Professional Documents
Culture Documents
CL vs Glasses
• Therapeutic CLs
• Cosmetic CLs
Sir JOHN HERSCHEL (1845) FIRST CL (all CLs are made of glass until 1930): Correction of
irregular corneal astigmatism (effect of the tear film). Shape of the CL is determined by
creating a mould made of clay
PRIMERA CL (todas las CL son de cristal hasta 1930): Corrección del astigmatismo corneal
irregular (efecto de la película lagrimal). La forma de la CL se determina creando un molde
hecho de arcilla.
DA VINVI: The first idea: to replace the anterior corneal surface with another surface of known
radius, with a liquid between both surfaces (Da Vinci does not manufacture any CL, only
presents the idea)
La primera idea: sustituir la superficie anterior de la córnea por otra de radio conocido, con un
líquido entre ambas superficies (Da Vinci no fabrica ninguna CL, sólo presenta la idea)
• Ernest Mullen (describes the refractive effect of the tear film under the CL) and
Theodore Obrig (uses sodium fluorescein to stain the tear film and assess CL fit)
Hydrogel CLs
• Spin casted
Disposable CLs
Silicone-hydrogel CLs
– Many others….
• In general:
– Children: spherical
– Adults: spherical
• Eccentricity:
• Asphericity: Q = - 2
Mean value of = 0.47
• So, to describe the corneal shape we should always provide a value of apical radius (or
2 if astigmatic) + a value of to show how this apical radius changes as we move from
the center.
• A single value of may not even be enough, as the rate of change may change towards
the periphery.
Un único valor de e puede no ser suficiente, ya que la tasa de cambio puede cambiar
hacia la periferia.
• Mean posterior corneal radius 10% less than anterior corneal radius
• Therefore it is thinner towards the center and thicker towards the periphery
• P = (n’ – n) / R
• n aqueous = 1.336
Keratometry
• Useful to determine the two main anterior radii of the central (3.6 mm in diameter
approx.) of the cornea
The cornea acts like a convex mirror and creates a virtual image, upright and smaller than the
object (First image of Purkinje, P1)
• Keratoconus
• Refractive surgery
Keratometer types
• Helmholtz: Measures both meridians at the same time. Fixed mires. Internal mobile
prism system. Move far back and approach slowly until “something is seen in focus”.
Move horizontally and vertically until mires are centered on the central cornea. Adjust
distance precisely. Inferior right mire must be single. The black cross of the eyepiece
must be inside the inferior right mire. Approach or separate mires to measure
horizontal and vertical at the same time. May need to rotate the body of the
instrument if astigmatism is not at 90° or 180°.
Mide ambos meridianos al mismo tiempo. Muros fijos. Sistema de prisma móvil
interno. Mover lejos y acercarse lentamente hasta que "se vea algo enfocado". Mover
horizontal y verticalmente hasta que los mires estén centrados en la córnea central.
Ajustar la distancia con precisión. El mires inferior derecho debe ser único. La cruz
negra del ocular debe estar dentro del mires inferior derecho. Aproximar o separar los
mires para medir la horizontal y la vertical al mismo tiempo. Puede ser necesario girar
el cuerpo del instrumento si el astigmatismo no está a 90° o 180°.
The keratometer must be calibrated with calibration spheres of know radii. Adjust the
eyepiece (very important to ensure the mires are seen in sharp focus at the correct (d)
distance. Adjust de position of the patient. The patient must fixate the light or fixation point
(so that the instrument is aligned with the corneal apex).
El queratómetro debe calibrarse con esferas de calibración de radios conocidos. Ajustar el
ocular (muy importante para asegurarse de que los mires se ven bien enfocados a la distancia
(d) correcta. Ajustar la posición del paciente. El paciente debe fijar la luz o el punto de fijación
(para que el instrumento esté alineado con el ápice corneal).
Notation
Types of astigmatism
• x 90° denotes the direction of the axis, therefore the meridian with the highest
curvature (maximum power) is at 180°
• We are saying the same as in Optometry, but in CLs we prefer to talk about corneal
“topography” instead of refractive power.
• In CLs we will not be using Javal’s rule because: It refers to keratometry astigmatism,
and we know that this is valid only for “normal corneas”. It is based on “statistics”.
Useful in general, not very useful for a particular patient. We can actually measure
posterior corneal astigmatism now!
Corneal Topography
Based on the Placido Disc, and the keratometry principle. The software of the topographer
determines radii from the distance (h’) between two consecutive reflected circles at different
degree intervals.
• Radii at each point. Therefore, we can know local changes, eccentricity, etc.
• Axial (or sagittal) map: radius of a point is “averaged” from that point to the visual axis.
Not good at showing local abrupt changes, smoother.
El radio de un punto se "promedia" desde ese punto hasta el eje visual. No es bueno
para mostrar los cambios locales abruptos, más suaves.
• Tangential map: radius of a point is compared to next point (from next ring). Very good
at showing local changes, such as exact position and shape of a keratoconus.
El radio de un punto se compara con el siguiente punto (del siguiente anillo). Es muy
bueno para mostrar cambios locales, como la posición exacta y la forma de un
queratocono.
Elevation map
• Long posterior ciliary nerves from the ophthalmic branch of the trigeminal nerve
• These nerves enter the cornea at anterior half of the stroma and ascend towards the
surface of the epithelium. They “normally” do not have myelin (from 1.5 mm of
limbus).
Estos nervios entran en la córnea en la mitad anterior del estroma y ascienden hacia la
superficie del epitelio. "Normalmente" no tienen mielina (a partir de 1,5 mm del limbo).
Corneal vascularization
• The cornea is an avascular tissue. Capillaries reach the limbal area and do not enter
the cornea (in normal conditions).
• When there is hypoxia (e.g., caused by a CL), capillaries may cross the limbus and enter
the cornea.
• Stroma
• Endothelium
Corneal epithelium
• Stem cells at the limbus able to regenerate the whole epithelium (important for
certain types of refractive surgery)
Stroma
• Composition:
– Collagen: type I
– Matrix
• Collagen fibers form about 250 collagenous lamellae (about 300 at the central cornea,
500 at the periphery).
• Collagen fibers are separated about 57 nm at the center of the cornea, about 62 nm at
the periphery.
• These fibers and lamellae provide biomechanical properties to the cornea (resist IOP).
– Anterior: the anterior 100 µm of the stroma has 50% more lamellae than the
posterior stroma. Lamellae are thinner but intertwined. More resistant. Very
compact, no spaces. Highest density of keratocytes.
los 100 µm anteriores del estroma tienen un 50% más de laminillas que el
estroma posterior. Las láminas son más finas pero están entrelazadas. Más
resistentes. Muy compacto, sin espacios. Mayor densidad de queratocitos.
– Posterior: lamellae are thicker but disposed one upon the other with little
intertwining. Less resistant. It contains spaces where water from the aqueous
humor may enter.
Las láminas son más gruesas pero están dispuestas una sobre otra con poco
entrelazamiento. Es menos resistente. Contiene espacios por donde puede
entrar el agua del humor acuoso.
Corneal endothelium
• Single layer or cells without mitosis (does not regenerate if cells are lost)
• Minimum of 2800 cells /mm2 to implant a IOL of anterior chamber (and 3200 µm of
anterior chamber depth)
What is required for the cornea to be transparent?
• Polarographic technique.
• Sensor with a known volume of oxygen in contact with the anterior corneal surface.
• Oxygen moves from the sensor to the cornea. Volume / time.
• Epithelium, stromal, endothelium consumption: 2:2:1 ratio
• Stromal consumption only when keratocytes are active, otherwise ratio is 4:0:1
• Endothelium: to return water entering the posterior cornea to the aqueous humor
through the ion pump (the ion pump moves ions, not water: water follows along out of
the stroma to reduce and equilibrate concentration of ions).
• Remember: the cornea does not have blood vessels to bring oxygen and nutrients!
– Aerobic pathway:
• Tricarboxylic acid cycle, also known as the Krebs cycle (36 ATP)
• Products: CO2 and water, which migrate towards corneal surface and
tear film
• Oxygen restriction
Oxygen restriction
– CL diameter (RCL smaller than the cornea, soft CL cover the whole corneal
surface)
Tear exchange underneath the CL:
Physiological edema
• Closed eye: 8% of oxygen at the epithelium, but 10% needed for correct aerobic
metabolism….
• RESULT: corneal physiological (normal) edema of 2 to 4% on opening the eyes in the
morning. Water has entered the cornea at the endothelium / aqueous humor interface
and accumulated at the posterior stroma, leading to slight swelling (increase in corneal
thickness).
• Water tends to enter the cornea when there is hypoxia, leading to anaerobic
metabolism and lactate migration towards posterior cornea. Water from aqueous
humor enters to equilibrate concentration of ions (osmotic equilibrium). This happens
precisely when the endothelium has less energy for the ion pump!
• Energy reserve: 30 minutes
• Maximum edema on eye closure: at 4 hours
• Increased oxygen demand during the first 10 minutes after eye opening
Using a CL with low Dk/t at night (oxygen available on the cornea will be even
less than 8%)
• They originate at the level of the basal cells of the epithelium (abnormal mitosis due to
lack of energy)
• They may show fluorescein staining when they reach the corneal surface
• Once resolved, ghost vessels may remain that could refill very fast if hypoxia occurs
again
CONJUNCTIVA
• CONJUNTIVAL HYPEREMIA
– Toxicity to solution
• Papillary conjunctivitis
– Commonly related to dry eye and CL discomfort, and to LWE (see later)
Limbus
Eyelids
Blinking
– Correctly distribute tear film over the ocular surface: on eye closure, it
compresses the tear film like a curtain.
– Drain tear film through the lachrymal puncta at the nasal canthus.
• Each blink is associated with a loss of visual information of 400 ms: Blink suppression.
We are not aware of closing the eyes!
• Ocular dryness
• It gives uniformity to the corneal surface, covering the microvilli and plicae of the
surface epithelial cells. This facilitates corneal transparency.
• It provides O2 to the cornea and eliminates CO2, dead cells, dust particles....
Atmospheric O2 first diffuses to the tear film before reaching the cornea.
• It provides lubrification to the anterior surface, offering comfort during and between
eyeblinks.
• It contains antimicrobial agents to prevent
anterior ocular surface infection and provide
stability to natural ocular microbiome.
• The thickness of the tear film depends on the time since the last blink and on the
position on the cornea:
– It is recommended to carry out measurement 2 second after the last blink (for
example for corneal topography).
Lipid layer
– Polar internal layer with hydrophilic part in contact with the aqueous layer
Aqueous layer
• Composition:
Water: 98%
Electrolytes: K+, Ca2+ & Mg 2+
Vitamins A & C
Dead epithelial cells
About 2000 types of proteins, amongst which many bactericidal agents:
• Lysozyme: 25% of tear proteins. Decreases in chronic eye conditions. It
is synthesized in the epithelial cells of the main and accessory lacrimal
glands. It destroys mucopolysaccharides in the bacterial wall.
Lactoferrin: provides a boost to immunologic system
• IgA: prevents bacterial adhesion to epithelial cells
• Beta-lysine
Mucin layer
• It offers viscosity to the tear (soluble mucin mixed within the aqueous layer, MUC
5AC).
• It converts the corneal surface into an hydrophilic surface and prevents tear rupture by
coating precipitated lipids (transmembrane mucin, glycocalyx, MUC 1, 4, 16).
‒ Pathological alterations
CL wear (it may decrease osmolarity at the post-lens tear film but increase it at
the pre-lens tear film): related to evaporation.
Tear distribution
• Evaporation
– 36 items.
– Some questions are specific for contact lens related dry eye
• Normality values:
– It is necessary to check the continuity and regularity of the meniscus along the
whole the eyelid, especially in the temporal zone.
Schirmer test
– Schirmer II and III: with anesthesia and external stimulation of tear secretion
Osmolarity evaluation
• Dry eye: Over or equal to 308 mOsm/l or a difference of more than 8 mOsm/l between
eyes
• Mires of the Helmholtz keratometer or Placido disc: time until the appearance of
distortions in the image following a blink.
• Tearscope
• Based on the observation of interference patterns on the lipid layer of the tear film
(estimation of thickness of the lipid layer).
Ocular Thermography
DRY EYE
• Loss of homeostasis of the tear film: loss of equilibrium of the various components,
structures, etc.
• Ocular symptoms (e.g., OSDI ≥ 13)
• Tear film instability (e.g., reduced BUT < 10 s)
• Hyperosmolarity (≥ 308 mOsm/l)
• Ocular surface inflammation and damage
• Neuro-sensory abnormalities (reduced corneal sensitivity, disruption of nerve
terminals, etc.)
• Sjögren syndrome:
Inflammatory, autoimmune alteration of all mucous production glands, leading
to dryness of eyes, mouth...
• Normal reduction in tear production with age (ARDE, age-related dry eye):
Reduction of 50% at 50 years
Reduction of 75% at 80 years
• Ionic surface attracts water in wearers with good tear film, but attracts
proteins (mainly lysozyme) in dry eye
• Blinking
– Water content:
• All contact lenses restore their water content from their environment,
i.e., tear film
– Caused by migration of tear film towards the horizontal meniscus under the
CL, leaving areas of corneal exposure and dryness
• Patel and Murphy (1994): Air non-contact esthesiometer of area 0.8mm2. Patients
refer “cold” sensation rather than “touch” sensation.
• Long posterior ciliary nerves from the ophthalmic branch of the trigeminal nerve
• These nerves enter the cornea at the anterior half of the stroma and ascend towards
the surface of the epithelium. They “normally” do not have myelin (from 1.5 mm of
limbus).
• In the epithelium the nerve axons loose their Schwann cells (corneal epithelial cells
function as Schwann cells, so changes in epithelial cells, e.g., CL wear, refractive
surgery, affect nerve axons).
• Mechano-nociceptors: Δδ fibers that describe the intensity of the stimuli but not their
duration.
• Polymodal nociceptors: C fibers that describe intensity and duration of stimuli.
Conduction velocity higher than Δδ.
• Cold: Δδ and C.
• There are about 7000 nociceptors (pain receptors) per mm2
Corneal esthesiometry
• Apply nylon filament perpendicular to the corneal surface and increase force until
flexure of filament or patient response.
• Start with maximum length at 60 mm and progressively shorten the filament until
patient response (objective: reflex blinking or subjective: e.g., hand raised).
• When filament is shortened, by applying the same force, more pressure is transferred
to the ocular surface, as filament flexure resistance is increased.
• For each length of filament, there is a corresponding pressure value at the ocular
surface, but normally results are written as length of filament at first patient response.
• It is important to always use the same diameter of filament. Thicker filaments will have
higher flexure resistance.
Physiological variations
• Location: maximum sensitivity at the center of the cornea and reduction towards the
periphery, being lower at the top.
• Overlap of the reception fields of the nerve terminals (greater density at the center of
the cornea).
• Neuronal adaptation to the mechanical touch of the eyelid on the upper part of the
cornea.
• Diurnal variations: lower in the morning (IOP is also higher in the morning).
• Lower corneal esthesiometry with RCL (rigid corneal lenses) than soft lenses
Greater decline with years of RCL use, and slower recovery
• Important for orthokeratology (reduction of up to 50% of corneal sensitivity) and for
refractive surgery (time without CLs prior to surgery to allow at least partial corneal
recovery)
• Reduced sensitivity during the first 3-6 months after the intervention. Abnormal
response to pain: allodynia and hyperalgesia