You are on page 1of 1

FAR-2xx7

Shut Down checks

1) Check Supply

Remove this cover to expose power


connections. Test using a
multimeter to confirm supply
voltage is correct.

2) Disconnect HV PCB

Power system down / isolate from ships supply.

To open top cover of processor unit, remove 4 x screws and slide cover off.
Unit should then open like a book.

1 Disconnect cable No.1. Power system on.

Does system boot to standby?

• Yes, system now powers on. Please run


self-test and send us results.
• No, system still does not power on.
Please power system down / isolate from
2 power. Disconnect cable No.2. Power
system on. Does system now boot to
standby?