Christie Medical Physics & Engineering

The Christie NHS Foundation Trust


Withington, Manchester
M20 4BX, UK

Diagnostic X-Ray QC Equipment Performance Evaluation

Report No: 0009/LCLAR/22 Report Date: 8 March, 2022

Customer: Royal Lancaster Infirmary

Equipment: Unfors Solo Dose


Base unit serial no: 226376
Probe serial no: 189895
[our ref. Q2]
Date of test: 06/01/2022
Previous relevant reports: 0770/LCLAR/20
Sent to: Sarah Moreland, Acting Lead Radiographer, Stewart Preston,
Radiographer, Antony Outterside, Radiographer
Sharon Timperley, Head of Department

Debbie Crawford, Clinical Lead, Radiology, Imaging Services Manager


Karen Brakewell, Deputy Head of Department
Conor Clancy, Jacob Brazier, CMPE
Contact: Chris Wood on (0161) 446 3551
or e-mail christopher.wood27@nhs.net

Diagnostic Radiology and Radiation Protection Group

FS 37543

List of Measurements Performed


Outcome

Measurement Tolerance Pass Fail


Dose accuracy ±5% Pass
Timer accuracy ±1% or ±0.33 ms Pass

Conclusions
This unit is acceptable for:
• Checking radiographic outputs over the range 40 to 125 kV and 0.5 to 200 mAs
• Checking exposure times over the range 50 to 500 ms

Our next scheduled checks on this equipment will be due in January 2023.
Note
This check does not constitute a full calibration of the equipment but is intended to give a guide to its accuracy compared to
instruments whose calibration is traceable back to national standards and to act as a baseline for future checks on the
consistency of the equipment.