Hematology Global Technical Support Dept.

Statement:

Thanks for choosing Mindray hematology products, and this manual is developed only
for Mindray 5-diff hematology product BC-6800,

In order to continue our best service to you and have optimum performance on
instrument result, we need to maintain the instrument regularly.

Therefore we have designed the service maintenance kit, and this manual, which will
guide you to perform complete and regular maintenance step by step accordingly.

And we hope this manual will help and satisfy filed engineer and end-user query during
maintenance procedure on BC-6800.

Yours sincerely,

Mindray Hematology Technical Support Dept.

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns
the intellectual property rights to this Mindray product and this manual. This manual may refer to
information protected by copyrights or patents and does not convey any license under the patent
rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any
infringements of patents or other rights of third parties.

Mindray intends to maintain the contents of this manual as confidential information. Disclosure
of the information in this manual in any manner whatsoever without the written permission of
Mindray is strictly forbidden.

Mindray Hematology global technical support department has the right of final interpretation of
this manual and retained without prior notice and modified manual content rights
Mindray Hematology Global Technical Support Department reserves the right to change without
prior notice.

Page1of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

Revision history

Version Revision Description Revised Reviewed


history by by
1.0 Newly added The new version of the writing Carl Shui
1.1 Adjust the / Carl Shui
layout of the
maintenance
manual and
update the
part number of
kit
2.0 1. Newly 1. new-added steps for Phoenix Mars
added the replacing piercing probe Wang Zhang
content of assembly for half year
half year maintenance on 2.2;
and one 2. new-added steps for
year replacing NRBC
maintena Diaphragm pump for half
nce; year maintenance on 2.2;
2. Newly 3. new-added steps for
added the replacing pressure plate of
part list of rotation scanner for half
maintena year maintenance on 2.2;
nce kit for 4. new-added steps for
1 year; replacing thee pressure
3. Update pump filter for one year
the part maintenance on 2.2;
list 5. new-added steps for
number replacing the tubes for
one year maintenance on
2.2;
6. one year maintenance kit
is new-added. The one
year kit includes two half
year maintenance kit and
one open vial wipe, one
pressure pump filter and 4

page2of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

kinds of tubes;
7. Part No. of NRBC
diaphragm pump is
changed. New version is
801-3100-00059-00, old
version is
TCN12-J001-006.
8. Open vial wipe is added
into one year maintenance
kit, part No. is
041-005497-00;
9. Pressure pump filter is
added into one year
maintenance kit, part No.
is 082-000210-00;
10. Four kinds of tubes are
new-added into one year
maintenance kit. Part No.
are M6G-020002---( tube
between WC1 and TC1),
M6G-020011---( tube for
PV14/15/16, the length
are different)

page3of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

Table of Contents

Part of Preventive maintenance kit………………………………………………………….3

1. Tools and Consumables…………………………………………………………………...4


1.1 Instruction……………………………………………………………………………….4
1.2 Tools……………………………………………………………………………………...4
1.3 Consumables…………………………………………………………………………..4

2. Service Plan…………………………………………………………………………………5
2.1 Maintenance Frequency……………………………………………………………….5
2.2 Maintenance Contents…………………………………………………………………5
2.3 Maintenance Check List …...……………………………………….………………...27

3. Status Check………………………………………………………………………….……28

Appendix A …………………………………………………………………………………....30

page4of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

Part List of BC-6800 Preventive Maintenance Schedule

No. Part No. Description Position in the Life Remark


analyzer time
801-3201-00043-00 Beside the 28 30000
1 Piercing probe assembly
piercing cylinder weeks times
Beside the
5 801-3201-00065-00 Sheath fluid filter LF3 diluents 6 months \
heating bath
Inside the
Replace pressure plate
9 043-001082-00 barcode scanner 6 months
of rotation scanner
assembly
801-3100-00059-00 On the right side
10 NRBC Diaphragm pump 6 months
of rear panel
041-005497-00 Under the
11 Open-vial wipe 1 year
open-vial probe
082-000210-00 Inside the
12 Pressure pump filter 1 year
THOMAS pump
M6G-020002--- 120mm,
Tube. PU tube
13 between WC1 1 year
4mmX2.5mm,transparent
and TC1
M6G-020011--- Tube. PharMed, 50mm, use for
14 1 year
1/16"ODX1/8"ID PV14
M6G-020011--- Tubing. PharMed, 40mm, use for
15 1 year
1/16"ODX1/8"ID PV15
M6G-020011--- Tubing. PharMed, 85mm, use for
16 1 year
1/16"ODX1/8"ID PV16

Remark:Calculated as: 150 samples one day, 7 days one week, 52 weeks one year

page5of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

1. Tools and Consumables

1.1 Instruction

This manual introduces preventive maintenance for BC-6800, including maintenance contents,
maintenance frequency and parts that need to be replaced regularly.

1.2 Tools

Num Name Quantity Remark


1 Cross-headed screw driver 1 pcs
2 Flat tip screw driver 1 pcs
3 M2.5 Allen wrench 1 pcs
4 diagonal cutting pliers 1 pcs
5 Nippers 1 pcs
6 Blade 1 pcs
7 Hexagon socket screw key 1 pcs

1.3 Consumables

Num Name Quantity Remark


1 Gloves 2-3 pairs
2 Paper 1 Roll
3 Middle-sized Cable Tie 8 pcs
4 Cotton Swab 1 bag
5 Cleaning tissue 3 pcs
6 Probe Cleanser 1 bottle
7 5ml Syringe 1 pcs
8 7um Standard Particle To Check Optical
System Status
9 R&D Calibrator 1 Vial To Check Optic gain
10 Middle level Quality Control 2 Vials To Check Repeatability
11 High level Quality Control 1 Vial To Check Carry-over
12 Plastic cable tie To Tie the tubes

page6of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

2. Service Plan

2.1 Maintenance Frequency

The preventive maintenance for BC-6800 contains 3-monthly, half yearly and yearly interval.
3-monthly maintenance is finished by operator
Half-yearly and yearly maintenance is required to perform by BC-6800 certificated engineer

2.2 Maintenance Contents

page7of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Detailed Steps
3-monthly Maintenance(For user)
Item Maintenance contents Steps Tools and Remark
Consumables
Clean open vial sample Clear the bloodstain at the (1) Remove diluent tube and waste tube from Nippers
probe and probe wipe bottom of probe wipe the probe wipe; Cleaning tissue
Clean exterior surface of (2) Take out probe wipe; Probe cleanser
sample probe (3) Add some drops of probe cleanser onto the
bloodstain which is on the bottom of probe
wipe and keep soaking for 3 minutes;
(4) Clean probe wipe with water;
(5) Clean exterior surface of sample probe by
cleaning tissue with probe cleanser. Leave
probe cleanser to the sample probe for
soaking. Clean it with water after 3 minutes;
(6) Reassemble the probe wipe and connect
relevant tubes;
(7) Run background test for 3~5 times to make
sure there is no water remaining at sample
probe and probe wipe;

Page1of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Clean piercing probe Clean bloodstain on the (1) Clean the bloodstain with probe cleanser and Cleaning tissue Caution:
and probe wipe top of probe wipe leave probe cleanser to the sample probe for Probe cleanser 1】Blood stain in the
Clean bloodstain at the soaking. Clean it with water after 3 minutes; piercing hole must be
end of piercing probe (2) Wipe the bloodstain at the end of piercing cleared. If you cannot
probe via cleaning tissue with probe cleanser. clear it, please drip
Leave probe cleanser to sample probe for some drops of probe
soaking. Clean it with water after 3 minutes;; cleanser into it and
(3) Run background test for 3~5 times to make soak until the piercing
sure there is no water remaining at piercing hole becomes clean.
probe and probe wipe.

page9of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Clean open vial waste Clear the bloodstain (1) Take out the waste tray and add some drops Probe cleanser
tray of probe cleanser into it;
(2) Soak for 3~5 minutes, clean it with water and
then put it back;

page10of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Clean waste tray for Clear the bloodstain (1) Take out the waste tray and add some drops Probe cleanser
piercing of probe cleanser into it;
(2) Soak for 3~5 minutes, clean it with water and
then put it back;

Item Maintenance contents Steps Tools and Remark


Consumables

page11of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Probe cleanser Maintain the machine with (1) Click Menu-Service-Maintenance to enter this Probe cleanser Note:
maintenance probe cleanser screen; 1 】 Probe cleanser
(2) Execute probe cleanser maintenance for flow maintenance for
cell; fluidics will take 91
(3) Execute probe cleanser maintenance for SRV; minutes, consuming
(4) Execute probe cleanser maintenance for 35ml probe cleanser
BASO channel and NRBC channel; and 1L diluent;
(5) Execute probe cleanser maintenance for HGB 2 】 Flowcell, SRV,
bath and aperture; BASO bath, NRBC
(6) Execute probe cleanser maintenance for bath must be
fluidic(This is for the whole device)s; selected when you
execute probe
cleanser
maintenance.
Others are
optional, according
to your actual
situation;

Half-yearly maintenance
Item Maintenance Steps Tools and Remark
page12of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

contents Consumables
Clean SRV and waste Clean SRV (1) Turn off the analyzer and pneumatic unit. Wait for M2.5 Allen wrench
tray Clean waste tray 3 minutes until the pressure is released, then Cleaning tissue
start to disassemble SRV; Probe cleanser
(2) Open the front cover and support it with the Syringe
supporting bar;
(3) Take out the waste tray and wash it with water;

(4) Move the probe wipe to the bottom position Caution: If probe wipe
manually until it leaves the sample probe does not leave sample
completely; probe completely,
sample probe may get
bent when taking out the
outer plate of SRV;

page13of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(5) Cut the cable tie of SRV, loosen the fixing screw
of tighten screw, anticlockwise rotate it and take
out the tightening screw;

(6) Take out SRV assembly slowly, making middle


plate leave the fixing rod of cylinder. Rotate and
take out outer plate. Note that there is diluent in
page14of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

SRV, so please underlay paper to the SRV. When


operating, please be careful not to damage the
tubing;

Caution:
1】 Don’t pull out SRV
by force or add too
much force on the
tubes of inner plate.
(7) After moving middle plate out of guide rod, rotate 2】 Reagent may leak
middle plate to reduce the attraction from inner from SRV when
plate. Slide the middle plate to take it out. Note taking out it. Please
that there is diluent inside, please underlay paper prepare some paper
to the SRV. When servicing, don’t pull out SRV or dry tissue to save
by force in case of any damage of inner plate or clean it.
tubing. 3】 When taking out the
SRV, be careful not
to bend the sample
probe.
4】 The three plates of
SRV are stuck
page15of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

together tightly,
please pull and
rotate them slowly to
reduce liquid
attraction. Then
slide relatively to
take out the plates.

(8) Use distilled water or 1:5 diluted probe


cleanser to clean the surface of inner and middle
plates. Add the diluted probe cleanser to the
surface by syringe. Then use cleaning tissue to Caution:
clear crystal. After cleaning with probe cleanser, Please use probe
use distilled water to wash it again. The cleaning cleanser to do this
for all the 3 plates are the same. cleaning but not other
kinds of cleanser.
SRV has the function
of anticorrosion, but
we still need to wash
all the cleanser off in
case of problems
caused by other main
parts.
(9) During the cleaning make sure there is not any
page16of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

dirt or dust. Be careful not to scratch the surface.


If there is dirt on the surface, it may lead blood Caution:
leakage and get incorrect result. 1】 Make sure that the
(10) After cleaning, assemble SRV and waste tray in metal rod of middle
the reverse order, then insert sample probe to plate is placed
probe wipe and move probe wipe to the top of between the two
probe. When assembling middle plate, the flat stop pieces of the
end should be faced up and please put the metal cylinder, or it may
rod between the two stop pieces of cylinder; cause error.
2】 Make sure to move
the probe wipe to
the top position after
inserting the sample
probe. If the probe
wipe stays at the
lower position, the
analyzer will give
error” Probe wipe
(11) Turn on the analyzer and pneumatic unit, and action unidentified”
perform complete initialization to make sure the after startup and test
SRV is filled with diluent. Execute background will not be allowed;
test and make sure that all the parameters are in 3】 After SRV is
the normal range. If initialization fails, please assembled back,
execute whole device priming in order to fill SRV connect and fix the
with dilluent; tubes around as
(12) Run QC test to check if the analyzer is in control. what it was before.
Make sure background, carryover, repeatability
page17of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

are normal and scatter gram is in normal


condition;

page18of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Clear crystal on syringe Clear crystal on syringe (1) Open right door and chamber assembly Cleaning tissue Caution:
panel; 1】Cleaning is only for a
(2) Check the end point of plunger for 2.5ml little crystal. If there is
syringe and 250ul syringe. If there is leakage much, you need to
from the inside, please replace the syringe replace the syringe;
assembly. If there is crystal, please use
cleaning tissue to clean it;
(3) Open the front cover and check the 100ul
syringe plunger’s end point. If there is
crystal, please remove it and clean;
(4) After the cleaning, execute several times of
test. Check the syringes, make sure there is
no leakage;

page19of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Check floater for waste Check floaters of WC1 and (1) Open right side door and chamber assembly Probe cleanser Caution:
chamber WC2 panel; Syringe 1 】 When performing
Check the floater of waste (2) Check floaters for WC1 and WC2. If the probe cleanser
tank outside of the floater is dirty, remove one tube which is on cleaning for WC1 and
machine the top cover of the chamber and use WC2, liquid surface
syringe to add probe cleanser (concentration should be higher than
more than 40%) into the chamber. Soak for floater;
10 minutes until it is clean; 2】 Before cleaning
(3) If the customer uses waste container, please waste container floater,
take out the cap assembly and check if there please turn off the
is any dirt stuck to the floater. Soak with analyzer in case
probe cleanser for 10 minutes until it is analyzer enters
cleaned; sleeping mode and
waste inside is poured
out;

page20of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Check buffer bath Check if there is any (1) Open the right door and left door; Cross-headed
crystal or liquid inside (2) Open chamber assembly panel (right side), screwdriver
check buffer bath TC1 which is beside WC1
and SCI chamber. If there is much crystal or
liquid inside, check if there is reverse
pouring;
(3) Open left panel and check TC3 which is
under vacuum release valve. If there is much
crystal or liquid inside, please find out if any
part has problem, troubleshooting;

page21of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Check air filter Check if the air filter is full (1) Open left door and gas valve assembly Cross-headed
of liquid panel; screwdriver
(2) Check air filter which is above power
supply unit. There should not be much
liquid inside. If there is a lot of liquid
inside, the air drier or the drain valve for
waste may be defective;

page22of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Check tube pincher Check tube pincher (1) Open the front panel; Cross-headed
(2) Check if the shape is changed or bent or screwdriver
cannot pinch sample tube tightly. If there is
any problem, please replace the pincher;

Item Maintenance contents Steps Tools and Remark


Consumables
Check WBC mixer Check mixer noise (1) Open the front panel; Cross-headed
Check if there is any (2) Check if there is abnormal sound or big noise screwdriver
crystal at the mixer when the mixer is working. If yes, please add Probe cleanser
some lubricating grease. Be careful not to Cleansing
drop the grease into WBC reaction bath and tissue
check the belt if it is too loose or tighten;
(3) Check if there is any crystal or impurity on
the metal panel which is above mixing unit. If
yes, clean it with cleaning tissue and probe
cleanser;

page23of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Check silicon tube of Check silicon tube of (1) Open the right door and chamber assembly Cross-headed
pinch valve pinch valve panel; screwdriver
(2) Turn off the pneumatic unit;
(3) Check the silicon tube of PV11~PV13 if there
is anything wrong at the pinching position,
such as leakage, breakage and so on. If it is
OK, please change the pinching position,
otherwise, you have to replace it.

page24of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

page25of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Check tubing for HGB Check tubing for HGB (1) Open the front cover; Cross-headed
bubble mixing bubble mixing; (2) Take out shielding cover of HGB bath; screwdriver
Check the function of HGB (3) Check the bubble mixing tube below HGB
bubble mixing and waste bath. It is connected with a T connector.
draining; Make sure the inner Teflon tube is 4mm out
of the connector in the red circle;
(4) Run sample test, make sure HGB bubble
mixing is normal and waste can be drained
smoothly;
Normal liquid level

page26of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Item Maintenance contents Steps Tools and Remark


Consumables
Check tube of reagent Check the tube of reagent (1) Check the tube at the back of the analyzer /
cap assembly cap assembly and make sure the tubes for all the lyses,
diluents are not sharply bent and no
leakage;
(2) Tidy up the tubes to make sure they will not
be pressed and bent;
(3) Check tubes for fluorescence reagent to
make sure they are well connected;

Check piercing probe Check piercing probe (1) Remove the air tubes for piercing cylinder; /
attrite status attrite status (2) Push the piercing probe to let it pass through
the probe wipe manually. Check the attrite
status of piercing probe tip. If it is seriously
attrite, please replace it.

Item Contents Steps Tools and Remark


Consumables
Replace piercing probe Replace piercing probe (1) Open the front cover and support it with the Hexagon
assembly assembly supporting bar; socket screw
(2) Loose the screws and take out the cover of key
piercing probe assembly; Cross-headed
screwdriver

page27of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(3) Loose Diba connector and the tubes


connected to the assembly and loose the
screws as the pictures show ;

page28of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(4) Replace the piercing probe assembly and


re-assemble the screws, Diba connector and
tubes;
Caution: Be careful for the sharp piercing needle
(5) Test several samples to check whether the
assembly is fixed well

Replace Sheath fluid Replace sheath fluid filter (1) Turn off the pneumatic unit; Cross-headed
filter LF3 (2) Open right side door and chamber assembly screwdriver
panel;
(3) Take out the tubes connected to the sheath
fluid filter outlets and loose the two screws as
the picture shows ;

page29of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(4) Replace the sheath fluid filter and


re-assemble the screws and tubes;
(5) Test several samples to check whether the
filter leak or not

Replace pressure plate Replace pressure plate of (1) Open the front cover and support it with the /
of rotation scanner rotation scanner supporting bar;
(2) Take out the old rotation pressing head

page30of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(3) Replace and re-assemble ;


(4) Test several samples to check whether the
rotation pressing head is fixed or not

Replace NRBC Replace NRBC (1) Turn off the pneumatic unit; Cross-headed Be caution for direction
Diaphragm pump diaphragm pump (2) Open right side door and chamber assembly screwdriver of diaphragm pump
panel; and the tube
(3) Loose the screws and take out the tubes connection, one for
connected to the diaphragm pump; liquid system, the other
for pneumatic system

page31of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(4) Replace and re-assemble the diaphragm


pump.

page32of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Yearly Maintenance (For Engineer)

Item Maintenance contents Steps Tools and Remark


Consumables
Clean Fan for whole Clean the heat (1) Take out the heat dissipation net from the /
device dissipation(for rear panel back panel of analyzer;
fan) and filtering net for (2) Clean it with water;
whole device (3) Re-assemble it after it is dry;

Pull out

Item Maintenance contents Steps Tools and Remark


Consumables

page33of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Check crystal for Check if there is crystal on (1) Open the front cover and chamber assembly Cross-headed
solenoid valve solenoid valve panel at the right side; screwdriver
(2) Check if there is any crystal on the valves. If Cleaning tissue
yes, remove it and clean relevant part;

Replace open vial wipe Replace open vial wipe (1) Move the probe wipe to the bottom position /
manually until it leaves the sample probe
completely;

(2) Take out the wipe block and replace it;


(3) Re-assemble the probe wipe and check the
position of the wipe and probe.
Replace the pressure Replace the pressure (1) Turn off the pneumatic unit; Cross-headed
pump filter pump filter (2) Open the cover of pneumatic unit; screwdriver
(3) Loose the pump filter as follow pictures showed;

page34of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(4) Replace the filter and re-assemble the


pneumatic unit
Replace the tubes Replace the tube of WC1 Turn off the pneumatic unit and open the right cover Cross-headed
and TC1; and take out the tubes as the follow pictures showed screwdriver
PV14,PV15,PV16 and replace with new one
(1) Tube between WC1 and TC1

page35of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

(2) Tube for PV14

(3) Tube for PV15

Open the front cover and support it with supporting


bar. Open the RBC sheath flow impedance bath
cover, take out the tube for PV16 and replace with
page36of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

new one

page37of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

2.3 BC-6800 Preventive Maintenance Check List


Hospital: Dept.: Operator:
Engineer in Charge: S/N: Tel.:
Frequency Item Jan Feb Mar Apr May June July Aug Sept Oct Nov Dec
Clean open vial probe & probe wipe
Clean piercing probe & probe wipe
3-Monthly Maintenance
Clean open-vial waste tray
(For User)
Clean piercing waste tray
Probe cleanser maintenance
Clean share valve and waste tray for SRV
Clean crystal in syringe(Just in case)
Check waste chamber floater
Check buffer bath
Check air filter
Check tube pincher for auto loading mode
Half-Yearly Maintenance Check mixer of WBC bath
(For Engineer) Maintain the tube for pinch valve
Check the tube for HGB bubble mixing
Check reagent cap assembly
Check piercing probe life status to replace
Replace the Filer LF3
Replace pressure plate of rotation scanner
Replace DP4 Diaphragm pump
Yearly Maintenance Clean the Fan for whole device
(For Engineer) Clear some crystal of solenoid valves

page38of32
血球产品全球技术支持部
Hematology Global Technical Support Dept.

Check piercing probe life status to replace


Replace the Filer LF3
Replace pressure plate of rotation scanner
Replace DP4 Diaphragm pump
Replace open vial wipe
Replace pressure pump filter
Replace tube between WC1 and TC1
Replace tube for PV 14
Replace tube for PV 15
Replace tube for PV 16

page39of32
血球产品全球技术支持部

Hematology Global Technical


Support Dept.

3. Status Check

During PM visit, please kindly check working status according to the following table :

NO. Inspection Item Description Reference Result Remark


250KPa 250±20 KPa □Pass □Fail
-70KPa NO Error Message □Pass □Fail
160KPa 160±5 KPa □Pass □Fail
Pressure/Vacuum
70KPa 70±5 KPa □Pass □Fail
40kPa 40±5 KPa □Pass □Fail
-40KPa -40±5 KPa □Pass □Fail
HGB Blank voltage HGB Blank Voltage 4.5±0.1 V □Pass □Fail
Blood sensor 2.30~2.40V □Pass □Fail
Sensor voltage setting Fluorescence reagent 2.30~2.40V □Pass □Fail
sensor

NO. Inspection Item Description Reference Value Result Remark


≤ 0.1 × 10 / L
9
WBC □Pass □Fail
≤ 0.2 × 10 / L
9
Background DIFF-WBC □Pass □Fail
≤ 0.2 × 10 / L
9
(Open vial-Whole NRBC-WBC □Pass □Fail
12
blood RBC ≤ 0.02× 10 / L □Pass □Fail
12
Autoloader-Whole RBC-o ≤ 0.02× 10 / L □Pass □Fail
blood HGB ≤1g/L □Pass □Fail
≤ 5 × 10 / L
9
Openvial-Prediluted) PLT □Pass □Fail
≤ 5 × 10 / L
9
PLT-o □Pass □Fail
Particle Quantity 2000~6000 □Pass □Fail
FS CG position 34.00~46.00 □Pass □Fail
Standard particle
FS CV ≤2.0% □Pass □Fail
(Latex Debug)
SS CG position 51.00~71.00 □Pass □Fail
SS CV ≤17.2% □Pass □Fail
Standard particle Particle Quantity 2000~6000 □Pass □Fail
(Latex Cal) FS CV ≤3.9% □Pass □Fail
Diff channel Shape and / □Pass □Fail
Fresh blood scatter
position
gram
normal

Page1of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

Baso channel Shape and / □Pass □Fail


position
normal
NRBC channel Shape and / □Pass □Fail
position
normal
Ret channel Shape and / □Pass □Fail
position
normal
WBC ≤1.0% □Pass □Fail
RBC ≤1.0% □Pass □Fail
Carry over
HGB ≤1.0% □Pass □Fail
Open vial-Whole
HCT ≤1.0% □Pass □Fail
blood
PLT ≤1.0% □Pass □Fail
(Autoloader-Whole
RBC-O ≤1.5% □Pass □Fail
blood)
WBC-D ≤1.0% □Pass □Fail
WBC-N ≤1.5% □Pass □Fail
Scatter gram Group CG Meet / □Pass □Fail
verification position requirement
WBC ≤2.5% □Pass □Fail
Repeatability
RBC ≤1.5% □Pass □Fail
Autoloader-Whole
HGB ≤1.0% □Pass □Fail
blood
MCV ≤1.0% □Pass □Fail
(Open vial-Whole
HCT ≤1.5% □Pass □Fail
blood)
PLT ≤4.0% □Pass □Fail

Attention:
1、Check if the rubber head of rotation barcode scanner presses at the center position of sample tube(Needs to be checked
every 6 months) ,
2、Lubricate the lead screw of syringe motor every year.
For the other operation parts like optic gain calibration & performance test, which were not mentioned by this
manual, please refer to BC-6800 Operator/Service Manual & Training materials.

page30of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

Appendix A
Group position for normal scatter gram for fresh blood:
A.1 DIFF Scatter gram

5/7

4/7

1/7

1/5 2/5 3/5 4/5


A.2 BASO Scatter gram

page31of32
血球产品全球技术支持部

Hematology Global Technical Support Dept.

A.3 NRBC Scatter gram

A.4 RET Scatter gram

page32of32