Span calibratior

(7) Calibrate

1. Input to Select the Calibration Mode.

FUNC-CRN
E-1-CEN
Digital Zero is released and Zero Tracking is inhibited. Put display in Gross
weight mode.

2. Zero Calibration

(a) Make sure there is no excess load on the scale.

(b) Make sure the STAB. lamp is ON. (Calibration should be done when scale is
stable.)
(c) Do Zero Calibration.

FUNC.-CRN-1-CEN Calibration Mode.

ZERO- CEN

(d) cErrê or LoR d The initial Tare is


beyond the zero adjustment range.
(e) CErr3or- L oRd The initial Tare is minus.
CErr9 The weight value is not stable and
Calibration stopped.

3. Span Calibration

(a) Put a Balance Weight on the loadcell (or scale). In order to


get ideal linearity of the scale, the
Balance Weight must be more than 50% of the Capacity and within the Capacity.
(b) Make sure the STAB. lamp is ON.
(cC) Input the Balance Weight.

FLUNC-CEN -CEN Calibration Mode.

CEN
Balance Weight