Professional Documents
Culture Documents
42 Reproduction (photocopying) of editorial content of this journal is prohibited without publisher's permission.
Journal of Chromatographic Science, Vol. 27, January 1989
Chlorine trapping were recorded. With these data, the moles of air sampled were
A suitable trapping agent for chlorine in air or water samples calculated, assuming the air to be an ideal gas saturated with
–
must convert all of the chlorine to Cl for quantitative analysis water vapor.
by I C . The agent must neither produce nor contain ions that When not sampling air, the bottle portion of the impinger
have long elution times nor produce peaks that interfere with was removed, and a tygon tube connected the impinger nozzle
– –
the Cl peak. Although KI converts all of the chlorine to C l , to a caustic bubbler. This allowed a continuous flow of air
–
it was rejected because the excess I had a peak elution time through the system to eliminate wall effects on the chlorine.
of 18 min. Sodium thiosulfate was rejected also because the ex Prior to sampling, the air flow was momentarily stopped, the
2 –
cess S O had a peak elution time of 21 min. Sulfamic acid
2 3 tygon tube was removed from the nozzle, and the bottle portion
was unsuccessful because its elution time interfered with the of the impinger containing 25 m L of trapping solution was
–
Cl peak. Sulfur dioxide was not tried because of the difficulty installed.
in using a gaseous reagent. Sodium hydroxide was rejected be Water was sampled for chlorine in one of two ways. More
–
cause it produces more than Cl from its reaction with chlorine. concentrated solutions were sampled by placing approximately
Sodium sulfite worked well as a trapping agent. Sulfite rapidly 5 m L of trapping solution, typically 5 to 15 drops of 1 Μ N a S O 2 3
–
reduced chlorine quantitatively to Cl under either acidic or trapping agent in 5 m L of distilled water, in a 100–mL volumetric
basic conditions. The mechanism is probably p H dependent. flask. Five to ten grams of water sample were added by weight
In this work, the p H varied from very acidic conditions when and the flask was sealed. The used trapping solution was then
trapping chlorine in water samples to basic conditions for air diluted to 100 m L for later analysis by I C . In weaker solutions,
samples bubbled through sulfite solutions. Standard oxidation- usually the case in scrubbing operations, chlorine was trapped
reduction potentials indicated the likelihood that sulfite, besides by simply adding a few drops of 1 Μ N a S O trapping agent
2 3
–
reducing C l to C l , could reduce either H O C l under acidic
2 to a minimum 200–mL sample.
–
conditions or O C 1 " under basic conditions to Cl (13,14). In
addition to discussions of a general nature on chlorine and sulfite Analysis by IC
chemistry (15-17), other evidence indicating the ability of sulfite –
Trapped samples were analyzed for Cl on a Model 2000
to reduce chlorine through its hydrolysis products in acidic (18) i / S P ion chromatograph (Dionex) with a Model SP4200 com
and basic (19) environments has been reported, although the puting integrator (Spectra-Physics) to record elution times and
mechanisms were not certain. The present work was consistent response areas. The column was an H P I C - A S 4 A anion separa
with these findings. tor (Dionex) with an H P I C - A G 4 A guard column (Dionex).
–
Furthermore, the Cl produced by trapping with sodium Eluent (0.00075 Μ N a H C O , and 0.0022 Μ N a C O ) was 2 3
sulfite was easily measured by I C , and all ions had peak elution pumped through the column at 2 m L / m i n . The regenerant for
times of less than 6 min. The produced HCl was very stable. the ASF–1 suppressor cell (Dionex) was 0.025 Μ H S O at 2 2 4
Although sodium bisulfite was not tried as a trapping agent, m L / m i n . The conductivity cell was operated at 100ΜSfor a 1–V
it would probably work similarly to N a S O . A l l chlorine trap
2 3 signal to the computing integrator. The integrator attenuation
ping solutions were freshly prepared from 1 Μ N a S O trapping 2 3 was 1024 m V for full scale deflection, and the peak threshold
agent to provide excess reagent for the stoichiometric conver value was 2000. A stable base line was obtained within 1 h of
sion of chlorine to H C l . startup of the equipment. The sample size was 50 μL injected
by glass syringe.
Sampling The N a S O reagent used in preparing trapping solutions con-
2 3
43
Journal of Chromatographic Science, Vol. 27, January 1989
tained an unknown impurity that appeared on chromatograms Gas samples were collected on a volume basis, and the chlorine
–
at the F elution time. This counteracted the water dip, which mole fraction was calculated by converting the trapped mass
–
had about the same elution time. A sample strength of about of Cl into equivalent moles of chlorine and dividing by the
–
0.003 Μ N a S O resulted in an impurity peak size that effec
2 3
moles of gas sampled. The trapped mass of Cl was the pro
–
tively nullified the water dip and gave a straight base line from duct of the measured Cl concentration of the diluted trapped
–
injection to the Cl peak. A l l samples, including standards, sample multiplied by its volume.
were dosed as needed with N a S O to a concentration of 0.003
2 3
For water samples collected on a weight basis, the trapped
–
Μ to eliminate the water dip and give consistent base lines on mass of Cl was determined as it was for air samples. This
–
all chromatograms. trapped mass of Cl was then divided by the sample volume
Ten HCl standards were prepared over the concentration taking the density of the solution as 1 g / m L to give chlorine
range 18.23 to 1.823 m g / L . This range was selected because, concentration. When the chlorine in water samples was trapped
–
on an equivalent Cl basis, it matched Cl concentrations ex –
by adding a few drops of 1 Μ N a S O to a minimum 200–mL 2 3
–
pected in trapped samples. Seven successive analyses of an in sample, the chlorine concentration was equal to the Cl con
dividual standard gave area responses with a coefficient of varia centration corrected for dilution.
–
tion (RSD) of 0.4%. This value was well within the chromato The column gave excellent separation of Cl from other
graph manufacturer's operational limit of an R S D value of 3% anions. Chromatograms of trapped chlorine solutions clearly
– 2 – 2 – 2 –
(20). Table I gives the instrument calibration data. Only two showed C l , excess S O , and produced S O . The S O
3 4 3
2 –
measurements of each standard were made because all R S D to S O conversion could not be correlated quantitatively
4
2 – 2 –
values of standards were within 3 % . Although area response because S O slowly forms S O
3 in water in the absence of
4
to standards could vary up to 2% over an 8–h period, samples chlorine. Analysis of trapped samples repeated three weeks after
and standards were usually analyzed within 10 min of each an initial analysis gave identical results showing at least a three-
other. A typical chromatogram is shown in Figure 2. week shelf life for trapped samples.
The standard concentration vs. area curve for the calibration
data in Table I deviates slightly from linearity at the low end Testing the method
of the concentration range. This was possibly the result of a The I C method for analyzing chlorine in gas samples was
large peak threshold value eliminating a larger percentage of tested by using it to measure the composition of a certified
a smaller peak. C l / N gas mixture of 530 ppm ±2% (cylinder pressure was
2 2
(integirator counts χ 10 ) –6
HClConcn Standard
(mg/L) Sample Mean deviation
44
Journal of Chromatographic Science, Vol. 27, January 1989
between the method of this work and a standard iodometric Results and Discussion
method (wet test) (21) by using both procedures to measure the
concentration of chlorine/water solutions. The iodometric The I C method of analysis for C l in air was tested on the
2
method was chosen over others because it is a well-established certified C l / N mixture of 530 ppm ±2%. Analyses for five
2 2
procedure for determining chlorine over a wide range of con samples taken directly from the cylinder spanning five days gave
centrations and it requires only standard laboratory equipment an average composition of 538 ppm C l with a standard devia
2
The comparison of the water analysis method with the wet Agreement between the wet test and the I C method for the
test was made at four different chlorine/water concentrations. analysis of chlorine in distilled water was also very good, as
Test chlorine/water solutions were prepared by bubbling both shown in Table II. Trials 1, 2, and 3, representing measurements
pure chlorine from a standard 150-pound cylinder and the cer made on three different days, showed a percent difference be
tified C l / N mixture mentioned above through distilled water
2 2 tween the two methods of less than 1%. In addition, standard
in an equilibrium cell at 20°C and at 30°C. In each case, water deviations of concentrations determined by wet test and I C for
samples were taken at equilibrium. Preliminary experiments in the same trials were, respectively, 0.6 and 1.4 m g / L . O f the re
dicated that bubbling pure chlorine through water at 600 m L / m i n maining three trials for different operating conditions, only Trial
for approximately 45 min would produce a saturated solution; 4 showed a larger difference between the methods than was ex
however, to insure equilibrium in the cell, pure chlorine was pected. The difference found for Trial 4, more than 4%, was
bubbled for two hours before sampling. Equilibrium was con probably because the equilibrium cell had not been adequately
firmed when hourly I C analysis of water samples from the cell, flushed between Trials 3 and 4, which were made sequentially
made after two hours of bubbling, showed no change in com on the same day. A s a result, the distilled water charged to the
position. When bubbling with the certified C l / N mixture, the
2 2 equilibrium cell for Trial 4 probably became contaminated with
–
liquid was assumed to be saturated with chlorine when hourly a small amount of Cl remaining in the cell from the previous
analysis of the exit gas of the cell by our I C method was un trial. This would cause the Trial 4 analysis by IC to be larger
changed. For this case, analysis of the exit gas showed no change than that by the wet test. The chlorine concentration range
in composition after bubbling for approximately 4 h at 600 covered for these samples was 160 to 7000 m g / L .
m L / m i n . In either case, once equilibrium between gas and liquid As part of the comparison of the water analysis method with
had been reached, water samples were taken in duplicate for the wet test, the exit gas of the equilibrium cell was analyzed
analysis, one by I C and the other by wet test. by the I C method to determine equilibrium when bubbling with
The wet test for chlorine is based on the formation of iodine the certified C l / N mixture. This was approximately one year
2 2
from the reaction of iodide ion with chlorine: after the original analysis of the certified C l / N mixture, and
2 2
cylinder pressure at this time was about 1400 psig. Eight analyses
Eq. 6
were made of the exit gas (i.e., the certified gas mixture after
The iodine is then reduced by sodium thiosulfate with starch having been bubbled through the cell for equilibrium conditions
as an indicator: at 20°C and 30°C). One analysis of a sample taken directly from
the supply cylinder of the certified gas mixture also was made.
Eq. 7 These nine analyses spanning three days gave an average com
position of 512 ppm chlorine with a standard deviation of 3
The C l concentration is given by:
2
ppm C l . The analysis of the sample taken directly from the
2
1. Ten milliliters of 2% by weight K I , 2 m L of glacial acetic tures stored in steel cylinders are known to change concentra
acid, and approximately 1 m L of a standard 1% starch tion by adsorption or desorption of chlorine as cylinder pressure
solution were added to a 100-mL volumetric flask which changes. The 512 ppm C l concentration is included as operating
2
45
Journal of Chromatographic Science, Vol. 27, January 1989
Table II. Comparison of IC and Wet Test Analysis of 2. J.L. Morrison. Recovery scrubber for waste chlorine gas. Tappi
Chlorine/Water Samples from Equilibrium Cell 51(12): 124A–25A (1968).
Cell Cl con centratioms 3. G.Y. Pan, J.F. DeGraw, and J.J. Renard. Tappi Proceedings, 1982
2
4. I.F. Rees. Bleach plant scrubber for chlorine and chlorine dioxide.
temperature
Trial Gas a
Wet test IC Difference 6 Appita 34(6): 471–74 (1981).
(°C)
5. A laboratory investigation of an iodometric method for determining
1 30 pure 5467 5480 0.24 chlorine and chlorine dioxide in pulp and paper industry workplace
2 30 pure 5390 5407 0.32 atmospheres. NCASI Technical Bulletin 409. September 1983.
30 pure 5481 5429 –0.95 6. G.C. White. Handbook of Chlorination, 2nd ed. Van Nostrand
3
4 30 512 ppm 159 166 4.40 Reinhold, New York, 1986, pp. 214–55.
5 20 pure 7072 6904 –2.38 7. G.C. White. Handbook of Chlorination, 2nd ed. Van Nostrand
CD
1. L.M. Bungey. Bleach plant atmosphere improved at CFP by scrub Manuscript received December 29, 1987;
bing off-gases. Pulp & Paper 55(7): 106–11 (1981). revision received July 25, 1988.
46