Professional Documents
Culture Documents
Objective To investigate the effect of different pulse oximetry (SpO2) target range settings during automated frac-
tion of inspired oxygen control (A-FiO2) on time spent within a clinically set SpO2 alarm range in oxygen-
dependent infants on noninvasive respiratory support.
Study design Forty-one preterm infants (gestational age [median] 26 weeks, age [median] 21 days) on FiO2 >0.21
receiving noninvasive respiratory support were subjected to A-FiO2 using 3 SpO2 target ranges (86%-94%, 88%-
92%, or 89%-91%) in random order for 24 hours each. Before switching to the next target range, SpO2 was manu-
ally controlled for 24 hours (washout period). The primary outcome was the time spent within the clinically set alarm
limits of 86%-94%.
Results The percent time within the 86%-94% SpO2 alarm range was similar for all 3 A-FiO2 target ranges (74%).
Time spent in hyperoxemia was not significantly different between target ranges. However, the time spent in severe
hypoxemia (SpO2 <80%) was significantly reduced during the narrowed target ranges of A-FiO2 (88%-92%; 1.9%,
89%-91%; 1.7%) compared with the wide target range (86%-94%; 3.4%, P < .001). There were no differences between
the 88%-92% and 89-91% target range.
Conclusions Narrowing the target range of A-FiO2 to the desired median ±2% is effective in reducing the time
spent in hypoxemia, without increasing the risk of hyperoxemia. (J Pediatr 2018;■■:■■-■■).
Trial registration www.trialregister.nl: NTR4368.
P
reterm infants often require supplemental oxygen to prevent hypoxemia, a condition that has been associated with organ
damage and an increased mortality.1-3 However, too much oxygen supplementation resulting in hyperoxemia may lead
to systemic oxidative stress and long-term complications such as retinopathy of prematurity.4,5 For these reasons, oxygen
saturation is continuously measured in preterm infants with pulse oximetry (SpO2) aiming to keep it within a safe range by
manually titrating the fraction of inspired oxygen (FiO2). However, studies in preterm infants have shown that SpO2 targeting
is a clinical challenge with patients spending only 50% of the time within their intended SpO2 range because of clinical insta-
bility and the limited time nurses have to adjust the amount of oxygen.6,7 To improve SpO2 targeting, systems for automated
FiO2 (A-FiO2) control have been developed.
These closed-loop A-FiO2 systems are increasingly incorporated into standard neonatal ventilators and consist of an oxy-
genation monitoring device (pulse oximeter), gas (air/oxygen mixing) delivery device, and an algorithm that determines the
timing and magnitude of the FiO2 adjustments. Previous studies have shown that these algorithms improve maintenance of
oxygenation within an SpO2 target range compared with routine or dedicated manual FiO2 (M-FiO2) control.8-12 However, in
these studies, the SpO2 target set in the A-FiO2 arm was relatively wide and the SpO2 alarm setting was identical to the setting
used during normal clinical care. This approach may not use the full potential of the closed-loop A-FiO2 systems. The aggres-
siveness of adjustments is proportional to the deviation of the measured SpO2 from the set SpO2 limits. Therefore, narrowing
the SpO2 target range during A-FiO2 might increase the time spent within the SpO2 alarm limits even further and make the
algorithm more effective.
This study aimed to investigate the effect of different SpO2 target range set-
tings during A-FiO2 on time spent within a (wider) clinically desired SpO2 range
in oxygen-dependent infants on noninvasive respiratory support. We hypoth-
esized that narrowing the upper and lower limits of the A-FiO2 target range would From the 1Department of Neonatology, Emma Children’s
Hospital, Academic Medical Center Amsterdam;
result in an increased proportion of time spent within the (wider) clinically set 2
Department of Pediatrics, Division of Neonatology,
Leiden University Medical Center, Leiden, The
SpO2 alarm limits and reduced episodes of both hypoxemia and hyperoxemia. Netherlands; and 3Faculty of Biomedical Engineering,
Czech Technical University in Prague, Kladno, Czech
Republic
T.B. is an independent clinical research consultant and
provides services to the manufacturer of an automated
FiO2 system (CareFusion). A.v.K. received travel support
A-FiO2 Automated FiO2 and a speaker fee from CareFusion. The other authors
declare no conflicts of interest.
FiO2 Fraction of inspired oxygen
M-FiO2 Manual FiO2 0022-3476/$ - see front matter. © 2018 Elsevier Inc. All rights
SpO2 Pulse oximetry reserved.
https://doi.org10.1016/j.jpeds.2018.01.077
Table I. Patient characteristics at time of study Table III. Time and episodes of normoxemia, hypox-
Variables N = 41 emia, and hyperoxemia in SpO2 target subgroups*
Gestational age — wk 26 (25-27)* A-FiO2 M-FiO2
Birth weight — g 835 (692-978)* 86%-94% 86%-94%
Male sex — no. (%) 20 (49) % time in SpO2 range 74 (67-81)† 57 (50-67)
Amsterdam — no. (%) 31 (76) % time >SpO2 range 12 (5-23)† 30 (23-40)
Age at data collection — d 21 (12-29)* % time SpO2 >98% 0.4 (0.1-2)† 3.7 (2-9)
nCPAP/NIPPV — no. (%) 34 (83)/7 (17) % time <SpO2 range 14 (9-18) 14 (10-16)
FiO2 0.28 (0.25-0.31)* % time SpO2 <80% 3.4 (1.6-4.9)‡ 3.9 (2.5-5.9)
Episodes 1 min SpO2 >98%, (n/24 h) 1 (0-11)† 44 (29-86)
nCPAP, nasal continuous positive airway pressure; NIPPV, nasal intermittent positive pressure
ventilation.
Episodes 1 min SpO2 <80%, (n/24 h) 7 (2-12)† 18 (11-25)
*Median and IQR. Median FiO2 (%) 25 (22-29)‡ 27 (24-31)
Median SpO2 (%) 91 (90-92)‡ 93 (91-93)
Manual FiO2 adjustments/24h 1 (0-3)† 81 (51-115)
*Time in range includes time when SpO2 >94 and FiO2 = 0.21. Median and IQR.
invasive mechanical ventilation and, therefore, (temporarily) †P < .001, ‡P < .05 automated vs manual.
exited the study. The reasons for clinical deterioration were nec-
rotizing enterocolitis (n = 2) and sepsis (n = 4). Four of these hyperoxemia by narrowing the target range during A-FiO2
infants restarted the study once they recovered and were control to the desired median ±1%.
transitioned back to noninvasive support. Including those 4 Hypoxemia, assessed by the proportion of time spent below
infants, the 120-hour study period was completed in 41 infants the intended SpO2 alarm range (<86%), was significantly
without any adverse events. Of these infants, 7 received nasal reduced by narrowing the target range during A-FiO2 control
intermittent positive pressure ventilation and 34 nasal con- to either 88%-92% or 89%-91% (both 10%) compared with
tinuous positive airway pressure at the start of the study. The the widest 86%-94% target range (14%). Consistent with this
patient characteristics are shown in Table I. finding, the proportion of time and the number of pro-
Protocol deviations occurred in 3 infants and consisted in
longed episodes of severe hypoxemia were significantly reduced
2 patients of a short episode of incorrect SpO2 target range
by narrowing the SpO2 target range during A-FiO2 control.
setting during A-FiO2 control. In 1 patient, the SpO2 target range There were no differences between the 88%-92% and 89%-
was set at 86%-94% twice, resulting in missing data on the 89%- 91% target ranges. The increased hypoxemia during the 86%-
91% target range. In 4 cases, technical failure of the data logger 94% SpO2 target range is also illustrated in the distribution
occurred and this was corrected by adding extra time to data of SpO2 values in the Figure, showing a clear left shift com-
collection, assuring that data were available for a 24-hour mea- pared with the narrower SpO2 target ranges. The median SpO2,
surement period. FiO2, and number of manual FiO2 adjustments did not differ
Table II summarizes the primary and secondary out- between the 3 SpO2 target range subgroups.
comes in the 3e A-FiO2 target range subgroups. The propor-
Comparison of the two 24-hour periods with M-FiO2 control
tion of time spent within the 86%-94% SpO2 alarm range was
showed no significant difference in outcomes of oxygen-
not significantly different between the 3 A-FiO2 target range
ation, and for this reason all M-FiO2 periods were averaged
subgroups. This was also true for the outcomes assessing time for further analyses. Table III shows the oxygenation out-
and episodes above the SpO2 range (hyperoxemia), although comes for the subgroups of infants when assigned to the widest
there was a trend to more episodes of prolonged severe target range during A-FiO2 control (86%-94%) and M-FiO2
control using the same SpO2 alarm range. The time spent within
the intended SpO2 range significantly increased during A-FiO2
control (74%) compared with M-FiO2 control (57%, P < .001).
Table II. Time and episodes of normoxemia, hypox-
In addition, the proportion of time spent in severe hyperoxemia
emia, and hyperoxemia in SpO2 target subgroups*
was significantly reduced during A-FiO2 control compared with
A-FiO2 A-FiO2 A-FiO2 M-FiO2 control. This was reflective of a marked decrease in
86%-94% 88%-92% 89%-91%
the number of prolonged episodes spent in severe hyperoxemia
% time in SpO2 range 74 (67-81) 72 (64-83) 73 (58-85) during A-FiO2 control (1 [0-11]) compared with M-FiO2
% time >SpO2 range 12 (5-23) 19 (9-35) 17 (6-31)
% time SpO2 >98% 0.4 (0.1-1.7) 0.5 (0.2-3.0) 0.8 (0.2-3.5) control (44 [29-86], P < .001). The effect on hypoxemia was
% time <SpO2 range 14 (9-18)† 10 (6-13) 10 (6-13) less pronounced, showing only a significant reduction in the
% time SpO2 <80% 3.4 (1.6-4.9)† 1.9 (0.9-3.7) 1.7 (0.8-4.3) number of episodes with prolonged severe hypoxemia during
Episodes 1 min SpO2 >98%, 1 (0-11) 4 (0-13) 12 (0-36)
(n/24 h) A-FiO2. Consistent with these findings, the distribution of SpO2
Episodes 1 min SpO2 <80%, 7 (2-12)‡ 3 (1-6) 4 (1-9) values during manual control showed a right-sided shift com-
(n/24 h) pared with A-FiO2 control (Figure). The median FiO2 and SpO2
Median FiO2 (%) 25 (22-29) 26 (23-30) 27 (24-31)
Median SpO2 (%) 91 (90-92) 92 (91-93) 92 (91-93) over the 24-hour period was significant lower during A-FiO2
Manual FiO2 adjustments/24h 1 (0-3) 1 (0-3) 0 (0-3) compared with M-FiO 2 control. Manual adjustments of
*Time in range includes time when SpO2 > 94 and FiO2 = 0.21. Median and IQR.
FiO2 were infrequent during A-FiO2 control and significantly
†P < .001, ‡P < .01 compared with other target ranges. lower compared with M-FiO2 control.
Optimal Target Range of Closed-Loop Inspired Oxygen Support in Preterm Infants: A Randomized Cross-Over Study 3
Figure. SpO2 histograms. Histogram of SpO2 exposure for the target ranges of A-FiO2 control and M-FiO2 control.
One study compared a wider (87%-93%) with a narrower late to clinical outcome will have to be addressed in future ran-
(90%-93%) SpO2 target range using the same A-FiO2 system.14 domized clinical trials.
Narrowing the target range resulted in significantly less time In conclusion, this study shows that narrowing the target
in hypoxemia and more time in moderate hyperoxemia. Al- range of A-FiO2 to the desired median ±2% is effective in re-
though the finding of less hypoxemia is consistent with our ducing the time spent in hypoxemia, without increasing the
study, the increased risk of hyperoxemia is not. This might be risk of hyperoxemia. Future studies need to evaluate the impact
explained by the fact that our study used the same median SpO2 on clinical relevant outcomes. ■
across all target ranges. In the study by Wilinska et al, the
median SpO2 of the narrower range (90%-93%) was higher Submitted for publication Aug 9, 2017; last revision received Dec 31, 2017;
than the wider range (87%-93%) and this may have caused accepted Jan 27, 2018
more time in hyperoxemia.14 Reprint requests: Maria Elisabeth Nicoletta van den Heuvel MD, Department
of Neonatology (H3-113), Emma Children’s Hospital/Academic Medical
Despite differences in patient characteristics, mode of re- Center, PO Box 22660, 1100 DD, Amsterdam, The Netherlands. E-mail:
spiratory support, and SpO2 targets, most of the findings in m.e.vandenheuvel@amc.uva.nl
this study are in agreement with other studies comparing
automated with manual control and using the same
algorithm.8-10,13-15 However, the difference in hyperoxemia References
between automated and manual control, especially the number 1. Carlo WA, Finer NN, Walsh MC, Rich W, Gantz MG, Laptook AR, et al.
of prolonged episodes of hyperoxemia, was more profound Target ranges of oxygen saturation in extremely preterm infants. N Engl
J Med 2010;362:1959-69.
in the present study compared with previous studies.13,15 This
2. Schmidt B, Whyte RK, Asztalos EV, Moddemann D, Poets C, Rabi Y, et al.
was mainly due to significantly more hyperoxemic episodes Effects of targeting higher vs lower arterial oxygen saturations on death
during manual control. This difference may have been caused or disability in extremely preterm infants: a randomized clinical trial. JAMA
by the fact that, in contrast to other studies, the nurses were 2013;309:2111-20.
unaware that SpO2 recording continued during the 2 manu- 3. Stenson B, Brocklehurst P, Tarnow-Mordi W. Increased 36-week sur-
ally controlled washout days. Previous studies have shown vival with high oxygen saturation target in extremely preterm infants. N
Engl J Med 2011;364:1680-2.
that the nursing staff has a tendency to tolerate hyperoxemia 4. Askie LM, Darlow BA, Davis PG, Finer N, Stenson B, Vento M, et al. Effects
to avoid spells of hypoxemia.16 This finding also suggests of targeting lower versus higher arterial oxygen saturations on death or
that studying A-FiO2 under nonblinded conditions may un- disability in preterm infants. Cochrane Database Syst Rev 2017;(4):
derestimate the true treatment effects of A-FiO2 control on CD011190.
5. Askie LM, Henderson-Smart DJ, Irwig L, Simpson JM. Oxygen-saturation
SpO2 targeting.
targets and outcomes in extremely preterm infants. N Engl J Med
Our study has some limitations. First, preterm infants require 2003;349:959-67.
supplemental oxygen for many weeks. Data collection was 6. Hagadorn JI, Furey AM, Nghiem TH, Schmid CH, Phelps DL, Pillers DA,
evaluated during a relatively short period that may not be et al. Achieved versus intended pulse oximeter saturation in infants born
representative for this prolonged period of support. Second, less than 28 weeks’ gestation: the AVIOx study. Pediatrics 2006;118:1574-
82.
data collection (5-second data logging) with a SpO2 value
7. Hagadorn JI, Sink DW, Buus-Frank ME, Edwards EM, Morrow KA, Horbar
averaged over 8 seconds might understate the rate of epi- JD, et al. Alarm safety and oxygen saturation targets in the Vermont
sodes of hyperoxemia and hypoxemia longer than 1 minute, Oxford Network iNICQ 2015 collaborative. J Perinatol 2017;37:270-
but should not affect the differences comparing the different 6.
target ranges. Third, the infants were all supported by a non- 8. Claure N, Bancalari E, D’Ugard C, Nelin L, Stein M, Ramanathan R, et al.
Multicenter crossover study of automated control of inspired oxygen in
invasive modality and had to be clinically stable during the
ventilated preterm infants. Pediatrics 2011;127:e76-83.
entire 5 days of study. The results may differ in less stable 9. Claure N, D’Ugard C, Bancalari E. Automated adjustment of inspired
infants and/or on invasive respiratory support. Finally, this oxygen in preterm infants with frequent fluctuations in oxygenation: a
study was done with 1 algorithm incorporated in 1 neonatal pilot clinical trial. J Pediatr 2009;155:640-5, e1-2.
ventilator. This might limit extrapolation of our results to 10. Claure N, Gerhardt T, Everett R, Musante G, Herrera C, Bancalari E. Closed-
loop controlled inspired oxygen concentration for mechanically venti-
other ventilators. Our results would only be generalizable to
lated very low birth weight infants with frequent episodes of hypoxemia.
other automated oxygen control devices having a similar Pediatrics 2001;107:1120-4.
algorithm to that used in the Avea ventilator, with SpO2 11. Hallenberger A, Poets CF, Horn W, Seyfang A, Urschitz MS. Closed-
error calculated in relation to the upper and lower limit of loop automatic oxygen control (CLAC) in preterm infants: a random-
the set target range, rather than the midpoint. Future studies ized controlled trial. Pediatrics 2014;133:e379-85.
12. Plottier GK, Wheeler KI, Ali SK, Fathabadi OS, Jayakar R, Gale TJ, et al.
need to confirm this.
Clinical evaluation of a novel adaptive algorithm for automated
Reducing time spent in hypoxemia by narrowing the SpO2 control of oxygen therapy in preterm infants on non- invasive
target range during automated control may have important respiratory support. Arch Dis Child Fetal Neonatal Ed 2017;102:F37-
clinical implications. Hypoxemia is associated with an in- 43.
creased mortality in preterm infants 2 and with adverse 13. van Kaam AH, Hummler HD, Wilinska M, Swietlinski J, Lal MK, te Pas
AB, et al. Automated versus manual oxygen control with different satu-
neurodevelopmental outcome at 18 months.17 Our results also
ration targets and modes of respiratory support in preterm infants. J Pediatr
suggest that using a too narrow SpO2 target range during A-FiO2 2015;167:545-50, e1-2.
control does not improve efficacy of automated oxygen control 14. Wilinska M, Bachman T, Swietlinski J, Kostro M, Twardoch-Drozd M.
and might even have adverse effects. How these results trans- Automated FiO2-SpO2 control system in neonates requiring
Optimal Target Range of Closed-Loop Inspired Oxygen Support in Preterm Infants: A Randomized Cross-Over Study 5
respiratory support: a comparison of a standard to a narrow SpO2 control 16. Claure N, Bancalari E. Closed-loop control of inspired oxygen in pre-
range. BMC Pediatr 2014;14:130. mature infants. Semin Fetal Neonatal Med 2015;20:198-204.
15. Lal M, Tin W, Sinha S. Automated control of inspired oxygen in venti- 17. Poets CF, Roberts RS, Schmidt B, Whyte RK, Asztalos EV, Bader D, et al.
lated preterm infants: crossover physiological study. Acta Paediatr Association between intermittent hypoxemia or bradycardia and late death
2015;104:1084-9. or disability in extremely preterm infants. JAMA 2015;314:595-603.