现场操作培训记录卡
On-site Operation Training Record Card
培训项目
Training project
水质检测仪 bante900 多功能产品说明书
Water Quality Meter bante900 Multi-function Instruction
(电导率测定 Conductivity Measurement)
培训日期 培训老师
Training Date Trainer
巡回检查的内容和标准(Contents and standards of patrol inspection):
Setup
Call 1
1 点校准
point 1 Calibration
1) 按 Mode 键直至显示 COND 模式。
1) Press “Mode” until it shows COND mode.
COND Measure
0.0000μs
25.0℃
2) 在参数设置模式设置校准点为 1 个。
2) Set the calibration point number is 1 under the parameter setting mode.
3) 用除盐水清洗电导电极,再用少量标准液冲洗。
3) Rinse the electrode with demineralized water, and then rinse it with a small amount of
standard solution.
4) 按 Call 键,仪表进入校准模式。
4) Press “Call”, instrument enters calibration mode.
5) 将电导电极浸入校准液中缓慢搅拌,仪表自动侦测当前校准液并显示默认 校准值
(例如: 1413μs)
5) Soak the electrode into the standard solution and slowly stir. Instrument will
automatically measure the standard solution and shows the default calibration value
印尼庞卡兰苏苏火电项目部
PANGKALAN SU SU COAL FIRED STEAM POWER PLANT PROJECT UNIT 3&4
2X200MW
COND Calibration
1413μs
Call 1
7) 等待数值稳定后,屏幕显示 END,仪表返回测量模式,校准完成。
7) After the value is stable, the screen will show “END”, instrument will return to
measurement mode, and calibration is completed.
COND Calibration
END
2 点校准:
point 2 Calibration:
1)确保在设置菜单中设置了校准点为 2 点。
1) Make sure that the calibration point number is set to be 2 under the parameter setting
mode.
2) 用除盐水清洗电导电极,再用少量标准液冲洗。
2) Rinse the electrode with demineralized water, and then rinse it with a small amount of
standard solution.
3) 按 CAll 键,仪表进入校准模式。
3) Press “Call”, instrument enters calibration mode.
4) 将电导电极浸入校准液中缓慢搅拌,仪表自动侦测当前校准液并显示默认 校准值
(例如: 1413μs)
印尼庞卡兰苏苏火电项目部
PANGKALAN SU SU COAL FIRED STEAM POWER PLANT PROJECT UNIT 3&4
2X200MW
4) Soak the electrode into the standard solution and slowly stir. Instrument will
automatically measure the standard solution and shows the default calibration value
COND Calibration
1413μs
CAl 1
6) 等待数值稳定后,屏幕自动显示 CAll 2,仪表等待第 2 点校准。
6) After the value is stable, the screen will show “Call 2”, instrument is ready for point 2
calibration.
COND Calibration
μs
CAl 2
7) 用除盐水清洗电极,再用少量校准液冲洗。
7) Rinse the electrode with demineralized water, and then rinse it with a small amount of
standard solution.
8)将电导电极浸入下一个校准液中,仪表立即显示当前校准液的默认值
(例如:12.88mS/cm)。
8) Soak the electrode into the next standard solution and instrument will automatically
measure it and shows the default calibration value of this standard solution (for
印尼庞卡兰苏苏火电项目部
PANGKALAN SU SU COAL FIRED STEAM POWER PLANT PROJECT UNIT 3&4
2X200MW
example: 12.88mS/cm).
9) 如果需要,按↑或↓键修改校准值,或直接按 Enter 键,Calibration 图标开始闪烁。
9) If there is a need, press ↑ or ↓ to change the calibration value, or directly press “Enter”,
Calibration sign will be flashing.
COND Calibration
培训签到 SIGNATURE