5 Monitor

Japanese Patent Application No. 2009-159089 PM-717


Model PM-717 is the continuous PM 2.5 monitor using beta attenuation
method. PM 2.5 mass concentration is evaluated using a unique humidity
control system that is implemented in relation to FRM procedure of
USEPA. PM-717 enables the precise measurement of PM 2.5 with
high-equivalency of FRM by the superior operating system technology.

PM2.5 Monitor
PM-717

Sample
It is necessary to keep the humidity of air sample lower, to Air
minimize mass variation of PM by moisture absorption or
deliquescence. PM-717 applies a unique system that
controls the humidity by mixing the dry air with sample. PM2.5
This system prevent to loss the excessive evaporation from Cyclone
the volatile matters in PM without heating the sample. Dry Air

KIMOTO realized the precise measurement of PM 2.5 with Air Mixer


high-equivalence of FRM by the continuous sampling for 24
Conditioned
hours without changing a filter spot. Furthermore, it is Air
unnecessary to replace the filter roll for one year. Dryer

PTFE tape filter that has few effected by moisture and gas
components.

Easy operation by large color LCD touch panel.

USB memory is available for data retrieve. PTFE


Tape
Filter Detector Pump
Large capacity memory saves measuring data for 1 year.

Optional cover tape roll protects sample spots for


speciation analysis. Pump
Standard Specifications Size (Desktop type and rack case type)
Measuring object PM2.5 in the ambient air

Measuring method Beta ray attenuation method (Unit: mm)


Measuring range 24h value (Dairy): 0 - 200 µg/m3, 485 490
1h value (as provisional): 0∼1000 µg/m3
Repeatability ±2% of equivalence film’s value

230
Inlet with cyclone,
Linearity ±5% of equivalence film’s value Temp./RH sensor
Span drift with ±10% of mass concentration Desktop type (weight : approx. 15kg)
standard air sample
Zero drift with clean air ±2μg/m3(24h value)

Sample flow rate 2.27LPM


485 590
Flow control Volumetric flow control

Filter media PTFE tape roll (20mm x 30m)


Beta ray source 14C, less than 10 MBq
Inlet with cyclone,
Detector Plastic scintillation detector Temp./RH sensor

Sampling inlet Inlet with PM2.5 cyclone

875
LCD display Measuring data, Output range, Alarm, Performance
and Control information

Internal data memory Measuring data (1min. data, 1h data), Message

Digital IN/OUT Ethernet, USB1.1, RS232C

Analog output for DC 0∼1V: Rack case type (weight : approx. 40kg)
telemeter(Optional) 24h value(Daily): 0 - 200 µg/m3,
1h value (as provisional): 0 - 200 / 500 / 1000 µg/m3
3 ranges auto/manual switching
*Starting range is selectable.
*Fixed range is also available.
Contact signals: Output range, Adjustment,
Power off, Failure report etc.
640 780
Power requirement AC 100 V, 50/60 Hz, approx. 450 VA

Withstand voltage test AC1000V 50/ 60Hz for 1min.

Insulation resistance More than 5 MEG


*For further information, please contact us. (3000)

Flow Diagram
1000

Sample air

PM2.5
Temp/RH Pressure Cyclone
sensor for sensor for
outdoor outdoor Beta source / Dry air
85

Sampling head
T,H P Air mixer
Conditioned air 540 690

Outdoor case type (weight : approx. 85kg)

T,H Temp/RH
sensor
Valve for sampling Product Code
head movement

PM-717-
F PTFE tape
filter
Flow Flow sensor Detector
controller for for sample P T1 Desktop type
sample Pressure Pump A1 Rack case type
Dryer
sensor
S1 Outdoor case type

R Serial recorder μRP1800


0 No use
F
T Analog telemeter communication unit TU07
Flow sensor Flow controller Pump for dry 0 No use
for dry air for dry air air circulation

KIMOTO ELECTRIC Co., Ltd.


3-1 FUNAHASHI-CHO, TENNOJI-KU, OSAKA 543-0024, JAPAN
PHONE: +81-6-6768-3401 FAX: +81-6-6764-7040
Home page: http://www.kimoto-electric.co.jp
E-mail: sales@kimoto-electric.co.jp