
9-16
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
Reminder When the STA Compact® declares a quality control as out of range, the control has already been
run 3 times.
If a control result is out of the range of acceptable results, for a single determination, control is rerun
in duplicate and for a duplicate determination, it is rerun in single (see description chapter 9.4.1).
In the CONTROL OUT OF RANGE window, out of range results are displayed in red and the range of
acceptable results is displayed (see description chapter 9.5.2).
2Screen of quality control results displayed with the CONTROL OUT OF RANGE WINDOW, cursor
positioned on Postpone Decision.
ACTIONS RESULTATS
" Type your access code, then confirm with STA Compact®reruns the quality control so
the key long as the consistency check does not lead
to a blocking of the sample pipetting.
Return to the quality control results of the
selected control level.
This function is not available if the control status is «confirmed», «to be validated» or «not done». It
is only possible for running controls (yellow square at the left of the test abbreviation).
2Screen of quality control results displayed (graphic form) for a selected level (see access
chapter 9.2).
9-17
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
ACTIONS RESULTATS
Print
Run
Cancel
Change Range
Delete All Results
Return to Result List
with the cursor positioned on Return to
Result List.
" Move the cursor to Cancel with K key, The quality control running for the selected
then confirm with the key level is cancelled.
" Press the Esc key Return to QUALITY CONTROL screen (test
list).
The colored square at the left of the
abbreviation for the selected test is
displayed in red (not done).
This procedure is possible only if the operator knows the access code to critical functions, (see
chapter 1.5). It concerns only quality control results of the controls for today zone (see description
of this zone chapter 9.3.2, point 9).
2Screen of quality control results displayed (graphic form) for a selected level (see access
chapter 9.2).
ACTIONS RESULTATS
" Press the F1 key The quality control results of the selected
level are displayed as tabular form.
Please enter
your access code:
Validate Esc Quit
9-18
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
ACTIONS RESULTATS
" Type your access code, then confirm with The following window is displayed:
the key
Current Controls Suppression
DEL Suppress
INS Restore
F10 Save
Esc Abandon
The cursor is positioned on the first result of
the CURRENT CONTROLS zone
" For each result to be deleted, move the Each result which has to be deleted is
cursor to it with the L key, then press the displayed in grey.
Delete key
" Press the Esc key Return to the quality control results as
graphic form.
This operation deletes all the quality control results for the selected level.
For this operation, the operator must know the access code to critical functions (see chapter 1.5).
This function is not available if the control status is «in progress» or «to be validated».
2Quality control results screen displayed (graphic form) for a selected level (see access
chapter 9.2).
9-19
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
ACTIONS RESULTATS
Print
Run
Cancel
Change Range
Delete All Results
Return to Result List
with the cursor positioned on Return to
Result List
" Press the D key (Delete All Results) The following window is displayed:
" Or move the cursor to Delete All Results
with the K key, then confirm with the Please enter
key your access code:
Validate Esc Quit
" Type your access code then confirm with The following window is displayed:
the key
DELETE ALL CONTROL RESULTS
accept (YES/NO) xxxx ?
" Type YES then confirm with the key All quality control results of the selected
level are detected.
" Press the Esc key, then confirm with the Return to QUALITY CONTROL screen (test
key list).
This operation is possible only if the STA Compact® is in stand-by mode (no measurement in
progress) and if all printouts previously generated have been performed.
2Screen of quality control results displayed (graphic form) for a selected level (see access
chapter 9.2).
9-20
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
ACTIONS RESULTATS
Print
Run
Cancel
Change Range
Delete All Results
Return to Result List
with the cursor positioned on Return to
Result List
This operation is possible only if the STA Compact® is in stand-by mode (no measurement in
progress) and if all printouts previously generated have been performed.
2Screen of quality control results displayed (graphic form) for a selected level (see access,
chapter 9.2).
9-21
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
ACTIONS RESULTATS
" Press the F1 key The quality control results of the selected
level are displayed as tabular form.
PRINTOUT
EXECUTE
Printer Test
Form Feed
Quit
with cursor positioned on EXECUTE
" Press the Esc key Return to the screen of quality control
results as graphic form.
2Screen of quality control results displayed (graphic form) for a selected level (see access
chapter 9.2).
ACTIONS RESULTATS
" Press the F1 key The quality control results of the selected
level are displayed as tabular form.
9-22
0931033L - July 2009
REFERENCE MANUAL 9
Quality Control
ACTIONS RESULTATS
" Press the F11 key The two following windows are displayed:
TRANSMITTING
then,
TRANSMISSION COMPLETED
Esc= Clear this message
All today’s quality control results for the
selected level are transmitted to the Host
computer with their rspective running date
and time.
" Press the Esc key Return to the table of quality control results
for the selected level.
" Press the Esc key Return to the screen of quality control
results as graphic form.
9-23
0931033L - July 2009
9 REFERENCE MANUAL
Quality Control
9-24
0931033L - July 2009
REFERENCE MANUAL 10
Table of contents
10-i
0931033L - July 2009
10 REFERENCE MANUAL
Table of contents
10-ii
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
System
Tests
Products
Stop Sample Pipetting YES
On-Line Transmission YES
On-Line Printout YES
with the cursor on System.
" Then press one of the following keys for The SYSTEM STATUS screen is displayed.
the screen required:
The TEST STATUS screen is displayed.
" (System)
The PRODUCT STATUS screen is displayed.
" T (Test)
" P (Product)
10-1
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
To exit from the STATUS screens, press the Esc key and the TEST PANEL screen will be displayed
again.
Legend:
1 Esc Options
The Esc key opens the following window:
Change Date
Change Time
Level Detection Management
Quit
10-2
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
Needle No.3, measurement block: the temperature must be within 36.5°C to 37.5°C.
5 Level detection option for each needle: No.1 (samples), No.2 and No.3.
The operator can change the options of the level detection from managed to not managed (and
vice versa), see chapter 10.1.2.4.
6 Number of functioning hours of STA Compact® since the last service call.
7 Number of functioning hours of STA Compact® since the STA Compact® installation.
When the availability reaches 0, a message warns the operator. If the maintenance is not done, the
number becomes negative and is displayed in red.
When the value of the field Syringe is 0:
- Each result is given with an alarm indicating that the maintenance has not been done.
- The Syringe Teflon tip must be replaced.
10-3
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
When the value of the field Light source (not available on STA Compact®CT) is 0:
- The results have no specific alarm because the light intensity level is always controlled.
- It is advised to change the Halogen lamp as a preventive measure.
11 No piercing kit. If a cap piercing kit is installed, please refer to the corresponding notice.
ACTIONS RESULTS
" Press the Esc key (Options) The following window is displayed:
Change Date
Change Time
Level Detection Management
Quit
with the cursor on Quit.
" Press the L key (Change Date), then The TYPE NEW DATE window is displayed:
confirm with the key
TYPE NEW DATE
Format mm*dd*yyyy
The separator key (*) and the date display
format are defined at the page 2 of the
GLOBAL OPTIONS screens, see description
chapter 10.2.3.4.
" Type the new date with respect to the The date is updated.
displayed format and separator then
The following window is displayed:
confirm with the key.
For this exampletype the month
(2 characters), the displayed separator, the Change Date
day (2 characters), the displayed separator, Change Time
the year (4 characters), then confirm with Level Detection Management
the key. Quit
with the cursor on Quit
10-4
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
" Press the Esc key (Options) The following window is displayed:
Change Date
Change Time
Level Detection Management
Quit
with the cursor on Quit.
" Press the K key (Change Time), then The TYPE NEW TIME window is displayed:
confirm with the key
TYPE NEW TIME
Format hh*mm*ss
The separator key (*) and the date display
format are defined at the page 2 of 2 of the
GLOBAL OPTIONS screens, see description
chapter 10.2.3.4.
Type the new time with respect to the The time is updated.
displayed format and separator then The following window is displayed:
confirm with the key.
For this example: type the hour
(2 characters), the displayed separator, the Change Date
minutes (2 characters), the displayed Change Time
separator, the seconds (2 characters), then Level Detection Management
confirm with the key. Quit
with the cursor on Quit.
10-5
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
" Press the Esc key (Options) The following window is displayed:
Change Date
Change Time
Level Detection Management
Quit
with the cursor on Quit.
" Press the K key (Level Detection The LEVEL DETECTION MANAGEMENT
Management) then confirm with the window is displayed:
key
LEVEL DETECTION MANAGEMENT
Needle No.1
Needle No.2 Controlled/Not Ctrl.
Needle No.3 Esc Quit
" Press the key to suppress level detection The following warning window is displayed:
for needle No.1 (samples)
WARNING:
Needle No.1 Level Detection
Management
INHIBITION OF CONTROL REQUESTED
You must DECANT ALL THE TUBES in the
sample drawer
All results will be flagged
Cancel CONFIRM
" Press the J key to CONFIRM and validate The message “not managed” appears.
with the key
The software does not warn the operator
" Decant all the tubes in the sample drawer when liquids are not detected. All the results
are flagged with an alarm code.
10-6
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
" Press the L key to Needle No.2, then The following warning window is displayed:
confirm with the key to suppress level
detection for needle No.2 WARNING:
Needle No.2 Level Detection
Management
INHIBITION OF CONTROL REQUESTED
All results will be flagged
CANCEL CONFIRM
" Press the J key to CONFIRM Then press the key to validate.
The software does not warn the operator
when liquids are not detected. All the results
are flagged with an alarm code.
" Press the L key to Needle No.3, then The following window is displayed:
confirm with the key to suppress level
detection for needle No.3 WARNING:
Needle No.3 Level Detection Management
INHIBITION OF CONTROL REQUESTED
All results will be flagged
CANCEL CONFIRM
" Press the J key to CONFIRM, then press The message “not managed” appears.
the key to validate The software does not warn the operator
when liquids are not detected. All the results
are flagged with an alarm code.
In order to inform the operator that level detection has been supressed for a needle, a warning
message is displayed upon opening of the sample drawer or upon opening of the product drawer.
10-7
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
NOTE This screen corresponds to a consistency check between the workload for STA Compact®
(number of tests to perform except the blocked ones) and the needs to complete this task:
- Products (cuvettes, washing solution, controls, calibrators, diluents, reagents, ...)
- Validated calibrations
- And validated quality controls
It is displayed:
- Upon request from the Status menu
- During closing of the drawers (sample and product), see chapter 6.12
- When tests are inserted in the Patient Files from the FILE PROCESSING screens (Confirmation /
Printout / Transmission menu or TEST PANEL), see description chapter 8.3
- When rerunning tests (see chapters 8.3 and 8.7.4)
- When running calibrations, see chapter 7.3.1
- When running quality controls, see chapter 9.4.1
If after the consistency check, one of the needs to complete the workload is not met, then all the
sample pipetting (sample plasma, controls and calibrators) will be blocked and the “BLOCKED SAMPLE
PIPETTING” message will be displayed in white on a red background at the bottom of the screen.
In this case, the operator can reactivate the sample pipetting for the tests which meet all the
conditions (correct calibration, quality control, volume and stability for all requested products) by
pressing the ESC key and then the Y key.
Otherwise, if after the consistency check, all conditions are met, then the sample pipetting continues
or starts again, the ESC key returns directly to the TEST PANEL screen.
This screen is updated every 5 minutes.
10-8
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
Legend:
- Directly if the “BLOCKED SAMPLE PIPETTING” message is not displayed at the bottom of the
screen
- Or after a confirmation window if the “BLOCKED SAMPLE PIPETTING” message is displayed at the
bottom of the screen (see description chapter 6.12)
4 For the cuvettes and the washing solution (STA< - Cleaner Solution), quantity currently available
5 For the cuvettes and the washing solution (STA< - Cleaner Solution), quantity theoretically
remaining once all the assays have been performed.
When the values are positive (surplus), they are displayed in grey. When the values are
negative (quantity not sufficient to perform all the assays), they are displayed in red.
6 Total number of tests (including the blocked tests) that the STA Compact® has to run without
taking into account calibrations and controls
8 Estimated end of work (hour and minutes) for the required tests, if all necessary products
(reagents, diluents, controls, Desorb Solutions, or calibrators) have been loaded and if their
stability is not expired. The estimated end of work is updated every 5 minutes, according to the
loading of the products.
10-9
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
11 Indication of the positioning zone of the products: D1 (sample drawer for the diluents) or D2
(product drawer for all the other products). When there are more than one vial for the same
product, the volumes of the vials are added and only the addition will be displayed with the recall
of the positioning zone
12 Margin: usable quantity theoretically remaining once all the assays have been performed.
The margin includes all the vials of the given product that are in the drawer.
A positive value indicates a surplus, it is displayed in grey
A negative value indicates missing products, it is displayed in red. These products must be
reloaded to allow the STA Compact® to finish all the tests.
In case of negative values, the zones: 9 (identification) and 10 (name), are also displayed in red.
14 Calibration status:
- Not done
- Running
- Confirm (to be confirmed)
- Control out (control out of range)
- Validated
When the test STATUS SCREEN is displayed, the cursor is positioned on the first product.
The J key moves the cursor to the test status area, then the L and K keys move the cursor from
one test to the other one.
10-10
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
16 The “BLOCKED SAMPLE PIPETTING” message is displayed in white on a red background at the
bottom of the screen when the sample pipetting is blocked whether:
- Voluntarily by the operator (Stop Sample Pipetting function, see description in chapter 10.1.5.1)
- Or automatically by the STA Compact® after the consistency check; if the required assays cannot
be done
Legend:
4 Alphanumeric identification of the product as defined in the TEST SETUP screens (see chapter
5.1), normally displayed in blue; it will be displayed in red in case of insufficient volume, liquid
level detection problem or expired stability.
5 Name of the product as defined in the TEST SETUP screens (see chapter 5.1).
10-11
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
Subsequently, after each sample has been taken, the vial volume is decreased by the sample
volume as specified in the TEST SETUP screens (see chapter 5.1).
When the volume reaches the Min. Volume specified in the TEST SETUP screens (see
chapter 5.1), then the Min. Volume value is displayed in red as well as the alphanumeric
identification of the product.
If, compared to the theoretical decrease in volume, the STA Compact® does not detect the
liquid level, then the displayed volume is 0 in red; the alphanumeric identification of the required
product is displayed in red too, and the product is not useable.
9 Theoretical stability end for each product specified with the month, the day and the time.
To calculate this value, the STA Compact® uses the stability (t) specified in the TEST SETUP
screens (see chapter 5.1). This stability can be edited during the loading step (see chapters 6.5,
6.6 and 6.7).
From the loading of each product (t) the STA Compact® does the following sum:
The displayed color of the stability end and the alphanumeric identification depends on the
following parameters :
10.1.5 Options
The Status menu enables the management of the 3 following options:
- Pipetting stop, see chapter 10.1.5.1.
- On-line transmission of the results, see chapter 10.1.5.2.
- Automatic printout of each patient file when all the related tests are done, see chapter 10.1.5.3.
10-12
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
This option is automatically set to YES by the STA Compact® if after inserting new assays to run
(sample tubes, controls or calibrations), one of the following conditions is not satisfied:
- Sufficient volume
- No risk of stability overdue
- Calibration done and validated
- Quality control done
The operator may then decide to unblock some tests (see chapter 10.1.3) by changing the option to
NO.
The operator may also use this option to stop the sample pipetting by changing the option from NO
to YES. In this case, all sample pipetting (including calibrators and controls) is blocked. However, STA
Compact® finishes the assays in progress for all the cuvettes still on the incubation plate.
If YES is chosen for this option, the other printouts (printout of a file selection...) can be performed
only according to special conditions.
10-13
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
" Press the G key (Global Options) The following window appears:
" Or Press the L key to reach Global
Options, then confirm with the key Please enter
your access code:
: Validate Esc: Quit
" Type your access code, then confirm with The first screen of GLOBAL OPTIONS
the key appears (see description below).
10-14
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
Legend:
1 Esc Opens the EXIT OPTIONS window (see description, chapter 10.2.5)
PgDn Displays the second page of GLOBAL OPTIONS.
10.2.2.2 Printout
Content of the 2 lines of the printout headings. These 2 lines usually identify the laboratory.
10-15
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
Non significant Digit (s) Leading / Ending Number of digits to delete on the left and/or on
the right (1 character)
10-16
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
10.2.2.5 Miscellaneous
10-17
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
Legend:
5 Definition of the time and date display format (see chapter 10.2.3.4)
10.2.3.2 Communications
The following parameters are used to define the characteristics necessary for a central computer
connection (ASTM Protocol):
Station Number (0...99) Allows host computer to identify the STA Compact® (2
characters).
10-18
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
Number of On Error Retries Number of times the STA Compact® sends each frame
after receiving a NACK message. For consistency with
standard E1381-91, this number is set to 6.
Verify Patient Data YES: The first four fields to identify a patient file are sent
by the host computer at the same time as the work list.
These fields can be checked during sample loading (see
File Acquisition chapter 10.2.3.3).
NO: The first four fields to identify a patient file are sent
by the host computer at the same time as the work list.
These fields can not be checked during sample loading
(see File Acquisition chapter 10.2.3.3).
Send Sequence Number Chose YES to send to the host computer the sequence
number the laboratory gives to each file.
10.2.3.4 Miscellaneous
10-19
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
10.2.4.1 Generalities
10-20
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
2GLOBAL OPTIONS screen displayed (page 1 or page 2), the cursor is positioned on data displayed
in blue.
ACTIONS RESULTS
" Enter the new information on the New data are displayed.
alphanumeric keyboard, then confirm with The cursor moves to the next data field.
the key
2GLOBAL OPTIONS screen displayed (page 1or 2), the cursor is positioned on a white parameter
and the message “Key DEL = Modification” is displayed at the bottom the screen.
ACTIONS RESULTS
" When the selected option is highlighted, The selected option is displayed.
press the key
The cursor moves to the next parameter.
" Press the Delete key The TIME FORMAT window appears:
TIME FORMAT
Separator
17:35:24
05:35:24 p
International Time
17 H 35 M 24 S
with the cursor on Separator.
Display of the separator key (:) according to
the last choice.
10-21
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
" Press the key The available separator keys are displayed in
the following window:
/ : - . ,
" Move the cursor to the required separator The TIME FORMAT window is displayed with
key the chosen separator key.
" Confirm with the key
" Press the L key to the required display The TIME ABBREVIATION window is
format. displayed with the cursor on the first letter
of the chosen format.
" Confirm with the key
" Press the Esc key The display is updated. Page 2 of the GLOBAL
OPTIONS screen is displayed.
10-22
0931033L - July 2009
REFERENCE MANUAL 10
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
" Press the Delete key The DATE FORMAT window appears:
DATE FORMAT
Separator
06/08/1993
1993/08/06
08/06/1993
For Year = 1993
Month = 8
Day = 6
with the cursor on Separator.
Display of the separator key (:) according to
the last choice.
" Press the key The available separator keys are displayed in
the following window:
/ : - . ,
" Move the cursor to the required separator The DATE FORMAT window is displayed with
key the chosen separator key.
" Confirm with the key
" Press the L key to the required date display The DATE ABBREVIATION window is
format displayed with the cursor on the first letter
of the selected format.
" Confirm with the key
(j or d= day, m = month, a or y = year)
" Press the Esc key The display is updated. Page 2 of the GLOBAL
OPTIONS screen is displayed.
10-23
0931033L - July 2009
10 REFERENCE MANUAL
STATUS and GLOBAL OPTIONS Screens
ACTIONS RESULTS
" Press the Esc key The EXIT OPTIONS window is displayed:
EXIT OPTIONS
Saving before quitting
Quit without saving
Do not quit
with the cursor on Save before quitting.
10-24
0931033L - July 2009
REFERENCE MANUAL 11
Table of contents
11 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
11-i
0931033L - July 2009
11 REFERENCE MANUAL
Table of contents
11-ii
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
11 Troubleshooting
An Emergency Stop can be performed by pressing ALT - F10. This will access the USER
MAINTENANCE menu.
Tests that are were in process prior to the emergency stop will be rerun when the analyzer restarts.
The power off procedure of the STA Compact® is described in chapter 3.6.2.
Follow the instructions given by the STA Compact® and if the problem persists, call your local
authorized Service Representative
11-1
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
01 Used Liquid Container " Check that the closing of the switch
Check presence detecting the presence of the bottle is
heard when it is replaced, see
chapter 3.5
" If the problem persists, call your local
authorized Service Representative
01 Used Liquid Container " Check the liquid level in the bottle of
Check maximum allowed used liquid, replace if necessary, see
volume chapter 3.5
" If the problem persists, call your local
authorized Service Representative
01 Cuvette Disposal Drawer " Check that the cuvette disposal drawer
Check presence is pushed fully home
" If the problem persists, call your local
authorized Service Representative
11-2
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
11-3
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
11-4
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
Cuvette Missing
11-5
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
11-6
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
11-7
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
Procedure:
" Press the Esc key to close the message
" Replace the lamp (see chapter 4.5.3)
The photometric tests will resume only if the lamp has been replaced.
MANAGEMENT OF ARM #2
Error 07.30.30
REMOVAL
Missing cuvette
Esc=Continue
" Press the Esc key to delete the error message.
When the STA Compact® goes to standby mode (no more measurement in progress), the following
message is displayed:
Start of test
Error 08.20.00
CUVETTE ERROR
At least one cuvette is still on
the measuring block
How do you want
to remove it?
F1=Automatic F2=Manual
Possible functions:
11-8
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
F1: Arm No. 2 (Suction Head) tries again to pick up the cuvette.
In case of defective suction tip or defective cuvette, this operation is impossible and may lead to
jammed situations.
F2: The operator can open the transparent panel and manually remove the cuvette which cannot
be automatically picked up (with usual precautions to biohazardous material).
Suction tip defective " Clean the suction tip (see procedure
chapter 4.2.8)
" Replace it if necessary (see procedure
chapter 4.5.4)
Cuvette in poor condition, for example: " Replace the roll of cuvettes
broken wings.
Bead present in a measurement or " Access the User Maintenance menu (see
incubation well, causing the cuvette to be chapter 4.7.1)
wrongly positioned
" Remove the transparent panel
" Remove the bead
Suction head incorrectly positioned on the " Call your local authorized Service
cuvette. Representative
11-9
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
RISK OF INJURY
In order to avoid risks of injuries by the
pneumatic jack:
- Do not use fingers.
- Use a tool to remove the cuvettes jammed
in the loading system.
Shuttle detection switch faulty " Call your local authorized Service
Representative
11-10
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
2 Check the rinsing volumes: run the function Rinsing Pump of User Maintenance menu (see
chapter 4.7.2)
3 Run a quality control and compare the result with the previous results
5 Verify the test setting: in the Test Setup menu, check the configuration of the test(s) causing a
problem and compare with the test settings recommended by Diagnostica Stago.
7 Run a quality control for a major test and compare the result to the previous results
9 Check that the appropriate weekly, monthly and quarterly maintenance has been carried
out
11-11
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
Short Short OK Contamination " Pass the mandrel through the needle
of needle No.3 No.3 to unblock it (see procedure
chapter 4.2.4)
" Check the suction in washing well No.3
(see procedure for cleaning washing wells
in chapter 4.2.4)
" Replace the vial of STA<- DESORB U.
" Call your local authorized Service
STA< Desorb Representative
U contamined
Short Short Short Needle No.3 " Check the rinsing of needle No.3 (see
or or or incorrect procedure chapter 4.7.2.2)
long long long pipetting " Pass the mandrel through needle No.3
to unblock it (see procedure chapter
4.2.4)
" Replace the Teflon tip of the syringe:
(see procedure in chapter 4.3)
" Call your local authorized Service
Representative
11-12
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
For a selected test, analyses Test has been blocked by Unblock the test (see
are not run. means of the File procedure chapter 8.7.3)
Modification menu (see
" View the TEST STATUS
chapter 8.7.3).
screen: the missing
Block is displayed in red on
products are displayed in
the TEST PANEL instead of
red (see chapter 10.1.3)
the result for this test.
" Load the missing products
Products are missing to run if necessary
the test.
" View the TEST STATUS
For this test, the calibration
screen, the calibration
is not confirmed.
status for the selected test
For the selected test, no is not: Validated (see
quality control has been description chapter 10.1.3)
run for more than
" Run the calibration of the
24 hours.
selected test
" View the TEST STATUS
screen, the quality control
status for the selected test
is not: Validated (see
description chapter 10.1.3)
" Run a quality control for
each selected test
None of the assays is run. The sample pipettings have " Load if necessary the
been blocked whether by missing products (see the
the Stop Sample list in the TEST STATUS
Pipetting function of the screen)
Status sub-menu (see
" Run the missing
description chapter
calibrations (see the list in
10.1.5.1) or from the TEST
the TEST STATUS screen)
STATUS screen (see
description chapter 6.12). " Run the missing controls
(see the list in the TEST
STATUS screen)
11-13
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
No response from keyboard The power cord is not " Connect the power cord
and STA Compact® does not connected. correctly
start.
The wall socket is not " Plug the STA Compact®
active. in elsewhere
11-14
0931033L - July 2009
REFERENCE MANUAL 11
Troubleshooting
Legend:
1 White zone = 15 mm
2 55 mm max.
11-15
0931033L - July 2009
11 REFERENCE MANUAL
Troubleshooting
11-16
0931033L - July 2009
REFERENCE MANUAL A
Safety appendix
A Safety appendix
A copy of the Material Safety data Sheet for the coolant (Frigoporteur T1) used in the STA Compact®
is attached.
Formulation of the Frigoporteur T1 coolant may use a pink dye or a blue dye.
The solutions are compatible and may be mixed.
A-1
0931033L - July 2009
A REFERENCE MANUAL
Safety appendix
A-2
0931033L - July 2009
PROFROID I n d u s t r i e s S. A. S
Z.I. des Paluds 178, rue du Fauge 13782 AUBAGNE Cedex
tél : +33 4 42 18 05 00 - FAX : +33 4 42 18 05 07
FRIGOPORTEUR T1
2/ Composition/information on ingredients
. Substance contributing to danger
Monoethylenglycol(CAS107.21.1) :Xn R22 -content 20-40%.weight by weight
Methanol (CAS-67.56.1) F and T :R 11- 23/24/25;R39/23/24/25 content 2.5-5%weight by weight
. Substances classified as "dangerous" -
3/ Hazards identification
Preparation classified as dangerous according to 1999/45/EC Directive,
Labelling information:
Hazard Symbol : Xn-harmful
Risk phrases: R20/21/22: Harmful by inhalation, in contact with skin and if swallowed.
R68/20/21/22: Harmful: possible risk of irreversible effects through inhalation,
in contact with skin and if swallowed.
Safety phrase: S36/37: Wear suitable protective clothing and gloves.
5/ Fire-fighting Measures
. Fire or explosion hazard: liquid with flash point over 55° C
. Fire-fighting equipment:
- recommended: CARBON DIOXYDE (CO2) or POWDER
- do not use: DIRECT WATER SPRAY
. Protective measures while fire-fighting:
Use autonomous breathing equipement and wear protective clothing
. Other recommandations: In case of fire, do not breathe the emitted fumes.
Protective clothing
Hygienic measures : Do not eat, drink or smoke when you use this product
Wash hands thoroughly when finished
Biological limit values: not available.
Engineering controls: not available.
Australian exposure standards: not available.
aspect Limpid liquid specific gravity (at 20°C) 1.03 (+/- 0.03)
colour Pink .
odor characteristic water solubility miscible
Product must be used outside or in a ventilated area and kept away from fire.
Incompatibility: keep product away from heavy acids and oxidizing agents
Under normal conditions of use, the product is not irritating to the respiratory tract
inhalation
This sheet completes the technical instructions but does not replace them
and does not give guaranteed or insurance data on the product properties.
This information is given according to our technical knowledge on the product
for the indicated date.
User's attention is drawn on the possible risks incurred
when a product is used for purposes other than those for which it has been designed.
This sheet does not, in any case, prevent the user from knowing
and following the whole regulatory texts related to his activity.
It is the user's responsability to determine the suitability of this information and to ensure that all
appropriate safety precautions are respected. We accept no responsability for indirect damages.
A
Access codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-16
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23, 5-41
Audible alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17
List of alarms associated to results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-41
ASTM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22, 10-18
Communication parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-18
Communication protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Connexion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Host computer communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-88
Automatic identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8, 6-13, 6-20
Bar code identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9, 6-13
Automatic rerun . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-15, 9-10
Average. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
B
Bar code
Problems regarding bar code reading . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-15
Bar code test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-62
C
Calculated Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-40
Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17, 5-18, 5-19, 6-1, 6-13, 6-15
Clot. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Colorimetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18, 2-20, 2-21
Cum
Cum daily . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Cum monthly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Cuvette disposal bag. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Cuvette disposal bag replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Cuvette rolls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Cuvettes
Number of cuvettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
CV (Coefficient of variation). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
D
Dates
Date format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-20
1
0931033L - July 2009
REFERENCE MANUAL
Index
E
Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4, 3-9, 3-17, 11-1, 11-7
Error history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-86
Error messages
Description of an error window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Error messsages
Problems regarding tube reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-15
F
Fuses
Location of the fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-50, 4-51
Replacement of the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-49, 4-51
G
Global options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-13
I
Immunology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18, 2-20, 2-21
Incubation-measurement zone. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
INR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10, 5-12, 5-22, 7-4, 7-11, 7-44
2
0931033L - July 2009
REFERENCE MANUAL
Index
L
Level detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14, 10-6
Loading
Generalities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Loading the roll of cuvettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Loading the Washing Solution Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Product loading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5, 6-8, 6-13
Sample loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6, 6-17, 6-20
M
Main Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Preventive maintenance schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Management
Level detection management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2, 10-6
Liquid level detection management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Product management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
Sample management. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
Manual Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11, 6-15, 6-18, 6-28
Manual identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8, 6-13, 6-20
N
Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12, 8-13
O
Offset corrector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
Optical density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
P
Patient files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Customer printout of validated files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-32
Routine mannagement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
3
0931033L - July 2009
REFERENCE MANUAL
Index
Q
QNS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8, 8-9
Quality Control
Description of the quality control screens . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Level of quality control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
Range of acceptable results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
R
Redilution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8, 5-12
Reference time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8, 7-14, 7-19
Restoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-34
Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10, 5-41
Choice of unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17
Graphic display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-65
Results display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Results status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Transmission of results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
Roll . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8, 3-13
Running control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9, 9-17
S
Samples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-39
Saving
Saving parameters from global options . . . . . . . . . . . . . . . . . . . . . . . . . . 10-24
Screens
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1, 10-13
Shut down
Extended shut down. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Routine shut down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Standard deviation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Stirring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4, 5-9
4
0931033L - July 2009
REFERENCE MANUAL
Index
T
TDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-90
Test
Teleloading tests from host computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-31
Test panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Test Setup
Dependent test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29, 5-34
Main test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28, 5-34
Methodology. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Test setup parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
Time
Modifying the system time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Modiying the time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Time format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-19
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
U
Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21, 5-25
Arbitrary unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17
Primary unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Unloading of products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-39
Unloading of samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-39
Urgent
Loading a STAT sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1, 8-9, 8-11
V
Validation section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
W
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2, 1-3, 1-5
Working File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1, 8-6, 8-23, 8-25, 8-26
Working List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44
5
0931033L - July 2009
REFERENCE MANUAL
Index
6
0931033L - July 2009