Professional Documents
Culture Documents
Third Edition
HOW THE PIPETTE STORY BEGAN...
HOW THE PIPETTE STORY BEGAN…
Over a century ago, Louis Pasteur invented the glass Pasteur pipette to reduce
contamination when transferring samples. The Pasteur pipette is still in use today.
The next notable advancement in pipetting occurred in the late 1950s with the
introduction of a handheld, piston-operated pipette as a safe alternative to potentially
dangerous mouth pipetting. The first handheld pipettes were fixed volume pipettes,
meaning that they had a pre-established volume setting. After fixed volume pipettes,
variable volume pipettes were introduced, which provided a more flexible stepper
volume setting.
In 1972, Dr. Warren Gilson invented the first adjustable volume pipette. As an original
error-preventing feature, the selected volume was now clearly displayed on the Gilson
pipette (direct digital readout). Today, Gilson precision pipettes are still the world
standard for accuracy, precision, and reliability.
Today, Gilson continues its creative efforts and offers innovative, robust, and reliable
pipettes to help scientists in their daily work.
The pipetting system is our core expertise, and we truly enjoy sharing this knowledge
and experience with you, and other pipette users, to help achieve your goals.
Filled with tips and information from our pipetting experts, this guide covers all aspects
of pipetting, from selection to proper maintenance. Examples given are based on our
own pipette range, but the techniques described are equally applicable to other brands
of pipettes and tips.
TABLE OF CONTENS
Appendix | 45
Appendix A: Pipetting Terms | 45
Appendix B: Example of a Performance Check | 47
Appendix C: Z Factor | 48
Appendix D: Evaporation Loss | 49
Appendix E: Chemical Resistance of Plastics | 50
FAQs | 51
• Push button
AIR-DISPLACEMENT PIPETTE
• Large volume
adjustment
knob
Thumbwheel
(also known as fine
volume adjustment ring)
Volumeter display
Tip ejector
Shaft
(also known as
tip holder)
Disposable tip
Identification
mark Sealing ring
Level
The updated line of mark
PIPETMAN Classic
is identified with an
NOTE underlined serial
number starting
from QG.
Figure 1
PIPETMAN® diagram*
*Pipettes with different designs are available. For more information, visit www.gilson.com.
• Push button
POSITIVE-DISPLACEMENT PIPETTE
•V
olume adjustment
knob
• Capillary and piston
ejector
1 Volumeter
5 display
0
T051401
Shaft
(also known as
1 capillary holder)
5
0
Disposable
capillary and piston
T051401
Serial number
Figure 2
MICROMAN® E diagram*
Piston
Shaft
Shaft
Disposable
piston
Disposable
capillary
Air cushion
Disposable tip
Piston seal
Sample
Sample
The following infographic shows how the piston determines the volume of air displaced and
subsequently the volume of sample aspirated.
1 2
1
0
0 Set Volume Prepare for Aspiration
The required The push button is
volume is set. pressed prior to sample
The piston aspiration. The piston
moves to the descends and expels a
appropriate volume of air equal to the
position. selected volume of liquid.
3 4
1 2
1 1
0 1 0
0 0 0
0 Set Volume Prepare for Aspiration
The required volume The push button
is set. The piston is pressed prior
moves down to the to sample aspiration.
appropriate start The piston descends
position. down to the end of the
capillary.
3 4
Table 1
Recommendations for pipetting different volumes
Molecular biology
Microbiology
Positive-displacement
M10, M25, M50 M100, M250, M1000
pipettes
% Systematic error
5
Regular air-displacement pipette
4.5
% Systematic error
4
3.55
Regular air-displacement pipette
3
4.5
2.54
2
3.5
1.53
1
2.5
0.52
1.5
0
MICROMAN® E positive-displacement pipette
1
0.5
0 0 40 60 80 85 90 95 100 % Glycerol
MICROMAN® E positive-displacement pipette
Viscosity/Density
0 40 60 80 85 90 95 100 % Glycerol
Figure 3 Viscosity/Density
Systematic Error (%) when using a MICROMAN® E positive displacement pipette vs. a regular air
% Random
displacement pipette error
3.5
Regular air-displacement pipette
3 Random error
%
2.5
3.5
Regular air-displacement pipette
23
1.5
2.5
12
0.5
1.5
01 MICROMAN® E positive-displacement pipette
0.5
0 40 70 80 85 90 95 100 % Glycerol
0 Viscosity/Density
MICROMAN® E positive-displacement pipette
0 40 70 80 85 90 95 100 % Glycerol
Viscosity/Density
Figure 4
Random Error (%) when using a MICROMAN® E positive-displacement pipette vs. a regular air
displacement-pipette
User-to-User Variability
Motorized pipettes can help reduce variability between users. There are many
factors that can affect your pipetting results, including volume setting, pipetting
technique, and the rate of aspirating and dispensing. With a motorized pipette
you can set the exact volume on the digital display — the motor uses the same
pipetting force every time and maintains the same rate of speed when aspirating
and dispensing a sample.
Aliquoting
To deliver several aliquots without refilling,
you may either choose the repetitive mode
of a motorized air-displacement pipette,
or use a positive-displacement repeater.
Repeaters enable up to 125 aliquots,
whereas the number of aliquots with
air-displacement motorized pipettes will
depend on the pipette volume.
For operations fewer than ten aliquots,
using a motorized air-displacement pipette
is likely the better option.
Pipetting techniques
Adjust the Volume Display
The forward mode is the standard way of pipetting with an air-displacement pipette like
PIPETMAN.
Pipetting techniques
1 2 3 4
Prepare Aspirate Dispense Purge
Hold the pipette Immerse the Place the pipette Wait one second, then
in a nearly vertical pipette tip in the tip at an angle depress the plunger to
position. Depress liquid.* Allow the (10° to 45°) against the second stop
the plunger plunger to move the inside wall position. This purge
smoothly to the up smoothly to the of the receiving stroke removes any
first stop position. rest position. Wait vessel. Depress the remaining sample
one second so that plunger smoothly from the tip. Remove
all the liquid has to the first stop pipette tip end from
time to move up position. sidewall by sliding
into the tip. it up the wall.
Rest position
First stop
Second
stop
(purge)
* The immersion depth of your tip can have a significant effect on your results (for depth per model, see table above). If the tip is
immersed too deeply, droplets will form on the outside of the tip and they will be deposited along with your sample. If the tip is
not immersed deeply enough, vortexing will occur and your pipette will not aspirate the selected volume.
Pipetting techniques
wiping of the pipette barrel (positive-displacement pipettes).
5
VOLUME IMMERSION DEPTH
Home
0.1 —1 µL 1 mm
Allow the plunger to
move up to the rest 1—100 µL 2–3 mm
position. 101—1000 µL 2–4 mm
Pre-Wet
To obtain greater uniformity and precision of dispensing, it
is better to provide identical contact surfaces for all aliquots.
This is done by pre-rinsing with the same liquid as the one
dispensed.
For pre-wetting, aspirate with the tip, and then dispense back
into the original reservoir or to waste.
Pre-wet again when adjusting the volume.
The reverse mode is only possible with air-displacement pipettes. It is sometimes used to
pipette slightly viscous liquids.
Pipetting techniques
1 2 3
Prepare Aspirate Dispense
Hold the pipette Immerse the pipette Place the pipette tip at
in a nearly vertical tip in the liquid. Allow an angle (10° to 45°)
position. Depress the the plunger to move against the inside wall
plunger smoothly up smoothly to the of the receiving vessel.
to the second stop rest position. Wait one Depress the plunger
position. second so that all the smoothly to the first
liquid has time to move stop position. Wait one
up into the tip. second.
Rest position
First stop
Second
stop
(purge)
Pipetting techniques
liquid that remains inside the tip while dispensing.
4 5
Re-Aspirate Purge
If the pipette tip is to be Wait one second and then
reused for the same sample, or purge. If the pipette tip is not
maintain the plunger in the to be re-used, depress the
intermediate position for plunger to purge position over
subsequent immersion for an appropriate waste container
the next pipetting cycle and and then eject the tip.
restart step 2 (Aspirate).
1 2 3
Prepare Aspirate Dispense
Press the plunger Immerse the capillary/piston Press the plunger
button to the first in the liquid.* Release the button to the first stop.
stop. The piston plunger, letting it move up The piston moves down
moves to the to the home position. The and expels the liquid
appropriate position. piston moves up and the out of the capillary.
ambient pressure forces the
desired volume of liquid
through the orifice into
the capillary.
Rest position
First stop
Ejection
* The immersion depth of your tip can have a significant effect on your results (for depth per model, see table page 20).
If the tip is immersed too deeply, droplets will form on the outside of the tip and they will be deposited along with your
sample. If the tip is not immersed deeply enough, vortexing will occur and your pipette will not aspirate the selected volume.
Pre-Wet
To pre-wet, aspirate with the tip and then dispense back
into the original reservoir or to waste.
Reuse of pipette tips may affect accuracy by up to 4% Use pipette tips only once
Pipetting Ergonomics
Take a Few Minutes to Get
Organized and Ensure You Have:
1. An appropriate posture
2. The right equipment
Gilson offers various pipettes with forces
adapted to user preferences. The forces
of PIPETMAN L are some of the lowest.
3. The appropriate technique
4. A good work organization and environment Download the Gilson Ergonomic
A good test for proper ergonomics is to see if Poster on www.gilson.com
you can rest your elbow comfortably on the work
surface. If not, your receptacle may be too low or
too high—find the right height.
Pipetting techniques
3. Take frequent short breaks. Change your sitting position. Lean back and relax
your shoulders and arms.
1. T
o facilitate uniform timing and motion, keep all necessary items within arm's
reach.
2. Place the most frequently used objects in front of you. The more rarely used
items can be placed a little further away from you.
3. The opening of the receptacle for used tips should be at the same height as
the end of your pipette.
Figure 5 Figure 6
SINGLE® pipette holder POWER CAROUSSEL® stand
Figure 7
Fit disposable pipette tips on single and multichannel pipettes
E
press the tip ejector push button. no-touch no-touch
disposal of tip disposal
To eject the tip from MICROMAN E, depress of capillary
the push button completely to the second and piston
stop. Discarded tips contain liquid residues,
particularly when a pipette is used in
reverse mode. Take suitable precautions
when discarding disposables. Does not apply
to 5 and 10 mL
pipettes
Autoclavable Tips
TOWERPACK
Tower of racks
RELOAD PACK
2 stacks of racks ECOPACK & EASYPACK
Bags of tips
TIPACK
BLISTER
REFILL
Figure 10
PIPETMAN tips packaging options
Figure 12
PIPETMAN® DIAMONDS tips guaranteed
traceability
preventing contamination
Types of Contamination and How to Prevent Them
●● Wipe lab bench before and after ●● Use the corrosion protection kit available
with an appropriate cleaner for your for PIPETMAN P1000 (PIPETMAN Neo,
application (cell culture, radioactive PIPETMAN G, and PIPETMAN L).
components, pathogenic samples, ●● For complete protection of the pipette,
etc.) choose MICROMAN E when working with
●● Work under a hood. non-aqueous liquids.
Pipette to Sample
Contaminated tips or a contaminated
pipette will contaminate your samples. Protected
air space
PREVENTION
●● Store pipettes vertically on a holder.
●● Eject tips into a designated container.
Piston
Filter
●● Follow laboratory protocol to clean Aerosol
seal
Non-protected
●● Use sterile tips when appropriate. air space
(aerosol)
●● Change the tip after each sample
to avoid cross contamination.
Figure 13
Solutions against contamination
preventing contamination
CONTAMINATION DECONTAMINATION
CLEANING GUIDELINES
CAUSES TECHNIQUES
Disassemble the lower part of your pipette. Fully immerse the con-
taminated parts* in an ultrasonic bath with a detergent or cleaning
solution recommended for laboratory instruments. It is strongly
Radioactive Detergent or cleaning recommended to rinse the pipette several times with water and dry
compounds solution it thoroughly.
Always make sure that radioactivity has decreased to
an acceptable level.
Aqueous solutions
and buffers Disassemble the lower part of your pipette. Rinse the contaminated
Water cleaning
parts thoroughly with distilled water and dry.
Acids/alkalis
If pipette brands other than Gilson are used, please make sure the material is compatible with
the appropriate cleaning solutions.
Autoclaving
This is a common method of sterilization. PIPETMAN DIAMOND Tips and parts of PIPETMAN
pipettes* may be sterilized in the laboratory under the following conditions:
moist heat/121°C/20 minutes/1 bar.
NOTE Autoclaving has a limited spectrum of action and will not destroy RNase, for example.
*P
IPETMAN parts can be autoclaved according to the User's Guide for the pipettes.
Refer to your model User's Guide for the specific parts and the recommended conditions.
100 ±3 ≤ 0.6 ±8 ≤3
P1000 500 ±4 ≤1 ±8 ≤3
1000 ±8 ≤ 1.5 ±8 ≤3
These specifications are defined for pipettes used in forward mode. The gravimetric method is
used with the temperature of the distilled water and all other conditions stabilized between
15°C and 30°C. The values given include all components of error due to both normal hand
warming and the changing of the tip.
To comply with the ISO 8655 standard, the specifications of the pipette must be within the
NOTICE
maximum permissible errors.
For all other damage, and for microvolume pipettes (2–10 μL), return the pipette for service.
mean volume
Pipetting technique
40
GILSON GUIDE TO PIPETTING
Calibration with the Gravimetric Method
The gravimetric method is recommended ●● Barometer with a standard uncertainty of
by pipette manufacturers and international max 0.5 KPa
standard organizations (ISO 8655). It is A calibrated barometer to check the
pipette service and maintenance
(293–296 K) ± 2.7
± 1.5
1 μL < V ≤ 10 μL 0.001 0.002 0.002
• Vessels
If the pipettes are used and Test equipment should correspond to the
therefore checked outside these following indications:
conditions, the weight of the
NOTE setting volume of water aspirated
will have to be corrected EXAMPLES
SAMPLE WEIGHING BALANCE OTHER
according to the conversion OF VOLUMES
RESERVOIR VESSEL RESOLUTION EQUIP.
INSTRUMENTS
table (μL/mg). See appendix.
P2 - P20
PM x20 Lid
0.1 to Ø 35 mm Ø 10.5 mm
Recommended Equipment
F2 - F10 0.001 mg Tweezers
20 μL H 50 mm H 13 mm
M10 to M25 Filters
M100
●● Calibrated thermometer with a
P100 - P200
standard uncertainty of max 0.2°C F25 - F200 > 20 to Ø 35 mm Ø 21 mm
0.01 mg Lid
A calibrated thermometer readable PM x300
M50 - M250
200 μL H 50 mm H 50 mm
is imperative that the performance of individual pipettes be compared regularly with the
manufacturer’s specifications.
Gravimetric analysis is a practical, widely used method for testing the performance (accuracy and
precision) of a pipette.
Accuracy check
Weekly to up Preventive Cleaning & Diagnosis:
End-users
to every three months maintenance - Visual inspection
- Function check
* Frequency should be adapted to the type of sample, the number of pipetting tasks, and the environmental conditions
of the laboratory.
Always check your pipette for mechanical faults (Refer to TWO-MINUTE INSPECTION
NOTE on page 37) before performing a gravimetric test.
Random Error*
Component of measurement error that
Air Cushion in replicate measurements varies in an
Also called "dead unpredictable manner.
volume", the air cushion Notes:
is the volume of air 1. Random error is equal to error minus systematic error,
Air cushion
located between the 2. Because only a finite number of measurements can
lower part of the be made, it is possible to determine only an estimate
pipette piston and the of random error.
surface level of the
sample. Systematic Error*
Component of measurement error that in
replicate measurements remains constant or
varies in a predictable manner.
Aliquot Notes:
Measured portion of a 1. Systematic error is equal to error minus random error,
homogeneous entity. 2. L
ike true value, systematic error and its causes cannot
be completely known.
A general term referring
to multiple samples of Comment: systematic error quantifies the error of
accuracy of a pipette.
any solution, mixture,
etc.
Good Laboratory Practice
Good Laboratory Practice (GLP) is concerned
with the organizational process and the
conditions under which laboratory studies are
planned, performed, monitored, recorded, and
reported.
Dispenser
An instrument
for delivering Measurand*
predetermined volumes Particular quantity intended to be measured
of liquid from a
Example: vapor pressure of a given sample of water at 20°C.
reservoir. The reservoir
may be integrated Note: the specification of a measurand may require
statements about quantities such as time, temperature,
into the instrument or and pressure.
connected externally.
APPENDICES
1 100 µL
0 instrument or measuring
0 system that provides Sample
guidance for its The appropriate
appropriate use. representative part of
Examples: a liquid to be analyzed.
1. 1 L as the value marked on a The term "test sample" is
single-mark volumetric flask used when necessary to
2. 100 μL as the setting avoid confusion with the
appearing on the volumeter of
Sample statistical term “random
a pipette
sample from population”.
Pipette/Pipetter
An instrument Sample Carryover
for transferring a
predetermined The portion of the
t dolore magna aliquam erat volutpat. Ut
odo consequat. Duis autem vel eum iriure
os et accumsan et iusto odio dignissim qui volume of liquid sample that is retained
et dolore magna aliquam erat volutpat. Ut
odo consequat. from one vessel to in the instrument after
sample delivery and that
t dolore magna aliquam erat volutpat. Ut
another. A pipetter is
odo consequat. Duis autem vel eum iriure
os et accumsan et iusto odio dignissim qui
reservoir. samples.
Note: Carryover from a
positivedisplacement pipette
is less than from an air-
displacement pipette.
True value
Precision* True value is a value that would be obtained
Closeness of agreement by a perfect measurement.
between indications
or measured quantity
values obtained by
replicate measurements Working range
on the same or similar Total volume and
objects under specified temperature range,
conditions. as well as ambient
Vmin Vmax conditions, for which
instrument performance
Repeatability* (of Results of is specified.
Measurements) Note: do not select volumes
Measurement precision under a set of outside recommended limits.
repeatability conditions of measurement.
Repeatability conditions include:
●● Same measurement procedure
●● Same operator/observer *Definitions abstracted from VIM (International Vocabulary
of Metrology).
●● Same measuring instrument
used under the same conditions
●● Same location
●● Repetition over a short period of time
Comment: for pipetting, variations due to the operator
(e.g., cycle time) are to be minimized.
evaporation loss ei that occurs during your (see table in Appendix C).
pipetting cycles.
Proceed as described in Appendix D to
determine ei
5. Evaluate accuracy
Systematic error (E):
2. C
hange the pipette tip and perform the 6. Evaluate precision (repeatability)
first weighing. Then, keep a regular cycle
Standard Deviation (SDw)
and perform the following ten
measurements:
Wr = 0.957 mg
W1 = 0.968 mg W6 = 0.966 mg
W2 = 0.960 mg W7 = 0.955 mg
W3 = 0.984 mg W8 = 0.972 mg
W4 = 0.942 mg W9 = 0.958 mg
W5 = 0.969 mg W10 = 0.967 mg
n: number of weighings
Wi weighing results
W = (0.968 + 0.960 + 0.984 + 0.942 + 0.969
+ 0.966 + 0.955 + 0.972 + 0.958 + 0.967) / 10
W = 0.964 mg
APPENDICES
PB = density of the balance weights. Use 8 g/cc for PB
Values of the conversion factor Z (μL/mg) as a function of temperature and pressure for
distilled water.
AIR PRESSURE
Temperature
HPA
°C
800 853 907 960 1013 1067
15 1.0018 1.0018 1.0019 1.0019 1.0020 1.0020
15.5 1.0018 1.0019 1.0019 1.0020 1.0020 1.0021
16 1.0019 1.0020 1.0020 1.0021 1.0021 1.0022
16.5 1.0020 1.0020 1.0021 1.0022 1.0022 1.0023
17 1.0021 1.0021 1.0022 1.0022 1.0023 1.0023
17.5 1.0022 1.0022 1.0023 1.0023 1.0024 1.0024
18 1.0022 1.0023 1.0024 1.0024 1.0025 1.0025
18.5 1.0023 1.0024 1.0025 1.0025 1.0026 1.0026
19 1.0024 1.0025 1.0025 1.0026 1.0027 1.0027
19.5 1.0025 1.0026 1.0026 1.0027 1.0028 1.0028
20 1.0026 1.0027 1.0027 1.0028 1.0029 1.0029
20.5 1.0027 1.0028 1.0028 1.0029 1.0030 1.0030
21 1.0028 1.0029 1.0030 1.0030 1.0031 1.0031
21.5 1.0030 1.0030 1.0031 1.0031 1.0032 1.0032
22 1.0031 1.0031 1.0032 1.0032 1.0033 1.0033
22.5 1.0032 1.0032 1.0033 1.0033 1.0034 1.0035
23 1.0033 1.0033 1.0034 1.0035 1.0035 1.0036
23.5 1.0034 1.0035 1.0035 1.0036 1.0036 1.0037
24 1.0035 1.0036 1.0036 1.0037 1.0038 1.0038
24.5 1.0037 1.0037 1.0038 1.0038 1.0039 1.0039
25 1.0038 1.0038 1.0039 1.0039 1.0040 1.0041
25.5 1.0039 1.0040 1.0040 1.0041 1.0041 1.0042
26 1.0040 1.0041 1.0042 1.0042 1.0043 1.0043
26.5 1.0042 1.0042 1.0043 1.0043 1.0044 1.0045
27 1.0043 1.0044 1.0044 1.0045 1.0045 1.0046
27.5 1.0044 1.0045 1.0046 1.0046 1.0047 1.0047
28 1.0046 1.0046 1.0047 1.0048 1.0048 1.0049
28.5 1.0047 1.0048 1.0048 1.0049 1.0050 1.0050
29 1.0049 1.0049 1.0050 1.0050 1.0051 1.0052
29.5 1.0050 1.0051 1.0051 1.0052 1.0052 1.0053
30 1.0052 1.0052 1.0053 1.0053 1.0054 1.0055
2 Cover the weighing vessel with its lid and place it on the balance using a pair of tweezers.
3 Aspirate a sample.
4 Tare the balance and take the weighing vessel out of the balance.
Replace the lid on the weighing vessel and then use tweezers to place the the vessel on the
7 balance.
Under normal conditions, this value is usually between 0.01 mg and 0.03 mg.
APPENDICES
Acetone ++ + - ++ ++ ++ ++ - ++ ++ + + ++
20 % ++ ++ + ++ ++ ++ N/A ++ ++ ++ ++ ++ ++
Acetic acid 50 % ++ ++ + ++ ++ - N/A + ++ ++ ++ ++ ++
100 % ++ ++ - ++ + - N/A - ++ ++ + + ++
10 % - ++ ++ ++ ++ - ++ ++ ++ ++ ++ ++ ++
Hydrochloric acid 20 % - + + ++ ++ - + ++ ++ ++ ++ + ++
37 % - - - ++ ++ - - + ++ ++ ++ - ++
20 % + + - ++ + - + ++ ++ ++ ++ + ++
Hydrofluoric acid
40 % - + - ++ - - - + ++ ++ ++ + ++
Formic acid 100 % ++ N/A - ++ ++ - + - ++ ++ N/A + ++
10 % ++ ++ + ++ ++ - ++ ++ ++ ++ ++ + ++
Nitric acid 30 % ++ + - + ++ - + ++ ++ ++ ++ - +
65 % ++ - - - + - - + + + ++ - -
20 % ++ N/A + ++ N/A - ++ ++ ++ ++ ++ + ++
Phosphoric acid
85 % ++ N/A - ++ N/A - ++ ++ ++ ++ ++ - ++
50 % ++ - + N/A N/A ++ ++ + ++ ++ N/A - ++
Propionic acid
100 % ++ - - N/A N/A + ++ - ++ ++ N/A - ++
20 % ++ ++ + + ++ + ++ ++ ++ ++ ++ + ++
Sulfuric acid 50 % ++ ++ - + ++ - + ++ ++ ++ ++ - ++
95 % ++ + - - - - - + + + ++ - +
20 % ++ N/A - N/A N/A + N/A ++ ++ ++ N/A ++ ++
Trifluoroacetic acid 80 % ++ N/A - N/A N/A - N/A + ++ ++ N/A + ++
100 % ++ N/A - N/A N/A - N/A - ++ ++ N/A - ++
Benzyl alcohol ++ ++ - N/A N/A + N/A - ++ ++ ++ - ++
Aniline ++ - + ++ N/A ++ N/A - + ++ N/A + ++
Butanol / Butyl alcohol ++ ++ ++ ++ N/A ++ ++ ++ ++ ++ N/A ++ ++
Chloroform ++ - - - N/A + - - + ++ + + -
Cyclohexane ++ ++ ++ - N/A ++ N/A ++ ++ - + ++ +
Diacetone alcohol ++ ++ + N/A N/A N/A N/A N/A N/A + N/A N/A ++
Methylene chloride ++ + - - N/A - - - + ++ ++ ++ +
Diethylene glycol ++ N/A ++ ++ ++ N/A N/A N/A ++ ++ ++ ++ ++
Dimethylformamide (DMF) ++ ++ - + ++ ++ ++ - ++ - ++ ++ ++
Dimethylsulfoxide (DMSO) ++ N/A - N/A N/A + N/A - ++ N/A N/A N/A N/A
Dioxane (1,4) ++ ++ - + N/A ++ ++ - ++ + N/A ++ +
Ethanol ++ ++ ++ ++ ++ ++ ++ ++ ++ ++ ++ ++ ++
Ether ++ ++ ++ + N/A ++ ++ ++ + ++ + ++ ++
Formaldehyde ++ ++ ++ ++ N/A ++ N/A ++ ++ ++ ++ ++ ++
Hexane ++ N/A ++ - + ++ ++ ++ + ++ + ++ ++
Hydrogen peroxide 50 % ++ N/A + ++ N/A ++ ++ ++ ++ ++ ++ ++ ++
Ammonium hydroxide 20 % ++ ++ ++ ++ N/A N/A + - ++ N/A ++ ++ ++
10 % ++ + ++ ++ ++ ++ + - ++ ++ ++ ++ ++
Sodium hydroxide
40 % ++ - + ++ ++ ++ + - ++ ++ ++ ++ ++
Sodium hypochlorite 15 % Cl + N/A + ++ ++ ++ ++ ++ ++ ++ ++ - +
Methanol ++ ++ ++ ++ + ++ ++ + ++ ++ ++ ++ ++
Methyl ethyl ketone ++ ++ - + ++ ++ ++ - ++ - + + ++
Pentane ++ N/A ++ - N/A N/A N/A ++ ++ ++ + ++ N/A
Tetrahydrofuran (TH-) ++ ++ + + + ++ + - - + + N/A +
Urea ++ ++ N/A N/A - ++ ++ N/A ++ ++ N/A ++ ++
Accurate but not precise Precise but not accurate Accurate and precise
notes