You are on page 1of 11

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/240642954

Thirty-Five Years of Mechanical Circulatory Support at the Texas Heart Institute

Article

CITATIONS READS
7 1,890

5 authors, including:

Courtney Gemmato Tim Myers


Beth Israel Deaconess Medical Center 116 PUBLICATIONS 4,856 CITATIONS
23 PUBLICATIONS 562 CITATIONS
SEE PROFILE
SEE PROFILE

All content following this page was uploaded by Courtney Gemmato on 16 November 2015.

The user has requested enhancement of the downloaded file.


Review
Thirty-Five Years of
Mechanical Circulatory
Support at the
Texas Heart Institute
An Updated Overview
Courtney J. Gemmato, BS Since the 1960s, the Texas Heart Institute has been intimately involved in the develop-
Matthew D. Forrester, BS ment of mechanical circulatory support devices (for example, ventricular assist devices,
Timothy J. Myers, BS aortic counterpulsation pumps, and total artificial hearts) for both short- and long-term
O.H. Frazier, MD
Denton A. Cooley, MD use. Here, we review the varied clinical experience with these technologies at the Texas
Heart Institute over the last 35 years. (Tex Heart Inst J 2005;32:168-77)

The human heart, about the size of a man’s fist,


pumps enough blood in the course of a lifetime to fill
the Rose Bowl. The wonderful thing about the heart
is that it replenishes itself. To duplicate that mechani-
cally is going to be a very difficult thing to do.

— Denton A. Cooley, MD*

R
eplacing the failing human heart has had a turbulent history. The 1st
human-to-human heart transplantation, performed by Christiaan Barnard
in 1967,1 marked an exciting advance in cardiac surgery. It revealed to the
medical community at large and to the public that failing hearts could be replaced,
at least technically, with healthy ones. However, the initial excitement was quickly
overshadowed by the difficulty of managing the attendant immunosuppressive
complications. By the early 1970s, only 4 institutions in the world were still
Key words: Equipment performing cardiac transplantation.2 Heart transplantation was reintroduced to a
design; heart, artificial; broader surgical population with the advent of the new immunosuppressive agent
heart-assist devices; heart cyclosporine in the early 1980s.3 This agent spared the nonspecific immune sys-
failure, congestive; heart
transplantation; history of tem and allowed the possibility of performing transplantation in patients who re-
medicine, 20th cent.; quired support with mechanical circulatory support (MCS) devices. Before this,
mechanical circulatory 3 patients supported with an MCS device as a bridge to transplant—two with an
support
artificial heart and one with a left ventricular assist device (LVAD)—had died of
overwhelming sepsis due to generalized immunosuppression while receiving aza-
From: The Cardiovascular thioprine. In March 1984, Frazier performed cardiac transplantation in a young
Research Laboratories,
Texas Heart Institute at St. woman who was septic with Staphylococcus aureus and Candida albicans at the time
Luke’s Episcopal Hospital, of transplantation. In spite of overwhelming sepsis, she survived the transplan-
Houston, Texas 77030 tation and was ultimately discharged safely from the hospital.4 This experience,
which showed that successful cardiac transplantation was feasible with cyclosporine
Address for reprints: use even in severely infected patients, stimulated research into the eventually suc-
O.H. Frazier, MD, Texas cessful use of MCS devices as a bridge to transplantation.
Heart Institute, MC 2-114A,
P.O. Box 20345, Houston, In 1966, DeBakey reported the 1st successful use of a ventricular assist device
TX 77225-0345 (VAD) for postcardiotomy support in a patient who was supported by the device
for 10 days.5 In 1969, Cooley became the first to implant an artificial heart (which
E-mail: was designed by Liotta) in a human being6 (Table I). Since the 1960s, the Texas
knowlin@heart.thi.tmc.edu Heart Institute (THI) has been involved in the development of other MCS de-
vices, including univentricular and biventricular devices, aortic counterpulsation
© 2005 by the Texas Heart ®
Institute, Houston *Oral communication to M.D. Forrester; 26 June 2001.

168 Mechanical Circulatory Support at THI Volume 32, Number 2, 2005


TABLE I. Summary of Mechanical Circulatory Support Systems Used at Texas Heart Institute

Year First Total No.


Used Patients FDA
Clinically Supported Original Regulatory
Device at THI at THI Intended Use Status

Liotta artificial heart 1969 1 Bridge to transplant N/A

Akutsu model III artificial heart 1981 1 Bridge to transplant N/A

TECO model VII ALVAD 1975 22 Postcardiotomy support N/A

Jarvik-7 TAH 1986 2 Bridge to transplant N/A

Hemopump 1988 43 Cardiogenic shock N/A

HeartMate IP LVAS
®
1986 87 Bridge to transplant PMA granted 1994

HeartMate VE/XVE LVAS


®
1991 141 Bridge to transplant; PMA granted 1998
destination therapy

Thoratec® PVAD 1987 89 Bridge to transplant; PMA granted 1995


postcardiotomy support

HeartMate® II LVAS 2003 8 Bridge to transplant IDE approved 2003

Jarvik 2000 Heart LVAD 2000 54 Bridge to transplant IDE approved 2000

AbioCor™ TAH 2001 5 Destination therapy IDE approved 2000

ABIOMED BVS 5000 VAD ®


1988 63 Postcardiotomy support; PMA granted 1993
bridge to transplant

ABIOMED AB5000™ VAD 2003 22 Postcardiotomy support; PMA granted 2003


bridge to transplant

TandemHeart® pVAD™ 2003 21 High-risk PTCA support; IDE approved 2002


cardiogenic shock

Levitronix® CentriMag® VAD 2003 17 Postcardiotomy support IDE approved 2003

ALVAD = abdominal left ventricular assist device; FDA = Food and Drug Administration; IDE = investigational device exemption;
LVAD = left ventricular assist device; LVAS = left ventricular assist system; PMA = premarket approval; PVAD = paracorporeal ven-
tricular assist device; pVAD = percutaneous ventricular assist device; TAH = total artificial heart; VAD = ventricular assist device

pumps, and total artificial hearts (TAHs). Historical- Utah in 1982, was used at THI in 2 patients in 1986.9
ly, these devices were used to provide short-term sup- In 1988, the Hemopump (Johnson & Johnson Inter-
port for patients in cardiogenic or postcardiotomy ventional Systems; Rancho Cordova, Calif ) was first
shock or as a bridge to transplantation. More recently, used at THI by Frazier and colleagues10 and was ulti-
they have come to be used to provide support during mately implanted in 43 patients over the next 5 years.
high-risk interventional procedures, as a bridge to re-
covery, or as destination therapy. Here, we review the Current Era of Mechanical Circulatory
varied clinical experience with these technologies at Support at Texas Heart Institute
THI over the last 35 years.
Long-Term Support Devices
Historical Use of Mechanical Circulatory Long-term MCS devices are used as a bridge to trans-
Support at the Texas Heart Institute plant, as a bridge to recovery, or as destination therapy.
Since 1969, VADs or TAHs have been implanted in Currently at THI, 6 systems are being used clinically
576 patients at THI. An abdominal left ventricular for long-term MCS. These include 5 VADs and a to-
assist device (ALVAD), the TECO Model VII (Ther- tally implantable artificial heart. Three of these systems
mo Electron Corp.; Waltham, Mass), was used to sup- are commercially available and 3 are investigational.
port 22 patients between 1975 and 1980.7 The Akutsu HeartMate® IP. The HeartMate implantable pneu-
model III artificial heart, developed at THI, was im- matic left ventricular assist system (HeartMate IP
planted in 1 patient in 1981.8 The Jarvik-7 TAH, de- LVAS, Thoratec Corp.; Pleasanton, Calif ) (Fig. 1) is a
signed at the University of Utah College of Medicine pulsatile blood pump that provides left ventricular sup-
and first implanted by DeVries at the University of port. The HeartMate IP is used in patients as a bridge

Texas Heart Institute Journal Mechanical Circulatory Support at THI 169


Fig. 1 HeartMate® implantable pneumatic (IP) left ventricular assist system. (Courtesy of Texas Heart Institute)

to transplant or bridge to recovery. The system consists and an improved version called the HeartMate ex-
of a pusher-plate blood pump, an interconnecting tended vented electric LVAS (HeartMate XVE LVAS)
drive line, and an external drive console.11,12 The pump (Fig. 2) are very similar to the HeartMate IP LVAS.
is positioned in the upper left abdominal quadrant, ei- The major difference between the IP device and the
ther intraperitoneally or preperitoneally. A percuta- VE devices is the method of actuation.12,13 The Heart-
neous drive line attached to an external console delivers Mate XVE is driven by an internal motor that requires
pulses of air that cause the pusher-plate and diaphragm 12 V of direct-current power. The motor controls the
to eject blood from the pump. A sensor, located in the pusher-plate and diaphragm within the pump. This
lower housing of the blood pump, monitors the posi- system is used as a bridge to transplant, as a bridge to
tion of the diaphragm in order to determine the stroke recovery, and as destination therapy. A percutaneous
volume. The pump can produce a stroke volume of 83 drive line attaches to a controller, which connects to a
mL and flows of up to 11.5 L/min. power base unit. The drive line also has a percuta-
The Texas Heart Institute has been involved in re- neous vent tube, the purpose of which is to equalize
search and development of the HeartMate LVAS the pressure in the motor chamber.12 The HeartMate
technology for more than 30 years. In 1978, the 1st XVE can be powered for short periods (4–6 hours) by
clinical trial of a predecessor of the HeartMate, the portable batteries, thus providing a distinct advantage
Model 7 LVAD, was initiated in 20 postcardiotomy over the pneumatic version and allowing more patient
patients at THI. The 1st clinical trial of the Heart- mobility.
Mate IP LVAS began in 1986, also at THI. The ini- Clinical trials of the HeartMate VE as a bridge to
tial trial was expanded to 17 centers in the late 1980s transplant began in 1991. This was the 1st device that
and early 1990s. In 1994, the US Food and Drug Ad- allowed untethered patients to be discharged in order
ministration (FDA) granted marketing approval for to return home and, in some cases, return to work,
the HeartMate IP LVAS as a bridge to transplant. The while awaiting transplantation.14,15 The FDA approved
HeartMate IP LVAS has been implanted in 87 pa- the HeartMate VE as a bridge to transplant in 1998
tients at THI. Since the development of an electrical- and approved the HeartMate XVE LVAS for the same
ly driven version of the HeartMate (described below), purpose in 2001. In light of the encouraging results of
use of the pneumatic version at THI has decreased the multicenter REMATCH trial,16 which compared
considerably. Nevertheless, our experience has dem- LVAD therapy and optimal medical management in
onstrated that the HeartMate IP LVAS is highly reli- patients ineligible for cardiac transplantation, the FDA
able for extended periods. approved another slightly modified version of the
HeartMate® VE and XVE. The HeartMate vented pump, the HeartMate® SNAP-VE, as destination ther-
electric LVAS (HeartMate VE LVAS; Thoratec Corp.) apy. In 2003, the HeartMate XVE LVAS was also ap-

170 Mechanical Circulatory Support at THI Volume 32, Number 2, 2005


proved by the FDA as destination therapy. The Heart- Indications for the Thoratec PVAD include acute
Mate VE or XVE LVAS has been implanted in 141 heart failure, postcardiotomy shock, and bridge to
patients at THI. transplant in cases of end-stage heart failure refractory
Thoratec® PVAD. The Thoratec paracorporeal ven- to maximal medical therapy. The system consists of a
tricular assist device (PVAD) system (Thoratec Corp.) blood pump, inflow and outflow cannulas, and a drive
(Fig. 3) is a pneumatically driven pump that can pro- console. The external pump and the cannulas rest on
vide either univentricular or biventricular support.17 the abdomen and connect to the drive console via
pneumatic and electric drive lines. The pneumatic con-
sole provides alternating negative and positive air pres-
sure that cause the pump’s blood sac to fill and eject
blood.18 The pump can produce a stroke volume of 65
mL and flows of up to 7 L/min.
Clinical trials with the Thoratec PVAD began in
1976, and the FDA granted marketing approval in
1995. In 1997, THI began using the Thoratec PVAD
in patients requiring right-, left-, or both-sided sup-
port for either short or long periods of time. The Tho-
ratec PVAD has been implanted in 89 patients at THI.
HeartMate® II. The HeartMate II left ventricular
assist system (Thoratec Corp.) (Fig. 4) is an implant-
able axial-flow blood pump that provides left ventric-
ular support. It is used as a bridge to transplant. The
system consists of a blood pump, a system driver, a
power base unit, and portable batteries. The blood
pump contains a single moving rotor powered by an
electromagnetic motor. A drive line, tunneled through
the patient’s abdomen, connects to the system driver,
which controls the motor. This system uses the same
power base unit and portable batteries as does the
HeartMate XVE. The HeartMate II has an operating
Fig. 2 HeartMate® extended vented electric (XVE) left
ventricular assist system. (Courtesy of Texas Heart Institute) range of 6,000–15,000 rpm and provides up to 10 L/
min of continuous cardiac output.18

Fig. 3 Thoratec ® paracorporeal ventricular assist device (PVAD) system. (Courtesy of Texas Heart Institute)

Texas Heart Institute Journal Mechanical Circulatory Support at THI 171


The Nimbus Corporation began development of
the HeartMate II in 1991. The internal components
of the device were based on those of the Hemopump
—the implantable, catheter-mounted axial-flow pump
that was first applied clinically at THI in 1988 but is
no longer used. The bearings of the HeartMate II are
blood lubricated. The 1st clinical trials were initiated
in Europe in June 2000. The initial clinical experience
was unfavorable and led to several design changes. The
FDA allowed the clinical trial to start again in 2003.
The 1st clinical use of the redesigned HeartMate II was
at THI in November 2003 in an 18-year-old man who
had dilated cardiomyopathy. The patient was success-
fully rehabilitated and was discharged in April 2004
after the FDA granted permission for study patients to
be discharged home to await transplant. The Heart-
Mate II has been implanted in 8 patients at THI and
in a total of 42 patients in the United States. Prelimi-
nary results indicate that the problems encountered in
the initial trial may have been resolved.
Jarvik 2000. The Jarvik 2000 Heart left ventricular
assist device (Jarvik Heart, Inc.; New York, NY) (Fig.
5) is an implantable axial-flow pump used for long-
term support as a bridge to transplant. The system Fig. 5 Jarvik 2000 Heart left ventricular assist device. (Cour-
consists of a blood pump, an external controller, and tesy of Texas Heart Institute)
portable batteries. The blood pump is approximately
the size of a C-cell battery and has a single moving im-
peller in the center of a titanium housing. A power
cable, tunneled through the abdomen of the patient, via a midline sternotomy, left thoracotomy, or sub-
connects to the external controller. Lead acid or lithi- costal (nonthoracic) approach. The various surgical
um ion batteries provide power to the pump. The approaches and the small size of the device allow it to
pump’s operating range is 8,000–12,000 rpm, which be implanted in patients who may not be candidates
can be manually adjusted to provide up to 6 L/min of for support by other LVADs.
continuous flow under optimal physiologic condi- Jarvik Heart, Inc., and THI began development of
tions.19 The Jarvik 2000 can be surgically implanted the Jarvik 2000 in 1988. Clinical trials of the Jarvik
2000 as a bridge to transplant began in April 2000. In
2003, the FDA granted permission for patients in that
study to be discharged home to await transplant. The
Jarvik 2000 has been implanted in 54 patients at
THI. Twelve of these patients have been discharged,
and one of them has been supported by the device for
more than 2 years.
AbioCor™ Total Artificial Heart. The AbioCor TAH
(ABIOMED, Inc.; Danvers, Mass) (Fig. 6) is a perma-
nently implantable replacement heart system that pro-
vides total circulatory support. The AbioCor TAH is
currently implanted in patients who are ineligible for
heart transplant, have a 30-day predicted mortality of
greater than 70% despite maximal medical therapy,
and have no other suitable treatment options.
The AbioCor system consists of 4 internal com-
ponents and 4 external components.20 The internal
components include the thoracic unit, a battery, a
Fig. 4 HeartMate® II left ventricular assist system. (Courtesy controller, and a transcutaneous energy transfer (TET)
of Thoratec Corp.) coil. The external components include a TET coil,
batteries, a TET module, and a bedside console. The

172 Mechanical Circulatory Support at THI Volume 32, Number 2, 2005


for up to 60 min. Thus, recipients of the AbioCor
TAH may be fully ambulatory. The pump’s motor op-
erates at 3,000–10,000 rpm and can generate a flow
of up to 8 L/min.20 One disadvantage of the device is
its large size, which limits its use in smaller patients
such as women, small men, and children. A smaller
version is being developed.21
Researchers at THI participated in preclinical tri-
als of the AbioCor TAH from 1993 to 2000. During
this period, the AbioCor went through several design
changes and was implanted experimentally in 43 calves.
In January 2001, the FDA granted permission to begin
a multicenter clinical trial of the AbioCor TAH with
15 patients. The Texas Heart Institute is one of 5 cen-
ters participating in this initial trial. Five patients at
THI and 14 patients overall have received the AbioCor
TAH. The longest surviving recipient at THI was sup-
Fig. 6 AbioCor™ total artificial heart. (From: Cohn LH, Edmunds ported by the device for 142 days. The longest surviv-
LH Jr. Cardiac Surgery in the Adult. 2nd ed. New York: McGraw- ing recipient overall was supported for 512 days.
Hill; 2003. p. 1512. Published with permission of The McGraw-
Hill Companies.) Short-Term Support Devices
Short-term circulatory assistance is often used to sup-
port patients in cardiogenic shock or postcardiotomy
thoracic unit consists of an energy converter and 2 shock, to provide protection during high-risk proce-
pumping chambers that function as the left and right dures, or to serve as a bridge to definitive therapy.
ventricles. Power is provided to the thoracic unit, and Four short-term devices are currently in clinical use at
the device’s internal battery is recharged transcutane- THI.
ously, by electromagnetic coupling of the internal and BVS 5000®. The BVS 5000 VAD (ABIOMED, Inc.)
external TET coils. The internal TET is positioned (Fig. 7) is a pneumatically driven, pulsatile, extracor-
under the skin in the upper left or right chest. The poreal, asynchronous pump that can be used for
external TET is attached with an adhesive to the skin short-term left ventricular, right ventricular, or biven-
above the internal TET and then connected to the tricular support.22 The device is used primarily as a
bedside console. Alternatively, power can be supplied bridge to recovery in patients who have exhausted all
to the AbioCor TAH by a pair of portable batteries other medical therapeutic options. The BVS 5000

Fig. 7 ABIOMED BVS 5000® circulatory support system. (Courtesy of Texas Heart Institute)

Texas Heart Institute Journal Mechanical Circulatory Support at THI 173


system consists of a dual-chamber blood pump, can-
nulas, and a drive console. The device is positioned at
the patient’s bedside and fills passively by gravity. A
bladder within the ventricular chamber is collapsed
by pulsed air delivered through drive-line tubing con-
nected to the external console. The BVS 5000 can
produce a stroke volume of 80 mL and flows of up to
6 L/min. Clinical trials of the BVS 5000 as treatment
for cardiogenic shock began at THI in 1988. The
FDA granted marketing approval of the BVS 5000 as
a postcardiotomy support device and as a bridge to
transplant in 1993. The BVS 5000 has been implant-
ed in 63 patients at THI.
AB5000™. The AB5000 circulatory support system
(ABIOMED, Inc.) is similar to the BVS 5000 in that it
uses the same drive console and tubing, but different
in that its prosthetic ventricle consists of a membrane
and 2 trileaflet valves. The indications for its use are
the same as those for the BVS 5000. Like the BVS
5000, the AB5000 can produce a stroke volume of 80
mL and flows of up to 6 L/min. In 2003, the FDA
granted marketing approval for use of the AB5000
device as postcardiotomy support and as a bridge to
transplant. The Texas Heart Institute is one of 20 cen-
ters where the device is now being used clinically. The Fig. 8 TandemHeart ® percutaneous ventricular assist device.
AB5000 has been implanted in 22 patients at THI. (Courtesy of CardiacAssist, Inc.)
TandemHeart® pVAD™. The TandemHeart pVAD
(CardiacAssist, Inc.; Pittsburgh, Pa) (Fig. 8) is a per-
cutaneous, continuous-flow, left ventricular assist de-
vice that provides short-term circulatory support. The
TandemHeart is used as support during cardiogenic
shock, as a bridge to long-term MCS, and for support
during high-risk percutaneous coronary interven-
tions. The system consists of a small blood pump,
transseptal and arterial cannulas, and a power base
controller. The pump contains a magnetically driven
rotor supported by a fluid-lubricated hydrodynamic
bearing. The TandemHeart can be inserted percuta-
neously in the cardiac catheterization laboratory or
surgically in the operating room. The TandemHeart
pVAD typically operates at speeds of 3,000–7,500
rpm and can provide a continuous flow of up to 4.0
L/min.
Development of the TandemHeart began in 1991.
In 2002, THI was one of 17 study sites that began en-
rolling patients in a phase II randomized clinical trial.
In July 2003, the FDA granted a 510(k) exemption
approving off-study use of the TandemHeart for up
to 6 hours in patients who did not meet the study
entry criteria. Three patients were enrolled in the trial,
and 18 others have been supported by the device un-
der the 510(k) exemption at THI.
Levitronix® CentriMag®. The Levitronix CentriMag
(Levitronix LLC; Waltham, Mass) (Fig. 9) is a con- Fig. 9 Levitronix ® CentriMag ® left ventricular assist device.
tinuous-flow left ventricular assist device with a mag- (Courtesy of Levitronix LLC)
netically levitated impeller. It can provide circulatory

174 Mechanical Circulatory Support at THI Volume 32, Number 2, 2005


support for up to 14 days in patients in cardiogenic systems such as the AbioCor TAH are under develop-
shock. The system consists of a small blood pump, ment, the internal batteries are able to provide suffi-
cannulas, a magnetic motor, and a bedside controller. cient power only for about 30 minutes. Hence, the
An advantage of this device is that it can be attached transcutaneous transmission of energy from an exter-
to cardiopulmonary bypass cannulas already in place. nal source is still required.
The Levitronix CentriMag system is capable of de- There has also been much discussion about the ap-
livering 9.9 L/min of flow at 5,500 rpm. The Texas propriate timing of VAD implantation and the se-
Heart Institute is one of 5 study sites that is now en- lection of appropriate recipients. Initially, MCS was
rolling patients in a phase I clinical trial that began in considered only after all other options had been ex-
2003. Four patients at THI have been enrolled in the hausted. Unfortunately, such a delay might allow
trial, and 13 patients have received the device under a multiple organ failure to progress until function is ir-
510(k) exemption. recoverable and prognosis is poor. However, recent ev-
idence indicates that MCS may in some cases improve
Discussion and restore end-organ function and thereby serve as a
bridge to recovery. As MCS becomes safer and more
The 1st implantation of an artificial heart by Cooley effective, it may become possible to place patients on
at THI was a last resort to prolong the life of a dying MCS earlier, thus improving their chances of func-
man. After an extensive surgery, the patient’s heart tional recovery. In cases of bridging to transplant, the
could not be weaned from cardiopulmonary bypass. use of MCS to restore adequate blood flow to organs
The heart was excised and replaced with a cardiac before they become permanently damaged is critical
prosthesis, which supported the patient for 64 hours for improving the prognosis after transplantation.
before a suitable donor heart became available for Therefore, the use of MCS at earlier stages in the pro-
transplantation. Without the Liotta heart, the patient gression of heart failure will likely become increasing-
would have died on the operating table, because a ly accepted.
suitable donor organ was not available at the time.5 Although bridge to transplant remains one of the
These circumstances have been repeated numerous principal uses for modern long-term MCS devices,
times since 1969. Whether the cause is postcardiot- other uses are envisioned as well. Myocardial recovery
omy failure, cardiogenic shock, or ischemic or idio- due to ventricular unloading and workload reduction
pathic cardiomyopathy, there is an inevitable point at with a VAD has been noted,25 especially in patients
which the human heart loses its ability to pump with acute cardiogenic shock. Reverse remodeling
blood despite maximal medical therapy. after long-term MCS has been noted too. Neverthe-
The clinical indications for MCS continue to ex- less, explanting MCS devices from a chronically sup-
pand in step with improvements in technology and ported patient is still typically performed only under
patient management. With time, devices are becom- extenuating circumstances such as infection or me-
ing smaller, easier to implant, more reliable, and more chanical malfunction. Further understanding of the
efficient. Implantation has been made faster and easi- phenomenon of reverse remodeling may lead to the
er by improvements in design and surgical techniques. use of long-term MCS in selected cases as a bridge to
In some instances, the smaller axial-flow VADs can recovery instead of a bridge to transplant.
be implanted without cardiopulmonary bypass.23 Be- Increasing experience with MCS has also highlight-
cause artificial surfaces and valves that come into con- ed the potential use of MCS devices as destination
tact with blood pose an inherent risk of thrombosis therapy. The AbioCor TAH is presently under inves-
and associated complications, the application of most tigation as destination therapy in patients who do not
MCS systems still requires concomitant systemic anti- qualify for heart transplantation and whose condi-
coagulation therapy. In some cases, however, the need tions are refractory to optimal medical management.
for anticoagulation and the risk of bleeding compli- The HeartMate XVE LVAD, which has been in use
cations have been reduced by the use of biocompati- in some form since the early 1980s, was approved
ble surfaces such as those in the HeartMate VE and for clinical use as destination therapy by the FDA in
XVE.24 Improved engineering has increased the dura- 2002. Even axial-flow LVADs such as the Jarvik 2000
bility and reliability of MCS devices, although there and the HeartMate II, which are now being investi-
still remains room for improvement. The increasing gated at THI and in Europe as bridges to transplant,
need for prolonged pump function is limited not only show the potential to be used as permanent long-term
by the finite mechanical lifetime of the devices, but therapy. Thus far, the experience with the Jarvik 2000
also by the current inability to provide an efficient, suggests that it may provide reliable support for sever-
inexhaustible, and portable power source. Most MCS al years. One patient in Europe26 has now been sup-
devices require externalization of either a pneumatic ported with a Jarvik 2000 LVAD for over 4 years.
or electrical drive line. Although totally implantable Here in the United States, 12 patients supported by

Texas Heart Institute Journal Mechanical Circulatory Support at THI 175


the Jarvik 2000 have been discharged from the hospi- not the ultimate solution to heart failure that many
tal, and 1 patient has been supported with the device thought it would be. Given the widely acknowledged
for more than 2 years. limitations of heart transplantation, MCS may evolve
Several recently introduced VADs provide short- beyond a temporary therapeutic alternative into an ac-
term MCS. Two of them in particular, the Tandem- cepted long-term, and perhaps even permanent, alter-
Heart pVAD and the Levitronix CentriMag, are now native. In the meantime, further research is warranted
being investigated at THI as part of multicenter clini- to develop MCS devices that more closely duplicate
cal trials. The potential clinical indications for short- the healthy human heart in terms of reliable function
term support with these devices include cardiogen- over long periods of time.
ic shock, high-risk percutaneous interventions, off-
pump coronary artery bypass grafting, and bridge to
definitive therapy. Both devices are small extracorpo- References
real circuits that can temporarily assume the work of 1. Barnard CN. The operation. A human cardiac transplant:
the ventricle. an interim report of a successful operation performed at
Our long, single-center experience with the devel- Groote Schuur Hospital, Cape Town. S Afr Med J 1967;41:
1271-4.
opment of MCS devices has imparted valuable infor- 2. Cooper DKC. Experimental development and early clinical
mation concerning the difficult problem of congestive experience. In: Cooper DKC, Miller LW, Patterson GA,
heart failure. It is clear, for example, that the function editors. The transplantation and replacement of thoracic
of the device must match the needs of the patient and organs: the present status of biological and mechanical re-
must be applied appropriately. Our early experience placement of the heart and lungs. 2nd ed. Dordrecht: Klu-
wer Academic Publishers; 1996. p. 153-60.
with the intra-aortic balloon pump, in which the de- 3. Borel JF, Feurer C, Gubler HU, Stahelin H. Biological ef-
vice was used in 34 patients over a 3-year period before fects of cyclosporin A: a new antilymphocytic agent. Agents
a single patient left the hospital after its application, Actions 1976;6:468-75.
dramatically illustrates the importance of the suit- 4. Frazier OH, Cooley DA, Okereke OU, Radovancevic B,
ability and timing of support.27 Today, the mechanism Chandler LB. Cardiac transplantation in a patient with sep-
ticemia after prolonged intraaortic balloon pump support:
of the intra-aortic balloon pump remains essentially implications for staged transplantation. Tex Heart Inst J
unchanged, and a 70% to 80% survival rate can be 1986;13:13-8.
expected when this technology is applied to the ap- 5. DeBakey ME. Left ventricular bypass pump for cardiac as-
propriate patient. The first use of the ALVAD led to a sistance. Clinical experience. Am J Cardiol 1971;27:3-11.
similar observation. The device worked well both ex- 6. Cooley DA, Liotta D, Hallman GL, Bloodwell RD, Leach-
man RD, Milam JD. Orthotopic cardiac prosthesis for two-
perimentally and clinically, but none of the 22 patients staged cardiac replacement. Am J Cardiol 1969;24:723-30.
who were supported with it survived. This experience 7. Norman JC, Duncan JM, Frazier OH, Hallman GL, Ott
aptly reinforces the importance of proper timing of de- DA, Reul GJ, Cooley DA. Intracorporeal (abdominal) left
vice application and its relationship to patient survival. ventricular assist devices or partial artificial hearts: A five-
Another important observation has been the diffi- year clinical experience. Arch Surg 1981;116:1441-5.
8. Cooley DA, Akutsu T, Norman JC, Serrato MA, Frazier
culty of implanting a biventricular pulsatile pump OH. Total artificial heart in two-staged cardiac transplanta-
(that is, the TAH). This technology’s current apogee tion. Cardiovasc Dis 1981;8:305-19.
of development is the AbioCor TAH. Its clinical effi- 9. Joyce LD, DeVries WC, Hastings WL, Olsen DB, Jarvik
ciency has been exemplary to date. Nevertheless, the RK, Kolff WJ. Response of the human body to the first per-
AbioCor TAH remains large and will not comfortably manent implant of the Jarvik-7 Total Artificial Heart. Trans
Am Soc Artif Intern Organs 1983;29:81-7.
fit within the chest cavity of most candidate patients. 10. Frazier OH, Wampler RK, Duncan JM, Dear WE, Macris
Therefore, its minimization with the use of continu- MP, Parnis SM, Fuqua JM. First human use of the Hemo-
ous flow technology, used so successfully in LVADs, pump, a catheter-mounted ventricular assist device. Ann
may play an important role in future clinical applica- Thorac Surg 1990;49:299-304.
tions of the TAH. 11. Frazier OH, Rose EA, McCarthy P, Burton NA, Tector A,
Levin H, et al. Improved mortality and rehabilitation of
Technological improvements and the increasing transplant candidates treated with a long-term implantable
availability of both short- and long-term MCS devices left ventricular assist system. Ann Surg 1995;222:327-38.
are broadening the use of MCS. The demand for 12. Frazier OH, Myers TJ, Radovancevic B. The HeartMate
MCS is increasing as improved medical therapies keep left ventricular assist system. Overview and 12-year experi-
patients with heart failure alive for longer periods of ence. Tex Heart Inst J 1998;25:265-71.
13. Petty M. Left ventricular assist systems. Medical Electronics
time and as nontransplant surgical techniques for 1999;April:34-7.
heart failure provide options for many patients who 14. Myers TJ, Catanese KA, Vargo RL, Dressler DK. Extended
would otherwise require transplantation. Despite med- cardiac support with a portable left ventricular assist system
ical or surgical management, sick hearts will eventually in the home. ASAIO J 1996;42:M576-9.
fail to pump enough blood to meet circulatory needs. 15. Myers TJ, Dasse KA, Macris MP, Poirier VL, Cloy MJ, Fra-
zier OH. Use of a left ventricular assist device in an outpa-
Simply replacing failing hearts with healthy ones is

176 Mechanical Circulatory Support at THI Volume 32, Number 2, 2005


tient setting. ASAIO J 1994;40:M471-5. device for cardiac failure. Perfusion 2001;16:13-8.
16. Rose EA, Moskowitz AJ, Packer M, Sollano JA, Williams 23. Frazier OH. Implantation of the Jarvik 2000 left ventricu-
DL, Tierney AR, et al. The REMATCH trial: rationale, de- lar assist device without the use of cardiopulmonary bypass.
sign, and end points. Randomized Evaluation of Mechanical Ann Thorac Surg 2003;75:1028-30.
Assistance for the Treatment of Congestive Heart Failure. 24. Scott-Burden T, Frazier OH. Cellular linings of ventricular
Ann Thorac Surg 1999;67:723-30. assist devices. Ann Thorac Surg 1995;60:1561-2.
17. Farrar DJ. The thoratec ventricular assist device: a paracor- 25. Khan T, Delgado RM, Radovancevic B, Torre-Amione G,
poreal pump for treating acute and chronic heart failure. Abrams J, Miller K, et al. Dobutamine stress echocardiogra-
Semin Thorac Cardiovasc Surg 2000;12:243-50. phy predicts myocardial improvement in patients support-
18. Frazier OH, Delgado RM 3rd, Kar B, Patel V, Gregoric ID, ed by left ventricular assist devices (LVADs): hemodynamic
Myers TJ. First clinical use of the redesigned HeartMate II and histologic evidence of improvement before LVAD ex-
left ventricular assist system in the United States: a case re- plantation. J Heart Lung Transplant 2003;22:137-46.
port. Tex Heart Inst J 2004;31:157-9. 26. Westaby S, Banning AP, Jarvik R, Frazier OH, Pigott DW,
19. Frazier OH, Myers TJ, Gregoric ID, Khan T, Delgado R, Jin XY, et al. First permanent implant of the Jarvik 2000
Croitoru M, et al. Initial clinical experience with the Jarvik Heart. Lancet 2000;356:900-3.
2000 implantable axial-flow left ventricular assist system. 27. Johnson MD, Holub DA, Winston DS, Brewer MA, Hibbs
Circulation 2002;105:2855-60. CW, Leachman CJ, Norman JC. Retrospective analysis of
20. Dowling RD, Gray LA Jr, Etoch SW, Laks H, Marelli D, 286 patients requiring circulatory support with the intra-
Samuels L, et al. The AbioCor implantable replacement aortic balloon pump. Cardiovasc Dis 1977;4:428-36.
heart. Ann Thorac Surg 2003;75(6 Suppl):S93-9.
21. Frazier OH, Dowling RD, Gray LA Jr, Shah NA, Pool T,
Gregoric I. The total artificial heart: where we stand. Car-
diology 2004;101:117-21.
22. Dekkers RJ, FitzGerald DJ, Couper GS. Five-year clinical
experience with Abiomed BVS 5000 as a ventricular assist

Texas Heart Institute Journal Mechanical Circulatory Support at THI 177

View publication stats

You might also like