Device Information

Parameter Settings
Order information
Customer ECONOMY PROCESS SOLUTIONS PVT LTD
Customer purchase order EPSPL/1213
Sales order number / Item 62/3016367493/1023385510
Internal order number / Item 3002266356/0050

Device information
Description iTEMP TMT162, Field Transmitter
TAG TT-112
Serial number TB044D97223
Order code TMT162-65A8/1F0
Extended order code TMT162-E21331BBA

Additional information
Box / Boxes: 2/2
TAG, metal TT-112

Electronics

Electronic type 4...20 mA, HART®


Device revision 4
Firmware version 04.01.11
Hardware version 04.01
Model name TMT162
Device type 0x11CE
SIL NO

Input 1

Sensor type Pt100, IEC 751


Sensor connection 3- wire

Output

Output signal direction 4...20 mA


PV-Mode PV=Ch1, Ch2 not
Measuring range 0...250 °C
Output failure mode <= 3.6 mA
Damping 0s
Ofset 0K

Statement

The device operates within the manufacturers stated tolerances

Aurangabad, 2022-11-09
Endress+Hauser Wetzer (India) Pvt Ltd
Plot No. M 171/173; MIDC, Waluj,
431 136 Aurangabad
www.endress.com Document ID: TB044D97223_2022 Page 1 of 1
Endress+Hauser Wetzer (India) Pvt Ltd
Plot No. M 171/173; MIDC, Waluj,
431 136 Aurangabad
Phone: +91 240 256 3600
Fax: +91 240 255 5179
www.endress.com

Factory calibration certifcate

Certifcate-No. TB044D97223
Date of issue 09.11.2022
Order code TMT162-65A8/1F0
Extended order code TMT162-E21331BBA
Serial number TB044D97223

Tagging (TAG), metal TT-112


Box / Boxes: 2/2

Standards used

Description of the standard ID-number of the standard Standard calibrated to


Keithley 2010 20150319 02/2023
Digistant 4462 20150321 02/2023
Burster 4530 20150320 02/2023

Procedure of calibration/measurement and ambient conditions

All measurements have been made in ambient controlled at 25°C +/- 3K and R.H. 0%...80%

Measurement results

Digital output: Pt100 IEC

Nominal Value Measured temperature Deviation Max allowed deviation

(°C) (°C) (K) (±K)

Channel 1
-200,0 - -200,02 -0,02 0,06
10,0 - 10,00 0,00 0,07
220,0 - 220,02 0,02 0,08
430,0 - 430,02 0,02 0,09
640,0 - 640,01 0,01 0,10
850,0 - 850,04 0,04 0,11

Analog output: Range 0...250 °C

Nominal Value Measured output Measured temperature Deviation Max allowed deviation

(°C) (mA) (°C) (K) (±K)


0,0 4,001 0,016 0,016 0,075
50,0 7,201 50,016 0,016 0,075
100,0 10,400 100,000 0,000 0,075
150,0 13,600 150,000 0,000 0,075
200,0 16,800 200,000 0,000 0,075
250,0 20,000 250,000 0,000 0,075

TB044D97223 / 09.11.2022 Page 1 of 2


Measurement results

The test unit has been calibrated and complies with the tolerances stated by manufacturer. This report has been created
automatically and is valid without a signature.

TB044D97223 / 09.11.2022 Page 2 of 2