Professional Documents
Culture Documents
2 Requirements 3
4 Getting Started 5
4.1 Importing the ICH Templates 5
4.2 Description of the ICH LOD LOQ (SN) eWorkflow template 6
4.2.1 eWorkflow General 6
4.2.2 Sequence General 6
4.2.3 Sequence Layout 8
4.2.4 Resulting Sequence 9
4.3 Customizing the ICH LOD LOQ (SN) eWorkflow template 9
4.3.1 eWorkflow General 11
4.3.2 Sequence General 11
4.3.3 Sequence Layout 11
Note: Indicates general information that points out specific items of interest.
• Processing Methods
• Report Template
• Custom Variables
• Example Sequence
1.3 ICH
The International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human
Use (ICH) provides recommendations on the validation of analytical procedures (Q2(R1)). Recommended tests include -
• Accuracy
• Detection Limit
• Linearity
• Precision
• Quantitation Limit
• Range
• Robustness
• Specificity
The calculation and reporting of all data requested by ICH can be very tedious and time consuming. The ICH method
validation templates provided in the Chromeleon 7 Extension Pack automate these steps and allow the quick and easy
creation of method validation reports.
Parameter Description
TEST Identifies the ICH test. This is an injection custom variable that can be entered in the injection list.
Identifies the last injection of a sequence for a particular ICH test. This is an injection custom variable that can
Test_End
be entered in the injection list.
User definable option to specify if the noise measurement is taken from the blank or the actual chromatogram.
Measure_Noise
This is a component custom variable that can be entered in the component table in the processing method.
User defined ratio to be used for the LOD calculation. This is a component custom variable that can be entered
LOD_Ratio
in the component table in the processing method.
User defined ratio to be used for the LOQ calculation. This is a component custom variable that can be entered
LOQ_Ratio
in the component table in the processing method.
User defined start time for the noise measurement of the injection. This is a component custom variable that
Noise_Start
can be entered in the component table in the processing method.
User defined end time for the noise measurement of the injection. This is a component custom variable that can
Noise_End
be entered in the component table in the processing method.
Test Specifies if the test will be run for the corresponding component or not.
Table 1: ICH Parameters in Chromeleon
2 Requirements
The ICH templates can only be used with Chromeleon 7.1 SR1 or higher.
It is possible to export the ICH results in the Adobe® Acrobat® PDF format. Acrobat Reader version 5.0 or higher is needed
to read these reports.
3 Known Issues and Limitations
3.1 Limitations
• The template Processing Method file can handle 5 calibration standard levels by default, but additional levels can
be added.
4. Select in which Data Vault the eWorkflow must be imported and click Start (Figure 1).
This eWorkflow contains all files for performing Limit of Detection and Limit of Quantitation
eWorkflow Description calculations based on the signal to noise ratio of the analyte peak. Please refer to the
attached user manual for information.
No instruments are associated with the ICH LOD LOQ (SN) Template eWorkflow. This
Instruments
eWorkflow can therefore be used for all instruments.
Processing Method ICH LOD LOQ (SN), containing the required custom variables for LOD LOQ
(Blank SD) calculations.
Report Template Method_Validation, containing all required calculations for LOD LOQ (Signal
Methods
to Noise).
View Settings ICH LOD LOQ to display chromatogram information in the Data Processing
Category.
Chromeleon backup file ICH LOD LOQ (SN) - Demo Data.cmbx, containing an ICH LOD LOQ
Attachments (SN) example
PDF document ICH LOD LOQ (SN) - User Manual.pdf
Table 2: Default eWorkflow General settings
Section 4.3 describes how these settings can be customized.
1
Default Report Template Method_Validation
2
Default View Setting ICH LOD LOQ
3 Defines the channel to be displayed as the default channel when opening an injection.
Default Channel
As this depends on the instrument and corresponding instrument method this is empty.
‘ICH LOD LOQ (SN)’ followed by current data/time (Example: ICH LOD LOQ (SN)
Sequence Name
20110511 161743)
ICH/LOD_LOQ is the directory that the resulting sequence is stored in. If the folder in
Path this directory path does not exist, it will be automatically created when the first sequence
is created.
All signature steps to submit electronic signatures are checked, to enable a qualified
Electronic Signature
user to electronically sign the sequence and submit, review or approve it.
Comment This sequence is created using the ICH LOD LOQ (SN) eWorkflow.
By default the sequence is setup to make 3 LOD LOQ (SN) blank injections and 2 calibration
1 standard injections at 2 concentration levels (total of 5 injections). However, it is not mandatory to
Sequence Header
follow this approach. You can add or remove injections to meet your requirements. Injections can
also be set as multiple injections from the same vial in the same autosampler vial position.
2
Sample Block <Empty>
3
Bracket <Empty>
4
Sequence Footer <Empty>
No. of inj. 1
Position <uneditable>
Volume 1µl
Status <uneditable>
Weight 1.0000
Dilution 1.0000
IntStd 1.0000
ID <empty>
01, 02, 03, 04, 05 (depending on level of calibration standard i.e. 01 for first calibration standard and
Level
02, 03 and 04 for other calibration standards)
Test_End <empty> except for last injection in list, which has LOD/LOQ (S/N) (End)
Comment <empty>
Table 5: Default sequence column settings
2. In the navigation pane select the ICH LOD LOQ (SN) eWorkflow template (Figure 7A).
Figure 7: Start editing the ICH LOD LOQ (SN) eWorkflow templates
3. In the work area click Edit (Figure 7B). The eWorkflow editor now opens.
4. To save the eWorkflow with a new name click File, Save As.
5. In the Save eWorkflow dialog select the Data Vault (Figure 8A) to store the eWorkflow and enter a name (Figure 8B)
for the new eWorkflow.
Tip: For an extensive explanation of modifying eWorkflows and defining the sequence layout, refer to the Chromeleon 7
online help.
4.3.1 eWorkflow General
While customizing the eWorkflow the state can be set to In Development. This way it is clear to
eWorkflow State users the eWorkflow is still being worked on. When the customization is finalized the state can
be set back to Approved for Use.
Add the instrument(s) that can run this ICH analysis. The eWorkflow can only be launched on
listed instrument(s).
Instruments
When no instruments are added, the eWorkflow can be used with all instruments accessible to
the user.
The current manual can be removed or other documents can be attached, for example a
Attachments
relevant SOP about running the ICH LOD LOQ (SN) application.
When instrument method(s) are added to eWorkflow General tab a list of corresponding
Default Channel
channels is available. Select a default channel.
If the default sequence name is not suitable, define a sequence name. Note that variables can
Sequence Name
be used for dynamic sequence names.
If the default ChromeleonLocal is not suitable, select the Data Vault where the sequences
Data Vault
should be stored.
If the default ICH/LOD_LOQ path is not suitable, select the folder location where the sequences
Path
should be stored. Note that variables can be used for dynamic folder names.
If electronic signatures are not used, uncheck the Submit electronic signatures option.
Electronic Signature
If electronic signatures are used, define the appropriate signature steps.
No. of inj. Enter the number of injections that should be injected from the same vial.
Position This is a non editable field, which will be filled when creating the sequence.
Select the correct instrument method. If the eWorkflow can be used on multiple instruments
Instrument Method
with different instrument methods, it’s better to leave this field empty.
When the processing method is updated and has a new name, select the correct processing
Processing Method
method.
Status <uneditable>
Weight 1.0000
Dilution 1.0000
IntStd 1.0000
01, 02, 03, 04, 05 (depending on level of calibration standard i.e. 01 for first calibration
Level
standard and 02, 03 and 04 for other calibration standards)
Test_End Ensure the last injection has the entry LOD/LOQ (S/N) (End)
This section provides a step-by-step guide for using the ICH LOD LOQ (SN) template.
2. Select the relevant eWorkflow (Figure 9A) and the instrument (Figure 9B) to use.
• Define the test level of each calibration standard by using the level field e.g. calibration standards at level 1 must have
01, etc.
• There is a column called “TEST” available. This is used to tell Chromeleon what test an injection is assigned to.
Injections must be assigned the test “LOD/LOQ (S/N)”.
• Finally, there is a column called “Test_End”. This is used to inform Chromeleon when the last injection for a specific
test has been made. You need to select the relevant value for the last injection i.e. for Standard 2 you need to use
“LOD/LOQ (S/N) (End)” (Figure 12).
Figure 12: Setting Custom Variables
Note: All information in the sequence table can be updated after the sequence has run.
Note: If you do not want to use the data from a specific test, ensure that the sample type is set to “Unknown” and that the
injection is not assigned a “TEST” value.
Note: When the option Run Sequence after creation is used in the eWorkflow wizard (see section 5.1.1, Launching the
ICH LOD LOQ (SN) eWorkflow, page 14), the sequence is already running.
Note: The processing method can have different layouts, consisting of different tabs and names. Therefore the
Processing Method tables and columns you see may look different from the ones in these pictures.
Tip: Choose the processing method layout appropriate for your application, or define your own layout by selecting
individual tabs via the Processing Method layout dialog ( ).
Eval. Type Allows you to choose the Y-axis for the linearity test.
Test Specifies if the test will be run for this component or not.
Defines if the noise will be measured in the current chromatogram (Current) or in the blank
Measure_Noise
chromatograms (Blank).
Noise_Start The start time for the noise calculation. Noise is measured in the blank injections.
Noise_End The end time for the noise calculation. Noise is measured in the blank injections.
Table 10: ICH LOD LOQ (SN) specific custom variables in the component table
3. Show the Chromatogram Pane and the Processing Method Pane using the corresponding tools in the Data
Processing Home Ribbon tab (Figure 18).
Figure 18: Panes Group of the Data Processing Home Ribbon tab
In Component Table:
4. Ensure that your component is correctly identified by updating the retention time and retention time window.
5. Decide if you want to test the LOD LOQ (Signal to Noise) of each component or not by updating the “Test” field.
6. Define if the noise will be measured in the current chromatogram (Current) or in the blank chromatograms (Blank).
7. Define ratio for calculation LOD. Recommended value is 3.
9. Define if the start of the period you wish to determine noise for each component.
10. Define if the end of the period you wish to determine noise for each component.
Note: When measuring the noise in the blank, it is recommended to choose a section of the chromatogram that
corresponds to where the peak is expected to appear. For example, the EP recommends measuring a range of 20 times
the peak width at half height around the expected retention time of the peak.
When measuring the noise in the current chromatogram, it is recommended to choose a section of the chromatogram
that contains no peaks.
Note: When only a few of several samples need reporting, select those samples, right click, and select Print or Export.
Note: The automatic reporting options are configured to print the ICH LOD LOQ (S/N) test sheet. Please refer to section
7 for instructions how to print the results of all ICH tests together.
This ensures that every file created has a unique name. To export the report follow the steps below:
2. Right click and select Export (Figure 21B). The Export dialog now opens (Figure 23).
Figure 23: Export dialog when exporting the ICH report
3. In the Report Template & Channel section, select the correct channel to export the data (Figure 23A).
Note: The automatic reporting options are configured to export the LOD LOQ (S/N) sheet for injections with TEST
custom variable.
6.1.1 Limitations
None.
6.2.1 Limitations
None.
3. Locate ‘Collate_Data.cmbx’ file on the Chromeleon 7 disc, in the folder \Extension Pack\ICH Templates. Click Open.
4. Select the correct Data Vault and directory that contains the method validation data.
2. In the Sequence search replace [ENTER_PATH] with the folder directory that contains the method validation (Figure
24A).
2. To print the report, select Print, or expand and select Report on the Query tool bar (Figure 26A). This will open
the Print dialog.
7. This will open the Export dialog (Figure 29). Ensure the option Apply to ‘Results of current injection query’ is
selected (Figure 29A).
Figure 31: Query Export dialog