Abstract: A number of medical and research communicating internal vital signs to the outer
applications require implantable devices to world. Providing power to these devices for long
locally stimulate internal organs and term is a challenging task [1]. Traditionally,
communicate the internal vital signals to the implantable batteries and percutaneous link
outer world. Wireless power transfer (WPT) power supply systems are used. However, the
technique is used to supply power to these batteries have limited energy storage and life
devices. The power is transferred wirelessly from span, and the percutaneous links across the skin
a stationary primary coil to a movable secondary impose infection risks [2]. For applications
coil. The secondary device is implanted in a involving rodents, battery size and weight is
rodent, which moves freely inside a cage around
more problematic for long duration experiments
which the primary is wrapped. However, the
(1 or more days). In some cases, implants use
continuously changing orientation of the rodent
rechargeable batteries, where those systems
leads to coupling loss/problems between the
primary and secondary coils. We propose employ wireless recharging [1].
configurations of the secondary coil employing The use of animal models for human disease
different size of ferrite rods placed at specific processes plays a vital role in biomedical
locations within the coil. Three dimensional research. A vast majority of the research on
finite element analysis (FEA) using COMSOL neural mechanisms of therapies is currently
software, is used to find the magnetic flux conducted using rodent models. Implantable
density distribution surrounding these secondary biotelemetry systems are effective tools for
configurations. The simulation results show a clinical medicine and also in animal research,
significant increase of flux through the coil using because they allow for the acquisition of
the ferrite arrangement, with improved coupling otherwise unavailable physiological data [1].
at most orientations. The secondary coil Rodents are important animals used for
configurations were constructed and experiments biomedical applications, and are widely used to
were conducted to test their performance. study various disease models [3]. When studying
Measurements show that ferrite rods improved the behavior of rodents and gathering biological
power transfer, where the long ferrite (4LF) data from them, miniature sensors and telemetry
configuration is highest received power. devices are needed. These devices must provide
Experiments show the maximum power continuous monitoring without limiting rodents
collected by 4LF was 127 mW when parallel to mobility or behavior. This work motivates to
the primary coil. power these sensors and devices.
This work proposes the use of a wireless power
Keywords: Wireless power transfer, Magnetic
transfer (WPT) system to continuously power a
resonance, Biomedical device, Finite element
rodent telemetric device. It was found that a
analysis, Implants.
number of researchers are investigating WPT for
rodents, as mentioned in [1, 4]. The proposed
1. Introduction
application is for a rodent implant device (RID)
that can act as a stimulator and sensor, for a
Implantable biomedical telemetry devices have
freely moving small rodent. The current RID
gained much attention for a variety of
prototype is suitable for subcutaneous
applications including generating stimulus
implantation in a rat, and is suitable as a head-
signals, monitoring the body, and