About this Manual
P/N: 01.54.000639
MPN: 01.54.000639010
Release Date: May, 2015
© Copyright EDAN INSTRUMENTS, INC. 2015. All rights reserved.

Statement
This manual will help you understand the operation and maintenance of the product better. It is
reminded that the product shall be used strictly complying with this manual. User's operation
failing to comply with this manual may result in malfunction or accident for which EDAN
INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
materials contained in this manual shall not be photocopied, reproduced or translated into other
languages.
Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information are contained in this manual, the user shall
not disclose such information to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of EDAN.
EDAN holds the rights to modify, update, and ultimately explain this manual.

Responsibility of the Manufacturer


EDAN only considers itself responsible for any effect on safety, reliability and performance of
the equipment if:
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by
persons authorized by EDAN, and
The electrical installation of the relevant room complies with national standards, and
The instrument is used in accordance with the instructions for use.
Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other
information to help qualified technician to maintain and repair some parts, which EDAN may
define as user serviceable.

Terms Used in this Manual


This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.

I
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
NOTE
A NOTE provides useful information regarding a function or a procedure.

II
Table of Contents
1 Warranty and Service ................................................................................................................. 1
2 Safety Guide................................................................................................................................. 2
2.1 Introduction ............................................................................................................................ 2
2.2 General Information ............................................................................................................... 2
2.3 Safety Precautions .................................................................................................................. 2
3 Installation ................................................................................................................................... 5
3.1 Environmental Requirements ................................................................................................. 5
3.2 Power Supply Requirements .................................................................................................. 5
3.3 Safety Requirements ............................................................................................................... 5
3.4 Installing the Doppler ............................................................................................................. 6
3.4.1 Putting the Doppler on a Flat Surface .............................................................................. 6
3.4.2 Installing the Doppler on a Trolley .................................................................................. 6
3.4.3 Installing the Charge Stand (SD3 PLUS, SD3 PRO) ....................................................... 6
3.5 Connecting to AC Power ........................................................................................................ 7
4 Overview ...................................................................................................................................... 8
4.1 Principle Introduction ............................................................................................................. 8
4.2 Appearance and Interfaces...................................................................................................... 8
5 Tests and Maintenance.............................................................................................................. 10
5.1 Functional Check .................................................................................................................. 10
5.1.1 Switching on Check ....................................................................................................... 10
5.1.2 Screen Check .................................................................................................................. 10
5.1.3 Keys Check .................................................................................................................... 10
5.1.4 Probe Check ................................................................................................................... 10
5.2 Performance Test .................................................................................................................. 11
5.3 Safety Test ............................................................................................................................ 11
5.3.1 Patient Leakage Current Test ......................................................................................... 12
5.4 Maintenance.......................................................................................................................... 12
5.4.1 Maintenance of the Main Unit ....................................................................................... 12
5.4.2 Maintenance of the Probe............................................................................................... 12
5.4.3 Maintenance of the Battery ............................................................................................ 13
6 Troubleshooting ......................................................................................................................... 14
7 Module Malfunction Verification ............................................................................................ 16
7.1 Verifying Malfunction of the Main Board ............................................................................ 16
8 Disassembling the Doppler ....................................................................................................... 18
8.1 Tools Required ...................................................................................................................... 18
8.2 Replacing Batteries ............................................................................................................... 18
8.3 Disassembling the Main Unit ............................................................................................... 19
8.4 Replacing the Main Board .................................................................................................... 20
8.5 Replacing the Speaker .......................................................................................................... 21
8.6 Replacing the Charge Adapter Board ................................................................................... 22
8.7 Disassembling the Charge Stand .......................................................................................... 22
8.8 Replacing the Charge Board of the Charge Stand ................................................................ 23
8.9 Replacing the Interface Board of the Charge Stand ............................................................. 24
Appendix 1 Renewal Parts .......................................................................................................... 25

III
SD3 Series Ultrasonic Pocket Doppler Service Manual Warranty and Service

1 Warranty and Service


Warranty
EDAN provides a one-year-warranty free of charge for EDAN's products (accessories are
included). The warranty period begins on the date the products are shipped to customers. The
warranty can be extended with a fee, but the extended warranty will not cover the accessories.
If a customer promptly notifies EDAN of customer's warranty claim hereunder, EDAN will either
repair, adjust or replace (with new or exchange replacement parts) the warranted products. EDAN
warrants that any service it provides to customers will be performed by trained individuals in a
workmanlike manner.
Limitation of Warranty
Direct, indirect or final damage and delay caused by the following situations for which EDAN is
not responsible may void the warranty:
 Groupware is dismounted, stretched or redebugged.
 Unauthorized modification or misuse.
 Damage caused by operating beyond the environmental specifications for the medical
product.
 Change or remove original serial number label or Manufacturer symbol.
 Improper use.
Contact Information
If you have any question about maintenance, technical specifications or malfunctions of devices,
do not hesitate to contact us.
EDAN Instruments, Inc.
TEL: +86-755-26898321, 26899221
FAX: +86-755-26882223, 26898330
E-mail: support@edan.com.cn

-1-
SD3 Series Ultrasonic Pocket Doppler Service Manual Safety Guide

2 Safety Guide

2.1 Introduction
This service manual is a reference for periodic preventive maintenance and corrective service
procedures for the SD3 series ultrasonic pocket Doppler.
It provides troubleshooting information, disassembly and assembly procedures, instructions for
functional testing and performance verification. It is intended for use only by technically
qualified service personnel.

WARNING
When performing a service procedure, follow the instructions as presented in this manual.
Failure to do so might damage the device, invalidate the product warranty or lead to
serious personal injury.

2.2 General Information


The SD3 series ultrasonic pocket Doppler (hereinafter called the Doppler) is designed in
accordance with the international safety requirements in IEC/EN 60601-1 for medical electrical
equipment. Classification information of the Doppler is as follows:
Anti-electric Shock Type: Internally powered equipment

Anti-electric Shock Degree:


Type B equipment
Main Unit: Ordinary equipment (Sealed equipment
Degree of Protection against without liquid proof)
Harmful Ingress of Water:
Probes: IPX8 Water Ingress Protection Code
Degree of Safety in
Equipment not suitable for use in presence of
Presence of Flammable
flammable gases
Gases:
Working System: Continuous running equipment
EMC: CISPR 11 Group 1 Class B

2.3 Safety Precautions


WARNING and CAUTION messages must be observed. To avoid the possibility of injury,
observe the following precautions during the operation of the device.

WARNING
1 The Doppler must be serviced only by authorized and qualified personnel. EDAN
does not assume any responsibility for damage or injury if modifications or repairs
are carried out by unauthorized personnel.

-2-
SD3 Series Ultrasonic Pocket Doppler Service Manual Safety Guide

WARNING
2 The service personnel must be familiar with the operation of the Doppler. Refer to the
SD3 Series Ultrasonic Pocket Doppler User Manual for details.
3 This device is not explosion-proof and can not be used in the presence of flammable
anaesthetics.
4 SHOCK HAZARD - Do not attempt to connect or disconnect a power cord with wet
hands. Make certain that your hands are clean and dry before touching a power
cord.
5 Do not touch the signal input/output connector and the patient simultaneously.
6 Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the system standard IEC/EN
60601-1. Anybody who connects additional equipment to the signal input connector
or signal output connector to configure a medical system must ensure that the
system complies with the requirements of the valid version of the system standard
IEC/EN 60601-1. If in doubt, consult our technical service department or your local
distributor.
7 Only use the probes provided by the manufacturer. Use and replace the substitutive
parts provided or recommended by EDAN only.
8 The device shall only be used when the battery cover is closed.

CAUTION
1 The main unit is designed for continuous operation and is ‘ordinary’. Do not immerse
it in any liquid (i.e. not drip or splash-proof).
2 Keep the device in a clean environment and avoid vibration during storage.
3 Do not sterilize the Doppler with autoclave or gas.
4 Electromagnetic Interference - Ensure that the environment in which the device is
operated is not subject to any source of strong electromagnetic emissions, such as
radio transmitters, mobile telephones, etc.
5 The following safety checks should be performed once every two years or as
specified in the institution’s test and inspection protocol by a qualified person who has
adequate training, knowledge, and practical experience to perform these tests.
 Inspect the equipment for mechanical and functional damage.
 Inspect the safety relevant labels for legibility.
 Verify that the device functions properly as described in the instructions for use.
 Test the pregnant woman’s leakage current according to IEC 60601 Limit: d.c
10 µA, a.c 100 µA.

-3-
SD3 Series Ultrasonic Pocket Doppler Service Manual Safety Guide

CAUTION
The leakage current should never exceed the limit. The data should be recorded in
an equipment log. If the device is not functioning properly or fails any of the above
tests, the device has to be repaired.
6 The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose
them together with house-hold garbage.

-4-
SD3 Series Ultrasonic Pocket Doppler Service Manual Installation

3 Installation
WARNING
The device should only be installed by qualified service engineers.

3.1 Environmental Requirements


Temperature: +5 ºC ~ + 40 ºC ( +41 ºF ~ +104 ºF)

Working Relative Humidity: 25% RH ~ 80% RH (non-condensing)

Atmospheric Pressure: 86 kPa ~ 106 kPa


Temperature: -20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
Transport and
Storage Relative Humidity: 25% RH ~ 93% RH (non-condensing)

Atmospheric Pressure: 70 kPa ~ 106 kPa

CAUTION
1 Ensure that the device is not subject to any source of strong electromagnetic
emissions, such as radio transmitters, mobile phones, etc.
2 Do not install the device in the presence of flammable anesthetics.
3 Keep the environment clean. Avoid vibration. Keep the device far from corrosive
medicine, dust area, high-temperature and humid environment.

3.2 Power Supply Requirements


SD3 LITE, SD3 and SD3 VASCULAR are powered either by three alkaline batteries or three
rechargeable NI-MH batteries. SD3 PLUS and SD3 PRO are powered by a lithium battery.
 SD3 LITE, SD3, SD3 VASCULAR:
Alkaline battery: AA, LR6, 1.5 V
Rechargeable NI-MH battery: AA, R6, 1.2 V
 SD3 PLUS, SD3 PRO:
Lithium battery: 3.7 V/2600 mAh

3.3 Safety Requirements


WARNING
1 Do not touch the signal input/output connector and the patient simultaneously.
2 Do not switch on the Doppler until all components have been properly installed and
verified.

-5-
SD3 Series Ultrasonic Pocket Doppler Service Manual Installation

3.4 Installing the Doppler


3.4.1 Putting the Doppler on a Flat Surface
Put the Doppler on a flat surface. Make sure that the surface does not vibrate and is free of
corrosive medicine and dust.

3.4.2 Installing the Doppler on a Trolley


The trolley is optional. If the trolley has been ordered, an Assembly Instruction will be delivered
with the trolley. Please refer to the Assembly Instruction for installing the Doppler on the trolley.

Figure 3-1 Installing the Doppler on a trolley

3.4.3 Installing the Charge Stand (SD3 PLUS, SD3 PRO)


The charge stand is optional. The charge stand can be put on a table or fixed on the wall.
On a table: Put the charge stand on a flat table.
On the wall: Aim the two bracket holes in the back of the charge stand to the wall, leave two
marks and drill in the marked place. Screw in each hole and make sure that the screws are fixed
in the wall. Aim the bracket holes to the screws on the wall and fix the screws into the holes.

Figure 3-2 On a table Figure 3-3 On the wall

-6-
SD3 Series Ultrasonic Pocket Doppler Service Manual Installation

WARNING
1 The charge stand can be fixed only on a concrete wall.
2 When you need to install the charge stand a second time, drill new holes and change
the screws.

3.5 Connecting to AC Power


The Doppler is internally powered equipment. Only when rechargeable batteries are low in
energy, connecting to AC power is required.
If the rechargeable NI-MH batteries are low in energy, take them out from the main unit, fit them
properly in the slots of the provided charger, and plug the charger into a power supply socket (AC
100-240 V, 50/60 Hz).
For SD3 PLUS and SD3 PRO, you can charge the Doppler directly or through the charge stand.
Charging the Doppler directly:
Connect the provided power adapter to the charge socket of the Doppler, and plug the power
adapter to a power supply socket (AC 100-240 V, 50/60 Hz).
Charging the Doppler through the charge stand:
Connect the provided power adapter to the charge socket of the charge stand, and plug the power
adapter to a power supply socket (AC 100-240 V, 50/60 Hz). The Doppler is charged when placed
in the charge stand.

WARNING
1 Make sure that the batteries are not connected reversely before plugging the charger
into the power socket.
2 The Doppler is not available for examination during charging.

-7-
SD3 Series Ultrasonic Pocket Doppler Service Manual Overview

4 Overview
4.1 Principle Introduction
The ultrasonic Doppler is the essential principle applied to the device.
The ultrasonic Doppler principle is as follows: When the ultrasound wave from the emitter
crystal encounters an active fetal heart, Doppler shift happens. As a result, there is a frequency
offset between the received ultrasound wave and the emitted ultrasound wave. The receiver
crystal receives the ultrasound wave and converts it to electric signals. The fetal heart sound is
acquired after the electric signals are amplified by the front end and processed by the
demodulation and filtering circuit. With envelope detection and autocorrelation algorithm, the
device calculates the fetal heart rate.

4.2 Appearance and Interfaces


The following figure shows the appearance of SD3 PRO, while other models are similar. You can
also refer to the SD3 Series Ultrasonic Pocket Doppler User Manual for details.

Screen
Probe socket

Audio socket
Volume
Battery control
compartment

Power
Charge socket button

Record/Play Charge stand Speaker


button socket

Figure 4-1 Appearance of SD3 PRO

-8-
SD3 Series Ultrasonic Pocket Doppler Service Manual Overview

Figure 4-2 and Figure 4-3 shows the main board interfaces of each model respectively.

Probe J6 J2(+) V
R Volume
J4 1 control
Earphone socket

Ba
ttery
J1(-)
Speaker J7

Figure 4-2 Main board interfaces of SD3 LITE, SD3 and SD3 VASCULAR

Probe detection

SW3
Probe J6 J2(+) V
R Volume
J4 1 control
Battery

Earphone or audio
cable interface J8

SW2
Charge socket J1(-)

Record/Play
J9
button

Speaker Charge board

Figure 4-3 Main board interfaces of SD3 PLUS and SD3 PRO (SW2 not available in SD3 PLUS)

-9-
SD3 Series Ultrasonic Pocket Doppler Service Manual Tests and Maintenance

5 Tests and Maintenance


5.1 Functional Check
A complete functional check should be performed if the device is suspected of malfunction or
each time after service. It is unnecessary to open the device enclosure for functional checks.

WARNING
Qualified service personnel should perform complete functional checks after the device is
installed to ensure proper running of the device.

5.1.1 Switching on Check

Press the POWER button on the front panel to switch on the Doppler. The OLED

screen displays .

If the probe is not connected or poorly connected, the OLED screen displays . You
should reconnect the probe properly.
When the probe is well connected, the OLED screen shows the probe frequency, for example,

5.1.2 Screen Check


Check if any symbol or character is missing on the OLED screen, or if there is any especially
bright or dark spot. Check if the OLED screen displays symbols normally.

5.1.3 Keys Check


Press the keys on the main unit to check if they work properly.

5.1.4 Probe Check


Switch on the Doppler. Touch the acoustic surface of the probe gently with your hand to check if
there is any sound coming from the speaker.
NOTE:
If any failure is detected, refer to chapter 6 Troubleshooting for details.

- 10 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Tests and Maintenance

5.2 Performance Test


This test checks the performance of FHR detection.
Required tool: Ultrasonic Doppler FHR Accuracy Tester System
Procedure:
1. Connect the tester and the device as shown below:

Ultrasonic Doppler FHR Accuracy Tester Main unit

Probe

2. Switch on the main unit.


3. Generate the heart rate 50 bpm, 90 bpm, 120 bpm, 150 bpm, 180 bpm, 210 bpm and 240 bpm
in sequence.
4. Check the FHR value displayed on the screen. If the error is within the range of ±2 bpm, the
test is passed.

5.3 Safety Test


The safety test should be performed by qualified personnel every year and each time after service.
The safety test described here is not a substitute for local safety test where it is required for an
installation or a service event. Additional tests may be required by your local standards and
regulations.
When performing a safety test, use a standard safety analyzer such as Fluke 601Pro series safety
analyzer or the equivalent. Perform the test according to your local regulations, for example, in
Europe according to IEC/EN60601-1 and in USA UL60601-1. For the test setup, please refer to
the Instructions for Use of the test equipment used.
You are recommended to record the results of the safety test.

- 11 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Tests and Maintenance

5.3.1 Patient Leakage Current Test

① —— DUT (Device Under Test) enclosure ④ —— Power supply ⑤ —— Applied part


P3 —— Plug, socket, or terminal that will be connected to the patient MD —— Measurement device

NOTE:
The circuit diagram is based on the Fluke 601Pro series safety analyzer.
This test measures the patient leakage current of the DUT and its applied part located within the
patient environment.
Allowable value:
Normal condition (NC): AC: 100 µA, DC: 10 µA (IEC/EN60601-1)

5.4 Maintenance
For details about basic cleaning and maintenance methods, refer to the Maintenance chapter in
the SD3 Series Ultrasonic Pocket Doppler User Manual. For further technical support, contact
service engineers of EDAN.

5.4.1 Maintenance of the Main Unit


Check if the equipment has any visible evidence of damage that may affect the patient and the
operator's safety or the Doppler's performance before each use. If any damage is detected,
replacement is recommended.
The gathering of dew on the screen may occur with abrupt temperature or humidity changes. A
stable environment is recommended.
Keep the exterior surface of the main unit clean, free of dust and dirt.
Do not scratch or damage the screen.

5.4.2 Maintenance of the Probe


Handle the probe with care. Avoid contacting the probe with hard or sharp objects. Wipe the
remaining gel off the probe after each use.

- 12 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Tests and Maintenance

5.4.3 Maintenance of the Battery


If the Doppler is not used for a long time, please remove the battery, and store the battery in a
cool and dry environment at a temperature within the range of 0°C ~ +45°C (+32ºF ~ +113ºF).
For rechargeable NI-MH batteries or lithium batteries, which have a service life of more than 300
charge-discharge cycles, we recommend the following:
 Store the battery with about 50% capacity. Do not store the battery in a fully charged or fully
discharged state.
 If the battery is stored for more than 6 months, charge it for about 2 hours once every 6
months.
 After the Doppler has been stored for a long time, fully charge the battery before using the
Doppler.

- 13 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Troubleshooting

6 Troubleshooting
A damaged PCB or part can be replaced. If replacement is necessary, refer to Chapter 8 for
details.

Phenomenon Possible Cause Solution


The speaker is muted. Turn up the volume.
Bad connection between the Make sure that the probe is well
probe and main unit. connected to the main unit.
The speaker does not sound. Main board defective. Replace the main board.
Speaker defective. Replace the speaker.
Probe defective. Replace the probe.
The volume is too low. Turn up the volume.
Apply coupling gel on the probe
The speaker sounds weak. No coupling gel is applied.
acoustic surface.
Replace the battery or charge the
The battery is in low energy.
rechargeable battery.
The probe is not placed at Locate the probe at a position
the optimum position. for optimal FHR examining.
Low sensitivity.
Apply coupling gel on the probe
No coupling gel is applied.
acoustic surface.
The screen displays nothing Main board defective. Replace the main board.
and the speaker does not
sound after the main unit is Replace the battery or charge the
The battery is in low energy.
switched on. rechargeable battery.
The speaker sounds but the
screen displays nothing or
displays wrong characters Main board defective. Replace the main board.
after the main unit is
switched on.
Make sure that the environment
The device is subject to in which the device is operated
strong electromagnetic is not subject to any source of
emissions. strong electromagnetic
emissions.
Speaker defective. Replace the speaker.
Probe defective. Replace the probe.
Loud noise. Main board defective. Replace the main board.
The probe has not been
Take out the probe from the
taken out from the probe
probe holder before switching
holder when the main unit is
on the main unit.
switched on.
Replace the battery or charge the
The battery is in low energy.
rechargeable battery.
Too much coupling gel on Wipe the residual coupling gel
the probe acoustic surface. off the probe after each use.

- 14 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Troubleshooting

The screen displays nothing Power adapter defective. Replace the power adapter.
or displays charge error Battery defective. Replace the battery.
when the Doppler (SD3
PLUS or SD3 PRO) is being Main board defective. Replace the main board.
charged. Charge stand defective. Replace the charge stand.

- 15 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Module Malfunction Verification

7 Module Malfunction Verification


If a module is suspected of malfunction, verify it by using the method described in this chapter.
NOTE:
For all the interfaces, the first pin is the one with a square solder pad.

7.1 Verifying Malfunction of the Main Board


To verify the main board, perform the following steps:
1) Disassemble the main unit by referring to section 8.3.
2) Check if pin-to-ground short circuit happens to the pins listed below using a multimeter. After
confirming that short circuit does not happen, power on the main board.
3) Measure the voltage of these pins using the multimeter.

No. Pin Reference result NOTE


1 TP2 (+5±5%) V
This pin is not available in SD3
2 TP4 (+9±5%) V
LITE or SD3 VASCULAR.
3 TP5 (+3.3±5%) V

TP2

- 16 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Module Malfunction Verification

TP4 TP5

4) If any of the results exceeds the reference range, the main board is considered defective.
Replacement of the main board is recommended.

- 17 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Disassembling the Doppler

8 Disassembling the Doppler


WARNING
1 The device enclosure can be opened only by qualified service personnel.
2 Switch off the device before disassembling. If the device is connected to AC power for
charging, remove the power adapter.
3 Each time after service, perform safety test before using the device again.

8.1 Tools Required


A cross-head screwdriver

8.2 Replacing Batteries


Replaceable Part Part No.
Alkaline Batteries 01.21.064086
Rechargeable NI-MH Batteries 21.21.064180
Lithium Battery 21.21.064198

WARNING
The direction of the battery should comply with the polar mark in the battery compartment.
Reversed connection is forbidden.

 SD3 LITE, SD3, SD3 VASCULAR:


1) Open the battery compartment by depressing the tab and pulling outward the battery cover.
2) Remove the batteries.
3) Put new batteries into the compartment.
4) Install the battery cover.

Alkaline batteries
or rechargeable
Battery cover NI-MH batteries

Tab

- 18 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Disassembling the Doppler

 SD3 PLUS, SD3 PRO:


1) Remove the screw fixing the battery cover with the cross-head screwdriver, then open the
battery compartment.
2) Remove the battery.
3) Put a new battery into the compartment.
4) Install the battery cover and fix it with the screw.

Lithium battery

Battery cover

8.3 Disassembling the Main Unit


The main unit consists of two parts: the upper unit and lower unit.
Switch off the main unit, open the battery compartment
and remove the batteries (see section 8.2).

Remove the 4 screws securing the upper and lower units


with the cross-head screwdriver.

Separate the upper unit from the lower unit.

Assemble the main unit in a reverse procedure.


 SD3 LITE, SD3, SD3 VASCULAR:

Upper unit
Lower unit

Main board

Speaker

- 19 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Disassembling the Doppler

 SD3 PLUS, SD3 PRO

Lower unit
Upper unit

Main board

Speaker

Charge adapter
board

8.4 Replacing the Main Board


Replaceable Part Part No.
SD3 main board 02.03.451495
SD3 LITE main board (also applied to
02.03.451497
SD3 VASCULAR)
SD3 PRO main board (also applied to
02.03.451453
SD3 PLUS)

Separate the upper unit from the lower unit


(see section 8.3).

Disconnect the cable from the main board.

Remove the 3 screws securing the main board with the


cross-head screwdriver.

Remove the interface panel and the volume control


button.

Remove the main board, then install a new one in a


reverse procedure.

- 20 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Disassembling the Doppler

 SD3 LITE, SD3, SD3 VASCULAR  SD3 PLUS, SD3 PRO

Volume control
button
Interface panel
Interface panel

Cable interface Cable interface

8.5 Replacing the Speaker


Replaceable Part Part No.
Speaker 01.14.038036

Separate the upper unit from the lower unit


(see section 8.3).

Disconnect the cable from the main board.

Remove the speaker, then install a new one in a


reverse procedure.

 SD3 LITE, SD3, SD3 VASCULAR  SD3 PLUS, SD3 PRO

Cable interface
Cable interface

Speaker

- 21 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Disassembling the Doppler

8.6 Replacing the Charge Adapter Board


(Applicable to only SD3 PLUS and SD3 PRO)

Replaceable Part Part No.


SD3 charge adapter board 02.02.451635

Separate the upper unit from the lower unit


(see section 8.3).

Disconnect the cable from the charge adapter board.

Remove the charge adapter board, then install a new


one in a reverse procedure.

Charge adapter board

Cable interface

8.7 Disassembling the Charge Stand


The charge stand is optional and applicable to only SD3 PLUS and SD3 PRO.
The charge stand consists of two parts: the front unit and rear unit.
Remove the connection with the AC power.

Remove the 4 screws securing the front and rear units


with the cross-head screwdriver.

Separate the front unit from the rear unit.

Assemble the charge stand in a reverse procedure.

- 22 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Disassembling the Doppler

Front unit Rear unit Rear unit Interface board Front unit

Charge socket Charge board

8.8 Replacing the Charge Board of the Charge Stand


Replaceable Part Part No.
Charge board of SD3 charge stand 02.02.451637

Separate the front unit from the rear unit


(see section 8.7).

Disconnect the cable from the charge board.

Remove the 2 screws securing the charge board with


the cross-head screwdriver.

Remove the charge board, then install a new one in a


reverse procedure.

Cable interface

Charge board

- 23 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Disassembling the Doppler

8.9 Replacing the Interface Board of the Charge Stand


Replaceable Part Part No.
Interface board of SD3 charge stand 02.02.451639

Separate the front unit from the rear unit


(see section 8.7).

Disconnect the cable from the interface board.

Remove the screw securing the interface board with


the cross-head screwdriver.

Remove the interface board, then install a new one in


a reverse procedure.

Interface board

Cable interface

- 24 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Renewal Parts

Appendix 1 Renewal Parts

CAUTION
Only connect the replaceable parts supplied by EDAN to the Doppler.

Part Part No.

SD3 main board 02.03.451495

SD3 LITE main board 02.03.451497

SD3 PRO main board 02.03.451453

SD3 charge adapter board 02.02.451635

Charge board of SD3 charge stand 02.02.451637

Interface board of SD3 charge stand 02.02.451639

SD3 2 MHz Probe 02.01.210721

SD3 3 MHz Probe 02.01.210722

SD3 4 MHz Probe 02.01.210875

SD3 5 MHz Probe 02.01.210876

SD3 8 MHz Probe 02.01.210877

Speaker 01.14.038036
Cable in the Doppler (SD3 PLUS and
01.13.036579
SD3 PRO)
Cable in the Doppler (SD3 LITE, SD3
01.13.036578
and SD3 VASCULAR)
Earphone 01.14.000831

Audio cable 01.13.036886

Normal carry case 01.56.465616

Alkaline batteries 01.21.064086

Rechargeable NI-MH batteries 21.21.064180

Lithium battery 21.21.064198

NI-MH battery charger 01.21.064218

- 25 -
SD3 Series Ultrasonic Pocket Doppler Service Manual Renewal Parts

Power adapter (American Standard) 21.21.064158

Power adapter (European Standard) 01.21.064161

Power adapter (Brazilian Standard) 21.21.064184

Power cable (Australian Standard) 01.13.03660

Power cable (Brazilian Standard) 21.13.036425

Power cable (English Standard) 01.13.036693