KS DCL Manual Original Casing - Rev01 - Organized
KS DCL Manual Original Casing - Rev01 - Organized
Importance Safeguards
All safety instructions must be read carefully & understood, before attempting to use any KS Product
(instrument, probe, equipment). This manual contains the safety instructions that must be observed in
order to avoid potential hazards that could result in personal injuries or could damage your (instrument,
probe, equipment). This safety instruction has been classified according to the seriousness of
the risk, as follows and all identified risks categorized accordingly.
Upon receipt of the instrument, the user should verify that the above listed accessories are included.
Website : http://www.korosispecindo.com
KOROSI SPECINDO
Model:
KS-DCL : Intrinsically Safe Corrosion Data logger
Physical Data:
• Instrument Weight : 0.40 kgs
• Instrument dimension : 96 mm x 48 mm x 80 mm
• Ambient Temperature : -200 C ~ +600 C
Feature:
• Input Signal : 4-20mA line (2 wire / 3 wire)
• Display type : 7 segment display, 4 digits
• Additional Feature : Modbus RS-485 converter (optional Modbus output)
Retrievable up to 4000 data with KS format via USB
IS (Intrinsic Safety) certified
Easy operation
Electrical Data
• Power Supply : 12 – 26 VDC input
• Power : app. 40mA normal operation, 60mA during USB copy
3. Instrument Installation
DCL is used for display from analog 4-20mA transmitter data to represent in actual value (mils / mpy unit
for ER transmitter or mm unit for UT thickness gauging), and can be installed in both Hazardous area (but
covered from water drops) and also in safety area.
3.1 Preparation
• Safety Procedure always ask for a person who will be in charged in the field, specifically an
Instrument, or Electrical Engineer who will take responsibility in the field during Installation of KS-
DCL.
• Follow wiring diagram and install it correctly before connecting to power supply
• KS-DCL must be protected for water drops, rain or potential damp environment
• Always Test all of your Equipment before you attempt to installed this in the field. Prepare any
tools/instrument such as, multimeter, 4-20mA/HART tester, etc.
• Bring all Documentation and Manual Book along to the field for conducting Installation, setting
parameter and Troubleshooting.
Note: Operator must concern to Local Safety Requirement, the safety requirements according to described
in operating and maintenance manual of equipment, the requirements of the relating explosion
proof standards as well as the requirements and instructions of EN60079-10, EN60079-14, EN60079-
17 latest standard.
3.2 Power supply
Adjust power supply output to its rating voltage before connect to KS-DCL.
Setting should be done in safety area, as it needs to open the front cover and access to the operation
button of KS-DCL.
OK
down
7 segment
Up
display
USB
port
The display of main screen will be between 0.000 to 25.00 based on the actual 4-20mA actual reading (if
already connected to 4-20mA loop)
(button OK Pressed)
(Button Back Pressed)
4800
9600
19200
38400
57600
115200
(button OK Pressed)
(Button Back Pressed)
high
Medium
Low
(button OK Pressed)
(Button Back Pressed)
Protected from changes,
button OK Pressed)
(success)
(Button Back Pressed)
To save parameter setting, press OK after change the value, and device will display if
successfully saved to EEPROM memory, and will display if error occurred and need to
repeat the setting procedure again.
When USB Flash drive is plugged in to KS-DCL USB slot, the display will change to
Press OK for downloading from internal memory to USB flash drive, the device will display the remaining
KS-DCL will count down the display until zero/ done. and will display “DONE” if data successfully
downloaded to USB Flash drive, otherwise it will show display “E0XX” if error occurred.
XX : exception code
WARNING : USB Flash drive downloading only can be performed in Non-hazardous area, and shall not be
done if KS-DCL is connected in Hazardous area as Fig. 3
4.4 Calibration
By defaults KS-DCL has been calibrated from the factory. make sure unclock the protection
from parameter menu before doing this calibration procedure, if you want to change the
calibration procedure as follows:
- connect KS-DCL to reference meter/current source at Loop+ and Loop- terminal
- set Current source to 4mA (you can use DCI -Tx as current source)
- at main screen/ default screen press & hold CANCEL + DOWN button until displayed “Zero”
- set current source to 20mA
- press & hold CANCEL + UP button until displayed “Span”
- finish, you can verify it by current loop test.
4.5 Maintenance
Note: Operator must concern to Local Safety Requirement, the safety requirements according to described
in operating and maintenance manual of equipment, the requirements of the relating explosion
proof standards as well as the requirements and instructions of EN60079-10, EN60079-14, EN60079-
17 latest standard.
5. Firmware
KS-DCL is a micro controller based unit and all its functions are firmware controlled. These functions are
accessed by issuing appropriate commands as per the command set.
• The cleaning of the enclosure of the KS-DCL unit must be done only with damped clothes because of the
danger of the electrostatic charging
• Do not replace the battery pack of the corrosion data logger KS-UT100 in a potentially explosive
atmosphere.
• In operation the local safety requirements, the safety requirements according to described in operating-
and maintenance manual of equipment, the requirements of the relating explosionproof standards as
well as the requirements and instructions of EN 60079-10:2003, EN 60079-14:2003, and EN 60079-
17:2003 standards must be kept
Appendix 1. Parameter value for Probe type