• Temperature-programmed desorption
• Pulse chemisorption
Innovative Design: AutoChem II 2920 Hardware Advantages The AutoChem II Technique
The AutoChem II features stainless- • Analysis gas may be introduced to temperature rapidly down to near During the TPR, a metal oxide reacts
steel construction, fully automated flow the carrier stream by a precision ambient, reducing analysis time with hydrogen to form a pure metal. This
and pressure control, an embedded automated loop. A calibrated, and increasing throughput. With reaction is referred to as a reduction; for
microprocessor with real-time control, conveniently located septum the CryoCooler option, analysis example, TPR of a catalyst containing
and an intuitive graphical user interface is also provided through which temperatures can be ramped over platinum. Argon, which has a very low Gas Flows into the
sample tube and
for reactor control. A temperature- analysis gas can be injected by the range of -100 °C to 1100 °C. relative thermal conductivity, is used
through the sample
controlled, stainless-steel flow path means of a calibrated syringe. as a component in the carrier gas. It
provides an inert and stable operating • Four internal temperature-controlled is blended in a fixed proportion with
environment, and reduces the potential • Thermal conductivity detector zones can be heated independently of hydrogen, the reducing gas with a much
for condensation in the flow path. (TCD) is capable of detecting minute each other to as much as 150 °C. higher thermal conductivity. Then the
differences in the concentration This prevents condensation in gas mixture flows through the analyzer,
• Twelve gas inlets (4 each for prep, of gases flowing into and out of the flow path and allows studies through the sample, and past the
carrier, and analysis) provide the the sample reactor. Its corrosion- to be performed with vapors. detector. When the hydrogen and argon
capability to perform sequential resistant filaments are operated at gas blend begins flowing over the sample,
experiments such as TPR/ constant temperature to prevent • The extremely low volume of the a baseline reading is established by the
TPO cycle experiments. thermal runaway, which can destroy internal plumbing minimizes peak detector. This baseline is established
filaments in other systems. spreading and significantly enhances at a low enough temperature so that
• Equipped with four high-precision, peak resolution. Furthermore, it no reduction of the sample occurs. The
independently calibrated mass flow • A clamshell furnace can heat the reduces the time lag between baseline level indicated by the detector
Before the reaction begins, the
controllers providing extremely quartz sample reactor up to 1100 °C. the sample reactions and the is that of the thermal conductivity of concentration of gases exiting
accurate, programmable gas The AutoChem II 2920 permits corresponding detector response. the two gases in their fixed proportion. the sample tube is unchanged
control. The ability to control any number of ramp rates and In other words, the proportion of gases after flowing through the sample
gas flow with accuracy assures sequences to facilitate customized flowing over the detector is the same
a stable baseline and accurate experiments. The included KwikCool as the proportion of gases entering the
determination of gas volumes. feature permits cooling the furnace analyzer because at the low temperature
there is no interaction with the sample.