You are on page 1of 182

Abacus 5

Hematology Analyzer
Service Manual
Rev 2.0

DIATRON MI PLC
39 Táblás street, Budapest, H-1097, HUNGARY
Tel.: +36 1 436 9800
Fax: +36 1 436 9809
www.diatron.com
E-Mail: support@diatron.com
Revision 2.0
2014-JUNE-02
2

Table of Contents
TABLE OF CONTENTS ............................................................................................................................................. 2
1 INTRODUCTION ............................................................................................................................................... 7
1.1 WHO SHOULD READ THIS MANUAL ................................................................................................................ 7
2 FOR YOUR SAFETY ......................................................................................................................................... 8
2.1 SPECIAL SYMBOLS USED IN THIS MANUAL................................................................................................... 8
2.2 GENERAL PRECAUTIONS ............................................................................................................................... 8
2.3 ENVIRONMENTAL FACTORS .......................................................................................................................... 9
2.4 ELECTRICAL REQUIREMENTS ...................................................................................................................... 10
2.5 SUGGESTED ON-LINE UPS ........................................................................................................................... 10
2.6 PROPER PLACEMENT OF ABACUS 5 ............................................................................................................. 11
2.7 WEIGHT REQUIREMENTS ............................................................................................................................. 12
2.8 WASTE DISPOSAL ........................................................................................................................................ 12
2.9 EMERGENCY SITUATIONS ............................................................................................................................ 13
3 STRUCTURE OF THE ANALYZER ............................................................................................................. 14
3.1 COMPONENTS LOCATED ON THE FRONT PANEL ............................................................................................ 14
3.1.1 Display screen and the touch sensitive surface ...................................................................................... 14
3.1.2 Start Button and LEDs ........................................................................................................................... 15
3.2 COMPONENTS ACCESSIBLE AFTER OPENING THE FRONT PANEL ................................................................... 15
3.2.1 Shear Valve Assembly ............................................................................................................................ 15
3.2.2 Sample rotor ........................................................................................................................................... 16
3.2.3 Main Dilutors ......................................................................................................................................... 16
3.2.4 Dilutor opto sensor boards..................................................................................................................... 17
3.2.5 Tube organizer ....................................................................................................................................... 17
3.2.6 Temperature Control Unit ...................................................................................................................... 17
3.2.7 Laser head .............................................................................................................................................. 18
3.2.8 Laser Head Assembly + Sample Injector ............................................................................................... 19
3.2.9 Laserdiode Driver Board ....................................................................................................................... 20
3.2.10 Pin Photodiode and Amplifier (OPTSENSOR_2v1) .......................................................................... 20
3.3 LEFT SIDE .................................................................................................................................................... 21
3.3.1 Valve boards........................................................................................................................................... 21
3.3.2 WBC/BASO Preheater Assembly............................................................................................................ 22
3.3.3 Counting chamber with electrodes and measuring aperture ................................................................. 22
3.3.4 HGB Measuring Head ............................................................................................................................ 23
3.3.5 Cell counter Amplifier Board ................................................................................................................. 24
3.3.6 Pressure sensor board ............................................................................................................................ 25
3.3.7 Reagent and Vacuum buffers .................................................................................................................. 27
3.3.8 Reagent Sensor Board ............................................................................................................................ 27
3.3.9 Opening the valve assembly plate .......................................................................................................... 27
3.3.10 Vacuum buffer .................................................................................................................................... 27
3.3.11 Pneumatic and Power Boards (PPB1 and PPB2) ............................................................................. 28
3.3.12 Pump assembly .................................................................................................................................. 29
3.4 RIGHT SIDE .................................................................................................................................................. 30
3.4.1 XY unit .................................................................................................................................................... 30
3.4.2 XYROpto Board ...................................................................................................................................... 31
3.4.3 Sampling needle ..................................................................................................................................... 31
3.4.4 Blood sensor ........................................................................................................................................... 31
3.4.5 Wash head .............................................................................................................................................. 32
3.4.6 Processor unit (LS-DACQ board with DimmBoard) ............................................................................. 32
3.4.7 Mini-ITX PC mother board (In Hardware block) .................................................................................. 38
3

4 ELECTRONIC BLOCK DIAGRAM .............................................................................................................. 39


4.1 HARDWARE BLOCK ..................................................................................................................................... 39
4.2 DATA ACQUISITION UNIT ............................................................................................................................ 41
4.3 AMPLIFIER BOARD ....................................................................................................................................... 43
4.4 OPTICAL HEAD ............................................................................................................................................ 44
5 OPERATION OF THE FLUIDIC SYSTEM .................................................................................................. 45
5.1 THE REAGENT SYSTEM ................................................................................................................................ 45
5.2 FLOW DIAGRAM OF MEASUREMENT ............................................................................................................. 46
5.3 INITIALIZATION OF THE FLUIDIC SYSTEM .................................................................................................... 48
5.4 REAGENTS PRIMING ..................................................................................................................................... 48
5.5 PIERCING PROCESS ...................................................................................................................................... 48
5.6 SAMPLING PROCESS ..................................................................................................................................... 48
5.7 NEEDLE WASHING PROCESSES ..................................................................................................................... 49
5.8 DILUTING PROCESSES .................................................................................................................................. 49
5.9 LYSING PROCESS ......................................................................................................................................... 50
5.10 RBC COUNTING PROCESS ............................................................................................................................ 50
5.11 WBC/BASO COUNTING .............................................................................................................................. 50
5.12 WBC 4DIFF COUNTING ............................................................................................................................... 51
5.13 CHAMBER DRAINING PROCESSES ................................................................................................................. 51
5.14 CLEANING (RINSING) PROCESSES................................................................................................................. 51
5.15 STANDBY PROCESS ...................................................................................................................................... 52
5.16 WAKE UP PROCESS ...................................................................................................................................... 52
5.17 SHUTDOWN PROCESS ................................................................................................................................... 52
6 ADJUSTMENTS ............................................................................................................................................... 53
6.1 XY UNIT MECHANICAL ADJUSTMENTS......................................................................................................... 53
6.1.1 XY horizontal movement......................................................................................................................... 53
6.1.1.1 Greasing the horizontal rods ........................................................................................................................... 54
6.1.1.2 Set opto flags for horizontal movement .......................................................................................................... 54
6.1.2 XY vertical adjustment ........................................................................................................................... 56
6.1.2.1 Basic operation ............................................................................................................................................... 56
6.1.2.2 Adjusting the needle’s vertical position .......................................................................................................... 57
6.1.3 Washing head settings ............................................................................................................................ 58
6.1.4 Needle setting ......................................................................................................................................... 59
6.1.4.1 Handmade setting ........................................................................................................................................... 59
6.1.4.2 Software made setting .................................................................................................................................... 60
6.2 SHEAR VALVE MECHANICAL SETTING.......................................................................................................... 60
6.2.1 Shear valve lower disc position setting .................................................................................................. 60
6.2.2 Shear valve upper disc setting ................................................................................................................ 61
6.2.3 Setting the opto sensors of the shear valve. ............................................................................................ 63
6.3 AMPLIFIER OFFSET SETTING ........................................................................................................................ 64
6.4 BLOOD SENSOR CALIBRATION ..................................................................................................................... 65
6.5 SAMPLE POSITION SETTING .......................................................................................................................... 67
6.6 PRESSURE SENSOR ADJUSTMENT ................................................................................................................. 69
6.7 SET SERIAL NUMBER.................................................................................................................................... 69
6.8 WBC PREHEATER ........................................................................................................................................ 69
6.9 SETTING LASER VALUES .............................................................................................................................. 69
6.9.1 Calibration procedure of optical heads ................................................................................................. 70
6.9.2 Short guide for laser values.................................................................................................................... 73
7 VERIFICATION PROCEDURES ................................................................................................................... 77
7.1 SELF-TEST ................................................................................................................................................... 77
7.1.1 Abacus 5 Self-Test Guide ....................................................................................................................... 78
7.1.1.1 General information........................................................................................................................................ 78
7.1.1.2 Electronic Self-Test parameters ...................................................................................................................... 79
4

7.1.1.3 Pneumatic Self-Test parameters ..................................................................................................................... 81


7.2 ERROR MESSAGES ....................................................................................................................................... 82
7.3 SERVICE MENU ........................................................................................................................................... 87
7.3.1 Service functions .................................................................................................................................... 89
7.3.1.1 Test functions ................................................................................................................................................. 90
7.3.1.2 Low level file management ............................................................................................................................ 90
7.3.1.3 Low level reboot (DimmBoard) ..................................................................................................................... 90
7.3.1.4 Pneumatical System – Initialize...................................................................................................................... 90
7.3.1.5 Network management ..................................................................................................................................... 91
7.3.1.6 RAW data saving mode .................................................................................................................................. 91
7.3.1.7 Export RAW files ........................................................................................................................................... 91
7.3.1.8 Windows Control Status ................................................................................................................................. 91
7.3.1.9 TCU fill/drain ................................................................................................................................................. 91
7.3.1.10 Flow cell fill/drain .......................................................................................................................................... 91
7.3.1.11 Blood sensor status ......................................................................................................................................... 91
7.3.1.12 Reset statistics ................................................................................................................................................ 91
7.3.1.13 Hard drive maintenance .................................................................................................................................. 91
7.3.1.14 Show advanced DB view ................................................................................................................................ 92
7.3.1.15 Small sample mode......................................................................................................................................... 92
7.3.1.16 Pre-diluted mode ............................................................................................................................................ 92
7.3.1.17 Monitoring ...................................................................................................................................................... 92
7.3.2 Service testing ........................................................................................................................................ 92
7.3.3 Service calibration ................................................................................................................................. 93
7.3.3.1 Enable scatter calibration................................................................................................................................ 93
7.3.3.2 Backup Calibration ......................................................................................................................................... 93
7.3.3.3 Restore Calibration ......................................................................................................................................... 93
7.3.3.4 Load Calibration ............................................................................................................................................. 93
7.3.3.5 Accept User factors ........................................................................................................................................ 94
7.3.3.6 Back................................................................................................................................................................ 94
7.3.3.7 Accept............................................................................................................................................................. 94
7.3.4 Stress measure ........................................................................................................................................ 94
7.3.5 Auto alignment ....................................................................................................................................... 95
7.3.6 AS (Auto-sampler) .................................................................................................................................. 95
7.3.6.1 Go to Pos ........................................................................................................................................................ 95
7.3.6.2 Reset ............................................................................................................................................................... 95
7.3.6.3 Repeat last command ...................................................................................................................................... 95
7.3.6.4 Go To Mixer ................................................................................................................................................... 95
7.3.6.5 Load to sampling position .............................................................................................................................. 96
7.3.6.6 Set up Barcode Reader ................................................................................................................................... 96
7.3.7 Adjustments ............................................................................................................................................ 96
7.3.8 Multiuser Settings ................................................................................................................................... 97
7.3.9 MDA view ............................................................................................................................................... 97
7.3.10 Software upgrade ............................................................................................................................... 98
7.3.10.1 Refresh data .................................................................................................................................................... 98
7.3.10.2 Change High Level software .......................................................................................................................... 98
7.3.10.3 Change Low level software ............................................................................................................................ 98
7.3.10.4 Change Low level boot ................................................................................................................................... 98
7.3.10.5 Change LSDACQ firmware ........................................................................................................................... 98
7.3.10.6 Change Optical Head Firmware ..................................................................................................................... 98
7.3.10.7 Change TCU software .................................................................................................................................... 98
7.3.10.8 Change AutoSampler software ....................................................................................................................... 99
7.3.11 Reagent Lock...................................................................................................................................... 99
7.3.12 QC Wizard ....................................................................................................................................... 100
7.3.13 Printer installation ........................................................................................................................... 100
7.3.14 Factory Settings ............................................................................................................................... 100
7.3.15 Service mode OFF ........................................................................................................................... 100
7.4 SOFTWARE SYSTEMS ................................................................................................................................. 101
7.5 GENERAL SW INSTALLATION GUIDE ......................................................................................................... 101
7.5.1 Install operating system Windows 2.0 install using pendrive............................................................... 103
7.5.2 Installing a Printer ............................................................................................................................... 107
5

8 INSTALLATION ............................................................................................................................................ 108


8.1 CHECK THE DELIVERY .............................................................................................................................. 108
8.2 PREPARE FOR INITIAL INSTALLATION ....................................................................................................... 108
8.2.1 Select a Suitable Location .................................................................................................................... 108
8.2.2 Make Any Special Arrangements.......................................................................................................... 108
8.2.3 Gather Your Peripherals Devices ........................................................................................................ 109
8.3 PERFORMING THE INSTALLATION .............................................................................................................. 109
8.3.1 Move the ‘Abacus 5’ to the Selected Location ..................................................................................... 109
8.3.2 Visual Inspection .................................................................................................................................. 109
8.3.3 Remove the Protective Card from the Shear Valve .............................................................................. 110
8.3.4 Connect the Optional Auto-Sampler .................................................................................................... 110
8.3.5 Connect the Reagents ........................................................................................................................... 111
8.3.6 Connect the Power Cord ...................................................................................................................... 112
8.3.7 Verify the ‘Abacus 5’ Computer Operation .......................................................................................... 112
8.3.8 Connect Peripherals ............................................................................................................................. 113
8.3.9 How to install a printer ........................................................................................................................ 113
8.3.10 Initializing the Optional Autosampler .............................................................................................. 114
8.3.11 Using the Settings Menu .................................................................................................................. 115
8.3.12 Set Up a Laboratory Information System (LIS) ............................................................................... 116
8.3.13 Set Up a Serial LIS Connection ....................................................................................................... 117
8.3.14 Set Up an Ethernet LIS Connection ................................................................................................. 117
8.3.15 Running Blank Samples and Blood Samples for the First Time ....................................................... 118
8.4 CHECKLIST FOR INSTALLING ABACUS 5 HEMATOLOGY ANALYZER ........................................................... 119
9 TROUBLESHOOTING .................................................................................................................................. 120
9.1 TROUBLESHOOTING CHECKLIST FOR ABACUS 5 HEMATOLOGY ANALYZER ....................................................... 120
9.1.1 Visual inspection .................................................................................................................................. 120
9.1.2 Powering up the system ........................................................................................................................ 121
9.1.3 Checking the software adjustments ...................................................................................................... 121
9.1.4 Checking the Hardware adjustments.................................................................................................... 122
9.1.5 Run a Self-test ...................................................................................................................................... 124
9.1.6 Run Blank measurements ..................................................................................................................... 124
9.1.7 Run Control measurements .................................................................................................................. 125
9.2 MECHANICAL PROBLEMS........................................................................................................................... 126
9.2.1 General guidelines to overcome motor or moving part related problems............................................ 126
9.2.2 Sample Rotor (SR or BOB) failures ..................................................................................................... 126
9.2.2.1 SR gives grinding noise and / or SW displays SR error messages ................................................................ 126
9.2.2.2 SR error appears during initialization process: ............................................................................................. 126
9.2.2.3 The SR does not turn into the analyzer even with open front panel.............................................................. 126
9.2.3 Needle mechanics, Vertical motor (MVert) problems .......................................................................... 127
9.2.3.1 The needle carriage keeps dropping back (down) at initialization ................................................................ 127
9.2.3.2 MVert cannot reach the optosensor (Up or Down) ....................................................................................... 127
9.2.4 Shear Valve (SV) related errors ........................................................................................................... 127
9.2.4.1 SV error at the first startup ........................................................................................................................... 127
9.2.4.2 Grinding noise after SV cleaning, (after SV reinstallation) .......................................................................... 128
9.2.4.3 SV leakage.................................................................................................................................................... 128
9.2.5 Dilutor errors ....................................................................................................................................... 128
9.2.5.1 Dilutor is making noise................................................................................................................................. 128
9.2.5.2 Dilutor is leaking .......................................................................................................................................... 128
9.2.6 A tube comes off of a valve ................................................................................................................... 129
9.2.7 Priming problems ................................................................................................................................. 129
9.2.7.1 The analyzer would not prime liquids .......................................................................................................... 129
9.3 ELECTRONICS RELATED PROBLEMS ........................................................................................................... 130
9.3.1 Touch screen / display errors ............................................................................................................... 130
9.3.1.1 No image on display ..................................................................................................................................... 130
9.3.1.2 Touch sensitive surface not working ............................................................................................................ 130
9.3.1.3 Touch (click) is inaccurate ........................................................................................................................... 130
6

9.3.1.4 Screen image is not normal, some parts of the software screen is not visible ............................................... 130
9.3.2 Communication errors ......................................................................................................................... 130
9.3.3 The analyzer does not power on ........................................................................................................... 131
9.3.4 I2C errors displayed at startup ............................................................................................................. 131
9.4 USEFUL INFORMATION AND TIPS ............................................................................................................... 132
9.4.1 High PLT background .......................................................................................................................... 132
9.4.2 Fast Starting up and Shutting down ..................................................................................................... 132
9.4.3 Removing bubbles from the Flow cell .................................................................................................. 132
9.4.4 Removing the DimmBoard ................................................................................................................... 132
9.4.5 Listen the sounds of the instrument ...................................................................................................... 133
9.4.6 Replacing small diameter tubes ........................................................................................................... 133
9.4.7 Swapping the pumps ............................................................................................................................. 133
9.4.8 Draining the reservoirs ........................................................................................................................ 133
10 MAINTENANCE ............................................................................................................................................ 134
10.1 USER MAINTENANCE ................................................................................................................................. 134
10.1.1 Daily maintenance ........................................................................................................................... 134
10.1.2 Weekly maintenance......................................................................................................................... 134
10.1.2.1 Shear Valve cleaning .................................................................................................................................... 134
10.1.2.2 Cleaning the wash head ................................................................................................................................ 134
10.2 PREVENTIVE MAINTENANCE ..................................................................................................................... 134
10.3 FLOW CELL CLEANING ............................................................................................................................... 141
11 REMOVAL AND REPLACEMENT PROCEDURES ................................................................................................. 144
11.1 OPENING THE INSTRUMENT ....................................................................................................................... 144
11.2 SHEAR VALVE ASSEMBLY ......................................................................................................................... 145
11.3 XY UNIT ................................................................................................................................................... 146
11.4 SAMPLE ROTOR ......................................................................................................................................... 149
11.5 DILUTORS .................................................................................................................................................. 150
11.5.1 Syringe replacement......................................................................................................................... 151
11.6 TCU MODULE ........................................................................................................................................... 151
11.7 VALVE BLOCK ........................................................................................................................................... 153
11.8 PRE-AMPLIFIER BOARD............................................................................................................................. 154
11.9 PUMP ASSEMBLY....................................................................................................................................... 154
11.10 HARDWARE MODULE ............................................................................................................................ 156
11.11 LASER HEAD ASSEMBLY....................................................................................................................... 157
11.11.1 Laser injector replacement procedure ............................................................................................. 160
11.11.1.1 Removing the laser for injector replacement: .......................................................................................... 161
11.11.1.2 Removing the injector: ............................................................................................................................ 162
11.11.1.3 Reinstalling the Laser Head: .................................................................................................................... 164
11.12 MEASURING BLOCK .............................................................................................................................. 164
11.12.1 Aperture removal: ............................................................................................................................ 165
11.12.2 WBC aperture removal: ................................................................................................................... 165
11.12.3 Measuring Chamber removal: ......................................................................................................... 166
11.12.4 HGB head removal .......................................................................................................................... 166
11.13 DISPLAY UNIT ....................................................................................................................................... 167
11.14 ABACUS5 FLUIDICS V2.1 - SV IN CHAMBER POSITION .......................................................................... 168
11.15 ABACUS5 FLUIDICS V2.1 - SV IN NEEDLE POSITION ............................................................................. 169
11.16 TUBINGS ............................................................................................................................................... 170
12 APPENDIX ...................................................................................................................................................... 181
12.1 RECOMMENDED KIT OF TOOLS................................................................................................................... 181
12.2 HOW TO SEND .RP FILES TO ASSESSMENT OF ANALYZER PERFORMANCE.................................................... 181
12.3 HOW TO USE THE „COLLECT” FUNCTION OF THE ABACUS 5 HEMATOLOGY ANALYZER ............................. 182
7

1 INTRODUCTION
1.1 Who should read this manual
This Service manual is intended for trained technicians or service engineers to be able to maintain and
repair Abacus 5 automated hematology analyzer. It contains the functional descriptions of the analyzer,
operation of the fluidic system, adjustments and settings, and very important information for the Service
Personnel about the service operations and possible problems.

If equipment operation is different from the manufacturer’s specifications and intended


use, the protection provided by the equipment may be impaired.
Misuse of equipment or use other than its intended purpose will invalidate conditions
of warranty. Accuracy and precision may also be impaired.
Follow the service procedures described herein to guarantee functional operation after
repairs and maintenance.

All information contained herein is the intellectual property of Diatron MI PLC and should not be used or
reproduced without prior agreement of Diatron MI PLC, the manufacturer.

Would you find things unclear, please, contact support@diatron.com for assistance.

For updated information regarding SW updates and Technical bulletins, please, visit our web site:
www.diatron.com

If you are a registered partner of Diatron, you can get a unique password from your sales department to
access download page on our web site.

User’s manual contains important information about the operation and measurement principles of
Abacus 5. Specifications, definition of parameters and user interface are also described therein,
consequently Service manual does not contain these data.

Use of the Service manual assumes knowledge of the information contained in the User’s manual.

Diatron Technical Support Team


8

2 For Your Safety


2.1 Special Symbols Used In This Manual

Label Meaning Explanation

Blood samples and analyzer waste are potentially infectious


Biohazard
materials.

Corrosive Reagents may cause corrosion or skin irritation.

Warning General warning of possible hazard conditions.

Sharp needle warning The sampling needle may be a hazard to the operator.

2.2 General Precautions

The sampling needle and other components inside the analyzer may cause injury, or can
get damaged if handled incorrectly. Only certified personnel should open the covers.
Running measurements with opened cover is not recommended due to the risk of possible
injury. Always wear safety gloves while performing maintenance actions.

The analyzer weighs 35kg (~77lbs). Please do not attempt to move it alone. The analyzer
should always be moved by two persons holding the analyzer by its sides in an upright
position.

Always use safe lifting procedures when lifting the analyzer.

Make sure to retain the original packaging material for safe transportation and storage in
the future.

To prepare the analyzer for shipping, storage or extended periods of inactivity, please drain
the reagents and repackage the Abacus 5 in its original packaging. Do not expose the
Abacus 5 to direct sunlight, extreme temperature or humidity (>80%).

The analyzer operates with chemically and biologically active reagents. Physical contact
with these reagents should be avoided. Please read reagent descriptions carefully for
possible emergency actions.
9

To ensure reliable operation and reliable results:


 Only genuine Diatron reagents should be used
 Required service maintenance should be performed as recommended in this
manual
 Only genuine Diatron service materials and spare parts should be used for repairs

Genuine reagents and service materials and spare parts are available from Diatron.

Only Diatron certified service personnel that have successfully completed the ‘Abacus 5
Service Training’ program are qualified to service the Abacus 5 analyzer.

Before operating Abacus 5 analyzer, all operators should complete an ‘Abacus 5 Operator
Training’ program. This program is offered by Diatron or by Diatron certified service
personnel.

Replacement materials or spare parts (tubes, valves, etc.) which might have been in
contact with human blood or reagents should be handled as a potentially biologically
hazardous and chemically dangerous material. All applicable laws and regulations must be
observed in the handling and disposal of these materials.

Abacus 5 is designed for laboratory operation. Mobile operation is not supported. Operate
Abacus 5 within the ambient temperature range described in User Manual section 2.4.

The IVD equipment complies with the emission and immunity requirements described in
relevant part of the IEC 61326 series.

This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference, in which case, you may need to take
measures to mitigate the interference.

Electromagnetic environment should be evaluated prior to operation of the device.

This analyzer contains electronic components. Please handle electronic waste


adhering to local or federal regulations.

CAUTION – Use of controls or adjustments or perfomance of procedures other than those


specified herein may result hazardous radiation exposure.

2.3 Environmental Factors


Operate Abacus 5 within the ambient temperature range of 15-30°C (59-86°F) and a relative humidity
range of 10% - 80%. Optimum operating temperature is 25°C (77°F).
10

Abacus 5 should be stored within the temperature range of 5-35°C (41-95°F). Avoid exposing the
analyzer to direct sunlight or to extreme high or low temperatures. If the analyzer was subjected to
extreme temperatures during shipment or storage, it must be placed for at least two hours in a room
whose temperature is within the operational range before installation or use.

Reagents should be stored at a temperature range of 15-30°C (59-86°F). Reagents may experience a
temperature range of at most 5-35°C (41-95 °F) for a maximum of 3 days.

The analyzer should be placed in a well-ventilated location. Operation at an altitude above 2000 meters
(6560 ft) is not guaranteed.

2.4 Electrical Requirements


The analyzer should only be operated from a wall outlet meeting these power input requirements:

 100-127VAC or 200-240VAC; 47Hz to 63Hz


 Power Consumption: maximum 400 VA (including power of the auto-sampler)
Please ensure that the wall outlet is also capable of supplying the power consumption of any additional
devices (such as a printer).

Use only the power cord supplied with the instrument. Avoid using extension cords. Abacus 5 comes
with a power cord appropriate for your power system. Proper use of the appropriate power cord assures
adequate grounding of the system. If the power is not reliable, contact your representative for options
such as the installation of an external UPS module.

Failure to properly ground the Abacus 5 bypasses important safety features and
may result in electrical hazard.

The instrument should not be placed near potentially interfering devices capable of emitting radio
frequencies (e.g. radio or television transmitters/receivers, radars, centrifuges, X-ray devices, fans, etc.).

This analyzer is designed to be safe for transient voltages to INSTALLATION CATEGORY II and POLLUTION
DEGREE 2.

2.5 Suggested on-line UPS


It is recommended to use a UPS (uninterruptible power supply) in line with the power input in the
following circumstances:

 if the main power is fluctuating


 in case of power black out happens frequently
11

Power consumption of the analyzer and optional Auto-sampler is 400VA max. If there is a power failure,
the analyzer should be able to complete the function in progress and perform a Windows shut-down to
secure data.

In order to provide the necessary power, it is recommended to use a UPS with the following minimum
specifications: 600 VA provides uninterrupted power for 10-15 minutes.

Diatron does not supply UPS for Abacus 5. Therefore, Diatron is not taking responsibility or
liability for the operation of Abacus 5 running on UPS.
If you need further information, please, contact technical support of the UPS vendor.

2.6 Proper Placement of Abacus 5


It is important to install the instrument in a suitable location, a poor location can adversely affect its
performance:
 Select a well-ventilated location near a power source and close to a suitable drain.
 Place the unit on a clean, level surface. Leave at least 0.5 m (18 inches) space on both sides and
above the instrument to access pneumatics.
 A minimum of 0.2 m (8 inches) must be maintained between the rear panel and the wall to allow
for heat dissipation and tubing clearance.
 Ensure there is enough clearance in front of the analyzer to open front panel. Allow enough
space if you want to use optional external keyboard, mouse or bar code reader.
 Your selected location should allow placement of the reagents in an unobtrusive location below
the laboratory bench that the instrument is placed on. Placement below the laboratory bench
also allows for storage of a spare set of reagents. Never place the reagents above the Abacus 5
analyzer.

Placing reagents above the Abacus 5 analyzer could result in reagent overflow and spilling.
Always put the diluent tank on the floor, never to the same level where the analyzer is.
12

1. Figure: Spacing of Abacus5 with Auto-Sampler and reagents

2.7 Weight Requirements


Abacus 5 weighs 35 Kg (77 lb) without the Auto-sampler.

Abacus 5 with optional Auto-sampler weighs 47Kg (104 lb). Adding an external keyboard, printer,
documents, etc. can bring the total weight up to 60 Kg (132 lb).

Please select a table, laboratory shelf, or other location which can support the weight of the Abacus 5
with all accessories and it is free from vibration.

To allow reliable operation and to provide a safe working environment, make sure that
the table supporting the unit is stable enough to carry the weight of the instrument and
accessories.

2.8 Waste Disposal


Waste of Abacus 5 analyzer contains human blood and reagents that are chemically and biologically
active, and should be considered to be a potential infection and biohazard threat. Safe laboratory
practices must be followed including the use of personal protective when operating Abacus 5 and
handling blood, reagents, and waste.

System waste should be handled as a potentially bio hazardous material. All


applicable laws, regulations, and laboratory practices should be followed in the
handling and disposal of waste.
13

2.9 Emergency Situations


Always follow all applicable laws and regulations with regards to emergency situations.

If Abacus 5 needs to be powered off due to an emergency situation (like fire, thunderstorm, etc.), follow
the procedures in User Manual section 7.3.4.

In case of fire, do not use water to extinguish the fire unless Abacus 5 is disconnected from
the electrical network!
14

3 Structure of the analyzer


Please, refer to the User’s manual for details on opening the front and side covers, and parts contained
there.

3.1 Components located on the front panel


The display panel and related components are covered with a metal plate, to avoid electronic
interference causing problems, and to protect sensitive electronics.

3.1.1 Display screen and the touch sensitive surface


Abacus 5 uses a color LCD with LED backlight touchscreen as main user interface.

The built-in touchscreen is a 4 wire resistive type. It is interfaced to the mainboard by a USB controller.
15

3.1.2 Start Button and LEDs

START Button board is mounted on the front panel. It contains a start button and two LEDs to illuminate
the START button. The LEDs are bicolor: red and green LEDs. When both red and green are switched on,
the resultant color is yellow.

3.2 Components accessible after opening the front panel


3.2.1 Shear Valve Assembly
Shear Valve assembly consists of the following parts:

 Shear Valve holding plate,


 Moving mechanism,
 Stepper motor,
 Ceramic Shear Valve,
 Shear Valve optoboard.
16

3.2.2 Sample rotor


Abacus 5 hematology analyzer has a sample rotor for safety and more precise sample handling.

Sample rotor unit uses a stepper motor, connected to the PPB through the XY opto board. The rotor has
micro switches for positioning.

The unit blocks itself in the home and end position with mechanical parts and has a special cap that
prevents the damage of the electronic and mechanic parts caused by any fluid.

Tube adapter Micro switches for positioning

3.2.3 Main Dilutors


Abacus 5 has two separate main dilutor modules. There are two stepper motors, a common motor opto
board, four syringes and piston rods with gear transmission in each module.

The software identifies and moves four dilutors (Dil1, Dil2, Dil3, Dil4), each dilutor consists of two
syringes. Dilutor 1 and 2 is in Dilutor module 1, dilutor 3 and 4 is in Dilutor module 2, as represented in
the picture below.

Main Dilutor Module 1 Main Dilutor Module 2

Dil 1 Dil 2 Dil 3 Dil 4


17

3.2.4 Dilutor opto sensor boards


Dilutors have their own separated opto-boards, located
directly in the units. On the front side of the optoboards there
are 4 optosensors and 4 control LED-s, on the back side are
mounted the motor connectors and the flat cable connector.

3.2.5 Tube organizer


This is the component that provides arrangement for the tubes going from the
valves to the dilutor, shear valve, optical head and Temperature Control Unit. It is
intended to allow easy identification and access to tubes for service related
cleaning procedures.

Four tubes have metal through tubes to allow easier removal and replacement
when necessary in case the TCU needs to be cleaned and rinsed.

3.2.6 Temperature Control Unit


18

Temperature Control Unit provides the necessary temperature for reagents used to create 4-DIFF
sample. It is able to heat or cool the sample and reagents, depending on the ambient temperature. It
contains a massive, molded aluminum block. There are multiple, curved and interconnected stainless
steel tubes (fluid paths) inside the TCU to ensure proper capacity.

Parts of TCU:

 in-line mixer, for mixing and homogenizing sample and specific reagents.
 Temperature Controller Board, to monitor the temperature of the aluminum block by power
transistors (heating elements) or Peltier cooling circuitry
 thermal insulation for temperature stability.

3.2.7 Laser head


Laser head is used for optical measurements of 4-DIFF and BASO parameters.

3
1. LASERDRV BOARD

2. LENS ASSEMBLY

3. OPTSENSE BOARD

4. SAFETY SWITCHES
1
5. SAMPLE INJECTOR

6. OPTICAL FLOW CELL

7. OPTICAL CABLE

2 4
6 7
5

Black anodized aluminum box works as safety cover. Two different micro-switches protect the technician
against direct exposure to beam:
19

 When cover holder screws are removed laser activity will be switched off by laser driver board.
 To turn on laser again, put the cover back, and fasten the screws so that both switches are
closed and the low-level SW should be restarted.
Optical measurement unit has a sheath and sample inlet, and a waste outlet from fluidic side, laser driver
cable, analogue output cable and auto-alignment cable from the electronics.

3.2.8 Laser Head Assembly + Sample Injector


Laser Head Assembly is responsible for detecting the 4diff and BASO cells from the prepared blood
sample.

The laser head is responsible for the precise illumination of the sample. The temperature controlled laser
diode source is mounted on a huge brass basis which holds it tight and also responsible for the cooling.
Just beside the laser diode aspheric and achromatic lenses performs the focusing of the laser beam.

For accurate and stable adjustment, this optical


unit is mounted on an aluminum block with a
sphere-to-cone contact. Powerful springs hold the
unit in place. Using stainless steel levers, the
direction of the laser can be rotated around two
axes. This means, the laser spot on the flow-cell
can be tilted in horizontal and vertical directions.
Precise linear motion stepper motors perform the
accurate setting of the laser.

Coarse adjustment of the laser can be made by


setting the rough adjustment screws.

The flow-cell unit is responsible for the precise


flow control and the pre-detection of the pulses.
An optically clear flow-cell (cross section for flow is
0.25x0.25mm) is mounted with 2 side cone-to-
cone connection into its holder. Below the flow-
cell an injector helps to insert the sample in the
middle of the main sheath streamline. The sheath
puffer opened to free air supplies the sheath and
the sample flow.

Cross section of the flow cell – sample injector


assembly
The size and position of the sampler needle, the different tube resistance of the sheath and sampler lines
and the applied vacuum result an about 40nm wide sample stream in the middle of the flow-cell.
20

Concentric ring shaped optical cable is also mounted to the flow-cell unit. This collects the scattered light
from the cells, and transfers it to the detection unit. Just before the insertion zone of the optical cable,
there is a laser dump for filtering direct laser beam.

3.2.9 Laserdiode Driver Board


The LASERDRV board is based on a PIC 24FJ64GA004 (Microchip)
microcontroller that performs all the control functions of the laser
diode and the communication with the LS-DACQ (DimmBoard).

The LASERDVR board incorporates a laser diode with built-in APC


(automatic power control). The laser power can be set by a
programmable digital potentiometer in the 1.8 – 5.5 mW range.
The laser can be switched on and off from program.

The laser diode temperature can be set, and then the


microcontroller will keep the diode temperature at the preset
value. As the temperature control uses only heating (2 transistors),
stable diode temperature can be maintained only above room temperature.

The LASERDRV board provides laser safety functions too. When removing the laser cover, the safety
switches cut off laser power immediately. In order to turn the laser on the cover must be in replaced and
the instrument must be turned off and back on.

The controlling interface between the LS-DACQ and the LASERDRV is I2C.

3.2.10 Pin Photodiode and Amplifier (OPTSENSOR_2v1)


When a blood cell in the diluted and lysed
blood stream crosses the focused laser light,
the light scatters and two pin photodiodes
sense the scattered light. The current,
generated by the two photodiodes has to be
handled separately, so two independent
analog channels are used on the OPTSENSOR
board. The photodiode’s current is amplified
by one transimpedance amplifier per channel.

Then the DC level is removed, and more


amplification is applied. The DC level of
output 0 and 1 (AOUT0, AOUT1) are clamped to +1V. The output span is 2V, so the output range is
+1V..+3V. This is appropriate for the A/D converter on the LS-DACQ, that has a 2V input voltage span,
only the offset has to be set according to the A/D’s requirement (+1.5V..+3.5V). The 3rd analog channel
21

(AOUT2) outputs the DC level of channel 0, amplified by 2. It can be used for the auto alignment of the
laser. The OPTSENSOR card contains the connectors and LEDs for AutoAlignment motors, as well as
position sensing.

3.3 Left side


3.3.1 Valve boards
The Abacus 5 system incorporates 44 valves. The valve boards
are controlled by two Pneumatic and Power Board (PPB)
boards. The used valve boards are the following:

2x Valve_1-5
2x Valve_6-12
2x Valve_13-18
2x Valve_19-22

4 valves are not used, and thus not installed. Valve coils are not installed on valve driver boards behind
either.

Note: Valves can be replaced individually, also the coils can be unsoldered and replaced but valve board
cannot be replaced individually. We only provide the whole valve assembly 40 valves as a spare part.
22

3.3.2 WBC/BASO Preheater Assembly

WBC/BASO
preheater
assembly

The WBC/Baso preheater assembly is located on the right side of the impedance measurement block,
near the WBC/BASO counting chamber. It consists of two stainless steel holding plates, thermal
insulation, heater block (see illustration below) and electronic board with heating transistors.

Thermostated on 36-37 °C and its function is to heat the


WBC/BASO dilution for better lysing. Diluted blood and lyse
reagent is pushed through the tubing of the heater block,
diluent used for rinsing and cleaning the chamber
also passes through this assembly thus the chamber itself is
warmer than the ambient temperature.

3.3.3 Counting chamber with electrodes and measuring aperture


Impedance method is used for determination of volume and number of cells. In this method a known
volume of dilution is drawn through a small aperture. Constant current is passed through the aperture
from one side to the other. When a cell passes through the aperture, it causes a change in resistance,
which generates a voltage pulse. The amplitude of the voltage pulse is proportional to the ratio of cell
volume per aperture volume. This is used to determine the volume of cells. The number of cells can be
obtained by counting the pulses.

In the instruments there are two cell-counter chambers: separate for RBC and WBC.

In the RBC chamber the instruments counts red blood cells, and uses no lyse at all in this chamber. It has
a smaller draining outlet made of plastic and its measuring tube contains a 70 µm-sized aperture.
23

In the WBC chamber the instrument counts all kind of WBC. It has a measuring tube with an aperture
size of 80 µm and a bigger draining outlet made of PTFE (Teflon).

Both chambers have a reference electrode and a draining outlet. The next picture shows the chambers
and the measuring tubes. The aperture is made of ruby and it is molded into the measuring tube.

RBC Reference WBC


chamber electrode chamber

Reference
electrode

RBC measuring tube WBC measuring tube


with the aperture (70m) with the aperture (m)
(no grooving) (1 grooving) Aperture

O-ring

3.3.4 HGB Measuring Head


Hemoglobin head is placed around the WBC measuring chamber in the instrument.

It contains: a light source (LED) at 540 nm wavelength and Photo Detector (TSL235). The Photo Detector
converts the light to frequency. The HGB concentration is a logarithmic function of this frequency
measured by the FPGA circuit of the COMB card.
24

TSL235
LED

Connection to the
amplifier

The analyzer performs enhanced Hemoglobin measurement technology for HGB measurement. The
output of HGB head is frequency (TSL235 detector is light to frequency converter). A digital counter in
the FPGA circuit counts this signal.

This counter counts up while the LED is on and counts down while the LED is off, the LED and the counter
directions are switched with a 250 Hz signal. This method provides “real time backlight correction”,
which makes the HGB measurement more precise in changing backlight environment situation as well.

There are two kinds of HGB measurements:

 Sample measurement (before RBC counting)


 Diluent measurement (in WBC washing phase)

The HGB result is calculated from these measurements by:

HGB  log (CNTdiluent light / CNTsample light)

Due to backlight correction, Abacus 5 is less sensitive to incident light changes.

3.3.5 Cell counter Amplifier Board


Amplifier board includes its own voltage regulators, connection interfaces to HGB head, to high voltage
board and to LSDACQ board. There is a current generator circuit on this board, which works from 50V
measuring voltage (generated by the High Voltage Board) and the probe voltage (DC) is amplified with a
voltage follower (output: ELV). Nominal measuring current is 870 µA.

Connection to:
CSA1 on COMB

Connection to:
HVB

Connection to:
COMB (DIGIO)
25

Connection to
the electrodes

Offset
potentiometer

Connection to
HGB head

Amplifier board includes one input connector for each measuring chamber (measuring electrode). There
is one opto switch (OPT1) and a relay (REL1) to connect high voltage to one of the probes with HSW
signal and to isolate the input of the amplifier. Test circuit allows generating test pulses (with TEST and
PLS signals through Q1, Q2 FETs) for checking proper operation of each amplifier channel.

Amplifier board includes a 3-stage main amplifier channel, which gains input signal to the 0...5 V range
(this is the input range of the A/D converter (IC10), which is placed on the LSDACQ card). The RSW signal
(with Q8 transistor) changes the input electrode through REL2 relay. There is an offset potentiometer, P1
in the third amplifier stage, manufacturer sets the correct offset voltage.

The Amplifier offset adjusted automatically by the SW. The service personnel can adjust it manually by
giving the offset value through the SW in service menu. If the auto offset is cannot put the value in range
the amplifier board is defective needs to be replaced.

3.3.6 Pressure sensor board

1 2 3

The Pressure Sensor Board incorporates three differential pressure sensors. The pressure values are read
out by the LS-DACQ card through I2C interface. The sesors are responsible for reading pressures of:
26

(1) vacuum for impedance count (“chamber”)


(2) vacuum used for draining chambers (“drain”)
(3) big puffer vacuum for optical aspiration (“big puffer”)
Upon replacing the pressure meter board, it is recommended to adjust pressure meter offset

Abacus 5 is equipped with sensitive pressure meters. These pressure meters are used to monitor and
control measurement, and chamber draining processes. In some cases, the pressure meters can develop
an offset value that can cause “Timeout” or “Chamber draining” errors. The new analyzer software can
compensate for this offset value drift.

Locate the “Start adjustment” button in the lower part of the screen – and tap it. The SW will perform a
process, where the pressure meters will be “vented” to atmospheric pressure – and this value will be
noted by the SW. This will minimize the occurrence of the “Timeout” or “draining” related errors. (On the
screen you will see the actual value of the pressure meter, corrected values can be seen only if a Self-test
is performed after the adjustments).

If the above process fails…

1. Perform a Low Level Reboot (service menu, service functions) to get low level PC in a default
known state
2. Run Test Function 11
3. Low Level Reboot 2 times
4. Perform Pressure Sensor Offset function (Service Menu, Adjustments)
5. Perform a 3-measurements based Calibration using normal level control blood
27

3.3.7 Reagent and Vacuum buffers

The lower part of the assembly plate holds six plastic cylinders, called chambers or buffers. The last one
(marked as ‘1’ in the image) holds vacuum used in the optical measurement process (for moving the
solution). This vacuum is always adjusted according to measured atmospheric pressure.

The four next chambers are used as temporary storage volumes for individual reagents for one
measurement cycle. The first two (in the front, marked as ‘5-6’ are linked in parallel to double the
capacity. Chambers 2-5 have internal floating (magnetic) level sensors. The volume of chambers 2-6
allow one measurement cycle to be performed after the reagent low warning.

3.3.8 Reagent Sensor Board

The Reagent Sensor Board monitors the liquid level in the reagent puffers continuously. If the liquid level
too high (puffer full), it signals to the PPB2 board and the software can stop the filling process.

3.3.9 Opening the valve assembly plate


The plate holding the valves and the measurement block can be folded out of the analyzer. The plate is
secured with 2 screws, similar to the ones securing the side panels. The screws are designed to stay on
the analyzer to avoid losing them.

3.3.10 Vacuum buffer


The glass chamber is located on the internal side of the assembly plate, and is used to store the vacuum
for the optical measurements.
28

3.3.11 Pneumatic and Power Boards (PPB1 and PPB2)

PPB card contains the main power regulator circuits, valve and motor driver circuits and other
connections for the fluidic and pneumatic system’s parts.

Power system generates +5V (Digital power), +8V (Printer power) and +12V (Motor and valve power)
from the single +12V DC input signal.

Motor driver part consists of six separated PIC micro-controllers with power drivers. Horizontal, Vertical
and Sample rotor motors have one combined ribbon cable connection. Main Dilutor (with two motors)
and Micro-dilutor have separated connectors.

Valve driver section is based on the valve driver PIC micro-controller and three 8-bit, powered output
shift registers (with built in protection diodes) and there are two common ribbon cable connections for
the 4 valve boards. The pump assembly has a separated Darlington driver circuit for more reliable
operation.

All microcontrollers have 2 LEDs: a yellow one and a green one.


29

The yellow one indicates motor moving or holding and active valve or pump moving. (it means current
flows into motors, valves or pump)

The green one has 3 states:

 dark: (after initialization phase) error state,

 blinking: communication in progress - normal state

 on (just lighting): OK - normal state

PPB boards have a small board on the bottom called PPB CON board. This board connects the PPB to the
LSDACQ board.

3.3.12 Pump assembly


Pump assembly generates regulated vacuum and drains the fluidic system. There are two pumps in the
system. They are connected to the two PPB boards.

For further information see chapter 11.9.


30

3.4 Right side


3.4.1 XY unit
This unit contains slides to move the sample sampling needle
in Horizontal and Vertical directions, two stepper motors,
XYR opto board, opto wheel, washing head and the sampling
needle. It moves the needle to the desired position: from
sampling position, to washing head, and by means of the
washing head it press the sample tube during the sampling
process .

Both stepper motors have optical end-switch sensors for detecting these positions. These are required
for correct initialization and error detection. All sensors have status LEDs to show actual conditions.

The Vertical motor works with a special opto wheel for detecting home & end positions. See the
Adjustment section of this manual to place this wheel to the proper position.

Greasing of the horizontal/vertical guiding rods should be done regularly using photolube A598.

It is recommended to check and repeat greasing of guiding rods every year, or after 30000
measurements. See section 6.1.1.1.
31

3.4.2 XYROpto Board


Horizontal and Vertical motors and the Sample Rotor unit have a common Opto-board.

Sample Rotor Motor


opto LEDs connections

Horizontal Horizontal Vertical status Vertical opto


opto sensors status LEDs LEDs sensors

The other (rear) side of the board contains the connection for the Sample rotor and a ribbon cable
connection to PPB#

3.4.3 Sampling needle


Sampling needle is assembled in the H&V moving unit and it makes the piercing and the sample
aspirations. Correct setting of sampling needle is necessary and very important (see Chapter 6.
Adjustments).

Be careful, the needle is very sharp and it can cause injury!

3.4.4 Blood sensor


Blood detector is a component that will allow determining if the sampling process was successful. It is
measuring the length of the sample. If the sample is found short, a “sampling error” is going to be
displayed. It also has an emergency mode in case the blood detector failed. In this case, the analyzer is
not going to consider blood detector data, but it will move the sample to a predefined position. Both
parts of the sampling control must be adjusted.
32

3.4.5 Wash head


Wash head is located at the bottom of the XY unit and it is for cleaning the
outer surface of the sampling needle. This washing process is made with
diluent reagent and the fluid is drained by the pump. The arrows on the
picture show the direction of diluent flow during sampling needle washing.

Replace washing according Preventive Maintenance procedure .

3.4.6 Processor unit (LS-DACQ board with DimmBoard)


The LS-DACQ board is based on a credit card size embedded PC (DimmBoard). The LS-DACQ implements
the following functions:

- Receiving commands from the Analytical Unit,


- Blood sampling and sample handling control,
- Motor and valve control,
- Measurements control,
- Amplifying and A/D converting of 4 input channels simultaneously,
- Data preprocessing,
- Transmitting data to the Analytical Unit,
- Interfacing the Laser Driver and the TCU,
- Interfacing the Start Button and Status LEDs,
33

The LSDACQ board incorporates a credit-card sized PC, named DimmBoard. The processor on the
DimmBoard is a 600MHz Pentium-class core, with 256Mbytes on-board RAM, and SD card controller.
This SD card contains the Low Level SW of the system which handles all the measuring processes. The
Low level SW can be updated from the service software update menu.

Flash (BIOS)

Vortex86DX
CPU

Edge connector
MicroSD
Card

Flash (BIOS)

FPGA interface

The FPGA is connected through the ISA bus of the DimmBoard. The FPGA implements several registers,
and the DimmBoard can reach these registers as memory-mapped.

Buffer memory

During optical measurement, a vast amount of data has to be transferred to the Analytical Unit through
the USB connection. It is necessary to use a buffer memory to store the data temporarily, because the
USB transfer speed may not be enough to transfer the data real-time. The LS-DACQ uses a 2 MB SRAM
memory as buffer. The SRAM is organized as FIFO (first-in first-out memory).
34

Connection with the Analytical Unit

The Analytical Unit is connected to the Data Acquisition System by a full-speed USB interface. The USB
interface is implemented by a FT2232L USB chip. The USB chip implements 2 channels, one of them
works as a virtual COM port, the other is a parallel data channel. The VCOM channel is used as a
command channel, through which the Analytical Unit sends commands to the Data Acquisition System.
The measurement data are sent to the Analytical Unit through the parallel data channel.

FPGA

The FPGA is a Xilinx Spartan II type. The FPGA preprocesses the digitized data of the optical
measurement and sends them to the Analytical Unit through the USB. It preprocesses the volumetric
impedance measurement data too, produces data packets, and sends them to the Analytical unit.

FPGA controls the SRAM, to make a FIFO data buffer of it. Implements an I2C interface to control the PPB
(Pneumatic and Power Board) boards.

Implements a MDA display for service purposes.

FPGA configuration

As the FPGA is SRAM based, it is configured after every power on. The program is stored in a Xilinx
configuration flash memory. The flash memory can be programmed through a JTAG port by a Xilinx
Parallel Cable or in-circuit from the Abacus 5 program.

Analog inputs

The LS-DACQ has 4 analog inputs (AIN0-AIN3). AIN0 and AIN1 for the 2 channel optical measurement,
AIN2 is for the auto alignment, AIN3 is for the volumetric impedance measurement. The input signal
range for AIN0 and AIN1 is 1V..+3V, the DC offset is programmable, to adapt it to the A/D converter’s
input level of +1.5V..+3.5V. The input signal range for AIN2 and AIN3 is 0V..+5V, and the gain is
programmable.

A/D converter

The A/D converter is a THS1007 type, four channel A/D, manufactured by Texas. The input voltage level
is +1.5V..+3.5V, so AIN0 and AIN1 can be connected after DC level setting, but AIN2 and AIN3 channels
require not only level conversion, but attenuation. The input amplifier and level converter perform these
functions. The sampling frequency is 1 MHz on all input channels.

Temperature and Power Voltage Measurement, I2C Interfaces

The LS-DACQ card incorporates a PIC microcontroller that is connected to the DimmBoard through the
FPGA. The PIC with a built-in I2C controller controls the Laser Driver Board, the Opto Sensor Board, the
Pressure board and the TCU.
35

The PIC measures the board temperature and there is an input for measuring an external temperature.
The PIC measures the board power voltages and the DimmBoard battery voltage.

Connectors

POWER Connector

The LS-DACQ board is powered directly by the PC power supply through a standard IDE power
connector. It supplies the board with +5V and +12V.

Power voltages on the LS-DACQ board:

+12V – Provides +12V to the Cell Counter Amplifier board (AJ5-MEAS), Pin Photodiode and Amplifier
board (OPTSENSOR), High Voltage Board (AJ-HVB) and the Laser Driver board (LASERDRV).

-12V – Generated by a DC-DC converter. Not used on the LS-DACQ, output to the Cell Counter Amplifier
board (AJ5-MEAS) and the Pin Photodiode and Amplifier board (OPTSENSOR).

+5V – Supply voltage to the DimmBoard, and to some other logic.

+3V3 – It is generated from the +5V by a low dropout voltage regulator. Supply voltage to the 3.3V logic,
among them to the FPGA IO pins.

+2V5 – It is generated from the +5V by a low dropout voltage regulator. FPGA core voltage.

PPB Connector

The PPB connector connects the 2 Pneumatic and Power Boards (PPB) to the LS-DACQ board.

DIGIT IO Connector

The Cell Counter Amplifier board (AJ5-MEAS) is connected to the LS-DACQ board through the DIGIT IO
Connector.

HVB Connector

The HVB connector connects the High Voltage Board (AJ-HVB).

ANALOG INPUT Connector

The ANALOG INPUT connector connects the Pin Photodiode and Amplifier board (OPTSENSOR) to the LS-
DACQ board.

AINCH2 Connector

The AINCH2 connector is not used in the actual design.


36

AINCH3 Connector

The AINCH3 connector connects the analog output of the Cell Counter Amplifier board (AJ5-MEAS) to the
LS-DACQ board.

XILINX JTAG Connector (not used on the field)

The FPGA configuration flash memory can be programmed through this connector by a Xilix Parallel
Cable.

DEBUG DISPLAY Connector (not used on the field)

Only for test and debug purposes.

LASER DRIVER Connector

The LS-DACQ board provides power and an I2C interface to the LASERDRV board through the LASER
DRIVER connector.

PRESSURE Connector

The LS-DACQ board provides power and an I2C interface to the PRESSMEAS board through the PRESSURE
connector.

TCU Connector

The LS-DACQ board provides an I2C interface to the TEMPCTRL board through the TCU connector. As the
TCU requires high current power, the TCU is powered directly by an IDE connector of the PC power
supply.

FRONT PANEL Connector

The FRONT PANEL connector connects the Front Panel board (STARTBUT) to the LS-DACQ board.

FLOPPY Connector

A standard 3.5” floppy drive can be connected. It is used only for test and debug purposes.

KEYBOARD Connector

A standard PS2 keyboard can be connected to the DimmBoard’s keyboard interface by this connector. It
is used only for test and debug purposes.

COM1 Connector

The DimmBoard’s COM1 port is output here. The Auto Sampler is connected to this port.
37

NTC TEMP Connector

An external temperature measuring NTC resistor can be connected, to measure external temperature.

USBA Connector

The DimmBoard’s USB upstream port is output here. It can be used for DimmBoard software upgrade.

USBB2 Connector

It is connected to the USB downstream port of the PIC microcontroller. Not used.

USBB1 Connector

It is the USB interface between the Data Acquisition System and the Analytical System. It is an USB
downstream connector.
38

3.4.7 Mini-ITX PC mother board (In Hardware block)

1. PS/2 mouse port (green). This port is for a PS/2 mouse.

2. Serial connector. This 9-pin COM1 port is for serial devices.

3. DVI connector. This 24-pin DVI port connects to a DVI monitor.

4 & 5. LAN (RJ-45) port. This port allows Gigabit connection to a Local Area Network (LAN) through a
network hub.

6. Line In port (light blue). This port connects a tape, CD, DVD player, or other audio sources.

7. Line Out port (lime). This port connects a headphone or a speaker. In 4-channel, 6-channel, and 8-
channel configuration, the function of this port becomes Front Speaker Out.

8. Microphone port (pink). This port connects a microphone.

9. USB 2.0 ports 3 and 4. These two 4-pin Universal Serial Bus (USB) ports are available for connecting
USB 2.0 devices.

10. USB 2.0 ports 1 and 2. These two 4-pin Universal Serial Bus (USB) ports are available for connecting
USB 2.0 devices.

11. VGA port. This 15-pin VGA port connects to a VGA monitor.

12. Serial connector. This 9-pin COM2 port is for serial devices.

13. PS/2 keyboard port (purple). This port is for a PS/2 keyboard.
39

4 Electronic block diagram


The analyzer hardware consists of two systems, an Analytical Unit and a Data Acquisition Unit. The two
autonomous systems communicate via an USB interface. A 400W ATX PC power supply generates the
necessary voltages for the two units.

4.1 Hardware Block


The function of the Analytical Unit is to implement the user interface, to start and control the
measurement processes, to receive and process the measurement data, handle the database, display,
store, and print the processed measurement data. Hardware Block is responsible to communicate with
the user, transfer data to the Low Level Software and store the settings and measured data.

The components of the Analytical Unit are:

- PC mainboard (mini-ITX mainboard).


- Mass storage device (250 GB Winchester)
- Power supply (400W)
- Audio amplifier and speaker
- Keyboard (optional)
- Mouse (optional)
- USB CD/DVD drive (optional)
- Mainboard back panel I/O ports
- LVDS cable
- Touch cable
- Internal USB cable
- LED display cable
40

10.4” LCD-
LED 600x800

Touch screen Data and Power connector Back light connector


connector

Touch screen USB


interface board Power supply

USB Power & Video Power supply connector

Mini-ITX main board

PS2 PS2 Mouse


Keyboard IDE 4 USB Audio USB

Keyboard Mouse 3.5” 250GB CD-DVD or Audio


optional HDD other eternal amplifier
optional device like board
printer
(optional)

Speaker

Data acquisition
system
41

4.2 Data Acquisition Unit


This unit executes the commands of the Analytical Unit. It controls all the processes of the sampling,
sample actuation, motor and valve control, measurement control and the data acquisition. It
preprocesses the raw sampled data and forwards it to the analytical unit.

The components controlled by the Data Acquisition Unit are:

- Laser diode driver and diode


- PIN photodiode and amplifier
- Cell counter amplifier and HGB head
- High voltage board (HVB)
- Pressure sensor
- TCU (Temperature Control Unit)
- 2 PPB boards (Pneumatic and Power Board)
- Stepper motor units with opto boards (dilutors, X-Y module, shear valve)
- 40 Valves
- 2 Pumps
- Reagent sensor board

See Data Acquisition “map” on the next page:


42

Reagent Sensor
Auto Alignment
27-33 valves

23-26 valves

12-16 valves

38-44 valves

17-22 valves

34-37 valves

Shear Valve
X-Y module
6-11 valves
1-5 valves

Dilutor 2
Dilutor1
Pump1

Pump2
I2C PPB1 I2C PPB2

PPBIF LS-DACQ board Impedance measurement


(I2C) Analog
connector Cell chamber +
counter electrode
amplifier
DIGITIO HGB head
connector

HVB HVB +50V


JTAG
connector board +150V
connector
(FW upgrade)
I2C Pressure sensor
Parallel port
(debug) I2C TCU

Analog
Floppy IF DimmBoard
connector Pin photo diode + amplifier
Debug

Laser driver
USBA Laser driver + laser diode
connector

PS2 Optical measurement


COM1 Start Button USBB
keyboard Connector

Autosampler HW
(optional) block
Start Button
+ LEDs
43

4.3 Amplifier board


This unit is responsible to gather information during impedance measurements. The amplifier board
receives the signal of cells gathered by the aperture and gains these analog signals before transferring it
to the LSDACQ.

The amplifier is using 50V DC for measurement and 150V DC for cleaning apertures. These voltages are
generated by the HVB (High Voltage Board).

Amplifier unit has a test function. Test generator switches pulses to the input of operation amplifiers
during the automated Self-test program.

HGB signals are distributed by Amplifier unit.


44

4.4 Optical Head


Optical head provide WBC differential result applying a hydrodynamic focused flow cell, a laser light,
light detectors and amplifiers to gain the detector signals.

The laser driver unit has two Safety switches on the Laser head cover fixing to turn off the laser light
when the cover is removed. For safety reason the laser diode remains off until the low level system or
the whole system is rebooted.

Three analog signals are provided on the output of Laser Amplifier board. The low angle is typical of the
size of white blood cells, the high angle is typical of the size of the cell complexity, DC is dependent on of
cleanness flow cell.
45

5 OPERATION OF THE FLUIDIC SYSTEM


This section describes the main fluidic steps of Abacus 5 measurement cycle. The following figures show
total measurement flow diagram and detailed descriptions of processes for understanding the fluidic
system work.

In Abacus 5 the cleaning process executed parallel to the measures and the standby process are
executed in the background. It means that the database and other functions (except pneumatic) are
accessible while the analyzer is performing the measurement cycle and while it is going to standby.

The cleaning process does not block the next measurement cycle, so after getting the results the next
measurement can be started.

5.1 The Reagent system


Name Description Function
Diatro•Dil 5P diluent Isotonic solution, used to dilute whole dilution of sample, rinsing
blood and quantitative and qualitative and cleaning the tubing
determination of RBC, WBC, PLT and HGB system; it is the sheath
concentration fluidic
Diatro•Lyse - 5P Reagent for stromatolysis of RBC and BASO and 4-part (LYM,
quantitative determination of WBC 5-part MON, NEU, EOS) lysing
differentiation (LYM, MON, NEU, EOS,
BAS) and HGB concentration
measurement of human blood.
Diatro•Diff - 5P Quantitative determination of leukocyte responsible for timed 4-
five-part differentiation (LYM, MON, NEU, Diff reaction
EOS, BAS)
Diatro• Hypocleaner CC Capillaries, tubing and chambers, Cleaning system (external
removing blood component precipitates. liquid, not part of standard
reagent system)
46

5.2 Flow diagram of measurement

START BLANK START Meas.

Needle to SR Sample in

Needle piercing

Cleaning processes Sampling Preparing for


making dilutions
Generating Needle washing
Vacuums (outside) HGB blank

SV to CP

Generating Dilutions in parallel Needle Pre-washing


Vacuums WBC, MIX, 4diff (Inside)

SV to NP

Making RBC dilution

SV to CP

Note: CP means “Chamber Position”; NP means “Needle Position”;


SV means Shear Valve; SR means Sample Rotor
47

Cleaning Needle Making 4diff RBC


(Inside) dilution + measurement

refilling sheath
puffer

Generating
vacuum SV to NP HGB
measurement
+ reset
vacuum

4diff Cleaning WBC


measurement measurement
MIX, RBC chambers,
TCU loops 1st time

Taking BASO
Regenerate
sample
vacuum

SV to CP

BASO Cleaning
measurement
WBC, RBC chambers
TCU loops 2nd time

RESULT SV to NP

START NEXT
48

5.3 Initialization of the Fluidic System


Fluidic initialization process performs the following steps:

 Checking the hardware conditions of the unit.

 Checking of pump and pressures sensor by generating measuring vacuum.

 Positioning all mechanical components by scanning moving range (with end-switches).

 Checking prime conditions of reagent buffers.

 Cleaning of aperture with high-pressure back-flush and high-voltage burning.

 Preparing for compulsory blank measurement.

5.4 Reagents priming


Sample measurement or most of the fluidic functions is not available if the reagents are not fully primed.
This can be done manually or automatically from the Maintenance menu, but the instrument also
performs automatic priming in several points of the pneumatic functions. (e.g. starting fill-up process,
Preparing for measurement, wake up…etc.)

The fluidic system is connected to the reagent containers in a closed fluidic way. This means when a
syringe is priming reagent from the internal reagent reservoirs, the generated vacuum will automatically
fill them up when the floating sensors detects low liquid level. Hereby the priming of the reagent during
the measurement is automatically.

During measurement the system automatically double-check the reagent levels in the buffers, and notice
the user at the end of that measurement, if any of the reagents are running low and replacement is
needed.

5.5 Piercing process


After the Start Button was pressed and all of the necessary initializations had done, the measurement
cycle starts with the piercing process.

The Sample rotor immediately turns in; the needle comes forward and starts to pierce the sample.

5.6 Sampling process


When the needle reaches the sampling position, a dilutor syringe takes out 100µl of blood. The blood
sample is separated from the diluent by air bubble. The air bubble was taken right after the needle
washing process of the previous measurement or during the preparing for measurement or wake-up
processes.
49

When the needle is pulled out of the sample tube and washed by the wash-head from outside, the
primary blood sample will be transferred into the Shear Valve loops.

5.7 Needle washing processes


When the primary sample is taken out, the needle will be washed outside by the wash-head using
diluent and continuous drawing.

During Diluting process, the needle will be washed through twice. First a preliminary flow of diluent
slowly washes out the remnant of the whole blood into the washing head. Finally 2,5ml of diluent cleans
out the needle with high speed.

When the needle is clean, the system will take air bubble into the end of the needle preparing for the
next sampling.

5.8 Diluting processes


After the primary blood sample was taken out and transferred to the first 3 SV loops, the SV rotates to
Chamber Position (CP).

The loops in the SV counting from the needle side are:

1st – WBC (~16μl) 2nd – MIX (~16μl) 3rd – 4diff (~40μl) || 4th – RBC (~16μl)

4-diff has a special loop; it is transparent and has higher volume.

The system makes the first step of RBC dilution, WBC dilution and 4diff pre-dilution in parallel.

WBC dilution is made by 2.5 ml mixture of diluent and lyse. The concentration of the lysing mixture is
approx. 1:4. The WBC dilution is mixed with air bubbles from the bottom of the WBC chamber.

During 4-diff pre-dilution the primary blood sample is forwarded with diluent into the TCU to preset of
the required temperature.

The first step of RBC dilution made by 2.2 ml of diluent and mixed with air bubbling from the bottom and
the side connector of the Mix chamber.

When the mix dilution is ready, a dilutor syringe moves the mix sample through the SV into the 4th
sampling loop. Then the SV rotates back to Needle Position (NP).

The RBC dilution is made in the RBC chamber with 2.5ml of diluent and mixed with air bubbling from the
bottom.
50

5.9 Lysing process


The dilution for the WBC and optical BASO count is the same. The WBCs are shrink by the quick lyse
reagent to its nucleus except the BASOPHILES. After the sample dilution measured in impedance way,
the remaining sample is transferred to the optical head for basophile count.

More complex lysing process is made on the 4diff dilution. It requires precise timing, volume control and
temperature conditions. The accurate temperature control is made by the TCU module.

In the TCU unit, there are loops for lyse, stopper, blood + lyse, pre-diluted blood mixture. Temperature
of these loops is controlled accurately.

After the 4-diff primer sample is pre-diluted and its temperature has been set, 3 syringes (lyse, stop,
diluent) make the precise dilution through the TCU unit, the Shear Valve 4-diff sample loop, ending in the
big vacuum puffer. Previously generated vacuum supports the smooth flow of the mixture.

The pre-diluted blood meets with the lysing reagent by a small “T” connector first. Right after this point
there is an inline mixer part in the TCU module, where the blood and lyse is mixed in its tube by flowing
through. After that the precisely mixed lysing dilution runs through one of the TCU heater loops for
better temperature conditions, and reaches the stopper “T” connector. The volume between lyse and
stopper “T” connectors is about 1ml, and this is the incubation zone for the lysing.

After adding the stopper reagent the mixture goes through another inline mixer, the SV 4-diff sampler
loop, and flows into the vacuum puffer. The SV 4diff sampler loop is volumetrically set by the software,
so the precisely lysed and stabilized 4diff mixture will be stopped in this loop, ready for measurement.

The parameters of the lysing process are the temperature of the dilution, the dilution ratios of the blood-
diluent-lyse-stopper mixture and the speed of the flow. These are determining the quality of the 4diff
dilution.

5.10 RBC counting process


The regulated vacuum aspirates RBC dilution (RBC) from the RBC chamber through the aperture. The
instrument counts the cells for 10 seconds in this case.

5.11 WBC/BASO counting


The regulated vacuum aspirates the WBC dilution from the WBC chamber through the aperture. The
instrument counts the cells for 10 seconds in this case.

After the WBC dilution has measured in capillary mode, the system transfers the remnant of the sample
to the BASO loop in the Shear Valve. The last parameter in measurement cycle is the optical BASO count.
The BASO count is measured by the optical head as like the 4-diff measurement.
51

5.12 WBC 4Diff counting


The regulated vacuum aspirates the WBC 4diff dilution and the sheath fluid from Sheath puffer, through
parallel tubes (one for sheath flow, one for sample flow), through the flow-cell into the vacuum
chamber. The instrument counts the cells for 6 seconds in this case.

During measurement the core diameter of the sample stream in the flow-cell is approx. 250µm. This is
determined by the tube resistance ratio between the sheath and the sampling lines.

A self-cleaning process is started after the 4-diff measurement ends. The system changes the insertion
point of the vacuum, which starts to wash back the sampler needle and the 4diff sample loop in the
Shear Valve with sheath fluid. This process prepares the optical head for the following BASO count.

5.13 Chamber draining processes


Chamber draining is made by a special pressure control. Pressure controlled draining starts with vacuum
generation in the puffer reservoir. The liquid aspirated from the chambers while buffer reservoir and
thus the pressure sensor is monitored in the draining tube. The instrument can detect the empty state of
the chamber by drop of vacuum. Error message is send if the vacuum loss was not detected indicating
the measuring chamber draining procedure was failed.

5.14 Cleaning (rinsing) processes


All cleaning processes during a measurement cycle uses mainly diluent, and for special cases small
volume of lyse, and stopper reagent (e.g.: cleaning the adequate loops in the TCU).

The instrument performs the cleaning processes parallel to the measurement. When any of the fluidic
part has finished its work with the whole or diluted blood, cleaning process will start.

The main cleaning processes during measurement are:

- Needle washing outside, 2x inside into washing head

- Mix chamber washing

- RBC, WBC chamber washing, back flush and high voltage burn of apertures

- 4diff diluting loop (Shear Valve + TCU loops) washing

- Optical head and sampling line washing with sheath fluid


52

5.15 Standby process


Because of the high throughput of the Abacus 5, the measuring cycle ends with empty WBC chamber,
and leaves the BASO sampling line is not cleaned. During continuous measurement the beginning of the
next measurement will automatically clean this loop.

Avoiding the BASO loop contamination the liquid is aspirated automatically from it two minutes after the
last measurement of a sample batch.

However, when there is no more measurement started, the instrument will go to stand-by mode after
some minutes.

Standby process performs cleaning, drains all vacuum chambers, drains all chambers then refill them
with diluent just above aperture level.

5.16 Wake up process


In this case the instrument prepares itself for the upcoming measurement cycle. Cleaning the wash head
and calibrating the blood sensor.

5.17 Shutdown process


The fluidic shutdown performs the following steps:

- Priming chambers with diluent to avoid drying out of aperture and prevent the chamber from
contamination.

- Needle is in up position and washed.

- All of the syringes are positioned down.

- Sample rotor moved out


53

6 ADJUSTMENTS
For proper functioning, it is vital to adjust the system correctly. Although Diatron production sets the
system in house, sometimes these adjustments needs to be done on the field also. Both mechanical and
software adjustments has to be done properly to avoid measurement results or pneumatic errors.

Please note that the lifetime of the mechanics system is long. If there's a problem with the mechanical
system, most cases a proper adjustment is enough to solve the problem.

6.1 XY unit mechanical adjustments


The XY unit has 4 critical parts to adjust. These 4 adjustments can be done separately, but some of them
can affect other settings. See the 4 attributes of the XY unit which has to be set properly:

 XY (needle) horizontal movement

 XY (needle) vertical movement

 Needle position

 Washing head position

If any of these four attributions are not set correctly, the system can stop with a pneumatic error or give
improper results. These adjustments can be tested in Service testing menu.

The movements of the XY unit made from Service testing menu are slower than
normally. Some cases the improper hardware settings are not appear during the
movement used by Service testing menu. To make sure that the mechanical part is
working properly, the system has to make an Initialization process (hardware init).
This process moves the mechanical part (XY) faster, highlighting the problems
made by the faulty settings or misalignments. Hardware init can be done in Service
functions screen or the system is making an automatic init process, after e.g.
Washing head setting process is finished. Use these functions to check mechanical
settings also.

6.1.1 XY horizontal movement


The proper horizontal movement is important to make sampling from the Sample rotor and from the
Auto-Sampler. The horizontal movement of the XY unit is made by one of the stepping motors and
positioned by two separate opto flags. The Front and Rear positions are defined by the settings of these
opto flags. The XY unit is moving on metal rods. These rods have to be greased after a period of time.
54

6.1.1.1 Greasing the horizontal rods

There are 2 horizontal rods in the XY which is responsible for proper and smooth movement of the
needle and the needle holding block. These rods are cylinder shaped. They have a fix position in both the
front wall and rear metal block in the XY. The rods are not fixed tight, they're loose and can wobble to
move together with the needle block.

The third rod is squire shaped and responsible to hold the vertical belt for the needle. The front of this
rod is not fixed in a metal block, it “hangs” free.

Guide rod

Square
shaped
rod

Guide rod

Test the movement of the needle to Front or Rear positions in Service testing menu. If there is any
grinding noise because of the improper lubrication, these three rods has to be lubricated.

Name (and part number) of the lubricant: Photolube 022 (A598)

6.1.1.2 Set opto flags for horizontal movement

If the XY unit hits the front wall of the mechanics (Front position, above the sample rotor) or stops in a
non centered position above the Auto-Sampler docking station (Rear position). Use a hexagon 2.0 mm
screw to loose the fixing screws holding the opto flags. Align them to make the XY horizontal movement
stop earlier or later, depending on the actual state of the flags.
55

Opto flags

Opto LEDs

Opto sensors

Fixing screws

Front wall

Use handmade horizontal positioning of the needle, place the needle to the correct position, then align
the opto flag to the opto sensor, see that the indication LEDs are turned on when the flag reached the
sensor.

When the needle reached the ‘Front’ position, the gap between the needle holding block and the front
wall of the XY unit should be around 0.5 mm.

When the needle reached the ‘Rear’ position, make sure that the two vertical rods -which are holding
the washing head can push the tube holders to the correct position on the Auto-Sampler's docking
station.
56

Needle holding block

Washing
head
Washing
head holder

AS docking

Tube

The opto sensors are fixed to the XY PCB and cannot be adjusted.

6.1.2 XY vertical adjustment


This setting is necessary for the vertical motor movements because this adjustment sets the opto end-
switches of the XY unit.

6.1.2.1 Basic operation

During testing the vertical movement of the XY, some strange noise can appear or maybe the vertical
movement is not performed correctly and it's causing the pneumatic error. These problems can be found
if we understand how the mechanics works.

Find the opto wheel (black cogwheel) at the rear side of the XY unit. Note that the vertical opto sensors -
next to the opto wheel- are using negative logics. The sensor is always triggered by the opto wheel,
except when the small holes on the wheel reaches the sensors. These wholes indicates the Up and Down
positions and the needle should stop when the wholes reaches the sensors. The opto LEDs are always on,
except when the needle reaches the ‘Up’ or ‘Down’ positions. The opto wheel is connected to the
vertical belt -by the middle square shape rod -and they are moving together. The needle is connected to
the belt by two screws. If the opto wheel reaches the ‘Up’ or ‘Down’ position, the needle should also
stop.
57

The problem is that the mechanical structure of the XY is making the needle stop before the whole on
the opto wheel is reaching the opto sensor. This can be happen in 'Down' position, because the needle
mechanics is reaching the position earlier than the opto sensor gets the signal to stop. Or in 'Up'
position, when the Washing head is already reached the highest position and hits the needle's vertical
block, while the opto sensor still didn't received the 'Up' position signal.

6.1.2.2 Adjusting the needle’s vertical position

The problem can be easily set by losing the two


screws behind the needle and let the needle
"release" the belt and fall down to "Down" position.
Because the screws are loose, the opto wheel and
the belt can move separately from the needle, the
needle itself is down and not moving.

Disengaged screws

Now the opto wheel can be set after the needle,


which means the opto wheel should reach Down
position. Press 'Down' in Service testing screen to move the opto wheel to 'Down' position. The needle is
already in 'Down' position. When the opto wheel is in 'Down' position, tight the two screws which makes
the needle "grap" the belt again. Now the vertical system is set together again, press 'Up' and 'Down'
buttons in Service testing screen to see that the movements are proper. Test the movement with Init
process also.

Adjustment is successful if LEDs goes on before moving part reaches end of mechanical range.

Changing the vertical position of the needle on the vertical belt can affect the position of the washing
head and the needle. Please perform a needle and washing head setting after vertical position is set.
58

6.1.3 Washing head settings


The position of the washing head is critical for the
system to make the sampling and needle washing
correctly. The washing head contains two metal
tubes to make the washing process. The bottom is
the inlet for diluent to wash the needle from
outside, the top is the outlet to aspirate the liquid
from the needle to the waste. During the washing
process, the system also pushes diluent to the
needle from the shear valve, ensuring that the
needle will be cleaned from the inside also. The
relative position of the needle and the washing head
is important to make the aspiration properly,
otherwise the diluent from the needle will flow out from the bottom of the washing head.

To set the Washing head’s position, go to Adjustments menu and press ‘Wash head’ button. The
message will appear to confirm that the washing head setting process will start. Press ‘Ok’. The system
will pull the needle and the washing head to Up position. Now the system holds the XY unit in the Up
position until the setting is ready. The washing head is in position if the distance between the top of the
head and the metal block above it is close to 1mm.

1 mm gap

This distance is not strict, but it’s enough to provide space for the washing head when the needle moves.
If washing head’s position is higher, it can hit the metal block when it’s raised to Up position and it can
hit the sample rotor and the tube holders if the position is too low. If the position is not correct and has
to be set, find the two screws on the top of the XY unit. These two screws are holding the two rods of the
washing head. Loosing these screws make the height settings enable.
59

1.5 hexagon screws to hold the


washing head leading rods

2.0 hexagon screws to


hold the needle

6.1.4 Needle setting


The height of the needle is important to make a good aspiration from the blood sample. The aspiration
whole is on the side of the needle, but for height setting, we use the bottom end of the needle. This end
of the needle is sharp, but during the procedure, there is no need to touch this part. For position
checking, the needle can be moved by hand or by the software.

6.1.4.1 Handmade setting

Manually push the needle to down position. Make sure that the needle is not hitting the sample rotor’s
plate. Check that the needle is completely down, then take a look at the end of the needle. The distance
between the sample rotor’s plate and the needle should be 1mm.

If the distance is not correct, find the two holding


screws at the other end of the needle. Use a
hexagon 2.0 mm tool to loose the screws and set
the needle’s position.

1mm distance

Note that when the needle is Down, the screws are


covered by the mechanics of the XY. The needle has
to be raised to have access to the screws.

The needle is sharp and it can cause injury! Always use gloves when needle setting
is performed! Make sure that the whole on the needle is facing the sample rotor.
60

6.1.4.2 Software made setting

In Adjustments menu, press ‘Needle’ button. The system will ask to confirm that the needle setting
should start. Press ‘Ok’. The needle will move down and up. The software will ask to confirm that the end
of the needle is visible at the bottom of the washing head. If the needle is not visible or it’s leaning out
too much, loose the holding screws to adjust. If setting is ready, press ‘Ok’.

6.2 Shear valve mechanical setting


This opto setting is necessary for the correct movement of the Shear Valve. Although the SV is stopped
mechanically in its end positions, the opto sensors provide the necessary feedback signal for the
electronics that the movement has been done. The setting of the mechanics and the opto sensors are
made by Diatron during production, sometimes these alignments has to be checked or realigned.

6.2.1 Shear valve lower disc position setting


For adjusting the lower disc, remove the thumb
screw and place the upper disc to a safe position on
the mechanics under the optical head. The lower
disc of the shear valve is positioned by two metal
plates. Setting up the plate on the right side is
critical.

Plate and disc connection surface

The plate is fixed by to screws. This is the basic


setting of the valve, if the disc is not fixed correctly,
the amount of sample can be different during each
measurement.

The disc and the plate should connect in the full length of the connection surface. The disc shouldn’t
wobble or move. If the setting of the disc is not proper, loosening the two screws on the bottom and
realigning the connection between the plate and the
disc should be the first step. This setting should be
set with a hexagon 2.5 mm screwdriver. The two
connection surface should be parallel. When the
lower disc is aligned, make sure about the following:

 the disc position is horizontal


 the plate fixing screws are tight
 the sides are parallel
 the disc meets the plate with full surface
 the lower disc cannot move
61

While making an alignment and the shear valve is disassembled, make sure that the internal surface of
the ceramics is clean, not scratched. If the plate and disc is aligned make sure that the second plate on
the left side is pushed to the disc. This plate ensures that the shear valve will not move away from the
aligning plate.

After the lower disc is aligned, replace the upper disc and fix the thumb screw back to position.

6.2.2 Shear valve upper disc setting


The upper disc and the mechanical stoppers (bumpers) are aligned by Diatron during QC and End of Line
procedures. If this setting is compromised, the mechanics should be aligned again. To have access to the
mechanical stoppers, the metal cover of the shear valve should be removed. The cover is fixed to its
place by the two inner screws behind the thumb screw.

The shear valve’s task is to change position to guide the sample and the reagents to the correct
direction. For this, the tubes should be perfectly aligned which means that the mechanics should stop
the turning of the discs in a correct position.

Bumper fixing screws

Shear valve upper disc

Plates holding the ceramic

The mechanic part is designed to change the motor’s circular motion to horizontal motion. There is a
horizontal drive behind the ceramic disc which can be shifted left or right by the stepping motor.
62

Horizontal drive

Shear valve moving nodes

Mechanical stoppers (bumpers)

This rod has two nodes which are surrounding and moving the upper disc’s rod. This is how the disk is
moving. The nodes can reach a stopping mechanic block on both sides of the mechanics. If nodes hit the
stoppers, the turning of the shear valve will stop. When the shear valve is stopped, the small metal tubes
on the upper and lower discs should be perfectly aligned to each other.

The shear valve has a free hole on the edge of each ceramics. This hole is not connected to the tubing
system and it’s easy to access. There is no special tool to check shear valve alignment; a simple paper clip
is enough to check the alignment in both positions. The paper clip should access both holes on the
ceramics. During the test, make sure the shear valve reached to bumper and cannot turn more.

The paper clip has a size and straight part just enough to push through the ceramics.

Opto sensors

Opto rod

Paper clip in alignment hole

Alignment hole

The lower disc has two holes to test both positions. Test one position at a time.

If the paper clip can go through both discs, it means the shear valve is in position.
63

If the paper clip cannot go through the holes the upper discs position has to be realigned. The bummer
fixing screw should be loose to make the bumper move free. Find the correct position where the paper
clip fits into both holes. If the position is ok, push the bumper to position until it hits the node of the
drive. If the positions are set, the bumpers have to be fixed again by tightening the screw.

Note that the shear valve needs to be pushed to the other position to have access to the bumper fixing
screws. Remove the paper clip to turn the shear valve with the horizontal drive.

Use the paper clip to see alignments in both positions (CP, NP). Make sure that the bumper holding
screws are well tight.

6.2.3 Setting the opto sensors of the shear valve.


There are two opto sensors for the shear valve to sense when the ceramics reached their positions. The
upper disc has a rod called the opto rod. This rod moves inside the opto sensors when the shear valve is
turned. The sensor should give the signal to the software to stop the motor because the positioning of
the shear valve is done.

The opto sensors are placed on a PCB called Shear


Valve OPTO (SVOPTO). The sensors can be
positioned separately; it doesn’t need to align both
if only one sensor is misaligned.

Shear valve LEDs

SVOPTO PCB

The sensors are connected to a LED, which turn on


when the sensors are triggered. Use these LED to
check alignment.

Turn the shear valve clockwise to the outer (chamber) position. Make sure that the shear valve is in the
maximal position and cannot turn more. Put the paper clip to the alignment hole. It should access both
holes easily. The outer sensor’s LED should be turned on in this position. Check that the LED is on,
position the LED PCB if it’s needed. Use hexagon 2.5 mm screwdriver.

If the sensor is in position and the LED is on, try to turn the shear valve counter-clockwise. Because the
paper clip is in the alignment holes, the shear valve will only turn in a very small amount. Find the
maximum position until the paper clip lets the shear valve turn. In this position, the LED should be still
on. Make sure that the LED is on and set the SVOPTO to make the LED light again. Remove the paper clip
and turn the shear valve to the inner (needle) position. Repeat the process to set the opto in this
position too. Move the shear valve by hand when aligning opto sensors, remove the paper clip when
done. Check the alignment by moving the shear valve in Service testing menu.
64

Note the in needle position, the SVOPTO is fixed with two screws. Loose and tight both screws during
alignments. The SVOPTO PCB screws can be accessed by the holes under the blood sensor.

6.3 Amplifier offset setting

Values
Offsetfor
values
manual
of Auto-offset checkbox
offsetting
the channels

Send settings button

Field for manual


offsetting

The analyzer contains several amplifier circuits to gain signals during the measurements. An ideal
amplifier gives zero mV output if input is has 0mV. In reality, the amplifier's output always has some
electric potential which is different than zero. This voltage is the offset of the amplifier and has to be
zeroed to avoid any effects on the measurement results. In Service menu Adjustments screen, it's
possible to set these values manually or enable the checkbox to use the auto-offset option. There are
three amplifier channels in the optical head, all three should be zeroed to avoid noise during optical
measurement.

CH0: Channel for Low angle gain

CH1: Channel for High angle gain

CH0 DC: Channel for optical head DC level.

Amplifier offset should be between ±5mV. It's possible to check offset values by pressing Read data
button. Reading the offset continuously will turn the start button red, indicating that the internal
65

communication is busy. The system cannot command the Low Level software until the offset reading is
Stopped.

Impedance amplifier offset can be adjusted separately, also by filling the potentiometer field in the
middle or by using Auto-offset checkbox.

Always press 'Send settings' button to finish adjustments. Offset setting are done by LSDACQ. If Auto-
offset is not working or values are not close to zero, replace the LSDACQ.

6.4 Blood sensor calibration


Abacus 5 contains a blood sensor to ensure that the value of the sample is the same during every
measurement. After every startup, the system calibrates the blood sensor automatically by making a
Wakeup process.

During wakeup, the system creates a bubble in the


wash head and pulls the bubble inside the blood
sensor. The sensor measures the number of steps that
the dilutor has to make to pull the bubble into the
sensor. The system repeats the process three times
and records the number of the steps. During the
wakeup, these numbers are displayed in MDA screen if
MdaOn checkbox is enabled.

Result of blood
sensor calibration

MdaOn checkbox
has to be enabled
66

Field to enter number of test


functions

Checkbox to enable or
disable blood sensor

Press MDA to view MDA


screen

If the number of steps are the same for two times (out of three), the sensor calibration is successful. If all
three values are different, the sensor calibration is failed. Failure of the calibration is displayed as a 'T'
flag after a blank, control or human measurement is finished and warning exclamation mark is also
shown on the bottom of the display. If blood sensor calibration problem appears, it's possible to run
Wakeup process directly from Service functions. See test functions list in section 7.3.1.1.

The sensor can have problems sensing the bubble's position because the sampling tube is not
transparent enough or it's dirty. In this case the tube between the needle and the shear valve has to be
replaced.

Note that the calibration value is strongly depends on the state and position of the sample tube. This
tube in the blood sensor is not fixed and its position can be changed by pulling the tube during a shear
valve adjustment or cleaning. The tube is set correctly if calibration value is in a range 44-49. If the value
is different, it can cause sampling problems, 'W' flag or it can happen that the sampling tube will be too
tense when the needle is piercing a sample from the Auto-Sampler.

Make sure that the tube between the shear valve and the blood sensor is horizontal, not too tense and
transparent.
67

6.5 Sample position setting


The measurement starts with a blood sample
aspiration. The blood sample needs to be moved to a
specific location through the shear valve. Open the
front cover of the analyzer and observe the beginning
of the measurement process. The below adjustment is
necessary to allow the sample reach this location – to
avoid sampling errors and thus avoid optical
measurement errors

During sampling, the blood sample must go beyond the


last loop of the shear valve, as indicated on the image.

The purple area is a schematic representation of the purple area on the shear valve in the Abacus5.
Please locate this tube in the analyzer (the tube is connected to the V41/1 tube in the vertical tube
organizer.

Run a control blood sample. If the sample position is


between 2 and 8 mm, then do not change the default
position. If the three loops are not filled with blood or
the sample stops more than 8mm beyond the shear
valve, the sampling settings has to be set in
Adjustments screen.

Length of sample position


used if blood sensor is
enabled

Default blood position value


is used if blood sensor is not
enabled

In Adjustments screen, the sample position can be set at Needle tube full length/sample position. If
blood sensor is enabled, the system uses the top value (70 by default) to position the sample. This value
can be calibrated by pressing Calib length button.
68

 If button is pressed, the system will create a bubble in the wash head and pulls it through the
sample tube and the three loops to the top of the shear valve. The software will ask whether the
bubble is visible at the top of the shear valve, this has to be checked visually.

 If the bubble is not arrived to the correct position, press Cancel to make the system (the dilutor)
pulls the bubble higher from the shear valve. The system will ask to check every new position
after Cancel is pushed.

 If the bubble is in position press Ok. The system will overwrite the old sample position value
(default 70) with the newly set value.

 Rerun a control sample. Observe the position. Run calibration again if sample position is not
proper or type in a new value by hand. If value is changed by hand, press Set length button to
save the new value.

If the blood sensor is disabled or not working properly, the system will use the bottom value so called
the default blood position. It is not possible to calibrate this value, but it can be modified by typing in a
new value. After the value is modified, press Set default blood position button to make new setting
accepted. This value should be in a range 0.190-0.220.

Make sure that the format of this entered value is correct and “0.” is visible. It's important to keep this
value under 0.22 because larger values will make the blood sample override the shear valve and will
enter to the reservoir.

Always use blood to test the sample position after the following:

 sample position value was calibrated

 sample position value was changed (with blood sensor)

 default blood position value was changed (without blood sensor)

 blood sensor was turned off or on

 sample tube was replaced or its position was realigned

 shear valve was disassembled

 XY adjustments was realigned (horizontal, vertical, wash head, needle)

 wash head was cleaned

Always test blood position from Auto-Sampler. Make sure that sample position is proper both from
Sample rotor and Auto-Sampler.
69

6.6 Pressure sensor adjustment


Similar to the amplifiers, the pressure sensors also has offsets. It is also possible to zero the pressure
sensor's offsets in Adjustments menu by pressing the Start adjustment button. The offset compensating
is done automatically, showing the offset results in mBars. These values must be 0 ± 20 mBars. If one
value is higher than 20mBars, the pressure board needs to be replaced.

The sensors can work with higher offsets than 20mBars, but if values are higher than 30mBars the
system will give a pressure sensor offset error and if the values are higher than 40mBars it will cause a
pneumatic error.

6.7 Set serial number


It is also possible to enter the instrument's serial number in Adjustments menu. Simply type in the
correct number and press Set serial button. The serial number of the instrument is stored in the
Hardware block, the LSDACQ and the DimmBoard. Serial number is also indicated in the label on the
back of the instrument.

Serial number has to be typed again if:


 Dimmboard was changed
 Low level software was reinstalled
 LSDACQ was changed
 HDD was changed
 High level software was reinstalled

6.8 WBC preheater


In Adjustments screen, it is possible to set the WBC preheater's temperature. The default value is 375
indicating that the temperature should be 37.5 °C. By pressing Read data button, it's possible to check
the preheater current temperature.

Make sure that the preheater's value is 375 (37.5°C), if it is not possible to set the temperature or values
are too high, check WBC preheater's cable or replace LSDACQ board.

6.9 Setting Laser values


Every laser head has unique values which have to be set in the instrument for proper working. These
values were set by Diatron, but has to be set again if the laser head was replaced. In Auto Alignment
screen it's possible to check or overwrite the values. The Read DC button and the laser values are
accessible, but other options are password protected. Password is asked every time when entering Auto
alignment menu.

Do not change the laser values unless it's necessary.


70

6.9.1 Calibration procedure of optical heads

1. Enter to Service/auto alignment menu (password:47715481) and set optical head values as
shown below:

Laser power: 120

Low angle gain: 130

High angle gain: 210

2. In service/service calibration menu, set Scatter X and Scatter Y values to 1,1. Press Accept when
done.

These values needs to be set to make a proper start for the laser to calibrate.
71

3. In settings/system/set menu, select Gravity X and Gravity Y columns on the right list, and put
them to the used columns on the left list by pressing the left arrow. Press done and save when
ready.

4. Do a control measurement using a control blood. In the database find the results of the
measurement. Check the Gravity X and Gravity Y values of the measurement and compare them
with the Gravity values shown on the blood sheet of the actual control lot.
72

5. If measured Gravity values (X and Y) are between 0,8 x Lot Gravity values (X and Y) and 1,2 x Lot
Gravity values (X and Y)(+-20%), then the Gravity values can be calibrated by software.

6. If measured Gravity X is lower than 80% of Lot Gravity X, raise low angle gain in auto alignment
menu to bring the measured Gravity X value closer to Lot Gravity X. If measured Gravity X is
higher than 120% of Lot Gravity X, then lower the low angle gain to bring the measured Gravity X
value closer to Lot Gravity X.

7. If measured Gravity Y is lower than 80% of Lot Gravity Y, raise high angle gain in auto alignment
menu to bring the measured Gravity Y value closer to Lot Gravity Y. If measured Gravity Y is
higher than 120% of Lot Gravity Y, then lower the high angle gain to bring the measured Gravity
Y value closer to Lot Gravity Y.

8. Repeat the process until the measured Gravity values are between 80% and 120% of Lot Gravity
values.

9. If values are ok, enter service calibration and check enable scatter calibration.
73

10. Press ok and browse the AS file for scatter calibration.

11. When ready, go to calibration menu and perform a calibration (min 3 measurements)

12. The new Scatter X and Scatter Y will appear. Press ok.

13. Calibration is done.

6.9.2 Short guide for laser values

This is a short description to give information about the values and settings of the optical head assuming
that user understand the operation of the optical measurement.

Concepts:

Scattergram Gravity X
Scatter scale X Gravity Y
Scatter scale Y Low angle gain
WOC4 High angle gain
WOCB DC level
Laser Power

Scatter diagram (Scattergram):

The scattergram is responsible to view the white blood cell


populations. The 4 diff scatter indicates the 4 differential
populations (LYM, MONO, NEU, EO), the Baso scatter shows the
BASOPHIL population. Both scattergrams are indicated as a
square, surrounding all the populations. In default, the scatter
has a size 1.00 x 1.00. This scattergram can be magnified, making
the cells / populations change positions. The size of the scatter
is defined by the scatter X (horizontal) and scatter Y (vertical)
values. These values are 1.00 in default and can be modified by
calibrating the scatter.

The values of Scatter X and Y are shown in service calibration menu or it can also be found on factory
settings menu.
74

Gravity values:

Gravity values are specific numbers for all type of control


blood for instruments with optical measurements. These
numbers indicates the center of gravity.
All cells (which is part of a population and not noise) are
counted and added horizontally, line by line and divided by
the number of lines shown in the scatter. This value is the
Gravity X value.
The Gravity Y is created the same way; all cells attached to
a population are added vertically and divided by the
number of columns shown in the scatter.

All controls have its own gravity values, these values are used as target reference during calibration. The
gravity values are stored in an AS file and can be uploaded or added manually before making a
calibration.
(After a measurement, the gravity values are indicated in the measurement database.)

WOC4 and WOCB values:

WOC4 (WBC Optical Count for 4 diff scatter): indicates the ratio of WBC cells included in the 4 diff
measurement. Min. value: 2.72
WOCB (WBC Optical Count for Baso scatter): indicates the ratio of WBC cells included in the BASO
measurement. Min. value: 5.44

These values are also displayed in the measurement database, showing that the number of cells during
measurement is enough.
If WOC4 is low (and WOCB is ok), it can indicate that the TCU is clogged and only a little amount of cells
entered the flow cell.
If WOCB is low (and WOC4 is ok), it can indicate that there is a partial clogging in the WBC chamber, not
allowing the cells to reach the flow cell.
If both values are low, it means that the flow cell is clogged (both 4diff and Baso cells aren’t sensed by
the laser) and needs to be cleaned by software or manually.
75

Neutrophil area center point:

If the measurements are displayed with Analyzer Spy


software, there is a chance to enlarge the image of the
actual measurement’s scatter. There is a small green
cross appeared on the scatter (both 4diff and Baso).
This marker is created to help the laser’s calibration
procedure, if there’s no control blood available for
measurement.
The marker indicates the Neutrophil populations
center point. With the use of this marker, the user can
manually set the scatterX and Y values in the Analyzer
Spy software. By enlarging the scatter with X and Y
scale scatter values, the Neu population field can
shifted to a position where the marker is the middle of
the Neutrophil population. This is a harsh approximate calibration showing the proper Scatter values. This
values can be set in the instrument manually in service calibration menu or factory settings.

Laser Power and gains:

In auto alignment menu, there is a chance to set the optical head values, laser power, low and high angle
gains. These values are set by Diatron QC, but these values can be modified to align the optical head
operation to the instrument.

Every optical head has an unique setting of values. These values can be set in auto
alignment menu and stored in the instrument’s memory. The values are not stored
inside the laser. If optical head is changed in the instrument, the new laser’s values
have to be set in auto alignment menu to make the new laser’s operation proper.
Also note that these values are reset to default, if the windows is reinstalled or a
hardware block is changed.

Laser power values (both for 4diff and Baso) indicate the current operating the Laser Diode. The higher
the laser power value, the lower the current flows though the diode. If the power values are increased,
the current will be decreased making the scatter smaller in both dimensions (X and Y).
Laser power value had a range from 54 to 96. Older (2012 and older) type of laser’s must keep this range
(54-96) with a standard low angle gain 48, and high angle gain 112.
In new type of optical heads (created in 2013 and after), the optics, flow cell size and amplifier is
modified.
This modifications increases the ranges both for power and gains. See ranges for new type of optical
heads below:
76

Laser power: 70 – 130

Low angle gain: 0 – 255

High angle gain: 0 – 255

Note that the lower the laser power value, the higher the Laser diode’s current, which can create NOISE
during laser blank measurements. These NOISE is indicated by warning flags n, N at the end of a
measurement.

The diode’s light is sensed by the detector. The area of


detection is masked by a metal plate to ignore the LED’s
direct light hitting the sensing zone. The zone is separated to
two different subzones. The internal zone responsible to
sense the low angle signals (size of cells), the outer larger
zone is for sensing the high angle signals (internal structure
of cells). Both incoming signals are very weak and must be
gained by an amplifier. The Low and High angle gains are the
gain values of the laser’s amplifier, these values also can be
set in auto alignment menu. In older optical heads, this
values are standard 48 and 112, but new optical head’s -with
3 stripes on the side- has a range of gain values.

The gain values are good to make small scatter alignments to improve the Gravity values and Scatter
sizes in small amount. Note that the values are directly proportional, the higher the value the higher the
gain is.

DC level:

The DC level value indicates the amount of light hitting the LOW ANGLE SENSING ZONE without any cells
involved. This value has a minimum and maximum range. Low DC level could mean that the beam of light
is misaligned, high DC level means dirty flow cell or optics.
77

7 Verification procedures
7.1 Self-test
The analyzer has built in test function to check and evaluate operation of internal modules and systems.
The function is accessible from the Main Menu, Diagnostics, Self-test function.

There are two subsets of tests: electronic and pneumatic. Each process takes approximately 1 minute,
and provides results of each tested parameter. You can select to run both test sets, by clicking on the
“Start Both” button. Accepted ranges are displayed below.

Big buffer time Small buffer time


Generate 3000 - 13000sec Generate1 1200 - 2800sec
Release 3000 - 6000sec Generate2 1200 - 2800sec
Release 700 - 1300sec
Big buffer drift Small buffer drift
Maximum 540 - 560mBar Maximum 225 - 235mBar
Minimum 530 - 560mBar Minimum 215 - 235mBar
Drift -5 - 15mBar Drift -5 - 15mBar
Pump status Null pressures
Pump1 1-1 Sheath -20 - 20mBar
Pump2 1-1 Capillary -20 - 20mBar
Chamber -20 - 20mBar
TCU Laser temperature/Optical
Reference 25 - 38°C Laser off 0 - 0.05V
Actual Reference +- 0.2 °C Laser on 0.06 - 0.8V
Sink 0 - 70°C
HGB LED Battery
HGB Dark 0 - 3000 pulses Battery voltage 2.7 - 3.3V
HGB light 3000 - 60000 pulses +12V 11.4 - 12.6V
-12V -12.6 - -11.4V
Electrode Noise/Pulse
Voltage 45 - 55V pls/5sec 0 - 2000 pulses
Current 620 – 680μA 20000pls 19990 - 20050 pulses
Offset -5.0 – 5.0mV

Would any value fall outside the above defined range, the SW will indicate it with a red, “Failed” string.
Correct and acceptable results are indicated with a “Passed” string.
78

7.1.1 Abacus 5 Self-Test Guide


The Abacus 5 Self-Test provides information about the actual status of hardware and software of the
analyzer.

This guide is intended for helping service engineers to correctly interpret Self-test result and to provide
guidelines in troubleshooting the instrument.

It is possible to perform electronic Self-Test only, pneumatic Self-Test cannot be performed separately.

7.1.1.1 General information

High level SW version: …….


Low Level SW version: …….

 Low and high level SW version must be identical. For proper functioning of the instrument the
two software versions has to be the same.

LSDACQ Firmware; LSDACQ PIC; Optical head; TCU SW versions; LSDACQ version:

 All these must have version numbers. If version number isn’t displayed that can indicate failure
of part or communication error.

 LSDACQ version number provide information about compatibility with certain electronic
components;

OPT sensor board; TCU board; Pressure board; Laser driver board:

 These are intelligent components which communicate with the low level software via I2C bus. All
boards must be „PRESENT” for correct operation of the analyzer.

 If any of these components is reported „NOT PRESENT” that indicate failure of part, cable or
connection problem.

Corrective action: Check cables and LSDACQ status LEDs


79

7.1.1.2 Electronic Self-Test parameters

TCU temperature(C°):

 Actual: normal operating temperature of TCU unit is 29+/- 0,2 C°. Lower/higher temperature
indicate failure of heater/cooler circuit.

Troubleshoot accordingly.

 Heatsink: cooling of TCU is done by Peltier elements. Heat generated on the other side of the
elements is dissipated by a heatsink and cooling fan. Failure of fan leads to overheating and
damage of Peltier elements.

Corrective action: check and/or replace TCU

 Laser DC level: Laser DC (Dark Current) is the reference value of the optical measurement circuit.
Measurement is performed on diluent. When the optical flow cell is clean and bubble-free, laser
light isn’t scattered by anything and only a very small signal should be measured by the
detectors.

 Laser off: DC level recorded without laser light. If out of range, probably the Optosensor card is
defective. In such case usually the Laser on value is out of range too.

Corrective action: Check laser head's cables, change Laser Head Assembly

 Laser on: DC level with laser light. If Laser off value is in range but Laser on value is high, it can
indicate: dirty flow cell, bubbles in the flow cell, misaligned laser.

Corrective action:

- run „Flow Cell Cleaning” followed by blank measurement to wash out any air bubble from the flow
cell and repeat Self-test. If DC level is OK => nothing else to do

- manually clean* the flow cell with bleach solution

- wipe the flow cell sides with lint-free lens paper dampened with isoprophyl-alcohol

- change Laser Head Assembly

Laser head must be opened for this operation! Power off the analyzer, and proceed with precaution;
take extra care not to damage or misadjust any internal component of Laser Head.
80

Electrode: These are the main parameters of impedance amplifier board.

 Voltage; Current: is out of range, the amplifier board or the high voltage board might be
defective

Corrective action: check or replace HVB board, check or replace amplifier board, check cables

 Offset: if voltage and current is in range, offset can be adjusted. If the offset cannot be adjusted
or it is unstable, check or replace the amplifier board and /or LSDACQ.

Corrective action: change the impedance amplifier board or LSDACQ

 Battery / Power supply: Monitors the supply voltage of LSDACQ board (+/- 12V) and the real-
time clock battery voltage.

Troubleshoot accordingly

Laser parameters:

 Temperature(C°): temperature of laser diode block, for information purposes only

 Laser Power: the value of laser power setting. It is not percentage (!), just a digital number.

DO NOT CHANGE THIS VALUE UNLESS IT'S NECESSARY OR YOU INSTALL A NEW LASER HEAD IN THE
ANALYZER. IN THAT CASE LASER POWER VALUE TO BE SET IS WRITTEN ON THE AUTOALIGNMENT
SHEET OF NEW LASER HEAD OR ON THE SIDE OF THE OPTICAL HEAD.

 HGB Head (electronic): HGB diode and sensor test.

 HGB dark: measured without light, if out of range probably the sensor is defective or too much
external light reached the sensor

 HGB light: measured with HGB diode on, if out of range that can indicate diode or sensor failure

Corrective action: check chamber and amplifier for salt buildups, replace HGB head

 WBC Preheater: if out of range it can be adjusted in „Service/Adjustments”

 LSDACQ: show the temperature of the Dimmboard


81

7.1.1.3 Pneumatic Self-Test parameters

 Big buffer time(msec): The time necessary to generate the required vacuum in the big(sheath)
buffer.
 Generate(P1+P2): failed indicate weak pump, leak in vacuum system, bad valve head or valve
coil, kinked tubing, etc.
 Release: time necessary to completely vent the buffer, if failed, valve #5, related tubing and vent
nozzle can be at the origin of failure
 Small buffer time(msec): The time necessary to generate the required vacuum in the
small(capillary) buffer.
 Generate: failed indicate weak pump, leak in vacuum system, bad or valve head or valve coil,
kinked tubing, etc.
 Release: time necessary to completely vent the buffer, if failed, valve #5, related tubing and vent
nozzle can be at the origin of failure
 Big buffer drift(mBar): Failed indicate leakage in the sheath buffer-related vacuum system.
Troubleshoot accordingly.
 Small buffer drift(mBar): Failed indicate leakage in the capillary buffer-related vacuum system.
Troubleshoot accordingly.
 Max vacuum P1(mBar/sec): The maximum value of vacuum that can be generated by the pump
in a given range of time. Failed indicate weak pump or problem in related valves or tubing.
 Max vacuum P2(mBar/sec): The maximum value of vacuum that can be generated by the pump
in a given range of time. Failed indicate weak pump or problem in related valves or tubing.
 Pump status: Failed indicate that the pump wasn’t sensed by the electronic system. Possible
pump, connection, cable or PPB board failure
 Pressures – buffers vented(mBar): Show the remaining pressure/vacuum in the buffers, after
the instrument released the vacuum. If (any)failed, check related tubing, valves, and pressure
sensor board, run pressure offset adjustment and repeat Self-test.
 Noise/Pulse: This is an electronic test but it require diluent to be present in the counting
chambers.
 pls/8 sec: if failed either electronic noise is picked up by the electrode or the amplifier board is
noisy. Troubleshoot accordingly
 20000 pls: a test circuit generate 20000 pulses and the counter number them. Failed indicate
defective amplifier board
 HGB head: Same test as in the electronic section, but with diluent in the counting chamber.
82

7.2 Error Messages


The analyzer checks the operations of several mechanic, fluidic and electronic parts during
measurement. The system shows the type of the error on the LCD display if any kind of malfunction is
detected.

When an error message pops up, the details contains the processes were involved when the error
appeared. The error codes also help to identify the problem.

Error code Description


0 EG_NO_ERROR
1 EG_BAD_VALVE_NUMBER
2 EG_BAD_PUMP_NUMBER
3 EG_PNEU_INIT
4 EG_BAD_CHAMBER_NUMBER
5 EG_ALL_VALVES_OFF
6 EG_NO_ANSWER
7 EG_CANCEL
8 EG_ABORT
10 EG_UNKNOWN
11 EG_ERROR_HANDLER
12 EG_NO_CONNECTION
13 EG_PRIME
14 EG_TEST_FUNCTION
15 EG_MEASURE_PUFFER
16 EG_MEASURE_PUFFER_SPEC
17 EG_CALIB_SENSORS
18 EG_TEST_MOTORS
19 EG_TEST_VALVES
20 EG_TEST_PUMPS
21 EG_PREP_FOR_SHIP
22 EG_CLEAN
23 EG_WASHHEAD_WASH
24 EG_BLANK
25 EG_STRESS
26 EG_EMPTYCHAMBER
27 EG_SHUTDOWN
28 EG_HARDWARE_INIT
29 EG_PNEUCONTROL
30 EG_PNEUCONTROL_VAR
31 EG_PNEUCONTROL_THR
32 EG_PNEUCONTROL_INIT
83

33 EG_MEASURECONTROL
34 EG_SV
35 EG_SEND_EVENT
36 EG_NEXT_MIX
37 EG_NEXT_PIERCE
38 EG_INIT_PORT
39 EG_INIT_FPGA
40 EG_INIT_PIC
41 EG_INIT_MDATA
42 EG_LOOP
43 EG_SAMPLER_DELEG
44 EG_MOVING_DELEG
45 EG_NMOVING_DELEG
46 EG_COMM_PARAM
47 EG_TEST_DISLPAY
48 EG_INIT_STATE
49 EG_FILE_NOT_FOUND
50 EG_FILE_NO_INFORMATION
51 EG_FILE_OPEN
52 EG_GET_FILE
53 EG_SEND_FILE
54 EG_CRC_CREATE
55 EG_UPGRADE
56 EG_TEST_SPEC
57 EG_ANSWER
58 EG_COMMUNICATION
59 EG_DRAIN
60 EG_NEEDLESETTINGS
61 EG_HARDCLEANING
62 EG_MAININIT
63 EG_WAITFORCONTROLEVENT
64 EG_S_MSGMLR_TIMEOUT
65 EG_NM_MSGMLR_TIMEOUT
66 EG_M_MSGMLR_TIMEOUT
67 EG_PRESSURE
68 EG_THREADS
69 EG_WAITFOREVENT
70 EG_SENDEVENT
71 EG_DELETE_FILE
72 EG_SETTINGS
73 EG_REAGENTS
84

74 EG_SAMPLER_CONSTRUCTOR
75 EG_SAMPLER_INIT
76 EG_SAMPLER_STOP
77 EG_SAMPLER_GETSTATE
78 EG_SAMPLER_SETSTATE
79 EG_SAMPLER_GETPOSITION
80 EG_SAMPLER_SETPOSITION
81 EG_SAMPLER_GETCONTAINERTYPE
82 EG_SAMPLER_CANCEL_WAITSTATE
83 EG_SAMPLER_WAITSTATE
84 EG_SAMPLER_ONLINE
85 EG_SAMPLER_SAMPLERDY
86 EG_SAMPLER_ERROR
87 EG_SAMPLER_FATALERROR
88 EG_SAMPLER_SENDPOS
89 EG_SAMPLER_BARCODE
90 EG_SAMPLER_INITDONE
91 EG_SAMPLER_STOPPED
92 EG_SAMPLER_TRAYDONE
93 EG_SAMPLER_PONG
94 EG_SAMPLER_SAMPLEMIX
95 EG_SAMPLER_BAROK
96 EG_SAMPLER_SENDTRAYINFO
97 EG_SAMPLER_SENDRACKINFO
98 EG_SAMPLER_UNKNOWNCMD
99 EG_SAMPLER_CMD
100 EG_SAMPLER_TRAY_DONE
101 EG_SAMPLER_STARTPOS
102 EG_AUTOSAMPLER_MEASURE
103 EG_FILL
104 EG_TEST_PIERCE
105 EG_MOVE_DILUTOR_1
106 EG_MOVE_DILUTOR_2
107 EG_MOVE_DILUTOR_3
108 EG_MOVE_DILUTOR_4
109 EG_BOB
110 EG_HORIZONTAL
111 EG_VERTICAL
112 EG_DILUFULL
113 EG_GET_BUBBLE
114 EG_BUBBLING
85

115 EG_MICRO
116 EG_MEAPUFF
117 EG_MEAPUFFBIG
118 EG_MEAPUFFSMALL
119 EG_PREDIL
120 EG_DIL_L
121 EG_SAMPLE_MOVE
122 EG_NEEDLE_WASHING
123 EG_WBC_STOP
124 EG_MIX_SAMPLE
125 EG_CHAMBER_MIX
126 EG_CHAMBER_RBC
127 EG_CHAMBER_WBC
128 EG_HGB_MEASURE
129 EG_BASO_MEASURE
130 EG_BASO_SAMPLE
131 EG_RBC_MEASURE
132 EG_WASHING_DIFF
133 EG_PREDRAIN
134 EG_SENDTHEEND
135 EG_WBC_MEASURE
136 EG_OPT_SAMP_CL
137 EG_SET_POSITION
138 EG_BURNRBC
139 EG_BURNWBC
140 EG_TIMEOUT
141 EG_TOTTI
142 EG_AAMVERT
143 EG_AAMHORI
144 EG_AUTO_ALIGNEMENT
145 EG_STOP_PUFFER
146 EG_LYSE_PUFFER
147 EG_DILUENT_PUFFER
148 EG_SHEATH_PUFFER
149 EG_PRIME_DILU
150 EG_PRIME_SHEATH
151 EG_PRIME_LYSE
152 EG_PRIME_STOPPER
153 EG_AS_ERROR
154 EG_RINSE
155 EG_PNEU_INIT_2
86

156 EG_DELAY
157 EG_WAKEUP
158 EG_STANDBY
159 EG_CHECKPRIME
160 EG_SELFTEST
161 EG_LASERHEAT
162 EG_LASERHEAT_INFO
163 EG_MACHINEINFO
164 EG_FILELIST
165 EG_SV_CLEAN
166 EG_AUTOSAMPLER
167 EG_EXTRA1
168 EG_EXTRA2
169 EG_EXTRA3
170 EG_UDILUTOR
171 EG_CHAMBERS
172 EG_CELL_CLOGGING
173 EG_FILL_LASER
174 EG_DEBUG
175 EG_REPAIR
176 EG_CANCEL_ANSWER
177 EG_CANCEL_AT_END
178 EG_PIC_ERROR
179 EG_SET_OFFSET
180 EG_AS_INFO
181 EG_REAGENT_LOCK
182 EG_RV_RD_ERROR
183 EG_RV_WR_FAULT_ERROR
184 EG_RV_WR_PROTECT_ERROR
185 EG_RV_ARG_ERROR
186 EG_RV_UNEXPECTED_ERROR
187 EG_RV_KEY_ERROR
188 EG_RV_CRC_ERROR
189 EG_RV_PERMISSION_ERROR
190 EG_RV_DEVICE_MISSING_ERROR
191 EG_RV_NOT_ENOUGH_MEASUREMENTS
192 EG_SERIALS
193 EG_OFFLINERINSING
194 EG_DC
195 EG_CLEANCELL
196 EG_CLEANDILU
87

197 EG_QCWIZARD
198 EG_QCTEST
199 EG_CHECK_ALTITUDE
200 EG_RELEASE_AIR
201 EG_AS_COVER
202 EG_BLOODSENSOR
203 EG_BLOODSENSOR_CALIB
204 EG_OVERNIGHT_CLEAN
205 EG_PRESSURE_OFFSET
206 EG_CANCEL_REPAIR
207 EG_DRAIN_TCU
208 EG_FILL_TCU
209 EG_DRAIN_FLOW_CELL
210 EG_FILL_FLOW_CELL
211 EG_XY
212 EG_EMPTYRCHAMBER
213 EG_EMPTYWCHAMBER

7.3 Service Menu


There is a Service menu for servicing and operation checking purposes; however it is hidden from the
standard user. To access the service menu:

With on screen keyboard active:

 Go to the Main Menu

 Tap and hold the Main Menu icon for appr. 6 seconds

 Accompanied by a “bing” sound, the password entry dialog will be displayed.

 Enter the service code (6484A5)

 The service menu icon will be displayed

With onscreen keyboard inactive

 Connect an external USB keyboard

 Go to the Main Menu

 Make sure the arrow is over the Main Menu icon

 On the external keyboard, type 6484A5


88

 Press enter

 The service menu icon will be displayed

Enter the following code to access service menu: 6484A5


Password is case sensitive

Once you typed in the correct service code, you will enter into the service menu. The system is designed
to provide you continuous access to service functions: you will have to enter the service code once for a
specific service session.

The button in the lower right corner says “Service mode OFF”. This is the EXIT point of the service menu.

Do not forget to end service operation by tapping the “Service mode off” button to prevent users
accessing the service menu.

The following screen displays available operations and functions. Refer to the menu screen above to
locate the description of the feature.
89

7.3.1 Service functions


90

7.3.1.1 Test functions

This item allows running specific test procedures. You have to enter the “ID” of the required service and
RUN the process. The list of available functions and their expected result is listed below.

ID Name Expected result


2 Go to Standby The system will go into standby independently of standby timer
status
3 Drain tubing system It is equivalent to tapping the “Drain all” function under Main
Menu / Maintenance / Drain all
4 Wakeup The system performs wakeup function.
15 Skip clean The system will assume Cleaning has been performed
61 Skip (next) Pneuinit The system will omit pneuinit
62 Skip (next) Rinse The system will omit Rinse (tube washing process)
63 Make PneuInit Mandatory The system will be forced to do a pneuinit (as if it was just
powered on)
64 Make Pneuinit and Prime All motors and the tube system is set to uninitialized state.
Mandatory (equivalent to power off)
65 Make Fill Mandatory Equivalent to first startup – the system will assume that all
pneumatical systems are empty.
66 Skip Pneuinit, Clean and Combination of various functions
Wakeup
67 Skip Pneuinit, Wakeup Combination of various functions
68 Skip Pneuinit, Rinse and Combination of various functions
Wakeup

7.3.1.2 Low level file management

Allows accessing files located in the DimmBoard’s SD card.

7.3.1.3 Low level reboot (DimmBoard)

This command allows restarting the Low Level software which controls the pneumatic system. This might
be necessary when the communication is lost between the High and the Low level systems. Typically:
when the Start button remains RED without pneumatically action and the system does not respond to
user commands. This function forces the DimmBoard controlled system into a known basic state.

7.3.1.4 Pneumatical System – Initialize

When Initialize starts, the system moves all mechanical components to their "standby" position. This
function allows starting up pneumatics after mechanical alignments or a system startup. The system will
perform an Init process after Wash head and Needle alignment or when Service testing menu was used.
91

7.3.1.5 Network management

This is the input area to define network related parameters in connection with LIS communication. The
parameters required for setting up the link should be acquired from the system operator of the network
you want to connect the analyzer to.

7.3.1.6 RAW data saving mode

Allows selection of RAW (unprocessed) data file save mode. Raw data files can be important for Diatron
support or R&D team to analyze and troubleshoot. Turn on Raw data saving mode to make the system
save measurement data during further measurements.

7.3.1.7 Export RAW files

Allows saving collected RAW data files to a removable storage media, typically USB Flash memory.

7.3.1.8 Windows Control Status

This setting controls whether the CTRL-ALT-DEL combination can be used to access system functions or
not. This setting is suggested to be turned OFF after installation of the analyzer at the end user’s site to
prevent unauthorized access to system files. This setting requires a system reboot.

7.3.1.9 TCU fill/drain

Allows to fill the TCU tubing system with reagents or remove all liquid from the tubing. Recommended to
use draining before and fill after TCU removing.

7.3.1.10 Flow cell fill/drain

Fills up or completely removes all liquid from the optical head. Use drain before removing laser head.

7.3.1.11 Blood sensor status

Turns on and off the blood sensor

7.3.1.12 Reset statistics

This function resets the statistic screen. The statistic contains information about the instrument's history.
It shows the number of startups, shutdowns, errors, etc. See section 16.4 in User Manual.

7.3.1.13 Hard drive maintenance

Makes a defragmentation on the hard disc. It can be used to make the system faster is database contains
too many measurements.
92

7.3.1.14 Show advanced DB view

Changes the database screen to contain more information

7.3.1.15 Small sample mode

Enables to use small sample mode if it's installed into the system.

7.3.1.16 Pre-diluted mode

Enables to use pre-diluted blood sample.

7.3.1.17 Monitoring

Allows to connect the system to an external computer by entering its IP address. If monitoring is
enabled, it's possible to see how the unit is used in real time on a PC or laptop which is located in
another area.

7.3.2 Service testing


Service Testing menu provides tools for checking
hardware.

 Valves - In the menu you can see buttons


(each represents a valve). The state of buttons
is similar to the state of valves.

 Motors - In the menu you can see buttons


(each represents a command to motor). By
pressing a button you can send the motor to
the requested end-position.

 Pumps - In the menu you can see buttons


(each represents a pump).By pressing the
buttons you can turn on and off the pumps.

 Also allows to monitor pressures in real time


on all three pressure sensors.
93

7.3.3 Service calibration


The analyzer provides a menu for Service calibration
purposes. This screen displays the actual calibration
factors, service level calibration factors and the user
level calibration factors. Service and User level factors
are multiplied and the system is using the result as the
actual calibration factor for each measured value.

Example for RBC:

RBCDisp. = FactRBC User * FactRBC Serv * RBCMeasured

Service calibration screen displays both factors for the


values measured in the chambers and in the optical
head.

7.3.3.1 Enable scatter calibration

This checkbox has to be enabled to make a scatter calibration. The system will ask for the scatter
calibration file (AS file) which contains the gravity values of the selected control blood. After the
AS file is selected, push calibration to measure with control blood. After measurements are
done, the system will display the new scatter values.

7.3.3.2 Backup Calibration

It is possible to save the calibration factors to a ‘bak’ file. This option is recommended before
reinstalling the operation system or the High Level software. Save the calibration file to a
pendrive.

7.3.3.3 Restore Calibration

It is possible to restore the saved calibration values from a pen drive by pushing Restore
Calibration.

7.3.3.4 Load Calibration

The system is capable to upload calibration factors made by the Analyzer Spy software. This is
useful when Backup calibration option was missed before software reinstall and all the factors
are reseted to 1.00. The Analyzer Spy software can create a fact file with all the defined
calibration factors. Load the fact file with Load Calibration and not with Restore Calibration.
94

7.3.3.5 Accept User factors

Pushing this button will reset the User factor values to 1.00. Reset values will be multiplied with
Service factors, creating new Service factors.

7.3.3.6 Back

Returns to Service menu screen.

7.3.3.7 Accept

Actual calibration factors can be modified by tapping the touch screen. After new values are
typed in, Accept has to be pushed to make the system use the new factors.

7.3.4 Stress measure


In Stress mode, the instrument performs measuring
cycles without sample (blank measurements)
continuously. This can be used for burn-in tests, or to
check pneumatic system after changing any main
fluidic parts.

You can have information about stability, cleanliness,


HGB operation, and counting time stability.

It is possible to Load last stress results to gather more


information of the analyzer.
95

7.3.5 Auto alignment


This function allows access to adjusting the optical
head’s laser alignment. The page is password
protected, but some functions are accessible without
entering the password.

 Laser settings can be typed in manually. After


values are entered, press Save values to
confirm the settings.

 By pressing Read DC, it is possible to check


the status of the optical head. DC level is also
indicated in Self-test menu. If DC level is in
range it means that the flow cell and the laser
diode is proper. If the DC is high or not
steady, flow cell cleaning needs to be done.

Always type in the new Laser settings if optical head


was replaced.

7.3.6 AS (Auto-sampler)
This function allows accessing and testing various
functions of the Auto Sampler.

7.3.6.1 Go to Pos

Pulls the selected rack position -defined by


the RackPos and SamplePos-in front of the
mixer.

7.3.6.2 Reset

Sets the Auto-Sampler to default position

7.3.6.3 Repeat last command

Repeats the previous action.

7.3.6.4 Go To Mixer

Pulls the selected rack position -defined by


the RackPos and SamplePos-into the mixer.
96

7.3.6.5 Load to sampling position

Pulls the selected rack position -defined by the RackPos and SamplePos-to the needle

7.3.6.6 Set up Barcode Reader

7.3.7 Adjustments
This page contains software and mechanical
adjustments. These adjustments and values have to
be set correctly for proper working.

 Optical channels offset

Reduces the offset voltages in the optical


head. Can be set by hand or automatically.

 Impedance amplifier offset

Reduces the offset voltages in the amplifier


board. Can be set by hand or automatically.

 Needle tube full length/sample


position

Sets the correct blood position by default or


by using Blood sensor.

 Pressure offset adjustment

Sets the pressure sensors to reduce the sensed pressure offsets during measurements.

 Needle & wash head adjustments

Sets the mechanical parts to position for checking and adjusting.

 Also possible to retype the system's serial number after a hardware (DimmBoard,
LSDACQ, Hardware block, HDD) change.
97

7.3.8 Multiuser Settings


In this page the Administrator can create new user
profiles or delete existing users.

More than one user can have access to the analyzer if


Multiuser mode enabled checkbox is set.

The user will have Administrator rights if the


Administrator checkbox is enabled.

7.3.9 MDA view


This screen allows monitoring activity of the Low Level
software in the DimmBoard. Regularly,
communication and pneumatic procedures post
status report on this screen.

Diatron Support may ask you to open this window and


report messages.

MdaOn checkbox has to be enabled to see the activity


of the Low Level software.
98

7.3.10 Software upgrade


The software of the Instrument can be upgraded using
a commercially available USB flash memory device.

Abacus 5 software can be upgraded with software


releases from Diatron. It is recommended to install
both upper (Windows XP) and lower (DimmBoard)
software from the same release package to keep
maximum performance and compatibility. However
software releases come in packages, it is possible to
install the upper and lower software independently.

7.3.10.1 Refresh data

Shows the serial number and all software and


firmware versions installed on the analyzer.

7.3.10.2 Change High Level software

Brings up the screen where the High Level software


file (msi file) can be selected (browsed). Make sure that the file is prepared on a pen drive attached to
the instrument.

7.3.10.3 Change Low level software

Brings up the screen where the Low Level software file (lls file) can be selected (browsed). Make sure
that the file is prepared on a pen drive attached to the instrument.

7.3.10.4 Change Low level boot

This button needs to be used when the system's software has to be updated from 1.3.xxx to 1.4.xxx

7.3.10.5 Change LSDACQ firmware

Brings up the screen where the LSDACQ firmware file can be selected (browsed). Make sure that the file
is prepared on a pen drive attached to the instrument.

7.3.10.6 Change Optical Head Firmware

The optical head's (laser) firmware can be updated here.

7.3.10.7 Change TCU software

The TCU board also contains software which can be updated here.
99

7.3.10.8 Change AutoSampler software

The Auto-Sampler has software similar to the Low Level software in the DimmBoard. This software can
be controlled by the High Level software. It is possible to update the software here is new software is
enable.

7.3.11 Reagent Lock


The analyzer can keep track of measurements and will
change (decrease) the number of available
measurements accordingly.

Definitely, some service actions require


measurements with reagents from the User’s licensed
tests. The analyzer will thus decrease the license
counter.

To make up for these measurements, the service


engineer can use the dedicated Service Reagent Key
to “give back” the used number of measurements to
the client.

Upon connection of the Service Reagent Key, the


numbers of available measurements on the key are
displayed. You can enter the number of tests you
need to provide for the end user to compensate them
for the sample runs related to the service operation.

Enter the number of measurements to grant, and click


on the Grant button. The selected number of tests
will be added to the counter of the analyzer, and at
the same time the number of available measurements
on the Service Reagent key will be decreased with the
granted number.

Make sure that the reagent key is connected properly


to the instrument.
100

7.3.12 QC Wizard
It's possible to run several automatic tests on the analyzer. QC Wizard is used by Diatron production to
ensure that all mechanical, pneumatic and electrical settings are correct.

QC wizard requires a 'lim' file containing ranges for all the tests.

7.3.13 Printer installation


Abacus 5 supports all Windows XP compatible printers.
To install a printer or a printer driver, you are going to
need either an external USB CD or DVD drive for the
original printer driver CD, or a USB flash drive with the
printer driver.

This option allows installing a printer on Abacus5.


Normally, users are not allowed to connect and install
printers, since all peripherals must be installed,
connected by the representative of the service
provider of the analyzer. The add printer button
initiates the Add printer Wizard from Windows. Follow
instructions on the screen to install the printer.

7.3.14 Factory Settings


Accessible to the Manufacturer only.

7.3.15 Service mode OFF


Clicking on this button will turn service mode off, and puts you back to the User Menu. To re-enter
Service menu, you need to type in the service code again.

Make sure you finish any service related actions requiring service menu access with clicking on this
button to keep the end user away from these functions.
101

7.4 Software systems


Abacus5 has a complex software system. This
system can be divided into two parts, low level
system and high level system. The low level
system (DimmBoard SW) is responsible for the
moving parts of the analyzer and the high level
system is responsible for the user interactions.
The high level system is running on a built-in
industrial PC with an operating system
(Microsoft Windows XP Embedded).

7.5 General SW installation


guide
 Download the install package from
www.diatron.com

DiatronOpticalFrame.msi

Upgrade.lls

 Copy all files to a USB memory stick.


 Power on the Analyzer – do not perform
pneumatic operations (like measurement).
 Connect the USB memory stick to the
Analyzer. Wait a few seconds until the LED
in the memory stick starts flashing.
 Go to Service Menu (enter the code:
6484A5)
 Go to SW upgrade. Press “Refresh data” to
see SW versions of individual components.
 Write down SW versions to a paper (or tap
Print icon to print the image).
 Upgrade software in the following order:

Low level SW – tap the “Change Low Level


SW”. Locate the lls file on the “E:” drive. (By
default, the USB stick should be the E: drive. If
not, please browse to the USB stick (F:, …)
102

 Select the file by tapping the checkbox next to the file, then clicking OK. The process will start –
you can see the progress on the progress bar.
 High Level SW- Click on “Change High Level SW” – the analyzer will ask you to locate the
DiatronOpticalFrame.msi file on the USB stick. Select the file by tapping the checkbox next to
the file, then clicking OK. The process will start. Click “next, next…install”.
 Wait for the process to end.
 Click on “Finish” (the new SW will start).

Note: Software restart (especially if there are more than 1000 samples present in the database) might
take several minutes. During this process, the screen will be blank (black – with the mouse pointer
displayed). Do not interrupt the process by turning the analyzer off.
103

7.5.1 Install operating system Windows 2.0 install using pendrive


Tools needed:

- Pendrive (at least 4 GB)


- Pendrive with high level software
- Windows iso image file 2.0 (winpe_x86_v2.0)
- Rufus v1.2.0(or higher) application
- Additional files directory
- External usb keyboard and mouse

Step 1.

To install windows from pendrive, a bootable pendrive has to be created first.

Copy the additional files directory and the software rufus_v1.2.0. to your hard disk.

Make sure that the windows 2.0 iso file (winpe_x86_v2.0) is also downloaded to the PC’s hard disk.

Step 2.

Insert the pen drive (4 GB or higher) into the PC

Step 3.

Run the rufus exe.

„Rufus” will automatically find the external device


attached to the PC.

Select NTFS for File system.

Cluster size is default.

Make sure that the „Quick format”, the „Create a


bootable disk using ISO image” and the „Create
extended label and icon files” checkboxes are
checked/marked. Click on the disk button to select
the windows ISO file. Browse and find
winpe_x86_v2.0.iso file. You can also set the
„New volume label” to a name of your choice.

If setting is ready, press Start button.


104

Step 4.

Let „Rufus” copy the ISO to the pendrive.

It will take approximately 80-120 minutes,


depending on the hardware.

Do not remove the pendrive during the copy!

Step 5.

When copying if finished, you have to


select the files in the Additional files
directory and copy it to the root of the
pendrive.

The pendrive is now a bootable windows


installer. The Pendrive should look like in
the image on the left.
105

Step 6.
Now we have to set the hematology analyzer to recognize to Pendrive.

Insert the pendrive to the hematology analyzer, also attach the USB keyboard and turn the analyzer
on. While the booting process is going, press „del” button to enter to BIOS.

In Advanced BIOS Features, set the first boot device to Hard disk.

On the same screen you have to enter to Hard Disk Boot Priority. Select the Pendrive to be first, so
the system will use the Pendrive as an external Hard Disk (use + and – buttons to change the order of
HDD).

If setting is ready, press F10 and save to exit from BIOS and restart.
106

Step 7.
Warning: In case of a new HDD, it can happen
that the „e” drive is not available. In this case,
remove the bootable pendrive and type in the
below commands into the console to create
proper partition on the HDD.

DISKPART
SELECT DISK 0
CLEAN
CREATE PARTITION PRIMARY SIZE=100000
FORMAT FS=NTFS QUICK
ACTIVE
ASSIGN LETTER = C
CREATE PARTITION PRIMARY
FORMAT FS=NTFS QUICK
ASSIGN LETTER = D
EXIT
EXIT
Replace the pendrive then wait until console
window returns.

If „e” drive exists type in „installpen” to start


the windows install. The system will offer to
options, first one is to recover the windows
the second one is a full format. The recover
format will keep the database, the full format
will erase all data from the HDD (which is not
a problem if the HDD is new). Choose number
1 or 2 then press enter to start the process.
The windows install/recovery would take
approximately 30 minutes. If the install is
done, remove the pendrive and type „exit” to
restart the system. If you didn’t remove the
pendrive, don’t press any key when it’s
The system now restarts and a message will offered, let the system start windows from
appear on the screen „Press any key to boot HDD. If the touch screen is not calibrated, you
from USB”. Press any key, than wait until have to use a mouse to find touch screen
windows command console window calibration on the desktop. Start touchcalib
appears. If you miss pressing any key, the and calibrate the screen, by pushing the 4
system will stop with an error. In this case, corners (crosses) on the touchscreen. If touch
restart the system with CTRL+ALT+DEL. As is calibrated, insert the second pendrive with
the windows console window appears and the high level software installer, start total
it’s ready to receive the commands, type in commander and run the high level software’s
„e:”. msi file.
Step 8.

After install is done, the high level software has to be set properly.

Please adjust or set:

- Calibration factors (load or restore calibration factors if it’s saved before)

- Laser setting (power, low and high angle gains have to be set again for proper scatter values)

- Offsets on amplifiers (optical measurement and impedance measurement offsets have to be set
(zeroed) in adjustments menu)

- Pressure sensor offsetting (also have to be set to make the pressure sensor work properly)

- Set the serial number of the instrument (the serial number is also stored on the HDD, after
windows install the serial have to be set again. Use adjustments menu to enter the same serial
number as on the DIMMPC/DIMMBOARD or on the sticker on the back of the instrument).

Windows and high level software install is done.

7.5.2 Installing a Printer


Abacus5 is using a Windows operating system. To add a new printer on Abacus 5 you have to install the
printer specific driver. The operating system has limited possibilities for user interaction affecting
operating system functions. Follow the steps below to install a printer, or use one of the printer install
functions from the service menu (see section 7.3.13).

1. Prepare the printer driver on a USB stick. (Either copy it from the install CD, or download it from
the printer manufacturer’s web site)
2. Connect a USB external keyboard.
3. Connect the USB stick with the printer driver to an available USB slot on the analyzer.
4. Make sure that Windows control function is Enabled in Service functions menu (section 7.3.1.8).
5. Press CTRL-ALT-DEL on external keyboard to access „Task manager”
6. Select „start new process”
7. A file dialog box appears. Browse the file system for the USB stick and locate the install package
of the printer. Run the application (printer driver). Follow instructions on the screen.
8. Upon completion the printer becomes available as an installed printer for Abacus 5
9. Exit task manager (close)
108

8 Installation
8.1 Check the Delivery
When you receive the ‘Abacus 5’ analyzer delivery, please ensure that the packaging is not damaged.
Check the bill of lading accompanying the package against your order documents and ensure that the
shipment is complete and that all documentation is in order. If you have ordered an optional
Autosampler, it will arrive in its own package. Please contact your sales representative, service
representative, or shipper if there are any discrepancies in the shipping documentation or any visible
damage to any of the packaging.

Please follow all applicable laws or regulations regarding the handling or opening of the ‘Abacus 5’
analyzer packaging.

8.2 Prepare For Initial Installation


Before you start the installation please ensure the following:

 Locate a suitable place for the ‘Abacus 5’ instrument


 Identify any additional laboratory personnel that will observe the installation process
 Have the contact information for your Diatron representative or service engineer
 Arrange for additional support (IT specialist, electrician, etc.) if necessary during the installation
 Understand and follow the analyzer General Precautions listed in section 2.2.

8.2.1 Select a Suitable Location


Select a location for the ‘Abacus 5’ analyzer that meets your laboratory requirements for safety,
ergonomics and efficient workflow. The location should also meet the environmental, electrical, and
safety requirements of the ‘Abacus 5’ listed in section 2.

It is important to install the instrument in a suitable location. A poor location can adversely
affect its performance. Consider the space and weight requirements listed in sections 2.6
and 2.7.

To allow reliable operation and to provide a safe working environment, make sure that the
table supporting the unit is stable enough to carry the weight of the instrument and
accessories. Reagents should never be placed above the analyzer to avoid spill hazards.

8.2.2 Make Any Special Arrangements


If you decide to route the reagent tubing through the tabletop, please ensure that any necessary holes
are drilled before the installation process starts.
109

If you plan to connect the ‘Abacus 5’ analyzer to any external devices (keyboard, mouse, printer, host
computer, etc.), please ensure that all necessary preparations (cable-channels, cable-binders, drilling
through tables, walls etc.) are complete before the installation begins.

8.2.3 Gather Your Peripherals Devices


Collect any external keyboard, mouse, bar code reader, or printer that you will be attaching to the
‘Abacus 5’ analyzer. Although the Windows® XP® Embedded operating system installed on the ‘Abacus 5’
analyzer is capable of automatically recognizing multiple peripheral devices, please ensure you have
available any installation disks or drivers provided by the device vendor. See section 3.4.7. to learn
peripheral connections on the hardware block.

8.3 Performing the Installation


Now that a location has been selected and all preparations are complete, you are ready to begin the
installation. Initial installation should only be done by Diatron certified service personnel. You can find
the installation checklist in section 8.4.

8.3.1 Move the ‘Abacus 5’ to the Selected Location


Always have two persons present and use safe lifting procedures when lifting the ‘Abacus 5’ analyzer.
Safely move the ‘Abacus 5’ analyzer, the accessory box, and the optional Autosampler (if ordered) and
the Autosampler accessories to the selected location. Keep the ‘Abacus 5’ analyzer in an upright position
and gently set it down in its new location.

8.3.2 Visual Inspection


Visually inspect the ‘Abacus 5’ before proceeding with the installation and verify that:
 The front panel is be free of cracks or scratches
 The display screen is free of cracks or scratches
 The top, sides, and back panels are free of dents or scratches

Open the front panel of the unit and visually verify that:
 The syringes are not cracked
 The shear valve has the protective card installed
 There is no fluid inside the tubing
 There is no salt buildup inside the tubing

Remove the side covers of the unit and verify that:


 The tubes on the valves are in position and not popped off from (note that valve # 4 has a free
connection)
 Cables are not loose and they are well connected
 The reagent puffers are not cracked and there is no salt buildup on their sides and bottom
 Chambers are not cracked
110

Visually inspect the Autosampler and verify that:


 The outside housing of the Autosampler is free of dents or scratches
 The transparent cover opens and closes smoothly
 The sample tray and the sample racks have no visible damage

8.3.3 Remove the Protective Card from the Shear Valve


The shear valve comes equipped with a protective plastic card between the ceramic disks of the shear
valve to prevent damage and adhesion during transportation.

To remove the protective card, perform the following steps:

 Open the front cover of the ‘Abacus 5’ analyzer


 Locate the white protective plastic card in the shear valve as shown in Hiba! A hivatkozási forrás
em található.
 Gently pull out the card
 Check and tighten the locking screw of the shear valve if necessary

2. Figure: Shear Valve Protective Card

8.3.4 Connect the Optional Auto-Sampler


To connect the optional Auto-Sampler to the ‘Abacus 5’ analyzer, perform the following steps:

 Unscrew the five screws and remove the small cover plate on the right side of the ‘Abacus 5’.
111

 Check that the connection surface is clean and there are no cables or other obstructions blocking
the opening.
 Gently push the Auto-Sampler into the ‘Abacus 5’ until the clamps are locked.
 Place the white racks with the tray into the Auto-Sampler.
 Close the Auto-Sampler's cover.
 Pull then push back the Sampler's tray to see that it can move front and back.

8.3.5 Connect the Reagents


Place the reagent containers as shown in section 2.6.

Open the protective package containing the reagent and waste tubes. When connecting or changing
reagent containers, please ensure that the reagent caps and tubes are protected and do not touch the
floor or other surfaces. This can lead to contamination of the reagents and the ‘Abacus 5’ analyzer.

To connect the reagents to the ’Abacus 5’ analyzer, perform the following steps:

 Push the color-coded reagent and waste tubes all the way on to the matching color-coded
reagent connectors on the back panel of the ‘Abacus 5’ analyzer.
o Green: Diatro•Dil 5P Diluent
o Orange: Diatro•Diff-5p
o Yellow: Diatro•Lyse-5p
o Red: waste container
 Route the reagent cap and tube to the matching reagent container, ensuring that the reagent
tubes are not bent, broken, twisted or blocked.
112

 Place the reagent or waste tube in the matching reagent or waste container and screw the cap
on to the container.

Only genuine Diatron reagents should be used with the ‘Abacus 5’ analyzer

The analyzer operates with chemically and biologically active reagents. Physical
contact with these reagents should be avoided. Please read reagent descriptions
carefully for possible emergency actions.

8.3.6 Connect the Power Cord


Before connecting the power cord, make sure that all of the ‘Abacus 5’ back panel switches and the
optional Autosampler switches are turned off:

 Turn off the main power switch (small switch) on the rear panel of the ‘Abacus 5’ analyzer near
the power connection to the ‘down’ position labeled ‘0’.
 If the optional Autosampler is installed, turn the power switch on the right side of the
Autosampler to the ‘off’ position labeled ‘0’.
Connect one end of the power cord to the power connection of the ‘Abacus 5’ analyzer, and the other
end of the power cord to an appropriate wall outlet.

The ‘Abacus 5’ analyzer should only be operated from a wall outlet capable of
meeting the electrical requirements listed in section 2.4.

8.3.7 Verify the ‘Abacus 5’ Computer Operation


We are now going to start up, then shutdown the ‘Abacus 5’ analyzer to ensure that the computer
system and software start and shut down correctly.

To power and shut down up the ‘Abacus 5’ analyzer, perform the following steps:

 Turn on the main power switch on the rear panel of the ‘Abacus 5’ analyzer near the power
connection to the ‘up’ position labeled ‘1’.
 Flip the power -on switch near the top of the rear panel of the ‘Abacus 5’ analyzer to the ‘up’
position.
 If the optional Autosampler is installed, do not turn on its power switch at this time.
 Wait until 'Abacus 5' boots up and the software loads. This will take a few minutes.
 Ensure that the software displays the main menu and that no warning or error messages are
displayed. Warnings are displayed on the bottom of the LCD as a yellow exclamation mark. Shut
down the ‘Abacus 5’ analyzer. DO NOT simply turn off the power switches to shut down the
‘Abacus 5’ analyzer. The analyzer requires a specific shutdown sequence. See chapter 7.3.2. in
User Manual for information about the shutdown procedure.
113

8.3.8 Connect Peripherals


Ensure that the ‘Abacus 5’ analyzer is completely powered down before connecting the peripherals.

To connect the peripherals, perform the following steps:

 Plug in any external keyboard, mouse, bar code reader into the appropriate port on the back
panel of the ‘Abacus 5’ analyzer.
 If a peripheral device requires its own power supply, plug it in now in a wall socket belonging to
the same socket group for proper grounding. Consult an electrician if you have any questions
about wall socket grounding.
 Turn on the main power switch on the rear panel of the ‘Abacus 5’ analyzer near the power
connection to the ‘up’ position labeled ‘1’.
 Flip the power-on witch near the top of the rear panel of the ‘Abacus 5’ analyzer to the ‘up’
position.
 If the optional Autosampler is installed, do not turn on its power switch at this time.
 Allow the computer inside the ‘Abacus 5’ analyzer a few minutes to start and initialize the
‘Abacus 5’ operating software.
 Perform the software installation of the peripherals devices. The Windows® XP® Embedded
operating system recognizes most of the peripherals without additional installation steps.
 Peripheral devices that require additional installation steps must be installed by a Diatron
certified service engineer.

8.3.9 How to install a printer


Before making the printer install, prepare the printer’s driver files to an USB drive. Make sure that the
driver files are for the printer you want to install. If the printer has an USB connection, there is a simple
way to do the installing:

 connect the USB drive containing the printer divers to the analyzer
 make sure that an external keyboard is also connected to the analyzer
 connect the printer to the analyzer via USB
 if windows supports the printer, after a few seconds, the printer install window should pop up.
Windows recognizes the newly attached hardware and it starts the installing process
automatically by popping up the install window. If you press the touch while the printer install
screen is loading, the window will hide behind the 'Abacus 5' software. Use ALT+TAB to switch
between the two screens
 in the popped up printer install screen, browse the driver files from the pendrive
 add all the necessary printer driver files if needed
 if the files are added properly and install is finished, the printer will appear in settings, printer
menu after „Refresh printer’s list” button is pushed (see user manual’s section 15.6).
 select installed printer and press save
114

If there is a problem during installing procedure or the printer install window is not popping up, there is a
way to install printers from service, printer install menu.

There will be three options for installation:


1, complete setup with SETUP.EXE file from printer driver CD or pendrive. Windows install.
Printer is not connected during installation
2, only driver installation with *.INF file. Will open Windows printer installation wizard procedure.
Printer connected and turned ON during installation
3, only driver installation with *.INF file. Will open Windows printer installation wizard procedure.
Printer is not connected during installation

8.3.10 Initializing the Optional Autosampler


If the optional Autosampler is not installed, skip this section. To initialize the Autosampler, perform the
following steps:

 Turn the power switch on the right side of the Autosampler to the ‘on’ position labeled ‘1’.
 Make sure that the cover of the Autosampler is closed and that the ‘Cover’ led changes to green;
 Double click the AS state on the left side of the Status Bar at the bottom of the screen to bring up
the ‘Autosampler info’ panel.
 Click the Reset button the ‘Autosampler info’ dialog as shown on the left figure below.
 The Autosampler performs a mechanical initialization. When this completes, ensure the HOME
message and the currently installed software version appears as shown circled in red on the
figure on the right below.
 Click the Ok button to close the ‘Autosampler info’ panel.
115

8.3.11 Using the Settings Menu


The settings menu allows modification of various system settings. Click the Settings icon on the Main
Menu to bring up the Settings screen.

The Settings screen contains various buttons that allow the user to change system settings. Each screen
contains a Back button and a Save button. The Back button returns to the Settings screen. The Save
116

button save any changes made in the screen. If the Back button is clicked without clicking the Save
button first, all changes made in the screen will be discarded. See description in the User Manual.

8.3.12 Set Up a Laboratory Information System (LIS)


'Abacus 5' analyzer supports uploading of measurement data to an LIS or host computer. If your
laboratory has an LIS system, the ‘Abacus 5’ analyzer can be connected to it using a serial or an Ethernet
connection. An LIS system is not required to operate the ‘Abacus 5’ analyzer.

Your LIS system must be configured to accept measurement results from the analyzer. ‘Abacus 5’ uses
the Diatron 3.1 protocol to communicate over a serial connection, and the HL7 (version 2.5 or higher) to
communicate over an Ethernet connection. See your LIS vendor to determine if your LIS system is
compatible with the ‘Abacus 5’ analyzer.

LIS configuration requires a moderate level of familiarity with computer settings and some
understanding of computer data communications. If you are not comfortable setting up the ‘Abacus 5’
LIS connection, consult your Diatron certified service engineer.

Use the ‘External devices’ button in the Settings screen to set up LIS connection options.
117

8.3.13 Set Up a Serial LIS Connection


To connect the ‘Abacus 5’ analyzer to an LIS system using a serial LIS connection, perform the following
steps:

 Ensure that your LIS system is compatible with the Diatron 3.1 protocol using a serial connection.
 Connect a serial cable (null modem or modem eliminator) between the COM 1 port on the back
panel of the ‘Abacus 5’ analyzer and the host system.
 Select the appropriate ‘Sending port baud rate’ on the ‘Abacus 5’ ‘External devices’ screen.
o Select 9600 baud if your serial cable is longer than 5m (~15’).
o Select either 9600 or 115200 baud if your serial cable is shorter than 5m (~15’).
 Check the ‘Automatic LIS’ check box if you want every result to be transmitted automatically to
the LIS. Uncheck if you want to manually select and transmit database records to the LIS.
 Uncheck the LIS check box.
 Uncheck the ‘Bidirectional LIS’ check box.
 Click the Save button to save your changes.
 Click the Back button to exit the External devices screen.
 Configure your LIS system to accept measurement results from the ‘Abacus 5’ analyzer using
these settings.
A copy of the ‘Diatron Communication protocol’ description is available from your sales representative or
by download from the Support section of the Diatron web site (http://www.diatron.com/).

8.3.14 Set Up an Ethernet LIS Connection


Connecting the ‘Abacus 5’ analyzer using an Ethernet connection requires connecting the analyzer to a
local area network. Consult your IT administrator about connecting the ‘Abacus 5’ analyzer to the local
area network of your laboratory. The ‘Abacus 5’ analyzer is configured at the factory to “Obtain an IP
address automatically.” If this setting is not suitable for your local area network and must be changed,
only a Diatron certified service engineer can complete the installation.

To connect the ‘Abacus 5’ analyzer to an LIS system using an Ethernet LIS connection, perform the
following steps:

 Ensure that your LIS system is compatible with the HL7 version 2.5 protocol or higher using an
Ethernet connection, and has been configured to accept ‘Abacus 5’ compatible HL7 messages.
 Check the ‘Automatic LIS’ check box if you want every result to be transmitted automatically to
the LIS. Uncheck if you want to manually select and transmit database records to the LIS.
 Check the LIS check box.
 Enter the IP address of the LIS host computer.
 Enter the Port of the host computer.
 Click the Save button to save your changes.
 Click the Back button to exit the External devices screen.
 Configure your LIS system to accept measurement results from the ‘Abacus 5’ analyzer using
these settings.
118

8.3.15 Running Blank Samples and Blood Samples for the First Time
The ‘Abacus 5’ analyzer requires a blank measurement to be run every day before the analyzer will allow
you to run blood or control samples. This measurement is also called Background which indicates that
the system is clean and ready to measure blood samples.

The transportation and packaging process sometimes causes minor particles to be present in the
pneumatic components of the analyzer. For this reason, five to ten initial blank measurements should to
be run during installation to flush the system.

Click the Measure icon on the top left side of the screen and run a blank measurement. The Start button
will change color to red. The fill procedure and blank measurement take several minutes to complete.
The start button will change color to green and the blank result screen will be displayed.

After the procedure completes, a blank result screen will be displayed. Click the ‘Start Again’ to run an
additional blank measurement. Repeat this procedure several more times until the blank results are no
longer flagged, or until the blank results are low enough to satisfy laboratory quality standards. If the
Blank measurement is satisfactory, press Accept to make other measurement modes enable.

Change the mode selector to Control and run a control measurement to see that the sample results are
correct. It's important to use a valid, not expired control sample with known target values and ranges.
See chapter 12 in User Manual for additional information about QC on the ‘Abacus 5’ analyzer.

Lastly, it's possible that a calibration procedure is needed on the ‘Abacus 5’ analyzer. See section 11 in
User Manual for additional details regarding ‘Abacus 5’ calibration.
119

8.4 Checklist for installing Abacus 5 hematology analyzer

1. Open the box, carefully take out and place the analyzer on a flat surface. The table or desk
should be able to bear at least 50 kg.
2. If the analyzer was stored in a cooler place than room temperature, allow 2 hours for
acclimatization, otherwise water condensation may happen.
3. Check the analyzer visually. Make sure that there is no visible damage. Open the front cover and
remove the side covers to check the electronics and the tubing. Replace the covers when done.
4. Remove the plastic card from the Shear Valve and check tightness of the locking screw.
5. If you have an Auto-sampler unit, remove the plate which covers the AS docking connector (from
the right side cover) and connect the Auto-sampler to the unit.
6. Place reagents preferably on lower level than the analyzer (max. 1 m difference). Never place
reagents to higher level than the unit itself!
7. Assemble and connect reagent and waste tubes to the corresponding containers and connectors.
8. Connect the power cord.
9. Turn on the analyzer, turn on and initialize the Auto-sampler.
10. Connect other peripherals if needed (keyboard, mouse and printer).
11. Check that all the software adjustments are proper. Use setting menu to adjust the software
according to the customer needs. Set the LIS connection (if the laboratory has any).
12. Check that all the hardware adjustments are proper. Use service/service testing menu to run or
test the pumps, valves and motors.
13. From ’Diagnostics/Self-test’ click on ’Start both’ button and wait for the result.
14. Click on ’Measure’. Pneumatic initialization and „fill” cycle will start. Automatic background cycle
will be performed (at first startup several blank runs might be needed till the results are
acceptable)
After two or three blank measurements, results should fall within the acceptable range, see
section 22.7.9 in User’s manual.
15. Run a control blood in ’Control’ mode or run QC procedure.
16. Check if results are within the acceptable range. Run calibration if necessary.
17. The Abacus 5 is ready for routine measurements.
120

9 Troubleshooting
This chapter will show how to do a step by step troubleshooting on the 'Abacus 5' analyzer. Following
the troubleshooting process ensures that the problem can be identified without making any harm on the
system. This chapter will helps find the root cause of the problem and navigate to the page where you
can find information how to solve the problem.

9.1 Troubleshooting checklist for Abacus 5 hematology analyzer


1. Visual inspection

2. Powering up the system

3. Checking software adjustments

4. Checking hardware adjustments

5. Run a Self-test

6. Run a Blank measurement

7. Run a Control measurement

9.1.1 Visual inspection


The first step during troubleshooting is to check the instrument visually. Some problems can be easily
defined visually before any sample run test, measurements or even to power up the system.

1. Check the front panel and the two side panels; look for scratch or any other damage.

2. Check the back panel of the analyzer, see that the reagents are connected properly, the tubes
are not misplaced.

3. See that the power cord is connected to the analyzer and plugged into the wall socket. The
power supply main switch is flipped to off position.

4. Open the front cover and remove the two side covers. See section 11.

5. Make sure that the tubing are not popped off from the valves, the chambers and reservoirs are
not cracked, there is no liquid on the bottom of the instrument and there are no salt buildups
around the pneumatic system. Check tube between blood sensor and shear valve, see that the
tube is transparent and not whites or has any other color. Replace the tube if it's not clean or
transparent enough. If tube was moved or replaced, the blood sensor and the sampling values
should be realigned. Please see section 6.4 and 6.5 to learn how to calibrate blood sensor and
sample position. Other tube color changes are not so critical, but if there is a tube in the system
121

with a very huge pigmentation, it should also be replaced. Please see tubing map in section
11.16.

6. Check all cable connections. Make sure that the cables are in position on the LSDACQ, PPB
boards, HVB panel and the valve PCB boards. Replace any damaged cable.

9.1.2 Powering up the system


After the visual inspection, it's time to see that the analyzer can turn on properly. Please follow the steps
below:

1. Make sure that the power cord is well connected to the analyzer and that the socket can provide
the proper electric power. Turn the power supply's switch to on position and check that the
power indicating LED on the motherboard is on. If the LED is not on, please check the power
cord. If the cord is ok, we must assume that the power supply is not functioning.

2. Flip the power-on switch on the top back of the analyzer to turn it on. Check the five green
voltage indicator LEDs on the LSDACQ. The five LEDs show that the necessary voltages are
present. If one of the LEDs is not lit, check power supply cable or replace LSDACQ board if
needed.

3. Wait until Windows and the High Level Software of the analyzer loads. If there is a problem
during the boot process or the analyzer's software cannot load, please see section 7.5.1.

4. There is a short audio welcome signal played by the analyzer after the software has loaded. If
you can hear the sound and the software screen is present on the display, then the powering up
was successful. If the display is showing windows, but not the High Level Software, please see
section 7.5.

9.1.3 Checking the software adjustments


Setting the software is important to make proper measurements. Every 'Abacus 5' software was set by
Diatron, but in some cases (windows recovery, high level sw reinstall) this setting has to be done again or
at least has to be checked. 'Abacus 5' analyzer has two software levels. The High Level Software is
responsible to communicate with the user and to give the user's commands to the Low Level Software.
The Low Level Software receives the User's commands and controls the pneumatic and mechanics
system of the analyzer. After low and high level analyzer software is loaded, check the following:

1. On the display, the High Level Software is present; the start button on the front panel is lighting
orange. If the start button is off, it means there is no communication between the two software
levels. See section 7.5 for communication error.

2. Enter to Service menu by entering the service password ‘6484A5’, then go into SW upgrade
screen. Press Refresh data. After a few seconds the analyzer should show all the software and
firmware versions in the system. Make sure that the analyzer has the same version of the Low
122

Level Software and High Level Software installed. If there is a problem occurring during
Refreshing data, please see section 7.3.10.

3. The software gives a warning mark if there is a problem noticed by the system. This is shown as
an exclamation mark in a yellow triangle on the bottom of the display. Double tap on the mark
shows the details of the problem. By pressing 'Go to' button, the software brings up the screen
where the problem can be solved. See section7.2 for detailed warnings.

4. Go to Service Functions screen in Service menu. Check that the necessary checkboxes like blood
sensor status and windows control are enabled. If blood sensor is disabled, please ask the user
about the reason. For blood sensor settings, please see section 6.4.

5. Go to Adjustments screen in Service menu. Check that the Auto-offset checkboxes are enabled in
Optical measurement offset and Impedance amplifier offset fields. Press 'Read data' to see that
the offset values are close to 0. Note that if 'Read data' is pressed, the start button will turn to
red, indicating the communication line between the two software levels are busy reading the
offset values. During the reading, no other commands can be done by the analyzer. Make sure
that the 'Stop data' button is pressed to stop continuous reading of the offset and make the
communication line free again. If offset values are very high during the reading and auto-offset
cannot lower them, it shows an offset error in the LSDACQ. This board has to be changed in this
case.

Check that WBC preheater value is 375 (37,5 °C). Press 'Read data' to see the current
temperature of WBC preheater. If the value is ok, press 'Stop data'.

Press Start adjustment at Pressure sensor adjustment. Adjustment will take the few seconds. The
start button will turn red to indicate that the system is busy. After the adjustment is done, check
that all three offset results are in range. The values shouldn't be higher than 20 mBars. If one of
the values is higher, replace the pressure sensor board.

6. Go to Auto Alignment screen. The system will ask a password, press cancel to enter without
giving the password. In this case, the password protected setting will be unable to adjust. Check
laser power, low level and high level gains. These are Laser Optics related values. Changing one
of them influents the WBC differential population positions on the scattergram. These are
factory defined values for the optical head and they should match to default setting of Laser
Optics in analyzer.

9.1.4 Checking the Hardware adjustments


The proper settings of the mechanical parts are important to make good measurements and to increase
the lifetime of the system. The mechanics are aligned by Diatron during production but some
realignment may need when some mechanical parts are replaced or repaired. These parts can be
checked by Service testing menu.
123

1. Press all 40 valves on. Check that every valve can turn on and off. Make sure that every valve can
be turned on together, than press 'All off' to turn them off. If one of the valves is not responding,
try to replace it with a new one. See section 7.3.2.

2. Press shear valve to Needle position (NP) and Chamber position (CP). Check that the shear valve
discs can turn to the correct position without any grinding noise and that there is no pneumatic
error during the test. For problems related to shear valve, see section 6.2.

3. Turn the sample rotor In and Out. Make sure that the front cover and the washing head is not in
the way. If you receive a pneumatic error, see chapter 6.1.

4. Press 'Rear' and 'Front' buttons at MHori test to move the needle back to the Auto-Sampler
station and back to sample rotor position. Listen the sound of the mechanics, if there is a
grinding noise or a pneumatic error, please see section 6.1.

5. Press 'Down' and 'Up' buttons to see that the needle can move down and up properly. Check
that there is not noise during the movement; the moving mechanics parts are not hitting the not
moving parts. See that the opto LEDs can indicate the needle positions properly. See section
6.1.2 for alignments.

6. Press 'Up' and 'Down' to all Dilutors. Check visually that the syringes can reach the top and
bottom positions without any problems. Look for micro bubbles at the pistons and for leakage at
the syringes. If there is a problem with a syringe and it needs to be changed, see section 6.1.2.

7. Go to Service Functions screen and press 'Wash head' button to see that the position of the
washing head is proper. The system wants you to confirm that you want to make a washing head
position checking/setting, press 'Ok'. If it is in a good position press 'OK' to make a mechanical
initialization of the XY unit. Press 'OK' when it's done. See section 6.1.3 for alignment
information.

8. Checking the needle's position. There is two way to see that the needle is in a good position.

o Move the needle down manually without using the software. If the needle is in down
position, check the distance between the end of the needle and the sample rotor's plate.
The distance should be around 1mm.

o In Service functions screen, press the 'Needle' button. The system wants you to confirm
that you want to make a needle position checking/setting, press 'Ok'. The system moves
the needle up, the needle tip is inside the washing head. If the needle is set correctly,
the end of the needle should be hanging out from the washing head. Press 'Ok' if it is
aligned well.

For needle setting procedure, see section 6.1.4.


124

9.1.5 Run a Self-test


Self-test can be run In Diagnostics menu. This test is fast and automatic, very useful for troubleshooting.
For proper functioning of the analyzer, a Passed selftest is as important as like a clean and acceptable
Blank measurement. For selftest results see chapter 16.1 in User Manual. If selftest is Failed, please see
Self-test guide in section 7.1.

9.1.6 Run Blank measurements


The Blank and sample measurement provides valuable information about the fluidics and electronic
system, but some setting may need to receive the information in Database:

1. If Database only provides those parameters that are important to the clinicians, the Database
parameter setting must be changed. In Setting menu go to System and press 'Set'. Please add
extra columns to the database. Make sure that the following columns are added.

o PrV RBV

o PrV WBC

o Warnings

o Gravity X

o Gravity Y

o Woc 4

o Woc B

Select these columns and add them from the right to the left column block then press Accept
and Save. If you don't press save the setting were not made and the result columns will not
appear in the database.

2. Perform Blank measurements. NOTE: If it's the first measurement after startup, the system
makes an init process, a wake up process and Rinse process. If there was an improper shutdown
before, the system also asks for cleaning. These processes can increase the time of the
troubleshooting and reagent consumption. That’s why it can be bypassed if you type in the
suitable code (66-69) in Service functions on Test functions field. See test functions at section
7.3.1.1. Type in the required test function's number and press 'Run test'. If it is done, the Blank
measurements can run without any other requirements.

3. If a blank measurement is ready, please check that the result is not flagged with warnings. Check
probe voltages. RBC probe voltage (PrV RBC) should be 10-13 Volts, WBC probe voltage (PrV
WBC) should be 20-27 Volts. If there are problems with probe voltages, replace apertures, check
measuring chamber and preamplifier; see section 11.12. If the blank measurement is stopped
with a pneumatic error, troubleshoot as a displayed error codes, see section 7.2. If there is a
problem with popping off tubing in the system, see section 11.16.
125

9.1.7 Run Control measurements


Finally, a measurement with a sample is the most complex testing from all. A measurement gives
information about sampling, calibration, nonfunctional modules, clogged tubing or valves. Measurement
with blood sample and please check the following:

1. The control blood needs to be run in control mode. Please select proper mode.

2. The control blood shouldn't be expired and should have an exact target and range for all values.

3. During sampling, please check visually if the sample position is proper at the shear valve. Please
see section 6.5 to learn more about sampling and setting sampling position.

4. After a measurement is done, check the steps below:

o The results of the control are in range.

o Cell numbers (Woc 4, Woc B) are proper on normal control. Woc numbers (White blood
cell Optical Count) indicates how many cells are present during optical measurement.
These values have only minimum number.

 Woc 4 min : 2.72

 Woc B min : 5.44

o Gravity values are close to the target. Improper or unstable gravity values can indicate a
calibration or an optical head problem. Scatter calibration must be performed. See
section 7.3.3 to learn more.

o Ignore probe voltages during Control measurement. Probe voltage values are different in
Control mode. To check probe voltages, use Blank or Human mode.

o Check that impedance measurement results are proper and close to target. If there is
problem with the results, make a calibration (User Manual section 11) or replace
appropriate part of measuring unit (section 11.12).

o Check HGB result. HGB is accomplished by a photometric measurement technology.


Incorrect HGB result can be caused by dilution or mixing failure in the WBC chamber.
Whole WBC (impedance) count is also incorrect in this case. But it's not related to the
apertures and to amplifier board. If there is a problem with HGB results, please clean the
measuring chamber and replace the HGB head; see section 11.12.4.

o The WBC differential (Ly, Mono, Neu, Eo, Baso) % results are within the limits, the values
are close to target. If there is a problem with the scatter, please do a scatter calibration,
see section 7.3.3, replace the injector in the optics, see section 11.11.1.
126

5. Do measurements from the Auto-Sampler if it is attached. Make sure that the sampler can move
in X and Y direction and that the sample position is proper. See section 7.3.6. to learn more
about the Auto-Sampler settings.

If the analyzer still has a problem despite of all the points of troubleshooting, please contact Diatron
Support Team.

support@diatron.com

www.diatron.com

9.2 Mechanical problems


9.2.1 General guidelines to overcome motor or moving part related problems
Make sure to follow this simple sequence to be able to locate sources of errors and to resolve problems
easily and correctly.

 Always make sure that the part with the problem can be moved freely and easily.
 Make sure that there is enough (and not excessive) amount of lubricant on the moving parts.
 Test the motor with the built in functions in service testing menu. It will always try to operate all
sensors, and will drive the motors with necessary current and power.
 Any Setting or adjusting (mechanical, opto sensor) should be done only after the previous steps
(cleaning, lubricating) did not solve the problem.

9.2.2 Sample Rotor (SR or BOB) failures

9.2.2.1 SR gives grinding noise and / or SW displays SR error messages

 The front panel alignment is not good.


 The front panel is not closed properly, the opening does not match with the sample rotor door.
Fully open or close the front panel to observe correct shear valve operation.
 There is liquid in the SR. If you can see traces of salt around the SR, then the switches might have
gotten damaged.

9.2.2.2 SR error appears during initialization process:

 Sample adaptor is removed from proper position.


 The cable of the SR is damaged, or not connected, or connection is not good.
 The motor is not working at all. Check motor cable and/or replace the motor.

9.2.2.3 The SR does not turn into the analyzer even with open front panel

 The SR door got stuck in the wash head. The wash head needs to be aligned correctly.
127

 Micro switches are not functioning inside the SR. Faulty switch needs to be replaced.
 PPB board is not controlling the rotor. PPB board replacement needed.

9.2.3 Needle mechanics, Vertical motor (MVert) problems

9.2.3.1 The needle carriage keeps dropping back (down) at initialization

 The wash head position -relative to the needle and to the SR- is not correct. The wash head
comes down too much, and even if lifted, leaves no room for the SR door to turn.
 Make sure that the vertical rod holding the wash head (if removed or modified) is inserted
correctly.

9.2.3.2 MVert cannot reach the optosensor (Up or Down)

 The wash head (when lifted) gets physically blocked by the non-moving parts of the needle
carriage (needle holding block). Set wash head, see washing head adjustments in chapter 6.1.3.
 The grip of the needle on the timing belt is in a wrong position. Set the needle and timing belt
connection. See chapter 6.1.2.2.
 The opto sensor is damaged, or the light path is blocked. Use a screwdriver, or a piece of paper
to verify operation.
 The timing belt is damaged or became too long, or got torn. In this case the XY unit has to be
removed and the belt should be replaced.
 The motor gives a strange noise (not “smooth”) even when moving. Check cable or connection
problem
 The problem comes up only when using Stastedt (or similar) sample tubes from sample rotor.
This could happen because the system is set to pierce a monovette type of tube, not a Stastedt
or the incorrect setting of opto flags on horizontal movement. In this case test that the motor
(carriage) can move freely and run MHori tests. Make sure that the needle height is set correctly.
See chapter 6.1.4.
 The problem comes up only when using Stastedt (or similar) sample tubes (AS position) because
the tube holders are not working properly. Check operation of the tube holders. They must
operate symmetrically, both tube holders should move easily and make sure that the rack can
move in and out freely. Also check that needle height is set correctly.

9.2.4 Shear Valve (SV) related errors

9.2.4.1 SV error at the first startup

 The SV cannot turn. Make sure that the pull tab has been removed, check free movement of the
upper disc and check operation of opto sensors by blocking the light path with e.g. a screwdriver.
 The SV is stuck because the last draining or preparing for shipment process was not performed
or it could not be finished. See that the motor can move the valve and it is not blocked “solid”.
128

Try to loosen the central screw closing the two disks and try to move the motor a few times in
service testing menu. If it does not help:
 use warm water to wet the side of the SV (to make some liquid get in between
the disks)
 you can also dip the SV into water if you can
 gently force it apart

9.2.4.2 Grinding noise after SV cleaning, (after SV reinstallation)

 The upper disc cannot reach opto position. In this case DO NOT adjust the optosensors! See that
lower disc is aligned correctly – loosen the fixing lever, and reseat the lower disc. See section
6.2.1.
 Check that the middle rod on the back of the upper disc is positioned between the two moving
nods. See section 6.2.2.

9.2.4.3 SV leakage

 The upper disc is not sitting well on the lower disc. Disassemble the shear valve, clean the
internal surface. Small particles or dirt can be stuck between the two discs, making a gap
between the ceramics causing a flow.
 The closing screw is not sitting completely on the upper disc. Remove than replace the screw
properly. Make sure there is no gap between the discs.

9.2.5 Dilutor errors

9.2.5.1 Dilutor is making noise

 Dilutor pistons are not attached correctly to the dilutor’s moving part. The piston has a different
size hole on the bottom; the larger hole has to face the instrument.
 One optosensor on the dilutor PCB is faulty. Try to move the dilutor up and down in Service
Testing menu. If noise is still present, replace the dilutor unit.

9.2.5.2 Dilutor is leaking

 The syringe or the top plastic of the syringe is cracked. In this case replace syringe.
 The “ribs” on the pistons are filled with liquid and they are white. The piston is leaking; the
whole syringe with piston should be replaced or greased with silicon grease. Do not grease the
black pistons, just replace them if needed.
129

9.2.6 A tube comes off of a valve


Depending on the location (ID) of the valve, it can be caused by:

 One of the valves are not opening well. Check all of the valves in Service Testing menu. Replace
the valve which is not responding.
 Shear valve discs are not aligned well. Adjust the shear valve, see section 6.2.
 Tube on the valve is too wide. Cut off the end of the tube and reconnect it to position.
 TCU tubing is blocked. Drain the TCU by using the option is Service Functions menu. Fill TCU and
see that the clogging in the TCU is removed. If the TCU tubing is still blocked, remove the TCU
and check the tubing inside.
 Laser flow cell is clogged. Perform a flow cell cleaning by software or manually.
 One of the chambers is clogged. Remove the metal plate from the side of the chamber and clean
it. Don't forget to drain the chambers in Maintenance before removing the front of the chamber.

If a tube comes off from a valve, it can be useful to check the tubing schematics. See the tubes
connected to the valve, follow the route of the tubing, note that the valves involved were open or close
state when the tube popped off. Also see shear valve position when tube pops off.

By using the schematics and following the route of the tubes, it is possible to find the clogging in the
system.

9.2.7 Priming problems

9.2.7.1 The analyzer would not prime liquids

 The respective reagent is out. Check the reagent status bar on the bottom. Change reagent if
needed.
 The reagent tubing is not connected to the proper inlet. Check both ends of the tube.
 The reagent tube is bended or broken. Check the tube.
 Tube popped off from a valve. It will make the vacuum level dropping immediately, so priming
won’t be successful. Reconnect the tube.
 There is not enough vacuum to fill up the reservoirs. Check all the valves in Service Testing menu.
Replace faulty or non-opening or closing valves.
 The pressure sensor is faulty. Check sensor offsets, see section 6.6.
 Reagent buffer reservoir is cracked and vacuum is escaping. Replace the reservoir.
 The float is not moving when the reagent is filled. In this case the float sensor is stuck to the
bottom. Gently touch or tap the buffer to make the float free.
 The reservoir is full, priming doesn’t stop, but the float is not sensing, even when it’s on the top.
In this case the reagent sensor board or the float sensor is damaged. Reset the Low Level
Software to stop priming and locate the reagent sensor board. See that the LED is sensing the
float’s signal, or connect the buffer’s cable to another pair of pins. See that the LED is on. If LED
is on, the float is ok and the board can be faulty and it has to be replaced. If the LED is not turned
on, the float is faulty and the reservoir needs to be replaced.
130

9.3 Electronics related problems


9.3.1 Touch screen / display errors

9.3.1.1 No image on display

 Display is white. LVDS data cable is removed from the touch screen or from the motherboard.
Reconnect the cable.
 Screen is black. Check the cables behind LCD and on the mother board. See that BIOS settings
are not modified to use different display device. See chapter 7.5.1., BIOS settings, Chipset/North
Bridge/Onchip/Select Display Device.

9.3.1.2 Touch sensitive surface not working

 When touch is not responding, it can happen that the sw is frozen or the display is faulty. The
touchscreen PCB is located on the back of the LCD. Remove the cover of the PCB and check the
board and the cable connection.
 Connect external USB mouse to the unit and check that the mouse cursor is available and
moving. If mouse is ok, remove and reconnect the Touch USB cable.

9.3.1.3 Touch (click) is inaccurate

 Connect external USB mouse and select touchscreen calibration in Maintenance menu.

9.3.1.4 Screen image is not normal, some parts of the software screen is not visible

 The analyzer uses a resolution 600x800. If there is a problem with the size of the screen check
the resolution in Windows. Leave the High Level Software by pressing CTRL+ALT+DEL to enter
Windows (only if Windows control checkbox is enabled in Service Functions).
 It is also possible that the resolution of the screen was changed in BIOS. Enter BIOS and check
the resolution in Chipset/North Bridge/Onchip/Select Display Device.

9.3.2 Communication errors


Check the communication by going into Service menu/Sw Upgrade and press Refresh button. If
communication is ok, all the available firmware and software will be listed. If there is a problem with
communication, the unit will send message “Pneumatic error” or “USB communication lost”.

 In most cases, the communication problem is created by the DimmBoard. All the electronics are
working well but the data cannot access the low level software. In this case check the
DimmBoard, locate the SD card inside the board and make sure that it’s not removed.
 Remove the DimmBoard from the LSDAQCQ, clean the connection surfaces and put it back.
Don’t need to turn off the unit, just restart the Low Level Software in Service Functions. The Low
Level Software will automatically turn on and start communication, when the connection is
available again.
131

 Check the start button. If it is red all time, it means the low level software is frozen or stuck in an
infinite cycle. Go to Service functions screen and restart/reset the low level sw. See section 7.3.1.
 If start button is blinking, the communication error is related to the Pneumatic system. Open the
valve block door and locate the two PPB panels. Check the cable connections; replace the small
PPB CON panel on PPB1 (left). If communication is still not ok, replace the PPB CON panel on PPB
2 (right). If the problem still not solved, replace PPB1 and then PPB2.
 The error message “Unit is not present” will notice that the communication line inside the unit is
broken. The line is a closed BUS between the LSDACQ (DimmBoard), the TCU, the pressure
sensor, the optical head and the front panel (touch). If one of these are damaged or not
connected by USB cable, the system gives a message about which unit is present or not. Locate
the faulty unit and see that the power supply is attached to it and check all their cables. See
section 9.3.4.

9.3.3 The analyzer does not power on


 If the analyzer is connected to the external power source and the main power switch is on, an
indicator LED on the motherboard should turn on to show that the power supply is working. If
the LED is not on, the power supply has to be changed. Remove the hardware module from the
analyzer and change the power supply. If the LED is on but the unit is not starting, the problem
can be the motherboard or the power-on switch. Follow the power-on switch’s cable to the
motherboard. Remove the cable from its position. Try to power on the motherboard by applying
a short circuit on that specific jumper of power-on switch with a screwdriver. If it starts, then the
switch is faulty.
 There is a “beep” sound coming from the instrument. Typically the motherboard gives this sound
if the board itself has a problem starting up or the memory card is not in place. Locate the
memory card and check its position. Replace motherboard and/or memory card if necessary.
 The power supply works, the motherboard starts, but the system would not boot up. If high level
sw and windows cannot start, the operation system has to be reinstalled or recovered. See
section 7.5.1.
 The power supply works, the motherboard starts, windows starts, but the User interface SW
does not start (just widows desktop appear). In this case the instrument doesn’t have a properly
installed High Level Software. Download the High Level Software from Diatron’s website (msi
file). Copy the software to a pen drive and insert it to the instrument. See software installation in
section 7.5. Make sure that the Low and High Level Softwares are the same version.

9.3.4 I2C errors displayed at startup


 Check the power and data connection of the board reported missing
 Most probably the power connector of the missing module is disconnected
 The second most likely cause is a misconnected, or not connected data cable
 A missing TCU power cable can cause a “multiple” I2C error when more than one modules are
reported missing. Check the power connection of the TCU module
132

9.4 Useful information and tips


9.4.1 High PLT background
After shipping it's possible that the PLT background is higher than normal for the first few background
runs.
 If the PLT level is not normal after a few runs, use a syringe to put Hypoclean CC inside the RBC
chamber and leave it there for a few minutes. Than make stress measurements.
 If the reagent tube was contaminated, and some bacteria begin to grow inside, once you put an
infected reagent tube into a new tank, by time it can become infected as well, i.e. the background
(PLT blank) becomes high. Wash the reagent tube - which is in connection with the reagent - with
1% of bleach solution, and then rinse with clean distilled water or diluent. It can avoid the bacteria
to grow inside.
 If tank is open – and cap is not installed or closed - external dust can make reagent dirty.
 Use stress measurements to find out that the problem is dirt inside the aperture or it is noise. If
the stress results are high but slowly decreasing, it most likely dirt inside the RBC chamber. If
results have a large diversity, it's caused by noise.

9.4.2 Fast Starting up and Shutting down


During troubleshooting, it is possible that the unit needs to be turned off and on multiple times. Before
shutdown, the unit has to make a cleaning process and after an improper shutdown, the instrument
requires a cleaning process when turned on again. The cleaning procedure is not always necessary during
troubleshooting and it can take a long time.

The cleaning procedure can be skipped by running program code 66 in Service Testing menu. Run this
program before shutdown to make it faster and to skip the cleaning process run the program before
running background measurement.

9.4.3 Removing bubbles from the Flow cell


The flow in the flow cell is very critical during the measurement. If bubbles appear in the tubing system
(the 4 meter tube and the flow cell), it can make improper scatter values. The easiest way to check the
amount of bubbles is to press valve #1 and #20 together. This will start the flow inside the 4 meter tube;
moving bubbles can be noticed easier. To start the flow in the flow cell, press valve #1 and #32 together
and check the laser waste tube for bubbles. If there are too many bubbles in the laser, check the tubing
and "T" connections in the tubing at valve #20 and #32. If bubbles are leaving the laser but not entering
into it, the injector must be cracked. Remove the laser and replace the injector.

9.4.4 Removing the DimmBoard


There is no need to turn off the instrument to remove the DimmBoard. Go to Service Functions menu
and press ‘Low Level Reboot’ button. After a few seconds, the start button will turn off which means the
low level sw is turned off to reboot. It is possible to remove the DimmBoard or the SD card while the Low
Level Software is off. After a few seconds, the software reboots and the communication will return. If
133

there is not DimmBoard in the LSDACQ, the communication will be lost until the board is replaced. After
replace, it is possible that the system needs a Low Level Software reboot to wake up.

9.4.5 Listen the sounds of the instrument


The unit keeps repeating the same process during measurements. The sound of the unit is very telling
and can warn the user or service engineer that something is not working properly. It is also useful to put
your ears on the front panel and listen the mechanics. Some noise can be discovered easier this way.

9.4.6 Replacing small diameter tubes


Some tubes in the unit are connected to a small plastic 'T' fitting. Replacing these tubes can be difficult
because the tube's diameter is smaller than the fitting. In this case a thin metal rod can help putting back
the tube to position. Put the rod's end into the 'T' fitting while the tube is on the rod. This way the fitting
will not broke while the tube is pressed onto it, because the metal rod is keeping it on position from
inside.

9.4.7 Swapping the pumps


For troubleshooting, it can be useful to swap the two
pumps. The instrument can work when the two
pumps are interchanged. It's not necessary to remove
the pumps, simply just remove pump 1 cable from
PPB1 and pump 2 cable from PPB2. Connect pump 1
(outer) cable to PPB2 and pump2 (inner) cable to
PPB1. After this, also change the tube connections
next to the reservoirs. By this it's possible to find out
which pumps has problems. Self-test results can help.

9.4.8 Draining the reservoirs


If there is a problem with a reservoir and it needs to be changed it has to be drained first. Draining is
made by the unit automatically by pressing drain in Maintenance menu. To save time, it is possible to
remove the tube from the top of the reservoir. The ambient air pressure will push the reagent out from
the reservoir faster than the draining procedure. Use this only when reservoir needs to be replaced.
134

10 Maintenance
10.1 User maintenance
10.1.1 Daily maintenance
Abacus 5 requires automated maintenance during daily routine. It is a user maintenance.

At the end of day a system cleaning procedure is recommended. To perform this cleaning, prepare a vial
containing 2-4 ml Diatro•Hypocleaner CC® and place it in the Sample Rotor. From Main Menu go to:
Maintenance/Functions and press Clean. This will initiate a cleaning cycle, all chambers, needle, shear
valve and the TCU is cleaned with diluted Diatro-Hypoclean®, than the system is rinsed with diluent.

10.1.2 Weekly maintenance


Weekly maintenance is for cleaning salt build-up from the most contaminated surfaces. It is
recommended to clean the Shear Valve and Needle Wash head after every 1500 samples.

10.1.2.1 Shear Valve cleaning

This procedure is detailed in User Manual section 17.6

10.1.2.2 Cleaning the wash head

This procedure is detailed in User Manual section 17.7

10.2 Preventive Maintenance


The procedures in this section are service maintenance.

Preventive maintenance frequency depends on the daily workload of the customer.

In the Abacus 5 User Manual and Service Manual the maintenance interval of moving parts and syringes
is determined as to be done after every 30.000 measurement cycles or twice per year.

As a general rule, at least 1 maintenance per year must be performed to ensure correct functioning
and performance according to specification.

Diatron recommends the following frequency for preventive maintenance by field service:

- for instruments with a workload of up to 100 samples/day


Preventive maintenance by field service should be performed once per year

- for instruments with a workload between 100 and 200 samples/day


Preventive maintenance by field service should be performed 2 times per year

- for instruments with a workload above 200 samples/day


135

Preventive maintenance by field service should be performed at least 3 times per year

To do list and materials needed for a preventive maintenance intervention

Note: there is no difference between the „6 months” or „1 year” preventive maintenance. The same
maintenance protocol must be followed and applied every time.

Procedure Duration: ~ 2 hours

Tools needed:

- hex screwdrivers (1.5, 2, 2.5, 3, 4)


- torx screwdrivers (2.5, 3)
- philips screwdriver (2)
- small size flat screwdriver
- forceps
- tweezers
- syringe (at least 20ml)

Materials needed:

- silicone grease (P/N: A599)


- teflon grease (P/N: A598)
- bleach (~200ml, Diatro Hypoclean CC and/or Diatro Hypocleaner )
- distilled water, ~500ml
- paper towels
- compressed air
- required protective safety equipment (gloves, lab coat, glasses, etc.)

Spare parts needed:

Maintenance KIT for A5


136

Contains:

Part number Description Quantity

A5330 Washing head Abacus 5 1

A543 4/1.8 Tygon tube, 1 m 4

A5433 1,27/2,23 Tygon tube (type: S 54-HL) 1

A5439-S 0,78/2,38 Tygon sampling tube for A5, 26 cm 1

A5-AJN014 Piercing needle (Abacus 5) 1

AP251 SCHOTT syringe with piston white (mounted, 2 ml) 6

J5P251 Schott syringe with piston black (mounted, 2ml) 2

Suggestion of spare parts which are good to have but not necessary at a PM visit:

- RBC/WBC Chamber J5P421


- RBC/WBC Aperture J5P431
- 2/2 Valve with Coil A504
- 3/2 Valve with Coil A505
- Pressure sensor board A5519

Preventive Maintenance Protocol:

1. First steps / Checking the state of the analyzer

1.1 turn on the analyzer, run background and check if values are within accepted range
1.2 run Self-test (electronic and pneumatic), if it is „successful” check if any value is close to the
accepted limit, if it is „failed” troubleshoot accordingly
1.3 in case of high „laser on” or „laser off” DC level run a background cycle and Self-test again. If DC
level values are still high do a flow cell cleaning, see section 10.3
1.4 from Service/Adjustments run „pressure offset adjustment” and check if values are within range;
change the pressure sensor board (P/N: A5519) if „offset values cannot be adjusted”, message
appears or if adjusted value is very close to the limit (refer to SM for limits)

Next step is to bring the analyzer into the proper state where maintenance procedures can be performed
without the risk of liquids spilling on parts. DO NOT skip the 3 procedures listed below:
137

1.5 from Service menu run „drain TCU” option; follow instructions in the SM regarding this
procedure
1.6 from Service menu run „drain Flow cell” option;
1.7 from Maintenance menu empty all 3 chambers(RBC/PLT, WBC, MIX)
At this point maintenance of parts can be started. Open the front cover and remove side covers of the
analyzer.

Inspect:

2. Valves
2.1 check for any dirt or deposit in the valve; if necessary clean inside of valve by pushing DI water or
bleach through it by means of a syringe
2.2 if any sign show that liquid spilled on the valve, remove the valve head to see if liquid got also in
the coil, if so, replace the coil
2.3 activate each valve from service menu; if switching of valve isn’t firm(smooth movement of
membrane) try to clean it or replace the valve head
2.4 check the exhausting tube/vent nozzle on valve #5 to be free of dust or any deposit; clean it with
DI water if necessary
2.5 special attention on valve #2: this valve and related tubing must be clean. In case of dirty
valve/tubing, flow of laser waste through valve is restricted and it has direct influence on optical
counts (!) Replace the valve and/or tubing if necessary
2.6 special attention on valves #34 and #35 as well; these valves and the related tubing must also be
clean and free of any deposit. Replace the valve and/or tubing if necessary

Proceed to step 3.

3. Tubes
3.1 all tubes must be clean inside; any deposit must be cleaned with bleach and rinsed with DI water
or replace the tubing
3.2 check tube connections on valves, chambers, SV, wash head, internal reservoirs etc. all
connection points must be tight, otherwise change that tube portion with new tube
3.3 as a rule if a tube cannot be cleaned sufficiently and/or its’ connection isn’t tight it is
recommended to be changed with new tube

Proceed to step 4.

4. Chambers
4.1 all chambers must be clean, free of scratches or cracks
4.2 removal of chambers is recommended to check if no liquid has spilled behind them and also to
see if the electrode is in good condition; change the chamber if any visible crack can be seen or if
the electrode can be moved by hand
4.3 reinstall the chambers

Proceed to step 5.
138

5. Dilutors
5.1 from service menu move dilutors one-by-one and check if large air bubbles are visible in the
syringes or if any leakage sign can be seen on or beneath them; mark the syringe(s) where you
saw large air bubbles getting in near the plunger
5.2 perform dilutor pistons cleaning and greasing procedure
- Follow instructions of section 10.2;
Black pistons should NOT be greased (!)

Use a forceps to close the tube above the syringe you are going to remove. This is to avoid reagent from
internal reservoirs spilling back to the reagent bottle. Remove one syringe at a time only!

5.3 cycle dilutors from service menu again, if any of the syringes does not seal properly
(air bubbles are still getting inside near the plunger) replace the syringe(s)

Diatron recommendation as preventive maintenance: replace syringes after 50.000 measurement cycles

5.4 clean dilutor optosensors from dust by means of compressed air


5.5 clean the driving pin of dilutors then lubricate them with teflon grease

Proceed to step 6.

6. Shear Valve
6.1 inspect tubes on SV for any deposit or discoloration, especially the „4diff waste” and „Laser
in”(refer to fluidic schematic of the analyzer to identify these tubes) tube ; lower section of laser
waste tube might have black color – it can be cleaned with bleach or that tube section can be
changed.

Note: when removing tubes from the SV ports use a small flat screwdriver to help the tube slide off from
the port; to avoid misconnection remove only 1 tube at a time

6.2 check tightness of tubes on SV ports by hand, all tubes connections must be firm; if any tube
comes off easily if pulled by hand that tube section should be changed
6.3 clean the SV optosensors from dust by means of compressed air
6.4 from service menu cycle the SV several times; if the movement isn’t smooth or noise is coming
from driving mechanism, remove the cover behind the SV and lubricate the cogged wheel/bar
with machine grease; apply a thin film of teflon grease to the two ends of the sliding clutch-bed
6.5 reinstall the driving mechanism cover
6.6 change the blood sensor tube(tube between the needle and SV)

Proceed to step 7.

7. Laser head
The laser head doesn’t need to be opened for regular maintenance. Opening the laser head will end the
warranty period for the optical head!
139

If „laser on” DC level was high during repeated Self-test cycles, it might be necessary to clean the Flow
Cell manually. This should be done only if „laser on” DC level is still high after the software given “flow
cell cleaning” procedure option was done.

If „laser off” DC level is high that might indicate failure of a component of the laser head. Contact Diatron
Support Team ( support@diatron.com ) for further information and assistance.

If any deposit can be seen in the „laser in” tube or if „laser on” DC level was high during Self-test
than do the following:

7.1 remove the thick, RED coded tygon tube(laser waste) from the tube organizer and connect a
syringe to it
7.2 prepare at least 10ml of 1/1 bleach solution(1 part bleach, 1 part DI water) in a cup
7.3 remove the „Laser in” tube from the SV and immerse it into the bleach
7.4 gently pull the syringe until at least half of the bleach is aspirated or until the „Laser in” tube is
clean
7.5 remove the „Laser in” tube from the bleach and pull ~ 20ml of air through it
7.6 reconnect the „Laser in” and „laser waste” tubes to their ports

Proceed to step 8.

8. TCU
The TCU unit is a key assembly and proper functioning of this part is vital for good performance
of the analyzer. Therefore with the occasion of periodic maintenance the TCU unit must also be
inspected by service engineer. Inspection is only possible if the TCU unit is removed from the
instrument. Follow instructions of the SM for removal of the TCU.

8.1 clean the fan and heat sink of the TCU from dust with compressed air
8.2 remove the in-line mixer cover and check the 2 tygon tubes for any deposit
8.3 if necessary clean or replace* the 2 tubes with 1,8x4,0 mm tygon tubes cut to the same length

* Tubes of the in-line mixer are the bottleneck for flow inside the TCU. Any deposit in these tubes will
cause overpressure in the 4diff sample preparation circuit which in worst case will result in tubes being
pushed off from valves and would cause liquid spills inside the analyzer. Natural ageing of tygon results in
loss of elasticity and small decrease of diameter. Ageing is speeded up by the aggressive reagents
(Lyse5P, Diff5P).

Diatron recommendation as preventive maintenance: change the in-line mixer tubes once per year.

8.4 reassemble the TCU unit to its place, follow instructions of the SM

Proceed to step 9.
140

9. Wash head and XY block


9.1 clean the sliding bars of the XY unit with a paper towel and apply a thin film of teflon grease on
them
9.2 clean the optosensor on the XY opto board with compressed air
9.3 verify the metal ports of the wash head are not loose, replace the wash head if any of the ports
can be moved by hand
9.4 verify the vertical movement of wash head neck inside of the XY block by lifting it up from the
needle holder

Attention: Sharp needle! Risk of injury! Biohazard!

9.5 vertical movement of the wash head must be easy without any obstruction, clean the wash head
neck if movement isn’t smooth
9.6 clean the bottom of wash head

Proceed to step 10.

10. Verification/adjustments
From service menu run:
10.1 „pneumatic initialization”
10.2 „fill flow cell”
10.3 „fill TCU”

Run shutdown then switch on the analyzer.

10.4 run background counts until background values are within accepted limit; several background
cycles might be necessary to achieve good values – this is normal behavior after tubes and
subassemblies have been drained and refilled
10.5 run Self-test (both) and check that all parameters are within range
10.6 if DC level is still high contact Diatron Support (support@diatron.com) for further assistance
10.7 from service menu perform wash head and needle position verification; adjust if necessary
10.8 adjust blood sensor default value(with blood sensor disabled)
10.9 calibrate the blood sensor
10.10 verify calibration with fresh control blood; calibrate the instrument if necessary
10.11 run human samples and check results
10.12 run repeatability test(at least 11 measurements from one fresh human blood with normal
values) and check CV%-s are within specification for all parameters

Preventive maintenance is done!


141

10.3 Flow cell cleaning


The optical measurement system of the Abacus 5 is strictly vacuum-based. Please do not apply
pressure to the flow cell!

Tools needed:

 small flat screwdriver


 Syringe(10-15 ml or bigger)
 small cup with Hypoclean CC or filtered bleach(approx.. 10-20 ml)

Procedure:

1. Open and secure front panel.


2. Run “empty flow cell” function.
3. Remove the laser waste tube from the connector located on left side of inner front panel. This
tube has red strip on it. Use a small flat screwdriver to help the tube come off from the metal port.
4. Connect the (empty) syringe to the tube.

5. Disconnect the sheath tube from the connector located on left side of inner front panel. This tube
has green strip on it and it is smaller in diameter. Use the screwdriver to help removal.
142

6. Pour Hypoclean CC or filtered bleach into a cup and place the end of the sheath tube in it.

7. Gently pull the syringe to make the bleach flow through the Flow Cell.
You can increase cleaning efficiency if you let air bubbles in the sheath tube while pulling.
8. Remove the sheath tube from the bleach and pull some air through the FC.
9. Reconnect sheath tube to its connector.
10. Remove the laser sample tube from upper disc of SV. This is the thinnest tube. Use the small flat
screwdriver to help removal from metal port.
143

11. Place the end of sample injector tube in bleach. Gently pull the syringe to clean the sample injector
tube from the inside.
12. Remove the sample injector tube from bleach, pull some air with the syringe, wipe the end of the
tube and reconnect it to its port on the SV.
13. Remove the syringe from the laser waste tube and connect the tube back to its connector.
14. Run “fill flow cell” function.
15. Check that background and sample results are acceptable.
144

11 Removal and replacement procedures


11.1 Opening the instrument
Both side covers of the instrument can be easily removed. This allows reaching of the fluidic system and
the mechanical parts. Other parts of the analyzer can be reached by lifting up the front panel.

Don’t forget to secure the front cover in the upper position after lifting,
as it may fall down if unsecured!

To take off the left cover:


Loosen the four milled-edge screws in the front and on the back of the analyzer.
Pull the cover to the left to take it off.

To take off the right cover:


The procedure is the same as in the case of the left
cover.

Milled-edge screws
145

11.2 Shear Valve Assembly


Shear Valve Assembly is located in middle of
front side of Abacus 5 behind the front cover.

Main parts of Shear Valve Unit:


 Ceramic valve
 Driver unit with a stepper motor
 SV Opto sensor board

Ceramic Valve
removal
 Drain SV
and the tubes on
it using Shear
Valve cleaning
function in the
Maintenance
screen
 Prepare a piece of absorbent paper to dry components if necessary
 Remove the thumb screw on the valve
 Remove the upper ceramic disc
 Remove the lower ceramic disc of the valve pulling it upward

Ceramic Valve install


 Put back lower ceramic disk
 Make sure the fixing plates hold the disc in tight enough, prevent
the lower disk rotating. Please realign the plates if it needs.
 Install the upper ceramic disk, making sure the longer horizontal
pin is between the two rods of the Shear Valve drive. So it can
move the upper ceramic disk. The other rod must be able to move
into opto sensors freely. Fix the Shear Valve with the thumb screw. Please do a Shear Valve Opto
SNSR alignment if the right side pin hits one of the opto sensors.
 Check the operation.
146

Shear Valve Opto Sensor board replacement


 Shear Valve opto sensors are mounted to the right side
on Shear Valve Assembly with 3 screws in such a way the
sensors can be set separately.
 remove screw behind the circular opening (through the
hole)
 Watch out the plastic washers.
 Remove the screws behind the oval opening (through
the opening)
 pull out the opto board and remove the 2 cables
 Install a new board.
 connect cables
 install big board (2 screws) use plastic washers
 install small board (1 screw) use plastic washers
 Make sure that the lower disc’s flat side is aligned, and
secure. You should not be able to move (twist) the lower
disc at all.
 Please do a SV alignment procedure 6.2.

Shear Valve Assembly removal


 Disconnect cable J7510 from SV opto board
 Disconnect SV motor cable from PPB
 Remove four screws on the front of SV assembly
 Move the SV assembly out of the instrument

11.3 XY Unit
XY unit contains the slides for Horizontal and Vertical movements, two stepper motors, XYR opto board,
opto wheel, sample rotor, wash head holder, wash head and the sampling needle. With the exception of
the sampling needle, wash head, wash head holder and MHori (stepper motor for horizontal movement),
for proper servicing the XY unit needs to be removed. The procedure is the following:

 Turn off the analyzer.


 Lift and secure the front panel, then remove the right cover.
 If the analyzer is equipped with an Autoloader, disconnect and remove the Autoloader module.
 Loosen the two sampling needle fixing screws and remove the needle.
 Sampling needle is sharp and represents biohazard! Take extra care and wear gloves when
servicing this part of the analyzer.
 Remove the baluster. Its lower screw can be access from the bottom of analyzer.
147

 Locate and unscrew the four screws from


the bottom of the analyzer, which hold
the Sample Rotor basis.

 Unscrew the two (upper and lower)


screws which hold the rear part of the XY
unit. One screw is located above the
MVert (the bigger stepper motor for
vertical movement and cap piercing), the
other one is located below the
optowheel and sliding bar, near the
same motor.
148

 Gently remove the wash head holder with


the wash head, by twisting it clockwise
and pulling it down until released.
 Disconnect electrical cables
 Remove the XY unit

Sampling needle
fixing screws

Upper screw

MVert
Washing head and
washing head holder
MHori

Sample Rotor basis (the Lower screw


screws are located on
the bottom of the
analyzer).

For mounting back the XY unit, follow the steps listed above in reverse order. Make sure to perform the
Sampling Needle adjustment procedure after reinstallation.
149

11.4 Sample Rotor


Sample rotor holds the sample tube and turns it inside the instrument for cap piercing. In order to
provide its proper operation with XY unit, they are equipped
together.

Sample Rotor removal:

 Remove XY unit with Sample Rotor as it is written in


previous item
 Remove two screws from the front of XY unit that
hold the Sample Rotor rods
 Separate the Sample Rotor
 Pull off the door. Please check rotor arm and the hole
of lower door plate rod position when put back the
door.
 Two end position switches can be removed and
motor of unit are accessible.
150

11.5 Dilutors
Abacus 5 has two separate main dilutor modules, located behind the front panel on the lower part of the
front plate. Each modules is secured by four hex screws and has a flat cable connection to PPB1
respectively PPB2. The removal/replacement
procedure is as follows:
1. Turn off the analyzer.
2. Lift and secure the front panel.
3. Remove the tubing from the syringes.
4. Unscrew the four hex screws and put them
in a safe place where you can find them
later.
5. Gently pull the unit toward you until you can
see the flat cable connector on the rear side
of the dilutor optoboard. Disconnect the cable.
6. Now the dilutor module is released and available
for maintenance or replacement.
To remount the dilutor unit, follow the steps described
above in reverse order.
151

11.5.1 Syringe replacement


 Remove two thumb screws from the syringe holder
bar and remove the holder bar from the top of the
syringes.
 Remove lower thumb screw from syringe piston
rod.
 Remove the syringe
Installing back a new syringe take care the direction of
piston rod. The wider hole must be turned inside the
system.

Correct Incorrect

11.6 TCU Module


The Temperature Control Unit is located below the Laser Head Assembly. It is connected to the Shear
Valve and to the tube organizer via Tygon tubing. The internal electronics are connected to the power
supply and to the LSDACQ board. In order to remove this unit, follow the steps below:

- turn off the analyzer


- If autoloader is installed, disconnect and remove the Autoloader module.
- drain the TCU module manually
 disconnect the BLACK tube and connect a 10ml syringe
152

 disconnect the GREEN tube and aspirate liquid from the TCU module using the
syringe
 disconnect the WHITE tube and
connect a 10ml syringe
 disconnect the RED tube and
aspirate liquid from the TCU
module using the syringe

- remove the screw above the shear valve


(securing the front of the TCU)

- remove the 4 screws in the back of the


analyzer securing the rear

- open the assembly plate on the left side of the


analyzer to have access to the TCU control board
cables
- locate and disconnect the power cable of the TCU
module
- disconnect the controlling cable coming from the
LSDACQ board
- pull the TCU out towards the back of the analyzer

To reinstall: follow the procedure in reverse order


153

11.7 Valve block

3/2 valves
Valves can be removed
together in one assembly
(valve, coil, board) or
individually one by one.

To remove the complete


assembly just unscrew the
6 mounting screw and
3/2 valves
disconnect the driver
cables on the back.

3/2 valves
A valve has two main
parts: the valve unit and
electromagnet coil.

The tubes are connected to the valve pins.


The coil is soldered to valve block board.
A plate fixes the main parts together on the coil site.
154

Pull with pliers

To remove individually you’ll need to use a plier to pull the


valve fixing plate out at the back and after that it is possible
to remove only the valve itself.

11.8 Pre-Amplifier Board


Pre-Amplifier Board is located inner side of the fluidics
panel, behind the Measuring Black. It is under a metal
cover that fixed to the fluidics panel with two screws. See
the picture.

11.9 Pump Assembly


1. Open the front cover.
2. Remove the left side cover by loosening the four fixing screws.
3. Loosen the two clamp-screws on the rear panel and open the side pneumatic board.
155

4. Pull off the connected tubes from inner fittings of Waste output and pump inlet tubes from
lower right side of flow Panel.

5. Pull off the connected cables (red-green-black) from the


PPB boards. The pump 1 (outer one) power cable is
connected to PBB1 (left board), the pump 2 (inner one)
power cable is connected to PBB2 (right board).
6. Unscrew the screws at the top of the pump fixing bar
and pull it up with the white plastic bar.
7. Lift off the pumps.
156

11.10 Hardware Module


The whole removing procedure is easier by opening the front cover and with removed side covers.

1. Unscrew the five screws connected on the rear plate.


2. The assembly is positioned and pushed down inside by a spring sheet. Gently pull out the
fixing plate until almost the whole module comes out. Be careful not to strain the power and
data cables connected to other parts inside.
3. If you want to completely remove the module you have to disconnect all the power and data
cables connected inside.
157

11.11 Laser Head Assembly


To remove the Laser Head:

Turn on the analyzer.

1. Press and hold the Menu icon until the on-screen keyboard pops up, then type in the service
password. Service button appears on the main screen, click on it to enter the service menu,
then click on Service Functions/Drain flow cell and wait until the process finish.
2. Turn off the analyzer (from rear start switch)
3. Open (lift up) and secure front cover.
4. Remove right side cover.
5. Unscrew the 4 holding screws of the top cover then remove it by lifting and sliding it
backwards.
6. Locate the three tube connections of the Laser Head.
#1 – laser sample tube; #2 – sheath tube; #3 – laser waste tube.

7. Gently remove the tubes from the metal ports. Use a small flat screwdriver to help sliding tubes
off.
158

8. CLOSE FRONT DOOR


9. Disconnect both cables from the optical head. Refer to picture 4 below.
(#1 – autoalignment cable*; #2 – optosensor board cable; #3 – laserdriver board cable)

Note: it is easier to disconnect the laserdriver board cable from the LSDACQ board. Units
produced after 2013.06.doesn’t have auto alignment cable.
159

1
2

10. Locate the four holding screws of the base-plate of the optical head. Refer to Picture 5.below.
Use a 2,5 mm hex screwdriver to remove the
screws. Take extra care of the washers.
11. Gently remove the optical head unit. To avoid
damage of the sample injector needle and
sheath inlets on the base of the flow cell
please always place the laser head assembly
on its’ side or upside-down. Do not ben tubes
too much.
160

11.11.1 Laser injector replacement procedure

Laser issues when injector needs to be changed:

No scatter available (both 4diff and Baso). No


scatter gram appeared after measurement,
warning flags x and X are displayed at warning
flags. In some cases the absence of the scatter
does not mean that the injector is faulty.
Sometimes the flow cell gets clogged making the
flow improper or unavailable. In this cases,
cleaning the flow cell could solve the problem,
using the default cleaning process in maintenance
menu (flow cell cleaning), or using a syringe. If
software cleaning or manual cleaning doesn’t
help, injector replacement is needed. In this case,
remove the laser and check the injector’s needle.
Sometimes the damage is not visible, so replacing
the injector is necessary.

It’s also an injector related issue, when the scatter


appears in both 4diff and Baso, but the graphs are
improper and values are abnormal. This can be,
because the injector needle is removed from
proper position and the sample flow inside the flow
cell is not in the middle. For a proper
measurement, it’s important that the sample flow
should be in the middle of the flow cell,
surrounded by the sheet flow which is injected next
to the sample needle.

In this case the reagent inlets has to be checked fist


to make sure that lyse and stopper reagents are
not replaced. If reagents are well attached, the
injector needle has to be checked and changed.
161

11.11.1.1 Removing the laser for injector replacement:

Step 1: Step 2:

Remove the holding screws of the optical head. Remove the three cables connected to the laser.

Step 3:

Disconnect the tubing from shear valve and the black plastic console
162

11.11.1.2 Removing the injector:

Step 1:

The injector is under the black protecting cap on the bottom of the laser. To access the injector, gently
remove the black cover by hand.

Step 2:

Remove the three holding screws completely. Use a flat-blade screwdriver (used when removing the
laser cables) to lift the injector from the injector’s housing. The injector is very well fitted on its place so
removing it requires some force.
163

Step 3:

Place the new injector inside the housing; make sure that the holes on the injector are looking at the
same direction as on the housing. Check that the injector has the black O ring(s) attached!

Step 4:

If injector is in position, replace the screws. Don’t tight the screws too hard because it can crack the
injector or the plastic pert of the flow cell. If the screws are well in place, put the cap back to position
(only the thin sample tube and the green sheet tube included, waste is separated).

Step 5:

Place the laser optics back into the instrument; connect the cables and the tubes. Run control and fresh
blood sample measurement to check that the laser works fine with the new injector. There were not
laser adjustments on the flow cell during this procedure, so there’s no need to realign the laser.

Step 6:

If scatter is visible and the values are acceptable, put the laser holding screws back to their place.

Please do a scatter calibration, if needed.


164

11.11.1.3 Reinstalling the Laser Head:

(Analyzer is turned off, right side and top covers removed, front door is CLOSED)

1. Connect laser driver board cable to laser head.


2. Place the Laser Head to its’ place. Guide tubes to exit towards the front.
3. Tighten the screws, make sure that ground cable is also in place and washers are installed.
3. Reconnect autoalignment* and optosensor cables.
4. LIFT UP AND SECURE FRONT COVER.
5. Reconnect tubes to the corresponding ports.
6. Turn on the analyzer; activate the service menu, then click on Service/Auto alignment.
Check the values of “laser power” (4 diff and baso) on “Laser settings” field. Compare these
values with the Auto alignment datasheet received with the new laser head. Enter new values
from the datasheet (only necessary if those are different from the analyzer’s values) and click on
Save values. Exit from service menu.
7. Click on Measure for background count, if necessary repeat background cycles until all values are
in acceptable range.
8. Prepare a freshly opened D-Check3P NORMAL level control. Run “Scatter Calibration
Procedure”.

11.12 Measuring Block


 Remove the left side cover by loosening the four fixing screws.
 Measurement Block is located middle upper side of flow panel, above the valve blocks. It
has WBC measuring chamber, RBC/PLT measuring chamber, Mix chamber and a drain
buffer.
165

11.12.1 Aperture removal:


WBC, 80µm and RBC/PLT, 70 µm apertures are behind metal front plate of measuring units mounted by
four screws. Two long ones (upper left and lower right) fixes the metal plate to the measuring chamber,
two short ones (upper right and lower left) fix the measuring chamber the back plate of the unit. See the
picture.

11.12.2 WBC aperture removal:


 Drain the chambers in maintenance
menu.
 Unscrew hex nut on the right site of
front panel that fixes the black metal
shield.
 Remove the black shield.
 Remove four screws from the unit.
 Remove black grounding wire and the
metal plate with two tube fittings.
 Aperture can be removed. Be careful
with the plastic parts, do not scratch
them.
166

11.12.3 Measuring Chamber removal:


 After two short screws (upper right and lower left)
and black grounding wire removal, the measuring
unit can be taken out.
 Pay attention: if long screws had been removed,
their square nuts on the back of measuring
chamber may be fallen out from the unit. See the
picture.

11.12.4 HGB head removal


 WBC measuring chamber must be removed, as it is written above.
 Pull off the HGB head from pre-amplifier connector.
167

11.13 Display unit


Display unit is installed on to the front panel of Abacus 5 with a metal mounting unit. The LED display is
controlled directly by Hardware block, the touch screen is driven by a Touch Controller board. It is
behind the display, under a metal cover.

Display removal:

 Turn the system off


 Remove the front cover fixing arm and two hinges
 Disconnect the grounding cable
 Take off the front panel and put it on flat surface
 Remove the metal cover of Touch Controller board
 Disconnect the cables
 Remove the metal mounting plate by taking out four screws
 Remove the LED display from the metal plate
168

11.14 Abacus5 Fluidics v2.1 - SV in Chamber Position


169

11.15 Abacus5 Fluidics v2.1 - SV in Needle Position


170

11.16 Tubings
Preparing of tubings:

= (A556)

= (A545) = (A546)

T0 T1

4 8 MEA V15/1 7 12 V2/2


V12/ 2
PUFFER
10 2

V1/3 V12/3

3,5 3,5
2 V 13/2 V12/1 T2
3,5 2
2 2 2,5 2,5
V4/3 V6/3 V8/3 25
10 6
P 3,5
3,5 3,5
3 2
2 2 2
V3/3 V5/3 V7/3 V23/2

V10/1 T3 T4
8 WBC
4 Cham 2 7 2 SVF12
2,5 22 BASO
Puffer
2
2

16 40 2
V15/2 V9/1 V11/1
V29/1 V11/2

V13/1 T5 T6
V36/2 22 4 V17/3
2
Washing 5 114 4 V14/1 3
head
5
V26/2

V20/2 T7
Dilu I. T8
4 puffer 9.

47 SHEAT
PUFFER 31 V16/1
Dilu II. 6.
1,7 puffer
V32/2
171

DILU II. DILU I.


PUFFER PUFFER T9

2 2

V43/1 2,5 18,5 8 20 3 9,3 V36/1

3 2 2

V41/1 2,5 V42/1 V40/1

10

V37/1

T10 RBC T11


MIX
2
2

V42/2 40 V29/2 40

9. 9.

V9/2 V10/2

MEA
T12 T13
PUFFER
V34/2
2,5
2 20 V3/2
MEA Puff
Small 6. 3 V35/2 7
MEA
PUFFER

Dilu I. T14 MEA T15


Puffer 4 Puffer
Small 10
20 V8/2 23 V23/1
. Dilu II 4 MEA
Puffer 10
Puffer
Small
172

CK1

CL1 CL2 CL3 CL4

Cham MIX RBC BASO

1 Puffer
CF2 CF3 CF4
1 2
2
CA3 CA4
MEA 3
PUFFER

C1 C2 C3 C4

1
1
3
2 2 1 3 2 2 2 2 2 2 2 2
2 1 1 2 1 1 1 1 1 1 1
V1 V2 V3 V4 V5 V6 V7 V8 V9 V10 V11 3
V12
V12 V12
V13 V12
V14 V12
V15 V12
V16 V12
V17 V12
V18 V19 V20

2 2 2 2 2 2 2 1 1
4
3 3 3 3 3 3 3 2 2
1 1 1 1 1 1 1

4 5
6
T6

7
T6 T6
5 8
5

2 1 1 3 1 1 1 2 1 3
9
3 3 3 3 1 3
1 2 2 2 2 2 2 1 2
10
T6

V23 V24 V25 V26 V28 V29 V30 V32 V33

V34 V35 V12


V36 VV37
12 V12
V38 V12
V39 V12
V40 V12
V41 V12
V42 VV43
12 VV44
12 11
1 1 2 2 2 2 2 2 2 2 2
2 2 3 3 3 3 3 3 3 3 3
1 1 1 1 1 1 1 1 1 12
T6
34

13
40

T12
14
T12

40
4 3 3 3 3 3 34 4 3 15

16

17

18

19
Mea
Sheath Stop Lyse Dilu I. Dilu II.
Puffer
Puffer Puffer Puffer Puffer Puffer
Small
40
34

1 1 1 1 1 1
2 2 2 2 2 2
40

34
40
173

CK1
= Kicsi
CL1 CL2 CL3 CL4
= Nagy
Cham MIX RBC BASO

1
Puffer
CF2 CF3 CF4
1 2
2
CA3 CA4
MEA
PUFFER 3
C1 C2 C3 C4
T3
T2 19 cm

T3
3 T2
T3 T3

T2

T2

T2

T2
T3
T2
T2

1
1
3
2 2 1 3 2 2 2 2 2 2 2 2
2 1 1 2 1 1 1 1 1 1 1

V1 V2 V3 V4 V5 V6 V7 V8 V9 V10 V11 3

T3
V12
V12 V12
V13 V12
V14 V12
V15 V12
V16 V12
V17 V12
V18 V19 V20

2 2 2 2 2 2 2 1 1
4
3 3 3 3 3 3 3 2 2
1 1 1 1 1 1 1

4 5
T2 6
T6

T2
7
T2

16 cm T3 T2
T6 T6
T2
5 8
5

8
11
5

3
2 1 1 3 1 1 1 2 1 3
9
3 3 3 3 1 3
1 2 2 2 2 2 2 1 2
10
T6

V23 V24 V25 V26 V28 V29 V30 V32 V33

V34 V35 V12


V36 VV37
12 V12
V38 V12
V39 V12
V40 V12
V41 V12
V42 VV43
12 VV44
12 11
1 1 2 2 2 2 2 2 2 2 2
2 2 3 3 3 3 3 3 3 3 3
1 1 1 1 1 1 1 1 1 12
T6
11
34

13
40

T12
14
T12

40
4 3 3 3 3 3 34 4 3 15

16

17

18

19
Mea
Sheath Stop Lyse Dilu I. Dilu II.
Puffer
Puffer Puffer Puffer Puffer Puffer
Small
40
34

1 1 1 1 1 1
2 2 2 2 2 2
40

34
40
174

12
CK1

CL1 CL2 CL3 CL4

12
Cham MIX RBC BASO

1
Puffer
CF2 CF3 CF4
1 12
2 2
CA3 CA4
MEA 3

19
33
PUFFER

C1 C2 C3 C4
T3
T2

T3
3 T2

33 T3 T3

T2

T2

T2

T2
28 T3
T2
T2

19
1
3
2 2 1 3 2 2 2 2 2 2 2 2
2 1 1 2 1 1 1 1 1 1 1
3

19
V1 V2 V3 V4 V5 V6 V7 V8 V9 V10 V11

T3
V12
V12 V12
V13 V12
V14 V12
V15 V12
V16 V12
V17 V12
V18 V19 V20

2 2 2 2 2 2 2 1 1
4
3 3 3 3 3 3 3 2 2
1 1 1 1 1 1 1

4 5
T2 6
T6

40 19 T2
7
T2
28

T2
T3
T6 T6 19
5 8
T2
33

8
11
5

3
2 1 1 3 1 1 1 2 1 3
9
3 3 3 3 1 3
1 2 2 2 2 2 2 1 2
10
T6

V23 V24 V25 V26 V28 V29 V30 V32 V33

V34 V35 V12


V36 VV37
12 V12
V38 V12
V39 V12
V40 V12
V41 V12
V42 VV43
12 VV44
12 11
1 1 2 2 2 2 2 2 2 2 2
2 2 3 3 3 3 3 3 3 3 3
1 1 1 1 1 1 1 1 1 12
T6
11
34

28 13
40

T12
14
T12

40
4 3 3 3 3 3 34 4 3 15
40

16

17

18

19
Mea
Sheath Stop Lyse Dilu I. Dilu II.
Puffer
Puffer Puffer Puffer Puffer Puffer
Small
40
34

1 1 1 1 1 1
2 2 2 2 2 2
40

40 34
40
175

12 CK1

CL1 CL2 CL3 CL4

12
Cham MIX RBC BASO
Puffer
1 CF2 CF3 CF4
1 12
2 2
CA3 CA4
MEA

19
36
33
PUFFER 3

42
C1 C2 C3 C4
T3
T2

T3
3 T2
36
33 42 T3 T3
28

T2

T2

T2

T2
T3
T2
T2

19
1
3
2 2 1 3 2 2 2 2 2 2 2 2
2 1 1 2 1 1 1 1 1 1 1
3

19
V1 V2 V3 V4 V5 V6 V7 V8 V9 V10 V11

T3
V12
V12 V12
V13 V12
V14 V12
V15 V12
V16 V12
V17 V12
V18 V19 V20

2 2 2 2 2 2 2 1 1
4
3
1
3
1
3
1
3
1
3
1
3
1
3
1 2 2 A5433
4 5
T2 6
36
42

T6

40 19 T2
14 7
T2
28

HF STOP 2 (20)
T2
T3
T6 T6 19
5 8
T2
33

14 5

8
11
5

3
2 1 1 3 1 1 1 2 1 3
9
3 3 3 3 1 3
1 2 2 2 2 2 2 1 2
17
14

10
T6

V23 V24 V25 V26 V28 V29 V30 V32 V33


14

V34 V35 V12


V36 VV37
12 V12
V38 V12
V39 V12
V40 V12
V41 V12
V42 VV43
12 VV44
12 11
1 1 2 2 2 2 2 2 2 2 2
2 2 3 3 3 3 3 3 3 3 3
1 1 1 1 1 1 1 1 1 12
T6 14
11
34

28 14 13
14
40

T12
14
T12

40
4 3 3 3 3 3 34 4 3 15
40

16

17

18

19
Mea
Sheath Stop Lyse Dilu I. Dilu II.
Puffer
Puffer Puffer Puffer Puffer Puffer
Small
40
34

1 1 1 1 1 1
2 2 2 2 2 2
40

40 34
40
176

12
CK1

CL1 CL2 CL3 CL4

12
Cham MIX RBC BASO
Puffer
1
CF2 CF3 CF4
1 12
2 2
CA3 CA4
MEA

19
36
33
PUFFER 3

42
C1 C2 C3 C4
T3

T2

T3
3 36 T2
T3 T3
33 42

T2
28

T2

T2

T2
T3

T2
T2

19
1
3
2 2 1 3 2 2 2 2 2 2 2 2
2 1 1 2 1 1 1 1 1 1 1
3

19
V1 V2 V3 V4 V5 V6 V7 V8 V9 V10 V11

T3
V12
V12 VV13
12 V12
V14 V12
V15 V12
V16 V12
V17 V12
V18 V19 V20

2 2 2 2 2 2 2 1 1
4
3
1
3
1
3
1
3
1
3
1
3
1
3
1 2 2 A5433
4 5
T2 14 6
36
42

T6

14
40 19 T2
14 7
T2
28

HF STOP 2 (20)

T2
15

T3
12

T6 T6 19
5 8
T2 HF LYSE 2 (20)
33

14 5

8
11
2
14
5

3
5

2 1 1 3 1 1 1 2 1 3
9
3 3 3 3 1 3
1 2 2 2 2 2 2 1 2
17
14

10
T6

V23 V24 V25 V26 V28 V29 V30 V32 V33


14

V34 V35 V12


V36 VV37
12 V12
V38 V12
V39 V12
V40 V12
V41 V12
V42 VV43
12 VV44
12 11
1 1 2 2 2 2 2 2 2 2 2
2 2 3 3 3 3 3 3 3 3 3
1 1 1 1 1 1 1 1 1 12
T6 14 14
11
34

28 14 13
14
40

T12
14
14
T12

40
T15

4 3 3 3 3 3 34 4 3 15
40

16

17

18

19
Mea
Sheath Stop Lyse Dilu I. Dilu II.
Puffer
Puffer Puffer Puffer Puffer Puffer
Small
40
34
T15

1 1 1 1 1 1
2 2 2 2 2 2
T15
40

40 34
40
177

1
2
3

4
5
6
7

9
10

11

12 38

13 37
37
37
35
14 35
35

15 17

16
L y se
L L yse
- EO
L - EO
17 D D D D

18

19
178

1
2
3

4
5
6
7

2 1
3
1
3
1
3
1
3 1
3
1
3
1
3
1
3
2 1
3
8
1 2 2 2 2 2 2 2 2 1 2

V2 3 V24 V2 5 V26 V27 V2 8 V29 V30 V3 1 V3 2 V33


9
V 34 V3 5 VV36
12 VV12
37 VV38
12 VV39
12 V1
V420 VV41
12 VV12
42 VV12
43 VV4
124

1 1
3
2
3
2
3
2
3
2
3
2
3
2
3
2
3
2
3
2 10
2 2 1 1 1 1 1 1 1 1 1

11

38
37
38 12

37
37 13
35

37
35

37 14
35

17
37 15
35
16

35 17
35
18

19
179

A5433

CK1
10
CL1 CL2 CL3
CL4 WF2 WF1
Ch a m MIX RB C BAS O WB C
Puff er
CF2 CF3 CF4

CA3 CA4

C1 C2 C3 C4

25
V 28 /1 40SVT4
T4 24
V18/ 1 30
V33 / 2 40SVT3
V44/1 37
V30/ 1 32
V2 /1 Las e
35 SVT1
V24/ 2 30

19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
30
24

30
40SVT4
40SVT3
35 SVT1

Upper
Las e

Under
14
32
37

11
15 1312 7
16
15

10 25 4, 5
7 SVT 1

8 26
1 6 7 27
23 4 5
4, 5 S V T 3
180

CK1

CL1 CL2 CL3


CL4
Ch a m MIX RB C BAS O
Puff er
CF2 CF3 CF4

CA3 CA4

C1 C2 C3 C4

A5439
2x 3cm

V20/ 1 400
V19 /1 26SVT2
47
54
45

V33 / 1 22
V24 /1 30
V25 / 2 30

19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1

30 47
45
22
54
Under 20
14 26SVT2
400

11 13 15
15 1312
Upper

2
30

16 10 24 12 14 16
25 11
25 10
8 26 9
1 67 27 26 8
23 4 5 7
6 5 3 21
4

Note: 400cm tube from V20 to SV upper disc is type A5433 1.27/2.23
181

12 Appendix

12.1 Recommended kit of tools


 PC standard keyboard (USB)
 Screwdrivers:
Slot Screwdrivers

Cross Slot Screwdrivers (Philips)

Hexagon Screwdrivers (3.5, 2.5, 2.0, 1.5 mm sizes)

 Pocket digital multimeter


 Diagonal Cutter (plier)
 Nipper

12.2 How to send .rp files to assessment of analyzer performance


Insert a pendrive (USB stick) in one of the USB ports. From ’Database’ press ’multiselect’ then select the
results you want to save( It is very important to save some human blood measurement results not just
controls), then press ’manage records’. A window will appear, press ’export’. The analyzer will search for
the pendrive, usually it will be displayed as drive E:.

Double click on the folder icon (just like in windows), the content of the pendrive will appear. Press ’OK’.

rp files will be saved. You can send these files on to Support for investigation of analyzer performance.
182

12.3 How to use the „Collect” function of the Abacus 5 hematology


analyzer

When is it needed?

- After servicing the analyzer (for documentation purposes)


- When requested by the service engineer (user can do it)
- When requested by Diatron Support

How to do it?

- Run a blank measurement*


- Run a patient sample*
- Run a control measurement*
- From ’Menu/Diagnostics/Self-test’ click on ’Start both’ button and wait for the result.
- Insert a USB Stick in one of the USB ports on back of the analyzer
- From ’Diagnostics/Information’ click on ’Refresh data’ then on ’Collect’ button.
- When the window with drive E:/ appears select the drive then click on ’OK’ button
- This function saves the following information in a .gz file:
o Self-test result
o Stress Blank result
o Last 10 Human measurement rp-file
o Last 3 Control measurement rp-file
o Last Blank measurement rp-file
o Event log
o Statistics
o User/Service/calibration factors
o SW – PIC – Firmware versions

Keep the file for your records or send it to support@diatron.com if asked by support staff.

* Not mandatory

end of document

You might also like