What information is required to set up most installed devices? Select all that apply.

AE Title

Host

IP Address

Port

Mark for follow up

What is the primary purpose of the Goaltab Generation Calibration?

Account for detector noise in dose calculation

Remove artifacts from images

Remove scatter

Account for heel effect in dose calculation

Mark for follow up

What is the AEC Scaling verification task in Caltool used for?

Verify the mAs is correct at various thicknesses

Verify there are no artifacts in images

Verify the KV is correct at various thickness

Verify proper dose tracking at various thicknesses

Mark for follow up

Why is the HVL Spot Check performed on the Dimensions PM?

To check if the tube is failing

To check if generator is failing

To test for tube failures

To verify the reported dose is correct

Mark for follow up

During an installation who is responsible for wiring the device to the electrical system?

The FSE

The site Biomedical team

The site electrician

The movers
Mark for follow up

How is the IP address of the Dimensions backed up?

Transfer from Node

The IT department backs it up

System Tools backup

Manually by the FSE

Mark for follow up

What is Hologic's preferred choice for a tomosynthesis support DICOM type?

CTO

BPO

SCO

BTO

Mark for follow up

How long must the detector be on before any calibrations for the detector can be performed?

1 hour

15 minutes

30 minutes

No wait time

Mark for follow up

What is the default 2D target dose for the Dimensions system?

1.2 mGy

1.3 mGy

1.45 mGy

1.6 mGy

Mark for follow up

What is the purpose of the QR Installed Device?


To get prior images from PACS

To send images to PACS

To get new procedures for the day

To update incorrect patient demographics

Mark for follow up

What is the proper method to clean the X-ray filters on the Dimensions?

Soft-bristled brush, alcohol if required

Degreaser

Canned compressed air

Leave them alone to prevent damage

Mark for follow up

Where is the tomo potentiometer (tomo pot) located?

Behind the detector

Left side of the gantry

Right side of the gantry

Back of the gantry

Mark for follow up

For any service call, how should you respond to the question, "Do I need a Medical Physicist?"

Tell the site that a Medical Physicist is needed

Tell the site to call the Physicist and report the service performed

The Medical Physicist is not needed

Call the Physicist and explain the service done for the site

Mark for follow up

What is the proper method for servicing the VTA on a PM?

Use a degreaser to remove grease and reapply

Reapply new grease over existing grease


Remove old grease completely and apply new grease

Reuse leftover grease from last PM and reapply

Mark for follow up

What is the most common cause of artifacts in Dimensions images?

Bad pixels

EMI

Debris in the beam path

Bad lines

Mark for follow up

When should the Biopsy Guidance Hardware calibration be performed? Select all that apply.

Every PM on the Affirm Upright

Every time an STX calibration is going to be performed

During the installation of the Affirm Upright

After major repairs of the Affirm Upright

Mark for follow up

How often should Filament Calibrations be performed during PMs?

During every other PM

Not required on a PM

During every PM

Only when a spot check fails

Mark for follow up

In which directory will the pixel mapping files be found for backup or artifact issues in the Dimensions?

C:\img

C:\gemini\data\HVL

C:\gemini\data\m35

C:\caltool_xfer

Mark for follow up


Which drawer is located at the top of the gantry?

HV Inverter

Power Distribution

HV Multiplier

PMC

Mark for follow up

Which of the following images types has a Generated 2D image included? Select all that apply.

Tomo

Combo HD

Combo

Tomo HD

Mark for follow up