Professional Documents
Culture Documents
5/F., 4thBldg.,NO.2,
st
Gaoxin C. 1 Ave.,
High-tech Industrial Park,
Nanshan,Shenzhen 518057, P.R. China
Tel: (+86) 755 26516667, 26952867
Fax: (+86) 755 26747910
Contents
CONTENTS
2.1.1 Function...............................................................................................................................16
2.1.2 Interface...............................................................................................................................17
2.2 CCU BOARD ........................................................................................................................ 17
2.2.1 Function...............................................................................................................................17
2.2.2 Front Interface ..................................................................................................................... 17
2.3 TFT CABLE CONVERTING BOARD ..........................................................................................19
2.3.1 Function...............................................................................................................................19
2.3.2 Interface...............................................................................................................................19
2.4. LCD ADOPTER..............................................................................................................................20
2.4.1. Function ....................................................................................................................................20
2.5 RECORDER BOARD.................................................................................................................. 21
2.5.1 Function .....................................................................................................................................21
2.6 DOM CARD ...................................................................................................................................21
2.6.1 Function .....................................................................................................................................21
T- 6807-30-02357 V1.0 I
Contents
5.1 GENERAL.............................................................................................................................31
5.2 ADJUST PROGRAM ............................................................................................................ 31
5.3 ADJUST PLACE ................................................................................................................... 32
6.4 MONTHLYMAINTENANCE..............................................................................................35
6.5 YEARLYMAINTENANCE..................................................................................................35
6.6 THE MAINTENANCE BEFORE TRANSPORTOR FOR INSTRUMENT THAT
WILL NOT BE USED FOR A LONG TIME......................................................................................35
II T- 6807-30-02357 V1.0
Instrument Introduction
1.1 STRUCTURE
1.1.1 Host
Control the process of sample collection, dilution and analysis, including the following units:
Hematology analyzer
1.1.2 Accessory
2. Diluent cap components, washing liquid cap components, hemolytic agent cap components,
1
Instrument Introduction
4. Waste barrels;
1.2 PURPOSE
It is used for detecting the parameters of RBC, WBC, HGB and differential counter.
The analyzer is used for the determination of the following 20 parameters and 3 histograms blood
2
Instrument Introduction
1.3 SPECIFICATIONS
According to the defence for deleterious liquid: General Device (Closing device of
non-defensive liquid)
anesthetic gas and air or gas mixture of oxygen and nitrous oxide: Do not use the
equipment in condition that using gas mixture of flammable anesthetic gas and air or
Principles of Measurement
Blood Cells are counted and sized by Electrical Impedance Method. Hemoglobin is
Parameters of Measurement
Basic parameters:
3
Instrument Introduction
Deviation
Calculated Parameters:
Hematocrit HCT %
Plateletcrit PCT %
Sampling Features:
RBC/PLT 1:44600
RBC/PLT 1:44500
WBC/RBC: 80μm
Display
Language
4
Instrument Introduction
English
Indicator
Alarm Indicator
Alarm
Interface
Recorder
When the liquid reagent is in the sensor the voltage is 3.0V ~ 5.0V;
When the liquid reagent is not in the sensor the voltage is ≤1.5V。
WBC Aperture voltage reference value :the 5 times average W.H.V.S value of blank test
RBC Aperture voltage reference value:the 5 times average R.H.V.S value of blank test.
Aperture voltage
Take electrode cable from CCU board and use the two pens of multimeter (which
5
Instrument Introduction
have been set the resistance value) to measure the Chamber shield and substrate
Insulation. The value should be more than 20MΩ。When you finish test, return the
Operating Ambient
Temperature: 15℃~35℃
Humidity: 10~90%
Ambient Storage
Temperature: -10℃~40℃
Humidity: ≤80%
Electric Specifications
Fuse: AC 250V 2A
6
Instrument Introduction
Dimensions: 394mm×381mm×500mm
Weight: 23kg
1.4 STRUCTURE
1 5 4 3 2
Figure 1-1
5. Thermal Recorder
7
Instrument Introduction
(9)
(5)
(4)
(6)
(2)
(7)
(3)
(8)
(1)
(18)
(17)
(16)
chart 1-2
8
Instrument Introduction
1.5 OPERATION
Figure 1-3
INFORMATION SECTION
1.5.2 Menu
You can operate the instrument by menu. Press the right key of the mouse, directing to the
function menu of the current window. Move the mouse to the required menu bar, and then press
the left key of the mouse, the required function will be selected. Pressing the left key out of the
9
Instrument Introduction
The common functions are showed at the top right corner of the screen as shortcut key. Move the
Sample Mode: Choose the sample mode among Venous Blood Mode, Capillary Blood Mode and
Prediluted Mode.
Diluent: Inject the diluent quantificationally by sample needle. This menu uses only in Prediluted
Mode.
Information: Enter the information window of the next sample, input and modify the
information.
Histogram: Enter the histogram adjustment window of the current sample. You can manually
Print: Output the data of the current sample to the appointed device.
About: Display information of the instrument such as software and copyright etc.
Shutdown: Startup the shutdown program to rinse the measurement unit automatically. Shutoff
the power supply when the screen displays “Please switch off”.
Function Menu
Review: Enter the review window. You can review, inquire or modify the measurement data of the
former sample.
Quality Control: Enter the quality control window to carry out quality control operation.
Setting: Enter the setting menu.
Calibration Menu
Manual Calibration: Enter the manual calibration window to calibrate the instrument manually.
10
Instrument Introduction
Auto Calibration: Enter the auto calibration window to adjust the instrument automatically.
QC Menu
QC Run: Enter the QC run window to carry out the selected QC files.
Setting Menu
Time: Enter the time setting window to set the time of the instrument.
Printout: Enter the printout window to set the functions such as print or report.
Config: Enter the config window to set the functions such as print, sleep, clean, and display etc.
Department: Enter the department window to set the department and abbreviation.
Operator: Enter the operator window to set the department, abbreviation, department and
operator.
Service Menu
Cleaning: Run the normal rinse process to rinse the measurement unit.
Park: Run the Park Program to rinse and empty the tubing.
Mechainc: Enter the Mechainc window to verify the functions of the instrument.
11
Instrument Introduction
The count principle of the instrument is based on the measurement of changes in electrical
a) The sample blood is diluted in a conductive liquid. As blood cells are non-conductive, the
b) When the diluent passes through the aperture sensor, electrodes are submerged in the liquid on
c) When cells pass through the aperture, the resistance between the electrodes increases as the
cell volume increasing, as is shown in figure 1-4.According to the Ohm Formulary: U=RI
increasing.
d) Passing through the magnification circuit, the voltage signal will be magnified and the noise
will be filtered, then you will gain the analytical results, as shown in figure 1-4.
e) One count pool and the detection circuit count the WBC. Another count pool and the detection
circuit count the RBC and PLT. The microprocessor of the instrument calculates and analyzes
the cells (WBC, RBC, PLT), and then gives out the histograms.
f) PLT count adopts advanced liquid, electronics and software system. It settles the repetitive
12
Instrument Introduction
Figure 1-4
Adding lyse in the blood, the red blood cell will rapidly be broken down and release hemoglobin.
Hemoglobin and lyse form a new mixture, which can absorb the wavelength of 540nm.Measure
the absorbency. Through the comparison of the absorbency between the pure diluent and the
When different types of cells pass through the aperture sensor, there will be different electrical
pulse height. Because of the evident difference of the cells sizes, the instrument is able to
differentiate the white blood cell, red blood cell and platelet by its pre-set program.
WBC 120~1000fL
RBC 82~98fL
PLT 2~35fL
The leucocytes disposed by lyse can be divided into three types, according to their volumes:
LYM 35~90 fL
MID 90~160 fL
13
Instrument Introduction
GRAN 160~450 fL
The parameters of the sample blood test can be described in three ways:
MCV derives from histograms and the instrument collecting and classifying the
RDW represents the volume distribution of the erythrocyte populations, derived from
the RBC histogram. It can be expressed by the coefficient variation of the erythrocyte
volume as RDW-CV, and the unit is %.It can also be expressed in standard deviation of
MPV is the average volume of individual platelets, derived from the PLT histogram. It
represents the mean volume of the PLT populations and can be expressed in fL.
PDW derives from the PLT histogram. It represents the geometry standard deviation (10
P-LCR derives from the PLT histogram. It represents the ratio of the larger PLT and the
unit is %.
HCT(%)= RBC×MCV/10
MCH(pg)= 10×HGB/RBC
MCHC(g/L)= 100×HGB/HCT
PCT(%)= PLT×MPV/10
14
Instrument Introduction
15
Hardware
Chapter 2 HARDWARE
CPU Board, CCU Board, cable adapter board, liquid tubing optical conpler optocoupler adapter
board, motor optocoupler adapter board, DOM card, LCD adapter board, transformer, recorder
2.1.1 Function
With CPU as the core, the computer system also includes some peripheral equipments, such as
DRAM, SRAM, besides there are some interfaces, such as USB interface, COM2 interface,
COM1 interface, LPT printer interface, power supply interface, mouse/keyboard interface, display
screen TFT cable interface, peripheral display interface, DOM interface. As the chart 2-1.
The basic components of the computer system are CPU, DRAM, SRAM. They also supplied the
16
Hardware
2.1.2 Interface
1 1、 COM1interface
2、 COM2 interface
3、 USB interface
2
4、 Mouse interface
3 5、 Keyboard interface
interface
interface
6
7 8 9
Chart 2-1
2.2.1 Function
17
Hardware
1
3 4 5 7 6
Chart 2-2
18
Hardware
11
10
8
Chart 2-3
2.3.1 Function
The function of TFT Cable Converting Board is to connect all blocks and CCU board, and to
adjust the sensitivity of liquid sensor through adjusting the trimmer on TFT Cable Converting
Board.
2.3.2 Interface
19
Hardware
2.4.1. Function
The main function of LCD is to connect screen and CPU with cable and TFT through FPC soft
cable. It is used to transmit signal from CPU to display screen. Display as following: chart 2-5.
Chart 2-5
20
Hardware
2.5.1 Function
The main functions of the recorder board are to receive and deal with the printing data, and then
transmit the data to thermal printer. At the same time, the recorder board drives the motor to print
Power supply
interface
Chart 2-6
2.6.1 Function
DOM card is used for deposition. It should make sure the DOM card is in correct way before
Drawing tablets
DOM jack
Drawing tablets
Chart 2-7
21
Tubing
Chapter 3 TUBING
Chart 3-1
22
Tubing
8 2、 Lyse sensor.
6
3、 Vacuum pump block.
1
4、 Dilution sensor.
7 2
5、 Waste liquid sensor.
4
6、 Dilution block
7、 Sample machine.
3
8、 Glass bulb
9、 Cleaner sensor.
9 5 Chart 3-2
The instrument includes four tubing sensor: diluent tubing sensor, rinse tubing sensor, lyse tubing
The tubing sensor, rinse tubing sensor, lyse tubing sensor are sued to check if the reagent enters
into the tubing system. If there is no reagent in the tube. The instrument will give alarm of “no
Waste liquid sensor is used to test if the waste liquid is full. If yes, there instrument will give alarm
reagent temperature sensor is used to check the reagent temperature in the tube. If the temperature
is too low the instrument will give alarm of “temperature too low”.
23
Tubing
This part consists of sample sucker structure and dilution structure. Sample sucker structure is
used to aspirate sample, add it into the probe structure and clean the sample needle. Dilution
The vacuum decelerating structure is working under the system software and is used to provide
power in the condition of electrifying. Vacuum pump is driven by vacuum decelerating structure
and produce negative pressure. So that to provide working power to all this tube system such as
waste emissions.
This part mainly refers to the tubing hose, connecting tube and electromagnetically operated
valve.
24
System Structure
Block
slider
Chart 4-1
25
System Structure
It is necessary to replace the sample needle wipe block, seal washer and synchronous driving belt
26
System Structure
It is necessary to replace the seal washer of the vacuum pump periodically. The replacing cycle is
one year.
27
System Structure
It is necessary to replace the seal washer of the piston block periodically. The replacing cycle is
one year.
28
System Structure
Chart 4-6
29
System Structure
Chart 4-7
30
Adjustment
Chapter 5 ADJUSTMENT
5.1 GENERAL
In main operation window of the instrument, click the right key of the mouse, pops up the menu.
Move the cursor to “Calibration/Manual Calibration”. Adjust the calibration factors to 100%.
Click the right key of the mouse, and pops up the menu, then move the cursor to
Click the right key of the mouse, pops up the menu, then move the cursor to
Use the calibration of Procan recommended. In main operation window, measure the calibration.
According to the measured histograms to adjust the gain of the passage. Adjust the gain of passage
WBC, RBC according to the Multi-Parameter Assayed Hematology Control Expected Results
Sheet.
In main operation window of the instrument, click the right key of the mouse, pops up the menu.
then move the cursor to “Service/Test”, enter the password “6800”,click “HGB Background/T”
31
Appendix 2 Debug record form
12
32
Maintenance
Chapter 6 MAINTENANCE
As other precision instrument, only careful daily service and periodic maintenance can the
instrument have a good working status, and can we get the reliable measurement results and have
less malfunctions. This chapter introduces some preventive methods for service and maintenance.
If you want to know more relative information, please contact the customer service department of
Procan.
According to the requirements for maintenance of the instrument during the using procedures, we
divide the preventive service and maintenance into the following types: daily, weekly, monthly,
Warning:
If the hospital or agency don’t perform a satisfying plan to maintain the instrument, there
Wipe the appearance of the instrument with neutral detergent or distilled water.
Caution:
Avoid using corrosive acids, alkali, and volatile organic solvent such as: acetone, aether,
chloroforms to wipe the appearance of instrument. Only can use the neutral detergent.
Caution:
33
Maintenance
The fuse is installed in the fuse-box on the side of power switch. Open the box to replace the fuse
expediently.
Warning:
Run
execute auto cleaning procedure according to the quantity of the sample to keep the
instrument in good working status. Set the auto cleaning procedure according to the
Shutdown
The instrument should be operated by auto maintenance before shut off the instrument.
You only need to clean the workbench and wipe the appearance of instrument when the
power-off.
b) In the main operation window, press the right key of the mouse, pops up the menu.
34
Maintenance
Operate as follows:
a) Turn off the power supply, and pull out the power line.
Take off the cover and the net. Clean them with neutral detergent brush and put them in the
shade.
c) Airing naturally .Brush away the dust clinging on the air filter grid at the rear panel of
instrument.
d) Put the air filter grid on the air filter net carefully. Planish them.
Notice:
Only can use the neutral detergent cleaning the air filter. Avoid airing it with heat.
It’s necessary to perform preventive maintenance once every year. Because of the
authorized engineer of Procan. Please contact the customer service department of Procan before
yearly maintenance.
If the instrument is not going to be used in 2 weeks or longer, or need to pack and transport,
a) In the main operation window, press the right key of the mouse, choose “Service/Park”
35
Maintenance
in the menu.
c) Rotate the bottle covers of rest reagents and store them as the reagent operation
introduction. User should perform efficient action to prevent the material from
d) Plug the stopples which were pulled out at the first installation into the corresponding
tube connector.
e) Pull out diluent connection tube, rinse connection tube, lyse connection tube and waste
connection tube and clean them with distilled water. Dry them in shady place, then
f) Pull out the power wire, packing it in plastics after cleaning with neutral detergent.
g) Put the instrument and parts packed in plastics into packing cases.
36
Troubleshooting
Chapter 7 TROUBLESHOOTING
technical
Make sure if any requirement
part is connected
correctly and OK
reliable.
END
Chart 7-1
37
Troubleshooting
Humidity: 10~90%
If the temperature is below 15 ℃,the activity of the reagent will reduce. It will weaken the capacity
of dissolving the blood. In a certain time, it can not break down the red blood cell membranes into
2fl cellular debris. At lower temperature conditions, the red blood cell debris will gather together.
It will affect the measurement of leukocyte, resulting in the count value of WBC increasing .It will
also affect the measurement of HGB, resulting in the count value of HGB increasing.
Sample Collection: Observe whether the operator uses correct method to collect the blood
/Chapter 3/Section 3.5 Preparation For Sample Collection), and whether the sample is stored in
Sample Storage: After the sample blood is well-mixed, test on the instrument promptly. As time
goes by, the volume and density of the cells will change. The environment and the patient
determine the varying rate of the volume and density. If the sample is placed a long time at room
temperature, the volume of RBC will increase, and the MCV, RDW, PDW will be affected, and the
Affection of WBC Classification: To get the exact measuring results, the venous blood sample
measurement must be completed within 4 hours. If the sample blood is placed a long time at room
temperature, the total number of WBC may change little, but the classification is inaccurate.
7.3 BACKGROUND CHECK
Background Check: This item can ensure the measurement and the calculated data won’t be
affected by reagent empty. If the background value exceeds the permissive area, the measured and
38
Troubleshooting
HCT ≤0.5% %
9
PLT
9
≤10×10 /L 10 /L
Background Test Method: When power-on, suggested to run background test, you can also
a) In the blood cell analyzer window, press “Info” button, and pops up the information
edit window. Then click the item textbox, change the ID as 999999999.then press
b) Background count in venous or capillary mode, press the “START” key below the
sample needle directly for background count. In prediluted mode, the sample needle
aspirates some non-contaminated diluent from the chamber for background count.
is also necessary to dilute sample with the diluent which is prior picked-up in this
method):
In main operation area, press the right key of the mouse, pops up the menu. In menu operation
area, select “Sample Mode/ Prediluted” to set the sample mode as prediluted mode.
In main operation area, press the right key of the mouse, pops up the menu. In the menu operation
area, select “Diluent”. Place a clean empty chamber or tube under the sample needle, press
“START” key to dispense the diluent into the chamber. Press the “START” key once, the diluent
is dispensed into the chamber or tube once (The quantity can be used for background count in the
39
Troubleshooting
3、 Electronic interference affects measurement and the calculated circuit, for example:
4、 Tubing Contamination
40
Lead label
Waste liquid
12 WST
sensor
13 switch STA
41
Debug record form
42
Debug record form
43
Debug record form
WBC accuracy ±
error range 6%
RBC accuracy ±
error range 5%
Test accuracy OK
HGB accuracy ±
Not
error range 4%
ok.
MCV accuracy ±
error range 5%
PLT accuracy ±
error range 14%
Printing characters, graphics should be correct, clear. The
functioning of institutions should be smooth. There is no abnormal □ OK
Record and alarm
sound, no paper block but with alarm function phenomenon, there □ Not ok.
are alarm.
Pinter and net Printing characters, graphics should be correct, clear.Meanwhile □OK
function the PC can accept complete and correct data and graphic. □Not ok.
WAperture
□OK □ Not ok.
estimation.
44
Troubleshooting
Fuse burn-out Fluctuating input AC voltage Use stabilized voltage power supply
Power supply board is short circuit Replace new power supply board
Bubbles The setup of count time is too long Reset the count time
The back cover and body of chamber lacks Replace the the back cover of sample
cover
The setup of count time is too short Reset the count time
Clogs
Aperture clogs Clean the aperture
Unclear display Incorrect connection of the signal line Re-connect the signal line
Abnormal HGB The WBC chamber is dirty , Dip in the chamber ,or adjust the gain
of HGB
background The battery is not get enough light Replace battery or WBC probe
voltage components
LBD Is Not Light Enough Replace the LBD or WBC probe
components
Some keys of the The CPU board or the keyboard is broken Replace the CPU board or the
keyboard is keyboard
irresponsive
45
Troubleshooting
Abnormal The sample is placed a too long time Measure the sample within 30
mid-sized cell minutes
percentage The anticoagulant is superfluous Reduce the quantity of anticoagulant.
No reagent, Fluid hose seal in bad way Check the connection and get rid of
liquid lack
Rinse(diluent/lyse)
The piston block leaks Replace the piston block or seal ring
empty alarm The rinse(diluent/lyse) sensor is broken ,or Replace the rinse(diluent/lyse) sensor
the connector is broken off
The count time is longer than setup time Reset the count time
WBC/ RBC clogs One sensor of the MTB board is broken Replace MTB board
There is no result There are problems on CCU board Replace or repair CCU board
of RBC, WBC The electrode cable is broken Replace or repair probe components
measurement
46